ct_iq

Upload: bbkanil

Post on 04-Jun-2018

227 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/13/2019 CT_IQ

    1/44

    CT Image Quality

    Wil Reddinger, M.Sc., R.T.(R)(CT)copyright April 1998

    OutSource, Inc. (all rights reserved)

  • 8/13/2019 CT_IQ

    2/44

    CT IMAGE QUALITY

    Wil Reddinger, M.Sc., R.T.(R)(CT)copyright April 1998

    One of the most important functions of a computed

    tomography (CT) system is to reproduce a three-

    dimensional structure and represent that structure as an

    accurate two-dimensional cross-section on a television

    monitor. There are several characteristics that effect how

    well a CT system performs this task. Spatial resolution,

    contrast resolution, linearity, noise and artifacts are the

    primary characteristics that effect image quality in CT.Enhancing or suppressing any of these characteristics

    depends upon the imaging interests and the region of the

    body being scanned. It is important that to realize that

    changing CT parameters such as section thickness,

    algorithms and field of view have a profound effect on

  • 8/13/2019 CT_IQ

    3/44

    the overall quality of the CT image. CT image quality is

    dependant upon balancing these characteristics andparameters to produce the best possible image for the

    anatomical region being scanned. Image noise and

    artifacts are the two biggest enemies of CT image

    quality. CT parameters can be manipulated to either

    decrease or eliminate the adverse effects of these imagequality characteristics. Generally, there is a trade-off

    when CT parameters are manipulated. For example, if a

    bone reconstruction algorithm is utilized to increase

    spatial resolution, image noise increases which degrades

    contrast or soft tissue resolution. Increasing technical

    factors such as mAs or kVp decreases image noise but an

    increase in patient dose occurs. The CT technologist's

    goal is to manipulate CT parameters according the

    imaging interest or situation.

  • 8/13/2019 CT_IQ

    4/44

    Spatial resolution is the CT system's ability to

    differentiate small objects that are adjacent to oneanother. The CT scanner's resolving power relies on how

    well small objects that are close together but have very

    different attenuation values or CT numbers are imaged.

    The point between two small objects with very different

    densities is considered to be a region of high frequencyor high contrast. An inherent problem with any CT

    system is that the edges or boarders of these high

    frequencies are blurred to a certain degree. Therefore,

    instead of imaging these two small structures as separate

    entities, the CT system "sees" them as one structure. A

    CT system that can image high frequency regions

    without blur results in smaller objects being resolved.

    Spatial resolution in conventional radiography is superior

    compared to the spatial resolution of computed

    tomography. For example, when evaluating a radiograph

  • 8/13/2019 CT_IQ

    5/44

    of the facial bones, it is easy to see that blurring does not

    occur between the maxillae and the maxillary sinuscavity. The area where the edge of the bone meets the

    sinus is not blurred unless the patient moved during the

    exposure. A CT section of the same region reveals a

    "fuzzy" or "hazy" appearance around the edges of the

    bones. There are parameters that a CT technologist canmanipulate to increase the spatial resolution when

    scanning high frequency regions. Utilizing a bone, sharp,

    high frequency or high pass algorithm during

    reconstruction mathematically enhances the edges of

    structures by diminishing structural blurring. However,

    the added algorithm produces statistical interference,

    which results in an increase in image noise. The increase

    in image noise decreases contrast or soft tissue

    resolution. Generally, this is an expected outcome

    because the scan involved visualizing the bone and the

  • 8/13/2019 CT_IQ

    6/44

    soft tissue demonstrated is sacrificed. Utilizing a bone

    window, a wide window width and a high window level,is necessary to visually enhance the edges of a structure

    (Figure 1).

    Decreasing section thickness narrows the collimated

    beam that concentrates the photons on to a smaller area,

    which increases spatial resolution. Zooming or targeting

    requires the operator to manipulate the displayed field of

    view. Zooming or targeting takes a selected region and

    spreads it over the entire matrix that is used to create and

    display the CT image. A 512 X 512 matrix yields 262,

    144 pixels or "little squares" that the image is displayed

    upon. If an image an entire abdomen is imaged the

    matrix contains all of the abdominal structures which fill

    the 262,144 pixels. If the CT technologist wants to

  • 8/13/2019 CT_IQ

    7/44

    Figure 1 shows the effect of window width and level

    on image noise.

