Page 1: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Page 2: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Release History

Part Number Release Publication Date

09936781 A November 2006

Any comments about the documentation for this product should be addressed to: User Assistance PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, Connecticut 06484-4794 U.S.A. Or emailed to: [email protected] Notices The information contained in this document is subject to change without notice. Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no warranty of any kind with regard to this document, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in connection with furnishing, performance or use of this material. Copyright Information This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this publication may be reproduced in any form whatsoever or translated into any language without the prior, written permission of PerkinElmer, Inc. Copyright © 2006 PerkinElmer, Inc. Produced in the USA. Trademarks Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are protected by law. PerkinElmer is a registered trademark of PerkinElmer, Inc. Clarus is a trademark of PerkinElmer, Inc.

Page 3: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Table of Contents

Chapter 1 Introduction About This Manual...................................................................................... 1-3 Conventions Used in this Manual................................................................ 1-5 Electromagnetic Compatibility (EMC) ....................................................... 1-6 Warning Signs on the Instrument ................................................................ 1-8 Chapter 2 Safety Practices Chapter Overview........................................................................................ 2-3 Precautions .................................................................................................. 2-5 Environmental Conditions........................................................................... 2-6 Electrical Safety .......................................................................................... 2-8 Moving the Clarus GC .............................................................................. 2-11 Labels on the Clarus GC ........................................................................... 2-12 ECD Radioactive Hazards......................................................................... 2-15 Safe Handling of Gases ............................................................................. 2-19 Hazardous Chemicals ................................................................................ 2-23 Definitions in Warning for Hazardous Chemicals .................................... 2-24 Chapter 3 System Description Overview of the Clarus 600 GC .................................................................. 3-4 About the Touch Screen .............................................................................. 3-6 Chapter 4 Before You Install a Column Column Installation Information ................................................................. 4-4 Protecting Your Column.............................................................................. 4-6 Chapter 5 Installing a Packed Column Summary ..................................................................................................... 5-4 Materials and Tools Required ..................................................................... 5-5 Packed Column Injector Overview ............................................................. 5-6 Chapter 6 Installing a Capillary Column Summary ..................................................................................................... 6-4 Materials and Tools Required ..................................................................... 6-5 Step A: Turn the Heaters Off: ..................................................................... 6-7 Step B: Connect the Column to the Injector:............................................... 6-8 Step C: Set the Carrier Gas................................................................. 6-56 Step D Leak Test All New Connections:............................................... 6-69 Step E: Condition the Column and the Mechanical Joint Between the Pre-column and Column:................................................... 6-72 Step F: Connect the Column to the Detector: ........................................ 6-78 Step G: Leak Test All New Connections:........................................... 6-82 Step H: Set up the Split Mode for a CAP or PSS Injector:................. 6-83 PSS and POC Operating Hints: ................................................................. 6-84 Calculating a Capillary Column Split Ratio .............................................. 6-87 Chapter 7 PreVent Overview ..................................................................................................... 7-3

Page 4: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing PreVent on an Injector................................................................. 7-5 Installing PreVent on a Detector ............................................................... 7-25 Installing PreVent on a Clarus MS Detector ............................................. 7-43 PreVent Operating Techniques.................................................................. 7-68 Chapter 8 PPC Fundamentals Introduction ................................................................................................. 8-3 Carrier Gas Control ..................................................................................... 8-5 Detector Gas Flow Control........................................................................ 8-30 Auxiliary Pressure and Flow Control ........................................................ 8-34 PPC Tips and Techniques.......................................................................... 8-38 Chapter 9 Maintenance Autosampler Maintenance........................................................................... 9-4 Installing a Syringe...................................................................................... 9-8 Replacing the Vial-Locator Mechanism...................................................... 9-9 Syringe Maintenance ................................................................................. 9-11 Injector Maintenance ................................................................................. 9-12 Changing the Charcoal Trap or Replacing Charcoal on the Split/Splitless CAP and PSS Injectors................................................... 9-33 ECD Maintenance ..................................................................................... 9-36 FID Maintenance....................................................................................... 9-44 PID Maintenance....................................................................................... 9-51 NPD Maintenance ..................................................................................... 9-57 FPD Maintenance ...................................................................................... 9-72 PPC Maintenance ...................................................................................... 9-87 Practical Hints ........................................................................................... 9-91 Chapter 10 Troubleshooting Error Messages .......................................................................................... 10-4 GC Troubleshooting ................................................................................ 10-12 Appendix 1 Nuclear Regulatory Commission Regulations Appendix 11 Ionization Potential

Page 5: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Introduction 1

Page 6: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Introduction

1-2

Page 7: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

1-3

About This Manual

This hardware guide is divided into following chapters:

Chapter 1 Introduction

This chapter contains a brief introduction on the instrument, the conventions and warnings used in the manual.

