comparison of the measurement of the phase change on reflection...
TRANSCRIPT
EUROMET Project 413: Inter-laboratory comparison of measurements of the phase
correction in the field of gauge block interferometry
Richard K. Leach, Anne Hart
National Physical Laboratory
Queens Road, Teddington
Middlesex TW11 0LW, UK
Abstract
A set of eight gauge blocks, two steel platens and two quartz platens have been circulated to ten
European laboratories, plus laboratories in the USA and Australia, in order to compare methods
for measuring a phase correction when obtaining the interferometric length of a gauge block.
Measurements were made of the central length and a phase correction for each gauge block.
The participants were free to use any method to measure a phase correction. The four methods
used to measure a phase correction were: the phase stack technique, optical scattering methods,
stylus profilometry and a mechanical contacting technique. The results for the central length
measurements showed a linear secular decrease in length as the project progressed - the total
change in length being about 30 to 40 nm. Within a 95% confidence level, the central length
measurements all agree, but the phase correction results vary by as much as 30 nm in the worst
case. Significant changes to the surface were recorded as the project progressed - the scratch
content increased and there was evidence of a reaction of the gauge block surfaces with the box
in which they were housed.
1. Introduction
The most accurate method for calibrating the length of a gauge block is to use optical
interferometry. However, interferometry measures the optical length rather than the mechanical
length of the gauge and an error in the length measurement may be introduced because of the
different bulk and surface characteristics of the gauge and the platen to which the gauge has
been wrung. These different characteristics, such as the surface roughness and the bulk optical
constants, will give rise to a difference between the length measured in an interferometer and
that measured using a mechanical, contacting technique. ISO 3650 (1998) states that a correction
must be applied to the measured length of the gauge to account for the differences in surface
characteristics and this correction is known as the ‘phase correction’.
The measuring instructions for this comparison, which involved ten European National
Measurement Institutes (NMIs) plus one laboratory in the USA and one in Australia, required
each participating laboratory to measure the deviation from nominal length and a phase
correction for eight gauges and four platens. This project is, therefore, a comparison of both
central length measurements and phase correction measurements. This paper presents a
summary of the results of this comparison.
2. Participants and timetable
2.1 The participants
The comparison was undertaken within the framework of EUROMET (project number 413)
with the National Physical Laboratory (NPL, UK) acting as the pilot laboratory. Table 1 lists, in
alphabetical order, the twelve laboratories which participated.
1 Bureau National de Métrologie - Laboratoire National d’Essais, 1 Rue Gaston Boissier, F-75015 Paris, France (BNM - LNE).
2 Centro Español de Metrologia, C/ del Afar 2, 28760 Tres Cantos, Madrid, Spain (CEM).
3 Commonwealth Scientific and Industrial Research Organisation, PO Box 218, Lindfield, NSW 2070, Australia (CSIRO).
4 Centre for Metrology and Accreditation, PO Box 239, FIN-00181, Helsinki, Finland (CMA
5 Danish Institute of Fundamental Metrology, Building 307, Anker Engelunds Vej 1, DK-2800 Lyngby, Denmark (DFM).
6 Instituto di Metrologia “G. Colonetti”, Strada della Cacce 73, 10135 - Torino, Italy (IMGC).
7 Justervesenet, Norwegian Metrology and Accreditation Service, Fetveien 99, N2007, Norway (JV).
8 National Institute of Standards and Technology, Metrology Building, Room B113, Gaithersburg, Maryland 20899-0001, USA (NIST).
9 National Physical Laboratory, Teddington, Middlesex TW11 0LW, United Kingdom (NPL).
10 Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany (PTB).
11 Portuguese Institute for Quality, Rua C à Avedida dos Trés Vales, 2825 Monte da Caparica, Portugal (IPQ).
12 Swiss Federal Office of Metrology, Lindenweg 50, CH-3003 Bern-Wabern, Switzerland (OFMET).
Table 1. Participating laboratories
2.2 Timetable
Each participant was given six weeks to carry out the measurements. The pilot laboratory
measured the gauges at the beginning, middle and end of the project. Table 2 shows the
timetable of the project.
