decontamination and sterilization, hugh oconnor.pdf
TRANSCRIPT
-
Hugh O'Connor AP 1
Decontamination Puzzle in CSSD/HSSU/CDU/LDU
Validation of Sterilsation processes !
-
Hugh O'Connor AP 2
The Documentation and Process Testing in the Hospital Setting
Plant centralised
Managers office
Tracability system
Humidity controller
Ergonomic Enginering Fire Exit
Enginers Workshop
Data logger
Extra Long Chamber BBS Sterilizer
-
Hugh O'Connor AP 3
CSSD IAP(Clean)Room
-
Hugh O'Connor AP 4
Decontamination
Decontamination is a process which removes or destroys
contamination and thereby
prevents micro-organisms or other contaminants reaching a susceptible
site in sufficient quantities to initiate infection or any other harmful
response.
A multitude of different processes are available for decontamination.
Choice of the
appropriate procedure for a given situation demands an understanding
of the processes available, the level of decontamination required and
the manufacturers instructions for decontamination.
Having determined the level of decontamination required, this should
be achieved by a method that combines reliability, cost effectiveness
and the minimum of hazard to the user.
-
Hugh O'Connor AP 5
Decontamination Testing
Sterilization/Disinfection is a process whose efficacy cannot be verified retrospectively by inspection or testing of the product.
For this reason Sterilization/Disinfection have to be validated before use.
The performance of the process must be routinely monitored and the equipment maintained.
-
Hugh O'Connor AP 6
The Philosophy of Testing
Three main principles
All sterilizers/washers are monitored to a planned programme of tests
All sterilizers/washers are subject to a planned program of PM
Expertise on all aspects of the testing should be available at three levels
1. Competent Management
2. Competent Users
3. Competent Test People
-
Hugh O'Connor AP 7
Situtation Prior to standards
Myriad of small benchtops
No automated Washers
No standardised testing or auditing
No documentation process
No auditing/Validation process
No dedicated facilities for decontam
No ongoing education process
-
Hugh O'Connor AP 8
Introduction of HSE/EN standards
Look at how best to reduce risk of infections from the reprocessing of medical devices especially Surgical instruments/endoscopes.
Look at what is best practice for procedures on cleaning, disinfection, sterilisation of such devices.
A healthcare institution should do everthing to prevent spreading of infections within the institution reason for manged CSSD.
-
Hugh O'Connor AP 9
Pilot review example
Introduction.
The pilot review was part of the process for implementing the HSE Code of Practice for Decontamination of Reusable Invasive Medical Devices Part 5b. The code is based on EU standards, directives and other international standards for decontamination of RIMD and health and safety legislation and covers recommended practices for decontamination in a local decontamination unit (LDU).
-
Hugh O'Connor AP 10
The Structure of the Pilot Review
Time taken to review each location.
Problems with questionnaire.
Information on water quality in each location.
Status and type of autoclave present.
Design and dimensions of relevant rooms. Record keeping.
Tracing systems.
Presence of clean and dirty areas.
Evidence of no validation, servicing of equipment
-
Hugh O'Connor AP 11
Some anomalies found in study
-
Hugh O'Connor AP 12
Findings of the Pilot Review
-
Hugh O'Connor AP 13
Findings of the Pilot Review
-
Hugh O'Connor AP 14
Positive Findings Pilot Review
-
Hugh O'Connor AP 15
Positive Findings Pilot Review
-
Hugh O'Connor AP 16
Positive Findings Pilot Review
-
Hugh O'Connor AP 17
Recommendations Pilot Review
Sign off on final version of Part 5b of code of practice.
Upgrade or centralisation.
Full national review of dental surgeries.
Estimate of cost of upgrading HSE dental surgeries up to the HSE Code of Practice.
Develop timetable for upgrading.
Funding issue.
-
Hugh O'Connor AP 18
Cost of implementing HSE COP
Labour Costs for in house testing
Labour Costs for in house user testing
Labour costs for water sampling
Labour costs for management of testing
Labour costs for auditing of testing
Labour costs for auditing of compliance
Labour cost to read and understand code of practice and standards
-
Hugh O'Connor AP 19
Brief Look at Other Audits
Healthcare Commission Report 2008 UK.
