communications impact of hall effect plasma

104
COMMUNICATIONS IMPACT OF HALL EFFECT PLASMA THRUSTERS by JAMES CLAUDE DICKENS, B.S.E.E., M.S.E.E A DISSERTATION IN ELECTRICAL ENGINEERING Submitted to the Graduate Faculty of Texas Tech University in Partial Fulfillment of the Requirements for the Degree of DOCTOR OF PHILOSOPHY Approved Co-;^hairperson of th§ Committee Co-C/ha/roerson of the Committee Accepted Dean of the Graduate School May, 1995

Upload: others

Post on 11-Feb-2022

2 views

Category:

Documents


0 download

TRANSCRIPT

COMMUNICATIONS IMPACT OF HALL

EFFECT PLASMA THRUSTERS

by

JAMES CLAUDE DICKENS, B.S.E.E., M.S.E.E

A DISSERTATION

IN

ELECTRICAL ENGINEERING

Submitted to the Graduate Faculty of Texas Tech University in

Partial Fulfillment of the Requirements for

the Degree of

DOCTOR OF PHILOSOPHY

Approved

Co-;^hairperson of th§ Committee

Co-C/ha/roerson of the Committee

Accepted

Dean of the Graduate School

May, 1995

J

T 3 ^ ACKNOWLEDGMENTS • ^ '* >z

([ I would like to express my appreciation to Dr. M. Kristiansen and Dr E.

O'Hair for their support and technical advice during this research project. I would

also like to thank the other members of my committee, Dr L. Hatfield, Dr. T. Trost,

and Dr. F. Curran for their guidance. I am also gratefijl to John Mankowski, for his

assistance in designing and building hardware necessary to complete the project.

I am especially indebted to J. Sankovic, E. Pencil, T. Haag, and D. Manzella,

who have provided many insights and assistance in completing this project. In

addition, without the use of NASA LeRC's vacuum tank facilities and equipment,

none of this work would be possible,

Finally, I would like to thank my family and especially my wife, Molly, who has

provided support and encouragement throughout my academic career.

11

T.^BLE OF CONTENTS

ACKNOWLEDGMENTS ii

ABSTRACT iv

LIST OF TABLES v

LIST OF FIGURES vi

CHAPTER

1. INTRODUCTION 1

2. PROBLEM DESCRIPTION 5

3. PLUME MODEL 9

4. MICROWA\^ INTERFEROMETER 22

5. PHASE SHIFT MEASUREMENTS 27

6. PHASE NOISE 48

7. CONCLUSIONS 60

REFERENCES 65

APPENTDIX

A. LANGMUIR PROBE PLOTS 67

B. MATHCAD WORKSHEETS 74

C PHASE SHIFT PLOTS 77

D. POWER SPECTRAL DENSITY PLOTS 85

111

ABSTRACT

A Hall effect thruster is an electric space propulsion device, in which a gas

(typically Xenon) is ionized and accelerated by a self-induced electric field. Because

the exhaust plume of a Hall effect thruster is an ionized gas, the thruster's plume can

affect the propagation of electromagnetic radiation. These effects could have a

significant impact on the channel capacity of satellite communication systems.

The first part of the study was devoted to developing a far field plume model

that can predict both the spatial and temporal number density in the plume of three

different Hall effect thrusters. The spatial dependence of the number density in the

plume was determined using a swept Langmuir probe. The temporal dependence was

determined using a high speed Langmuir probe positioned along the centerline of the

plume.

In an effort to verify the far field number density plume model, a sampling

microwave interferometer was developed, that can accurately measure the phase shift

of microwave signals propagating through the plume of a Hall effect thruster. The

interferometer was used to measure the phase shift of a 6 GHz microwave signal

during the startup of a Hall effect thruster with several different propagation paths A

comparison between the model predicted phase shift and the experimentally obtained

phase shift are made. A microwave spectrum analyzer was used to qualify and

quantify the plumes effect on the phase noise of a microwave signal propagating

through the plume.

IV

LIST OF TABLES

3.1. Plume model coefficients 1

A. 1. Operating parameters of the thrusters for the Langmuir probe and current plots in Appendix A 68

C 1 . Operating parameters of the thrusters for the phase shift plots and current waveforms in Appendix C 78

D. 1. Operating parameters of the thrusters for the power spectral density plots and current waveforms in Appendix D 86

V

LIST OF FIGURES

1.1. Cross-section diagram of a TAL plasma thruster 1

1.2. Cross-section diagram of a TML plasma thruster 2

3.1. Normalized number density profile of a D-55 thruster 11

3.2. Normalized number density profile of a SPT-100 thruster 11

3.3. Comparison of the measured plume profile of a D-55 thruster and the plume model of Equation (3,2) 13

3.4, Comparison of the measured plume profile of an SPT-100 thruster and the plume model of Equation (3,1) 14

3.5, Percent number density for a T-100 thruster 16

3.6, Percent peak current waveform for a T-100 thruster 16

3.7, Percent number density for an SPT-100 thruster 17

3.8. Percent peak current waveform for an SPT-100 thruster 17

3.9 Percent number density of a D-55 thruster 17

3.10. Percent peak current waveform of a D-55 thruster 18

3.11. FFT of the D-55 number density waveform 19

3.12. FFT oftheT-lOOnumber density waveform 20

3.13. Plot of the static plume model for an SPT-100 21

3.14. Plot of the complete plume model for an SPT-100 21

4.1. Schematic diagram of a basic microwave interferometer 23

4.2. Block diagram of a sampling microwave interferometer 24

5.1. Sketch of parallel propagation path (relative to the exit plane) 28

5.2. Parallel start-up phase shift at 0.5 m for a T-lOO 28

\i

5.3. Perpendicular start-up phase shift at 0,9 m for an SPT-100 :, 30

5.4. Perpendicular start-up phase shift at 0.9 m for a D-55 30

5.5. Sketch of 45 deg propagation path 31

5.6. Startup phase shift along a 45 deg propagation path for a D-55 32

5.7. Startup phase shift along a 45 deg propagation path for an SPT-100 33

5.8. Startup phase shift along a propagation path perpendicular to the exit plane of a D-55 33

5.9. Startup phase shift along a diagonal propagation path for an end-of-life SPT-100 ^ 35

5.10. Change in phase shift due to change in cathode flow rate for a D-55 (0.2mg/s at 2 sec linearly to 1,6 mg/s at 28 sec) 36

5.11. Measured phase transients of a 6 GHz signal propagating along a parallel

path (relative to the exit plane) through the plume of a T-100 37

5 12, Current waveform of the T-100 in Fieure 5,11 38

5 13 Predicted phase shift along a parallel propagation path 0.5 m down from the exit plane as a fianction of m, for the T-100 38

5.14. Measured phase transients of a 6 GHz signal propagating along a parallel propagation path through the plume of a SPT-100 40

5.15. Current waveform of the SPT-100 in Figure 5.14 40

5.16. Predicted phase transients of a 6 GHz signal along a parallel propagation path 0.9 m down from the exit plane as a function of m 40

5.17. Measured phase transients of a 6 GHz signal propagating along a 45 deg diagonal propagation path for a SPT-100 42

5.18. Current waveform of the SPT-100 in Figure 5.17 42

5.19. Predicted phase transients along 45 deg diagonal propagation path at 1 m down from the exit plane as a fijnction of m 42

VI1

5.20. Measured phase transients of a 6 GHz signal propagating along the parallel propagation path through the plume of a D-55 44

5.21. Current waveform of the D-55 in Figure 5.20 44

5.22. Measured phase transients of a 6 GHz signal propagating along a 45 deg diagonal propagation path for a D-55 44

5.23. Current waveform for Figure 5.22 45

5.24. Measured phase transients of a 6 GHz signal propagating along a 31 deg propagation path for an end-of-life SPT-100 45

5.25. Current waveform for Figure 5,24 45

5.26. Predicted phase transients along 31 deg diagonal propagation path at

1.8 m to the side of an SPT-100 as a fijnction of m 46

6.1. Phase noise test setup 50

6.2. Spectrum of a received signal with and without a T-lOO running

(1 MHz span) 51

6.3. Current waveform of the T-lOO in Figure 6.2 51

6.4. Spectrum of a received signal with and without an SPT-100 running

(200 kHz span) 52

6.5 Current waveform of the SPT-100 in Figure 6.4 52

6.6. Spectrum of a received signal with and without a D-55 running

(200 kHz span) 53

6.7. Current waveform of the D-55 in Figure 6.6 54

6.8. Spectrum of a received signal with and without a SPT-100 running along

the 45 deg diagonal propagation path(100 kHz span) 55

6.9. Current waveforms of the SPT-100 in Figure 6.8 55

6.10. Spectrum of a received signal along the 45 deg diagonal propagation path with a D-55 operating at 3.5 A and 4,5 A (200 kHz span) 56

6.11. Current waveforms of the D-55 in Figure 6,10 56 viii

6.12, Spectrum of a 6 GHz received signal with and without an end-of-life SPT-100 running (200 kHz span) 58

6.13, Current waveform of the end-of-life SPT-100 in Figure 6,12 58

6.14, Spectrum of a 2.8 GHz received signal with and without an end-of-life SPT-100 running (750 kHz span) 59

7.1. Phasor illustration of thruster induced phase noise and typical additive

noise 62

A.l. Percent number density of a T-lOO ( file: lang205 ) 69

A.2. Percent peak current of a T-lOO ( file: curr205 ) 69

A.3. Percent number density of a T-lOO ( file: lang202 ) 69

A,4. Percent peak current of a T-lOO ( file: curr202 ) 70

A. 5. Percent number density of a T-100 ( file: lang204 ) 70

A.6. Percent peak current of a T-lOO ( file; curr204 ) 70

A. 7. Percent number density of an SPT-100 ( file: lang526 ) 71

A. 8. Percent peak current of an SPT-100 ( file: curr526 ) 71

A.9. Percent number density of an SPT-100 ( file: lang522 ) 71

A.IO. Percent peak current of an SPT-100 ( file: curr522 ) 72

A l l . Percent number density of a D-55 (file: lang404 ) 72

A.12. Percent peak current of aD-55 ( file: curr404 ) 72

A.13. Percent number density of aD-55 (file: lang406 ) 73

A.14. Percent peak current of a D-55 (file: curr406 ) 73

C I . Start-up phase shift (along a parallel path at 0.5 m) for a T-lOO (file phasll6) 79

C.2. Current waveform of the T-lOO for Figure CI 79

IX

C,3 Phase shift of a 6 GHz passing through the plume of a T-100 ( parallel

path 0,5 m from the exit, file: phasl07) 79

C.4. Current waveform of the T-lOO for Figure C.3 80

C.5. Start-up phase shift (along a 45 deg diagonal path) for an SPT-100 (file phas614) 80

C.6. Current waveform of the SPT-100 for Figure C 5 80

C.7. Phase shift of a 6 GHz passing through the plume of an SPT-100 along a

45 deg diagonal path (file: phas613) 81

C.8. Current waveform of the SPT-100 for Figure C.7 81

C.9. Phase shift of a 6 GHz passing through the plume of an SPT-100 along a

45 deg diagonal path (file: phas612) 81

CIO, Current waveform of the SPT-100 for Figure C,9 82

C11. Start-up phase shift (along a 45 deg diagonal path) for a D-55

(filephas714) 82

C.12. Current waveform of the D-55 forFigureC.il 82

C.13. Phase shift of a 6 GHz passing through the plume of a D-55 along a 45

deg diagonal path (file: phas707) 83

C.14. Current waveform of the D-55 for Figure C.13 83

C.15. Phase shift of a 6 GHz passing through the plume of aD-55 along a 45

deg diagonal path (file: phas710) 83

C.16. Current waveform of the D-55 for Figure C.15 84

C17. Phase shift of a 6 GHz passing through the plume of a D-55 along a 45 deg diagonal path (file: phas713) 84

C.18. Current waveform ofthe D-55 for Figure C17 84 D. 1. Power spectral density of a 3.8 GHz signal with a parallel path 1 m down

stream from the exit of an SPT-100 (200 kHz span, file: an021) 87 D.2. Current waveform ofthe SPT-100 for Figure D.l 87

x

D,3. Power spectral density of a 6 GHz signal with a parallel path 1 m down stream from the exit of an SPT-100 (200 kHz span, file; an026) 87

D 4. Current waveform ofthe SPT-100 for Figure D.3 88

D.5. Power spectral density of a 6 GHz signal with a 45 deg diagonal path

through the plume of an SPT-100 (200 kHz span, file: an022) 88

D.6. Current waveform of the SPT-100 for Figure D. 5 88

D.7. Power spectral density of a 6 GHz signal with a 45 deg diagonal path through the plume of an SPT-100 (750 kHz span, file: an629) 89

D.8. Current waveform ofthe SPT-100 for Figure D.7 89

D.9. Power spectral density of a 6 GHz signal phase modulated with a 200 kHz tone propagating through the plume of an SPT-100 along a 45 deg diagonal path (750 kHz span, file: an632) 89

D 10. Current waveform ofthe SPT-100 for Figure D,9 90

D. 11. Power spectral density of a 6 GHz signal with a parallel path 1 m down

stream from the exit of aD-55 (200 kHz span, file: anOllb) 90

D,12, Current waveform ofthe D-55 for Figure D,l 1 90

D,13. Power spectral density of a 3.8 GHz signal with a parallel path 1 m down

stream from the exit of a D-55 (1 MHz span, file: an017b) 91

D.l 4 Current waveform ofthe D-55 for Figure D.l 3 91

D. 15. Power spectral density of a 6 GHz signal with a 45 deg diagonal path through the plume of a D-55 (750 kHz span, file: an636) 91

D.16. Current waveform ofthe D-55 for Figure D.15 92 D. 17 Power spectral density of a 6 GHz signal phase modulated with a 50 kHz

tone propagating through the plume of a D-55 along a 45 deg diagonal path (750 kHz span, file: an638k) 92

D.l 8. Current waveform ofthe D-55 for Figure D.l 5 92

xi

CHAPTER 1

INTRODUCTION

Hall thrusters are a type of electric space propulsion unit in which a gas

(typically Xenon) is ionized and accelerated by a self-induced electric field. There are

two main types of Hall thrusters currently being investigated, the thruster with

magnetic layer (TML) and the thruster with anode layer (TAL). Although the TML

and the TAL are structurally different, their theory of operation is the same. Both

types of Hall thrusters have four major components, an inner magnetic coil, an outer

magnetic coil, an anode/gas ejector, and a cathode assembly. In addition, the TML

has an insulator lining the discharge chamber. Cross-section diagrams ofthe TAL and

TML are shown in Figure 1 1 and Figure 1.2, respectively.

