Product Data SheetDAN-PDS-GC-700
Daniel® Danalyzer™ Model 700Gas Chromatograph
Model 700
Visit us online atwww.daniel.com
Features and Benefits
One package for fiscal metering or gas quality Custody transfer analysis from C6+ to C9+ Contaminant monitoring - trace H2S, CO2, oxygen,
etc. Combine measurements and reduce analysis cost
– Oxygen (0-2%)– Trace H2S (0-50 ppm)– CO2 (as fast as 90 seconds)– Helium/Hydrogen (0-10%)– Hydrocarbon Dew Point (±2°C)
Reduced installation costs 24 VDC power standard Integrated controller electronics Pipe-mount, wall-mount, or floor mount
Lower operation and maintenance costs No shelter or instrument air required Low helium and power consumption Longest gas chromatograph valve and column
warranties available
Unmatched measurement performance
Highest stated precision — ±0.25 BTU/1000 for broad ambient temperature C6+ analysis in -18° to 54°C (0°-130°F) environmental conditions
Wide dynamic range from percent to trace level components
Reliable performance over broad ambient temperatures -30° C to 60° C (-20°-140° F)
Easy to use
Compatible with Danalyzer Model 500 Gas Chromatographs
MON2000™ software setup and diagnostics Identical setup and operation for all models – no
additional training
Danalyzer™ Model 700 Gas Chromatographs offer the broadest range of analysis options available today in a field-mount gas chromatograph (GC). Whether it is heating value measurement, trace contaminant monitoring, pipeline integrity, or product quality/process control, the Danalyzer Model 700 is flexible enough to meet your analysis needs. Based on the same proven technology and software as the Danalyzer™ Model 500 series, the Danalyzer Model 700 design offers superior reliability and precision, lower installation and operating costs, greater application flexibility, and unmatched measurement performance.
Page 2
Product DatasheetDAN-PDS-GC-700
APPLICATIONS
Standard Natural Gas ApplicationsWe have made our most popular energy and gas quality applications standard. Applications may vary by components of interest, analysis time, reduced hardware (economics), or improved precision. Choose the one that is right for you, or we will create one that is customized for your specific application.
Energy Measurement (to C6+ and C9+)The Danalyzer Model 700 offers applications for energy measurement from C6+ hydrocarbon ranges to C9+ hydrocarbon ranges. Calculations based on GPA 2145 or ISO 6976 standards can be provided.
Standard Measurement RangesMethane 65 to 100-mole %Ethane 0 to 20-mole %Propane 0 to 10-mole %N-Butane 0 to 5-mole%Iso-Butane 0 to 5-mole %N-Pentane 0 to 1 mole %Iso-Pentane 0 to 1 mole%Neo-Pentane 0 to 1 mole%Hexane+ 0 to 0.7-mole%Nitrogen 0 to 20 mole %Carbon Dioxide 0 to 20 mole%Hexanes 0 to 1 mole%Heptanes 0 to 1 mole%Octanes 0 to 0.5 Mole%Nonane+ 0 to 0.5 mole%
Gas Quality AnalysisNatural gas contaminants reduce pipeline integrity over time. Most contaminants can be easily measured in the Model 700 for online quality assurance. Contaminant monitoring can be combined with energy measurements for complete custody transfer analysis. To the extent possible, these combined applications utilize independent gas chromatograph valves, detectors, and columns, for each primary measurement. This technique offers greater reliability, increased speed, and easier troubleshooting. This application approach also makes field upgrades and re-applications in the Danalyzer Model 700 easy by minimizing internal piping changes.
C6+ C9+
Hydrocarbon Dew Point MonitoringAccurate and reliable hydrocarbon dew point determination can now be made as an option to any C9+ extended analysis application. The DewCalc™ Hydrocarbon Dew Point software is completely integral to the gas chromatograph and provides dew point temperatures and cricondentherm for up to four user-entered pressures. Alternately, real-time dew point results can be provided by using analog or Modbus inputs from another device.
