corneal permeability in a redesigned corneal holder for the bovine cornea opacity and permeability...
TRANSCRIPT
Toxicology in Vitro 18 (2004) 853–857
www.elsevier.com/locate/toxinvit
Corneal permeability in a redesigned corneal holder for thebovine cornea opacity and permeability assay
J.L. Ubels a,b,*, J.A. Ditlev a, D.P. Clousing a, P.L. Casterton c
a Department of Biology, Calvin College, 3201 Burton St., SE, Grand Rapids, MI 49546, USAb Department of Ophthalmology, Wayne State University School of Medicine, Detroit, MI 48201, USA
c 3520 Vinewood SE, Grand Rapids, MI 49546, USA
Received 1 March 2004; accepted 9 April 2004
Abstract
The bovine cornea opacity and permeability assay (BCOP) is a proposed alternative to the Draize rabbit test for potential eye
irritants. In the standard BCOP, bovine corneas are mounted in a holder on a flat surface between two identical chambers. The flat
configuration of the standard holder does not conform to the normal curved shape of the bovine cornea and it comes into direct
contact with the cornea tissue. Mounting corneas in this holder causes extensive damage to both epithelial and endothelial corneal
cell layers. Our laboratory has designed a new holder that allows the cornea to maintain its natural curvature and does not damage
the cornea. Previous tests, using both the new and standard holders, and comparing corneal opacity, hydration and endothelial
morphology, have shown that the new holder is a significant improvement over the standard holder. The present study extends the
comparisons of the new and standard holders to measurement of corneal fluorescein permeability. The permeability (ng/cm2/min) of
intact corneas, corneas with no epithelium, and corneas treated with 1% NaOH, isopropanol, acetone, 30% trichloroacetic acid or
30% sodium dodecysulfate for either 1 or 10 min was determined by measuring fluorescence of samples taken from the endothelial
chamber after 90 min epithelial exposure to 0.04% sodium fluorescein. In all trials, the redesigned holders yielded not only lower
permeability measurements but also decreased measurement variability. The data provide further evidence that the new holder is an
improvement over the standard holder and should be incorporated into a new protocol for the BCOP.
� 2004 Elsevier Ltd. All rights reserved.
Keywords: Cornea; Ocular toxicology; Bovine cornea opacity and permeability assay
1. Introduction
The bovine cornea opacity and permeability assay
(BCOP), proposed by Gautheron et al. (1992), is an in
vitro method for testing the effects of potential irritants
on the cornea of a bovine eye. Earlier studies showed
that results obtained from the BCOP assay correlatewell with data from the Draize test (Gautheron et al.,
1994; Sina, 1994), but, as with all in vitro eye irritation
tests, the BCOP assay has not yet achieved regulatory
acceptance as an alternative to animal testing. Valida-
tion of this assay as predictive of human response to
irritants requires that the test be performed under
*Corresponding author. Tel.: +1-616-526-6219; fax: +1-616-526-
6501.
E-mail address: [email protected] (J.L. Ubels).
0887-2333/$ - see front matter � 2004 Elsevier Ltd. All rights reserved.
doi:10.1016/j.tiv.2004.04.005
optimal conditions, including the use of tissue that is in
excellent condition.
Since the development of the BCOP assay, our lab-
oratories have proposed several assay modifications.
Casterton et al. (1996) proposed that a spectropho-
tometer can be used for opacity measurement. By
determining corneal light absorbance at 570 nm, moreprecise measurements of opacity can be obtained. We
have also proposed that measurement of corneal
hydration and evaluation of the corneal endothelium be
added to the protocol. Ubels et al. (2000, 2002) showed
that the standard holders in which the bovine corneas
are mounted cause excessive damage to the delicate
corneal endothelial layer because the holder flattens and
wrinkles the cornea. The standard holders also causecorneal edge damage because the holder clamps directly
onto the cornea. To address these problems we devel-
oped a new, redesigned holder that does not contact the
854 J.L. Ubels et al. / Toxicology in Vitro 18 (2004) 853–857
cornea and allows it to maintain its normal shape. We
have previously tested this holder against the standard
holder, measuring both opacity and hydration as well as
endothelial morphology (Ubels et al., 2002). The resultsof these studies indicated that the redesigned holder did
not damage the cornea during the mounting process and
that corneal opacity and hydration were similar in the
two types of holder after any given treatment.
