Product Data Sheet DAN-PDS-GC-700 Daniel ® Danalyzer™ Model 700 Gas Chromatograph Model 700 Visit us online at www.daniel.com Features and Benefits One package for fiscal metering or gas quality Custody transfer analysis from C 6 + to C 9 + Contaminant monitoring - trace H 2 S, CO 2 , oxygen, etc. Combine measurements and reduce analysis cost – Oxygen (0-2%) – Trace H 2 S (0-50 ppm) – CO 2 (as fast as 90 seconds) – Helium/Hydrogen (0-10%) – Hydrocarbon Dew Point (±2°C) Reduced installation costs 24 VDC power standard Integrated controller electronics Pipe-mount, wall-mount, or floor mount Lower operation and maintenance costs No shelter or instrument air required Low helium and power consumption Longest gas chromatograph valve and column warranties available Unmatched measurement performance Highest stated precision — ±0.25 BTU/1000 for broad ambient temperature C 6 + analysis in -18° to 54°C (0°-130°F) environmental conditions Wide dynamic range from percent to trace level components Reliable performance over broad ambient temperatures -30° C to 60° C (-20°-140° F) Easy to use Compatible with Danalyzer Model 500 Gas Chromatographs MON2000™ software setup and diagnostics Identical setup and operation for all models – no additional training Danalyzer™ Model 700 Gas Chromatographs offer the broadest range of analysis options available today in a field- mount gas chromatograph (GC). Whether it is heating value measurement, trace contaminant monitoring, pipeline integrity, or product quality/process control, the Danalyzer Model 700 is flexible enough to meet your analysis needs. Based on the same proven technology and software as the Danalyzer™ Model 500 series, the Danalyzer Model 700 design offers superior reliability and precision, lower installation and operating costs, greater application flexibility, and unmatched measurement performance.

Upload: others

Post on 12-Sep-2021

31 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Product Data SheetDAN-PDS-GC-700

Daniel® Danalyzer™ Model 700Gas Chromatograph

Model 700

Visit us online atwww.daniel.com

Features and Benefits

One package for fiscal metering or gas quality Custody transfer analysis from C6+ to C9+ Contaminant monitoring - trace H2S, CO2, oxygen,

etc. Combine measurements and reduce analysis cost

– Oxygen (0-2%)– Trace H2S (0-50 ppm)– CO2 (as fast as 90 seconds)– Helium/Hydrogen (0-10%)– Hydrocarbon Dew Point (±2°C)

Reduced installation costs 24 VDC power standard Integrated controller electronics Pipe-mount, wall-mount, or floor mount

Lower operation and maintenance costs No shelter or instrument air required Low helium and power consumption Longest gas chromatograph valve and column

warranties available

Unmatched measurement performance

Highest stated precision — ±0.25 BTU/1000 for broad ambient temperature C6+ analysis in -18° to 54°C (0°-130°F) environmental conditions

Wide dynamic range from percent to trace level components

Reliable performance over broad ambient temperatures -30° C to 60° C (-20°-140° F)

Easy to use

Compatible with Danalyzer Model 500 Gas Chromatographs

MON2000™ software setup and diagnostics Identical setup and operation for all models – no

additional training

Danalyzer™ Model 700 Gas Chromatographs offer the broadest range of analysis options available today in a field-mount gas chromatograph (GC). Whether it is heating value measurement, trace contaminant monitoring, pipeline integrity, or product quality/process control, the Danalyzer Model 700 is flexible enough to meet your analysis needs. Based on the same proven technology and software as the Danalyzer™ Model 500 series, the Danalyzer Model 700 design offers superior reliability and precision, lower installation and operating costs, greater application flexibility, and unmatched measurement performance.

Page 2: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 2

Product DatasheetDAN-PDS-GC-700

APPLICATIONS

Standard Natural Gas ApplicationsWe have made our most popular energy and gas quality applications standard. Applications may vary by components of interest, analysis time, reduced hardware (economics), or improved precision. Choose the one that is right for you, or we will create one that is customized for your specific application.

Energy Measurement (to C6+ and C9+)The Danalyzer Model 700 offers applications for energy measurement from C6+ hydrocarbon ranges to C9+ hydrocarbon ranges. Calculations based on GPA 2145 or ISO 6976 standards can be provided.