90 % of acute hospitals failed to comply FULLY with Commision Hygene Code.
One in five failed to satify requirements of decontamination.
From April 2009 CQC will enforce requirement of CHQ for infection control.
Care Quality Commission can impose conditions and fines.
-
Hugh O'Connor AP 20
COP Recommends Centralisation
Centralise for patient safety
Provide a quality process that is consistent, controlled and validated.
Dedicated trained staff.
Improve working environment.
Update equipment.
Cost efficiency.
-
Hugh O'Connor AP 21
Some Advantages for User of CLDU !
Reduce high cost of decontamination operations.
Checking of water quality and storage tanks.
Costly disposal of empty distilled water bottles.
Savings on multiple site validation.
Savings on multiple service costs.
Centralisation of tracability.
Release staff from low productivity tasks.
Free staff to attend prohylactic treatments.
-
Hugh O'Connor AP 22
The Plan for CLDU may start like this
-
Hugh O'Connor AP 23
The Future CDU Department
Returns and
wash
Clean
process
Storage
Treatment
room
Treatment
room
Office
Treatment
room
clean corridor
dirty corridor
Patient
ward/waiting
area
Staff room
Sterilizers
-
Hugh O'Connor AP 24
Best Practice - CSSD
-
Hugh O'Connor AP 25
Area of Focus
Wash Room
Steriliser Loading Area
Steriliser Plant Room
Machinery and Equipment
-
Hugh O'Connor AP 26
How A Wash Room Should Look
-
Hugh O'Connor AP 27
Wash Room Aseptics
314 stainless inspection and dis-assembly tables
Negative AHU ventilation
White-rocked or similar environment
No access to IAP
Pass through hatch
-
Hugh O'Connor AP 28
Cart Wash (to solve MRSA ?)
-
Hugh O'Connor AP 29
Decontamination Services in LDU/CSSD
Dirty in/Clean out environment
Separation of Processes Environmental testing and control ACs WDs validated and tested to EN,COP 5 Reverse Osmosis water quality Designated decontamination staff Competency assessed training Computerised barcode traceability Documented ISO quality management system
Regular audit
-
Hugh O'Connor AP 30
The Team from COPS Infection
Control &
Microbiologist
Estates
Management
Planning
Servicing/repair
Routine testing
Manufacturer of
the ACs,WDs
Sterile Services
Authorized
Person/similar
Contractors
Testing/building
Test
Person
(Sterilizers)
Health Estates
The User
-
Hugh O'Connor AP 31
The Ideal Sterilization/Disnfection Process
Effective against all infective agents
Sterilant/Disinfectant and its residues non-toxic,bio-degradable
Not inactivating by soiling,protein,etc
Highly penetrative
Rapid in action
Cheap to install and operate
-
Hugh O'Connor AP 32
The Ideal Sterilization/Disinfectant Process
Requires no training, technical knowledge
No environmental or bio hazard
Can be continuously and independently monitored
Causes no damage to any of the products
Acceptable to National/European, International regulations
-
Hugh O'Connor AP 33
Critical Steps in the Reprocessing Process
Before any automated or manual disinfection or sterilisation procedures the critical steps to be followed are.
1. Clean the instruments manually, meticulously removing all organic material in or on the surgical instruments/endoscope.
2. Rinse all parts and lumens throughly with copious amounts of water and dry them before placing the scope in the reprocessing unit.
-
Hugh O'Connor AP 34
Critical Steps in the Reprocessing Process
4. The drying process here avoids dilution of the disinfectant concentration or interaction between the water and the sterilant/disinfectant.
5. Follow the manufacturers direction for reprocessing that specific brand and model of instruments. (Ensure endoscope can be reprocessed in automated manner not all can.)
6. If in doubt about capability of washer to reprocess particular models or brands of surgical instruments check with manufacturer to establish if these have been tested on the particular machine.
-
Hugh O'Connor AP 35
Critical Steps in the Reprocessing Process
Regardless of whether you manually reprocess or use automated processes, incorporate a final drying step in your reprocessing protocol, flush all channels with alcohol/Hot Air followed by purging of channels with air to remove alcohol.
-
Hugh O'Connor AP 36
Critical Steps in the Reprocessing Process
8. Visual inspection of the reprocessing procedure and reprocessed surgical instruments can identify conditions that may affect cleaning and disinfection processes.