Cathode Assembly

H / / ^ • I ' 5

Annular Anode

Xenon Gas Feed Lines -

Main Body Housing —' and Primars' Pole Piece

Insulator

/ / • / / / /

/ / ,• ./-l'

5?;^ - Magnetic Field Profiler '— Inner Magnetic Coil

\ /

• — Outer Magnetic Coil

Figure 1.1. Cross-section diagram of a TAL plasma thruster.

1

Cathode Assembly

V Xenon Gas Feed Lines —

Main Bod)- Housing —-" and Pnman Pole Piece

'' / /A"—' ' 'r'—r'-^— : ' > > \ V — > • • ••'

A ///'/// '/^/, 'y '• / - /y W ' X '/ '/ /, /'

KM . ^ • ^

Annular Insulator

^

/ y

'/•/y'y'//' ^yV//^''y' '?. ' / y / / /

/ / / / / / / / / y ' y ' -

/ / / ' • '^ y / , / y / / , / ' • y / y y ' y ' ,\

-Magnetic Field Profiler Inner Magnetic Coil

Insulator

Anode

y /

• / y / ' y / , ,

• / , '

y ' \

— Outer Magnetic Coil

Figure 1,2, Cross-section of a TML plasma thruster.

The cathode assembly is used as a source of free electrons to neutralize the

plasma exhaust plume and to provide electrons to the acceleration and ionization

regions in the discharge chamber. The inner and outer magnetic coils are biased with a

dc current to produce a radial magnetic field. When a voltage is applied across

cathode to anode, the electrons being emitted by the cathode assembly are swept

toward the anode. The radial magnetic field lines "trap" the free electrons and impede

their motion from the cathode to the anode and thus produce a cyclotron motion along

equipotential magnetic field lines. The Xenon gas, introduced through the anode,

difiFuses into the discharge and into an ionization region where the "trapped" electrons

collide with the Xenon atoms, exciting the atoms into an ionized state (preferably a

singly ionized state). The Xenon ions are then accelerated axially by the electric field

produced by the trapped electrons. When an electron collides with a Xenon atom, the

electron diffuses towards the anode, thus after a finite number of collisions, the

electron reaches the anode and must be replaced by additional electrons from the

cathode assembly to continue the process

The ceramic insulator lining the wall ofthe main discharge chamber in the

TML is utilized to prevent erosion ofthe magnetic pole pieces. The TAL relies upon

the magnetic field profile to prevent pole piece erosion The TML magnetic field

profile is broader than that ofthe TAL and has a peak value located within the thruster

body The area of acceleration is located in the area of peak magnetic field, thus for

the TML, the acceleration region is also contained within the thruster body In

contrast, the TAL has a sharp magnetic field profile with a peak value at the exit plane.

As a resuh the acceleration region for the T.AL is located at the exit ofthe thruster,

therefore, eliminating the need for a ceramic insulator

The performance of TML and TAL thrusters ofthe same electric power, has

been shown by Sankovic [1,2] and Garner [3] to be ver}' similar. Typical xalues of

specific impulse (ISP) for a 1,5 kW Hall effect thruster are in the range of 1500 s to

1800 s and they have an efficiency that can exceed 50 percent. In particular, three

1.35 kW Hall effect plasma thrusters have been characterized and will be the subject of

this research report. The single TAL thruster tested is the D-55 manufactured by

TSNIIMASH (Research Institute of Machine Building) in Russia, The two TML

thrusters tested are the SPT-100, manufactured by Fakel Enterprises and the T-lOO,

manufactured by the Scientific Research Institute for Thermal Process (NTITP), Both

of these thrusters are also manufactured in Russia

The first chapter of this paper discusses the potential impacts of Hall effect

thrusters on communication systems In the second chapter, a spatial and temporal

far-field plume model is developed. Chapter 3 describes a sampling microwave

interferometer developed to provide a way to verify the plume model and to quantify

the magnitude and frequency of phase oscillations and phase transients experienced by

communications signals propagating through the plume of Hall effect thrusters. In

Chapter 4, the experimentally obtained phase shift data is presented and discussed.

Chapter 5 describes an experiment used to determine the impact a Hall effect

thruster's plume can have on the phase noise of a communication signal. The results

of this experiment are also presented in that chapter The final chapter presents

conclusions that can be made from the research and recommendations to minimize the

thruster's impact on communication systems. In addition, suggestions for fijture work

are also discussed.

CHAPTER 2

PROBLEM DESCRIPTION

Potential applications for Hall effect thrusters include North-South station

keeping and orbit raising of communications satellites, deep space research missions,

and general positioning of military satellites. Because each of these applications

involves extensive data transfer between the satellite and a ground station or another

space vehicle, the adverse effects ofthe propulsion system on communications must be

characterized and minimized. Areas of potential impact include radiated EMI, signal

attenuation, beam aberrations, increased phase noise, and high-speed phase transients.

The scope of research presented in this paper is limited to the latter ofthe two impact

areas, increased phase noise and high-speed phase transients.

Satellite communication systems can utilize any one of several modulation

methods, including frequency-shift keying (FSK), phase-shift keying (PSK), multistate

frequency-shift keying (M-ary FSK), and muhistate phase-shift keying (M-ary PSK).

Each of these modulation schemes transmits information in the phase or frequency of a

RF carrier signal. Frequency bands typically allocated to communication satellites are

in the microwave range of 4 GHz and 6 GHz. Because satellite communication

systems must transmit hundreds of signals at a time utilizing either time division

multiplexing or frequency muhiplexing, any increase in phase noise or phase transients

can cause channel cross-talk in analog systems or increased bit error rates in digital

systems [4,5]. The resuh of either of these situations is a loss in channel capacity and,

therefore, an economic loss to the satellite owner.

The plume of a Hall effect thruster is composed primarily of ionized Xenon

atoms and an equal number of free electrons, thus forming a neutral plasma.

Therefore, to determine how the exhaust plume can impact communications, the

interaction between the plasma and a microwave signal must be examined. A long

known effect of free electrons in a plasma is the effect on the relative dielectric

constant and thus the local wavenumber The local wavenumber in the plasma, k, can

be expressed as.

k{x,y,z) = ^<^o^ -o)Ax,y,z)

(21)

where cOo is the carrier frequency and c is the speed of light. The local plasma

frequency cOp is given by.

coAx,y,z) = \n^{x,y,z)q-

^o^h (2,2)

where q is the charge of an electron, me is the mass of an electron and ne is the local

plasma electron number density. The phase difference between waves traveling along

a path with a plasma present and one without a plasma present can be expressed as.

^e=\[k,-k{x,yr^)\ix^k^\ 1-Jl n^{x,y,z)q'

^o ^o^h dX, (2.3)

where the integral is along the signal propagation path, ko is the wavenumber in free

space, and 8o is the permitti\'ity in free space. The phase shift expressed in Equation

(2.3) is static and does not represent a serious communication problem. If the local

number density, n^ is not temporally stable however, a modified phase shift equation is

given by.

Aeit) = k^j ^o £o^h

1_ | l _ ! ! i l M i l i i V ^ ^ (2.4)

In this equation, the phase shift is a fijnction of time due to the temporal dependence

ofthe local number density, n .

In a communication system using phase modulation, the transmitted signal can

be expressed as,

f{t), = Acos{(oJ^y{t)) (2.5)

where A is the amplitude, cOc is the carrier frequency, and y(t) is the time dependent

information to be transmitted. If the transmitted signal has a propagation path that

passes through a region of time-vary'ing plasma, the received signal will be modified

and have a form given by,

/ ( / ) , =/icos(^,/ + r(?) + A^(0) (2,6)

At the receiver end ofthe communication link, the extracted phase information will

contain the original information signal, Y(t), plus the additive phase shift signal, A9(t),

Because A9(t) is unknown and possibly random, extraction ofthe function from the

original signal is impossible and thus must be treated as an additive noise term. In a

similar analysis, it can be shown that a frequency modulated signal passing through a

7

temporally unstable plasma will have the same additive noise, A9(t) present in the

extracted information signal.

Depending on the magnitude ofthe phase shift and type of modulation scheme

used, severe communication system degradation can occur. Areas of system

degradation include decreased signal-to-noise ratio ofthe individual channels, cross­

talk between adjacent channels, increased bit error rates (BER), and complete frame

loss for the case of large phase transients [5].

8

CHAPTER 3

PLLTvIE MODEL

As a resuh ofthe previous analysis, it was determined that a comprehensive

plume model was required to quantify potential communication impacts. The model

must include both spatial and temporal representations ofthe electron number density

in the thruster far field (greater than 0.5 m from the exh plane). Although a near field

plume model is beneficial from a thruster development point of view, it is not

necessary for a communication impact study, because the satellite antennas are usually

never located closer than 0.5 m for mechanical reasons. The plume model developed

for this study was broken into two parts, the spatial representation and a temporal

modulation term.

The spatial model was based upon an analysis performed by Carney [6] on low

power arcjet thrusters. The analysis provided a model that assumed that the far flow

field was a freely expanding gas dynamic process. Because Hall effect thrusters have

lower mass flow rates, the assumption that the far flow field is still freely expanding is

certainly valid. As a resuh, the plume model was believed to be easily adaptable for

use in describing the number density in a Hall effect thruster's plume The general

form ofthe model is given by,

-[/.O-cos.f))]"

"{>•,&} = —: I T M - ' (3.1) ^ r COS [6}

where n and ). are coefficients to be determined from the plume profile data and n is

the on-axis number density at 1 m.

In an effort to determine n and /., swept Langmuir probe measurements were

made using both planar and spherical probes. The Langmuir probes were biased with

a constant dc vohage and swept 180 deg around the exit ofthe thruster under test.

The plume profile was obtained by using a computer-controlled data acquisition

system to capture the vohage across a 100 ohm resistor at a rate of 250 samples per

second. A detailed description ofthe system is given by Mankowski [7]. The effort in

this experiment was directed at verifying results previously reported by Manzella [8],

using a similar test apparatus, and Myers [9], using a Langmuir probe rake setup.

Figure 3 1 shows the normalized plume profile data obtained for a D-55 TAL

operating at its nominal vohage and current of 300 V and 4.5 A, respectively On the

same graph, data obtained by Manzella are also plotted for comparison. The

Langmuir probe was biased positive and swept about a radius of 0.6 m. The plume

data for the SPT-100 are shown in Figure 3.2. The plume profile ofthe T-lOO was

found to be similar to those ofthe SPT-100 and D-55 despite early claims of a

narrower beam width. In addition, changing operating parameters such as magnetic

field strength, cathode flow fraction, and discharge vohage had little, if any, effect on

the plume profile for the three thrusters. It can be seen in Figure 3.1 that the D-55

plume is de-focused and has an annular peak around the plume centerline. Although

the basic plume model cannot account for the defocusing, a modification to the model

resuhing in the new equation.

10

Measured Data • Manzella Data

Figure 3.1. Normalized number density profile of a D-55 thruster.

1.2

.-=0.8

a:

0.6

0,4

0,2 ^ ^

^

f 1

l^y^

n 4:]

ad f

o

a •III D

4

•1 [

• • '1

^ ^^f*!*^

ncsBuma 33CCn

-40 -30 -20 -10 0

deg

10 20 30 40

Measured Data ^ Manzella Data

Figure 3.2. Normalized number density profile of an SPT-100 thruster

11

''^'^'^~- rcos-(.) <'-^^^^^('y^^ (32)

where q and k are coefficients dependent upon the degree of de-focusing, can account

for the annular peaks. The plume profiles ofthe SPT-100 and T-lOO do not exhibit

the same defocusing and, therefore, Equation (3,1) can sufficiemly model the plume.

For the purpose of a communications impact study, the accuracy of Equation (3 1) in

describing the plume ofthe D-55 is sufficient and thus reduces the computation time

required to make phase shift calculations.

The plume model was fitted to the data with values of n ranging from 0,5 to

0,7 and X in the range of 40 to 60, additionally the values of q and k for the D-55 are

in the range of 500 to 1000 and 0,1 to 0,3, respectively. Optimal values of n, q, k and

X for the three thrusters running at their nominal operating points are given in

Table 3.1. Figure 3.3 shows a comparison between the measured plume profile ofthe

D-55 and the plume profile described by the model described by Equation (3.2).

Figure 3.4 gives a comparison between the plume model of Equation (3 1) and the

SPT-100 thruster. In both plots, the fit ofthe model to the measured profile is ver>-

tight and within experimental uncertainty In each ofthe following graphs, the ion

Table 3.1. Plume model coefficients.

Thruster Model D-55

SPT-100

T-lOO

n 0.65

0.65

0.68

X 35

33

37

q 0,3

N/A

N/A

k 1000

N/A

N/A

12

1,2

I" U

B 0,6

o N

o 2

0.4

0.2

1

J^

,' ^ n

\

\

-40 -30 -20 -10 0

deg

10 20 30 40

Measured Data ••° Plume Model

Figure 3.3. Comparison ofthe measured plume profile of a D-55 thruster and the plume model of Equation (3.2),

density distribution was given. An important assumption that was made during the

development of this far field plume model was that the electron and ion number

densities are approximately equal. Although this assumption is reasonable in the far

field, in the confined space of a vacuum tank significant differences between the two

number densities can be present due to charge exchange whh the walls. In addhion,

the near field electron and ion number densities can be different due to the magnetic

field leakage from the thruster. The resuh ofthe magnetic leakage field is a local

increase in the number density around the thruster body.