Standard Measurement RangesMethane 65 to 100-mole %Ethane 0 to 20-mole %Propane 0 to 10-mole %N-Butane 0 to 5-mole%Iso-Butane 0 to 5-mole %N-Pentane 0 to 1 mole %Iso-Pentane 0 to 1 mole%Neo-Pentane 0 to 1 mole%Hexane+ 0 to 0.7-mole%Nitrogen 0 to 20 mole %Carbon Dioxide 0 to 20 mole%Hexanes 0 to 1 mole%Heptanes 0 to 1 mole%Octanes 0 to 0.5 Mole%Nonane+ 0 to 0.5 mole%

Gas Quality AnalysisNatural gas contaminants reduce pipeline integrity over time. Most contaminants can be easily measured in the Model 700 for online quality assurance. Contaminant monitoring can be combined with energy measurements for complete custody transfer analysis. To the extent possible, these combined applications utilize independent gas chromatograph valves, detectors, and columns, for each primary measurement. This technique offers greater reliability, increased speed, and easier troubleshooting. This application approach also makes field upgrades and re-applications in the Danalyzer Model 700 easy by minimizing internal piping changes.

C6+ C9+

Hydrocarbon Dew Point MonitoringAccurate and reliable hydrocarbon dew point determination can now be made as an option to any C9+ extended analysis application. The DewCalc™ Hydrocarbon Dew Point software is completely integral to the gas chromatograph and provides dew point temperatures and cricondentherm for up to four user-entered pressures. Alternately, real-time dew point results can be provided by using analog or Modbus inputs from another device.

DewCalc™ software is loaded with features, including:

Choice of two equations of state, Peng-Robinson or Soave-Redlich-Kwong

Library of 93 hydrocarbon compounds for detailed breakdown of C9+ fraction

DewCalc report with analysis results and factors used

Custom ApplicationsIf our standard applications do not fit your needs, the Danalyzer Model 700 can be customized to meet your measurement requirements. Simply submit a completed application data sheet found at the end of this data sheet with your request, or ask our Application Engineers for assistance.

Page 3: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 3

Product DatasheetDAN-PDS-GC-700

Standard Danalyzer™ Applications and Repeatability

SUPERIOR PERFORMANCEThe Danalyzer Difference

Modular Analytical OvenAn element that is key to the Danalyzer Model 700 Gas Chromatograph’s application flexibility is the modular analytical oven. Using field-proven Danalyzer valves, columns, and detectors, the analytical oven has been redesigned to make it modular and scalable. Six- or ten-port valves and detectors mount to a triangular “kiosk” block which facilitates valve heating and pneumatic actuation. Three temperature control zones and up to three valves and two detectors operating independently provide extreme application flexibility and range. All components in the oven are completely accessible and serviceable in the field.

Gas Chromatograph ValvesAt the heart of the Danalyzer’s superior reliability are the six- and ten-port diaphragm/piston gas chromatograph valves. These pneumatic valves are guaranteed for the life of the gas chromatograph and specified to operate over 5 million times between service. By minimizing internal movement (1/1,000 inch) of the pistons, which never come in contact with sample, abrasive mechanical wear is virtually eliminated. This unique double-diaphragm design removes the need for all springs, o-rings, or lubrication. Valve service is performed by replacing a cost-effective diaphragm set, normally completed in less than 10 minutes.

DetectorsThe thermal conductivity detector (TCD) is the detector of choice for most applications due to its universal response to all compounds. The TCD found in the Danalyzer Model 700 Gas Chromatograph is able to go well beyond the normal measuring ranges seen in other designs by being able to do many applications

Application Code Application Name Components

MeasuredAnalysis

Time% Repeatability

@ Controlled temperature% Repeatability Over

0-130˚F

604 Standard BTU/CV C1 to C6+, CO2, N2 4 min ±0.0125 ±0.025

6GT Gas Turbine BTU/CV/Density

C1 to C6+, CO2, N2 2 min ±0.05 ±0.1

TS2 Gas Quality (H2S) Trace H2S (3 - 30 ppm) 90 sec ±2 ppm ±3 ppm

LF8 Landfill Gas BTU/CV C1, N2, CO2, O2 8 min ±0.1 ±0.2

LF3 Landfill Gas (mole-sieve) C1, N2, CO2, O2 3 min ±0.1 ±0.2

6PS Std C6+ with % H2S C1 to C6+, CO2, N2, H2S (%)