DewCalc™ software is loaded with features, including:
Choice of two equations of state, Peng-Robinson or Soave-Redlich-Kwong
Library of 93 hydrocarbon compounds for detailed breakdown of C9+ fraction
DewCalc report with analysis results and factors used
Custom ApplicationsIf our standard applications do not fit your needs, the Danalyzer Model 700 can be customized to meet your measurement requirements. Simply submit a completed application data sheet found at the end of this data sheet with your request, or ask our Application Engineers for assistance.
Page 3
Product DatasheetDAN-PDS-GC-700
Standard Danalyzer™ Applications and Repeatability
SUPERIOR PERFORMANCEThe Danalyzer Difference
Modular Analytical OvenAn element that is key to the Danalyzer Model 700 Gas Chromatograph’s application flexibility is the modular analytical oven. Using field-proven Danalyzer valves, columns, and detectors, the analytical oven has been redesigned to make it modular and scalable. Six- or ten-port valves and detectors mount to a triangular “kiosk” block which facilitates valve heating and pneumatic actuation. Three temperature control zones and up to three valves and two detectors operating independently provide extreme application flexibility and range. All components in the oven are completely accessible and serviceable in the field.
Gas Chromatograph ValvesAt the heart of the Danalyzer’s superior reliability are the six- and ten-port diaphragm/piston gas chromatograph valves. These pneumatic valves are guaranteed for the life of the gas chromatograph and specified to operate over 5 million times between service. By minimizing internal movement (1/1,000 inch) of the pistons, which never come in contact with sample, abrasive mechanical wear is virtually eliminated. This unique double-diaphragm design removes the need for all springs, o-rings, or lubrication. Valve service is performed by replacing a cost-effective diaphragm set, normally completed in less than 10 minutes.
DetectorsThe thermal conductivity detector (TCD) is the detector of choice for most applications due to its universal response to all compounds. The TCD found in the Danalyzer Model 700 Gas Chromatograph is able to go well beyond the normal measuring ranges seen in other designs by being able to do many applications
Application Code Application Name Components
MeasuredAnalysis
Time% Repeatability
@ Controlled temperature% Repeatability Over
0-130˚F
604 Standard BTU/CV C1 to C6+, CO2, N2 4 min ±0.0125 ±0.025
6GT Gas Turbine BTU/CV/Density
C1 to C6+, CO2, N2 2 min ±0.05 ±0.1
TS2 Gas Quality (H2S) Trace H2S (3 - 30 ppm) 90 sec ±2 ppm ±3 ppm
LF8 Landfill Gas BTU/CV C1, N2, CO2, O2 8 min ±0.1 ±0.2
LF3 Landfill Gas (mole-sieve) C1, N2, CO2, O2 3 min ±0.1 ±0.2
6PS Std C6+ with % H2S C1 to C6+, CO2, N2, H2S (%)
10 min ±0.05 ±0.1
6PO Std C6+ with % Oxygen C1 to C6+, CO2, N2, O2 10 min ±0.05 ±0.1
6HH Std C6+ with Helium/Hydrogen
C1 to C6+, CO2, N2, He/H2
5 min ±0.05 ±0.1
6TS C6+ with ppm H2S C1 to C6+, CO2, N2, H2S (ppm)
7 min ±0.0125/±2 ppm ±0.015/±3 ppm
905 Std C9+ C1 to C9+, CO2, N2 5 min ±0.025 ±0.05
HC9 Std C9+ and Hydrocar-bon Dewpoint
C1 to C9+, CO2, N2 5 min ±0.025 ±0.05
712 Standard C7+ BTU/CV C1 to C7+, CO2, N2 10-12 min ±0.025 ±0.05
Page 4
Product DatasheetDAN-PDS-GC-700
with low parts-per-million measurement requirements. This greatly simplifies the gas chromatograph design when a simple and rugged TCD can be used. The flame ionization detector (FID) for trace hydrocarbon measurements and the flame photometric detector (FPD) module for trace sulfur measurements are also available.
