comparison of pressurized solvent and reflux extraction methods for the determination of...
TRANSCRIPT
Comparison of pressurized solvent and reflux extraction methods for thedetermination of perfluorooctanoic acid in polytetrafluoroethylenepolymers using LC-MS-MS
Barbara S. Larsen,a Mary A. Kaiser,a Miguel Botelho,a Gilbert R. Woolera and L. William Buxtonb
Received 16th August 2004, Accepted 4th October 2004
First published as an Advance Article on the web 24th November 2004
DOI: 10.1039/b412609b
Both pressurized solvent extraction (PSE) and reflux extraction in various solvents were used to
select the most efficient system for the determination of the quantity of perfluorooctanoic acid
(PFOA) present in polytetrafluoroethylene polymers. After evaporating the solvent, PFOA was
determined using liquid chromatography tandem mass spectrometry. Ethanol, water and
methanol gave comparable results and were shown to be good solvents for this extraction.
Acetonitrile was a reasonable solvent using the reflux extraction method, but not with PSE.
Chloroform resulted in poor recovery for both extraction methods. PSE proved to be the more
efficient extraction method.
Introduction
In 1999 Moody et al. reported the presence of perfluorinated
compounds in groundwater impacted by fire-fighting
activities, even though the activities had concluded many
years before the analyses were conducted.1 Several investiga-
tors have reported that human sera contained low concentra-
tions (viz., ppb) of perfluorinated organic molecules [e.g.,
perfluorooctane sulfonate (PFOS), perfluorooctanoic acid
(PFOA)] in the general population.2–5 Other reports showed
that low levels of these and similar compounds could be found
in wildlife and in the environment.6–8
Organofluorine compounds have unusual properties that
make them difficult to measure.9 In addition, perfluorocar-
boxylates are used in the production of many fluoropolymers10
as a processing aid.11,12 Fluoropolymers are important inert
components commonly used in laboratory apparatus and
analytical laboratory instrumentation. If residual perfluoro-
carboxylates are present in analytical systems, their levels and
ease of extraction will be important in deciding the low levels
of quantitation needed, especially for exposure, environmental,
and toxicological studies.
In this study, we compare solvents and extraction methods
used for the determination of perfluorooctanoic acid (PFOA),
a common fluoropolymer processing aid, in two polytetra-
fluoroethylene (PTFE) polymers.
Based on the work of Vandenburg et al.,13 we selected
pressurized solvent extraction and boiling under reflux to
determine optimal conditions for these extraction measure-
ments. Since PFOA is quite soluble in water14 and since water
is a good solvent used to ascertain potential exposure to PFOA
from perfluoropolymers, water was selected for this study.
Ethanol has been used for studies involving food contact in
FDA studies.15 Acetonitrile and methanol are common
solvents. Chloroform was evaluated since it is a common
halogenated solvent.
Experimental
Apparatus
Reflux extractions were done with a 500-mL round-bottom
flask with glass magnetic stirring bar and a glass water
condenser. The pressurized solvent extraction was done with a
Dionex (Sunnyvale, CA) accelerated solvent extractor (ASEH
series 200).
Materials and reagents
Polytetrafluoroethylene fluoropolymer resins were obtained
from a commercial lot and an intermediate of a commercial lot
(average particle size approximately 500 mm). Liquid chroma-
tographic grade or analytical grade water, methanol, acetoni-
trile, chloroform and standard Ottawa sand were obtained
from EMD Chemicals, Inc. (Darmstadt, Germany). Ethanol
(100%) was obtained from Sigma–Aldrich (Milwaukee, WI). A
dual 13C enriched perfluorooctanoic acid was synthesized
in house.
Standards for the seven-point calibration curve were pre-
pared by dilution with methanol from a 1000 ppb (mg L21)
standard solution to concentrations of 0.5, 1, 5, 10, 25, 50, and
100 ppb.
Reflux extraction procedure
Five reflux extraction systems were run concurrently. One
hundred mL of solvent was added to each 500-mL round-
bottom flask with a glass stirring bar and reflux water
condenser. The solvent was refluxed for 20 min and the
solvent discarded. Then, 250 mL of the solvent of interest was
added to each flask. To one of the flasks was added 1.0 mL of
the recovery check standard that was subsequently used to
determine percent recovery of PFOA. Approximately 0.75 g of
each polymer was weighed (¡ 0.1 mg) and added to three of
the remaining flasks. The fifth flask had only solvent, which
PAPER www.rsc.org/analyst | The Analyst
This journal is � The Royal Society of Chemistry 2005 Analyst, 2005, 130, 59–62 | 59
Publ
ishe
d on
24
Nov
embe
r 20
04. D
ownl
oade
d on
28/
10/2
014
07:0
2:57
. View Article Online / Journal Homepage / Table of Contents for this issue
served as the solvent blank. The solvents in the prepared flasks
were refluxed for two hours, cooled and filtered through a
no. 40 filter paper (Whatman, Clifton, NJ). The solvents were
then evaporated on a Buchi rotary evaporator (Brinkmann,
Westbury, NY) to remove most of the solvent (approximately
10 mL remaining). The remaining solvent was then transferred
to either a 30 mL or 60 mL disposable collection vial and
evaporated to dryness under nitrogen on a temperature-
controlled heating block (ambient for chloroform, 45 uC for
methanol, 55 uC for ethanol and acetonitrile, and 85 uC for
water).