Laboratory Time schedule Date of
measurement Receipt of results
NPL 1/5/97 30/5/97 - OFMET 30/6/97 3/7/97 14/10/97 DFM 11/8/97 18/8/97 11/11/97 CEM 22/9/97 23/10/97 12/12/97 BNM - LNE 3/11/97 5/12/97 5/1/98 IPQ 15/12/97 30/1/98 12/2/98 NPL 2/2/98 19/2/98 - NIST 16/3/98 15/4/98 1/9/98 CMA 27/4/98 2/6/98 1/7/98 JV 8/6/98 16/6/98 -
3/7/98 9/7/98
IMGC 20/7/98 4/9/98 1/2/99 PTB 12/10/98 Not known 3/5/99 CSIRO 1/3/99 16/3/99 -
20/4/99 5/5/99
NPL 1/4/99 20/5/99 -
Table 2 The timetable
3. The gauge blocks and platens
The eight grade K steel gauge blocks were supplied with nominal lengths of 2,5 mm, 3,0 mm,
3,5 mm, 4,0 mm, 4,5 mm, 5,0 mm, 5,5 mm and 6,0 mm. The gauge blocks fulfilled the
requirements of the relevant specification standard (ISO 3650: 1998). Also supplied were two 50
mm diameter quartz platens and two 60 mm diameter steel platens, together with an adjustable
mount to be used for obtaining the correct orientation for the wedged quartz platens in an
interferometer. All the gauges were marked with their respective nominal lengths and the serial
number C289. The gauges were manufactured by Alan Browne Gauges Ltd and were
approximately six months old at the start of the comparison. When measured at NPL, all the
gauge surfaces wrung readily. The coefficient of thermal expansion was given by the
manufacturer as 10,6 x 10-6 K-1 and all participants used this value. Provided the gauges were
measured on the steel platens, no correction for the different optical constants was requested
(although PTB did make a correction). Where the quartz platen was used, a correction for the
optical constant was used (see equation 2.1 in Leach [1998]). The values for the optical constants
of the gauges and platens, n and k, were obtained at Southampton University using
ellipsometry (PTB also used an ellipsometer to measure n and k).
The steel platens were labelled PLBI7 and PLBI8 and the quartz platens were labelled A and B.
The quartz platens were polished so that their optical axes were perpendicular to their polished
faces in order to minimise the difference in thermal expansion coefficient between quartz and
steel.
4. The measurements
Measurements were made of the central length of each gauge block according to ISO 3650. A
phase correction was also measured for each gauge using any method desired. Length
measurements were carried out at 20 ºC ± 0,3 ºC. Instructions for handling and measuring the
gauges as well as for data reporting were distributed to all the participants. Only combined
standard uncertainties (k = 1) have been quoted in the results section.
5. Instrumentation
Table 3 lists the instruments used to measure both central length and a phase correction. NIST
did not measure central length. To determine a correction for the refractive index of the air, all
participants, with the exception of NIST who did not make interferometric measurements, used
the modified Edlèn formula (Birch and Downs 1993, Bönsch and Potulski 1998) along with
measurements of the temperature, pressure and humidity of the air. PTB and NPL also used
interference and phase contrast microscopy to examine some of the surfaces. Details of the
instruments are not presented here but can be found in the references provided.
Laboratory Interferometer used to measure central length/manufacturer
Methods used to measure phase correction
NPL Twyman Green/NPL, TESA (Pugh,
Jackson 1986)
Phase stack technique (Hume K J
1963), Total integrated scatter (TIS)
(Leach 1998)
OFMET Twyman Green/NPL, TESA Phase stack technique
CEM Twyman Green/NPL, TESA Phase stack technique
BNM-LNE Twyman Green/NPL, TESA Phase stack technique
IPQ Twyman Green/NPL, TESA Phase stack technique
NIST Not measured Spherical contact technique (Stoup,
Faust, Doiron 1998)
CMA Twyman Green/CMA (Ikonen, Riski
1993)
Phase stack technique,
Integrating sphere (TIS) (Ikonen, Riski
1993)
JV Twyman Green/NPL, TESA Phase stack technique
IMGC Twyman Green/NPL, TESA Phase stack technique
PTB Twyman Green/PTB Phase stack technique
Integrating sphere (TIS) (Bönsch 1998)
CSIRO Kösters interferometer Mechanical stylus technique (Thwaite
1977)
Table 3 Methods used by the participants to measure central length and phase correction
6. Results and discussion
6.1 Deviation form nominal length
The reference values which served for the comparison of length measurements were calculated
as the non-weighted mean over all measurements for each gauge block. The pilot laboratory,
which measured the length of the gauge blocks three times, thus contributes three times to this
average. Figures 1(a) and 1(b) show the deviations from the nominal lengths for each gauge
block. All the methods for measuring a phase correction have been included in this figure and
the standard deviations for all the results are quoted.