10 min ±0.05 ±0.1

6PO Std C6+ with % Oxygen C1 to C6+, CO2, N2, O2 10 min ±0.05 ±0.1

6HH Std C6+ with Helium/Hydrogen

C1 to C6+, CO2, N2, He/H2

5 min ±0.05 ±0.1

6TS C6+ with ppm H2S C1 to C6+, CO2, N2, H2S (ppm)

7 min ±0.0125/±2 ppm ±0.015/±3 ppm

905 Std C9+ C1 to C9+, CO2, N2 5 min ±0.025 ±0.05

HC9 Std C9+ and Hydrocar-bon Dewpoint

C1 to C9+, CO2, N2 5 min ±0.025 ±0.05

712 Standard C7+ BTU/CV C1 to C7+, CO2, N2 10-12 min ±0.025 ±0.05

Page 4: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 4

Product DatasheetDAN-PDS-GC-700

with low parts-per-million measurement requirements. This greatly simplifies the gas chromatograph design when a simple and rugged TCD can be used. The flame ionization detector (FID) for trace hydrocarbon measurements and the flame photometric detector (FPD) module for trace sulfur measurements are also available.

Micro-packed ColumnsDanalyzer micro-packed columns offer a superior combination of features found in both capillary and conventional packed columns — speed, sharp peak resolution, and low carrier gas consumption. In addition, this unique design provides for greatly extended column life and the longest warranty available on the market (five years on the standard C6+ natural gas set).

Stream Switching ModuleThe Danalyzer Gas Chromatograph offers an internal sample stream switching module. Available in four- or eight-stream versions, this module utilizes the same helium-rated 24-volt solenoids used in the analytical oven. It is temperature-controlled for operation above the sample dewpoint. For applications with widely varying stream composition, double block and bleed configuration is optionally available.

CONTROLLER ELECTRONICS AND COMMUNICATIONS

Modular ElectronicsThe controller electronics, option cards, and field termination boards are packaged conveniently in the lower enclosure of the Danalyzer Model 700. Separate power and output connections are made through this enclosure.

Compatible Electronics and SoftwareThe Danalyzer Model 700 uses the same CPU board as the Danalyzer Model 500, thus making features, options, and operation nearly identical. No new training is required.

Local Indication and Operation PanelAnalyzer health and valve status can be viewed through the glass cover of the lower enclosure. The panel displays green (OK), yellow (warning), and red (failure) LEDs, along with LEDs for all gas chromatograph valve actuations, power, and CPU health. Each valve can be actuated manually for simplified troubleshooting and fast system purging after maintenance.

Touch Key Local Operator Interface (Optional)The Model 700 local operator interface (LOI) allows for maintenance and operation of a Model 700 without a laptop or PC. The LOI is a state-of-the-art high resolution color display that is touch key infrared activated and supports all core GC operations.

Features of the LOI include:- Color LCD with QVGA (320 x 240 pixels) resolution- ASCII text and graphics modes- Auto-backlighting (adjustable)- 8 infrared-activated touch keys and screen saver - Eliminates external magnetic pen requirement - Maintains the Model 700 hazardous area

classifications- Operates in the same ambient temperatures as the

base Model 700- Complete GC status, control and diagnostics,

including full chromatogram display

Model 700 shown with optional touch-key LCD LOI

Page 5: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 5

Product DatasheetDAN-PDS-GC-700

Data archiving and reportsThe Danalyzer Model 700, with expandable solid-state memory, virtually eliminates the need for external data storage. Every analysis is time/date stamped and archived, for retrieval via the MON2000 software. Pre-configured reports can be displayed, printed, and/or stored internally. Results can be trended directly or exported easily in ASCII format.

Security - three levels of password protected security configurable to read/write or read only for third party access

Audit logs - data and event logging fully conforms to API report 21.1 for metering audit purposes and backup to primary systems (flow computer, SCADA, DCS)

Event logs - a continuous record of all operator changes, with time, date, and operator name recorded

Alarm logs - a continuous record of all historical alarms, time/date stamped with alarm state and description

Maintenance log - a “scratch pad” for keeping track of maintenance or testing performed on the gas chromatograph system

Archiving - the last 35 days of each analysis and 64 calibrations can be archived automatically by time/date stamp. Last chromatogram stored for each sample and calibration stream.