Accounting for the temporal dependence ofthe plasma plume was the next

step in refining the plume model. Because the discharge currents of all the Hall

13

1.2

I OS

I 0.6 u

04

0.2

HHffnl a««assfl9B5^

1 _

j r «

n^

^ ,

J /

- . 1 ' :

c 1 1 t ! 1 1

\ \

V ^ ^ ^ L

-40 -30

Measured Data ^ Plume Model

-20 -10 0

deg

10 20 30 40

Figure 3.4. Comparison ofthe measured plume profile of an SPT-100 thruster and the plume model of Equation (3 1).

effect thrusters tested have oscillations, the number density in the plume is also

oscillating. In order to account for the temporal oscillations, a modulation term was

added to the profile model. The modulation term has the form of a traveling wave

with a spherical wavefront originating at the center ofthe thruster exit plane. With the

addition ofthe modulation term to Equation (3 1) the resulting plume model has the

form,

-[A(l-cos(^))f

nir.eA^^-. T7T^(l-'wcos(A2;r/-r/:^)), ^ ' r'cos'(^)

(3,3)

where m is the percent number density fluctuation coefficient, f is the dominant

fluctuation frequency, and kp is the wave propagation constant given by,

14

^ = ^ ^ ^ , (3 4) v

F

where Vp is the plume exit velocity The coefficients m and f were determined using a

high-speed Langmuir probe setup. In this experiment, a spherical Langmuir probe was

placed in the plume centerline ofthe thruster under test The probe bias was swept

using a fijnction generator connected to a bipolar amplifier. The average centerline

number denshy, n^ was obtained by measuring the average saturation current across a

100 ohm bias resistor. The current density modulation coefficient, m, and the

modulation frequency, fm, were obtained by applying a constant dc bias on the probe in

the centerline. The current across the probe bias resistor was recorded using the high­

speed data acquisition system. A complete description ofthe Langmuir probe setup is

given by Mankowski [4],

As expected, the centerline number density, n^ was highly dependent on the

Xenon mass flow rate to the anode. The cathode gas flow rate also has a large impact

on the number density Under certain operating conditions, the cathode flow rate

would change the number density by 40 percent or more. Typical values of n,. in this

experiment were in the range of 0.4xl0'^ to 1 5xlO'^ under nominal operating

conditions for all the thrusters tested. These numbers are close to those obtained by

Myers [9]. The number density modulation coefficient, m, was obtained graphically

from the high-speed Langmuir current wa\eforms. Figure 3 5 shows the relative

number density as a ftinction of time on the plume centerline of a T-lOO thruster, and

Figure 3.6 shows the associated relative current waveform From the graphs it can be

15

seen that the number density for the T-lOO has a modulation coefficient, m, between

0 02 and 0.16, depending upon the level of current oscillations. In a similar format.

Figure 3.7 shows the relative number density for an SPT-100 with nominal operating

parameters. Figure 3.8 gives the relative discharge current waveform ofthe SPT-100.

The SPT-100 graph shows a peak number density fluctuation of 0.12. The number

density fluctuation and current variation graphs for the D-55 are shown in Figure 3,9

and Figure 3.10, respectively. Note the differences in the vertical axis scale for these

figures.

100

95 -

J 90 E Z

85 -

80

7"

• I. • • , ! ( -

\ -, •, • '• •

I ^ . I \

I

M,,:,5 • ) ! ^ 1

(

1 •'

0.1 0,2 0,3 0,4 0,5 0,6 time (ms)

0.7 0,8 0.9

Figure 3.5. Percent number density of a T-lOO thruster

100

60

I )

t.[ '' ^ /; :

J, , A . r :

1

I I -[ '

1 ! "

< ' 1

i f V

I L 0,1 0.2 0.3 0.4 0.5 0.6

time (ms)

0.7 0.8 0.9

Figure 3.6. Percent peak current waveform for a T-lOO thruster.

16

100

85 0.1 0.2

_L_ 0.8 0.3 0.4 0.5 0.6 0.7

time (ms)

Figure 3 7. Percent number density of an SPT-100 thruster.

0.9

100

c 3

u c .o

50

f I I

\ I

i l l ! !

' ;• i i i l !:

I ' l ihi iMi^

r i\ !!

M M i I 1 M ; I i I I M I i ,/ I h, I I ! ; 1 I I

I ! i

1 ; ' . .' I

'7 1/ V {! {' l V \ I'' \! ' I i M M M1111 i 1; : i i j !; ;; 1M, I 11 w i I 1M n n i \ M / 1/ w II / '' V. I :,i •' \! \: \l V \! ; V ' '' '• ' \' i ' 1/

± 0.1 0.2 0.3 0.4 0.5 0.6

time (ms)

0.7 0.8 0.9

Figure 3.8. Percent peak current waveform for an SPT-100 thruster.

0.1 0.2 0.3 0.4 0.5 0.6

time (ms)

0.7 0,8 0.9

Figure 3.9. Percent number density of a D-55 thruster.

17

100

•S 90

3 O

80

( '

f i l l i f "

M i l I'

70

V\

±

i i I' , f ) .

, i ^ ' ^ ' i l l i l l i . ' ( V ' V '•" 1 1 h I I ' i ' i ' I'l

Vi;;,i I 1 I ) I I I M l I n " • ' t ^

0.1 0.2 0.3 0.7 0,8 0.9 0,4 0,5 0,6 time (ms)

Figure 3.10. Percent peak current waveform for aD-55 thruster.

The D-55 graph shows a number density fluctuation of 3 %. The Langmuir

probe data presented in the previous graphs are representative data only. The

modulation coefficient for each thruster is dependent upon the exact operating

parameters, discharge fiher, and thruster age. Addhional Langmuir probe and current

waveform pairs are presented in Appendix A. The number density modulation factor,

m, can be approximately related to the discharge current fluctuation by a current

sensitivity factor, Csens, given by.

C.„„. = m C

(3.5) mod%

where Cniod% is given by,

C C = 1 - — (3,6)

where C^n and Cmax are the minimum and maximum currents in a given time window.

Values of Cn,odo.. for the three thrusters are in the range of 20 % to 25 %, whh a value

18

of 22.5 % typical for the SPT-100 and T-lOO thrusters and 25 ° o typical for the D- >

thruster.

The primary modulation frequency, f , for the three thrusters was obtained

both graphically and by using an FFT algorithm Although the primary modulation

frequencies ofthe SPT-100 and T-lOO are clearly defined, the number density

waveform ofthe D-55 appears to have no distinguishable primary frequency. As seen

in Figure 3.11, the FFT ofthe D-55 number density waveform is broadband and has

several peaks in the 10 to 70 kHz range. In contrast, the FFT ofthe T-lOO waveform,

shown in Figure 3 12. has a distinct frequency spike at 22 kHz Likewise, the

SPT-100 has a distinct number density oscillation of 27 kHz, .AJthough the

modulation frequency is not as dependent upon the operating parameters as the

modulation index is, small variations in the modulation frequency are observable \\ ith

changes in the magnetic field for each thruster

I 08 C

c.0.6

•V 0.4

E z

02

i! '

1 ' I

I!:: ''Mi '' I

I i\ I J

i l 1 1 I

|!i

I ii ' , . 1 ' M

f I! " ' I I.**! 1 I •!Ui ! I , ' J '

i / l ^ ' I i i U i|i

20 40 60 80 100

kHz

i:c 140 160 180 200

Figure 3.11. FFT ofthe D-55 number denshy waxeform

19

•= 0,8

^0.6 v.

0,4

Z 0,2 I }K

I ,. < I I

20 40 60 80 100 120 140 160 180 20C

kHz

Figure 3 12. FFT of the T-100 number denshy waveform.

The one remaining parameter not determined is the exh velocity. Because the

exit velocity is not easily measured directly, an inference had to be made that the exit

velochy of all three thrusters was similar The reasoning behind this is the fact that the

three thrusters have similar specific impulses and efficiencies, both of which are

directly related to the exit velocity. Whh this assumption, the exit velocity data

obtained by Manzella [10] for the SPT-100 can be applied to the other two thrusters.

The reported exh velocity ofthe SPT-100 is 15 km/s Although an exact exh velocity

for the D-55 and T-lOO has not been determined, the overall effect of this

approximation is negligible when viewed from a communication impact point of view.

With the spatial and temporal coefficients determined for each ofthe three

thrusters, a complete far-field plume model can be assembled A visual representation

of Equation (3.1) using coefficients determined for the SPT-100 is shown in Figure

3 13 Figure 3.14 is a visual representation of the complete plume model described by

20

Equation (3.3) for the SPT-100. Although many ofthe plume model coefficients are

dependent upon the thruster operating parameters, the number density modulation

factor, m, shows the greatest dependence on variations in the thruster operating

parameters.

nt

Figure 3.13. Plot ofthe static plume model for an SPT-100 (5 m square)

nt

Figure 3.14. Plot ofthe complete plume model for an SPT-100 (5 m square).

21

CHAPTER 4

MICROWAVE INTERFEROMETER

A microwave interferometer is a device that is used to obtain the line

integrated number density of a plasma by measuring the phase shift that a microwave

signal experiences when passing through the plasma. Due to the dispersion properties

of a plasma on RF signals, as discussed in Chapter 2, a microwave signal passing

through a plasma undergoes a phase shift relative to a signal propagating in free space.

If the phase shift can be determined and the integration path is known, the line-

integrated number denshy can be extracted using Equation (2 4), In this research, a

microwave interferometer was used for two purposes. The first was to verify the

plume model developed in Chapter 3 and the second was to measure phase transients

of a microwave communication signal propagating through the plume of a Hall effect

plasma thruster.

A schematic diagram of a microwave interferometer is shown in Figure 4,1

The basic system consists of a microwave oscillator, directional coupler, mechanical

phase shifter, and a DC coupled mixer. In the basic microwave interferometer, the

output of a microwave oscillator is fed into a directional coupler, where a portion of

the energy is splh off from the main output. The main output ofthe coupler is passed

on to a horn antenna and is propagated across the plasma volume where h is received

by another horn antenna. The output ofthe receiving antenna is then input into the RF

port of a mixer. The microwave signal split off by the directional coupler is

22

Mecnan.ca Phase Adjuster"

Decoupled . Horn Antenna

Mixer ^ , ^ / /

1 • ^ V

Thruster Plume

>

Horn Antenna

Directional Coupler

-—.

Isolator ^/r~\^\

Microwave Oscillator

. A . Output

Figure 4.1. Schematic diagram of a basic microwave interferometer.

passed through a mechanical phase shifter and into the LO port ofthe mixer. Because

the mixer is DC coupled the IF port will contain a vohage proportional to the sine of

the phase difference between the RF and LO ports. Thus, from the output ofthe

interferometer setup, the phase difference between the signal propagating around the

plasma and the one going through the plasma can be obtained [11].

Although the basic microwave interferometer is adequate for some diagnostics,

it has many problems and limitations when compared to the measurement requirements

of this project. In particular, the far field plasma volume ofthe Hall thrusters on which

the diagnostics are to be performed is large compared to a typical laboratory plasma

where microwave interferometry has tradhionally been used. Because ofthe large

volume, the distance across which the interferometer will be used are large, thus

causing large signal losses in the reference leg of a conventional interferometer,

particularly at higher operating frequencies ( > 10 GHz ). In an effort to overcome the

23

signal loss problem, a microwave interferometer was developed v\ hich uses a low

frequency oscillator as the reference source

The phase sampling microwave interferometer, shown in Figure 4 2. uses a

100 MHz crystal oscillator as a reference source, thus reducing the power loss along

the reference leg ofthe interferometer. In this system, a 6 GHz multiplied cavity

oscillator was phase locked to the reference source using a sampling phase detector

The loop fiher was adjusted to provide a 3 dB bandwidth of 500 kHz. The loop gain

was set sufficiently low to allow a minimum DC phase error of 30 degrees relative to

the reference source. By allowing the large DC phase error, an external control

voltage was added to the automatic frequency control (AFC) vohage, thus providing

an inexpensive electronic phase adjustment in the system. The output ofthe phase

locked oscillator was then transmitted through the plume of a Hall thruster using two

horn antennas. The received signal was input into another sampling phase detector

where its phase was compared to that ofthe 100 MHz reference source. The

SRD Pulse Generator

Mechanical Phase Adjuster

Thruster Plume

Horn Antenna Horn Antenna

100 MHz Xtal. Osc

BRD Pulse Generator

X

6 GHz Multiplied Cavity Osc, w/AFC

^

^ \

Phase Detector Output •Electronic Phase Adjustment

Figure 4.2 Block diagram of a sampling microwave interferometer.

24

output ofthe receiving sampling phase detector was then fihered for aliasing and

amplified and buffered to increase the signal-to-noise ratio at the measurement site

The buffered output ofthe interferometer was connected to a high-speed

analog to dighal (A-to-D) computer board in a 486 computer. The phase data were

collected using a commercial software package that was programmed to provide a

customized interface for this system. Because the maximum sampling frequency ofthe

A to D board was 20 MHz, the system bandwidth was limhed to the 10 MHz Nyquist

frequency. Although the data acquisition system bandwidth is an order of magnitude

lower than that ofthe sampling interferometer, the expected phase shift frequencies are

in the range of DC to 150 kHz and, therefore, the data acquisition system bandwidth

does not significantly limit the performance ofthe overall system.

The primary limitation ofthe system was the high quantization error associated

with using an eight bh A-to-D converter in the data acquishion system Because the

interferometer can uniquely distinguish any phase difference between -90 and +90

degrees, the output ofthe data acquishion system was limited to steps of 0,7 degrees.

In order to overcome this problem, the output amplifier ofthe interferometer was

made adjustable to provide a means of controlling the ftill-scale output vohage. Three

gain levels were chosen to provide fijll scale readings of + - 90 degrees, + - 45

degrees, and + - 25 degrees. These fijll-scale readings correspond to phase

quantizations of 0.7, 0.35, and 0.098 degrees, respectively The background noise

level ofthe interferometer output is highly dependent upon the stability ofthe horn

25

antenna mounts and RF cable vibrations Typical levels of noise were in the range of

0.1 degrees to 0,5 degrees.

The system was calibrated m-situ using a mechanical phase adjuster in the

reference leg. The reference phase was adjusted through a length equal to 360 degrees

at 6 GHz, while the output ofthe interferometer was recorded From the recorded

output, the peak deviations, both positive and negative, were determined, and were

used as coefficients in a fijnction used to convert the output ofthe interferometer to

degrees. Because the output ofthe interferometer is proportional to the sine ofthe

phase difference, the output must be transformed using an inverse sine fijnction ofthe

form.

Phase{l\^,^^ = sin V

ir.l (41)

where \',nt is the interferometer output voltage, V.nnax is the maximum interferometer

output voltage (corresponds to a 90 degree phase difference), ^ ntInln is the minimum

output voltage (corresponds to a -90 degree phase difference), and Phase(\',ni) is the

phase difference in degrees.

