compact series · • automation programming in labview, python, matlab, .net, etc. ......

20
Frequency range: 300 kHz - 18 GHz Wide output power adjustment range: -55 dBm to +10 dBm* Dynamic range: 130 dB (10 Hz IF bandwidth) typ. Measurement time per point: 30 μs per point, min typ.* • Up to 16 logical channels with 16 traces each max Automation programming in LabView, Python, MATLAB, .NET, etc. • Models available in 50 Ohm • Up to 200,001 measurement points • Multiple precision calibration methods and automatic calibration ® EXTEND YOUR REACH TM Compact Series M Models * Depending on model USA: +1.317.222.5400 London: +44 75 03 68 21 13 Singapore: +65.6323.6546 Latin America: +1.954.706.5920 631 E. New York St | Indianapolis, IN | 46202 www.coppermountaintech.com

Upload: others

Post on 24-Jul-2020

21 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

• Frequency range: 300 kHz - 18 GHz• Wide output power adjustment range: -55 dBm to +10 dBm*• Dynamic range: 130 dB (10 Hz IF bandwidth) typ.• Measurement time per point: 30 µs per point, min typ.*• Up to 16 logical channels with 16 traces each max

• Automation programming in LabView, Python, MATLAB, .NET, etc.• Models available in 50 Ohm• Up to 200,001 measurement points• Multiple precision calibration methods and automatic calibration

®

E X T E N D Y O U R R E A C H T M

Compact SeriesM Models

* Depending on model

USA: +1.317.222.5400London: +44 75 03 68 21 13

Singapore: +65.6323.6546Latin America: +1.954.706.5920

631 E. New York St | Indianapolis, IN | 46202www.coppermountaintech.com

Page 2: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

Small Size, Basic Software Features

M Series VNAs deliver metrology-grade performance in a more economical package that excludes a number of advanced features: Vector Mixer Calibration, TRL Calibration, Frequency Offset, Time Domain, and Gating. These Vector Network Analyzers are small, can be powered by battery, and are ideal for use in laboratory and production testing in a variety of applications including filter tuning, antenna test and characterization, amplifier testing, etc.

Copper Mountain Technologies’ USB VNAs are next generation analyzers designed to meet the needs of 21st Century engineers. Our VNAs include an RF measurement module and a processing module, a software application which runs on a Windows or Linux PC, laptop, or tablet, connecting to the measurement hardware via USB interface.

This innovative approach delivers high measurement accuracy and enables users to take advantage of faster processors, newer computers and larger displays. USB VNAs have lower Total Cost of Ownership and fewer potential failure points. These instruments are smaller and lighter, can go almost anywhere, are very easy to share and eliminate the need for data purging or hard drive removal in secure environments.

1

Page 3: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

“CMT devices are lightweight, compact and a necessary tool for technical sales or engineers on the go. The software interface allows users to test RF products with any standard computer system. This is a revolution and a relief in terms of space occupied in the lab, measurement reliability and dynamic range. CMT provides the highest level of timely and attentive customer care.”

Jessy CavazosIndustry Director, Frost & Sullivan

The Whole Solution

All our products come with a standard three-year warranty from date of shipment. During that time we will repair or replace any product malfunctioning due to defective parts or labor.

While we pride ourselves on quality of our instruments, should your VNA malfunction for any reason, we will gladly offer a loaner unit while we service yours. With our USB VNAs where all data is stored on your PC, a simple swap of the measurement module assures uninterrupted workflow. So you will experience little or no downtime.

Warranty, Service, & Repairs

Our team of applications engineers, service technicians, and metrology scientists are here to help you with technical support, application-specific recommendations, annual performance testing, and troubleshooting or repair of your CMT instruments.

Our engineers will work with your team to augment your in-house capabilities. We can write custom applications and test software, develop test automation scripts and help with integrated RF system testing. We can design and provide an RF switching network specific to your requirements; electro-mechanical, solid-state, or PIN diode-based. If the S-parameter measurement fixture involves challenging conditions for repeatability and accuracy we can assist with measurement uncertainty analysis.

An extensive library of technical materials including application notes, tips on performing VNA measurements, sample automation scripts, and how-to videos are available on our website www.coppermountaintech.com and YouTube channel, CopperMountainTech.

Our engineers are an extension of your team

Copper Mountain Technologies’ Indianapolis calibration laboratory is accredited in accordance with the recognized international standard ISO/IEC 17025:2017 and meets the requirements of ANSI/NCSL Z540-1994-1. All reference standards and equipment in the laboratory are traceable to National Institute of Standards and Technology (NIST) or international equivalent.

Should you prefer to perform the annual testing yourself or use a third party, contact us for information or questions on performing these procedures. Additionally, the VNA Performance Test (VNAPT) software application is available for third party laboratories without restriction. Use of VNAPT to execute performance tests is optional, but the software is designed to automate and streamline VNA performance testing, including automatic generation of test reports. Please contact Copper Mountain Technologies or your local distributor for recommended calibration options.

