Technical Information TI 193C/07/en No. 51500279 Due to the modularity of its design, the Liquisys S CLM 223 / 253 transmitter can be adapted to a wide range of customer needs. The basic version, which provides simple measuring and alarm signalling functions, can be equipped with additional software and hardware modules to match specific applications. Retrofitting of expansion modules is also possible. Areas of application Ultrapure water Water treatment Ion exchangers Reverse osmosis Cooling water desalinisation Sewage Benefits at a glance Measuring transmitter in field or panel-mounted housing Universal application Conductivity and specific resistance via software switching Transmitters for conductive or inductive measurement Ultrapure water compensation acc. to IEC 746 (conductive only) Simple handling Logically arranged menu structure with plain text in 6 languages facilitates instrument configuration Large, two-line display indicates measured value and temperature at the same time Calibration via CAL key Safe operation Overvoltage (lightning) protection according to EN 61000-4-5 Direct access for manual contact control User-defined alarm configuration for alarm contact and error current The basic unit can be extended with: 2 or 4 contacts for use as: Limit contacts (also for temperature) P(ID) controller Timer for simple rinse processes Complete cleaning with Chemoclean Plus package: Any current output configuration via table Automatic initiation of cleaning in case of alarm or limit violation Ultrapure water monitoring with pre-alarm acc. to USP 23 (conductive only) Polarisation detection (conductive only) Concentration measurement (% / TDS) Temperature compensation via table Live check of measuring cell HART ® or Profibus PA 2nd current output for temperature Hauser + Endress Nothing beats know-how Conductivity Measurement liquisys S CLM 223 / 253 Conductivity and Resistance Transmitter Quality made by Endress+Hauser ISO 9001 Liquisys S CLM 223 Liquisys S CLM 253

Upload: others

Post on 27-Jan-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

  • TechnicalInformationTI 193C/07/enNo. 51500279

    Due to the modularity of its design, theLiquisys S CLM 223 / 253 transmittercan be adapted to a wide range ofcustomer needs. The basic version,which provides simple measuring andalarm signalling functions, can beequipped with additional software andhardware modules to match specificapplications. Retrofitting of expansionmodules is also possible.

    Areas of application• Ultrapure water• Water treatment• Ion exchangers• Reverse osmosis• Cooling water desalinisation• Sewage

    Benefits at a glance• Measuring transmitter in field or

    panel-mounted housing• Universal application

    – Conductivity and specific resistancevia software switching

    – Transmitters for conductive orinductive measurement

    – Ultrapure water compensation acc.to IEC 746 (conductive only)

    • Simple handling– Logically arranged menu structure

    with plain text in 6 languagesfacilitates instrument configuration

    – Large, two-line display indicatesmeasured value and temperature atthe same time

    – Calibration via CAL key• Safe operation

    – Overvoltage (lightning) protectionaccording to EN 61000-4-5

    – Direct access for manual contactcontrol

    – User-defined alarm configuration foralarm contact and error current

    The basic unit can be extended with:

    • 2 or 4 contacts for use as:– Limit contacts (also for temperature)– P(ID) controller– Timer for simple rinse processes– Complete cleaning with Chemoclean

    • Plus package:– Any current output configuration via

    table– Automatic initiation of cleaning in

    case of alarm or limit violation– Ultrapure water monitoring with

    pre-alarm acc. to USP 23(conductive only)

    – Polarisation detection (conductive only)– Concentration measurement (% / TDS)– Temperature compensation via table– Live check of measuring cell

    • HART® or Profibus PA• 2nd current output for temperature

    Hauser+EndressNothing beats know-how

    Conductivity Measurementliquisys S CLM 223 / 253

    Conductivity and Resistance Transmitter

    Quality made byEndress+Hauser

    ISO 9001

    Liquisys S CLM 223

    Liquisys S CLM 253

  • Details Liquisys S CLM 223 / 253 providesa solution for all drinking waterprocessing, process water and sewagetreatment applications.

    Conductive or inductiveTwo instrument versions for measurementwith conductive (two-electrode) measuringcells or inductive measuring cells areavailable. The use of inductivemeasuring cells that are less sensitive tosoiling than conductive measuring cellsis recommended for concentrationmeasurement or adhering media.

    Features of the basic version (CD / ID) :

    Measurement of conductivity orspecific resistanceThis is selected via the menu. Duringmeasurement, the value measured canbe displayed in the other measuringmode. The temperature is displayed atthe same time if desired.