Time (days)
-50-40-30-20-10
01020304050
0 100 200 300 400 500 600 700 800
2.5 mm gauge block
Mean deviation = 2.6 nmStandard deviation = 15 nm
-50-40-30-20-10
01020304050
mea
sure
d
leng
th /
nm
Mean deviation = -32.0 nmStandard deviation =15 nm
3.0 mm gauge block
-50-40-30-20-10
01020304050
Dev
iatio
n
from
m
ean
3.5 mm gauge block
Mean deviation = -5.7 nmStandard deviation =14 nm
-50-40-30-20-10
01020304050
4.0 mm gauge block
Mean deviation = 16.3 nmStandard deviation =17 nm
NPL1
OFM
ETD
FM
CEM
BNM
-LNE
IPQN
PL2
CM
AJV IM
GC
PTB
CSIR
O
NPL3
Figure 1(a). Deviations from mean measured length for 2.5 mm to 4.0 mm gauge blocks where � and � denote gauges measured on steel platens and quartz platens respectively, using stack method for phase correction, ∆ denotes gauges measured on steel platens using TIS for phase correction and � and � denotes gauges measured on steel platens and quartz platens respectively using surface roughness for phase correction.
Time (days)
-50-40-30-20-10
01020304050
0 100 200 300 400 500 600 700 800
4.5 mm gauge block
Mean deviation = -14.3 nmStandard deviation =13 nm
-50-40-30-20-10
01020304050
mea
sure
d l
engt
h /
nm
5.0 mm gauge block
Mean deviation = -40.7 nmStandard deviation =15 nm
-50-40-30-20-10
01020304050
Dev
iatio
n f
rom
m
ean
5.5 mm gauge block
Mean deviation = 23.8 nmStandard deviation = 14 nm
-50-40-30-20-10
01020304050 6.0 mm gauge block
Mean deviation = -39.8 nmStandard deviation = 19 nm
NPL1
OFM
ETD
FM
CEM
BNM
-LNE
IPQN
PL2
CM
AJV
IMG
C
PTB
CSIR
O
NPL3
Figure 1(b). Deviations from mean measured length for 4.5 mm to 6.0 mm gauge blocks where � and � denote gauges measured on steel platens and quartz platens respectively, using stack method for phase correction, ∆ denotes gauges measured on steel platens using TIS for phase correction and � and �denotes gauges measured on steel platens and quartz platens respectively using surface roughness for phase correction.
It is clear from figures 1(a) and 1(b) that there has been a secular change in length for each
gauge block of approximately 30 to 40 nm. A linear fit could be removed from the data to take
account of this secular change in length. However, at this stage of the analysis only the raw
results are considered.
It is impossible to say whether the secular length change is due to chemical and physical
changes in the bulk material, i.e. ageing, or due to wear during handling and repeated
wringing. The condition of the gauges was measured at NPL at the start of the project using
Nomarski phase contrast microscopy. Figure 2 shows a micrograph of a typical gauge surface
before the start of the project.
Figure 2. Nomarski micrograph of a gauge surface before the project began (NPL).
Figure 3. Interference micrograph of a gauge surface during the project (PTB).
The polishing and lapping marks can just be made out in this figure, but the main structures are
the bubbly marks that appear to be topographic features above the mean plane. PTB used
differential interference microscopy and Mirau microscopy to observe a gauge surface towards
the end of the project and noticed the dominant topographical features shown in figure 3 which
now appear to be valleys in the surface. However, measurements using the NPL stylus
instrument, NanoSurf IV, prove that these apparently topographical structures are in fact
crystal grains, having different phase changes on reflection, embedded in the bulk material. The
difference between the condition of the gauges in figures 2 and 3 illustrates that, as would be
expected, they have been scratched during use. Note that the micrographs presented in this
paper are for qualitative information only.
Figure 4 shows the surface of the 3,5 mm gauge at the end of the project measured using the
Nomarski microscope. A new structure is now apparent in the form of the lighter bubbly marks
shown in the figure. Repeated cleaning of the gauge surface did not clear these marks. It was
also noted that only half of the gauge was covered with this structure. To date, the only
explanation of this phenomenon is that the gauges have reacted with either the wooden
material of their box or its treatment process (wood varnish). This structure is not visible to the
un-aided eye. Other gauges also display this effect to a lesser extent.