Annual Calibration

2

Page 4: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

Software Application

The software application takes raw measurement data from the data acquisition (measurement) module and recalculates into S-parameters in multiple presentation formats utilizing proprietary algorithms. These new and advanced calibration and other accuracy enhancing algorithms were developed by our metrology experts. Our software can be downloaded free from our website, used on an unlimited number of PCs using either Linux or Windows operating systems, and enables easy VNA integration with other software applications and automation.

The software application features a fully functioning Demo Mode, which can be used for exploring VNAs’ features and capabilities without an actual measurement module connected to your PC.

Software application is part of the VNA

Measured parameters: S11, S21, S12, S22All models also measure absolute power of the reference and received signals at the port.

Number of measurement channelsUp to 16 independent logical channels: each logical channel is represented on the screen as an individual channel window. A logical channel is defined by such stimulus signal settings as frequency range, number of test points, or power level.

Data tracesUp to 16 data traces can be displayed in each channel window. A data trace represents one of such parameters of the DUT as S-parameters or input power response. Memory tracesEach of the 16 data traces can be saved into memory for further comparison with the current values.

Data display formatsLogarithmic magnitude, linear magnitude, phase, expanded phase, group delay, SWR, real part, imaginary part, Smith chart diagram and polar diagram display formats are available.

Measurement Capabilities

3

Page 5: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

Dynamic Range

AutoscalingAutomatic selection of scale division and reference level value to have the trace most effectively displayed.

Electrical delayCalibration plane moving to compensate for the delay in the test setup. Compensation for electrical delay in device under test (DUT) during measurements of deviation from linear phase.

Phase offsetDefined in degrees.

Sweep FeaturesSweep type: Linear frequency sweep and logarithmic frequency sweep are performed with fixed output power. Linear power sweep is performed at a fixed frequency.Measured points per sweep: Set by the user from 2 to 200,001.Segment sweep features: A frequency sweep within several independent user-defined segments. Frequency range, number of sweep points, source power, and IF bandwidth can be set for each segment.Output Power: Source power from -55 dBm to +5 dBm with a resolution of 0.05 dB. In frequency sweep mode power slope can be set up to 2 dB/GHz to compensate for high frequency attenuation in fixture cables.Sweep Trigger: Trigger modes: continuous, single, or hold. Trigger sources: internal, manual, external, bus.

4

- 135 dB - 132 dB

Trace FunctionsTrace displayData trace, memory trace, or simultaneous indication of data and memory traces.

Trace mathData trace modification by math operations: addition, subtraction, multiplication or division of measured complex values and memory data.

Typical dynamic range of 130 dB is achieved from 300 kHz through the top of the frequency range (at 10 Hz IF bandwidth). Seen here is the maximum dynamic range achieved when using IFBW 1 Hz and an output power level of 5 dBm.

Page 6: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

Software Application

5

Limit testing is a function for automatic pass/fail based on measurement results. Pass/fail is based on comparison of the trace to the limit line set by the user and can consist of one or several segments.

Each segment checks the measurement value for failing either the upper or lower limit, or both. The limit line segment is defined by specifying the coordinates of the beginning (X0, Y0) and the end (X1, Y1) of the segment, and type of the limit. The MAX or MIN limit types check if the trace falls outside of the upper or lower limit, respectively.

Limit Testing

Frequency Scan SegmentationThe VNA has a large frequency range with the option of frequency scan segmentation. This allows for optimal use of the device to realize the maximum dynamic range while maintaining high measurement speed.

The power sweep feature turns compression point recognition, one of the most fundamental and complex amplified measurements, into a simple and accurate operation.

Power Scanning & Compression Point Recognition

Page 7: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

Allows the user to mathematically simulate the DUT parameters after virtual connection through a fixture circuit between the calibration plane and the DUT. This circuit is described by an S-parameter matrix in a Touchstone file.

EmbeddingAllows users to mathematically exclude from the measurement result the effect of the fixture circuit connected between the calibration plane and a DUT. This circuit should be described by an S-parameter matrix in a Touchstone file.

De-Embedding

This function converts the S-parameters measured at a 50Ω port into values which would be seen if measured at a test port with arbitrary impedance.

Port Impedance ConversionThis function allows for conversion of measured S-parameters to the following parameters: reflection impedance and admittance, transmission impedance and admittance, and inverse S-parameters.

S-Parameter Conversion

Page 8: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

7

Data OutputAnalyzer StateAll state, calibration and measurement data can be saved to an Analyzer state file on the hard disk and later recalled into the software program. The following four types of states are available: State, State & Cal, Stat & Trace, or All.

Channel StateA channel state can be saved into the Analyzer state. The procedure is similar to saving of the Analyzer state, and the same types are applied to channel state saving. Unlike Analyzer state, channel state is saved into the Analyzer volatile memory (not to the hard disk) and is cleared when power to the Analyzer is switched off. For channel state, there are four memory registers A, B, C, D. Channel state saving allows the user to easily copy the settings of one channel to another one.