Pressurized solvent extraction procedure
Five pressurized solvent extractions were performed concur-
rently. The instrument’s solvent reservoir was filled with the
appropriate solvent and the lines rinsed four times with
approximately 5 mL of solvent. The cells were filled with
Ottawa sand to approximately 3 mm from the top. Each cell
was closed and loaded onto the instrument. The cells with sand
were preconditioned. The preconditioning cycle was set to
1500 psi (10.3 kPa), 150 uC, 7 min heating time, 100% volume
flush, 240 second purge for one cycle. The solvents were
then discarded and new collection vessels set in place.
Approximately 5/6 of the sand was removed from three of
the vessels and a weighed quantity of polymer added to each
vessel. Each cell was refilled to 3 mm from the top with the
sand and the lid secured. To the fourth sand-filled cell 1.0 mL
of recovery check standard was added on top of the sand and
the lid secured. The fifth sand filled flask served as the solvent
blank. The extraction conditions were set to 1600 psi
(11.0 kPa), 150 uC, 7 min heating time, 100% volume flush,
240 s purge for 4 cycles. The collection vials were removed
when the extraction was complete and placed in a heating
block where they were evaporated to dryness under nitrogen
(ambient for chloroform, 45 uC for methanol, 55 uC for
ethanol and acetonitrile and 85 uC for water).
Analytical method
The concentration of PFOA in the reconstituted extracts was
determined using high performance liquid chromatography
coupled with negative ion electrospray tandem mass spectro-
metry (LC-MS-MS) (Micromass Quatro Ultima, Beverly,
MA). The extracts were reconstituted by adding 1 mL of
100% methanol to the vial and shaking for 30 min on a wrist-
action shaker. The methanol was then placed into a 5-mL
volumetric flask, and brought to volume with the LC mobile
phase A, as described below. The dual 13C-enriched standard
(final concentration 50 ppb) was added as an internal standard
to all the reconstituted samples.
The analyte was separated using an Agilent 1100 liquid chro-
matograph (Wilmington, DE) modified with low dead-volume
internal tubing. A guard column, Hypersil C18 2 6 50 mm
(Thermo Keystone, Bellefonte, PA), was installed between
the mixer and the autoinjector. Twenty-five microliters of
the reconstituted extract were injected onto a Hypersil ODS
2.1 6 200 mm (Thermo Keystone, Bellefonte, PA) at a flow
rate of 0.3 mL min21 and maintained at 60 uC. Duplicate
injections were made for all samples.
Initial gradient conditions were 15% mobile phase B, where
mobile phase A is 2 mM ammonium acetate–1% methanol and
B is 100% methanol. A linear gradient was used from 15–67%
B over 16 min. The conditions were returned to 15% B for two
additional minutes. Typical elution time for PFOA was
approximately 16.5 min. Fig. 1 shows a typical total ion
chromatogram of the internal standard and the PFOA analyte.
In negative ion mode, PFOA is observed as an anion at
413 [CF3(CF2)6COO2]. The internal standard is observed at
415 [CF3(CF2)513CF2
13COO2].
Quantitative analysis was performed using selected ion
monitoring for the transition of 413 . 369 (loss of CO2) for
the analyte and 415 . 370 (loss of 13CO2) for the internal
standard. Any samples that fell outside the calibration were
diluted appropriately and reanalyzed. A seven-point linear
calibration curve was prepared (not including zero) for
external calibration. Each calibrant concentration set was
run in duplicate and bracketed the samples. A typical
calibration curve consisted of all fourteen calibration points.
With no weighting, the acceptance criterion for the calibration
curve required a correlation coefficient, R, ¢0.992.
A methanol blank was run after each 100 ppb standard. The
acceptance criterion was set so that the lowest level PFOA
standard’s area was at least five-times greater than the area of
the methanol (solvent) blank. (No quantifiable area was
observed in the blank.)
The limit of quantitation (LOQ) was set at 0.5 ppb, the
concentration of the lowest calibration standard. A signal
between the blank background and the LOQ was defined as
not quantifiable (NQ).