Standard reports include:

Average reports - hourly, 24-hr., weekly, monthly and variable averages

Analysis reports - physical property calculations by component and group for analysis and alarms

Raw data report - retention times, peak areas, detector number, method, integration start/stop, and peak width for the analysis

Calibration report - raw component data, new response factors, retention times, and deviation for last calibration

Final calibration report - results of final calibration, response factor and retention time adjustments

ENVIRONMENTAL CHAMBER TESTINGWe provide the most thorough gas chromatograph testing in the world. Prior to shipment, we test nearly every Danalyzer Model 700 in our walk-in environmental test chambers, where temperatures are cycled between 0°F and 130°F for a minimum of 24 hours. This is all part of our commitment to provide analyzers that are capable of providing reliable measurements in the field.

Superior measurement stability in extreme climates is tested for every Model 700 Process Gas Chromatographs before it ships.

1034

1035

1036

1037

1038

1039

1040

Stated Repeatability (oven temperature): ±0.05% CVRepeatability as tested: ±0.0165%

ELAPSED TIME

BTU

70oF

0oF

130oF

COLD RAMP COLD SOAK HOT RAMP HOT SOAK

BTU Repeatability In Environmental Chamber

0:00

1:07

2:14

3:21

4:28

5:35

6:43

7:50

8:57

10:0

4

11:1

1

12:1

8

13:2

5

14:3

3

15:4

0

16:4

7

17:5

4

19:0

1

20:0

8

21:1

6

22:2

3

71 naJ 04 eD g C

Model 700 C + Application - Pipeline Gas9

Page 6: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 6

Product DatasheetDAN-PDS-GC-700

MON2000™ SOFTWAREThe Danalyzer Model 700 Gas Chromatograph is designed to operate unattended. If adjustments are needed, our exclusive MON2000™ software allows complete control of your gas chromatographs – either locally or remotely. From within MON2000, a user can:

Review and modify analytical settings on one screen

Upload and display multiple chromatograms on the screen for comparison

Upload and trend any of the measured results

Export data for use in other third-party applications

Overlay multiple chromatograms for troubleshooting and calibration

Check original calibration against last calibration

The MON2000 software is Windows®-based software designed to make analyzer configuration, maintenance and data collection easy. With intuitive drop-down menus and fill-in-the-blank tables, even new users can quickly navigate through the software. The MON2000 software can display both current and multiple archived chromatograms on the screen streamlining the time needed to perform routine analyzer maintenance.

MON2000 also has a number of tools built in to help users manage their analyzers such as:

Automatic recording of alarms in a log file

Event logs that provide a continuous record of all operator changes with time and user name stored

Maintenance log scratch pad for keeping track of maintenance or testing done

Download and install the latest GPA 2145 and ISO 6976 physical property factors

Data collected from the Danalyzer Model 700 can be stored and displayed in a wide range of options such as trend lines on the screen and logs automatically docu-menting all changes made to the gas chromatograph. Data can also be exported in formats compatible with most third-party Windows® applications.

Simple drop-down menus

Powerful Yet Simple – MON2000 Software

Connect to any GC with a mouse click

Full featured chromatogramdisplay

Summary of application event status and time

Automatic listing of measuredcomponents

Quickly addchromatogramsto overlay

Save chromatogramsto hard drive

Display resultsInteractive cursur

Manually adjust X/Y scale

Page 7: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 7

Product DatasheetDAN-PDS-GC-700

Secure Modbus Connectivity

For online gas chromatographs, Modbus continues to be the preferred choice to connect to a gas chromato-graph network. Our Modbus design avoids the use of central interface cards or computers that can act as a single-point of failure in the Modbus link. Instead, the flow computer or SCADA system can “talk” directly to each gas chromatograph to gather the data needed. Furthermore, the register and coil addresses can be easily customized to meet the specific data structure of the flow computer or SCADA system. There is also a program built into the MON2000 workstation to test the Modbus link if troubleshooting is needed.