Trace Data CSV FileThe Analyzer allows the user to save an individual trace’s data as a CSV file (comma separated values). The active trace stimulus and response values, in its current format are saved to a *.CSV file.

Trace Data Touchstone FileAllows the user to save S-parameters to a Touchstone file. The Touchstone file contains frequency values and S-parameters. Files of this format are industry-standard for most circuit simulator programs. The .s2p files are used for saving all S-parameters of a device. The .s1p.s1p files are used for saving S11 or S22 parameters of a 1-port device. The Touchstone file saving function is applied to individual channels. In addition, the software can be used as a Touchstone file viewer, which allows the user to graphically display and work with previously saved Touchstone files.

Screenshot captureA print function is provided with a preview feature, which allows for viewing the image to be printed on the

screen, and/or save it to a file. Screenshots can be printed using three different applications: MS Word, Image Viewer for Windows, or the Print Wizard of the Analyzer. Each screenshot can be printed in color, grayscale, black and white, or inverted for visibility or to save ink. The current date and time can be added to each capture before it is transferred to the printing application, resulting in quick and easy test reporting.

Software Application

Page 9: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

8

Calibration

CalibrationCalibration of a test setup (which includes the VNA, cables, and adapters) significantly increases the accuracy of measurements. Calibration allows for correction of errors caused by imperfections in the measurement system: system directivity, source and load match, tracking, and isolation.

Calibration methodsThe following calibration methods of various sophistication and accuracy are available:

• Reflection & transmission normalization• Full one-port calibration • One-path two-port calibration• Full two-port calibration

Reflection and transmission normalizationThis is the simplest calibration method; however, it provides reduced accuracy compared to other methods.

Full one-port calibrationMethod of calibration performed for one-port reflection measurements. It ensures high accuracy.

One-path two-port calibrationMethod of calibration performed for reflection and one-way transmission measurements, for example for measuring S11 and S21 only. It ensures high accuracy for reflection measurements, and moderate accuracy for transmission measurements.

Full two-port calibrationThis method of calibration is performed for full S-parameter matrix measurement of a two-port DUT, ensuring high accuracy.

Mechanical Calibration KitsThe user can select one of the predefined calibration kits of various manufacturers or define a new calibration kit.

User CalibrationAutomatic Calibration ModulesElectronic, or automatic, calibration modules offered by CMT make calibration faster and easier than traditional mechanical calibration.

Sliding load calibration standardThe use of a sliding load calibration standard allows for a significant increase in calibration accuracy at high frequencies compared to the fixed load calibration standard.

“Unknown” thru calibration standardThe use of a generic two-port reciprocal circuit instead of a characterized Thru in full two-port calibration allows the user to calibrate the VNA for measurement of “non-insertable” devices.

Defining of calibration standards Different methods of calibration standard definition are available: standard definition by polynomial model and standard definition by data (S-parameters).

Error correction interpolationWhen the user changes any settings such as the start/stop frequencies or the number of sweep points, compared to the settings at the moment of calibration, interpolation or extrapolation of the calibration coefficients will be applied.

Power calibrationPower calibration allows more stable power level setting at the DUT input. An external power meter should be connected to the USB port directly or via a USB/GPIB adapter.

Receiver calibrationThis method calibrates the receiver gain at the absolute signal power measurement.

Page 10: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

9

Automation

Measurement Automation

COM/DCOM interfaceThe VNA software provides a COM/DCOM (ActiveX) interface, allowing the instrument to be used as a part of a larger test system and in other specialized applications. The VNA program runs as a COM/DCOM server, while the user program runs as a client. COM/DCOM is able to be used with Windows OS only.

SCPI via TCP SocketAlternatively a TCP socket is provided for automation from either localhost--the same machine running the VNA software application--or from a second PC connected by an IP network. The SCPI command is largely compatible with legacy instruments, maximizing code reuse for existing test automation platforms. SCPI via TCP Socket is able to be used with either Windows or Linux operating systems.

SCPI via HiSlip Based on VXI-11, the HiSlip interface uses the same SCPI command set but further allows for instrument discovery and provides ease of automation through Visa library of your choice. SCPI via HiSlip is able to be used with either Windows or Linux operating systems.

LabVIEW compatibleThe device and its software are fully compatible with LabView applications, for ultimate flexibility in user-generated programming and automation. LabVIEW is able to be used with Windows OS only.

Our command set is modeled after industry-standard legacy equipment; porting code is straightforward and we can help. Complete installation of any CMT software comes with multiple programming examples and guides installed in the C:\VNA\S2VNA\ Programming Examples and Guides directory on Windows or ~/Documents/VNA directory on Linux.

CMT software includes many features that other vendors offer as options, including S-parameter Embedding and De-Embedding. No integrated PC means faster data processing turnaround and regular updates that are easy to install. Less complexity in the VNA leads to fewer points of failure that cost you production/development time.

Plugins can add wide ranges of functionality and can be developed upon request. Examples include streamlined production applications, functionality to trigger with external generators, and virtual circuit matching modeling.

