cost emf - med (action bm1309): european network for...

4
C European network f Safety of a Researcher: Elena Lucano. email: lu Home Institution: Department of In University of Rome, Italy. Contact: M Host Institution: IT`IS foundation, Z STSM Reference: COST-STSM-BM13 STSM dates: FROM19 th October 201 Abstract: Magnetic Resonance Imaging ( practice. One of the clinical haza to the radiofrequency (RF) field deposited power at the implant such as: magnitude and phase o goal for this STSM, the power exposures was evaluated using t A. Purpose of the STSM Traditionally, RF-induced heat temperature measurements in been used increasingly to comp in patients with medical devices the ability of elongated conduc locally deposit it in the tissue ne is used to calculate the electrom ISO 10974 describes a methodo purposes of the work planned f methods for MRI-RF-induced im Determine the applicability of t (Edition 2), for 3 T evaluations; 3 evaluation of implants during M diversified incident field charact conductive wire).More details ab B. Work Description 1) As a first step of this experien systems was gained in the lab. T RF field at 64 MHz and 128 MH system was used to excite the systems are part of the radiated 2) As a part of my experienc phantoms, such as the ASTM a COST EMF - MED (Action BM1309): for innovative uses of EMFs in biomedical ap STSM Report: active implantable medical devices in 3T MRI [email protected] nformation Engineering, Electronics and Telecommun Micaela Liberti email: [email protected] Zürich(CH). Contact: Niels Kuster email: [email protected] 309-29736 16TO15 th January 2016 (MRI) is a radiological imaging technique wid ards related to the MRI is the tissue heating effec d in a patient with an active implantable medica t electrodes is highly dependent on the incident of the electric field tangential to the conductive w deposition of generic and clinical implants und the transfer function approach together with test ting has been evaluated by experimental gel-filled phantoms. In recent years, computa plement the experimental results to evaluate the s undergoing MRI. The clinical hazard of AIMD in ctive structures to pick up the RF energy during ear the electrodes. As part of the evaluation, com magnetic (EM) field generated by the RF coil used ology for numerical assessment of in vivo exposu for the STSM were: 1) Gain insight and gather k mplant heating characterization, modeling, and the four-tier method, currently defined for 1.5 3) design a novel experimental testing method fo MRI at 1.5 T and 3.0 T, by means of power depo teristics (magnitude and phase of the electric fie bout the experience will follow. nce proficiency in using different measurements The exposure systems (ZMT, Zurich Switzerland) Hz were the MITS1.5 and the MITS3.0, respect e implant under test and measure its transfer d testing setup of AIMD for safety evaluation. ce the MITS1.5 and MITS3.0 were used to ch and the ellipsoidal phantom. Measurements of pplications nications, “Sapienza”, thz.ch dely used in clinical ct that can occur due al device (AIMD).The t field characteristics wire. As a part of the der more than 1000 ting in silico. methods, such as ational modeling has e RF-induced heating n MRI originates from g MRI exposure and mputational modeling d during MRI. The TS ure of the AIMD. The knowledge of current model validation; 2) T by the ISO 10974 or safety assessment osition evaluation for eld, tangential to the s setup and exposure used to produce the tively. While the piX r function. All these haracterize different the electromagnetic

Upload: others

Post on 06-Oct-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: COST EMF - MED (Action BM1309): European network for …cost-emf-med.eu/wp-content/uploads/2016/03/STSM-Report... · 2016. 3. 4. · COST EMF European network for innovative uses

COST EMF

European network for innovative uses of EMFs in

Safety of active implantable medical devices in 3T MRI

Researcher: Elena Lucano. email: [email protected]

Home Institution: Department of Information Engineering, Electronics and Telecommunications, “Sapienza”,

University of Rome, Italy. Contact: Micaela

Host Institution: IT`IS foundation, Zürich(CH).