Chapter 2 Safety Practices

Important safety information for the Clarus GC is provided in this chapter.

Chapter 3 System Description

This chapter contains information on the components of the instrument, how it works and instrument specifications.

Chapter 4 Before You Install a Column

This chapter contains general information regarding column installation and the basic procedures you should know in order to install a column.

Chapter 5 Installing a Packed Column

This chapter contains procedures for connecting a packed column to the packed column injector and setting the carrier gas flow using PPC and/or manual pneumatics.

Chapter 6 Installing a Capillary Column

This chapter contains procedures for connecting a capillary column to a Capillary Injector (CAP); a Programmed Split/Splitless Capillary Injector (PSS); and a Programmed On-Column Capillary Injector (POC). It also describes how to set the gas flows using PPC and/or manual pneumatics.

Page 8: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Introduction

1-4

Chapter 7 PreVent

PreVent™ is an enhanced capillary inlet system for the Clarus GC that uses columns with an inside diameter (i.d.) between 0.25 mm and 0.53 mm.

Chapter 8 PPC Fundamentals

This chapter describes the fundamentals of Programmable Pneumatic Control (PPC™) and how to use PPC to control your injectors.

Chapter 9 Maintenance

Maintenance and cleaning procedures for the various components of your instrument are provided.

Chapter 10 Troubleshooting

Performance checks, troubleshooting information, software and system error messages along with practical hints for running for the instrument are provided.

Appendix I: U.S. Nuclear Regulations

Regulations from the Nuclear Regulatory Commission.

Appendix II: Ionization Potential

Ionization potentials for representative compounds.

Page 9: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

1-5

Conventions Used in this Manual Normal text is used to provide information and instructions.

Bold text refers to button or tab page that is displayed on the touch screen.

All eight digit numbers are PerkinElmer part numbers unless stated otherwise.

Notes, cautions and warnings Three terms, in the following standard formats, are also used to highlight special circumstances and warnings.

NOTE: A note indicates additional, significant information that is provided with some procedures.

CAUTION

We use the term CAUTION to inform you about situations that could result in serious damage to the instrument or other equipment. Details about these circumstances are in a box like this one.

WARNING

We use the term WARNING to inform you about situations that could result in personal injury to yourself or other persons. Details about these circumstances are in a box like this one.

Customer Service

PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484-4794 U.S.A

Tel: 1 (800) 762-4000 or (203) 762-4000 Internet: http://www.perkinelmer.com

Page 10: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Introduction

1-6

Electromagnetic Compatibility (EMC)

United States (FCC) This product is classified as a digital device used exclusively as industrial, commercial, or medical test equipment. It is exempt from the technical standards specified in Part 15 of the FCC Rules and Regulations, based on Section 15.103 (c).

European Union All information concerning EMC standards will be in the Declaration of Conformity and these standards will change as the European Union adds new requirements.

European Union Industrial Environment The 220/230/240 V/50 Hz. Clarus GC has been manufactured for use in the European Union and is intended for the industrial environment. The instrument is to be connected to a mains power network supplied from a high or medium-voltage transformer dedicated for the supply of an installation feeding a manufacturing or similar plant.

Industrial environments are characterized by the existence of one or more of the following conditions:

-industrial, scientific and medical (ISM) apparatus are present

-heavy inductive or capacitive loads are frequently switched

-currents and associated magnetic fields are high

These are the major contributors to the industrial electromagnetic environment and as such distinguish the industrial from other environments. The instrument is not intended for connection to a public mains network supplying residential, commercial and light-industrial locations.