The present study is a continuation of our evaluation
of the new holder. We have now measured the perme-
ability of the cornea to Na-fluorescein under control
conditions and following exposure to irritants in boththe standard and redesigned holders. Because the cornea
is not damaged in the new holder, it was expected that
for any condition, permeability would be lower when
using the new holder compared to the standard holder.
2. Materials and methods
2.1. Corneas
Corneas were excised from slaughterhouse-derived
bovine eyes within 2 h of enucleation. Those that were
obviously damaged or that stained with fluorescein werediscarded. Corneas with absorbance at 570 nm of
greater than 0.1 after mounting in the holders and 15
min of pre-incubation were not used in experiments.
2.2. Corneal holders––volume and permeability measure-
ments
Both the standard and redesigned corneal holders
used in this study have been previously described (Ubels
et al., 2002, 2003). The new holder is illustrated in Fig. 1.
In the standard holder, the cornea is clamped between
two chambers of approximately equal volume and hasan exposed surface area of 2.26 cm2. In the new holder
the volume of the posterior (or endothelial) chamber is
Fig. 1. The new corneal holder for the BCOP assay. The cornea with a
5 mm scleral rim is mounted with the endothelial side of the sclera
contacting the O-ring of the half-chamber on the left. The corneal
epithelium faces the half-chamber on the right. The chamber does not
make contact with the cornea.
larger than that of the anterior (or epithelial) chamber
and the exposed area of the cornea is 10.04 cm2.
In the standard BCOP protocol, corneal permeability
is determined by measuring optical density at 490 nm ofthe solution in the endothelial chamber after a 90 min
incubation with fluorescein in the epithelial chamber
(Gautheron et al., 1992; Casterton et al., 1996). This OD
reading is used directly in calculating the in vitro score.
Because of the differences in holder dimensions, this
method could not be used in comparing the standard
and redesigned chambers. Alternatively, we calculated
an actual permeability rate in ng fluorescein/cm2 ofcornea/min during the 90 min incubation period. To do
this it was necessary to know the volumes of the
chambers. This was calculated by mounting a cornea in
the chamber, introducing a known amount of methylene
blue into each side and calculating the chamber volumes
according to the dye dilution principle. To measure the
total amount of fluorescein crossing the cornea, fluo-
rescence of samples taken from the endothelial chamberwas measured in a Packard FluoroCount Microplate
Fluorometer (Packard Instruments, Meridian, CT) and
the fluorescein concentration was determined from a
standard curve.
2.3. Experimental protocol
Intact corneas or corneas with epithelium removed by
scraping with a Gill knife were mounted in the standard
and redesigned holders. Endothelial chambers were fil-
led with minimum essential medium without phenol red
(MEM, Sigma). Na-fluorescein (Sigma, St. Louis, MO)
dissolved in MEM at 0.04% was placed in the epithelialchamber and the corneas were incubated for 90 min at
35 �C. The contents of the endothelial chamber were
mixed well, fluorescence of a sample from the endothe-
lial chamber was measured and fluorescein permeability
was calculated.
To compare effects of irritants on cornea permeability
between the two chambers, corneas were mounted and
pre-incubated for 15 min in MEM. The corneal epithe-lium was then exposed to a test substance for 1 or 10
min. Following exposure, the test material was removed,
the epithelial surface was rinsed with balanced salt
solution to remove the test substance and the corneas
were incubated with the fluorescein solution for 90 min,
followed by measurement of fluorescein levels in the
endothelial chamber. Effects of acetone, isopropyl
alcohol (IPA), 1% NaOH, 30% trichloroacetic acid(TCA) and 30% sodium dodecylsulfate (SDS) on cor-
neal permeability were compared between the two cor-
neal holders. These compounds were chosen to represent
categories of organic solvents, bases, acids and surfac-
tants. Each experiment was conducted using five corneas
and the t-test ðp6 0:05Þ was used to make relevant
comparisons. Data in the figures were grouped not
45
Std. Holder*
J.L. Ubels et al. / Toxicology in Vitro 18 (2004) 853–857 855
according to related categories but by range of values
for clarity of illustration.