STSM Reference: COST-STSM-BM1309

STSM dates: FROM19th

October 2016

Abstract:

Magnetic Resonance Imaging (MRI) is a radiological imaging technique wpractice. One of the clinical hazards related to the MRI is the tissue heating effect that can occur due to the radiofrequency (RF) field in a patient with an active implantdeposited power at the implant electrodessuch as: magnitude and phase of the electric fieldgoal for this STSM, the power deposition of genericexposures was evaluated using the transfer function approach together with testing

A. Purpose of the STSM

Traditionally, RF-induced heating has been evaluated bytemperature measurements in gelbeen used increasingly to complement the experimental resultsin patients with medical devices undergoing MRI. The clinical hazard of AIMD in MRI originates from the ability of elongated conductive structures to pick up the RF energy duringlocally deposit it in the tissue near theis used to calculate the electromagnetic (EM) field generated by the RF coil used during MRI. The TS ISO 10974 describes a methodology for numerical assessment of purposes of the work planned for the STSM were: 1) methods for MRI-RF-induced implant heating characterization, modeling, and model validation; 2Determine the applicability of the four(Edition 2), for 3 T evaluations; 3) design a novel experimental testing method for safety assessment evaluation of implants during MRIdiversified incident field characteristics (magnitude and phase of the electric field, tangential to the conductive wire).More details about the experience will follow.

B. Work Description

1) As a first step of this experience proficiency in using different measurements setusystems was gained in the lab. The exposure systemsRF field at 64 MHz and 128 MHz system was used to excite the implant under test andsystems are part of the radiated testing

2) As a part of my experience the MITS1.5 and MITS3.0 were used to characterize different phantoms, such as the ASTM and the ellipsoidal phantom. Measurements of the electromagnetic

COST EMF - MED (Action BM1309):

for innovative uses of EMFs in biomedical applications

STSM Report:

Safety of active implantable medical devices in 3T MRI

[email protected]

Department of Information Engineering, Electronics and Telecommunications, “Sapienza”,

University of Rome, Italy. Contact: Micaela Liberti email: [email protected]

IT`IS foundation, Zürich(CH). Contact: Niels Kuster email: [email protected]

BM1309-29736

October 2016TO15th

January 2016

Resonance Imaging (MRI) is a radiological imaging technique widely used in clinical of the clinical hazards related to the MRI is the tissue heating effect that can occur due

radiofrequency (RF) field in a patient with an active implantable medical device (AIMD).deposited power at the implant electrodes is highly dependent on the incident

of the electric field tangential to the conductive wireower deposition of generic and clinical implants under more

exposures was evaluated using the transfer function approach together with testing

induced heating has been evaluated by experimental methods, such as temperature measurements in gel-filled phantoms. In recent years, computational modeling has been used increasingly to complement the experimental results to evaluate the in patients with medical devices undergoing MRI. The clinical hazard of AIMD in MRI originates from the ability of elongated conductive structures to pick up the RF energy duringlocally deposit it in the tissue near the electrodes. As part of the evaluation, computational modeling is used to calculate the electromagnetic (EM) field generated by the RF coil used during MRI. The TS ISO 10974 describes a methodology for numerical assessment of in vivo exposure of the AIMD. purposes of the work planned for the STSM were: 1) Gain insight and gather knowledge of current

induced implant heating characterization, modeling, and model validation; 2Determine the applicability of the four-tier method, currently defined for 1.5 T by the ISO 10974 Edition 2), for 3 T evaluations; 3) design a novel experimental testing method for safety assessment

evaluation of implants during MRI at 1.5 T and 3.0 T, by means of power deposition evaluation for t field characteristics (magnitude and phase of the electric field, tangential to the

.More details about the experience will follow.

As a first step of this experience proficiency in using different measurements setusystems was gained in the lab. The exposure systems (ZMT, Zurich Switzerland)

and 128 MHz were the MITS1.5 and the MITS3.0, respectivelyexcite the implant under test and measure its transfer function

adiated testing setup of AIMD for safety evaluation.

my experience the MITS1.5 and MITS3.0 were used to characterize different ch as the ASTM and the ellipsoidal phantom. Measurements of the electromagnetic

pplications

Department of Information Engineering, Electronics and Telecommunications, “Sapienza”,

[email protected]

idely used in clinical of the clinical hazards related to the MRI is the tissue heating effect that can occur due

medical device (AIMD).The incident field characteristics

the conductive wire. As a part of the under more than 1000

exposures was evaluated using the transfer function approach together with testing in silico.