26

CR'\PTER 5

PHASE SHIFT MEASLTIEMENTS

The microwave interferometer was used to verify the accurac\ ofthe plume

model developed in Chapter 3. In order to accomplish this, phase shift measurements

were made at several transmission angles and distances relative to the thruster plume

The measured phase shifts were then compared to the theoretical values obtained with

the plume model. During these sets of experiments, the interferometer was nulled

whhout the thruster running. The acquisition system recorded the output voltage at a

sampling frequency of 250 Hz for 32 s. The thruster was started approximately 5 s

after the data acquisition was inhiated. The acquired waveform was then transformed

using Equation (3.3) The resulting waveform gives the phase difference between a

microwave signal propagating with and without the thruster plasma plume present

The first set of tests was conducted on axis with the signal propagation path

parallel to the thruster exh plane. The distance from the exit plane was 0,5 m, A

sketch ofthe test setup is shown in Figure 5 1, Due to time constraints in the vacuum

tank facilities at NASA Lewis Research Center, the T-lOO was the only thruster tested

at this distance. Figure 5.2 shows the measured phase shift of a 6 2 GHz signal during

the startup of a T-100. The T-lOO exhibits a large over-shoot before stabilizing to a

value of 45 deg. The overshoot is probably due to ionization ofthe Xenon atoms

diffijsing out the end ofthe thruster before ignition. After ignhion,

27

/ > « '.'"..'..'...

2,0 m

Figure 5.1. Sketch of parallel propagation path (relative to the exit plane).

100

80

60

00

40

20

0 -1

1

IL p. 1 V

t|i>.'VWffA«JMV^ '^H^VV^fl^Jy ' ' • ^ i i i L

^ ^ ^**^*Asy,r>y,>*A

l iJV^

AitJT

1.8 2.686 3.571 6.229 7.114 4.457 5.343

time (s)

Figure 5.2. Parallel start-up phase shift at 0.5 m for a T-100.

28

the accelerated ions collide with the neutral atoms, causing the atoms to ionize and

undergo a momentum change in the direction ofthe exhaust plume Using the

Mathcad worksheet contained in Appendix B, the line-integrated phase shift of

Equation (2.3) was calculated using the number density profile described by

Equation (3.1). A phase shift of 45 deg corresponds to a 1 m centerline number

density, no of 0.6 x 10 ^ m'^

A similar start-up plot for the SPT-100 is shown in Figure 5 3. In this plot the

start-up phase shift is approximately 27 deg. The propagation path geometry' is similar

to that in Figure 5.1 except that the distance down from the thruster exit was changed

to 0.9 m because a different vacuum tank was used. Usine the same Mathcad

worksheet for perpendicular propagation paths, a 25 deg phase shift gives a

corresponding value of no of 0.62 x 10 ^ m" . The start-up plot for a D-55 is given in

Figure 5.4. The propagation path was identical to that ofthe SPT-100 test In this

plot, the phase shift is approximately 27 deg at 6.2 GHz. The corresponding value of

no at this distance and propagation path is 0.65 x lO' m ' The D-55 and SPT-100 did

not exhibh the same number density overshoot, during startup, at 0 9 m as did the T-

100 at 0.5m.

The next set of start-up tests used the propagation path shown in Figure 5.5.

The path crossed through the plume plasma on axis at a 45 deg angle and was received

1 m down stream from the thruster exit plane. The main limitation with this

configuration was the finite tank size which caused some ofthe plume to deflect back

29

into the propagation path With the additional particles along the propagation path,

the measured phase shift should be greater than that predicted by the model

40

30

^ 20

I'M* I t)i/i .r'

wm'^i A

in ^ H I

' , 1 II' M M H' l ji ^ ^ ^ ^

3.714 6.5"1 286 4 429 5 143 5.857

time (s)

Figure 5.3. Perpendicular start-up phase shift at 0.9 m for an SPT-100.

30

20

M 10

-10

" • . ^ . — ^

^.^- -J-f.J.'^'^'' v-.- _- V

2.429 2.857 3.286 3."14

time (s)

4.143 4.571

Figure 5.4. Perpendicular start-up phase shift at 0.9 m for a D-55

30

Figure 5.6 shows the startup phase shift for the D-55. The measured phase

shift shown in the figure is approximately 50 deg. Appendix B contains a Mathcad

worksheet developed to calculate the phase shift along diagonal propagation paths.

Using this worksheet, a predicted phase shift of 50 deg for this thruster corresponds to

a 1 m centerline number density, rio, of 0.66 x 10 m . Although the increase of no

J6 „ -3 (from 0.65 x 10 m " to 0.66 x 10 m ) between the perpendicular and diagonal

propagation paths is not beyond experimental error, the number does indicate that the

finite tank size can be causing an artificial increase in the plume number density along

the propagation path. The startup phase shift waveform for the SPT-100 along the

same propagation path is shown in Figure 5.7. The SPT-100 shows a phase shift

Figure 5.5. Sketch of 45 deg propagation path.

31

bU

50

40

^ 30

20

10

0

1

i

i 1

' • _ * / - ^ — ^ •

6.286 6,571 7.429 7.714 6.857 7.143 time (s)

Figure 5.6. Startup phase shift along a 45 deg propagation path for a D-55

8

of approximately 52 deg along the diagonal propagation path. A centerline number

density of 0 68 x lO' m" corresponds to this phase shift. As whh the D-55 diagonal

propagation path startup waveforms, the SPT-100 shows a slight increase in no due to

the fimte tank size.

The next set of startup traces was taken whh a propagation path on axis 0.5 m

to the side and perpendicular to the exh plane of a D-55. The receiving horn was

placed 2 m from the exh plane, and as was the case for the diagonal propagation path,

the finite tank size can cause an artificial increase in the number density along the

propagation path. Because the tank diameter is approximately 2 5 m, and the D-55

thruster has a significant ion flux at 45 deg, deflected ions will cross back into the

propagation path at a distanced of 1 m and greater away from the exit plane. The

phase shift for the D-55 perpendicular startup shown in Figure 5.8 is approximately

32

60

40

20

0 •*"~ • ~ 2,5

I ' . L 1 h ^ ' ' ;

,i. ^^^^AVV' '."'''V // rvvr '' - ' > •,W'^'>V

3,5 5,5 4 4,5 5 time (s)

Figure 5 7. Startup phase shift along a 45 deg propagation path for an SPT-100.

25

20

15

r

1

i

1

i

. J

f 1

1

i t * ; " " ' '

r

'.••.T-'^-'-^j,""rAi»u*->, . . . . '-> . -

0.5 1.571 2,643 3,714 4.786

time (s)

5,857 6,929

Figure 5.8. Startup phase shift along a propagation path perpendicular to the exh plane of a D-55

33

22 deg. Changing the propagation angle to 0 deg and the exit distance to 2 m in the

diagonal Mathcad worksheet, resuhs in a value of 1 x lO' m' for no. The predicted

phase shift along this propagation path with no equal to 0.6 x lO'Ss 15 deg.

The final startup test was conducted on an "end-of-life ' SPT-100 thruster in a

large tank facility at NASA Lewis, Unlike the previous tests, this test was conducted

in conjunction with a plume deposhion experiment and therefore cylindrical

collimators were placed along a semi-circle in the thruster's exhaust plume. Due to

limitations on time and on the number of suitable mounting points in the tank, the

transmitting and receiving horn antennas were positioned such that the propagation

path was in the same plane as the collimators. In particular, the propagation path was

on axis and started 1.8 m to the side ofthe thruster and traveled at a 31 deg angle

relative to the exit plane, whh the receiving antenna 6 m from the exit plane. Because

the collimators partially blocked the accelerated ions along the propagation path, the

phase shift measured was expected to be smaller than predicted by the plume model.

Figure 5,9 shows the startup phase shift for the end-of-life SPT-100 The

oscillatory startup was caused by an incorrectly set current limit and is not a normal

phenomena. The measured phase shift is approximately 9,5 deg This corresponds to

an no of 0.5 x 10 ^ m' to achieve the same phase shift according to the plume model.

Although the lower phase shift is probably due to the collimators, inaccuracies in the

antenna placements and the fact that the thruster was tested at the end-of-life could

also account for the difference. Overall however, the predicted phase shift is within 15

percent ofthe model prediction,

34

12

10

^. 6 i ln./i i l ' l '

(f'14

+t TTTTT l\Ul'

M,

, ! !•

^.^'f-

T T

-H-

' .lililllhj M - ' : I

P i l i ^

il ' II II !i II

I t 1

, . (r

28,85" 35.'14 42 5"1 49 429 56.286 63 143 70

time (s)

Figure 5.9 Startup phase shift along a diagonal propagation path for an end-of-life SPT-100,

The last ofthe low speed (i,e,, several seconds) phase shift measurements were

made on the D-55 thruster. During these tests, the cathode flow rate was changed

from 0,2 mg/s to 1.6 ma s while the thruster was running. The interferometer was

zeroed whh the thruster running at the lower cathode flow rate The propagation path

was the same diagonal path described previously for the D-55 tests. In Figure 5.10 the

phase shift changes 29 deg between the minimum cathode flow rate and the maximum

flow rate. Based upon the previous resuhs and addhional tests conducted away from

the designed operating point ofthe thruster, the value of no was determined to have a

dependence on the cathode flow rate greater than ± 20 percent ofthe nominal value

Although significant charge exchange with the tank walls will occur at the two cathode

flow rate extremes, this does indicate the influence ofthe cathode on the thruster

plume denshy

35

30

20

Ob

10

0

/

\ 1

/

i

> •

^V^-jr^.-nri — , — 1 r — - - • — ^ - * — - L ^ - Y . • •4^.^.^^>-^

0 5.714 11.429 17.143 22 857 28.571 34.286

time (s)

Figure 5,10, Change in phase shift due to change in cathode flow rate for a D-55 (0,2mg/s at 2 sec linearly to 1.6 mg/s at 28 sec).

40

The next set of experiments was designed to provide information on the

frequency and magnitude of phase fluctuations that a microwave communication signal

would experience propagating through the plume of a Hall effect thruster. During

these tests, the microwave interferometer was adjusted to provide an average phase

shift of zero while the Hall thruster was operating under steady-state conditions The

output ofthe interferometer was sampled and recorded at a rate of 10' samples per

second, resuhing in a maximum detectable frequency of 5 MHz, The recorded output

was then transformed using Equation (4,1) and plotted to illustrate graphically the

transient phase shifts. In addhion, the current waveform was also acquired and is

plotted for comparison whh each phase shift plot. Because these tests were conducted

in conjunction with the start-up tests discussed previously, the previous section should

36

be referred to for a complete description ofthe propagation path, as only a brief

description ofthe path will be given in this section.

Figure 5.11 shows the measured phase shift transients caused by a T-100

plasma plume with the parallel propagation path discussed previously. Figure 5.12 is

the associated current waveform taken at the same time Figure 5.11 indicates that the

T-100 is causing phase transients in excess of 20 deg peak to peak, A comparison of

Figure 5.11 and 5.12 shows a distinct correlation between the amplitude and frequency

ofthe current oscillations to those in the phase shift plot In a previous section, it was

shown that the T-100 exhibited a current sensitivhy factor Csem of 0,225, which relates

the current oscillations to the number denshy modulation factor, m In a time window

from 3.1 ms to 3.3 ms, Cmod% is found to be 0.5 which resuhs in a modulation factor,

m, of 11 percent. As a comparison, the predicted phase oscillation using the temporal

plume model are shown in Figure 5,13, The plot shows the predicted

20

4 0

-10

-20

, ' • I 1 . ' il 1

', '-^ >f. • 1' , t I '

II

f

f

{

1,

t " * r ' 1

. . 1 - ' ' i

li 1

i

V

t

1

U 1 '

1 ' 1

1 ,

> . 1

"• 1

, ' 1

t>

( ' ' , ' ' .' •'

1 1 1

1 ''' ^ • • , 1 - ,

1 >

r ( 1 i: (1 "

• . ' , ' • -

' - 1 V

1 '

' P./ -»

1 1

1 ' /I

' " A 1 .. ' W

!• ' '

j '

1

1

i\

11 1

* t>

2,5 3,5 time (ms)

4.5

Figure 5.11. Measured phase transients of a 6 GHz signal propagating along a parallel path (relative to the exit plane) through the plume of a T-100.

37

10

l i l

I I I ! , , , . , . , , | l i | , ,

1

1

' 1 rl: > 1 1 . i 1

•1

1

I , , h ' .

• ,1

' i ' I ' r , '

1 . 1 ' I

'hi' 1 ' ' '

J I

.11-1. A I ' V - '

Ml 1

2,5 35

time (ms)

4,5

Figure 5.12. Current waveform for the T-100 in Figure 5 11

60

55

50 . , r ! i\ I 1

' I I ,

iL 45

40

35

A A ;. /' A ' A l\ : > M ' I 1

TTTT iIlM/ Il !i I i I i M i

V V ^ . i!

30 1 25 5 10 15 20

modulation factor m (%)

Figure 5.13. Predicted phase shift along a parallel propagation path 0.5 m down from the exh plane as a fijnction of m, for the T-100

3u

phase oscillation as the modulation factor m is varied from 0 percent to 30 percent

From the graph h is found that a value of 11 percent for m results in a predicted phase

shift of 9 deg. This value is almost identical to the measured value in the same time

window. Likewise, other time windows can be examined with similar results.

Because the current oscillations ofthe T-100 are not constant, a direct comparison

cannot be made, however, the predicted phase shift is whhin a reasonable error ofthe

38

measured phase shift, given the instability ofthe thruster and the complexity ofthe

problem.

The phase transient data and associated current waveform for the SPT-100

along the parallel propagation path are shown in Figure 5 14 and Figure 5,15,

respectively Because the SPT-100 thruster was configured to operated with the

current to the electromagnets in series with discharge voltage, no magnetic field

augmentation was possible As a result, the current oscillations during this test were

considered excessive and not a true representation ofthe potential stability ofthe SPT-

100 thruster Subsequent results, presented below, will show that the SPT-100 is

highly dependent upon the discharge power supply design and filter In addition, the

use of a separate power supply for magnetic field augmentation has been shown to

reduce the amplitude ofthe current oscillations [3] To compare the accuracy ofthe

plume model and the current sensitivity factor, Csens, the predicted phase shift along the

given path for values of m ranging from 0 to 0,3 is shown in Figure 5 16, From Figure

5.15 the value of Cn,od% is determined to be 0.9, resuhing in a predicted modulation

factor, m, of 21 percent. A predicted phase shift of 9 deg corresponds to a 21 percent

modulation factor using Figure 5.16, This resuhs in an overall phase shift prediction

error of 10 percent given only the current waveform

During the 45 deg diagonal propagation path testing ofthe SPT-100, the

magnetic field coils were connected to a separate power supply, thus allowing the

magnetic field to be augmented. The separate supply allowed the thruster to be

adjusted to provide a more stable current waveform. Figure 5 17 shows the phase

39

10

^ 0

-5

-10

I—TT nn—rr •^TT

rtm i i i ' * ' '

M-1 1 I I

0.5

1 I

J L - T T - ^

TT

r

• 1 , I

I , > I . I TT -U-

2.5 1 1,5 2

time (ms)

Figure 5,14. Measured phase transients of a 6 GHz signal propagating along a parallel propagation path through the plume of an SPT-100,

10

8 • '

1

i l , 1

• • ' • ! ; i / '

j 1 1 ' 1 , ' t > . 1 . '

Ii 1 1 M

,! i l '

' ' ; ' • : ' "

, 1 • > 1 ^ ' V V » . . ' ,

i l ! I | (

i

. ' ' ' . > 1 • i » . , 1

. 1 , i . • 1

1

i '• i ' i I

,

1 • » 1 . 1 1 , 1 " ' ' * '

1 t

(

, !