  • 8/13/2019 CT_IQ

    8/44

    increase the spatial resolution of an individual vertebra,

    the vertebra is "Zoomed" or "Targeted". The result isthat the vertebra fills the 262,144 pixels thereby

    increasing the imaging accuracy of the structure.

    Primarily, zooming or targeting decreases partial volume

    averaging of a region which increases the accuracy of the

    CT numbers in a region of interest measurement. Thereare 262,144 pixels used to display the image but zooming

    or targeting decreases their overall size, which decreases

    the possibility of many tissues occupying a single pixel.

    Zooming or Targeting is not the same as image

    magnification. Image magnification is usually achieved

    by turning a dial or trackball, which results in pixel

    stretching. Although, this technique may be visually

    appealing it does not change the amount of tissue

    occupying a single pixel. The ultimate situation would

    be to fill each individual pixel with one tissue type.

  • 8/13/2019 CT_IQ

    9/44

    However, computer and display matrices are limited in

    size so it is not likely to happen. Other factors thatinfluence spatial resolution include pixel size, which is

    influenced by the chosen, scanned field of view and

    matrix size, width of the detector, spacing between

    detectors, number of projections or views obtained and

    focal spot size.

    Several methods that quantify or measure spatial

    resolution include the point-spread function (PSF), line

    spread function (LSF), edge response function (ERF) and

    the modular transfer function (MTF). The MTF is the

    most commonly used method to measure the spatial

    resolution capabilities of a CT system, which is

    graphically depicted. The MTF demonstrates the

    frequency components of a structure in line pairs per

    centimeter (lp/cm). A line pair phantom consists of lead

  • 8/13/2019 CT_IQ

    10/44

    strips that are placed at different widths apart. The more

    lines that can be visualized separate from one another,the better the systems spatial resolution. Spatial

    resolution depends upon imaging high frequency

    structures located a very small distance apart. In figure

    ** an MTF value of 1.0 represents a complete, without

    blurring, transfer of an object through the CT system to amonitor. An MTF value of 1.0 corresponds to imaging

    large structures that can very easily be imaged accurately

    by most CT systems. Therefore, the value of 0.1, which

    represents small, high frequency or density structures, is

    used to evaluate the spatial resolution capabilities of a CT

    system. It is known that a CT system does not have

    problems faithfully imaging large structures such as the

    liver or brain. Because of the inherent blurring of small,

    high density structures like IAC's it is important that

    individual's purchasing CT systems should inquire how

  • 8/13/2019 CT_IQ

    11/44

  • 8/13/2019 CT_IQ

    12/44

    Figure 2: Modulation Transfer Function (MTF) Graph

  • 8/13/2019 CT_IQ

    13/44

    CT scanner to image these slight differences is known as

    low contrast delectability, which is used to describecontrast resolution in computed tomography. Essentially,

    contrast resolution describes the CT systems ability to

    discriminate between two or more anatomical structures

    that attenuate "nearly" the same amount x-ray photons.

    CT is superior compared to conventional radiography indiscriminating the absorption differences between

    structures that have similar but slightly different

    absorption characteristics. The liver, kidneys and psoas

    muscles are large structures that have different densities

    and atomic numbers but the absorption characteristics ofthe structures are not very different. Conventional

    radiography of the abdomen reveals the image structures

    as primarily "gray shaded" shadows and are some are

    superimposed. The position of the structures can be

    determined but the actual attenuation values cannot be

  • 8/13/2019 CT_IQ

    14/44

    distinguished from one another. Computed tomography

    not only eliminates the superimposition of structures but

    absorption differences can be determined. Absorption

    properties of any tissue are represented by linear

    attenuation coefficients, which are numbers that describe

    the absorption characteristics of the tissues imaged. It is

    important to remember that the linear attenuationcoefficient is an absorption measurement and it is