Results and discussion
In the extraction of the PTFE polymers with the various
solvents, a recovery check standard was run. The recovery
check standard vessel contained just the solvent with
500 ng mL21 of PFOA added in order to determine the
recovery of PFOA in the solvent. Table 1 summarizes the
results of the recovery check standard for each solvent used in
this study. The recoveries of PFOA for methanol, water and
ethanol for both extraction methods were acceptable (80–
120%). With acetonitrile acceptable recovery was observed
using reflux extraction: however, using PSE as the extraction
system, acetonitrile had no detectable recovery of PFOA. The
results indicated that there was no recovery of PFOA using
chloroform by either the reflux extraction or pressurized
solvent extraction.
The initial extraction results with chloroform stabilized with
1% hexane were unexpected and additional trials using
different stabilizers were performed. The chloroform modified
with 1% or 5% ethanol did not extract PFOA quantitatively or
reproducibly. The initial extraction in the PSE was followed by
a re-extraction with methanol, showing an improvement in the
recovery to 57%. This indicates that the chloroform extraction
was ineffective. Recoveries for the water and alcohols ranged
from 89% to 104%.
To ensure that the low sample recoveries observed in
chloroform were not a result of loss during evaporation of the
solvent, a sample of PFOA (500 ng mL21) in chloroform was
60 | Analyst, 2005, 130, 59–62 This journal is � The Royal Society of Chemistry 2005
Publ
ishe
d on
24
Nov
embe
r 20
04. D
ownl
oade
d on
28/
10/2
014
07:0
2:57
. View Article Online
dried overnight in the hood. An identical sample was
evaporated to dryness with nitrogen and heat. These samples
were reconstituted and analyzed by the analytical method. The
observed recoveries were similar, 103% and 94%, respectively.
This indicates that the low recoveries in chloroform extracts
are not the result of losses during the evaporation or
reconstitution process, but are due to inefficient extraction of
PFOA in chloroform stabilized with either 1% hexane or 1%
ethanol. The experiment with chloroform containing 5%
ethanol showed modest recovery, indicating that the ethanol
plays a role in the extraction efficiency.
Tables 2 and 3 show data for the spike recovery and the
extraction results for the two different fluoropolymers (PTFE)
comparing reflux extraction with pressurized solvent extrac-
tion. The results in Table 2 are comparable to those shown in
Table 1. For water and alcohols under these conditions, PSE
percent recoveries were generally the same as the reflux
extraction method. The amount of analyte extracted and
detected using the reflux extraction method appeared to be the
same level for the water and alcohols (Table 3). The level of
analyte detected using the water and alcohols was greater
under the same PSE conditions than that with acetonitrile and
chloroform. PSE is the more efficient extraction method for
determining PFOA content in PTFE polymers (Table 3). The
PTFE sample labeled polymer I is from a commercial lot and
polymer II is the intermediate of a commercial lot, showing
that the finishing steps significantly reduce the PFOA
concentration in the commercial polymer. When the chloro-
form extraction was followed by methanol extraction and the
two quantities added together (Polymer II, Table 3), the
combined solvent results agreed with the water and alcohols
extraction results under the same PSE conditions (1580 ppb).
Table 1 Check standard percent recovery
Solventa Refluxb PSEb,c
Methanol 97 95Acetonitrile 89 4.1 (0.31)Ethanol 104 99Water 103 96Chloroform–1% hexane 0.9 2.8 (0.06)Chloroform–1% ethanol 60* 0 (58)Chloroform–5% ethanol 42* 45 (11)a Chloroform was purchased with either 1% hexane or 1% ethanol asthe stabilizer and the % ethanol was increased by volume to 5%.b Results are an average of two measurments and * denotes anaverage of three measurements. c A second solvent extraction withmethanol was made with the extraction efficiency indicated in theparenthesis
Table 2 Check standard percent recovery from PTFE samples
SolventRefluxtemperature/uC
Polymer I Polymer II
Reflux PSE Reflux PSE
Methanol 65 93.1 88.4 92.9 89.2Acetonitrile 82 91.3 0 86.4 7.9Ethanol 78 99.1 85.5 104 92.5Water 100 104 104 104 106Chloroform 61 3.1a 0a 36.1b 59.5b
a Chloroform stabilized with 1% hexane was used for theseextractions. b The results reported is the sum of the initial extractionusing chloroform stabilized with 1% ethanol followed by a secondextraction with methanol.
Table 3 Concentration of PFOA (ppb) extracted from PTFEpolymers
SolventRefluxtemperature/uC
Polymer I Polymer II
Reflux PSE Reflux PSE
Methanol 65 42 140 485 1420Acetonitrile 82 46 NQa 412 241Ethanol 78 46 135 535 1100Water 100 46 59 493 1040Chloroform 61 NQa,b NQa,b 59c 1580c
a NQ is nonquantifiable, indicating that the observed signal was lessthan the lowest calibrant. b Chloroform stabilized with 1% hexanewas used for these extractions. c The result reported is the sum ofthe initial extraction using chloroform stabilized with 1% ethanolfollowed by a second extraction with methanol.