ANALYZER NETWORKING AND DATA COMMUNICATIONDanalyzer Gas Chromatographs can be configured in a number of networking and data communication schemes to meet the communication requirements of the natural gas industry. Options include modem, Ethernet, as well as multi-drop RS-485 networks. It is even possible to set up automatic polling for data collection if desired. Data communication options include simple analog and discrete signals as well as Modbus serial links. To preserve the integrity of the analysis data, all Danalyzer Model 700 Gas Chromatographs are capable of storing up to 35 days of analysis or calibration data in the event of loss of communication. All Danalyzer Gas Chromatographs are designed to operate unattended. Occasionally, adjustments to the analyzer analytical method or a review of possible alarms may be needed. Using our exclusive MON2000 software loaded on either a PC or laptop running Windows®, you’ll have complete control of your gas chromatographs – either locally or remotely.

Networking Flexibility

Whether you want to network gas chromatographs throughout the plant or simply link a single gas chro-matograph to a flow computer system, the Danalyzer Model 700 can be configured to handle most of the scenarios:

Choice of Ethernet or RS-485 network

Can use same network to connect Danalyzer Model 700, Model 500, and Model 1000 Gas Chromato-graphs

Able to connect multiple PC workstations using MON2000

Connectivity to flow computer or SCADA systems using industry standard protocols such as Modbus and OPC

MODEL 1000

P Q R

MODEL 500

OPTIONAL LAPTOPFOR LOCAL

COMMUNICATION

- RS 232C- RS 485/422- ETHERNET

MODEL 700

REDUNDANT RS 485/422

UP TO32 GC’S

PRINTER / LOGGER

ETHERNET - UNLIMITED NUMBER OF GC’S

MON2OOOWORKSTATION

- RS 232C- RS 485/422- ETHERNET

MODEL 500

MODEL 700

INTERNET

MON2OOOWORKSTATION

DCS OPC SERVER

OPC SERVER

MON2OOOWORKSTATION

- RS 232C- RS 485/422- ETHERNET

MODEL 500

MODEL 1000

P Q R

MODEL 500

MODEL 700OPTIONAL LAPTOP

FOR LOCALCOMMUNICATION

- RS 232C- RS 485/422- ETHERNET

MODEL 700

MODEL 700

Simplified Network Communications - Multiple serial ports with Ethernet

option can be field configured to provide redundant data highways.

Page 8: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 8

Product DatasheetDAN-PDS-GC-700

ACCESSORIESWe offer standard and custom gas chromatograph accessories to meet your installation and application requirements.

Gas Chromatograph System Enclosures Complete range of gas chromatograph enclosure

solutions ranging from simple sun-shields and cabinets to skids systems and shelters

API 14.1 compliant solutions with heating of sample lines and calibration gas

Standard Auxiliary Equipment Sample probes / pressure regulators / filters for a

wide range of sample handling requirements Carrier gas systems with dual manifold regulators for

uninterrupted operation Calibration gases and heating blankets

GAS CHROMATOGRAPH ACCESSORIES Laptop Computer Shelf mounts easily - folds down

on side (PN 2-3-0700-143 )

"Condulet" Junction Box for quick laptop PC (db9 connector) access

Custom-Engineered Sample SystemsAny gas chromatograph is only as good as the quality of the sample it measures. Every sample system for Danalyzer Gas Chromatographs is custom-engineered for the specific requirements of the application. Common features include:

Heated and open panel designs

All components rated for the area classification

Automatic calibration / validation available as an option

Variety of sample probes to extract a reliable and stable sample

3 - Sided Shelter

Sample Probe

Carrier Manifold

Calibration Gas with Heating Blanket

Multi-stream Heated Sample Conditioning System

Condulet JunctionBox

Laptop ComputerShelf

Page 9: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 9

Product DatasheetDAN-PDS-GC-700

Danalyzer™ Model 700 Gas ChromatographCSA Certified Danalyzer GC, 6 or 10 port valves, single or dual detectors

Ordering Matrix (must select only 1 option for each column)77X C W DC 4 SN N S4 612 1N S (example)

Code ApprovalsCA

CSAATEX

Code MountingW Wall-mountP Pipe-mountF Floor-mount

Code PowerDC 24 VDCAC Universal 85 to 240 VAC

Code Digital/Analog I/O4 5 Digital In; 4 Digital Out; 4 Analog In; 4 Non-isolated Analog Out8 5 Digital In; 4 Digital Out; 4 Analog In; 8 Analog Out, 4 Non-Isolated/4 Isolated