9. Keep records of the use of each instrument /endoscope, showing the patient upon which it was last used, the type of procedure and the reprocessing method and unit used.
-
Hugh O'Connor AP 37
Critical Steps in the Reprocessing Process
10. Regular review of training given to reprocessing staff, including additional staff training and supervised practice each time a new instrument model is introduced into the facility.(PQ)
11. Independent audits by an Authorised Person to verify compliance with procedures and processes.
-
Hugh O'Connor AP 38
First StepValidate the Washer/Disinfector
Mop available
Detergent available
Temp of water
Method to change water
Method to change mop
Check efficacy
-
Hugh O'Connor AP 39
Post Design, Wash Room example
-
Hugh O'Connor AP 40
How the Washer Works
s
t
a
n
d
b
standby
Initial wash
-
Hugh O'Connor AP 41
How the Washer works
Print to Cleanroom
Instruments Dry
If not clean not washed
Test with Ninhyrdin
Check for water spotting
Check all critical Parameters
-
Hugh O'Connor AP 42
Ao Lethality with Thermal Washers
Standard EN DIN ISO15 883-1, Annex a
Moist predicatable lethal effect on micros
WD must guarantee moist effect on micro
All inner outer surfaces on RMD are wet
Furniture and wall of WD reach temp
Ao is the equivilent time in seconds@80c
D value is the time to reduce micro count to 10%.the Z value temp to reduce D 90%
-
Hugh O'Connor AP 43
Bio-burden reduction required for cleaning
Water is the solvent and carrier in which soil is going to disolved,suspended and transported, away from the items to be washed. It provides the environment in which all cleaning actions take place
Mechanical action such as wiping,brushing,spraying water under pressure or ultrasonic waves
Chemical action,detergent+water used to soak and suspend germs,dirt,also kill and protect
Heat improves the diluting power of the water and detergent in order to prevent the coagulation of tissues and blood
-
Hugh O'Connor AP 44
Bio-burden reduction required for cleaning
MIs are used in wounds of patients,should have no micros or foreign particles
Left on instruments can cause Sepsis or septicimia
If such a dirt particle enters blood stream recovery and morbidity can be effected
Leaving food(fungi)blood(Hep B) pus (bacteria ,gangerine) allows micro to divide and multiply
-
Hugh O'Connor AP 45
Bio-burden reduction required for cleaning
In Summary
Remove all visable dust and dirt
Removal of breeding ground for surviving micros
Protection against corrosion
Ensure safe free movement of equipment and material
Thats why we do efficacy tests
-
Hugh O'Connor AP 46
Bio-burden reduction required for cleaning
Washing bioburden lesser challenge
Process is checked with 5,5,5 process
1. 5 kinds of micros
2. In 5 minutes
3. With a factor of 10 to the 5(100,000)
In otherword after process 5 mins the no of survivors of 5 kinds of micros is 1/100,000 of the original no 1 million then 10 (bacilus subtillus)
-
Hugh O'Connor AP 47
Bio-burden reduction required for cleaning
Fibrin layer leftover
Simulates blood film
Not clean not sterile
All instruments suspect
Quarintine loads
Call HELP Engineer required
-
Hugh O'Connor AP 48
Efficacy Study - Pre Wash
Before machine washing.
Orthophedic intruments pr soiled.
Tosi and Brownes test soil.
-
Hugh O'Connor AP 49
Cleaning Eefficacy tests on WDs
Tosi 95% water soluble proteins + 5% water insoluble fibrin.
Marginal Fail blood+fibrin remain.
Absolute fail with some washers and detergent.
Pass, acceptable with at 67 mins of main wash.
-
Hugh O'Connor AP 50
Efficacy Study - Post Wash
Visual cleaned
Ninhydrin test
Washing must remove all visible evidence of soiling
All washer must be subjected to performance tests
-
Hugh O'Connor AP 51
Closer Visual Inspection after Manual Process .
-
Hugh O'Connor AP 52
Closer Visual Inspection after Automatic Washer Disinfection process !