• Segmented frequency sweeps• Linear/logarithmic sweeps

Automation Features• 16 channels max. with up to 16 traces each• Marker math

• Power sweeps• Multiple trace formats

• Limit tests

We maintain code examples and guides in the following languages:• MATLAB• Visual Basic (Excel)

Automation Languages

• C++*• Python*

• LabVIEW• And many more

*Available for use with Linux operating system

Page 11: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

10

Software Plug-InsAll Copper Mountain Technologies VNAs include support for executable software add-on modules or plug-ins. With plug-ins, customers and CMT support engineers can develop extensions to the base software launched from inside the main application menu. Place your executable into the /Plug-ins/ subfolder of your VNA’s installation path, and then use the System->Plug-ins menu sequence to launch.

Most plug-ins are developed based on specific customer’s needs. We also offer source code for many plug-ins to help you get started with creating your own plug-ins or as a jumping off point for automation projects.

Our most popular plug-in, Manufacturing Test, supports incorporating VNA software into your manufacturing QMS:

• Streamline production test processes.• Ensure consistency of test process across multiple operators and workstations.• Easily create and manage pass/fail limits across multiple workstations. Pass/fail limits and instrument configuration are

stored in a human-readable plaintext “specifications” file which can be maintained by an authorized test engineer.• Organize test results for subsequent retrieval and analysis.

Page 12: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

11

M5045 Specifications1

160 mm

44 mm

Impedance 50 OhmTest port connector type N, femaleNumber of test ports 2Frequency range 300 kHz to 4.5 GHzFull frequency accuracy ±5·10⁻⁶Frequency resolution 1 HzNumber of measurement points 2 to 200,001Measurement bandwidths (with 1/1.5/2/3/5/7 steps) 1 Hz to 100 kHzDynamic range² 125 dB (130 dB typ.)

Primary Specifications

Accuracy of transmission measurements⁴ Magnitude / Phase0 dB to +10 dB ±0.2 dB / ±2° -45 dB to 0 dB ±0.1 dB / ±1°

-65 dB to -45 dB ±0.2 dB / ±2° -85 dB to -65 dB ±1.0 dB / ±6°

Accuracy of reflection measurements⁵ Magnitude / Phase -15 dB to 0 dB ±0.4 dB / ±3°

-25 dB to -15 dB ±1.0 dB / ±6° -35 dB to -25 dB ±3.0 dB / ±20°

Trace noise magnitude (IF bandwidth 3 kHz) 0.002 dB rmsTemperature dependence 0.02 dB/°C

Measurement Accuracy

Directivity 46 dBSource match 40 dBLoad match 46 dB

Reflection tracking ±0.10 dBTransmission tracking ±0.08 dB

Effective System Data

Image shows actual size

[1] All specifications subject to change without notice. [2] The dynamic range is defined as the difference between the specified maximum power level and the specified noise floor. The specification applies at 10 Hz IF bandwidth. [3] Reflection and transmission measurement accuracy applies over the temperature range of (73 ± 9) °F or (23 ± 5) °C after 40 minutes of warming-up, with less than 1 °C deviation from the full two-port calibration temperature, at output power of -5 dBm. Frequency points have to be identical for measurement and calibration (no interpolation allowed). [4] Transmission specifications are based on a matched DUT, and IF bandwidth of 10 Hz. [5] Reflection specifications are based on an isolating DUT. [6] Specification applies over entire frequency range, at output power of 0 dBm. © Copper Mountain Technologies - www.coppermountaintech.com - Rev. 2020Q2

Power range -55 dBm to +5 dBmPower accuracy ±1.5 dBPower resolution 0.05 dBHarmonic distortion⁶ -20 dBcNon-harmonic spurious⁶ -20 dBc

Test Port Output

Directivity 15 dBSource match 15 dBLoad match 15 dB

Uncorrected System Performance

Noise floor -130 dBm/HzDamage level +23 dBmDamage DC voltage 35 V

Test Port Input

Page 13: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

12

Port 10 MHz Ref In/OutExternal reference frequency 10 MHzInput level -1 dBm to 5 dBmInput impedance 50 OhmConnector type BNC, female

Frequency Reference Input

Port Ext Trig OutMaximum output current 20 mAOutput level

Low level voltage 0.0 VHigh level voltage 3.5 V

Polarity positive or negativeConnector type BNC, female

Trigger Output

Port Ext Trig InInput level

Low threshold voltage 0.5 VHigh threshold voltage 2.7 V

Input level range 0 V to + 5 VPulse width ≥2 µsPolarity positive or negativeInput impedance ≥10 kOhmConnector type BNC, female

Trigger Input

Operating system Windows 7 and aboveCPU frequency 1.5 GHzRAM 1 GBInterface USB 2.0Connector type USB BPower supply 110-240 V, 50/60 HzPower consumption 14 WInput power 9 V DC to 15 V DCInput power consumption DC 12 W