Micro-packed ColumnsDanalyzer micro-packed columns offer a superior combination of features found in both capillary and conventional packed columns — speed, sharp peak resolution, and low carrier gas consumption. In addition, this unique design provides for greatly extended column life and the longest warranty available on the market (five years on the standard C6+ natural gas set).
Stream Switching ModuleThe Danalyzer Gas Chromatograph offers an internal sample stream switching module. Available in four- or eight-stream versions, this module utilizes the same helium-rated 24-volt solenoids used in the analytical oven. It is temperature-controlled for operation above the sample dewpoint. For applications with widely varying stream composition, double block and bleed configuration is optionally available.
CONTROLLER ELECTRONICS AND COMMUNICATIONS
Modular ElectronicsThe controller electronics, option cards, and field termination boards are packaged conveniently in the lower enclosure of the Danalyzer Model 700. Separate power and output connections are made through this enclosure.
Compatible Electronics and SoftwareThe Danalyzer Model 700 uses the same CPU board as the Danalyzer Model 500, thus making features, options, and operation nearly identical. No new training is required.
Local Indication and Operation PanelAnalyzer health and valve status can be viewed through the glass cover of the lower enclosure. The panel displays green (OK), yellow (warning), and red (failure) LEDs, along with LEDs for all gas chromatograph valve actuations, power, and CPU health. Each valve can be actuated manually for simplified troubleshooting and fast system purging after maintenance.
Touch Key Local Operator Interface (Optional)The Model 700 local operator interface (LOI) allows for maintenance and operation of a Model 700 without a laptop or PC. The LOI is a state-of-the-art high resolution color display that is touch key infrared activated and supports all core GC operations.
Features of the LOI include:- Color LCD with QVGA (320 x 240 pixels) resolution- ASCII text and graphics modes- Auto-backlighting (adjustable)- 8 infrared-activated touch keys and screen saver - Eliminates external magnetic pen requirement - Maintains the Model 700 hazardous area
classifications- Operates in the same ambient temperatures as the
base Model 700- Complete GC status, control and diagnostics,
including full chromatogram display
Model 700 shown with optional touch-key LCD LOI
Page 5
Product DatasheetDAN-PDS-GC-700
Data archiving and reportsThe Danalyzer Model 700, with expandable solid-state memory, virtually eliminates the need for external data storage. Every analysis is time/date stamped and archived, for retrieval via the MON2000 software. Pre-configured reports can be displayed, printed, and/or stored internally. Results can be trended directly or exported easily in ASCII format.
Security - three levels of password protected security configurable to read/write or read only for third party access
Audit logs - data and event logging fully conforms to API report 21.1 for metering audit purposes and backup to primary systems (flow computer, SCADA, DCS)
Event logs - a continuous record of all operator changes, with time, date, and operator name recorded
Alarm logs - a continuous record of all historical alarms, time/date stamped with alarm state and description
Maintenance log - a “scratch pad” for keeping track of maintenance or testing performed on the gas chromatograph system
Archiving - the last 35 days of each analysis and 64 calibrations can be archived automatically by time/date stamp. Last chromatogram stored for each sample and calibration stream.
Standard reports include:
Average reports - hourly, 24-hr., weekly, monthly and variable averages
Analysis reports - physical property calculations by component and group for analysis and alarms
Raw data report - retention times, peak areas, detector number, method, integration start/stop, and peak width for the analysis
Calibration report - raw component data, new response factors, retention times, and deviation for last calibration
Final calibration report - results of final calibration, response factor and retention time adjustments
ENVIRONMENTAL CHAMBER TESTINGWe provide the most thorough gas chromatograph testing in the world. Prior to shipment, we test nearly every Danalyzer Model 700 in our walk-in environmental test chambers, where temperatures are cycled between 0°F and 130°F for a minimum of 24 hours. This is all part of our commitment to provide analyzers that are capable of providing reliable measurements in the field.
Superior measurement stability in extreme climates is tested for every Model 700 Process Gas Chromatographs before it ships.