    dependant on thickness of a material, density of a

    material, atomic number and photon energy. A kVp

    change results in a change of a structures linear

    attenuation coefficient even if the same structure is beingimaged. This is critical because the CT computer

    eventually assigns each pixel that contains a linear

    attenuation coefficient a number known as a Hounsfield

    Unit (HU) or a CT number. A CT operator can utilize a

    region of interest (ROI) measurement to visually display

  • 8/13/2019 CT_IQ

    15/44

  • 8/13/2019 CT_IQ

    16/44

    Figure 3: Region of Interest (ROI) measurements

  • 8/13/2019 CT_IQ

    17/44

    section thickness is reduced the lead jaws of the

    collimator aperture closes which reduces the number of

    photons reaching the detector. The finely collimated

    beam is also filtered before it goes through the patient.

    These special filters called bow-tie filters absorb weaker

    energy photons, which reduce the possibility of scattered

    photons being detected by the CT detectors. An x-raybeam consists of polychromatic photons or photons

    having different energies. The bow-tie filters serve to

    absorb lower energy photons primarily to "lessen" the

    effects from the polychromatic nature of an x-ray beam.

    If the x-ray beam was not heavily filtered the accuracy ofCT numbers would be sacrificed because less desirable

    photon energies would probably scatter and be included

    in the attenuation value of the tissue scanned. The ideal

    situation would be to have monochromatic x-ray photon

    energies pass through the patient and into the detectors.

  • 8/13/2019 CT_IQ

    18/44

    This is a virtually impossible thing to do but heavy

    filtration and thin collimation of the x-ray beam makes

    the situation better. When section thickness is reduced

    the number of photons concentrated on the area of

    interest is reduced. Decreasing the number of photons

    results in the image becoming more "grainy". The

    "grainy" appearance of the section is due to image noise.For example, During a CT scan of the abdomen a

    technologist may be required to begin the procedure

    utilizing a 10 millimeter (mm) section thickness. The

    abdomen protocol may require the technologist to reduce

    the section thickness to 5mm if the pancreas is a specificarea of interest. When comparing the 10 mm and 5 mm

    sections, note that the 5 mm sections appear to be

    grainier in appearance than the 10 mm sections. This

    results from limiting the number of photons used to

    create the 5 mm sections compared to the 10 mm

  • 8/13/2019 CT_IQ

    19/44

    sections. Increasing the number of photons to the 5 mm

    sections may be accomplished by increasing mAs or

    kVp. However, the increase in technical factors

    increases patient dose.

    A soft tissue, standard or smooth algorithm is used

    during the reconstruction process to enhance soft tissueand contrast resolution. Soft tissue, standard or smooth

    algorithms are low-frequency or low pass algorithms.

    Low-frequency or low pass algorithms mathematically

    accentuate the soft tissues or where there are subtle

    attenuation differences between the tissues being imaged.

    The use of these algorithms requires a narrow window

    width and a low window level to properly be visualize on

    a television monitor. Detector systems must be sensitive

    to all the possible interactions between photons and

    imaged tissues. The number of levels of information that

  • 8/13/2019 CT_IQ

    20/44

    a detector can detect is called the dynamic range of a

    detector. The higher the dynamic range the better the

    detector will be able to see and discriminate between

    small differences in attenuation.

    Morgan (1983) defines linearity as " Property of a

    detector characterized by an output electrical current thatis exactly linearly proportional to the input radiation

    incident on the detector." Linearity is utilized to describe

    the performance of the CT system. Linearity describes

    the accuracy between the linear attenuation coefficient

    and the computer assigned CT number. The CT number

    scale assigns the linear attenuation coefficient of water

    the CT number zero. Bushong (1997) describes that the

    phantom utilized for the quality assurance of linearity of

    a CT system utilizes several materials placed in different

    positions throughout a water phantom. These materials

  • 8/13/2019 CT_IQ

    21/44

  • 8/13/2019 CT_IQ

    22/44

    contains no information. Noise is characterized by a

    grainy appearance of the image. Many authors describe

    noise as a salt and pepper pattern on the CT image. If a

    quality assurance phantom is comprised of a known

    material like water, when the phantom is used to evaluate

    the quality of the image, it is expected that every portion

    of that phantom would have the CT number zero. Due tothe statistical fluctuation in every scan it is impossible for