System & Power

Recommended factory adjustment interval 3 years

Factory Adjustment

4

6 3/10 inches

1 3/4 inches

Image shows actual size

Port 10 MHz Ref In/OutInternal reference frequency 10 MHzOutput reference signal level at 50 Ohm impedance 1 dBm to 5 dBmConnector type BNC, female

Frequency Reference Output

Time per point 70 µs typ.Port switchover time 1 ms

Measurement Speed

Operating temperature +5 °C to +40 °C (41 °F to 104 °F)Storage temperature -50 °C to +70 °C (-58 °F to 158 °F)Humidity 90 % at 25 °C (77 °F)Atmospheric pressure 70.0 kPa to 106.7 kPa

Environmental Specifications

Page 14: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

M5065 Specifications1

160 mm

44 mm

Impedance 50 OhmTest port connector type N, femaleNumber of test ports 2Frequency range 300 kHz to 6.5 GHzFull frequency accuracy ±5·10⁻⁶Frequency resolution 1 HzNumber of measurement points 2 to 200,001Measurement bandwidths (with 1/1.5/2/3/5/7 steps) 1 Hz to 100 kHzDynamic range² 125 dB (130 dB typ.)

Primary Specifications

Accuracy of transmission measurements⁴ Magnitude / Phase0 dB to +10 dB ±0.2 dB / ±2° -45 dB to 0 dB ±0.1 dB / ±1°

-65 dB to -45 dB ±0.2 dB / ±2° -85 dB to -65 dB ±1.0 dB / ±6°

Accuracy of reflection measurements⁵ Magnitude / Phase -15 dB to 0 dB ±0.4 dB / ±3°

-25 dB to -15 dB ±1.0 dB / ±6° -35 dB to -25 dB ±3.0 dB / ±20°

Trace noise magnitude (IF bandwidth 3 kHz) 0.002 dB rmsTemperature dependence 0.02 dB/°C

Measurement Accuracy

Directivity 46 dBSource match 40 dBLoad match 46 dB

Reflection tracking ±0.10 dBTransmission tracking ±0.08 dB

Effective System Data

Image shows actual size

[1] All specifications subject to change without notice. [2] The dynamic range is defined as the difference between the specified maximum power level and the specified noise floor. The specification applies at 10 Hz IF bandwidth. [3] Reflection and transmission measurement accuracy applies over the temperature range of (73 ± 9) °F or (23 ± 5) °C after 40 minutes of warming-up, with less than 1 °C deviation from the full two-port calibration temperature, at output power of -5 dBm. Frequency points have to be identical for measurement and calibration (no interpolation allowed). [4] Transmission specifications are based on a matched DUT, and IF bandwidth of 10 Hz. [5] Reflection specifications are based on an isolating DUT. [6] Specification applies over entire frequency range, at output power of 0 dBm. © Copper Mountain Technologies - www.coppermountaintech.com - Rev. 2020Q213

Power range -55 dBm to +5 dBmPower accuracy ±1.5 dBPower resolution 0.05 dBHarmonic distortion⁶ -20 dBcNon-harmonic spurious⁶ -20 dBc

Test Port Output

Directivity 15 dBSource match 15 dBLoad match 15 dB

Uncorrected System Performance

Noise floor -130 dBm/HzDamage level +23 dBmDamage DC voltage 35 V

Test Port Input

Page 15: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

Port 10 MHz Ref In/OutExternal reference frequency 10 MHzInput level -1 dBm to 5 dBmInput impedance 50 OhmConnector type BNC, female

Frequency Reference Input

Port Ext Trig OutMaximum output current 20 mAOutput level

Low level voltage 0.0 VHigh level voltage 3.5 V

Polarity positive or negativeConnector type BNC, female

Trigger Output

Port Ext Trig InInput level

Low threshold voltage 0.5 VHigh threshold voltage 2.7 V

Input level range 0 V to + 5 VPulse width ≥2 µsPolarity positive or negativeInput impedance ≥10 kOhmConnector type BNC, female

Trigger Input

Operating system Windows 7 and aboveCPU frequency 1.0 GHzRAM 512 MBInterface USB 2.0Connector type USB BPower supply 110-240 V, 50/60 HzPower consumption 14 WInput power 9 V DC to 15 V DCInput power consumption DC 12 W

System & Power

Recommended factory adjustment interval 3 years

Factory Adjustment

6 3/10 inches

1 3/4 inches

Image shows actual size

Port 10 MHz Ref In/OutInternal reference frequency 10 MHzOutput reference signal level at 50 Ohm impedance 1 dBm to 5 dBmConnector type BNC, female

Frequency Reference Output

14

Time per point 70 µs typ.Port switchover time 1 ms

Measurement Speed

Operating temperature +5 °C to +40 °C (41 °F to 104 °F)Storage temperature -50 °C to +70 °C (-58 °F to 158 °F)Humidity 90 % at 25 °C (77 °F)Atmospheric pressure 70.0 kPa to 106.7 kPa