    Temperature compensation selectionsavailable are linear (α = 0 means notemperature compensation) or specialultrapure water compensation (NaClcurve) according to IEC 746. Thereference temperature is 25 °C.

    Different alarms are required dependingon application and operator. Thereforethe Liquisys S CLM 223 / 253 permitsindependent configuration of thealarm contact and error current foreach individual error. Unnecessary orundesirable alarms can be suppressedin this manner.

    Up to four contacts can be used aslimit contacts (also for temperature), toimplement a P(ID) controller or forcleaning functions.

    Direct manual operation of thecontacts (bypassing the menu)provides quick access to limit, control orcleaning contacts, permitting speedycorrection of deviations.

    The serial numbers of the instrumentand modules and the order code canbe called up on the display.

    The cell constant can be edited andeven calibrated for demanding specialapplications.

    The CS / IS version provides additionalfunctions :

    Inductive measuring cells must normallybe installed in pipes at a requiredminimum distance from the pipe wall. Ifthis minimum distance is not observed,the measured value changes. The built-in adaptive calibration feature can beused to compensate for this once thecell is installed.

    In order to display wide measuringranges while still achieving a highresolution in certain ranges, the currentoutput can be configured as requiredvia a table. This permits bilinear orquasi-logarithmic curves, etc.

    The sensor and process check systemalso indicates any polarisation of theconductive measuring cell. Since thereis no subdivision into severalmeasuring ranges (the full range from0 ... 2000 mS/cm is available at alltimes), the measuring range is limitedby the cell constant only. Polarisationdetection then determines whether or notthe working range determined by themeasuring cell is exceeded.

    The live check issues an alarm whenthe measuring cell signal does notchange over a defined period of time. Thismay be caused by blocking, passivation,separation from the process, etc.

    Ultrapure water monitoring according toUSP 23 means that the uncompensatedconductivity and the temperature aremeasured and compared to a table. TheLiquisys S CLM 223 / 253 (conductive)comes with this function, and theuser-adjustable pre-alarm indicatesundesirable operating values in due time.Full compliance with USP 23 requiresthe use of a precisely calibratedmeasuring cell, for example, the CLS 15.

    The conversion from conductivity toconcentration is effected using fouruser-definable concentration curves.This permits concentrations to bedisplayed directly in % or ppm, i.e. asTDS (total dissolved salts solids).

    E 017E 123- - - -

    onoffon

    offonoff

    2.4 / 22 mA

    offon

    Timer

    t

    off

    offon

    onChemoclean

    t

    Instrument DATA

    eS rNr I1

    1 563483

    0

    20

    mA

    t0

    20

    mA

    T T TµS/cm µS/cm

    10

    15

    20

    0.9

    1.0

    1.1

    45

    50

    55

    75

    80

    85

    1.8

    1.9

    2.1

    USP 23

    2

  • Basic version andplus package

    Measuring and controlsystem

    A complete measuring systemcomprises:• the conductivity transmitter Liquisys S

    CLM 223 or CLM 253• a conductivity measuring cell with or

    without integrated temperature sensor

    • a CYK 71 or CLK 5 cable permanentlyattached to the sensor

    Options:• an extension cable CYK 71

    (conductive) or CLK 5 (inductive) and• junction box VBM

    Measuring ranges ofconductivity measuringcells

    ENDRESS+HAUSERLIQUISYS

    CAL REL

    REL1

    REL1

    ALARMREL2

    REL2E

    ENDRESS+HAUSERLIQUISYS

    CAL REL

    REL1

    REL1

    ALARMREL2

    REL2E

    Up to 4 relays

    Measuringcable

    Liquisys S CLM 253

    CLS 15

    CLS 21 withintegrated Pt 100

    Immersionassembly CLA 111

    Alarm contact

    Signal output 1:conductivity

    2 digital inputs

    Power supply (mains)

    Aux. 15 V supply voltage

    LM2X3SYS.CDR

    Signal output 2:temperature (optional)

    Liquisys S CLM 223CLS 50

    Complete measuring systemswith Liquisys S CLM 223 / 253

    100.01 0.1 1 10 100 1 100 1000

    µS/cm

    k = 10 (CLS 30)