    this to occur. In the case of too little radiation, too few

    photons reach the detectors. As a result, the variance of

    CT numbers pixel to pixel is quite large. When a ROI is

    selected and the average CT number is displayed, thestandard deviation is quite large. The level of noise in an

    image is recorded as the standard deviation in an ROI

    measurement. The larger the standard deviation, the less

    accurate the average CT number of the ROI. An analogy

    would be that of asalary survey. If only 5 people were

  • 8/13/2019 CT_IQ

    23/44

    interviewed and their annual salary recorded, an average

    salary could be calculated. The variance, or deviation,

    between each salary listed would in all likelihood be

    quite large and the less the average would reflect any one

    of the five. On the other hand, if 100 people were

    interviewed, the deviation between each salary listed

    would not be as great as it was with just 5 people. Theaverage salary calculated would more accurately reflect

    the salary of the group(Figure 4).

    CT system manufacturers have minimum and maximum

    values for allowable differences in CT numbers of a

    water phantom. Generally the range of +3 to -3

    difference in CT numbers is negligible. Many of the

    spiral/helical scanners have a +4 to -4 allowable range

    because of the mathematical process of interpolation is

  • 8/13/2019 CT_IQ

    24/44

    Figure 4: Noise as a result of low radiation dose

  • 8/13/2019 CT_IQ

    25/44

    included with the image reconstruction process. The

    major types of noise include quantum noise, electronic

    noise and computational noise. Quantum noise is a result

    of too few photons reaching a detector after being

    attenuated by the body. Any factor that limits the

    number of attenuated photons at the detector will

    increase image noise. Anatomical structure size,reduction of slice thickness without increasing technical

    factors, decreasing pixel size and scatter radiation are all

    factors that contribute to image noise. Electronic noise is

    noise contained within the image that can be caused by

    vibrations of any of the physical components, especiallythe rotational components or power fluctuations.

    Computational noise is primarily caused by all the

    statistical fluctuations that occur from the reconstruction

    mathematics that are essential to produce a CT image. A

    CT image involves millions of mathematical equations

  • 8/13/2019 CT_IQ

    26/44

    being solved at approximately the same time.

    Spiral/helical CT data uses a mathematical technique

    called interpolation along with a reconstruction algorithm

    to produce axial image data from a volume of data. The

    process of adding a mathematical technique makes the

    computer work harder, which results in the CT image

    suffering by becoming grainier in appearance. It may be

    very difficult to actually distinguish which characteristic

    is responsible for the grainy appearance of an image but

    noise is always there.

    "An Artifact is any distortion or error in the image that is

    unrelated to the subject being studied (Morgan 1983)."

    Wolbarst (1993) describes artifacts as aberrations that

    arise at the interfaces of materials significantly different

    from the radiologic properties of the structures being

    scanned. There are many causes of artifacts that degrade

  • 8/13/2019 CT_IQ

    27/44

    image quality as well as hide areas of pathology.

    Artifacts can appear as geometrical inconsistencies,

    blurring, streaks or inaccurate CT numbers. Streak

    artifacts are the most common distortions or errors that

    affect the quality of CT images. Motion, metallic

    objects, out-of-field, edge gradient effects, high-low

    frequency interfaces, equipment malfunctions and

    sampling errors are all causes of streak artifacts.

    Motion artifacts occur primarily because during

    reconstruction the mathematical algorithm is unable to

    solve for the inconsistencies in attenuation. Imagereconstruction relies on the computer's ability to place

    attenuation values onto a matrix, which is nothing more

    than grid that has rows and columns of little blocks or

    pixels. If motion occurs during the scan the computer is

    unable to place an attenuation value at the proper

  • 8/13/2019 CT_IQ

    28/44

    corresponding "address" so it can only manipulate the

    data it is given. The result usually appears as blurring or

    streaking of the object being scanned.

    Metallic objects such as dental fillings or prosthesis

    cause a streaking effect on an image. The appearance

    may mimic a "splashing" effect from the middle of thestructure towards the edges of the object scanned. The

    primary reason that streaks occur from metal objects is

    because the objects exceed the attenuation values that CT

    system can faithfully image(Figure 5).