-
Hugh O'Connor AP 53
Conclusion from Study
Disinfection kills or removes most parhogenic microorganisms
Washer with validated performance will achieve expected efficacy
If washing fails then Not clean Not sterile
-
Hugh O'Connor AP 54
Washer at 5th Gear
-
Hugh O'Connor AP 55
Ao Letahlity for WDS
Disinfection at bacteria,fungi,mycobateria, heat sensitive viruses.
Ao value 600 is defined or 600 s=10@80
The Ao 600 can be achieved 90 c@1 min
Heat resistant viruses Hep B need Ao 3000
Need hold time of 90 c for 10 min
Does not eliminate impeccable cleaning of MD which is precondition for Disinfection
-
Hugh O'Connor AP 56
Reduction in number of Live Spores in WD
Temperature T in C
Ao in Seconds Ao in mins
70 6000 100.00
75 1897 25.00
80 600 10.00
85 190 3.16
90 60 1.00
90 Ao (3000) 25 minutes
-
Hugh O'Connor AP 57
A o Trace of 3000 @ 93 c
-
Hugh O'Connor AP 58
Washer Disinfectors Issues
Oprator Safety (door interlocks)
Consistent Cleaning Effect (Tosi clean)
High Temp Wash Stage >= 90c@1 min
Consistent Detergent solution
Predetermined wash stages
Cycle will include drying stage
Independent monitoring of critical events
-
Hugh O'Connor AP 59
Water Quality
The water supply may be hard ,it may contain environmental mycobacterium
Mycobacterium chelonae,pseudomonas,
and other Gram negative bacilli may contaminate the instrument.
Sterile water ids recommended for the final rinsing off all types of endoscopes.
-
Hugh O'Connor AP 60
Decisions on Water?
Two options to obtain a supply of water of suitable microbial quality.
1. The use of pre-sterilized (pyrogen free) bottled water for stand alone machines
2. The use of pre-treated pyrogen free water connected to mains machines
The type of pre-treatment is determined by an audit of the supply water.
-
Hugh O'Connor AP 61
Reversed Osmosis Filtered System for clean water and clean steam!
Carbon Filter
Ionise Water
5,2,1 micron reduce
.1 micron final water
High rejection due to High Conductivity
Require Potable/main < 150 msc
Heat remove NCGS and residual air
-
Hugh O'Connor AP 62
Water Supply Factors
The effect of the installation of any water treatment device on these parameters
The location of the water treatment device
(Note: failure to perform routine maintenance can undermine the effectiveness of the washer/water supply)
-
Hugh O'Connor AP 63
List of Tests for Washer Disinfector EN15883
Test
Daily/weekly Quarterly/Bi-Annual Annual
Efficacy
Temperature
Safety Tests
Verification on instruments
Clean strainers
Chemical Purity
Chemical Dosing
Final Rinse BI Test
-
Hugh O'Connor AP 64
How a Sterilizer Works !
-
Hugh O'Connor AP 65
Some of the Maths of Sterilization
-
Hugh O'Connor AP 66
Reduction in number of Live Spores in D values at 121 c
Spores of D 121 c (minutes) Z- value
Clostridium Botulinum
0.204 10
Clostridium sporogenes
0.81.4 13
Bacillus steramoph
2.0 6
Bacillus subtilis o.4 10
IM0 2.5 18.6
-
Gravity Displacement Autoclave
Arrange load inside the chamber and insert the temperature probe Close door Open steam valve Steam enters the chamber Air and condensate pushed through the drain Temperature in chamber increases When load temperature reaches required sterilization temperature the timing starts When hold time is complete the steam valve is closed Autoclave is allowed to cool until temperature
-
Hugh O'Connor AP 68
Steriliser Loading Area
-
Hugh O'Connor AP 69
How a Sterilizer Works !
A steam sterilizer uses steam as a carrier of thermal energy or heat Latent Heat,enthalpy
The steam softens the outer layer of the micros which permits the thermal energy or heat to enter the organism and come in contact with the proteins in the organism
Heating the proteins causes changes in the in the nature of the proteins which is called Denaturing,Coagulation or poaching of proteins.
-
Hugh O'Connor AP 70
How a Sterilizer Works !
Just as an egg white changes from liquid to solid when heated.
The proteins in side the cells also change when heated (the cytoplasm membrane rendering the cell no longer viable)
Firstly the proteins do their job feeding , programming the cells essential functions.