1034
1035
1036
1037
1038
1039
1040
Stated Repeatability (oven temperature): ±0.05% CVRepeatability as tested: ±0.0165%
ELAPSED TIME
BTU
70oF
0oF
130oF
COLD RAMP COLD SOAK HOT RAMP HOT SOAK
BTU Repeatability In Environmental Chamber
0:00
1:07
2:14
3:21
4:28
5:35
6:43
7:50
8:57
10:0
4
11:1
1
12:1
8
13:2
5
14:3
3
15:4
0
16:4
7
17:5
4
19:0
1
20:0
8
21:1
6
22:2
3
71 naJ 04 eD g C
Model 700 C + Application - Pipeline Gas9
Page 6
Product DatasheetDAN-PDS-GC-700
MON2000™ SOFTWAREThe Danalyzer Model 700 Gas Chromatograph is designed to operate unattended. If adjustments are needed, our exclusive MON2000™ software allows complete control of your gas chromatographs – either locally or remotely. From within MON2000, a user can:
Review and modify analytical settings on one screen
Upload and display multiple chromatograms on the screen for comparison
Upload and trend any of the measured results
Export data for use in other third-party applications
Overlay multiple chromatograms for troubleshooting and calibration
Check original calibration against last calibration
The MON2000 software is Windows®-based software designed to make analyzer configuration, maintenance and data collection easy. With intuitive drop-down menus and fill-in-the-blank tables, even new users can quickly navigate through the software. The MON2000 software can display both current and multiple archived chromatograms on the screen streamlining the time needed to perform routine analyzer maintenance.
MON2000 also has a number of tools built in to help users manage their analyzers such as:
Automatic recording of alarms in a log file
Event logs that provide a continuous record of all operator changes with time and user name stored
Maintenance log scratch pad for keeping track of maintenance or testing done
Download and install the latest GPA 2145 and ISO 6976 physical property factors
Data collected from the Danalyzer Model 700 can be stored and displayed in a wide range of options such as trend lines on the screen and logs automatically docu-menting all changes made to the gas chromatograph. Data can also be exported in formats compatible with most third-party Windows® applications.
Simple drop-down menus
Powerful Yet Simple – MON2000 Software
Connect to any GC with a mouse click
Full featured chromatogramdisplay
Summary of application event status and time
Automatic listing of measuredcomponents
Quickly addchromatogramsto overlay
Save chromatogramsto hard drive
Display resultsInteractive cursur
Manually adjust X/Y scale
Page 7
Product DatasheetDAN-PDS-GC-700
Secure Modbus Connectivity
For online gas chromatographs, Modbus continues to be the preferred choice to connect to a gas chromato-graph network. Our Modbus design avoids the use of central interface cards or computers that can act as a single-point of failure in the Modbus link. Instead, the flow computer or SCADA system can “talk” directly to each gas chromatograph to gather the data needed. Furthermore, the register and coil addresses can be easily customized to meet the specific data structure of the flow computer or SCADA system. There is also a program built into the MON2000 workstation to test the Modbus link if troubleshooting is needed.
ANALYZER NETWORKING AND DATA COMMUNICATIONDanalyzer Gas Chromatographs can be configured in a number of networking and data communication schemes to meet the communication requirements of the natural gas industry. Options include modem, Ethernet, as well as multi-drop RS-485 networks. It is even possible to set up automatic polling for data collection if desired. Data communication options include simple analog and discrete signals as well as Modbus serial links. To preserve the integrity of the analysis data, all Danalyzer Model 700 Gas Chromatographs are capable of storing up to 35 days of analysis or calibration data in the event of loss of communication. All Danalyzer Gas Chromatographs are designed to operate unattended. Occasionally, adjustments to the analyzer analytical method or a review of possible alarms may be needed. Using our exclusive MON2000 software loaded on either a PC or laptop running Windows®, you’ll have complete control of your gas chromatographs – either locally or remotely.