1! 1

1 11

1 ! !

'' 1

^1

' 1 1. I • ! 1 1 . ! 1

0.5 1.5 1 ^

time (ms)

Figure 5 15 Current waveform for the SPT-100 in Figure 5 14

35

30

^ 25

20

15

-^"^ JL A

^' V V f^ f\ A A A '\ l\ \ i\

i I

mnn V I' I . 1 • M M ! l i 1 y i' ! n I 1

V I

1 10 15 20

modulation factor m ("o) 25 30

Figure 5 16, Predicted phase transients of a 6 GHz signal along a parallel propagation path 0,9 m down from the exh plane as a function of m

40

transient waveform recorded during this test with the associated current waveform in

Figure 5.18. Examining the current waveform in a time window from 0.8 ms to 0.9

ms gives a value of 0.375 for Cn,od%, resuhing in a predicted modulation factor, m, of 8

percent. Figure 5.19 shows a predicted phase transient of 5 deg for this value of m

along the described propagation path. In Figure 5 17, the measured phase shift during

this time window is approximately 5 deg, thus showing that the phase shift prediction

by the model is accurate.

During the operation ofthe D-55 thruster, the current oscillations were small

compared to those ofthe SPT-100 and T-100 thrusters. As a result, the predicted and

measured phase transients were also smaller. Figure 5.20 shows the measured phase

transients for the D-55, whh the previously described parallel propagation path.

Figure 5.21 is the associated current waveform for Figure 5.20. Because the plume of

the SPT-100 and the D-55 are similar, the predicted phase shift plot in Figure 5.16 can

be used for model verification ofthe D-55. In the time window of 1 ms to 1.5 ms the

value of Cmod% is 0.275, resulting in a predicted value of 7 percent for m Referring

back to Figure 5.16, a value of 7 percent for m corresponds to a predicted phase shift

transient of approximately 4 deg, which compares to the 5 deg phase shift measured

during this time window.

The next set of phase transient data is for the D-55, and was taken along the

diagonal 45 deg propagation path. Figure 5.22 shows the measured phase transients

along this path and, likewise, the associated current waveform is shown in Figure 5.23

In a similar argument as before, the SPT-100 predicted phase transient plot

41

•^ 0

, , . ^ . .

• • . P l ^

'I ih.

— i -

!...., 1: •/ (

11( I

1 . 1 I r

\'::vj —TH

iM

' • • ' • ' • ' ' - h j r : ' ^•'iijli^

I . < l i

. ' . . > - ' •

' • . , :

0,5 •> « 1 15 : time (ms)

Figure 5.17 Measured phase transients of a 6 GHz signal propagating along a 45 deg diagonal propagation path for an SPT-100

10

8

, 1 ' l . l .

r. ' I

, , - : ' - | ' > '

• ! , i ' <

1

\

t 1 r . . 1

1 1 1

( • '

i 1

Ij'l

• • 1

' 1 1 ( ' - 1

\

' 1

!

1

• '}•:

iM ' l i , ,

- • • • • • i , • • , • ^ • •

1 0 417 0 833 1 t6' 083 1 25

time (ms)

Figure 5 18 Current waveform for the SPT-100 in Fiaure " P

6C

55

.2 50

45

40

' ^ . - / u W i l l i l l i i ' n

I ! > I

^^v i ' i ' v . .V i i j |H jy j | j j i

10 15 20 modulation factor rr, ('o)

3:

Fieure 5,19. Predicted phase transients along 45 deg diagonal propagation path at 1 m down from the exh plane as a fijnction of m

42

for the 45 deg diagonal propagation path will be used for comparison to the model In

a time window from 0.4 ms to 0.7 ms, the value of C od is 0 2, which gives a

predicted modulation factor, m, of 5 percent. Figure 5.19 gives a predicted phase shift

transient value of 3 deg for this value of m. A comparison ofthe 3 deg prediction and

the measured 1.5 deg shows a 50 percent error Although this is a large percentage

error, the difference error is small compared to the static phase shift that the transients

are added to.

The final phase transient data were taken on the end-of-life SPT-100 in the

large vacuum tank facility. The propagation path was the 31 deg diagonal path

described previously. During this test, an integrated power processing unit, that

contained all the necessary power supplies and mass flow controllers, was used to

operate the thruster. The power processing unh was designed and optimized

specifically for the SPT-100 and is the same type that will be used on fijture space

missions involving the SPT-100. Figure 5.24 shows the measured phase transient

waveform recorded during this test. The associated current waveform is shown in

Figure 5.25. A significant decrease in the amplitude ofthe current oscillations over

previous tests involving the SPT-100 is apparent. As a result, the measured phase

transients are expected to be smaller in amplitude as well. Figure 5.26 shows the

predicted phase transients along the propagation path for values of m from 0 to 0.3.

From Figure 5.25 a value of Cmod% can be calculated to be 0 275 in the time window

from 0 to 1.5 ms. This corresponds to a predicted modulation factor, m, of 6.1

percent. In Figure 5.26, an m factor of 6,1 percent corresponds to a predicted phase

43

.^ 0

-2

:iiiuMiML m Ti ijjjhift .jiLllJ-MJMilul r iM

0.5 15 time (ms)

Figure 5.20. Measured phase transients of a 6 GHz signal propagating along the parallel propagation path through the plume of a D-55

10

'•'f i''>t'^^^''"•^'' ;'.^'jj4iv,.ic>->^ ;,F,.v„-.>'..' - ,'vAi' ^^[ ihSlTi i rv; '

0.5 1 15 :

time ims)

Figure 5,21, Current waveform for the D-55 in Figure 5 20

= 1

fJl E Ul iiUI i ! I l

JLOI n 0.1 0.2 0,3 0.4

time ims i

0.6

Figure 5,22 Measured phase transients of a 6 GHz signal propagating along a 45 deg diagonal propagation path for a D-55

44

10

» , - w

T !iL-i:^ W-.' ' iU'\ . j '4. '--- ' .

— • • ? ' ' « • i I i l-W' / ' -' ' '*, ' l I I " ' . * - • 1 •-. 'J :i=X

, li J ^ , •< ^ - «

0.1 0.2 0,5 06 0,3 0 4 time (ms)

Figure 5,23 Current waveform for Figure 5,22,

0,7

w iMl ^!'i'

m .f'V •4

fall \

',T"V ^11', i.i ?

fe^ ''t;V ' :f

-1

time (ms)

Figure 5.24. Measured phase transients of a 6 GHz signal propagating along a 31 deg propagation path for an end-of-life SPT-100.

10

^il11|l|#' \m^ Miii\ii\W imM^kMmfW^ 1^ j t 4yL,.

1 2 3 4 5 6 tmie (ms)

Figure 5.25. Current waveform for Figure 5 24.

45

114

11.3

ac

11.2

11.1 10 15 20

modulation factor m (%)

25 30

Figure 5.26, Predicted phase transients along 31 deg diagonal propagation path at 1.8 m to the side of an SPT-100 as a fijnction ofm

shift of 0.08 deg. In comparison, the measured phase transients during the same time

interval are greater than 0,2 deg, which corresponds to an error in excess of 40

percent. Again, due to the small phase transients being investigated along this

propagation path, the difference in the two values is small enough to make engineering

decisions about the potential impacts ofthe thruster based upon the developed plume

model.

Further examination of Figure 5 24 shows that an unexplainaed phase jump in

excess of 1,5 deg occurs at 1.75 ms, which correlates to a small current drop at

approximately 1 7 ms. The small current drop is probabK due to the closing and

opening of a thermal valve used in the regulation ofthe mass flow to the anode. The

phase transient that followed the current drop is much larger than can be explained by

the model, and may be a random occurrence caused by system noise or most likely by

factors not considered in the model, such as material randomly flaking from the

discharge chamber, perturbing the thruster's plume. Previous studies on the SPT-100

46

have shown that random periodic flaking of deposited material on the discharge

insulator occurs after several hundred hours of testing [12, 13] The researchers

conducting these studies have theorized that the material is due to the vacuum tank

facilities and will not be present in the space environment In any event, this

phenomenon may require fijrther investigation to build a more accurate description of

the potential communications impact of this thruster.

47

CHAPTER 6

PHASE NOISE

In Chapter 5, h was shown that the phase of a microwave signal is changed as

it propagates through the plume of a Hall effect thruster. Although the phase

fluctuations were found to be oscillatory in many instances, a random or non-

deterministic component was present in all ofthe recorded phase transient

measurements. Due to these random phase fluctuations, the power spectral density of

f(t)r in Equation (2.6) will show broading around the carrier frequency, ©c,

proportional to the amplhude and spectral content ofthe phase fluctuations caused by

the plasma plume. Although an increase in phase noise will impact the performance of

communication systems in different ways, depending upon the application and

modulation scheme, the most notable is the limitation on how close adjacent

communication channels can be spaced in the frequency domain. Because the system

degradation due to phase noise cannot be overcome by addhional transmitter power,

any increase in phase noise due to the transmitting media would be considered a

source of concern for most communication systems. It should be noted that the phase

•noise should not be mistaken for the typical noise figure that is often stated for

communication systems, which can be reduced by increased transmitter power

Typically, the system phase noise is dominated by the phase noise ofthe

transmitter and receiver oscillators. In high quality satellhe communication systems,

the microwave oscillators are designed to provide low phase noise and are usually

state ofthe art. When determining the system phase noise, a communication system

48

integrator will typically use the output phase noise ofthe oscillators and include a

small increase due to the amplifiers, mixers, and other components in the system.

Overall, however, the peripheral components contribute a small fraction to the phase

noise when compared to the phase noise ofthe local oscillator Historicalh', the phase

noise increase due to the transmission media was limited to random atmospheric

disturbances such as lightning strikes and is therefore not typically considered a major

contributor to the system phase noise parameter. Because the plume of a Hall effect

thruster has been shown to change the phase of an RF signal propagating through it,

an investigation to determine if the plume significantly increases the phase noise ofthe

RF signal was made

To qualify and quantify the effects ofthe Hall thruster's plume on RF signal

phase noise, a test setup was devised which transmits a microwave signal through the

plume using a low phase noise cavity multiplied oscillator as the transmitting source

A pair of microwave horn antennas were placed across the plume, with the source

oscillator feeding the transmitting antenna and the output ofthe receiving antenna

being sent to a microwave spectrum analyzer, A diagram ofthe test setup is shown in

Figure 6 1 During the operation of this test setup, a baseline ofthe power spectral

density ofthe received signal, propagating along a given path, was obtained without

the thruster running using the spectrum analyzer. The data were downloaded and

stored for future plotting using a laptop computer. After the baseline signal was

obtained, the thruster was started and allowed to stabilize. After stabilization, the

49

Spectrum Analyzer

il D D D D D D D D D

O

, /

' \

Horn Antenna

Thruster Plume

^ . • ? • •

Horn Antenna 5 QHZ r.lultiplied Cavity Osc. Phase

Locked to a 100 MHz Xtol. Osc.

Figure 6.1. Phase noise test setup.

power spectral density ofthe received signal was acquired for comparison to the

basehne spectrum. In each ofthe following spectrum plots, the dashed line is the

baseline signal, and the solid line is the received spectrum with the thruster running.

As with the phase transient measurements, the phase noise measurements were made

in conjunction whh the startup phase measurements and thus the propagation paths

described in Chapter 5 should be referenced, as only a brief description will be given in

this section.

Figure 6.2 shows the received power spectral denshy of a 6 GHz signal passing

through the plume of a T-100 along the parallel propagation path, 0.5 m from the exh

plane. Because the settings ofthe spectrum analyzer, such as the frequency span,

resolution bandwidth, and video bandwidth affect how the spectrum ofthe signal is

displayed, the settings of each plot were adjusted to provide the best representation of

the difference between the baseline signal the signal passing through the plume ofthe

thruster. In addition, each plot is normalized to 0 dBm at its peak. Figure 6.3 shows

the current waveform acquired at the same time as the spectrum plot in Figure 6 2 .\

comparison ofthe baseline spectrum to the spectrum

50

E T3

3

"S.

20

-40

-60

->/v^- ' ^

-80 6183.5 6183.667 6183.833 6184

Frequency (MHz)

6184167 6184333 6184.5

T-100 Running Baseline

Figure 6.2 Spectrum of a received signal with and without a T-100 running (1 MHz span).

Figure 6.3. Current waveform ofthe T-100 in Figure 6.2.

ofthe received signal with the thruster running shows that a significant increase in

the phase noise has occurred. At a carrier offset of 250 kHz the increase is greater

than 45 dBm. Although this is not a propagation path that is likely to occur on a

commercial satellite, it does indicate the potential for communication system

51

degradation. Figure 6.4 shows the spectrum comparisons for the SPT-100 with a

parallel propagation path 0.9 m from the exit plane. Figure 6.5 shows the current

waveform acquired at the same time as the data in Figure 6.4. Note that the primary

frequency ofthe discharge current oscillations is approximately 25 kHz. This

corresponds to the distance from the carrier to the first periodic peak in the spectrum

6182,9 6182,933 6182.967 6183 6183.033 6183.067

Frequency (MHz)

6183.1

SPT-100 Running Baseline

Figure 6.4 Spectrum of a received signal with and without an SPT-100 running (200 kHz span).

Figure 6.5. Current waveform ofthe SPT-100 in Figure 6.4.

52

ofthe received signal. It also is the distance between subsequent peaks in the

received spectrum, which follows the well-known communication theory of phase

modulated signals. In addition to the periodic peaks, a significant increase in the

phase noise across the entire 200 kHz span is present.