    CLS 50

    k = 1 (CLS 21)

    k = 0.1 (CLS 12 / 13 / 15 / 19)

    k = 0.01 (CLS 12 / 13 / 15 / 19)

    mS/cm

    CD-CELL.CDR

    Measuring ranges ofconductive andinductive conductivitymeasuring cells

    Basic version With plus package (CS / IS version)

    MEASUREMENT

    CALIBRATION of cell constantCALIBRATION of residual coupling (ind. cells)

    Read instrument DATA

    Linear CURRENT OUTPUTCURRENT OUTPUT simulation

    1 programmable ALARM CONTACT(contact and error current)

    Adaptive CALIBRATION (inductive cells)

    POLARISATION detection(two-electrode cells)LIVE CHECK of sensor

    CURRENT OUTPUT programmable (table)

    Additional features Additional features

    2 CHANGEOVER CONTACTS for– Measuring parameter limit– Temperature limit– P(ID) controller– Timer for cleaning

    Measurement according to USP 23 withpre-alarm (two-electrode cells)Automatic cleaning triggered by alarm or limitviolation

    Additional features Additional features

    2 more CHANGEOVER CONTACTS (total of 4) for– Measuring parameter limit– Temperature limit– P(ID) controller– Chemoclean cleaning

    (water and cleaning agent)

    Cleaning triggered externally or automaticallyby alarm or limit violation

    Ala

    rmsi

    gnal

    ling

    Con

    trolli

    ngC

    lean

    ing

    3

  • Operation

    Everything at a glanceThe display simultaneously shows thecurrent measured value and thetemperature – the essential processdata. Brief informational texts in theconfiguration menu provide assistancewith parameter configuration.

    Intelligent and simpleAll instrument control functions arearranged in a logical menu structure.Following access code entry, theindividual parameters can be easilyselected and modified as needed.

    SETUP 1

    SETUP 2

    ALARM

    CHECK

    RELAYS

    α TABLE

    CONCENTRATION

    SERVICE

    CURRENT OUTPUT

    Selection of measurement type and unit, adjustment of measuredvalue filter

    Adjustment of current output assignment, curve andsimulation for verification

    User adjustment of errors to switch relay or generate error current(S version: automatic cleaning trigger in the event of an alarm)

    Adjustment of polarisation detection (conductive) and live check alarm

    Up to 4 relays may be configured as limit contacts, P(ID) controller, timeror for chemical cleaning (S version: automatic cleaning trigger in caseof limit violation / CS version: USP 23)

    Definition of table for temperature compensation (up to 10 value pairs)α

    In addition to conductivity measurement, the concentration of a substancecan be displayed in different units with user-defined assignments tothese base units (e.g. % / ppm TDS)

    Selection of language version, display of software or hardwareserial numbers, entry of access codes

    Temperature measurement and temperature compensation

    CALIBRATIONCalibration of cell constant; calibration of residual coupling (inductive cells);for IS version also adaptive calibration

    Current temperature display can be suppressed by pressingthe plus key

    Display of active errors (sequence according to priority)

    MEAS.VALUE

    TEMPERATURE

    Display

    LM2X3OP.CDR

    Overview of Liquisys SCLM 223 / 253 menu.This overview covers allthe options that can beinstalled (see page 2Details).

    1734mS/cm

    ENDRESS+HAUSERLIQUISYS

    CAL REL

    REL1

    REL1

    ALARMREL2

    REL2E

    Setpoint

    R362

    Liquid crystal displayfor display ofmeasured values,configuration dataand current menu field

    4 main control keys forcalibration and instrumentconfiguration

    Key for switchingbetween automatic andmanual relay operation

    Key for manualrelay switching

    LED indicatorsfor switched limitoutputs

    LED indicator foralarm function

    LCD-LF-E.CDR

    Field for userlabelling

    User interface:Display and keys

    4

  • Electrical connection All connections to the panel-mountedinstrument CLM 223 are established viathe terminal strips on the rear.In the case of the field instrumentCLM 253, all wires (including themeasuring cell cable) are connected toterminals in a separate wiringcompartment.

    All the wiring can remain in place if theinstrument needs repair because allpossible repairs are limited to assemblyreplacement. So, dismantling theinstrument and rewiring are no longernecessary.