Now the thermal energy disrupts these essential functions and ZAPS the reproduction nature of the programming.
-
Hugh O'Connor AP 71
How a Sterilizer Works !
Sterilization with steam depends on properly elevated Saturated Steam at a sufficient time and temperature
90%95% dryness factor
Sterilization of steam depends on properly elevated Temp,sufficent moisture and sufficient time.
Just as it takes three mins to boil an egg then in cooking and sterilization the higher the compatible temperature the faster the task is completed
-
Hugh O'Connor AP 72
How a Sterilizer Works !
The critical factor in the destruction of microorganisms is the relationship between
Temperature 134 c too 137 c
Pressure 2.12 bar gauge
Time of exposure 3 minutes too 3.5
-
Hugh O'Connor AP 73
How a Sterilizer Works !
Air Removal by Vacuum pump that draws air out while injecting steam into the chamber
As more pressure enters the chamber the steam and air from a mixture
-
Hugh O'Connor AP 74
Condensation formation in the chamber of a porous load steriliser during sterilizing.
Surgical instrument
Pack 7kg +-5%
Steam supply
2.2 barg
135.8c
chamber
Condensate
At 135.8c
Drain (condensate to condenser and steam trap
And vacuum pump at 20mb is flashed of
-
Hugh O'Connor AP 75
Mixtures of air and steam
According to Daltons law of partial pressures, the pressure exerted by a mixture of gases is the sum of the individual pressures which each exert if occupied a given space alone.
For ex if a space containing steam and air at 4 bar this could be 3bg of steam and 1bg of air.
The resulting tmp would correspond to 3bg of saturated steam instead of required tmp for 4 bar
This is why air venting,steam traps,air detectors are essential.
-
Hugh O'Connor AP 76
Reduction in number of Live Spores in D values at 121 c
Spores of D 121 c (minutes) Z- value
Clostridium Botulinum
0.204 10
Clostridium sporogenes
0.81.4 13
Bacillus steramoph
2.0 6
Bacillus subtilis o.4 10
IM0 2.5 18.6
-
Hugh O'Connor AP 77
When a pre-determine pressure is reached (getinge 1.11 bg),the drain opens and the pressurized steam and air rush from the chamber creating a Vacuum.
How a Sterilizer Works !
H
w
-
Hugh O'Connor AP 78
How a Sterilizer Works !
Raising ,heating time
Once the air is removed from the load and chamber,steam is injected until the thermal sensor probe (drain) reaches the pre-selected temperature (134 c)
Sensor
drain
-
Hugh O'Connor AP 79
Handpiece Failures Processing with N cycle
-
Hugh O'Connor AP 80
How a Sterilizer Works !
Exhausting time(post-vacuum) Holding time complete the vauum will exhaust to > 70 milibar.
Steam injector closed
Vacuum stay open untill cycle complete
Hepa bacterial filter opens
Sterile filtered air is injected untill 1.014 bg atmospheric is reached
> 70mb
-
Hugh O'Connor AP 81
How a Sterilizer Works !
Exposure Time (Plateau time)
When thermal sensor registers (134.2c) the steam injection will close and the actual sterilization time will be activated.
Sensor drops bellow 134.2 c, the steam injector opens and tops up.