12 5 Digital In; 4 Digital Out; 4 Analog In; 12 Non-isolated Analog Out

Code Serial Communications/Local Keypad-LCDSN Standard/no keypad - LCD {(2) RS232, (1) RS422/486 Factory/Field configurable}EN Extended/no keypad - LCD {(4) RS232, (2) RS422/486 Factory/Field configurable}

Code Internal Modem and/or EthernetN No Modem or EthernetM Modem onlyE Ethernet onlyB Modem and EthernetX Special (consult factory)

Code Internal Stream Switching (Total max. # streams = sample + calibration streams)

S4 Standard – Up to 4 streamsS8 Standard – Up to 8 streams D4 Double block and bleed - Up to 4 streamsD8 Double block and bleed - Up to 8 streamsES External stream switching - (consult factory)

X Special (consult factory)

Code Oven Assembly/Application612 BTU/CV C6+ / 12 min.604 BTU/CD C6+ / 4 min.6GT BTU/CV C6+ / 2 min./TurbineTS2 Trace H2S / 3 min.*LF8 Landfill Gas / 8 min.LF3 Landfill Gas / 3 min. (mole-sieve)6PS Std. C6+ with % H2S6PO Std. C6+ with % Oxygen6HH Std. C6+ with Helium/Hydrogen6TS C6+ / 12 min and ppm H2S / 7 min.*905 Std. C9+ and HC Dew

X Special

Code Sample Conditioning (sample streams only)1N 1 stream plate1G 1 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off2N 2 stream plate2G 2 stream plate, with Fast Loop/Membrane Filters/Liq. Shut-off3N 3 stream plate3G 3 stream plate, with Fast Loop/Membrane Filters/Liq. Shut-off4N 4 stream plate4G 4 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off5N 5 stream plate5G 5 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off6N 6 stream plate6G 6 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off7N 7 stream plate7G 7 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-offO Heated Sample System (Order Separately)X Special (consult factory)

Code Application SoftwareS Std. GPA 2172/GPA2145I ISO 6976;1995

X Special (consult factory)

Manuals: (one set included with each system)Hardware manual - P/N 3-9000-521Software manual - P/N 3-9000-522

All Danalyzer Model 700 Gas Chromatographs must be crated for shipment.

* Users with rich natural gas applications may require heated sample system ovens and sample transport tubing to insure that the sample is kept above its sample dew point. Consult the factory to determine if heated sample system is needed.

Page 10: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 10

Product DatasheetDAN-PDS-GC-700Application Data SheetEvery Danalyzer Model 700 Gas Chromatograph is custom built for the specific application requested. To request a free quotation, simply fill out the information below and send it to your world area contact information located on the back cover of this datasheet.

(For more than two streams, make copies of this page)

Stream Composition Units Stream 1 Stream 2

Min Normal Max Measure Min Normal Max Measure

Your Reference #:

Project Name:

Process Unit:

Stream Name(s):

Hazardous Area Classification:

Name:

Title:

Company:

Address:

Phone #:

Fax #:

Email:

Stream 1 Stream 2Stream TemperatureStream PressureStream Phase (vapor/liquid)Stream Contaminants

Distance to GC from sample point

Power: 100-240 VAC ( ) 24 VDC ( )Mounting: Wall ( ) Pipe ( ) Stand ( )Data communication Analog output ( ) Modbus ( ) OPC ( ) Printer ( ) Modem ( )Analyzer network: Ethernet ( ) RS-485 Multi-Drop ( )Other options: Carrier Gas ( ) Calibration Gas ( ) Start-up ( ) Training Class ( )Special Instructions:

Page 11: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 11

Product DatasheetDAN-PDS-GC-700

OUTLINE AND MOUNTING DETAILS

SPECIFICATIONSPower:

Standard: 24 VDC, (24.5-26.5 VDC) Optional: 90-264 VAC, 47-63 Hz 80 watts start-up; 33 watts nominal

Environment: -29° to 60°C (-20° to 140°F) Enclosure Protection Rating: IP66 and NEMA 4XDimensions (without sample system):

Wall-mount: 914mm H x 462mm W x 483mm D (36” H x 18.2” W x 19” D) Pipe-mount: 914mm H x 462mm W x 645mm D (36” H x 18.2” W x 25.4” D) Floor-mount: 1483mm H x 419mm W x 577mm D (58.4” H x 16.5” W x 22.7” D)