Susceptibility to RF Interference With the exception of the Flame Ionization Detector (FID), an RF field strength of 10 V/m between 80 MHz. and 1000 MHz. with 80% modulation at 1 kHz. may cause a deflection on the chromatographic detector baseline that exceeds its

Page 11: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

1-7

normal pattern. This implies that if a transmitting device, such as a walkie-talkie carried by a security guard, is use near the detector, a spike or peak on the chromatographic baseline may occur. If you are concerned that such an event may occur, PerkinElmer recommends that walkie-talkie restriction notices be posted in the vicinity. Cell phones, beepers, and other similar devices operate in a much higher frequency range and do not cause interference.

Page 12: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Introduction

1-8

Warning Signs on the Instrument

Alternating current.

Protective conductor terminal.

Off position of the main power source.

On position of the main power source.

Page 13: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

1-9

Warning, hot surface

Warning, risk of electric shock.

Warning, Risk of Danger. Consult the manual to determine the nature of the potential hazard and any actions which have to be taken. Also, consult the manual for allowable gases and maximum permitted pressures.

Page 14: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Introduction

1-10

Page 15: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices 2

Page 16: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-2

Page 17: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-3

Chapter Overview

This chapter describes the general safety practices and precautions that must be observed when operating the Clarus 600 GC.

This advice is intended to supplement, not supersede, the normal safety codes in the user's country. It is also a supplement to the PerkinElmer standard Safety and Health Policy. The information provided does not cover every safety procedure that should be practiced. Ultimately, maintenance of a safe laboratory environment is the responsibility of the analyst and the analyst's organization.

Please consult all manuals supplied with the Clarus GC and accessories before you start working with the instrument. Carefully read the safety information in this chapter and in the other manuals supplied. When setting up the instrument or performing analyses or maintenance procedures, strictly follow the instructions provided. The Clarus GC should be used in accordance with the instructions provided in this manual. If used otherwise, the protection provided by the instrument may be impaired.

Generic Warnings

Before installing or operating the Clarus GC, read the following information concerning hazards and potential hazards. You should ensure that anyone involved with installation and/or operation of the Clarus GC is knowledgeable in both general safety practices for the laboratory and safety practices for the Clarus GC. Get advice from your safety engineer, industrial hygienist, environmental engineer, or safety manager before you install or use this instrument.

Heated Zones

Heated zones should be treated with caution, for example, injector caps and detectors. Avoid physical contact with the injector caps. The detector cover may get hot, especially if flame ionization detectors are operated at high temperatures. As a general rule, allow heated zones to cool before attempting to work in the oven, injector, or detector areas.

Page 18: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-4

CAUTION

THERMAL RUNAWAY PROTECTION: The Clarus GC software shuts down the instrument if any heated zone exceeds 470 °C. Should this occur, the following error message is displayed:

INSTRUMENT SHUTDOWN xxx THERM RUNAWAY where xxx is the heated zone

Call your PerkinElmer Representative.

Instrument shutdown also occurs if there is a PRT (Platinum Resistance Thermometer) or MPU (Micro Processor Unit) failure. In these cases the following error message is displayed:

INSTRUMENT SHUTDOWN xxx PRT ERROR where xxx is the failed zone.

Call your PerkinElmer Representative.

Page 19: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-5

Precautions

WARNING

Be sure that all instrument operators read and understand the precautions listed below. It is advisable to post a copy of the precautions near or on the instrument shelf.

CAUTION

The protection provided by this equipment may be impaired if the equipment is used in a manner not specified by PerkinElmer.

The following precautions must be observed when using the Clarus GC: • This equipment requires no specified inspection or preventive maintenance

to ensure the continuous functioning of its safety features.

• Be sure that the power line voltage of the Clarus GC corresponds to the voltage used in your laboratory.

• Never remove the side panels of the Clarus GC without shutting down the instrument and disconnecting the instrument power cord from line power.

• Do not immerse the purge gas exit line in a liquid, as the liquid may be drawn back into the sample holder.

• Only high quality purge gases should be used with the Clarus GC. Minimum purity of 99.995% is recommended. A high quality filter-dryer accessory is recommended for the removal of any moisture from the purge gases.

Page 20: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-6

Environmental Conditions

Operating Conditions

CAUTION

The Clarus GC is designed for indoor use only.