The next spectrum plot, shown in Figure 6.6, is for the D-55 with a parallel

propagation path. Figure 6.7 shows the associated current waveform for Figure 6.6.

Because the D-55 is characterized by broad band discharge oscillations, the number

density modulation has no primary frequency. As a result, the spectrum in Figure 6.6

shows broad phase noise with little sideband development. In addition, the level of

phase noise produced is significantly less than that ofthe SPT-100 or T-100. The

lower phase noise is attributed to the magnitude ofthe current oscillations being

much smaller. Because the spectrum analyzer used has a limited dynamic range of

approximately 75 dB at a 200 kHz span, an accurate comparison ofthe phase noise

618-4 966 6184.999

D-55 Running Baseline

6185.033 6185,066 6185099 Frequency (MHz)

6185133 6185,166

Figure 6.6 Spectrum of a received signal with and without a D-55 running (200 kHz span).

53

-;-/-nr;.'', \ ^ , ; / . A, . • 1

;• / .a.-• . ' . • -

• I ' I •

1

1

* •

V •

0.1 0.2 0.3 0.4 ms

0.5 0.6 0.7

Figure 6.7 Current waveform ofthe D-55 in Figure 6.6.

whh and whhout the thruster running is not possible. However, the minimum increase

in the phase noise level is clearly greater than 15 dB,

The spectrum plot for the SPT-100 along the 45 deg diagonal propagation

path is shown in Figure 6,8 with the associated current waveforms shown in

Figure 6,9. In these figures, the data acquired for two operating condhions ofthe

SPT-100 are shown for comparison. When operating whh the clearly unstable

discharge current shown in Figure 6.9 the power spectral density ofthe 6 GHz signal

shows a significant increase in sideband development approximately 30 kHz from the

carrier, corresponding to the primar>' frequency ofthe discharge current oscillation. In

addhion, an overall increase in the phase noise level is also evident. In contrast, the

more stable operating condition shows a significant decrease in the sideband

development as well as an overall decrease in the noise level. Even with the SPT-100

operating with a stable discharge current, the phase noise level is increased by 30 dB

or more along this propagation path.

54

6184.95 6184.967

- SPT-100 Stable — SPT-100 unstable

Baseline

6184.983 6185

Frequency (MHz)

6185.017 6185 03? 6185,05

Figure 6.8 Spectrum of a received signal with and without an SPT-100 running along the 45 deg diagonal propagation path( 100 kHz span).

ms

Figure 6.9. Current waveforms ofthe SPT-100 in Figure 6.8.

The next set of plots show the power spectral density and current waveforms

with a D-55 operating at a discharge current level of 3.5 A and 4.5 A. The

propagation path ofthe 6 GHz signal was the same 45 deg diagonal path used in the

SPT-100 test. Figure 6.10 shows a distinct increase in the phase noise level with the

55

D-55 operating at a discharge current level of 4.5 A compared to 3.5 A. The phase

noise level at 4.5 A is greater than 4 times the phase noise level at 3.5 A. The current

waveforms, shown in Figure 6.11, show no significant increase in the level of current

oscillations between the two current levels.

-80 6184.9 6184.933

D-55 3.5 A D-55 4.5 A Baseline

6184.%7 6185 6185.033 Frequency (MHz)

6185.067 61851

Figure 6.10 Spectrum of a received signal along the 45 deg diagonal propagation path with a D-55 operating at 3.5 A and 4.5 A (200 kHz span).

Figure 6.11. Current waveforms of the D-55 in Figure 6.10.

56

The final phase noise data are for the end-of-life SPT-100 operating in the

large vacuum tank facility at N.^SA Lewis Research Center. The propagation path for

this test was normal to the exh plane and 1.8 m to the side ofthe thruster body The

receiving antenna was 6 m from the exit plane. Figure 6 12 shows the received

spectrum of a 6 GHz signal propagating along the above described path Figure 6.13

shows the current waveform acquired during this test. As with the previous end-of-

life SPT-100 current waveforms, the magnitude ofthe current oscillations are small

compared to newer SPT-100, used in other tests. Although age may be a factor in the

reduced current oscillation level, the refinement ofthe discharge power supply output

filter is the most likely reason for the improvement. As a result, the signal phase noise

level is correspondingly lower In addition, the lower number density along this path

will also contribute to a lower phase noise level. The highest spectral noise level

occurs at approximately 25 kHz from the carrier, which corresponds to the primary

frequency ofthe discharge oscillation.

To verify the effects of transmission frequency, the spectrum of a 2 8 GHz

signal, propagating along the same path was acquired Because the phase shift caused

by a plasma is greater at lower frequencies closer to the plasma frequency, the phase

noise of a 2.8 GHz signal would be expected to increase more than that of a 6 GHz

signal, with all other parameters equal As can be seen, the measured power spectral

density ofthe 2.8 GHz signal, shown in Figure 6.14, has a significant increase in phase

noise over the 6.2 GHz signal. At 2.8 GHz, the end-of-life SPT-100 caused an

increase in the signal phase noise in excess of 40 dB, while at 6 2 GHz the thruster

57

thruster caused less than a 10 dB increase. It should be noted that the power spectral

densities in Figures 6.12 and 6.14 were acquired within minutes of each other, with

the same operating conditions and level of current oscillations.

-80 6212.9 6212.933 6212.967 6213 6213.033

Frequency (MHz)

SPT-100 Running Baseline

6213.067 6213.1

Figure 6.12 Spectrum of a 6 GHz received signal with and without an end-of-life SPT-100 running (200 kHz span).

^AA/V^VVV^U^''' V\ A/VVY-U->'VW /\rV|/^Vvi/VA/v / O A A ^ V ^ Y ^ 4/V\/\/vA^VV-v-'

0.2 0.4 0.6

ms

0.8

Figure 6.13. Current waveform ofthe end-of-life SPT-100 in Figure 6.12.

58

2801.625 2801.75 2801.875 2802 2802.125 Frequency (MHz)

— SPT-100 Running Baseline

2802.25 2802.375

Figure 6.14 Spectrum of a 2.8 GHz received signal with and without an end-of-life SPT-100 running (750 kHz span).

59

CHAPTER 7

CONCLUSIONS

An important engineering model was developed, which allows a satellite

system integrator to predict the impacts of Hall effect thrusters on communication

systems. Although the basic plume model is probably valid for all types of Hall effect

thrusters, plume parameters were determined for three thrusters in particular, the D-

55, SPT-100, and T-100. The plume model developed predicts the stationary electron

number denshy in the plume far field. The model also accounts for the temporal

dependence ofthe electron number denshy caused by the discharge current

instabihties. Langmuir probe measurements, both high speed and swept, were

performed to determine the spatial and temporal parameters ofthe plasma plume

In an effort to verify the plume model and refine it if necessary, a microwave

interferometer was designed and built. The interferometer uses a phase samphng

technique which allows accurate phase shift measurements to be performed across

large plasma volumes. Although the interferometer was designed to be broadband

(1 GHz to 20 GHz), the majority ofthe measurements were made at 6 GHz due to the

availability of low noise sources and the applicability ofthe measurements to the

current 6 GHz satellite communication links. The noise level ofthe interferometer

system was typically less than .05 deg at 6 GHz.

The sampling microwave interferometer was used to perform startup phase

shift measurements on the three thrusters. Using several propagation paths, the phase

60

shift predicted by the plume model was compared to the measured phase shifts along

the same paths. The plume model accuracy during the startup tests was typically

better than 90 percent. The largest source of error was most likely from the effects of

a finite test chamber, causing charge exchange and plume reflux.

In addhion to the startup phase measurements, the interferometer was used in a

high-speed mode to determine the amplitude and frequency of phase oscillations and

transients caused by the plasma plume on a microwave signal. Due to the dependence

ofthe number denshy on the discharge current, the frequency and amplitude ofthe

phase shift oscillations correlated closely to those ofthe discharge current The

factors that affected the phase shift oscillations the most were the propagation path,

discharge current oscillations, and cathode mass flow,

A microwave spectrum analyzer was used to acquire the power spectral

density of a microwave signal after passing through the plume of a Hall effect thruster

In each ofthe tests conducted, the phase noise increased by a minimum of 10 dB, and

in some instances by more than 50 dB for a 6 GHz signal. As with the phase transient

measurements, the amount of phase noise increase was dependent upon the

propagation path and discharge stability Due to the limited dynamic range ofthe

spectrum analyzer, an accurate measurement ofthe phase noise increase was not

possible, therefore a best-case estimate was made by comparing the noise floor level of

the analyzer to the power spectrum obtained with the thruster running.

An important distinction should be made as to the origin ofthe phase noise in

the carrier signal. Due to the time varying dielectric constant caused by the thruster's

61

plasma plume, the actual phase ofthe carrier signal is being changed. The noise

contributions caused by amplifiers, mixers, and other devices are typically caused b\

additive noise signals. As a resuh, an increase in the original signal amplitude (before

the noise is added) will give a better signal-to-noise ratio. In contrast, because the

carrier wave itself is being modified (the phase ofthe wave is varied in a random or at

least non-deterministic manner), any increase in amplitude will not increase the signal

to noise ratio.

A phasor illustration ofthe plume induced phase noise and an additive signal

noise is shown in Figure 7.1. From the illustration, h can be seen that the additive

noise signal causes both a phase modulation and an amplitude modulation. In

comparison, the thruster plume causes only a random phase modulation. As a result

the plume's impact will affect phase or frequency modulated signals the most.

Thruster Induced Phase Noise

Carrier Signal

Random Phase Jitter

Typical Additive Noise

Noise Signal X <

Resulting Signal

Carrier Signal

Figure 7.1. Phasor illustration of thruster induced phase noise and typical addhive noise.

The majority of all satellite communication systems use some form of phase or

frequency modulation. Depending upon the type of modulation scheme utilized and

the transmission data rate, the effects ofthe thruster induced phase modulation may be

62

overcome by increasing the signal to noise ratio. In the worst case such as .\l-array

modulation, where muhiple phase states are used to transmit the data, a complete loss

ofthe communication channel is possible if the magnitude ofthe thruster induced

phase shifts exceed the phase spacing between adjacent states

For the case of a typical north-south station keeping mission, the thruster's

plume should not significantly impinge upon the communication system's uplink and

down link paths. As a result, the most likely effect ofthe thruster induced phase noise

will be a slight reduction in the overall signal-to-noise ratio present at the receiver. In

the case of large phase transients caused by arcing or large current instabilities, the

possibihty exists for intermittent frame losses, again depending upon the modulation

scheme and antenna placement (relative to the thruster plume). As a general rule,

significant increases in phase noise or the occurrence of large phase transients at

frequencies of 4 GHz and above will be avoided if the transmission path lies outside of

a 45 deg cone for a distance of 8 m away from the thruster's exit. At lower

frequencies or higher modulation constellation densities (16 or 256 Q.A\I), a larger

avoidance angle may be required.

The following points should be remembered when assessing the overall impact

of a Hall effect thruster on communications systems. The level of phase noise and the

magnitude of phase transients is propagation path dependent, and the plumes' effects

are reduced at higher frequencies In addition, the level of discharge current

oscillations relate directly to the amplitude and frequency ofthe phase shifts

63

Although for most communication satellhe missions, the placement ofthe

antennas can be made so as to prevent the communication signals from passing

through the plume, exploratory research probes may require the data signal path to

pass through the plume. In this case, the carrier frequency should be selected as high

as possible with all other factors considered. In addition, several other possibilities

exist to reduce the plumes' effects. The greatest impact in reducing the plumes'

effects can be realized by stabilizing the discharge current. During the tests on the

end-of-life SPT-100, the discharge current was very stable and thus the levels of phase

noise and phase oscillations were minimized. The filter network in the power

processing unit appeared to be optimized for the thruster at this point in its life

Earlier tests on the same thruster at NASA Jet Propulsion Lab showed significant

levels of discharge current oscillations at different stages of thruster life [3]. Research

should be conducted to find a way of reducing the oscillations throughout the life of

the thruster. Another area to be considered is the reduction of discharge current, with

a corresponding increase in the discharge voltage. By reducing the mass flow, the

electron number density in the plume should be reduced if double or triple ionization

can be avoided. The lower number density means that the magnhude ofthe phase

oscillations are reduced for a given number denshy modulation factor.

64

REFERENCES

1. J. M. Sankovic and T. W Haag, "Operating Characteristics ofthe Russian D-55 Thruster with Anode Layer," NASA Technical Memorandum 106610, 30th Joint Propulsion Conference AIAA-94-3011, Indianapolis, IN, June 27 - 29, 1994.

2. J. M. Sankovic, J. A. Hamley, and T W Haag, "Performance Evaluation ofthe Russian SPT-100 Thruster at NASA LeRC," NASA Technical Memorandum 106401, 23rd International Electric Propulsion Conference, IEPC-93-094, Seattle, WA, September 13 - 16, 1993.

3. C. E. Garner, J. E. Polk, K. D. Goodfellow and J R, Brophy, "Performance Evaluation and Life Testing ofthe SPT-100,'" 23rd International Electric Propulsion Conference, IEPC-93-091, Seattle, WA, September 13 - 16, 1993

4. Greenstein, Larry J. and Shafi, Mansoor, Micros are Digital Radio, IEEE Press, New York, 1988

5 Bhargava, Vijay K., Haccoun, Da\id, Matyas, Robert and Nuspl, Peter P., Digital Communications By Satellite, John Wiley & Sons, New York, 1981

6. Lynnette M. Carney, "E\-aluation ofthe Communications Impact of a Low-Power Arcjet Thruster,"' 24th Joint Propulsion Conference, AI.AA.-88-3105, Boston, MA, July 11-13, 1988.

7. John Mankowski, Master's Thesis, Texas Tech Univershy, 1995.

8. David H. Manzella, NYMA, Inc. at NASA LeRC, Personal Communication, 1994,

9. Roger M. Myers and David H, Manzella, "Stationary Plasma Thruster Plume Characteristics," 23rd International Electric Propulsion Conference, IEPC-93-096, Seattle, WA, September 13-16, 1993.

10. David H. Manzella, "Stationary Plasma Thruster Ion \'elocity Distribution," 30th Joint Propulsion Conference, AIAA-94-3141, Indianapolis, IN, June 27-29, 1994.