Environmental Specifications

Page 16: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

Accuracy of transmission measurements⁴ Magnitude / Phase300 kHz to 6.5 GHz

0 dB to +10 dB ±0.2 dB / ±2° -45 dB to 0 dB ±0.1 dB / ±1°

-65 dB to -45 dB ±0.2 dB / ±2° -85 dB to -65 dB ±1.0 dB / ±6°

6.5 GHz to 8.0 GHz0 dB to +10 dB ±0.2 dB / ±2° -40 dB to 0 dB ±0.1 dB / ±1°

-60 dB to -40 dB ±0.2 dB / ±2° -80 dB to -60 dB ±1.0 dB / ±6°

8.0 GHz to 8.5 GHz0 dB to +10 dB ±0.2 dB / ±2° -35 dB to 0 dB ±0.1 dB / ±1°

-55 dB to -35 dB ±0.2 dB / ±2° -75 dB to -55 dB ±1.0 dB / ±6°

Accuracy of reflection measurements⁵ Magnitude / Phase -15 dB to 0 dB ±0.4 dB / ±3°

-25 dB to -15 dB ±1.0 dB / ±6° -35 dB to -25 dB ±3.0 dB / ±20°

Trace noise magnitude (IF bandwidth 3 kHz) 0.002 dB rmsTemperature dependence 0.02 dB/°C

Measurement Accuracy

M5090 Specifications1

Impedance 50 OhmTest port connector type N, femaleNumber of test ports 2Frequency range 300 kHz to 8.5 GHzFull frequency accuracy ±5·10⁻⁶Frequency resolution 1 HzNumber of measurement points 2 to 200,001Measurement bandwidths (with 1/1.5/2/3/5/7 steps) 1 Hz to 100 kHzDynamic range²

300 kHz to 6.5 GHz 125 dB (130 dB typ.)6.5 GHz to 8.0 GHz 120 dB (125 dB typ.)8.0 GHz to 8.5 GHz 115 dB (120 dB typ.)

Primary Specifications

Directivity 46 dBSource match 40 dBLoad match 46 dB

Reflection tracking ±0.10 dBTransmission tracking ±0.08 dB

Effective System Data

300 kHz to 6.5 GHzDirectivity 15 dB

Source match 15 dBLoad match 15 dB

6.5 GHz to 8.5 GHzDirectivity 12 dB

Source match 15 dBLoad match 15 dB

Uncorrected System Performance

Power range300 kHz to 8.0 GHz -55 dBm to +5 dBm8.0 GHz to 8.5 GHz -55 dBm to +3 dBm

Power accuracy ±1.5 dBPower resolution 0.05 dBHarmonic distortion⁶ -20 dBcNon-harmonic spurious⁶

300 kHz to 6.5 GHz -20 dBc6.5 GHz to 8.5 GHz -15 dBc

Test Port Output

160 mm

44 mm

Image shows actual size

[1] All specifications subject to change without notice. [2] The dynamic range is defined as the difference between the specified maximum power level and the specified noise floor. The specification applies at 10 Hz IF bandwidth. [3] Reflection and transmission measurement accuracy applies over the temperature range of (73 ± 9) °F or (23 ± 5) °C after 40 minutes of warming-up, with less than 1 °C deviation from the full two-port calibration temperature, at output power of -5 dBm. Frequency points have to be identical for measurement and calibration (no interpolation allowed). [4] Transmission specifications are based on a matched DUT, and IF bandwidth of 10 Hz. [5] Reflection specifications are based on an isolating DUT. [6] Specification applies over entire frequency range, at output power of 0 dBm. © Copper Mountain Technologies - www.coppermountaintech.com - Rev. 2020Q215

Page 17: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

Noise floor300 kHz to 6.5 GHz -130 dBm/Hz6.5 GHz to 8.0 GHz -125 dBm/Hz8.0 GHz to 8.5 GHz -122 dBm/Hz

Damage level +23 dBmDamage DC voltage 35 V

Test Port Input

Port Ext Trig InInput level

Low threshold voltage 0.5 VHigh threshold voltage 2.7 V

Input level range 0 V to + 5 VPulse width ≥2 µsPolarity positive or negativeInput impedance ≥10 kOhmConnector type BNC, female

Frequency Trigger Input

Port 10 MHz Ref In/OutExternal reference frequency 10 MHzInput level -1 dBm to 5 dBmInput impedance 50 OhmConnector type BNC, female

Frequency Reference Input

Port 10 MHz Ref In/OutInternal reference frequency 10 MHzOutput reference signal level at 50 Ohm impedance 1 dBm to 5 dBmConnector type BNC, female

Frequency Reference Output

Port Ext Trig OutMaximum output current 20 mAOutput level

Low level voltage 0.0 VHigh level voltage 3.5 V

Polarity positive or negativeConnector type BNC, female

Frequency Trigger Output

Operating system Windows 7 and aboveCPU frequency 1.0 GHzRAM 512 MBInterface USB 2.0Connector type USB BPower supply 110-240 V, 50/60 HzPower consumption 14 WInput power 9 V DC to 15 V DCInput power consumption DC 12 W