    REAR223.TIF

    Left:Liquisys S CLM 223,connections on the rearof the instrument

    Right:Liquisys S CLM 253,rear of instrument withreplaceable electronicsbox (E)

    15 V85

    86

    S

    10 ... 50 V

    93

    94

    L+

    L–

    L

    N

    ∼ =

    41

    42

    43

    47

    57

    51

    54

    48

    58

    52

    55

    49

    59

    53

    56

    mA

    10 ... 50 V

    81

    34

    33

    82

    ϑ

    ϑϑ

    mA

    32

    31

    11

    84

    84

    15

    16

    83

    83

    12

    13

    κ

    Conductivemeasuringcell

    Relay 1*

    Temperaturesensor Pt 100

    Power supply

    ANLF.CDR

    Aux. voltageoutput

    Digital input 1

    Signal output 1:conductivity

    Relay 2*

    Alarm*

    Relay 3*

    Relay 4*

    Digital input 2

    Inductivemeasuringcell

    optional

    (alternative)

    Signal output 2:temperature(optional)

    *Contactstatus shown:no current orfault present

    bl

    bl

    rdrd

    ye

    gn

    wt

    rd

    wt

    Electrical connection ofLiquisys S CLM 223 / 253

    FELD2.CDR

    E

    5

  • DimensionsENDRESS+HAUSERLIQUISYS

    CAL REL

    REL1

    REL1

    ALARMREL2

    REL2E

    XM223-2D.CDR

    90

    139

    6

    96

    Dimensions ofLiquisys S CLM 223

    ENDRESS+HAUSERLIQUISYS

    CAL REL

    REL1

    REL1

    ALARMREL2

    REL2E

    XM253-2D.CDR

    157

    247

    115

    154

    11

    ∅6

    170

    70

    70

    Dimensions ofLiquisys S CLM 253

    6

  • Mounting ofLiquisys S CLM 223

    The panel-mounted version is mountedusing the supplied tensioning screws.The required overall mounting depth isapprox. 175 mm.

    Mounting ofLiquisys S CLM 253

    There are several mounting options forthe field instrument:• Mounting on a square-tube mounting

    post• Mounting on cylindrical pipes• Wall mounting with fastening screws

    Weather protection cover CYY 101 isrequired for outdoor installation. Thiscover is compatible with all fieldinstrument mounting options.

    Proceed as follows to install theinstrument on a square-tube mountingpost (universal upright post CYY 102 orupright post of suspension assemblyholder CYH 101):➀ Install the weather protection cover

    on the upright post first.➁ Attach the field instrument to the

    weather protection cover from therear.

    CYY101.CDR

    300

    270

    320

    Weather protectioncover CYY 101(see Accessories)

    XM223MON.CDR

    Mounting cutoutaccording toDIN 43700

    92+

    0.5

    92+0.5

    90

    33

    96

    57

    ca. 25 139

    max. 45

    6

    Flat gasket(included in delivery)

    Installation of panel-mounted housing

    YY101102.CDR

    60 × 60

    150

    1495

    120

    Universal upright postCYY 102 or identicalupright post ofsuspension assemblyholder CYH 101(square tube, seeAccessories);mounting of weatherprotection cover andfield instrument

    7

  • Mounting ofLiquisys S CLM 253(continued)

    XM253WAN.CDR

    157

    115

    16

    ∅8

    Wall mounting;Screws: ∅ 6 mmAnchors: ∅ 8 mm

    MASTSATZ.CDR

    90

    70

    30

    ∅8

    ∅ 6

    90

    70

    9070

    ∅8

    M 6

    60

    Post mounting kit forinstallation of fieldhousing on cylindricalpipes (horizontal orvertical, max. ∅ 60 mm;see Accessories)

    XM253MAS.CDR

    Mounting on cylindricalpipes (shown withweather protectioncover installed on right)

    8

  • Technical dataGeneral specifications Manufacturer Endress+Hauser

    Product designation Liquisys S CLM 223Liquisys S CLM 253

    Input Measured quantities conductivity, resistance, temperature

    Conductive conductivity/resistance measurement

    Measuring range conductivity: 0 ... 2000 mS/cm (uncompensated)resistance: 0 ... 86 MΩ ·cm

    Usable cell constant k = 0.0025 ... 99.99 cm–1

    Maximum length of cable to measuring cell conductivity: 100 mresistance: 15 m

    Measuring frequency conductivity: 299.75 ... 1077.6 Hzresistance: 32.5 ... 425 Hz