11 Auciello, Oriando and Flamm, Daniel L., Plasma Diagtiostics. Academic Press, Inc. San Diego, CA, 1989

65

12. T Randolph, G. Fischer, J. Kahn, H Daufman, \ ' Zhurin, K Kozubsky, \ ' Kim, "The Mitigation of Discharge Oscillations in the Stationary Plasma Thruster," 30th Joint Propulsion Conference, Al.AA-94-2857, Indianapolis, IN, June 27-29, 1994.

13. T. Randolph, M. Day, V Kim, H Kaufman, V. Zhurin, K, Kozubsky, "Facility Effects on SPT Thruster Testing," 23rd International Electric Propulsion Conference, IEPC-93-093, Seattle, W A, September 13-16, 1993,

66

APPENT)IX A

L.\NGMLTR PROBE PLOTS

67

c a.

<

a. c t 3 O

"O

c cd a> o

3

E 00 c cd

•o c cd

C/5

5 j = .«—>

x: C M o

• < — '

E cd k> cd

a. 00 cd

a. O

<

Cur

rent

w

avef

orm

F

igur

e

Lan

gmui

r P

robe

Fig

ure

Ano

de

Flo

w R

ate

mg/

s

Cat

hode

F

low

Rat

e m

g/s

Inte

rnal

M

agne

t C

urre

nt (

I)

Ext

erna

l M

agne

t C

urre

nt (

I)

Dis

char

ge

Vol

tage

(V

)

Dis

char

ge

Cur

rent

(A

)

Thr

uste

r T

ype

Fig

ure

A.2

. F

igur

e A

. 1.

45.3

0.

4 N

/A

6.0

300

4.5

T-1

00

Fig

ure

A 4

F

igur

e A

3.

35.3

0.

4 N

/A

2.8

300

3.4

T-1

00

Fig

ure

A 6

. Fi

gure

A. 5

. 44

.9

0.4

N/A

5.

0 30

0 4.

5 T

-100

Fig

ure

A 8

. F

igur

e A

. 7.

5.1

N/A

N

/A

N/A

40

0 4.

5 S

PT

-100

Fig

ure

A 1

0.

Fig

ure

A.9

. 4.

9 N

/A

N/A

N

/A

300

4.0

SP

T-1

00

Fig

ure

A.1

2.

Fig

ure

All

. 4.

0 0.

6 1.

0 4.

0 30

0 3.

5 D

-55

Fig

ure

A 1

4.

Fig

ure

A. 1

3.

3.4

0.6

1.0

5.0

300

3.0

D-5

5

68

100 T-J ' I

I 80 £ Z

60

IM . I 1' i!

i; ,-ll! >J •

I I I , , ' ( . 1 1

. , 1 , , ^ li

' ' I M It

„ ! , • • • ! ' • M I I I

r i'l;'. ' I f

1 I

1 ; / ' i^ l\iiV^ • i l l ' > •, 1 ' (

H'P I ' M l t

0.1 0.2 0.3 07 0.8 0.9 0.4 0 5 0 6 time (ms)

Figure A,l. Percent number density of a T-100 ( file; lang205 ),

100

c t 3 o

50

I' I;

1 I I I

' ' I I , , ' i

V I , '

V V u W r ,

1, il

I ]

'! r il

" ' M I , 1 ; i II

'vj j l i i U ' ' ' * ' • '

I

I ' M

1

0,1 0,2 0,3 0,7 08 0,4 0,5 0.6 time (ms)

Figure A,2, Percent peak current of a T-100 ( file: curr205 ),

0.9

100

80

c

& 6 0 - 1 ^ E 3

z 40

20

^V ' ^'^' '''^'^i ' •^' ' ' • '^V, . ; J:' •••••v^^ ''' •'•V / ' ^

1

I

I , ' I

0,2 0,4 0,6 1 4 1,6

H

0,8 1 12

time (ms)

Figure A,3. Percent number density of a T-100 ( file lang202 )

1.8

69

100

3 U ^ 50

m

^

r

y ^

\

±

• i . „ '• i I

i : 'v ' I

0.2 0.4 0.6 1.4 1.6 0.8 1 1.2

time (ms)

Figure A.4. Percent peak current of a T-100 ( file: curr202 ).

1.8

time (ms)

Figure A.5. Percent number density of a T-100 ( file: lang204 ).

100

c u t 3

^ 50

h i '

Ji i; il ' I' ii hi h i i f I

< " J 1 I V

' ' I V

l l i } \ ^ ' ^ ' . i ; V ; i ^ '

± 0.2 0.4 06 0.8 1 1.2

time (ms)

1.4 1.6 1.8

Figure A.6. Percent peak current of a T-100 ( file: curr204 ).

70

100

Figure A. 7. Percent number density of an SPT-100 ( file: lang526 ).

100

3 o 50 -

I . 1

\ I

I

• 1 i i •I i i

i

, \.

l i

I ' ! I 1 j ! 1 li i 1 !i

• h4\\ !l : ' i I M

I I

0.1 0.2 0.3 0.7 0.8 04 0.5 0.6 time (ms)

Figure A.8. Percent peak current of an SPT-100 ( file: curr526 )

0.9

100

I 95

E 3 ^ 90 e

85

- ; y

\ i l I ! 1 !

\ : ,

- • y ^ r

i I

\ l! ., II i I

. , I. I ' ^ i 1 i

/ , i M h

In l i l t

! . ;

0.1 0.2 0.3 0.4 0.5 0.6

time (ms)

0.7

i\ ! I l

I ! '

iM V , I i

i l

0.8 0.9

Figure A.9. Percent number density of an SPT-100 ( file: lang522 )

71

100

M /•

I I M i

I

| i ' I ' '

' i i l > I I

1 TT

I ;

I 1

a' y ^ n y y y i y i v , y i. 0.1 0.2 0.3 0.4 0.5 0.6

time (ms)

Figure A.IO. Percent peak current of an SPT-100 ( file: curr522 ).

100

99 — 1 !

c u

y 98t-

1 3 Z

97

96

T 1 rJ-\:: I

l i '

J

" V ' J

if J ' • ' i i ' • I . I \ i .

» . ' I , . ' ^ .

0.1 0.2 0,3

time (ms)

0,4 05

Figure A l l . Percent number density of a D-55 ( file: lang404 ).

0.6

100

c 90

3 O

»° 80

70

•\l\ \ <

" ' ' ! I

M^liiii;;; i i

I .

i i i l

i;i \''['\:\-'' 11 •

1 i u W , ' ' •! ;; I I I I |, ' I - r H ' ' ,' M i ' • P

' 1 I i II ,:

; i I t

r I

i. * : )

0.1 0.2 0.3

time (ms)

0.4 0.5 0.6

Figure A.12. Percent peak current of a D-55 ( file: curr404 ).

72

100

c

2 98 E 3 Z

96

\. r f

f •' 1 ' .'J

'f • i l l I J • / I ',1 k 'ul L, . L , 'J . 1, 1. L^ _

t 1 I ^ihi 11 , y I

. I .

H I .

' T J

• . , - • - * - ,

0.1 0.2 0.3

time (ms)

0.4 05 0.6

Figure A.13. Percent number density of a D-55 ( file: lang406 ).

100

t u

^ 80

60

I ^ » 1.

.1 I , I

'\ :

1 1

^

I > ! 1

/ I

I ! I ' t

0.1 0.2 0,3

time (ms)

04 0.5

Figure A. 14. Percent peak current of a D-55 ( file: curr406 ).

0 6

73

APPENTDIX B

MATHCAD WORKSHEETS

74

Mathcad Worksheet for Parallel Propagation Path (relative to the exit plane)

HQ 65 10 Centerline number density at 1 m rrig 9 1093897 10'^' mass of an electron

^m ^^^'^ Frequency of discharge oscillations c 300 10 Speed of light

f 62 10 Frequency ofthe RF signal e^ 8 854187817 10"'^ Free space permittivity

Vp = 15 10 Exhaust ion velocity q - 160217733 10"'^ Charge of an electro

m - .0001 Number density modulation factor n .65 ?.n - 35 Model coefficients

A p — Number density oscillation wavelength k — ^m '• P

2 2 X C 2. T r^x.z) X -c z 6(x,z) alan - ; o = f 2 - /. - k — -

z, f A

HQ g-'((l-cos(e(x,z)))>ji>" n(i ,x ,z) ' l mcos ' f j^2 71 I - k r(x,z) ) Plume model

r(x,z)^ cos(e(x ,z) )^- ,01

cop(t,x,z) —— Calculates the local plasma frequency m e ^ o

1 ' (op(t,x,z)"' Calculates the time dependent radian ' P^ ' ) = ' ^^ , 2 ^"^ phase shift of an RF signal,

. 1

ph(t.2) - ^^^"^ Converts the radian phase shift to degrees deg

z = 1 Distance dov^n stream from the thruster exit plane

ph(0,z) =23,56 Static phase shift of an RF signal propagating parallel to exit plane at a distance z dov\/n stream from the thruster's exit.

75

Mathcad W^orksheet for a Diagonal Propagation Path (relative to the exit plane)

HQ -.65 10 Centerline number density at 1 m m 9 1093897 10'^' mass of electron

f j ^ 30 10^ Frequency of discharge oscillations c 300 10^ Speed of light

f 6.2 10 Frequency ofthe RF signal e^ 8 854187817 10''^ Free space permittivity

V : 15 10 Exhaust ion velocity q 1,60217733 10"'^ Charge of an electro

m - ,0001 Number density modulation factor n - .65 Xn = 35 Model coefficients

^'p 2 n >. _ —^ Number density oscillation v\/avelength k = —^

m - p xO = 45 Starting x value 0t = 45 Propagation angle

x(z) r(z) ^ .,,x(z) - z

.2 2 6(z) alan © f 2 TI \ z

n ^ ^-,((l-cos(e(z)))/ji) . n(t,z) = fl - mcos ' f j^2 T: t k p r ( z )

r(z)^ cos(e(z) )^- .01

x(z)

c

f

- xO- lan(0tdeg)z

Plume model

I 2

(i)p{i,z) ' Calculates the local plasma frequency

"^e^o

l l

2,2 (op(t,z) 1 I Calculates the time dependent radian

dp(0 11 1 - M 2—j j ' ^ phase shift of an RF signal. [ \ « / j

.01

dp(i) ph(t) Converts the radian phase shift to degrees

deg

ph(0) = 52.951 Static phase shift of an RF signal propagating along a diagonal path through the plume of a Hall effect thruster.

76

APPENT)IX C

PHASE SHIFT PLOTS

77

u ><

'•B c a> a. a. <

W5

> «3

C

t 3 O

•o c

IE 1/3

C/5 C«

a.

C/3

0)

o

B CC Cd a. 00

c CO

u. o

cd

y f O —

3 > • —

<c ;E O

so S

Ig

ur

C3 U . J= a.

y 4—>

O CS ^

3 ^ E u.

y D •>->.

Cat

hodi

lo

w R

a m

g/s

u. _

« M M ^ ^ "^^^

tcm

a ag

nc

rent

£ 2 5 u

^ ^ • ^ ^ ^ H

tcm

a ag

nct

rent

(

Ui ^ p u

y GO y b to C3 f? r-^

Dis

ch

Vol

t;

y 00 t i u C c3 y ^-v

Dis

c C

ui

ko-B O. . i ; t i

CL 00

y •J^ y

Thm

s Ty

p

<N

u y

3 OO

u.

^^ u

y I -

3 00

Urn

m m

( N

N/A

J^

"^

o o m

»r) r-i

" y ^^ cd b. cd

CL

100

1

T t

u £i 3 00

u.

. r ,

P b. 3 00

u.

</", r*-,

">*

^^

N/A

( N ro

o o r",

>n r-,

5 W V «

s cd

O .

100

1

f -

\C r

^

§) U .

w~i u y u. 3 OO

u.

vC >/",

<

z

N/A

oc r",

O c m

p i/- ,

OO o T3

•n T t

O o

T-I

0 . CO

oc U

y

S) ^

r

y k. 3 OO

u.

\C LT,

<

z

N/A

oc r".

O c r

C

>ri

OO y

T3

«n T f

C o

T-1

a.

^

O y

5j 'LZ

a-'

y b> 3 00

£

lA ,

>/",

<

Z N

/A

W-)

r*",

O C m

ir-,

00 y

• o

»n •^

.—, ^ I-l

0 . cr>

rsi

U y

5b iZ

^^ ^ M

U u £. 3 OO

ul

<N >/",

^C

— •

r-,

r-i

^ o r ,

O ir^i

00 y

•o I T )

"^

I T ) 1

" '

O y

M r ,

r* . ^mt

1) u. 3 op

^

r " '

V.C

I /",

__ —"

o o f*",

r-•^

00 y

t 3

»n r t

1

\C'

y

op f ,

*/", .«

y u. 3 op

tz

I /",

CN

rsi

^^

t r , r-

o

^ o r«-,

^« — ( N

00 y

•o m "^

I / -1

oc 1

w y

3 OO

u.

r-^ M

y u. 3 00

U-

— i

i r ,

rs)

^

(Nl

m

r~,

o c m

m o •^

00 y

•o »n T f

78

6':-

4.457 5 34?

time (s)

Figure C.l . Start-up phase shift (along a parallel path at 0.5 m) for a T-100 (file phas 116)

10

,#J(fti irtv.*i» 'iW *i(iH^ *-V^ ^ ^

1.8 2.686 5.571 4 457 5 34?

time (s)

6.229 7.114

Figure C 2 Current waveform of the T-100 for Figure C.l

time ;ms)

Figure C.3. Phase shift of a 6 GHz passing through the plume of a T-100 ( parallel path 0.5 m from the exit, file: phas 10")

79

10

) 0.357 0.714 1.071 1.429 1.786 2.14?

time (ms)

Figure C.4. Current waveform ofthe T-100 for Figure C.3.

2.5

100

80

60

f T vili',^'l''<''^»^'^" " ^ r«^w»n>'^ii*vvjji,<''it^*uw«>'^'' t. wi,w^Wiui' Vi)U'«^»«w>.'n'*»'rtVi

CO •J

40

20

>'

O'' 1.143 2.286 3.429 4.571

time (s)

5.714 6.857

Figure C.5 Start-up phase shift (along a 45 deg diagonal path) for an SPT-100 (file phas614).

2.286 5.714

:t[;

6857 3.429 4571

time (s)

Figure C.6. Current waveform ofthe SPT-100 for Figure C.5

80

10

] i

CC

•S 0

-10

' ) ^

( , (

y '• ^ . /

0.114 0.229 0.343 0 457

time (ms' 0.571 0.686 08

Figure C.7. Phase shift of a 6 GHz passing through the plume of an SPT-100 along a 45 deg diagonal path (file: phas613).