System & Power

6 3/10 inches

1 3/4 inches

Image shows actual size

Recommended factory adjustment interval 3 years

Factory Adjustment

16

Time per point 70 µs typ.Port switchover time 1 ms

Measurement Speed

Operating temperature +5 °C to +40 °C (41 °F to 104 °F)Storage temperature -50 °C to +70 °C (-58 °F to 158 °F)Humidity 90 % at 25 °C (77 °F)Atmospheric pressure 70.0 kPa to 106.7 kPa

Environmental Specifications

Page 18: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

Accuracy of transmission measurements⁴⁴ Magnitude / Phase300 kHz to 10 MHz

0 dB to +10 dB ±0.2 dB / ±2° -35 dB to 0 dB ±0.1 dB / ±1°

-55 dB to -35 dB ±0.2 dB / ±2° -75 dB to -55 dB ±1.0 dB / ±6°

10 MHz to 7 GHz0 dB to +10 dB ±0.2 dB / ±2° -50 dB to 0 dB ±0.1 dB / ±1°

-70 dB to -50 dB ±0.2 dB / ±2° -90 dB to -70 dB ±1.0 dB / ±6°

7 GHz to 16 GHz0 dB to +10 dB ±0.2 dB / ±2° -45 dB to 0 dB ±0.1 dB / ±1°

-65 dB to -45 dB ±0.2 dB / ±2° -85 dB to -65 dB ±1.0 dB / ±6°

16 GHz to 18 GHz0 dB to +5 dB ±0.2 dB / ±2° -40 dB to 0 dB ±0.1 dB / ±1°

-60 dB to -40 dB ±0.2 dB / ±2° -80 dB to -60 dB ±1.0 dB / ±6°

Accuracy of reflection measurements⁵⁵ Magnitude / Phase300 kHz to 10 GHz

-15 dB to 0 dB ±0.4 dB / ±3° -25 dB to -15 dB ±1.0 dB / ±6° -35 dB to -25 dB ±3.0 dB / ±20°

10 GHz to 18.0 GHz -15 dB to 0 dB ±0.5 dB / ±4°

-25 dB to -15 dB ±1.5 dB / ±10° -35 dB to -25 dB ±5.5 dB / ±30°

Trace noise magnitude (IF bandwidth 3 kHz)300 kHz to 9 GHz 0.002 dB rms9 GHz to 18 GHz 0.004 dB rms

Temperature dependence300 kHz to 7 GHz 0.02 dB/°C7 GHz to 18 GHz 0.04 dB/°C

Measurement Accuracy

M5180 Specifications1

Impedance 50 OhmTest port connector type N, femaleNumber of test ports 2Frequency range 300 kHz to 18 GHzFull frequency accuracy ±5·10⁻⁶Frequency resolution 1 HzNumber of measurement points 2 to 200,001Measurement bandwidths (with 1/1.5/2/3/5/7 steps) 1 Hz to 300 kHzDynamic range²

300 kHz to 10 MHz 115 dB10 MHz to 7 GHz 130 dB (135 dB typ.)7 GHz to 12 GHz 125 dB (130 dB typ.)

12 GHz to 16 GHz 122 dB (125 dB typ.)16 GHz to 18 GHz 116 dB (120 dB typ.)

Crosstalk²ᵃa300 kHz to 5 GHz -5 GHz to 7.5 GHz -120 dB typ.

7.5 GHz to 8.5 GHz -110 dB typ.8.5 GHz to 15 GHz -120 dB typ.15 GHz to 18 GHz -100 dB typ.

Primary Specifications

300 kHz to 10 GHzDirectivity 46 dB

Source match 40 dBLoad match 46 dB

Reflection tracking ±0.10 dBTransmission tracking ±0.08 dB

10 GHz to 18 GHzDirectivity 42 dB

Source match 38 dBLoad match 42 dB

Reflection tracking ±0.10 dBTransmission tracking ±0.08 dB

Effective System Data

200 mm/7.9 inches

65 mm/2.5 inches

Image shows 78% scale

[1] All specifications subject to change without notice. [2] The dynamic range is defined as the difference between the specified maximum power level and the specified noise floor. The specification applies at 10 Hz IF bandwidth. [2a] Uncorrected crosstalk is defined at maximum specified output power level. Dynamic range of the analyzer may be limited on the lower end by either crosstalk or noise floor. [3] Reflection and transmission measurement accuracy applies over the temperature range of (73 ± 9) °F or (23 ± 5) °C after 40 minutes of warming-up, with less than 1 °C deviation from the full two-port calibration temperature, at output power of 0 dBm. Frequency points have to be identical for measurement and calibration (no interpolation allowed). [4] Transmission specifications are based on a matched DUT, and IF bandwidth of 10 Hz. [5] Reflection specifications are based on an isolating DUT. [6] Specification applies over entire frequency range, at output power of 0 dBm. © Copper Mountain Technologies - www.coppermountaintech.com - Rev. 2020Q217