Networking Flexibility
Whether you want to network gas chromatographs throughout the plant or simply link a single gas chro-matograph to a flow computer system, the Danalyzer Model 700 can be configured to handle most of the scenarios:
Choice of Ethernet or RS-485 network
Can use same network to connect Danalyzer Model 700, Model 500, and Model 1000 Gas Chromato-graphs
Able to connect multiple PC workstations using MON2000
Connectivity to flow computer or SCADA systems using industry standard protocols such as Modbus and OPC
MODEL 1000
P Q R
MODEL 500
OPTIONAL LAPTOPFOR LOCAL
COMMUNICATION
- RS 232C- RS 485/422- ETHERNET
MODEL 700
REDUNDANT RS 485/422
UP TO32 GC’S
PRINTER / LOGGER
ETHERNET - UNLIMITED NUMBER OF GC’S
MON2OOOWORKSTATION
- RS 232C- RS 485/422- ETHERNET
MODEL 500
MODEL 700
INTERNET
MON2OOOWORKSTATION
DCS OPC SERVER
OPC SERVER
MON2OOOWORKSTATION
- RS 232C- RS 485/422- ETHERNET
MODEL 500
MODEL 1000
P Q R
MODEL 500
MODEL 700OPTIONAL LAPTOP
FOR LOCALCOMMUNICATION
- RS 232C- RS 485/422- ETHERNET
MODEL 700
MODEL 700
Simplified Network Communications - Multiple serial ports with Ethernet
option can be field configured to provide redundant data highways.
Page 8
Product DatasheetDAN-PDS-GC-700
ACCESSORIESWe offer standard and custom gas chromatograph accessories to meet your installation and application requirements.
Gas Chromatograph System Enclosures Complete range of gas chromatograph enclosure
solutions ranging from simple sun-shields and cabinets to skids systems and shelters
API 14.1 compliant solutions with heating of sample lines and calibration gas
Standard Auxiliary Equipment Sample probes / pressure regulators / filters for a
wide range of sample handling requirements Carrier gas systems with dual manifold regulators for
uninterrupted operation Calibration gases and heating blankets
GAS CHROMATOGRAPH ACCESSORIES Laptop Computer Shelf mounts easily - folds down
on side (PN 2-3-0700-143 )
"Condulet" Junction Box for quick laptop PC (db9 connector) access
Custom-Engineered Sample SystemsAny gas chromatograph is only as good as the quality of the sample it measures. Every sample system for Danalyzer Gas Chromatographs is custom-engineered for the specific requirements of the application. Common features include:
Heated and open panel designs
All components rated for the area classification
Automatic calibration / validation available as an option
Variety of sample probes to extract a reliable and stable sample
3 - Sided Shelter
Sample Probe
Carrier Manifold
Calibration Gas with Heating Blanket
Multi-stream Heated Sample Conditioning System
Condulet JunctionBox
Laptop ComputerShelf
Page 9
Product DatasheetDAN-PDS-GC-700
Danalyzer™ Model 700 Gas ChromatographCSA Certified Danalyzer GC, 6 or 10 port valves, single or dual detectors
Ordering Matrix (must select only 1 option for each column)77X C W DC 4 SN N S4 612 1N S (example)
Code ApprovalsCA
CSAATEX
Code MountingW Wall-mountP Pipe-mountF Floor-mount
Code PowerDC 24 VDCAC Universal 85 to 240 VAC
Code Digital/Analog I/O4 5 Digital In; 4 Digital Out; 4 Analog In; 4 Non-isolated Analog Out8 5 Digital In; 4 Digital Out; 4 Analog In; 8 Analog Out, 4 Non-Isolated/4 Isolated
12 5 Digital In; 4 Digital Out; 4 Analog In; 12 Non-isolated Analog Out
Code Serial Communications/Local Keypad-LCDSN Standard/no keypad - LCD {(2) RS232, (1) RS422/486 Factory/Field configurable}EN Extended/no keypad - LCD {(4) RS232, (2) RS422/486 Factory/Field configurable}
Code Internal Modem and/or EthernetN No Modem or EthernetM Modem onlyE Ethernet onlyB Modem and EthernetX Special (consult factory)