-
Hugh O'Connor AP 82
Steriliser Guidelines
EN 554/EN 17665 Planned and Documented maintenance
HTM 2010/HTMO5 PT2 & 4 guidance on maintenance
MDA DB200 Vacuum Sterilizers
MDA DB9605 Benchtop Sterilizers
EN 13060 European Standard < 60 litres
DB9804 Periodic testing
HTM2031 Clean steam for sterilization
COP HSE Part 5a,5b
Manufacturers Guidelines especially Type S
-
Hugh O'Connor AP 83
Test Description for HSE COP Pt 5b Pre
Evaluations and Audits
Non
Vacuum
Autoclave
Vacuum
Autoclave
Washer
Disinfector
Ultra Sonics
Yearly Safety Test
Yes
N/a
N/a
Air Leakage Test (automatic)
n/a*
N/a
N/a
N/a
Air Leakage Test (manual- temp. and pressure sensors connected)
n/a
N/a
N/a
N/a
Automatic Control Test
Yes
N/a
N/a N/a
N/a
Door and Door Interlocks
Yes
N/a
Yes
N/a N/a Verification of Calibration of Autoclave
Instruments
Yes
N/a
N/a N/a
Yes
Cleaning Efficacy Tests
N/a
N/a
Yes
N/a N/a Detergent Reproducibility Test
N/a
N/a
N/a
Yes
Function Test
N/a
N/a N/a
N/a
N/a
Air Detection System Performance Test for a Full Load
n/a
N/a
N/a
Thermometric Test for a Full Ld
N/a
N/a N/a
N/a
N/a N/a Thermometric Test for Disinfection
N/a N/a
Yes
Ultrasonic Activity Test
N/a
Yes
N/a
Porous Load Dryness Test
n/a N/a
Yes
N/a N/a
N/a
Full Performance Qualification Test
N/a
N/a
N/a
n/a
Tests for Performance Requalification (as required)
N/a
N/a N/a
n/a
n/a
Air Leakage Test (automatic) (sensors removed)
n/a
N/a
N/a
Air Detection System Function Test (automatic)
n/a
N/a
N/a
N/a
Steam Penetration Test
n/a
N/a
Load Dryness on Reference Load
N/a
N/a
N/a N/a
N/a N/a Steam Quality Tests (Commission)
N/a
Water Quality (Ph,Hardnes,Cd)
Yes
N/a
N/a N/a Photographic Evidence
N/a
N/a
N/a
N/a
-
Hugh O'Connor AP 84
Test Description for HSE COP Pt 5b Post
evaluations and Audits (Annual)
Non
Vacuum
Autoclave
Vacuum
Autoclave
Washer
Disinfector
Ultra Sonics
Yearly Safety Test
Yes
Yes
Yes
Yes
Air Leakage Test (automatic)
n/a*
Yes
N/a
N/a
Air Leakage Test (manual- temp. and pressure sensors connected)
n/a
Yes
N/a
N/a
Automatic Control Test
Yes
Yes
Yes
Yes
Door and Door Interlocks
Yes
Yes
Yes
Yes
Verification of Calibration of Autoclave Instruments
Yes
Yes
Yes
Yes
Cleaning Efficacy Tests
N/a
N/a
Yes
Yes
Detergent Reproducibility Test
N/a
N/a
Yes
Yes
Function Test
N/a
Yes
Yes
Yes
Air Detection System Performance Test for a Full Load
n/a
Yes
N/a
N/a
Thermometric Test for a Full Ld
N/a
Yes
N/a
N/a
Thermometric Test for Disinfection
Yes
Yes
Ultrasonic Activity Test
Yes
Yes
Yes
Porous Load /Dryness Test
n/a
Yes
Yes
Yes
Full Performance Qualification Test
Yes
Yes
Yes
n/a
Tests for Performance Requalification (as required)
Yes
Yes
n/a
n/a
Air Leakage Test (automatic) (sensors removed)
n/a
Yes
N/a
N/a
Air Detection System Function Test (automatic)
n/a
Yes
N/a
N/a
Steam Penetration Test
n/a
Yes
Load Dryness on Reference Load
N/a
Yes
Yes
Yes
Steam Quality Tests (Commission)
Yes
Water Quality (Ph,Hardnes,Cd)
Yes
Yes
Yes
Photographic Evidence
N/a
N/a
Yes
Yes
-
Hugh O'Connor AP 85
Validation a Planned and Documented Process
-
Hugh O'Connor AP 86
Sterilizer in Action
-
Hugh O'Connor AP 87
If not clean cannot be sterilized
All viable viruses and contaminants removed
All detergents removed
Visualy clean
Test for protein presenc
Test for efficay of wash
-
Hugh O'Connor AP 88
Specifications for WD
Fully automatic pre-determined Wash Cycle.
Automatic/verfiable Detergent dosing.
Three rinses + final rinse (90 c at 1 minute).
Single manually/double glass door.
Electricaly /Steam heated/Stainless 314-sl.
Event type printout with independent sensor (temp,cycle time,dose) for traceabilty.
Single Shot clean Final rinse system (R/O,DI)
Compliant to EN-158833-1-2, COP 5.
