Connex Spot Monitor, CSM

Service and Repair training

Page 2: Connex Spot Monitor, CSM

Program

• Introduction

• Models and properties

• Operation – Screens & indicators

– User menu’s

– Advanced menu’s

– Configuration

• Service – Troubleshooting

– Repair

– Functional verification

• Quiz

Page 3: Connex Spot Monitor, CSM

CSM

• A Spot monitor for NIBP, SpO2 and temperature

• Suitable for Neo, Ped and Adults

• Touchscreen user interface

• Various connectivity options

• Can be supplemented with an Accessory Power Management (APM) stand

Page 4: Connex Spot Monitor, CSM

Model comparison

7100 Value

7300 Bluetooth

7400 WiFi Ready

7500 WiFi

Co

nn

ectivity

WiFi X

Bluetooth X

USB X X X X

Ethernet X X X X

SpO

2 O

ptio

ns

Nonin X X X

Nellcor X X X

Masimo X X X

Parameter

Co

mb

inatio

ns

BP Only X

BP/Temp X X

BP/SpO2 X X X X

BP/SpO2/Temp X X X X

Custom Scoring X X X

Hardware Upgradable No Yes Yes Yes

Software Upgradable Yes Yes Yes Yes

Page 5: Connex Spot Monitor, CSM

Model number configuration

Model SpO2 Temperature - Power Cord

71 = Value 73 = Bluetooth 74 = WiFi Ready 75 = WiFi

W = Nonin C = Covidien (Nellcor) M = Masimo X= None

T = SureTemp E = Braun Pro6000 Ear X = None

- A = Denmark B = North America C = China G = Argentina K = South Korea N = India/UAE P = Thailand T = Taiwan Y = Italy Z = Brazil 2 = Europe 3 = Israel 4 = United Kingdom 5 = Switzerland 6 = Australia 66 = Australia (Orange) 7 = South Africa

Example: A Connex Spot Monitor configured to include Bluetooth connectivity, Covidien SpO2, and Braun Pro6000 thermometry, shipping to North America would be 73CE-B

73 C E - B

Page 6: Connex Spot Monitor, CSM

Power supply

• Mains – Internal 100 – 240 V AC in APM

– Without APM, use external adapter: MS3, 35W

• Battery – 2 cell, Li-Ion smart battery inside the CSM

– Supplemented by a 9 cell Li-Ion battery in APM

– The hours of use depend on installed options, battery age and capacity, display brightness settings, etc

• Protection type: CSM Class 2, APM class l, BF

Page 7: Connex Spot Monitor, CSM

Communication

• USB: – Mini B

• Ethernet

• WiFi: – Internal, incl. antenna

– IEEE 802.11 a/b/g/n

– Security: WEP, WPA, WPA2

– Encryption: AES, PSK, EAP (various types)

• Bluetooth

Page 8: Connex Spot Monitor, CSM

CSM Controls, Indicators & Connections

1. 7” touchscreen

2. SureTemp module (can also be Braun Pro6000)

3. Battery charge / Power indicator 1. Off: mains not connected

2. Red: mains connected and charging

3. Green: mains connected, battery fully charged

4. Power button

Page 9: Connex Spot Monitor, CSM

CSM Controls, Indicators & Connections

1. Battery compartment

2. NIBP connection

3. USB client port

4. USB port

5. Power

6. Ethernet

7. Nurse call

8. SpO2 connection

9. SureTemp connection

Page 10: Connex Spot Monitor, CSM

APM indicators and connections

1. Battery compartment

2. Battery charge status

1 Led: 0 – 19% charge

2 Leds: 20 – 39% charge

3 Leds: 40 – 59% charge

4 Leds: 60 – 79% charge

5 Leds: 80 – 95% charge

6 Leds: 96 – 100% charge

3. Light power switch

Page 11: Connex Spot Monitor, CSM

APM indicators and connections

1. Printer port (future use)

2. 2 USB ports

3. USB cable to CSM

4. Power cable to CSM

5. Mains connection

6. Ground lug

7. Mounting recess

8. Battery cover screw

9. APM light

Page 12: Connex Spot Monitor, CSM

Device Operation, startup

• Runs POST upon starting up

• Confirm – LED’s flashing

– Splash screen

– Beep (buzzer) and Chime (speaker)

– Display: all panes appearing (after login screen if in use)