Code Internal Stream Switching (Total max. # streams = sample + calibration streams)
S4 Standard – Up to 4 streamsS8 Standard – Up to 8 streams D4 Double block and bleed - Up to 4 streamsD8 Double block and bleed - Up to 8 streamsES External stream switching - (consult factory)
X Special (consult factory)
Code Oven Assembly/Application612 BTU/CV C6+ / 12 min.604 BTU/CD C6+ / 4 min.6GT BTU/CV C6+ / 2 min./TurbineTS2 Trace H2S / 3 min.*LF8 Landfill Gas / 8 min.LF3 Landfill Gas / 3 min. (mole-sieve)6PS Std. C6+ with % H2S6PO Std. C6+ with % Oxygen6HH Std. C6+ with Helium/Hydrogen6TS C6+ / 12 min and ppm H2S / 7 min.*905 Std. C9+ and HC Dew
X Special
Code Sample Conditioning (sample streams only)1N 1 stream plate1G 1 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off2N 2 stream plate2G 2 stream plate, with Fast Loop/Membrane Filters/Liq. Shut-off3N 3 stream plate3G 3 stream plate, with Fast Loop/Membrane Filters/Liq. Shut-off4N 4 stream plate4G 4 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off5N 5 stream plate5G 5 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off6N 6 stream plate6G 6 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off7N 7 stream plate7G 7 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-offO Heated Sample System (Order Separately)X Special (consult factory)
Code Application SoftwareS Std. GPA 2172/GPA2145I ISO 6976;1995
X Special (consult factory)
Manuals: (one set included with each system)Hardware manual - P/N 3-9000-521Software manual - P/N 3-9000-522
All Danalyzer Model 700 Gas Chromatographs must be crated for shipment.
* Users with rich natural gas applications may require heated sample system ovens and sample transport tubing to insure that the sample is kept above its sample dew point. Consult the factory to determine if heated sample system is needed.
Page 10
Product DatasheetDAN-PDS-GC-700Application Data SheetEvery Danalyzer Model 700 Gas Chromatograph is custom built for the specific application requested. To request a free quotation, simply fill out the information below and send it to your world area contact information located on the back cover of this datasheet.
(For more than two streams, make copies of this page)
Stream Composition Units Stream 1 Stream 2
Min Normal Max Measure Min Normal Max Measure
Your Reference #:
Project Name:
Process Unit:
Stream Name(s):
Hazardous Area Classification:
Name:
Title:
Company:
Address:
Phone #:
Fax #:
Email:
Stream 1 Stream 2Stream TemperatureStream PressureStream Phase (vapor/liquid)Stream Contaminants
Distance to GC from sample point
Power: 100-240 VAC ( ) 24 VDC ( )Mounting: Wall ( ) Pipe ( ) Stand ( )Data communication Analog output ( ) Modbus ( ) OPC ( ) Printer ( ) Modem ( )Analyzer network: Ethernet ( ) RS-485 Multi-Drop ( )Other options: Carrier Gas ( ) Calibration Gas ( ) Start-up ( ) Training Class ( )Special Instructions:
Page 11
Product DatasheetDAN-PDS-GC-700
OUTLINE AND MOUNTING DETAILS
SPECIFICATIONSPower:
Standard: 24 VDC, (24.5-26.5 VDC) Optional: 90-264 VAC, 47-63 Hz 80 watts start-up; 33 watts nominal
Environment: -29° to 60°C (-20° to 140°F) Enclosure Protection Rating: IP66 and NEMA 4XDimensions (without sample system):
Wall-mount: 914mm H x 462mm W x 483mm D (36” H x 18.2” W x 19” D) Pipe-mount: 914mm H x 462mm W x 645mm D (36” H x 18.2” W x 25.4” D) Floor-mount: 1483mm H x 419mm W x 577mm D (58.4” H x 16.5” W x 22.7” D)
Mounting: Free-standing (standard); wall-mount or pipe-mount (optional)Approximate Weight (without sample system): 80 kg (175 lbs.)Area Safety Certification Options:
CSA: Class 1, Division 1, Groups B, C, D, T4 USA: Class I, Zone 1, Ex d IIB+H2, T4, Enclosure Type 4 Canada: Class I, Zone 1, Ex d IIB+H2, T4, Enclosure Type 4ATEX: Ex II 2G Ex d IIC T4 Tamb = 60°C IECEx: Ex d IIC T4 Gb Tamb = 60°C
Oven: Airless heat sink, maximum 130°C (266°F)Valves: 6-port and 10-port diaphragm chromatograph valves. Other types of
valves may be used, such as liquid injection or rotary valves, depending on the application
Carrier Gas: Application-dependent. Typically zero-grade helium, nitrogen, or hydrogen
4“ OR 2” POLEMOUNT OPTION
25.4
22.4
26.25REMOVAL
CLEARANCETYPICAL
SAMPLE PLATE
16.7518.2
56.6
1.56
12.0
SAMPLE PLATE
4X Ø .50 THRU
36.8
16.5 22.1
58.4
SIDE VIEW FLOOR MOUNT UNITFRONT VIEW FLOOR MOUNT UNIT
FRONT VIEW WALL MOUNT UNIT
POLE MOUNT UNIT
22.715.50
13.6
33.8RECOMMENDEDFROM GROUND
OR FLOOR
Sample Input Pressure Range (recommended): 15-20 psigCarrier Gas Input Pressure Range (recommended): 90-100 psigDetector: Thermal conductivity detector (TCD), flame ionization detector
(FID), TCD/TCD or TCD/FID dual detector configurations possible; flame photometric detector (FPD) available (see FPD module data sheet)
Gating Options: Fixed-time automatic slope sense gating of peaksStreams: Up to eight streams (including calibration stream) internally
controlled; up to 12 streams externally controlledChromatograms stored/archived internally: Most recent analysis and/or
calibration per stream Analog Inputs: Four inputs filtered with transient protection (user scalable and
assignable)Analog Outputs: Four non-isolated analog outputs, 4 – 20 mA (standard); four
additional isolated analog outputs or eight additional non-isolated analog outputs (optional)
Serial Communication Ports: Three serial ports standard with option for a total of seven. Depending on the port, choice of RS-232, RS-422, and RS-485 is available as well as the Modbus protocol
Digital Inputs: Five digital inputs; user assignable Digital Outputs: One gas chromatograph alarm and four user-assignable
outputs (2 form A, 3 solid-state optically isolated)Parallel Printer Port: One parallel port available for printed reportsInternal Modem (optional): Field-configurable; 300 to 33.6k baud rateEthernet Card (optional): 10 mbps with RJ-45 portRAM (DATA) Memory Capacity: 128 Mb RAM (standard)
Page 12
Product DatasheetDAN-PDS-GC-700
Emerson Process ManagementDaniel Measurement and Control, Inc.
World Area HeadquartersHouston, Texas, USAT: 713-827-6380, F: 713-827-3865USA Toll Free [email protected]
Calgary, Alberta, Canada, T: 403-279-1879, F: 403-236-1337
Singapore - Asia Pacific, T: +65-6777-8211, F: +65-6770-8001
Stirling, Scotland - Europe, Mid-East, Africa, T: +44-1259-727-222, F: +44-1786-433-606
Daniel Measurement and Control, Inc. and Rosemount Analytical are a wholly owned subsidiaries of Emerson Electric Co., and are divisions of Emerson Process Management. The Daniel name and logo are registered trademarks of Daniel Industries, Inc. The Emerson logo is a registered trademark and service mark of Emerson Electric Co. All other trademarks are the property of their respective companies. The contents of this publication are presented for informational purposes only, and while every effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by Daniel’s terms and conditions, which are available upon request. We reserve the right to modify or improve the designs or specifications of such products at any time. Daniel does not assume responsibility for the selection, use or maintenance of any product. Responsibility for proper selection, use and maintenance of any Daniel product remains solely with the purchaser and end-user. Daniel products are protected by United States and international patents and patents pending.