CAUTION

Do not operate in a Cold Room or a refrigerated area. The Clarus GC operates most efficiently under the following conditions:

• Ambient temperature is 10 °C to 35 °C (50 °F to95 °F). The GC will operate safely between 5 °C and 40 °C (41 °F and 104 °F).

• Ambient relative humidity is 20% to 80% non-condensing.

• Operating altitude is in the range of 0 to 2 000 m.

WARNING

The Clarus GC is not designed for operation in an explosive environment.

Pollution Degree

The Clarus GC will operate safely in environments that contain nonconductive foreign matter up to Pollution Degree 2 in IEC 1010-1.

Cleaning the Instrument

Exterior surfaces may be cleaned with a soft cloth, dampened with a mild detergent and water solution. Do not use abrasive cleaners or solvents.

Page 21: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-7

Cleaning and Decontamination

Before using any cleaning or decontamination methods except those specified by PerkinElmer, users should check with PerkinElmer that the proposed method will not damage the equipment.”

Clarus GC Touch Screen

For optimum performance, the Clarus GC’s touch screen may require periodic re-calibration. The interval between re-calibration may be affected by exposure to combined heat and humidity conditions (ambient conditions between 30 °C / 50% RH and 35 °C / 80% RH).

Storage Conditions The Clarus GC may be stored under the following conditions:

• ambient temperature is -20 °C to +60 °C (-4 to 140 °F)

• ambient relative humidity is 20 to 80%, non-condensing

• altitude is in the range 0 to 12 000 m.

General Laboratory Safety Your laboratory should have all equipment ordinarily required for the safety of individuals working with chemicals (fire extinguishers, first-aid equipment, safety shower and eye-wash fountain, spill cleanup equipment, etc.).

Page 22: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-8

Electrical Safety

The Clarus GC contains high voltage. To prevent the risk of shock, unplug the line cord from the AC outlet and wait at least one minute before opening or removing any instrument panels.

The instrument has been designed to protect the operator from potential electrical hazards. This section describes some recommended electrical safety practices.

CAUTION This unit contains protective circuitry. Contact PerkinElmer

Service before performing any AC line tests.

WARNING

Safety Interlocks PerkinElmer provides a number of safety interlocks on its instruments that monitor such things as doors, gas pressures, temperatures of various components .Do not attempt to defeat these interlocks; you may compromise your own or someone else’s safety.

WARNING

Connect the GC to an AC line power outlet that has a protective ground connection. To ensure satisfactory and safe operation of the GC, it is essential that the protective ground conductor (the green/yellow lead) of the line power cord is connected to a true electrical ground. Any interruption of the protective ground conductor, inside or outside the GC, or disconnection of the protective ground terminal may impair the protection provided by the GC.

Page 23: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-9

WARNING

Do not operate the GC with any covers or parts removed.

WARNING

To avoid electrical shock, disconnect the power cord from the AC outlet before servicing. Servicing on the GC is to be performed only by a PerkinElmer service representative or similarly trained and authorized person.

WARNING

Do not attempt to make adjustments, replacements or repairs to this GC except as described in the user documentation.

CAUTION

To ensure adequate cooling of the instrument electronics, do not obstruct the gap at the base of the GC, and leave at least a 6-inch clearance between instruments.

Ensure that the power cord is correctly wired and that the ground leads of all electrical units (for example, recorders, integrators) are connected together via the circuit ground to earth. Use only three-prong outlets with common earth ground connections.

Servicing of incoming AC line components in your laboratory should be performed only by a licensed electrician.

Page 24: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-10

WARNING

Lethal voltages are present at certain areas within the instrument. Installation and internal maintenance of the instrument should only be performed by a PerkinElmer service engineer or similarly authorized and trained person. When the instrument is connected to line power, opening the instrument covers is likely to expose live parts. Even when the power switch is off, high voltages can still be present. Capacitors inside the instrument may still be charged even if the instrument has been disconnected from all voltage sources.

The instrument must be correctly connected to a suitable electrical supply. The supply must have a correctly installed protective conductor (earth ground) and must be installed or checked by a qualified electrician before connecting the instrument.

WARNING

Any interruption of the protective conductor (earth ground) inside or outside the instrument or disconnection of the protective conductor terminal is likely to make the instrument dangerous. Intentional interruption is prohibited.