• In case of a POST failure, the CSM will display an appropriate message / alert

Page 13: Connex Spot Monitor, CSM

Device Operation, Home Screen

1. Status Bar: Clinician info, time, Connection info, profile, battery info (see next slide)

2. Content: each parameter (set) has it’s own pane, touching the panes triggers various functions

3. Primary navigation: Navigation to other screens

Page 14: Connex Spot Monitor, CSM

Device operation, battery status indicator

• Connected to mains, battery charging or full (estimated charge is displayed as a percentage of capacity)

• running on battery power , remaining capacity of all batteries: – 1–4 bars (represents % charge left) – Estimated charge time remaining

• When battery is running low, CSM will give 2 warnings: – Amber low priority, appr. 30 mins left – Red high priority, critically low

Page 15: Connex Spot Monitor, CSM

Device operation, user screens

• Patient screen: enter patient info

• Alarms screen: general alarm settings and setting alarm limits for parameters (only available in intervals profile)

• Review screen: review, send and delete recorded data

• Settings screen: user and advanced settings

Page 16: Connex Spot Monitor, CSM

Device operation, Alarms

• The CSM can present physiological and technical alarms

• 4 different alarm priority levels: – Information: blue background, 1 or 2 pulse tone

– Very low priority: cyan background , 1 or 2 pulse tone

– Low and medium priority: amber background , 3 pulse tone

– High priority: red background , 10 pulse tone

• Alarms are traceable via the log file and are maintained for 14 Log files

Page 17: Connex Spot Monitor, CSM

Device settings, Intervals

• Intervals can be set as:

– Automatic with a defined interval

– Stat: a 5 minute continuous measuring

– Program: Programs 1 – 5 are configured in the advanced settings, users can only run them, Program 6 can be configured by the user

Page 18: Connex Spot Monitor, CSM

Device settings, Profiles

• The following user profiles are available (model dependent): – Spot profile

– Office profile

– Intervals profile

• Spot and Office profiles lack intervals alarms etc.

• Office can use Bluetooth or USB connectivity

• Spot and Intervals can use USB, Ethernet and WLAN

Page 19: Connex Spot Monitor, CSM

Device settings, Device

• Display brightness from 1 to 10

• Power down and sleep (alt.: touch power button)

• Display lock

Page 20: Connex Spot Monitor, CSM

Device settings, Date & Time

• Date & Time may be automatically set by a network configuration

• To set manually: change values and touch “Set Time”

• Manual settings will be overwritten by network settings

Page 21: Connex Spot Monitor, CSM

Device settings, Clinician

• This screen allows a clinician to sign in and out of the system if the system is able to retrieve this info via the network

• Or to enter clinician details

• The clinician info will be displayed on the top left of the home screen

Page 22: Connex Spot Monitor, CSM

Advanced settings, Status

• Displays information about firmware, config files, hardware and connectivity status

• Allows logging in to the advanced settings menu

Page 23: Connex Spot Monitor, CSM

Advanced settings, General

• Language selection

• Date

– Set the Date format

– Use Network Time protocol (will disable manual date & time entry on settings tab)

Page 24: Connex Spot Monitor, CSM

Advanced settings, Parameters

• NIBP settings: view, units, tube type, algorithm, Target pressure (Step mode only)

• Temp: units, display option for other unit

• Averaging: Create different programs for NIBP averaging, Office profile only

• Program: Create interval programs, Intervals profile only

Page 25: Connex Spot Monitor, CSM

Advanced settings, Network

• Radio: specify ESSID, Radio band, network alarms

• Wireless Security: specify security and encryption

• Server: Specify method to connect to (Connex) Server + test option

• Bluetooth: Specify the device name as seen by other BT devices (7300 only)

Page 26: Connex Spot Monitor, CSM

Advanced Settings, Service

• Ethernet stats: LAN connection information

• Radio Stats: WLAN connection information

• General: Set device info and touchscreen calibration

• Config File: Current config info and load or save a configuration

• Log File: Save the log to a USB drive

• System: Reset to defaults and upgrade firmware

Page 27: Connex Spot Monitor, CSM

Lab

• Use

• Menu

• Advanced menu

Page 28: Connex Spot Monitor, CSM

Recommended service intervals

Page 29: Connex Spot Monitor, CSM

Troubleshooting, symptoms and solutions

• Symptoms are unexpected behavior that do not necessarily trigger an alarm or message

• Refer to the symptom and solutions table in the service manual

• This table will help you to deduct the cause of the issue and suggests actions to resolve it

Page 30: Connex Spot Monitor, CSM

Troubleshooting, technical alarms

• Technical alarms appear as messages in the status bar and follow the same logic as physiological alarms

• Most tech alarms are low (Cyan) or very low (blue) priority

• The messages can be traced in the DFU and service manual for possible causes and suggested action