Fig. 1 LC–MS–MS of a PTFE extract showing in the upper trace the internal standard (415 . 370), lower trace the PFOA analyte (413 . 369).
This journal is � The Royal Society of Chemistry 2005 Analyst, 2005, 130, 59–62 | 61
Publ
ishe
d on
24
Nov
embe
r 20
04. D
ownl
oade
d on
28/
10/2
014
07:0
2:57
. View Article Online
Conclusions
Ethanol, methanol and water efficiently extracted PFOA from
PTFE polymers. PSE was able to extract more analyte in a
shorter time period than reflux extraction with water,
methanol and ethanol. Acetonitrile as the solvent gave
reasonable recoveries with reflux extraction but not with
PSE. Chloroform gave unacceptable results by either extrac-
tion method. This result supports the use of water, ethanol, or
methanol for extracting PFOA from PTFE matrices.
Acknowledgements
The authors thank Charles F. Koerting, Donald A. Wilson
Jr. and Gregory A. Urove for their advice and assistance.
Barbara S. Larsen,a Mary A. Kaiser,a Miguel Botelho,a
Gilbert R. Woolera and L. William Buxtonb
aDuPont Corporate Center for Analytical Sciences, ExperimentalStation 402/5321, P.O. Box 80402, Wilmington, DE 19880-0402, USA.E-mail: [email protected]; Fax: 302 695-1053;Tel: 302 695-8435bDuPont Fluoroproducts, P.O. Box 80713, Wilmington, DE 19880-0713,USA
References
1 C. A. Moody and J. A. Field, Environ. Sci. Technol., 1999, 33,2800.
2 K. J. Hansen, L. A. Clemen, M. E. Ellefson and H. O. Johnson,Environ. Sci. Technol., 2001, 35, 766.
3 Z. Kuklenyik, J. A. Reich, J. S. Tully, L. L. Needham andA. M. Calafat, Environ. Sci. Technol., 2004, 38, 3698.
4 K. Kannan, S. Corsolini, J. Falandz, G. Fillman, K. S. Kumar,B. G. Loganathan, M. A. Mohd, J. Olivero, N. V. Wouwe,J. H. Yang and K. M. Aldous, Environ. Sci. Technol., 2004, 38.
5 K. Inoue, F. Okada, R. Ito, S. Kato, S. Sasaki, S. Nakajima,A. Uno, Y. Saijo, F. Sata, Y. Yoshimura, R. Kishi andH. Nakazawa, Environ. Health Perspect., 2004, 112, 1204.
6 M. M. Schultz, D. F. Barofsky and J. A. Field, Environ. Eng. Sci.,2003, 20, 487.
7 K. Kannan, S. Corsolini, J. Falandysz, G. Oehme, S. Focardi andJ. P. Giesy, Environ. Sci. Technol., 2002, 36, 3210.
8 J. W. Martin, M. M. Smithwick, B. M. Braune, P. F. Hoekstra,D. C. G. Muir and S. A. Mabury, Environ. Sci. Technol., 2004, 38,373.
9 J. W. Martin, S. A. Mabury, K. Kannan, U. Berger, P. De Voogt,J. Field, J. Franklin, J. P. Giesy, T. Harner, D. C. G. Muir,B. Scott, M. A. Kaiser, U. Jarnberg, K. C. Jones, H. Schroeder,M. Simcik, C. Sottani, B. Van Bavel, A. Karrman, G. Lindstromand S. Van Leeuwen, Environ. Sci. Technol., 2004, 38, 248A.
10 J. G. Drobny, Technology of Fluoropolymers, CRC, Boca Raton,FL, 2001.
11 The Society of the Plastics Industry, Inc., Guide to the SafeHandling of Fluoropolymer Dispersions, Washington, DC, 2001.
12 E. Kissa, Fluorinated Surfactants and Repellents, Marcel Dekker,NY, 2nd edn., 2001.
13 H. J. Vandenburg, A. A. Clifford, K. D. Bartle, J. Caroll andI. D. Newton, Analyst, 1999, 124, 397.
14 The Society of the Plastics Industry, Inc., Detecting andQuantifying Low Levels of Fluoropolymer Polymerization Aids–AGuidance Document, Washington, DC, 2003.
15 U.S. Food and Drug Administration, Guidance for Industry:Preparation of Food Contact Notifications and Food AdditivePetitions for Food Contact Substances: Chemistry Recom-mendations, Rockville, MD, 2002.
62 | Analyst, 2005, 130, 59–62 This journal is � The Royal Society of Chemistry 2005
Publ
ishe
d on
24
Nov
embe
r 20
04. D
ownl
oade
d on
28/
10/2
014
07:0
2:57
. View Article Online