Page 19: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

300 kHz to 7 GHzDirectivity 15 dB

Source match 12 dBLoad match 15 dB

7 GHz to 14 GHzDirectivity 10 dB

Source match 10 dBLoad match 12 dB

14 GHz to 16 GHzDirectivity 8 dB

Source match 10 dBLoad match 12 dB

16 GHz to 18 GHzDirectivity 6 dB

Source match 10 dBLoad match 12 dB

Uncorrected System Performance

Noise floor300 kHz to 10 MHz -115 dBm/Hz10 MHz to 7 GHz -130 dBm/Hz (135 dBm/Hz typ.)7 GHz to 12 GHz -125 dBm/Hz (130 dBm/Hz typ.)

12 GHz to 16 GHz -122 dBm/Hz (127 dBm/Hz typ.)16 GHz to 18 GHz -120 dBm/Hz (125 dBm/Hz typ.)

Damage level +23 dBmDamage DC voltage 35 V

Test Port Input

Power range300 kHz to 16 GHz -40 dBm to +10 dBm16 GHz to 18 GHz -40 dBm to +6 dBm

Power accuracy ±1.5 dBPower resolution 0.05 dBHarmonic distortion⁶⁶ -15 dBcNon-harmonic spurious⁶⁶

300 kHz to 16 GHz -20 dBc16 GHz to 18 GHz -15 dBc

Test Port Output

Port Ext Trig InInput level

Low threshold voltage 0.5 VHigh threshold voltage 2.7 V

Input level range 0 V to + 5 VPulse width ≥2 µsPolarity positive or negativeInput impedance ≥10 kOhmConnector type BNC, female

Frequency Trigger Input

Port 10 MHz Ref In/OutExternal reference frequency 10 MHzInput level -1 dBm to 5 dBmInput impedance 50 OhmConnector type BNC, female

Frequency Reference Input

Port 10 MHz Ref In/OutInternal reference frequency 10 MHzOutput reference signal level at 50 Ohm impedance 1 dBm to 5 dBmConnector type BNC, female

Frequency Reference Output

Port Ext Trig OutMaximum output current 20 mAOutput level

Low level voltage 0.0 VHigh level voltage 3.5 V

Polarity positive or negativeConnector type BNC, female

Frequency Trigger Output

Operating system Windows 7 and aboveCPU frequency 1.0 GHzRAM 512 MBInterface USB 2.0Connector type USB BPower supply 110-240 V, 50/60 HzPower consumption 32 WInput power 9 V DC to 15 V DCInput power consumption DC 25 W

System & Power

Recommended factory adjustment interval 3 years

Factory Adjustment

18

Time per point 30 µs typ.Port switchover time 0.2 ms

Measurement Speed

Operating temperature +5 °C to +40 °C (41 °F to 104 °F)Storage temperature -50 °C to +70 °C (-58 °F to 158 °F)Humidity 90 % at 25 °C (77 °F)Atmospheric pressure 70.0 kPa to 106.7 kPa

Environmental Specifications

Page 20: Compact Series · • Automation programming in LabView, Python, MATLAB, .NET, etc. ... solid-state, or PIN diode-based. If the S-parameter measurement fixture involves ... Use of

®

Technology is supposed to move. It’s supposed to change and update and progress. It’s not meant to sit stagnant year after year simply because that’s how things have always been done.

The engineers at Copper Mountain Technologies are creative problem solvers. They know the people using VNAs don’t just need one giant machine in a lab. They know that VNAs are needed in the field, requiring portability and flexibility.

Data needs to be quickly transferred, and a test setup needs to be easily automated and recalled for various applica-tions. The engineers at Copper Mountain Technologies are rethinking the way VNAs are developed and used.

Copper Mountain Technologies’ VNAs are designed to work with the Windows or Linux PC you already use via USB interface. After installing the test software, you have a top-quality VNA at a fraction of the cost of a traditional analyzer. The result is a faster, more effective test process that fits into the modern workspace. This is the creativity that makes

Copper Mountain Technologies stand out above the crowd.

We’re creative. We’re problem solvers.

!"#$" !"#%" !"#&# !"'(#)*+,-+./0123.4+ !""#$%&#'(#)*+#,%& !""#$%&#'(#-*+#,%& !""#$%&#'(#.*+#,%& !""#$%&#'(#/.#,%&

5673*38+9+*: 0!! 1#0"! 1#0!" 1#0"" 0!! 1#0"! 1#0!" 1#0"" 0!! 1#0"! 1#0!" 1#0"" 0!! 1#0"! 1#0!" 1#0"";0.38</123.4+ /!"#231#'45* /!"#231#'45* /!"#231#'45* /!+#231#'45*

USA: +1.317.222.5400Singapore: +65.6323.6546

631 E. New York St | Indianapolis, IN | 46202www.coppermountaintech.com

Latin America: +1.954.706.5920London: +44 75 03 68 21 13