0
5
10
15
20
25
30
35
40
1 min.1%NaOH
10 min.1% NaOH
1 min.Acetone
10 min.Acetone
Intact
Fluo
resc
ein
Upt
ake
(ng/
cm2 /m
in)
New Holder
*
*
*
Fig. 3. Permeability of corneas exposed to 1% NaOH or acetone for
1 or 10 min. * Significantly different than permeability in new holder
(t-test, p6 0:05, n ¼ 5). Permeability of corneas treated for 10 min is
significantly different than permeability of corneas treated for 1 min in
each type of holder.
12
14
16
/min
)
Std. HolderNew Holder
*
3. Results
Under all conditions tested in this study, corneal
fluorescein permeability measurements were higher in
the standard cornea holder than in the redesignedholder. There was also a measurable flux across the in-
tact cornea when fluorescein was detected using a fluo-
rimeter. In the new holder permeability of the intact
cornea was 0.13 ng/cm2/min, but this increased about
threefold, to 0.37 ng/cm2/min in the standard holder.
Removal of the cornea epithelial barrier by scraping
resulted in a permeability of 31.7 ng/cm2/min in the new
holder, but this increased to 74.2 ng/cm2/min in thestandard holder (Fig. 2).
Acetone, isopropanol (IPA) and 1% NaOH all caused
increased corneal permeability to fluorescein with
markedly greater permeability measurements in either
holder following the 10-min exposures (Figs. 3 and 4).
Permeability after these treatments never reached the
level of permeability of a cornea without epithelium.
After 1 or 10 min treatments with NaOH or IPA, cor-neal permeability was significantly higher in the stan-
dard holder than in the new holder. Permeability was
higher in the standard holder after a 1-min treatment
with acetone. After a 10-min exposure to acetone, cor-
neal permeability was also higher in the standard holder
but the difference was not statistically significant. It is
0
20
40
60
80
100
Intact no Epi 1 min.30%SDS
10 min.30% SDS
Fluo
resc
ein
upta
ke (n
g/cm
2 /min
)
Std. Holder
New Holder
*
*
*
Fig. 2. Permeability of intact corneas, corneas with the epithelium
removed by scraping and corneas exposed to 30% SDS for 1 or 10 min.
* Permeability in the standard holder is significantly different than in
the new holder (t-test, p6 0:05, n ¼ 5).
0
2
4
6
8
10
1 min.IPA
10 min.IPA
1 min.30% TCA
10 min.30%TCA
Intact
Fluo
resc
ein
upta
ke (n
g/cm
2
*
**
Fig. 4. Permeability of corneas exposed to isopropanol (IPA) or 30%
trichloroacetic acid (TCA) for 1 or 10 min. *Significantly different than
permeability in new holder (t-test, p6 0:05, n ¼ 5). Permeability of
corneas treated with IPA for 10 min is significantly different than
permeability of corneas treated for 1 min in each type of holder.
noted, however, that the variability of the response in
the standard holder (coefficient of variation¼ 0.42) was
much greater than in the new holder (coefficient of
variation¼ 0.066).
856 J.L. Ubels et al. / Toxicology in Vitro 18 (2004) 853–857
Exposure of the cornea to 30% TCA for either 1 or 10
min had no effect on permeability in the new holder. In
the standard holder treatment with TCA for 1 min
caused an approximately threefold increase in perme-ability compared to an intact cornea (0.92 vs 0.37 ng/
cm2/min), but a 10 min exposure had no effect on per-
meability compared to the intact cornea in the standard
holder. Corneal permeability to fluorescein after TCA
treatment was always very low, but was higher in the
standard holder than in the new holder after a 1 min
exposure (Fig. 4).