When working with the instrument:

• Disconnect the instrument from all voltage sources before opening it for any adjustment, replacement, maintenance, or repair. If afterwards, the opened instrument must be operated for further adjustment, maintenance, or repair, this must only be done by a PerkinElmer Service engineer.

• Whenever it is possible that the instrument is no longer electrically safe for use, make the instrument inoperative and secure it against any unauthorized or unintentional operation. The electrical safety of the instrument is likely to be impaired if, for example, the instrument shows visible damage, has been subjected to prolonged storage under unfavorable conditions, or has been subjected to severe stress during transportation.

Page 25: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-11

Moving the Clarus GC

The Clarus GC weighs 69 kg (152 lb). Improper lifting can cause injury to the back. If the instrument must be moved, we recommend that at least two people carefully lift the instrument in order to move it.

Page 26: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-12

Labels on the Clarus GC

Shut off hydrogen and cap unusedflame detector fittings to preventaccumulation of hydrogen in ovenand possible explosion.

Warning

Warning Hot Surface

Do not depressthe oven safetyswitches.

WarningDo not touch theoven fan while itis moving.

Warning

I f the instrument top is o pened ,the alignment o f A utosamplertower must be verified . Refer to your Clarus 600 Manual.

Caution

Do not touch thehot meta l surfaceof the oven.

Warning

Disconnect supply cord and turn off inlet gases before opening top.Recheck integrity of rear gasconnections after closing.

Warning

DETECTOR CONTAINSRADIOACTIVE MATERIAL(SEE TAG ON DETECTOR)

CAUTION

ECD Certificate Label

Page 27: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-13

Caution

Chimney Vent Power Cord

D o not restrict air in take or exhaust.Caution

230 VAC ~ 50/60HzMAX POWER 2400 VA

EN 55011-Class A, Group 1

PerkinElmer

Nxxxxxxx

Xxxxxxx

precisely.Shelton, CT 06484 USA

Part No.

Serial No.

Clarus 600Gas Chromatograph

Do not install closer th an 6mmto a wall o f combustible material.

Caution

Warning Hot Surface

Caution

Grounding circuit continuity is vitalfor safe operation of equipment.Never operate equipment withgrounding conductor disconnected

Warning

This unit contains protective circuitry.Contact PerkinElmer qualified service personnel before performing any AC line tests.

Caution

Disconnect AC power cord from outletbefore removing any cover or parts .Do not operate the instrument with anycover or parts removed.

Warning

Consult the proper section in the manual for the allowable

gases and maximum permitted pressures

Page 28: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-14

Warning Hot Surface

Page 29: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-15

ECD Radioactive Hazards

WARNING

To assure that removable radioactive contamination on the external parts of the ECD remains at a safe level, the United States Nuclear Regulatory Commission requires that: • The ECD must be wipe tested at least once every six months.

• A record of the results must be maintained for NRC inspection.

United States Government Regulations for ECDs

NOTE: To repair an Electron Capture Detector cell requires a specific license issued by the U.S. Nuclear Regulatory Commission (NRC) and/or in some states by the equivalent state agency. For further information on obtaining a license, contact the Customer Service Department at PerkinElmer, Shelton, Connecticut, or the NRC Material Branch, Office of Nuclear Materials, Safety and Safeguards, Washington, DC 20555.

All USNRC regulations can be obtained through the internet at www.nrc.gov/reading-rm/

NOTE: These instructions are for ECD cell purchasers who are not specifically licensed to handle radioactive materials.

The Clarus GC Electron Capture Detector model (Part No. N610-0063) contains a maximum of 15 mCi of Nickel 63 (Ni 63), a radioactive material. Your possession and use of this detector is governed by 10 C.F.R. Section 31.5 which is reproduced in Appendix I. Under the provisions of that regulation you are deemed a "General Licensee."

Your possession and use of the detector cell may also be regulated by the state where you are located. The requirements of state regulatory agencies are substantially similar to those contained in NRC regulation 10 C.F.R. Section 31.5, but they may differ in some respects. It is suggested that you procure a copy of the regulations of your particular state. (Supplement 2 in Appendix I contains a list of the "Agreement States" which have been granted authority by the U.S.