-
Hugh O'Connor AP 89
Specification AC
Fully automatic pre-determined Ster Cycle.
Steam generated internally/externally.
Full active Vacuum system.
Single manually/automatic opertated door.
Electricaly /Steam heated.
Event type printout with independent sensor (pressure,temp) for traceabilty.
Air detector a posibility for ADFT.
Single Shot clean steam system (R/O).
Pressur vesel marked to EN61010-2-041.
-
Hugh O'Connor AP 90
List of Tests for Sterilizer/Autoclave EN285/COP HSE 5b
Test daily Weekly Annual
Bowie Dick Test
Weekly Tests
Verification of instruments
ACT
Vacuum Leak test Not applicable to type N cycle (non-vacuum)
Air Detector Tests If applicable
Small Load Test
Full Load Test (solid)
Pressure Tests 24 months
BI+ Steam Quality commissioning
Ap sign off
-
Hugh O'Connor AP 91
Management Responsibility
The fundamental cause of this disaster is to be found in human failings, ranging from simple carelessness to poor management of men and plant. (Sir Cecil Clothier on the report into the deaths of five patients due to a faulty sterilizer)
-
Hugh O'Connor AP 92
Your Responsibility
Purchasing,maintenance and inspection.
Regulatory and legal aspects
Sources of further advice (AP) and information
Technical aspects and safety considerations
The hazards of circumventing safety aspects
-
Hugh O'Connor AP 93
Key Information You Need From the Supplier
Compliance of equipment with regulatory and legal requirements.
Information on the technical and safety aspects of steriliser.
Knowledge on the hazards of circumventing safety features
-
Hugh O'Connor AP 94
Key Issues for the User
The processing of wrapped goods.
The processing of instruments with lumens or cavaties.
The processing of porous materials.
The hazards of infection resulting from inadequate processing.
MDA (93/42/EEC) all medcial devices should be safe and fit for their intended purpose.
-
Hugh O'Connor AP 95
Traceability
-
Hugh O'Connor AP 96
Independent Monitoring & Verification System
For EN Sterilization
Real time data logging & archiving of microprocessor and independent data
Real time data comparison with automatic machine alarm and shutdown
On-screen machine operation mimic
Spreadsheet and graphical data representation
Individual cycle / file storage with unique identifier
Bar coding interface facility options
-
Hugh O'Connor AP 97
Requirements for Checks and Balances
-
Hugh O'Connor AP 98
To State of Sterility & Grace
-
Hugh O'Connor AP 99
If we dont the risk is
-
Hugh O'Connor AP 100
Your Responsibilty lets do a bungi jump
-
Hugh O'Connor AP 101
Whats the legal Requirement
Just as in the case of patient charts, present German
regulations decree that results documentation be kept for 30
years. (German Civil Code - BGB; limitation period after
which compensation claims become statute-barred). Claims
from the doctor-patient contract or from the hospital admission
contract become statute-barred after 30 years pursuant to
Article 195 of the German Civil Code. Because German
legislation recognises the principle of shifting the burden of
proof, which states that hospitals must prove that no damage
was inflicted on the patient, this documentary obligation also
applies for the CSSD.
-
Hugh O'Connor AP 102
Overall Advice
HSE COP Pt5b, NHS website gives clear guidance on Steriliser/Sterilisation Guidance.
Need bench tops with one shot system.
Standards are varied but should be driven by user professionals and patients.
Trial machines before purchase.
Thermometric tests supersedes biological tests.
PQ and validate new machine to reflect medical device needs
-
Hugh O'Connor AP 103
Summary The current requirement of BS EN544,
HTM01/02, DB9804, and EN 13060/EN 15883 are that successful decontamination management is achieved by a combination of personnel who have responsibilities for audit, maintenance, and validation of decontamination equipment through its working life.
You and Me
-
Hugh O'Connor AP 104
Validation
Kill, Kill,Kill them all, Until nothing is alive
Ghengis Khan
-
Hugh O'Connor AP 105
List of Tests for Washer Disinfector EN15883-1/COP5
Test daily Weekly Annual
Efficacy test
Spray arms
Verification of instruments
Sump strainers
MIS lumen test
Small Load Test
Full Load Test
Dose pump check
Pump tests
Final rinse water tests