    Inductive conductivity measurement

    Measuring range 0 ... 2000 mS/cm (uncompensated)

    Usable cell constant k = 0.0025 ... 99.99 cm–1

    Maximum length of cable to measuring cell 60 m (CLK 5)

    Measuring frequency 2 kHz

    Minimum distance for 0 / 4 ... 20 mA signal meas. value betw. 0 and 199.9 µS/cm: 20 µS/cmmeas. value betw. 200 and 1999 µS/cm: 200 µS/cmmeas. value betw. 2 and 19.99 mS/cm: 2 mS/cmmeas. value betw. 20 and 2000 mS/cm: 200 mS/cm

    Temperature measurement

    Temperature sensor Pt 100

    Measuring range –35 ... +250 °C

    Temperature offset range ±20 °C

    Temperature compensation

    Compensation types linear, NaCl, table, (conductive only:) ultrapure water

    Range –35 ... +250 °C

    Reference temperature 25 °C

    Digital inputs 1 and 2

    Voltage 10 ... 50 V

    Current consumption max. 10 mA

    Output Conductivity/resistance signal output

    Current range 0 / 4 ... 20 mA, galvanically separated;error current 2.4 / 22 mA

    Load max. 500 Ω

    Output range adjustable

    Isolation voltage max. 350 Vrms / 500 V DC

    Overvoltage (lightning) protection according to EN 61000-4-5:1995

    Temperature signal output (optional)

    Current range 0 / 4 ... 20 mA, galvanically separated

    Load max. 500 Ω

    Output range adjustable, ∆ 10 ... ∆ 100% of upper range value

    Isolation voltage max. 350 Vrms / 500 V DC

    Overvoltage (lightning) protection according to EN 61000-4-5:1995

    Auxiliary voltage output

    Output voltage 15 V ± 0.6 V

    Output current max. 30 mA

    Contact outputs (floating changeover contacts)

    Switching current with ohmic load (cos ϕ = 1) max. 2 A

    Switching current with inductive load (cos ϕ = 0.4) max. 2 A

    Switching voltage max. 250 V AC, 30 V DC

    Switching power with ohmic load (cos ϕ = 1) max. 1250 VA AC, 150 W DC

    Switching power with inductive load (cos ϕ = 0.4) max. 500 VA AC, 90 W DC

    9

  • Technical data(continued)

    Output (continued) Limit contactor

    Pickup / droput delay 0 ... 2000 s

    Controller

    Function (adjustable) pulse length / pulse frequency controller

    Controller response PID

    Proportional band Kp: 0.10 ... 10.00

    Period for pulse length controller 0.5 ... 999.9 s

    Frequency for pulse frequency controller 60 ... 180 min–1

    Alarm

    Function (switchable) steady / fleeting contact; normally closed / open

    Alarm threshold adjustment range conductivity / resistance / temperature:complete measuring range

    Alarm delay 0 ... 2000 s

    Accuracy Conductivity measurement

    Deviation of indication1 max. 0.5% of measured value ± 4 digits

    Reproducibility max. 0.2% of measured value ± 2 digits

    Measurement deviation1, conductivity signal output 0.75% of current output range

    Resistance measurement

    Deviation of indication1 max. 0.5% of measured value ± 4 digits

    Reproducibility max. 0.2% of measured value ± 2 digits

    Measurement deviation1, resistance signal output 0.75% of current output range

    Temperature measurement

    Resolution 0.1 °C

    Deviation of indication1 max. 1.0% of measuring range

    Measurement deviation1, temperature signal output max. 1.25% of current output range

    Ambient conditions Ambient temperature (nominal operating conditions) –10 ... +55 °C

    Ambient temperature (limit operating conditions) –20 ... +60 °C

    Storage and transport temperature –25 ... +65 °C

    Relative humidity (nominal operating conditions) 10 ... 95%, non-condensing

    Protection class of panel-mounted unit IP 54 (front), IP 30 (housing)

    Protection class of field housing IP 65

    Electromagnetic compatibility interference emission acc. to EN 50081-1:1992interference immunity acc. to EN 50082-2:1995

    Physical data / design Dimensions of panel-mounted unit (H × W × D) 96 × 96 × 145 mm