    Many older CT number scales assign the number 1000 as

    the top of the number scale. The number 1000 coincides

    with the attenuation value of cortical bone, which is

    primarily the densest structure in the human body. Many

  • 8/13/2019 CT_IQ

    29/44

  • 8/13/2019 CT_IQ

    30/44

    alloys used to make dental fillings or prosthesis have

    attenuation values greater than cortical bone. Therefore,

    the computer can only assign the highest value it knows.

    These metallic type objects exceed the dynamic range of

    the detectors in the detector array. The result is the

    metallic object is unable to be faithfully image resulting

    in streak artifacts. Many newer CT systems include

    attenuation values higher than bone. CT number scales

    have been expanded to include objects that have a CT

    number as high as 4,000.

    Out of field artifacts are caused by anatomy that is outside of the selected scan field of view. For example, if a

    persons chest measures 54 centimeters (cm) and the

    maximum scan field of view is 48cm, the CT system can

    only image 48 of the 54 cm. Unfortunately, The "extra"

    tissues are blocking detectors and attenuating photons.

  • 8/13/2019 CT_IQ

    31/44

    The CT system can only account for the 48 cm scan

    circle and is unable to avoid the blocked detectors.

    Streak artifacts occur throughout the entire image, which

    degrades diagnostic accuracy and the overall quality of

    the CT image(Figure 6).

    Edge gradient and high-low frequency interface aresimilar causes of streak artifacts present on an image.

    Edge gradient streak artifacts generally occur when the

    edges of a "sharp" high-density object interface with a

    smooth surface. A streak artifact originates from the

    edges of the high frequency structure. Edge gradientartifacts frequently occur from the edges between bone

    and soft tissue. For example, on a CT image of the pelvis

    the ischial spines are sharp high frequency structure that

    interface with adjacent muscle tissue, which is a low

    frequency structure. A thin black streak artifact arises

  • 8/13/2019 CT_IQ

    32/44

    Figure 6: Streaking caused by tissue being outside the scan

    circle (scan FOV).

  • 8/13/2019 CT_IQ

    33/44

    from the edge of the bone. The streak emanating from a

    thin biopsy needle is an example of edge gradient streak

    artifacts. Streak artifacts arise from materials or

    structures when a high-density material such as barium

    interfaces with a low-density material such as air. This

    type of streak artifact commonly occurs in the

    gastrointestinal tract is filled with barium and air. The

    artifact arises due to the inability of the CT system to

    "transfer" the information accurately. In the previous

    discussion of spatial resolution, it was explained that CT

    systems inherently have problems with imaging

    structures that rapidly change from high to low densities(Figure 7).

    Equipment malfunctions such as tube arching, electrical

    malfunctions and detector malfunctions produce streak

    artifacts on a CT image. Tube arching artifacts give rise

  • 8/13/2019 CT_IQ

    34/44

    Figure 7: Steaks due to dense contrast material in the bowel.

  • 8/13/2019 CT_IQ

    35/44

    to many streaks and may appear to look like a "lightening

    storm". Streak artifacts that arise from malfunctioning

    detectors generally appear as straight line streaks. Edge

    gradient streak artifacts are contained within the

    anatomical part and are not always straight line streak

    artifacts. A malfunctioning detector causes a straight

    black line streak artifact on a scout image. Streak

    artifacts may also arise if too few views or projections of

    an object are obtained. These types of streak artifacts

    have been virtually eliminated because CT scanners are

    capable of producing hundreds of thousands of views or

    projections. The more views, or projections, "taken" ofan object, the more accurate the image reconstruction.

    Beam hardening artifacts occur when the low energy

    photons are absorbed leaving the high-energy photons to

    strike the detectors. When x-ray photons strike a high-

  • 8/13/2019 CT_IQ

    36/44

    density structure such as bone, lower energy photons are

    absorbed which results in further hardening of the beam.

    Hardening the beam results in a more penetrating beam.