Figure 4 Nomarski micrograph of the 3,5 mm gauge surface at the end of the project (NPL).
6.2 Phase correction
The results for the phase corrections for the steel gauges on steel platens are presented in figure
5(a) and 5(b). The reference values which served for the comparison of phase correction
measurements were calculated as the non-weighted mean over all measurements for each
gauge block. Note that the guidelines stated that the 2,5, 3,5, 4,5 and 5,5 mm gauges should be
used on platen PLBI7 (or quartz platen A) for one phase stack and the 3,0, 4,0, 5,0 and 6,0 mm
gauges should be used on platen PLBI8 (or quartz platen B). For this reason if the phase stack
method has been used the values for a phase correction will be the same for each gauge in the
stack. This is clearly shown figures 5(a) and 5(b) where the relative dispositions of the results
are the same for all gauges measured on the same platen.
-25-20-15-10
-505
10152025
phas
e c
orre
ctio
n /
nm
3.5 mm gauge block - steel platen PLBI7
Mean phase correction = -12.6 nmStandard deviation = 8 nm
-25-20-15-10
-505
10152025
Dev
iatio
n f
rom
m
ean
4.5 mm gauge block - steel platen PLBI7
Mean phase correction = -12.6 nmStandard deviation = 9 nm
-25-20-15-10-505
10152025 5.5 mm gauge block - steel platen PLBI7
NPL1
OFM
ETD
FM
CEM
BNM
-LNE
IPQ
NPL2
CM
A
JV IMG
C
PTB
CSIR
O
NPL3
Mean phase correction = -12.7 nmStandard deviation = 8 nm
Time (days)
-25-20-15-10-505
10152025
0 200 400 600 800
2.5 mm gauge block - steel platen PLBI7
Mean phase correction = -12.9 nmStandard deviation = 7 nm
Figure 5(a) Deviations from mean phase correction for gauge blocks measured on steel platen PLBI7 where � denotes the stack method for measuring phase correction, ∆ denotes the TIS method for measuring phase correction and � denotes the stylus method for measuring phase correction.
Time (days)
-25-20-15-10
-505
10152025
0 200 400 600 800
3.0 mm gauge block - steel platen PLBI8
Mean phase correction = -11.9 nmStandard deviation = 6 nm
-25-20-15-10-505
10152025
phas
e c
orre
ctio
n /
nm
4.0 mm gauge block - steel platen PLBI8
Mean phase correction = -11.9 nmStandard deviation = 6 nm
-25-20-15-10-505
10152025
Devi
atio
n f
rom
m
ean
5.0 mm gauge block - steel platen PLBI8
Mean phase correction = -12.0 nmStandard deviation = 6 nm
-25-20-15-10
-505
10152025
6.0 mm gauge block - steel platen PLBI8
NPL1
OFM
ET
DFM
CEM
BNM
-LNE
IPQN
PL2
CM
AJV IM
GC
PTB
CSIR
O
NPL3
Mean phase correction = -12.0 nmStandard deviation = 6 nm
Figure 5(b) Deviations from mean phase correction for gauge blocks measured on steel platen PLBI8 where � denotes the stack method for measuring phase correction, ∆ denotes the TIS method for measuring phase correction and � denotes the stylus method for measuring phase correction.
The results for the phase corrections for the gauges on quartz platens are shown in figures 6(a)
and 6(b).
Time (days)
-30-25-20-15-10-505
10152025
0 100 200 300 400 500 600 700 800
2.5 mm gauge block - quartz platen A
Mean phase correction = 32.6 nmStandard deviation = 10 nm
-30-25-20-15-10-505
10152025
phas
e c
orre
ctio
n /
nm
3.5 mm gauge block - quartz platen A
Mean phase correction = 31.6 nmStandard deviation = 10 nm
-30-25-20-15-10
-505
10152025
Dev
iatio
n f
rom
m
ean
4.5 mm gauge block - quartz platen A
Mean phase correction = 32.5 nmStandard deviation = 10 nm
-30-25-20-15-10-505
10152025 5.5 mm gauge block - quartz platen A
NPL1
CEM
IPQ
NIST
CM
AJV IM
GC
CSIR
O
NPL3
Mean phase correction = 32.5 nmStandard deviation = 10 nm
Figure 6(a) Deviations from mean phase correction for gauge blocks measured on quartz platen A where � denotes the stack method for measuring phase correction, � denotes the NIST method for measuring phase correction and � denotes the stylus method for measuring phase correction.