experimental methods, such as filled phantoms. In recent years, computational modeling has

to evaluate the RF-induced heating in patients with medical devices undergoing MRI. The clinical hazard of AIMD in MRI originates from the ability of elongated conductive structures to pick up the RF energy during MRI exposure and

As part of the evaluation, computational modeling is used to calculate the electromagnetic (EM) field generated by the RF coil used during MRI. The TS

exposure of the AIMD. The Gain insight and gather knowledge of current

induced implant heating characterization, modeling, and model validation; 2) ly defined for 1.5 T by the ISO 10974

Edition 2), for 3 T evaluations; 3) design a novel experimental testing method for safety assessment power deposition evaluation for

t field characteristics (magnitude and phase of the electric field, tangential to the

As a first step of this experience proficiency in using different measurements setup and exposure used to produce the

, respectively. While the piX transfer function. All these

my experience the MITS1.5 and MITS3.0 were used to characterize different ch as the ASTM and the ellipsoidal phantom. Measurements of the electromagnetic

Page 2: COST EMF - MED (Action BM1309): European network for …cost-emf-med.eu/wp-content/uploads/2016/03/STSM-Report... · 2016. 3. 4. · COST EMF European network for innovative uses

STSM Report

COST EMF - MED (Action BM1309):European

field were collected in several sessions(DASY52, SPEAG, Zurich Switzerland).of several implants used for research and clinical purpose (i.e. 3 generic straight leads, 3 generic coiled leads, more than 6 leads for cardiac stimulation, more than 6stimulation), the work was performed at 64

3) The main contribution of the experience was method for safety assessment of implant during MRI exposure. In fact tsystems used in the standards elongated devices (>80 cm) cannot currently beAIMD, the research project wasdiversification of induced electric field inside adjustment of the RF birdcage driving vectorpolarizations into the RF coil system. The approach (MITS 1.5, ZurichMedTech AG, Zurich, Switzerland). The feasibility test 128 MHz system.

magnetic field polarization was characterizedexposures, defined by the relative magnitude and phase of the two feeding results, 10 out of 120 polarizatiselected polarization dataset, a more refinewas performed, with the extension to other 10 measurements pointsphantom. For the measured points indicated in figure 2a a 360probe was performed, obtaining the information about the polarization of the field in the selected point (figure 2b). The polarization data returned informaticomponents (i.e., blue for x, red for y and green for z in figure 2b) and total magnitude (i.e., yellow in figure 2b) of the magnetic field. The procedure was repeated for the 9 selected points and for the 10 selected polarization settings, for a total of 90 field polariza

A full-wave FDTD simulations using Sim4Life (ZMT, Zurich Switzerland)numerical replica of the RF system loaded with the PP. The

Figure 1: a) Measurement setup: RF coil loaded with

the Phase Phantom (PP) in b)

a)

Figure 2: a) measurement points selected inside the PP; b) magnetic field polarization for one of the selected

point in (a). The polarization was obtain by performing a

850 mm

740 mm

a) b)

Elena Lucano

European network for innovative uses of EMFs in

several sessions using the dosimetric and near-field assessment system(DASY52, SPEAG, Zurich Switzerland).The piX system was used to characterize the tof several implants used for research and clinical purpose (i.e. 3 generic straight leads, 3 generic coiled leads, more than 6 leads for cardiac stimulation, more than 6 leads for spinal cord

, the work was performed at 64 MHz and 128 MHz.

The main contribution of the experience was the development of a novel experimental testing method for safety assessment of implant during MRI exposure. In fact the current measurement

used in the standards are subjected to several limitations and testing of extremely cannot currently be performed. In the context of the safety study for

was focused on a novel radiated testing method by utilizing induced electric field inside a RF birdcage coil system through careful control and

adjustment of the RF birdcage driving vector. The field diversity was produced imposing different system. The approach was performed in the 64

(MITS 1.5, ZurichMedTech AG, Zurich, Switzerland). The feasibility test is to

The work was conducted for generic leads and for clinical leads used in cardiac and spinal cord stimulation. An innovative phantom named Phase Phantom (PP) was used as part of the testing procedure (Figure 1)implemented to guarantee controlledconditions for the lead under test, in terms of magnitude and phase of the induced electric field, as well as to accommodate long leadeddevices while minimizing reflections and perturbation to the device under test (DUT)first step for the field diversity evaluation

field polarization was characterized at the isocenter of the RF system relative magnitude and phase of the two feeding