Mounting: Free-standing (standard); wall-mount or pipe-mount (optional)Approximate Weight (without sample system): 80 kg (175 lbs.)Area Safety Certification Options:

CSA: Class 1, Division 1, Groups B, C, D, T4 USA: Class I, Zone 1, Ex d IIB+H2, T4, Enclosure Type 4 Canada: Class I, Zone 1, Ex d IIB+H2, T4, Enclosure Type 4ATEX: Ex II 2G Ex d IIC T4 Tamb = 60°C IECEx: Ex d IIC T4 Gb Tamb = 60°C

Oven: Airless heat sink, maximum 130°C (266°F)Valves: 6-port and 10-port diaphragm chromatograph valves. Other types of

valves may be used, such as liquid injection or rotary valves, depending on the application

Carrier Gas: Application-dependent. Typically zero-grade helium, nitrogen, or hydrogen

4“ OR 2” POLEMOUNT OPTION

25.4

22.4

26.25REMOVAL

CLEARANCETYPICAL

SAMPLE PLATE

16.7518.2

56.6

1.56

12.0

SAMPLE PLATE

4X Ø .50 THRU

36.8

16.5 22.1

58.4

SIDE VIEW FLOOR MOUNT UNITFRONT VIEW FLOOR MOUNT UNIT

FRONT VIEW WALL MOUNT UNIT

POLE MOUNT UNIT

22.715.50

13.6

33.8RECOMMENDEDFROM GROUND

OR FLOOR

Sample Input Pressure Range (recommended): 15-20 psigCarrier Gas Input Pressure Range (recommended): 90-100 psigDetector: Thermal conductivity detector (TCD), flame ionization detector

(FID), TCD/TCD or TCD/FID dual detector configurations possible; flame photometric detector (FPD) available (see FPD module data sheet)

Gating Options: Fixed-time automatic slope sense gating of peaksStreams: Up to eight streams (including calibration stream) internally

controlled; up to 12 streams externally controlledChromatograms stored/archived internally: Most recent analysis and/or

calibration per stream Analog Inputs: Four inputs filtered with transient protection (user scalable and

assignable)Analog Outputs: Four non-isolated analog outputs, 4 – 20 mA (standard); four

additional isolated analog outputs or eight additional non-isolated analog outputs (optional)

Serial Communication Ports: Three serial ports standard with option for a total of seven. Depending on the port, choice of RS-232, RS-422, and RS-485 is available as well as the Modbus protocol

Digital Inputs: Five digital inputs; user assignable Digital Outputs: One gas chromatograph alarm and four user-assignable

outputs (2 form A, 3 solid-state optically isolated)Parallel Printer Port: One parallel port available for printed reportsInternal Modem (optional): Field-configurable; 300 to 33.6k baud rateEthernet Card (optional): 10 mbps with RJ-45 portRAM (DATA) Memory Capacity: 128 Mb RAM (standard)

Page 12: Daniel® Danalyzerâ„¢ Model 700 Gas Chromatograph

Page 12

Product DatasheetDAN-PDS-GC-700

Emerson Process ManagementDaniel Measurement and Control, Inc.

World Area HeadquartersHouston, Texas, USAT: 713-827-6380, F: 713-827-3865USA Toll Free [email protected]

Calgary, Alberta, Canada, T: 403-279-1879, F: 403-236-1337

Singapore - Asia Pacific, T: +65-6777-8211, F: +65-6770-8001

Stirling, Scotland - Europe, Mid-East, Africa, T: +44-1259-727-222, F: +44-1786-433-606

Daniel Measurement and Control, Inc. and Rosemount Analytical are a wholly owned subsidiaries of Emerson Electric Co., and are divisions of Emerson Process Management. The Daniel name and logo are registered trademarks of Daniel Industries, Inc. The Emerson logo is a registered trademark and service mark of Emerson Electric Co. All other trademarks are the property of their respective companies. The contents of this publication are presented for informational purposes only, and while every effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by Daniel’s terms and conditions, which are available upon request. We reserve the right to modify or improve the designs or specifications of such products at any time. Daniel does not assume responsibility for the selection, use or maintenance of any product. Responsibility for proper selection, use and maintenance of any Daniel product remains solely with the purchaser and end-user. Daniel products are protected by United States and international patents and patents pending.