The permeability of corneas exposed to 30% SDS for1 min in the standard holder increased markedly to 96.5
ng/cm2/min, a level equal to permeability of corneas
without epithelium in the standard holder (Fig. 2). In
contrast, a 1 min exposure to SDS in the new holder
increased permeability to only 2 ng/cm2/min. A 10 min
exposure to 30% SDS increased the corneal permeability
to the range of 20–30 ng/cm2/min with no significant
difference between the two holders. The coefficient ofvariation was however higher in the standard holder
(0.28) than in the new holder (0.19).
4. Discussion
This study demonstrates that corneal permeability is
lower in the re-designed corneal holder for the BCOP
assay than in the standard holder under control condi-
tions or following treatment that breaks down the epi-
thelium. It is also shown that in both holders, when
corneas are exposed to compounds that compromise the
epithelial barrier a reduced treatment time of 1 mincauses less damage to the cornea than a 10 min treat-
ment. The exception to this observation is the effect of
SDS, as discussed below.
We previously suggested that the standard corneal
holder, which clamps onto clear cornea, causes edge
damage to the epithelium and endothelium (Ubels et al.,
1998, 2000, 2002), and we have confirmed this by his-
tology (data not shown). This edge damage would beexpected to increase corneal fluorescein permeability
and the data from the present study confirms this. The
large differences in permeability between the standard
and new holders when the epithelium is removed,
damaged by scraping or exposed to chemicals suggests
however that the advantage of the new holder is not due
to elimination of edge damage alone. In our previous
studies we showed that the standard holder causesextensive damage to the corneal endothelium due to
wrinkling of the cornea (Ubels et al., 2000, 2002). It is
well known that damage to the endothelium increases
corneal permeability (Araie, 1986a,b; Watsky et al.,
1989). We therefore conclude that the lower perme-
ability of corneas with a damaged epithelium in the new
holders is largely due to elimination of the mechanical
damage to endothelium that is caused by the standard
holder. It should be noted that corneal edge damage
may also allow greater penetration of test substances
into the cornea, thereby accounting for some of thedifference between the standard and new holders. The
exception to these observations is the effect of TCA. As
previously reported (Casterton et al., 1996) and con-
firmed in the present study, the permeability of corneas
treated with TCA is very low because TCA chemically
fixes the corneal epithelium rather than breaking down
the epithelial barrier (Ubels et al., 1998).
SDS, a surfactant, has a detergent effect on cellmembranes and extracellular matrix. A 1 min exposure
to SDS in the standard holder completely removed the
corneal epithelium, which explains the high permeability
of these corneas. In the new holder it was observed that
a 1 min exposure to SDS did not completely remove the
corneal epithelium, accounting for the lower perme-
ability in the new holder. Loss of the epithelium in the
standard holder may be accelerated by edge damagewhich allows more rapid penetration of SDS to the basal
cell layer of the epithelium. The relatively low perme-
ability of the corneas in the standard holder after a 10
min exposure to SDS was unexpected; however, it was
observed that after 10 min SDS not only removed the
epithelium, but also damaged the underlying stroma.
This may have had the effect of reducing permeability
compared to corneas with epithelial damage alone.The lower permeability of corneas treated with ace-
tone, 1% NaOH or IPA for 1 min compared to 10 min in
either holder is in agreement with our previously re-
ported data on opacity and hydration after reduced
treatment times (Ubels et al., 2000). We have suggested
that the 10 min treatment with test substances in the
BCOP protocol is an exaggerated exposure compared to
that which would occur accidentally in humans. On theother hand the bovine corneal epithelium is much
thicker than the human epithelium (10–12 cell layers
compared to 5–7 cell layers) with the result that a re-
duced treatment time might lead to underestimates of
damage when bovine eyes are used to screen potential
irritants. This problem requires further investigation.
In our report introducing the redesigned holder we
were unable to demonstrate a difference in cornealopacity or hydration between the standard and new
holders (Ubels et al., 2002). It was clear, however, that
the new holder eliminates edge damage and damage to
the endothelium. Therefore we argued for use of the new
holder based on morphology, since it is essential to be-
gin an in vitro toxicology assay with tissue that is in
optimal condition. This is a principle that has been
followed for many years in corneal physiology andpharmacology (Dikstein and Maurice, 1972; Edelhauser
and Maren, 1988; Klyce and Crosson, 1985; Riley,
1985). We have now demonstrated that using the rede-
signed corneal holder does in fact provide lower
J.L. Ubels et al. / Toxicology in Vitro 18 (2004) 853–857 857
permeability measurements. The permeability values
reflect effects of treatment alone that are not affected by
mechanical damage to the cornea. We believe that these
permeability data are more accurate and less variablethan those obtained when using the standard holder.