10 out of 120 polarizations were selected for testing of induced electric fielda more refined incident magnetic field characterization

was performed, with the extension to other 10 measurements points (Figure 2For the measured points indicated in figure 2a a 360

ospatial rotation of the magnetic field

obtaining the information about the polarization of the field in the selected point (figure 2b). The polarization data returned information about the spatial profile of the three components (i.e., blue for x, red for y and green for z in figure 2b) and total magnitude (i.e., yellow in figure 2b) of the magnetic field. The procedure was repeated for the 9 selected points and for the 10

ted polarization settings, for a total of 90 field polarization measurements.

wave FDTD simulations using Sim4Life (ZMT, Zurich Switzerland) was computed using a numerical replica of the RF system loaded with the PP. The obtained numerical data

a) Measurement setup: RF coil loaded with

b)

Figure 2: a) measurement points selected inside the PP; b) magnetic field polarization for one of the selected

point in (a). The polarization was obtain by performing a 360o spatial rotation of the magnetic field probe.

235 mm

440 mm

P a g e | 2

for innovative uses of EMFs in biomedical applications

field assessment system piX system was used to characterize the transfer function

of several implants used for research and clinical purpose (i.e. 3 generic straight leads, 3 generic leads for spinal cord

the development of a novel experimental testing he current measurement

limitations and testing of extremely the context of the safety study for

a novel radiated testing method by utilizing the through careful control and

produced imposing different MHz RF coil system be repeated in the

The work was conducted for generic leads and cardiac and spinal cord

An innovative phantom named Phase Phantom (PP) was used as part of the

(Figure 1). The PP was implemented to guarantee controlled exposure conditions for the lead under test, in terms of magnitude and phase of the induced electric

, as well as to accommodate long leaded-devices while minimizing reflections and perturbation to the device under test (DUT). As

ield diversity evaluation, the at the isocenter of the RF system for 120 different

relative magnitude and phase of the two feeding ports. Based on the induced electric field. For the

incident magnetic field characterization inside the PP (Figure 2a) inside the

spatial rotation of the magnetic field obtaining the information about the polarization of the field in the selected

on about the spatial profile of the three components (i.e., blue for x, red for y and green for z in figure 2b) and total magnitude (i.e., yellow in figure 2b) of the magnetic field. The procedure was repeated for the 9 selected points and for the 10

was computed using a numerical data was used to

Figure 2: a) measurement points selected inside the PP; b) magnetic field polarization for one of the selected

spatial rotation of the magnetic field probe.

Page 3: COST EMF - MED (Action BM1309): European network for …cost-emf-med.eu/wp-content/uploads/2016/03/STSM-Report... · 2016. 3. 4. · COST EMF European network for innovative uses

STSM Report Elena Lucano P a g e | 3

COST EMF - MED (Action BM1309):European network for innovative uses of EMFs in biomedical applications

extract the tangential component of the electric field along the implant routing paths inside the PP for the calculation of the power deposition at tip of implant. The different polarization settings were numerically obtained by imposing for each measured point the simulated polarization to be equal to the measured one. For the purpose of validating the numerical data, the magnetic field mapping inside a 300 x 300 mm

2 area was performed for each polarization setting, and data were compared.

Once the coil was characterized in terms of field polarization, the power deposition at the tip of an implant was evaluated for each polarization setup both numerically and through measurements. On the measurements side, a surface area of 7x80mm

2 was measured 1 mm on top of the tip lead for

each polarization setup. On the simulation side, the electric field tangential to the lead trajectory was evaluated from the numerical data, calibrated for each polarization setup. And the power at the tip of the implant was calculated based on the transfer function approach, by the formula:

� = �� ℎ(�)��(�) ��� ∙ �� ℎ(�)��(�)

���∗

whereℎ is the piX model(transfer function) of the studied lead, and tan is the tangential component of the local incident electric field. Both quantities are defined along length l for length unit dl. The πX model was measured at both 64 and 128 MHz using the piX system available in the research laboratories.