10'

1 I

^ ; \

I I

\ i

I \

I'.

0.114 0.229 0.343 0 45''

time (ms)

0.5-1 0 686 0.8

Figure C.8. Current w aveform ofthe SPT-100 for Figure C.

A. -^ 0

( ( L

; •

-1 T-n^ A ' i ' » ^ 1 ^ ^ ' • YT • I- I. • ! I • •

\ ..

-A' 0.114 0.229 0 343 0 45'

time ! ms 1

0.571 0 686 0.8

Figure C.9. Phase shift of a 6 GHz passing through the plume of an SPT-100 along a 45 deg diagonal path (file phas612)

81

10

v -v^ K f

- i —

0.114

• ' ' . % • •.»>'. J . - ^-^f,.- :;=a: - N / %.

0.229 0.343 0.457 0.571 0.686 time (ms)

Figure CIO. Current waveform ofthe SPT-100 for Figure C.9.

08

100

80

60

40

20

i r-^

•- "*'. , ' -,l."«'^'V_»^_^ "._/-.

1 i

i

• • V -

1

i 1

0.714 1 429 143 857 5-1 4.286

time (s)

Figure C. 11 Start-up phase shift (along a 45 deg diagonal path) for a D-55 (file phas714).

10

' - • ' • -

^*]

!

1

1

S,'**' >.'vfv'Y* ''-..v*^\'.|*">>'.'Vii

1

j

i^'*<ry.>H'^'''.'^'.*'V< k',v.'^'>'>'Vf« l|^,^*•-VV•.'„*,>^lTl

0.714 3.571 4.286 1.429 2.143 2.857 time (s)

Figure C.12. Current waveform ofthe D-55 for Figure C. 11

82

10

00

^ 0

-10

: ; 1

^ — * l

• ^ V '"

1

1 i

' ' 1

'' ' ' J / }

, , , . , . - .

• •

'

0.114 0.229 0 343 0.457 0.571 0 686

time (ms)

Figure C.13. Phase shift of a 6 GHz passing through the plume of a D-55 along a 45 deg diagonal path (file: phas707).

0.8

10

r .1 f' '-m.

1 i f '

1 ! ' ' ''

r

1

i' II 1' • . ii (1 ;

. - i i i l l ! ' , !

' [ . r

(•

. { • .• • V

J. . .-

v.'.V.-.^^

ii

r

0.114 0.229 0.343 0.457

lime (ms)

0.571 0 686 0.8

Figure C.14. Current waveform ofthe D-55 for Figure C.13.

0.5

w\ ir.nr:ii r. \#R^ 4 ji>j,,ijrnF.irT::r. Ti\ f-fm. f WTf 1 ) •

so ^ -0 .5

\ (I I

-1.5 0.114 0.571 0.686 0.229 0.343 0 457

time (ms)

Figure C.15. Phase shift of a 6 GHz passing through the plume of a D-55 along a 45 deg diagonal path (file phas710).

0.8

83

K^Tl'-^-A-*-,

^^^^—^——

(

'^-, ' W ' Y ^ ^ V ^ y •^^-.-^..^^^,'^._^-^^^.^^., ' '^-.vw^>.^..^*Y^,^/..^

0 0.114 0.229 0.343 0.457 0.571 time (ms)

Figure C.16. Current waveform ofthe D-55 for Figure C.15

0.686 08

f l iL

# ^ 4 L 'a. [I jr T ^

-ft; m T:

-1

-2

M

i¥Wti

;i I

L'f r'J b'

0.114 0.229 0.343 0.457 0.571 time (ms)

0.686 0.8

Figure C.l 7. Phase shift of a 6 GHz passing through the plume of a D-55 along a 45 deg diagonal path (file: phas713).

10

1 / 1

• ^ ' . " . ' • " i l l . "

i

0 0.114 0.229 0.343 0.457 0.571 0.686 0.8

time (ms)

Figure C.18. Current waveform of the D-55 for Figure C. 17

84

APPENDIX D

POW^R SPECTR.\L

DENSITY PLOTS

85

T3 C

a. <

CA

a > ed

• 4 — >

c t 3 U

C ed

Cu

c "O

"cd u . •<—<

a. C/3 k-

o

c/3

a; 1/3

2 -4-1

O

U.

• 4 - 1

E cd k> ed a. 00

c '•C ed k. a> o. O

Q jj

ed

H

Cur

rent

w

avef

orm

Fi

gure

Pha

se S

hift

F

igur

e A

node

Fl

ow R

ate

mg/

s

Cat

hode

Fl

ow R

ate

mg/

s

Intc

mal

M

agne

t C

urre

nt (

I)

Ext

cmal

M

agne

t C

urre

nt (

I)

Dis

char

ge

Vol

tage

(V

)

Dis

char

ge

Cur

rent

(A

)

Pro

pa­

gati

on

Path

Thr

uste

r T

ype

Figu

re D

2

Fig

ure

D. 1

. 4.

3 N

/A

N/A

N

/A

300

3.5

Para

llel

S

PT

-100

Fig

ure

D.4

. F

igur

e D

.3.

4.3

N/A

N

/A

N/A

30

0 3.

5 Pa

rall

el

SP

T-1

00

Fig

ure

D 6

F

igur

e D

.5.

3.3

N/A

N

/A

N/A

40

0 3.

5 P

aral

lel

SP

T-1

00

Figu

re D

.X

Figu

re D

.7.

CN

N/A

N

/A

3.2

300

4.48

45

deg

S

PT

-100

Fig

ure

D. 1

0 Fi

gure

D.9

5.

1 N

/A

N/A

4.

0 30

0 4.

45

45 d

eg

SP

T-1

00

Fig

ure

D 1

2.

Figu

re D

. 11.

oc

0.6

4.0

300

4.5

Para

llel

D

-55

Fig

ure

D 1

4 Fi

gure

D.I

3.

4.8

o

1.0

5.0

300

4.5

Par

alle

l D

-55

Fig

ure

D 1

6 Fi

gure

D 1

5 5.

0 0.

2

( N

2.0

300

4.5

45 d

eg

D-5

5

Fig

ure

D.l

8.

Fig

ure

D.l

7.

0.2

2.0

2.0

300

4.5

45 d

eg

D-5

5

86

_ - 2 0 ? c •o

.. •5 -40

E <

-60

-80

\

A

/vj

/

/ \ '

/ V.

\

\ A / \

\

- \

r

A ^ V ^

3889.9 3889.933

— SPT-100 Running Baseline

3889.967 3890

Frequency (MHz)

3890.033 3890.067 3890 1

Figure D.l. Power spectral density of a 3.8 GHz signal with a parallel path 1 m down stream from the exit of an SPT-100 (200 kHz span, file: an021).

10

r 11 ^ '"'

M / 1 -' ' \,

' • • ' f '

;i 11 i 'M

.' V 'v

r

i! M ' ; ; ' i !^

,/ V

1

'' !\ :' 1 ; i

' '• i :

/ : ( 1 : ; J

Tj

M h -•y 'v'

A

'1 A '

•J

t

'1 'i fl '•:\i\i\

•• ^.' V

)! ii i: M ( W 1

V '- \

0.125 0.25 0.375 0.5 0.625 0.75

time (ms)

Figure D.2. Current waveform ofthe SPT-100 for Figure D.l.

0.875

-20^

• | -40

c E <

-60

-80

Xii ' ' til ll

.J 1

. / ./l-

/ L /1 rr ^^

< 1 > I l ' - '

6183.629 6183662

SPT-100 Running Baseline

\AS

6183.696 6183.729 6183.762 6183.796 6183.829

Frequency (MHz)

Figure D.3. Power spectral density of a 6 GHz signal with a parallel path 1 m down stream from the exit of an SPT-100 (200 kHz span, file an026).

87

. '• ':

(i l\ l\

ai' . V 'v/ ^

A u 'I'l

1

1 \/ \j ^

r

' '1 !1 m \y'

M l i l i

\l\l\i u ^

• ,

1 .' . .

M MM

'

1 ! ' 1

1 ' 1 r - .

r,

" r ( H i ' i i ,

-

1

' I l

11 / • ; V

0.125 0.25 0"5 0.375 0.5 0.625 time (ms)

Figure D.4. Current waveform ofthe SPT-100 for Figure D.3

0.875

-40

< -60

-80

' . • V ; . ' . - '•

' ' ' . ' • '

f - < . s ' • 1

/

.' ' '- \ /

' I

J

1

r !1 -~

•- t • • ' 1

r- . r-.'

-'-. . A ' -i 1-

1

6183.65 618?.683 6183.717 618375 6183.783

Frequency (MHz)

— SPT-100 Running Baseline

6183.817 6183 85

Figure D.5. Power spectral density of a 6 GHz signal with a 45 deg diagonal path through the plume of an SPT-100 (200 kHz span, file: an022).

10

8

6

4 ,1 ^ A ^ • V

:A ; '''

A

1 t

' - •

il . 1 \

i ' ^

' 1 1

\

-

' \

i ( ••

'' 1 1

-

,/",

II n

1

M

1 1 t

' 1 vy

•-A

f 1

1 \ -

0.125 0.25 0.75 0.375 0.5 0625 time (ms)

Figure D.6. Current waveform ofthe SPT-100 for Figure D.5

0.875

88

-.-20 E

1-40 c E < -60

-80

— — - — - ^

i

i i 1

./A^A-.A-'A' '.'.'*' . . ' 1 . , '

: i j

i

1 • l \ • • V . '-,,.

\4< .A"^'--'^'--v- - — ' . _ — _ _ _

6184.625 6184.75 6184.875 6185 6185.125

Frequency (MHz)

~ SPT-100 Running Baseline

6185.25 6185.375

Figure D.7. Power spectral density of a 6 GHz signal with a 45 deg diagonal path through the plume of an SPT-100 (750 kHz span, file: an629).

10

8

6

4 ^— Y ^

'

• - ^ A •• A S-... ^J •--vV . ; ' . '^-V,- /V,

0.1 0.2 0.6 0.7 0.3 0,4 0,5

time (ms)

Figure D.8. Current waveform ofthe SPT-100 for Figure D 7

0.8

-20

-40

c E

-60

-80

1 3; i:

•J

f

1; 1

i • ' 1

' i ii M i l

1

r 1

> i. ii

1,

6184.5 6184.667

SPT-100 Running Baseline

6184.833 6185 6185.167 6185.333 6185.5

Frequency (MHz)

Figure D.9. Power spectral density of a 6 GHz signal phase modulated with a 200 kHz tone propagating through the plume of an SPT-100 along a

45 deg diagonal path (750 kFIz span, file: an632).

89

" I W > \ir\

! I

I LJT

M M i ; M

/ '0 l-

' I I i

l\i\i

I '

i \ )

) M ^

t ' I

dl

0.1

I . f 1 /

' ^ I

MLU

0.2 0.6 0.3 0,4 0.5

time (ms)

Figure D.IO. Current waveform ofthe SPT-100 for Fieure D 9

07 0 8

-20

•5 -40

-60

-80

•''•' r ' 1 ' .

r'

/

r '.• • . , ' . ' "

"'

1 ' - '

V

' ' ' . " • 1 • ' , • . '

"'-, ' ' • ' A . ' - ' ; r

1

6184.966 6184.999 6185.033 6185.066 6185,099 Frequency (.MHz)

D-55 Running Baseline

6185,133 6185 166

Figure D 11. Power spectral density of a 6 GHz signal with a parallel path 1 m down stream from the exit of a D-55 (200 kHz span, file: anOl lb).

10

8

6

4 \ J^y/^\r^^\' '- -w'"' '^f^>^^J^y^s-

, ,"-/^-- v^- . ' "'*^ — ' ^ , ^ ~ . - . j - ^ • , _ , ' _ . ^ . w ^ ^

,

0.125 0.25 0.625 0.75 0.375 0.5

time (ms)

Figure D. 12. Current waveform ofthe D-55 for Figure D 11

0 875

90

^ - 2 0 E

u 1 - 4 0

c E <

-60

-80 •AA.'V-'-/ '- ' ''V

/

I r'

\ ,

3890.187 3890.354 3890,52 3890,687 3890 854

Frequency (MHz)

D-55 Running Baseline

3891.02 3891 187

Figure D.I3. Power spectral density of a 3.8 GHz signal with a parallel path 1 m down stream from the exit of a D-55 (1 MHz span, file an017b).

10

8

6

4 ' / . ^./^,,^-v^. ^^' l ^ / V v ^ - > ^ "• ; r. ,_ - fv."^",, v . - ' ^ ^ "»'.'' '^ ' - . ^ ^ V * ^ " •'" ---V. _ - .-.—.-/. _, . . - . • ^ ^ > ' v . - - ~ \ - i . A / - J - ^ ' . ' - . ' ^

0.125 0.25 0.75 0.375 0.5 0.625

time (ms)

Figure D.14. Current waveform ofthe D-55 for Figure D.13.

0.875

u T> -S

E < -60

-80

. . ' . ^ ^ \ _ ^ > _ ^ - ^ ' ^

1 ^'^i-^•\/'^'':••

1

1

I , . . . , . ' 1 '

1 1

1

;

. ..., . ' \

i i 6184.625 6184.75

D-55 Running Baseline

6184.875 6185 6185,125 6185 25 6185,375

Frequency (MHz)

Figure D.l 5. Power spectral density of a 6 GHz signal with a 45 deg diagonal path through the plume of a D-55 (750 kHz span, file: an636)

91

10

t •]

; A : ; M . ^ M J ' M " • ' •^' 1 . ^ r

/.^A^^A. A M. . , , - . • M ' .. 1

0.1 0.2 06 0.3 0 4 0.5

time (ms)

Figure D.16. Current waveform ofthe D-55 for Figure D. 15

0.7 08

6184.625 6184.75

D-55 Running

Baseline

6184.875 6185 6185.125 6185,25 6185,375

Frequency (MHz)

Figure D 17. Power spectral density of a 6 GHz signal phase modulated with a 50 kHz tone propagating through the plume of a D-55 along a

45 deg diagonal path (750 kHz span, file: an638k).

10

M.-'MV •:X^\.: I, n

0.1 0.2

• r ..J- ^-^K-

-f. ^ - . •! •V ^ - ^ - - . K . ,;:i^^

0.3 0.4 0.5

time (ms)

0,6 0.7 08

Figure D.18. Current waveform ofthe D-55 for Figure D.15,

92