Time (days)
-30-25-20-15-10-505
10152025
0 200 400 600 800
3.0 mm gauge block - quartz platen B
Mean phase correction = 29.9 nmStandard deviation = 10 nm
-30-25-20-15-10-505
10152025
phas
e c
orre
ctio
n /
nm
4.0 mm gauge block - quartz platen B
Mean phase correction = 30.2 nmStandard deviation = 10 nm
-30-25-20-15-10
-505
10152025
Dev
iatio
n f
rom
m
ean
5.0 mm gauge block - quartz platen B
Mean phase correction = 29.3 nmStandard deviation = 10 nm
-30-25-20-15-10-505
10152025 6.0 mm gauge block - quartz platen B
NPL
DFM
CEM
IPQ
NIST
CM
A
JV IMG
C
CSIR
O
NPL3
Mean phase correction = 29.6 nmStandard deviation = 10 nm
Figure 6(b) Deviations from mean phase correction for gauge blocks measured on quartz platen B where � denotes the stack method for measuring phase correction, � denotes the NIST method for measuring phase correction and � denotes the stylus method for measuring phase correction.
7 Conclusions
If the assumed linear secular change in length is removed from the results then, at 95%
confidence level, the deviations from nominal length are in agreement for all of the
measurement techniques used on all of the gauge blocks. It is reasonable to make this
assumption based on examination on figures 1(a) and 1(b). This shows that the choice of platen
material (quartz or steel) and surface roughness measurement method (stack, scattering or
mechanical stylus) does not appear to effect the overall result. The only exceptions to this
statement are one result for the 5,5 mm gauge block and the final result on the 3,5 mm gauge
block measured at NPL which shows clear evidence of significant changes to the surface that
have occurred during the project.
The largest difference between phase corrections determined by the participants was
approximately 30 nm. This is a relatively large figure compared to the interferometric length
measurement uncertainties which are usually quoted. There is no obvious difference in the
variability of the values for a phase correction for the quartz and steel platens.
In general the results obtained with scattering instruments are in good agreement with the
results obtained using the phase stack technique. The exceptions to this are the final NPL results
where surface changes invalidate this comparison and all NPL measurements on platen PLBI7.
The reason for this latter disagreement is not fully understood at this stage, but may be due to
the reliance of the NPL TIS method on the type of steel (or finishing technique) that is used
during its calibration.
Acknowledgements
The authors would like to thank all their colleagues in the participating laboratories: Kari Riski
(CMA), Joaquín Rodríguez (CEM), Nick Brown (CSIRO), Geneviève Lipinski (BNM-LNE), Jes
Henningsen (DFM), Maria Paola Sassi (IMGC), Helge Karlsson (JV), John Stoup (NIST),
Gerhard Bönsch (PTB), Fernanda Saraiva (IPQ) and Reudi Thalmann (OFMET) for performing
the measurements, supplying all the results and providing valuable comments. They would
also like to thank their colleagues in the Dimensional Metrology Section at NPL: Teresa
Grimshaw, Andrew Lewis and Keith Jackson.
References
ISO 3650: 1998 Geometrical product specifications (GPS) - Length standards - Gauge blocks
(ISO: Geneva)
Birch K P, Downs M J 1993 An updated Edlén equation for the refractive index of air Metrologia
30 155-162
Bönsch G 1998 Interferometric calibration of an integrating sphere for determination of the
roughness correction for gauge blocks Proc. SPIE 3477 152-160
Bönsch G, Potulski E 1998 Measurement of the refractive index of air and comparison with the
modified Edlén’s formulae Metrologia 35 133-139
Hume K J 1963 Engineering Metrology (Macdonald: London)
Ikonen E, Riski K 1993 Gauge block interferometer based on one stabilised laser and a white-
light source Metrologia 30 95-104
Leach R K 1998 Measurement of a correction for the phase change on reflection due to surface
roughness Proc. SPIE 3477 138-151
Pugh D J, Jackson K Automatic gauge block measurement using multiple wavelength
interferometry Proc. SPIE 656 244-250
Stoup J R, Faust B S, Doiron T D 1998 Minimising errors in phase change correction
measurements for gauge blocks using a spherical contact technique Proc. SPIE 3477 161-172
Thwaite E G 1977 Phase correction in the interferometric measurement of end standards
Metrologia 14 53-62