C. Results

a)

b)

c)

Figure 3: a) field polarization comparison between

measurements and simulations for one of

polarization setup tested; measured (b) and

simulated (c) magnetic field map inside the Phase

Phantom for one of the polarization settings tested.

The percentage error between measurements and

simulation for the selected result was 1.57 %.

Page 4: COST EMF - MED (Action BM1309): European network for …cost-emf-med.eu/wp-content/uploads/2016/03/STSM-Report... · 2016. 3. 4. · COST EMF European network for innovative uses

STSM Report Elena Lucano P a g e | 4

COST EMF - MED (Action BM1309):European network for innovative uses of EMFs in biomedical applications

dB Point1 Point2 Point3 Point4 Point5 Point6 Point7 Point8 Point9

Pol1 1.620 1.626 1.621 1.629 1.534 1.533 1.627 1.622 1.541 Pol2 1.026 1.041 1.029 1.044 1.032 1.013 1.023 1.129 1.102 Pol3 1.448 1.441 1.445 1.460 1.443 1.459 1.449 1.439 1.446 Pol4 1.274 1.260 1.270 1.282 1.278 1.358 1.352 1.260 1.337 Pol5 1.268 1.275 1.269 1.269 1.268 1.262 1.266 1.277 1.269 Pol6 1.005 0.981 1.002 0.931 1.018 0.924 1.013 0.976 0.998 Pol7 0.962 0.959 0.962 0.961 0.962 0.955 0.961 0.959 1.031 Pol8 0.627 0.544 0.625 0.613 0.621 0.643 0.632 0.602 0.623 Pol9 0.481 0.476 0.478 0.506 0.529 0.506 0.454 0.603 0.478

Pol10 0.518 0.522 0.515 0.500 0.537 0.515 0.534 0.521 0.590

Table 1: field percentage error of simulated dataset with respect to the measured plane for the selected 10 polarizations (in rows) imposing the measured polarization to the different 9 points (columns). For each polarization deviation within points was less than 4%.

Comparison of field polarization distribution inside the phantom showed that a small amount of field distortion can occur in different point of the phantom. This is due to interaction of the field with the phantom and to physical limits of any exposure system. This is not to be considered a problem if similar distortion can easily be predicted by simulated data. The numerical dataset was indeed able to replicate the polarization distribution of the measurements with an overall error under 2 dB (as reported in Table 1). Hence it was verified that it is possible to measure only one point inside the phantom to be account for the correct field polarization in the numerical data. This result was considered very important in the contest of safety assessment of an AIMD. Validation of numerical results is in fact fundamental allowing for trustable numerical evolutions of clinical scenarios that cannot be accounted for experimentally. Figure 3 reports an example of polarization comparison between measurements and simulations, with the relative magnetic field distribution in the 300 x 300 mm

2 area inside the phantom. The comparison was performed for all the 10 polarizations

selected, imposing to the numerical dataset the measured polarization in the 9 locations of the phantom (Figure 2). For each comparison the three field component and the magnitude were considered. Hence 27 magnetic field distributions were compared for each polarization setting for a total of 270 field distribution.

Once the numerical data were sufficiently validated against measurements, the transfer function obtained by the piX system was used to numerically evaluate the power at the tip of an implant for all the different polarization settings (i.e. more than 1000).Analyses for the different polarizations at 64 MHzresulted in a dynamic range of 10 dB in the deposited power of the implant.

D. Future collaboration with host institution

Analysis at 3 T needs to be finalized. Further investigation shall focus on the experimental implementation of the method. If proven practical, the proposed method is in compliance with the radiated immunity tests required in ASTM 2182 [2] and ISO/TS 10974 [3] and may improve the safety assessment of medical implants.

E. Expected Publications

Preliminary results of this work will be submitted in the form of a conference paper for the BioEM meeting that will be hold in Gant (Belgium) in June 2016. A complete work is expected to be submitted to a peer review journal in the next months.

Confirmation by the host institution of the successful execution of the STSM:

We confirm that Elena Lucano has performed the research work as described above.

NielsKuster

Micaela Liberti

Elena Lucano