The use of the new holder will reduce the risk of false
positives due to erroneously high corneal permeability
when using the BCOP assay to screen substances for
potential ocular irritancy.
Acknowledgements
This study was supported by a grant from the Access
Business Group, Alticor Corp., Ada MI, USA.
References
Araie, M., 1986a. Barrier function of corneal endothelium and the
intraocular irrigating solutions. Archives of Ophthalmology 104,
435–438.
Araie, M., 1986b. Carboxyfluorescein. A dye for evaluating the corneal
endothelial barrier function in vivo. Experimental Eye Research 42,
141–150.
Casterton, P.L., Potts, L.F., Klein, B.D., 1996. A novel approach to
assessing eye irritation using the bovine corneal opacity and
permeability test. Journal of Toxicology––Cutaneous and Ocular
Toxicology 15, 147–163.
Dikstein, S., Maurice, D.M., 1972. The metabolic basis to the fluid
pump in the cornea. Journal of Physiology (London) 221, 29–41.
Edelhauser, H.F., Maren, T.M., 1988. Permeability of human cornea
and sclera to sulfonamide carbonic anhydrase inhibitors. Archives
of Ophthalmology 106, 1110–1115.
Gautheron, P.D., Dukic, M., Alix, D., Sina, J.F., 1992. Bovine corneal
opacity and permeability test: an in vitro assay of ocular irritancy.
Fundamental and Applied Toxicology 18, 442–449.
Gautheron, P., Giroux, J., Cottin, M., Audegond, L., Morilla, A.,
Mayordomo-Blanco, L., Tortjada, A., Haynes, G., Vericat, J.A.,
Priovano, R., Gillio Tos, E., Hagemann, C., Vanpary, S.,
Deknudt, G., Jacobs, G., Prinsen, M., Kalweit, S., Spielmann,
H., 1994. Interlaboratory assessment of the bovine corneal opacity
and permeability (BCOP) assay. Toxicology in Vitro 8, 381–
392.
Klyce, S.D., Crosson, C.E., 1985. Transport processes across the
corneal epithelium. Current Eye Research 4, 323–331.
Riley, M., 1985. Pump and leak in regulation of fluid transport in the
rabbit cornea. Current Eye Research 4, 371–376.
Sina, J.F., 1994. Validation of the bovine corneal opacity-permeability
assay as a predictor of ocular irritation potential. In vitro
Toxicology 7, 283–289.
Ubels, J.L., Erickson, A.M., Zylstra, U., Kruelen, C.D., Casterton,
P.L., 1998. The effect of corneal hydration on opacity in the bovine
cornea opacity and permeability assay. Journal of Toxicology––
Cutaneous and Ocular Toxicology 17, 197–220.
Ubels, J.L., Pruis, R.M., Sybesma, J.T., Casterton, P.L., 2000. Corneal
opacity, hydration, and endothelial morphology in the bovine
cornea opacity and permeability assay using reduced treatment
times. Toxicology in Vitro 14, 379–386.
Ubels, J.L., Paauw, J.D., Casterton, P.L., Kool, D.J., 2002. A
redesigned corneal holder for the bovine cornea opacity and
permeability assay that maintains normal corneal morphology.
Toxicology in Vitro 16, 621–628.
Ubels, J.L., Paauw, J.D., Casterton, P.L., 2003. Evaluation and
refinement of the bovine cornea opacity and permeability assay. In:
Katz, S., Salem, H. (Eds.), Alternative Toxicologic Methods. CRC
Press LCC, Boca Raton, FL, pp. 107–113.
Watsky, M.A., McDermott, M.L., Edelhauser, H.F., 1989. In vitro
corneal endothelial permeability in rabbit and human: the effects of
age, cataract surgery and diabetes. Experimental Eye Research 49,
751–767.