    Installation depth approx. 175 mm

    Dimensions of field housing (H × W × D) 247 × 170 × 115 mm

    Weight of panel-mounted unit max. 0.7 kg

    Weight with field housing max. 2.3 kg

    Display LC display, two lines, five and nine digits, with statusindicators

    Materials

    Housing of panel-mounted unit polycarbonate

    Front membrane polyester, UV-resistant

    Field housing ABS PC Fr

    Power requirements Supply voltage 100 / 115 / 230 V AC +10 / –15%, 48 ... 62 Hz24 V AC/DC +20 / –15%

    Power consumption max. 7.5 VA

    Fuse protection fine-wire fuse, medium time-lag, 250 V / 1 A

    1acc. to IEC 746-1, for nominal operating conditions

    Subject to modifications.

    10

  • Accessories Mounting accessoriesType Features Order number

    Weatherprot. coverCYY 101

    For mounting on field housing, for outdoor useDimensions (H × W × D): 320 × 300 × 270 mmMaterial: stainless steel 1.4301

    CYY 101-A

    Universalupright postCYY 102

    Square tube for mounting of field housingDimensions (H × W × D): 1495 × 60 × 60 mmMaterial: stainless steel 1.4301

    CYY 102-A

    SuspensionassemblyholderCYH 101

    For immersion assemblies with suspensionbracket, e.g. CYA 611;for installation on basin or channel rimMaterials: stainless steel 1.4301 / PE / PP / PA

    CYH 101-A

    Pendulumframe

    For pendulum suspension of assemblies, e.g.CYA 611Material: stainless steel 1.4301

    50080196

    Postmounting kit

    Kit for mounting of field housing on horizontalor vertical pipes (∅ max. 60 mm)Material: stainless steel 1.4301

    50086842

    Assemblies

    Type Features Applications

    CYA 611 Immersion assembly with G 1, G ¾ or NPT ¾"thread

    Sewage, water

    DipsysCLA 140

    Immersion assembly with flange DN 80 PN 16,ANSI 3" 150 lbs or JIS 10K 80A; measuringcell holder with bayonet lock

    Sewage, paperindustry

    Two-electrode conductivity measuring cells (selection)

    Type Features Applications

    CLS 15 Cell constant: k = 0.1 cm–1 or 0.01 cm–1

    Process connections: NPT ½", 1½" clampCondensate, ionexchangers

    CLS 21 Cell constant: k = 1 cm–1

    Process connections: G 1, DN 25 andDN 40 dairy pipe fitting, 2" clamp

    Service water,boiler feed water

    CLS 30 Cell constant: k = 10 cm–1

    Process connections: G 1, DN 25 andDN 40 dairy pipe fitting

    Service water,concentrationmonitoring

    Inductive conductivity measuring cell

    Type Features Applications

    CLS 50 Cell constant: k ≈ 2 cm–1Process connections: G ¾, NPT 1", DN 50 flange(incl. 5 m attached cable)

    Concentrationmeasurement,product monitoring

    Cables

    Type Features Order number

    CYK 71 Standard measuring cable for connection oftwo-electrode-measuring cells with andwithout Pt 100 (also usable for extension)

    50085333

    CLK 5 Extension cable for inductive measuring cells 50085473

    VBM Junction box 50003987

    11

  • Product structure

    Endress+Hauser GmbH+Co.- Instruments International -

    P.O. Box 2222D-79574 Weil am Rhein

    Tel. (07621) 975 - 02Fax (07621) 975345

    TI 193C/07/en/01.99Printed in Germany / DT / CV5 (51500279)

    Hauser+EndressNothing beats know-how

    Version

    CD Conductivity/resistance measurement (two-electrode cell)CS Conductivity/resistance measurement (two-electrode cell) with additional functionsID Conductivity measurement (inductive cell)IS Conductivity measurement (inductive cell) with additional functions

    Power supply

    0 230 V AC1 115 V AC5 100 V AC8 24 V AC/DC

    Measurement output

    0 Conductivity/resistance1 Conductivity/resistance and temperature3 Profibus PA5 Conductivity/resistance with HART6 Conductivity/resistance with HART and temperature

    Contacts

    05 No additional contacts10 2 contacts (limit / P(ID) / timer)15 4 contacts (limit / P(ID) / Chemoclean)

    CLM 223- complete order code for CLM 223

    CLM 253- complete order code for CLM 253

    Conductivity and resistance transmitter Liquisys S CLM 223 / CLM 253