    If the photons then traverse over a region that consists of

    low densities structures such as the brain, the resultant

    effect is a thick streak artifact that appears across the

    region. This occurs because the effective energy of the

    x-ray beam is increased when it passes through the

    tissues being scanned. When the energy of the beam is

    changed during the scanning process, the linear

    attenuation coefficients of the tissue that the "harder"

    beam travels across are changed. If an ROI measurementwere taken in the area of the dark streak one would find

    that the CT numbers would be lower in that region. For

    example, when imaging the posterior fossa, beam-

    hardening artifacts are found in the area between the

    bone and soft tissue. If an ROI measurement were

  • 8/13/2019 CT_IQ

    37/44

  • 8/13/2019 CT_IQ

    38/44

  • 8/13/2019 CT_IQ

    39/44

    occurs due to one or several misaligned or miscalibrated

    detectors in the detector array of the rotate-rotate CT

    system. The tube and detector array rotate at the same

    time. The tube and detectors are physically connected to

    one another. If the tube shifts slightly it can cause

    misalignment of the CT system. Therefore, a "bad" view

    is created for every photon that the misaligned or

    miscalibrated detector or detectors "sees". The result is

    non-uniform information becoming a part of the image

    that manifests as a series of rings. The detector or

    detectors must be re-aligned or re-calibrated to eliminate

    ring artifacts(Figure 9)

    .

    The partial volume effect or volume averaging is when

    two or more different tissue types occupy the same pixel

    and are averaged together. The ultimate scenario for CT

    would be for one tissue type to occupy one pixel. Due to

  • 8/13/2019 CT_IQ

    40/44

    Figure 9: Ring artifact(s)

  • 8/13/2019 CT_IQ

    41/44

    the limitations of CT matrices, which determine how

    many pixels, makes up a CT image this is virtually

    impossible to accomplish. For example, if one pixel

    contained three separate tissues that had CT numbers of

    60, 80 and 200, the ROI measurement of that pixel would

    be approximately 113. The tissues are averaged which

    produces a number that is inconsistent with the three

    tissues that were evaluated. Partial volume averaging is

    always present and can never be entirely eliminated.

    Utilizing smaller section thicknesses or smaller displayed

    field of views may increase the accuracy of CT numbers.

    CT image quality is dependant upon the balancing of

    parameters relative to image resolution. Balancing image

    quality by manipulating and altering CT parameters

    depends upon the region or the condition being scanned

    with respect for patient dose. Image quality is also

  • 8/13/2019 CT_IQ

    42/44

    dependant upon limiting or eliminating the image

    degrading effects of noise and artifacts.

  • 8/13/2019 CT_IQ

    43/44

    Reference List

    Berland, Lincoln L. (1987). Practical CT Technology and

    Techniques. New York: Raven Press.

    Bushberg, Jerrold T., Seibert, J. Anthony, Leidholdt, Edwin M.,

    and Boone, John M. (1994). The Essential Physics of Medical

    Imaging. St. Baltimore: Williams & Wilkins.

    Bushong, Stewart. (1993). Radiologic Science for Technologists.

    5th edition. St. Louis: CV Mosby Publishers.

    Behrman, Richard H. Editor. (1994) Study Guide to Computed

    Tomography Advanced Applications. Greenwich, Connecticut:

    Clinical Communications Inc.

    Curry, Thomas S, Dowdey, James E., and Murry, Robert C.

    (1990). Christensen's Physics of Diagnostic Radiology. 4th edition.

    Philadelphia: Lea & Febiger.

    Morgan, Carlisle L. (1983). Basic Principles of Computed

    Tomography. Baltimore: University Park Pres

  • 8/13/2019 CT_IQ

    44/44

    Scroggins,D. , Reddinger,W. , Carlton,R., & Shappell,A. (1995).

    Computed Tomography Review. Philadelphia: J.B. Lippincott

    Seeram, Euclid. (1994). Computed Tomography: Physical

    Principles, Clinical Applications and Quality Assurance.

    Philadelphia: W.B. Saunders Company.

    Wolbarst, Anthony B. (1993). Physics of Radiology. St.

    Norwalk, CT: Appleton & Lange.