Page 31: Connex Spot Monitor, CSM

Troubleshooting, getting the log files

• All technical alarms are maintained in the log

• The log can be saved to a USB drive and reviewed on a PC – To save the log to a USB drive: go to Service > Log file and

touch Save to USB

– On the USB drive, look for the zip file with the correct time stamp,

– within that zip file you will find another zip file for each day

– within that zip file you will find multiple log files for that specific day (the highest number is the latest one)

– These files can be opened with Wordpad

Page 32: Connex Spot Monitor, CSM

Troubleshooting, reading the log files

• The log messages follow this format:

<PRI>ISODATE(sp)MSGHDR:(sp)MSG(new-line)

– <PRI>: Facility and Priority number

– ISODATE: datestamp in this format: yyyy-mm-ddThh:mm:ss.mmm+-ZONE

– MSGHDR: Message header and [PID]

– MSG: text of the actual message

Page 33: Connex Spot Monitor, CSM

Troubleshooting, deciphering <PRI>

• To get the facility number: – Round down PRI/8 to an integer

• To get severity number: – PRI - (Facility x 8)

• Facility and Severity codes can be traced in the service manual

Example 1:

PRI = 13

Facility = 1

Severity = 5

Example 2:

PRI = 92

Facility = 11

Severity = 4

Page 34: Connex Spot Monitor, CSM

Troubleshooting, Message

• The message header indicates the program or process

• The message indicates the actual issue

• If the message also contains a code, then this message can also be found in the error messages table in the manual

Page 35: Connex Spot Monitor, CSM

CSM Service options

• Troubleshooting may lead to the requirement to replace some internal parts like boards, cables etc.

– For this follow the disassembly and reassembly instructions in the next slides

• After service and also as a preventive periodic maintenance, a full functional verification is required.

Page 36: Connex Spot Monitor, CSM

Part replacement

• Individual parts like boards, pump etc. are not to be repaired but replaced only

• Welch Allyn supplies service kits, these kits include parts that are usually replaced together .

• The service kits including the material numbers are listed in the service manual

Page 37: Connex Spot Monitor, CSM

Disassembly precautions

• Use ESD protection!

• Make sure to have a clean work space

• Make sure to have all the required tools:

– Phillips #1 Screwdriver

– Torx-10 Screwdriver

– Torque driver

• Before opening the device, always remove mains and battery first

Page 38: Connex Spot Monitor, CSM

Locking Connector

ZIF Connector

Friction Connector

Connector Types

The CSM uses these different cable connectors:

Page 39: Connex Spot Monitor, CSM

Disassembly Procedures, Backhousing

Page 40: Connex Spot Monitor, CSM

Disassembly Procedures, Backhousing

SureTemp and Pro6000 dock procedures are the same

Page 41: Connex Spot Monitor, CSM

Disassembly Procedures, Speaker

Disconnect cable and slide speaker out

Page 42: Connex Spot Monitor, CSM

Separate main PCBA from fronthousing

• Disconnect touchscreen flat-cable

• Lift antenna out of it’s holder

• Lift complete PCBA including connector panel (SpO2 insert) from fronthousing

Page 43: Connex Spot Monitor, CSM

Disassembly, SpO2 insert

• Remove 2 screw from the front of the assembly

• Disconnect the mains cable connector from the PCBA

• Disconnect the mains connector from the SpO2 insert

Page 44: Connex Spot Monitor, CSM

Disassembly, SpO2 board

• Procedures differ per SpO2 type

Nonin Covedien (Nellcor) Massimo

Page 45: Connex Spot Monitor, CSM

Disassemble, Radio & Antenna

• After removing the two screws, draw the board straight out to disconnect the mini-USB connector

• For the blue-tooth board, follow the same procedure

Page 46: Connex Spot Monitor, CSM

Disassemble, LCD

• Pull out the LCD flex cable (friction) • Turn assembly over and lift LCD as indicated

Page 47: Connex Spot Monitor, CSM

Disassemble, LCD

• Remove the gaf tape (1)

• Carefully lift the bale to unlock the flex cable (2)

• Remove the LCD

1

2

Page 48: Connex Spot Monitor, CSM

Disassemble, Pump Assembly

• Disconnect tubing manifold and connector

Page 49: Connex Spot Monitor, CSM

Disassemble, Pump Assembly

• Remove 2 screw that hold the pump assembly

• Pump assembly can now be removed

Page 50: Connex Spot Monitor, CSM

Disassemble, Bezel

• The rubber bezel is locked with the 6 bosses on the housing

Page 51: Connex Spot Monitor, CSM

Disassemble, Power switch

• To remove the power switch:

– Peel off the switch from the housing

– Pull the switch and flex cable through the opening

Page 52: Connex Spot Monitor, CSM

Disassemble, APM

• To replace the battery, loosen the screw that secures the battery door

• The APM itself cannot be disassembled and is replaced only

Page 53: Connex Spot Monitor, CSM

Reassembly

• Reassemble in reverse order

• Make sure to insert cables correctly and lock where applicable

• Use Torque driver to prevent damage to housing, see toque specification table

Page 54: Connex Spot Monitor, CSM

Reassembly, torque value table

Page 55: Connex Spot Monitor, CSM

Hands-on: Disassemby and reassamble

Let’s take it apart (and put it back together again)

Lab, Disassembly & reassembly

Page 56: Connex Spot Monitor, CSM

Welch Allyn Service tool (WAST)

• The Welch Allyn Service tool works with the CVSM, CIWS, CSM, CP150 and ProBP and future Platform devices and can be downloaded from the Welch Allyn website

• The tool comes with Calibration process locked which requires a license to unlock

• After repair training (online or classroom), a Gold license (Device specific) will be supplied to the trainee (not to the organization he or she works for)

Page 57: Connex Spot Monitor, CSM

Welch Allyn Service tool (WAST)

• The tool is used for the following: – Review device info – Schedule device maintanance – Install software updates and upgrades – View logs (event- and error) – User administration – Basic device check (NIBP accuracy, leak and over-

pressure test) – Program device SN after main PCB exchange (Gold

license only) – Full functional verification and calibration (Gold license

only)

Page 58: Connex Spot Monitor, CSM

Post repair actions, Main PCBA replacement

• After replacement of the main PCBA, use the WAST to:

– Program the device sn and host controller sn (sticker on the main PCBA)

– Program Device model number

– Re-initiate the NIBP module

– Update device firmware to the latest version

– Perform a Full functional calibration using WAST with a gold license

Page 59: Connex Spot Monitor, CSM

Program device sn’s and model number

• In WAST, double click the device

• Go to Configure – Current settings – change

• Enter device sn, Host controller sn and model number and click save

Page 60: Connex Spot Monitor, CSM

Update device firmware

• In WAST, double click device, go to upgrade tab

• Click check for upgrades (internet connection required)

• Select the device and click upgrade (all)

Page 61: Connex Spot Monitor, CSM

Initialize NIBP

• In WAST, double click the device and go to verify and calibrate

• Select NIBP sensor and click Begin

• Click Initialize NIBP board and confirm

• Follow instructions on screen to perform:

– the Power Calibration Test – Primary/Safety

– the Pressure Calibration – Primary/Safety

Page 62: Connex Spot Monitor, CSM

Functional verification

• Perform the Basic functional verification using the Welch Allyn Service Tool (WAST) Standard Edition • As periodic preventive maintenance perform the Basic Functional

Checks as defined in the service manual.

• Perform the complete functional verification using the Welch Allyn Service Tool (WAST) with the Gold license: – If the device has been dropped – If there is doubt about it’s functioning – If the device has been opened – If parts have been replaced

• Electrical safety test is not required on the CSM device itself due to the power supply is in the power cord assy. However if you use the APM with your device you should perform EST. The APM is not repairable only replaced but annual check should be done.

Page 63: Connex Spot Monitor, CSM

Functional verification

• The WAST help file displays all required tools and connections

Page 64: Connex Spot Monitor, CSM

Functional verification

• In WAST, double click the device and go to verify and calibrate

• To verify and calibrate the complete device, select the device and click perform all

• Alternatively you may perform each module separately but device calibration date will not be changed nor a calibration certificate generated.

• Follow the instructions on screen to perform all tests and calibrations

Page 65: Connex Spot Monitor, CSM

Hands-on: Functional verification

Let’s make sure it’s working as it should

Lab, Functional verification

Page 66: Connex Spot Monitor, CSM

Configuration tool

• The configuration tool allows users to create configurations online which can be deployed to the devices

• After logging in, the tool allows to create new, edit and download existing config files

• After downloading the config files can be deployed through the WAST or via a USB flash drive

• Configurations can also be deployed to devices that are being ordered

Page 67: Connex Spot Monitor, CSM

Configuration tool

http://config.welchallyn.com

Page 68: Connex Spot Monitor, CSM

Summary

• Connex Spot Monitor

• NIBP, SpO2, Temp

• Profiles

• Connectivity

• Service & repair options

• WAST – Updates

– Functional verification

– Configuration

Page 69: Connex Spot Monitor, CSM
Page 70: Connex Spot Monitor, CSM