Page 30: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-16

Nuclear Regulatory Commission to regulate the possession and use of radioactive material.)

It is required that you be familiar with regulation 10 C.F.R. Section 31.5 (Appendix 1 in the Hardware Guide 0993-6781). Following are summaries of its requirements.

Labels Do not remove any of the labels attached to the ECD cell or any of the labels attached to your Clarus Gas Chromatograph that refer to the ECD cell. Follow all instructions and abide by all precautions provided by the labels and in user instruction manuals referred to by the labels.

Leak Testing

You are obligated under U.S. federal and state regulations to make certain that the ECD cell is wipe-tested for leakage of radioactive materials at intervals of no longer than six months, and that the analysis of these wipe tests is conducted by a person specifically licensed to do so, either by the U.S. Nuclear Regulatory Commission or by an Agreement State. The analyses can be performed by the firm listed below:

National Leak Test Center P.O. Box 486 North Tonawanda, New York 14120

ECD Cell Failure or Damage

If a leak test detects more than 0.005 μCi (microcurie) of removable radioactive material on the surface of an ECD cell, or if the cell itself is damaged in such a way as to indicate that it may no longer adequately shield the radioactive material inside, you must immediately suspend operation of your chromatograph until the cell has been repaired or disposed of by a person specifically licensed to do so. Any such incident must be reported by you to the Regional Office, Inspection and Enforcement, U.S. Nuclear Regulatory Commission.

Page 31: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-17

Reporting Radiation Incidents, Theft or Loss

Please read Regulation 10 C.F.R. Section 20.2201 and 20.2202. These describe your duties should the radioactive material (Ni 63) in the ECD cell be lost, stolen, or released, or should any person be exposed to radiation.

Other ECD Requirements

Regulation 10 C.F.R. Section 31.5 (see Appendix I) does not permit you to abandon the ECD cell or export it. It may not be transferred except to a person specifically licensed to receive it. Within thirty days of such a transfer, you must report to the Director of Nuclear Material Safety and Safeguards, U.S. Nuclear Regulatory Commission, Washington, D.C. 20555, the name and address of the transferee. However, no report is needed to transfer a defective ECD cell to PerkinElmer in order to obtain a replacement.

You may transfer the ECD cell to another general licensee, like yourself, only when it remains at the same location to which it was shipped by PerkinElmer. Give the transferee a copy of these instructions and the regulations in Appendix I, and report to the commission as required in Regulation C.F.R. Section 31.5.

WARNING

NEVER DISMANTLE THE ECD CELL!! You can remove the ECD cell from the GC for repair.

United Kingdom Regulations

In the U.K., registration is required under the Radioactive Substances Act of 1960, for anyone keeping or using radioactive materials. Application should be made to any one of the following governing bodies:

ENGLAND Department of the Environment Queen Anne's Chambers

Page 32: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-18

Tothill Street London, SW1H 9J4

SCOTLAND Scottish Development Department 21 Hill Street Edinburgh, EH2 3J4

WALES Welsh Office Cathay's Park Cardiff, CF1 3NG

NORTHERN Ministry of Development IRELAND Parliament Building

Storemont Belfast, Northern Ireland

Page 33: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-19

Safe Handling of Gases

When using hydrogen, either as the combustion gas for a flame ionization detector or as a carrier gas, special care must be taken to avoid buildup of explosive hydrogen/air mixtures. Ensure that all hydrogen line couplings are leak-free and do not allow hydrogen to vent within the oven.

Ventilation

Adequate ventilation must be provided, particularly if a liquid nitrogen or carbon dioxide subambient accessory is in constant use. When analyzing hazardous compounds, such as pesticides, it may be necessary to arrange for venting of detector effluent into a fume hood.

Using Hydrogen

WARNING

Flame Ionization Detectors (FID) and Flame Photometric Detectors (FPD) use hydrogen as fuel. If the hydrogen is turned on without a column attached to the injector and detector fittings inside the oven, hydrogen could diffuse into the oven creating the possibility of an explosion. To avoid possible injury, DO NOT TURN ON THE HYDROGEN UNLESS A COLUMN IS ATTACHED AND ALL JOINTS HAVE BEEN LEAK TESTED.

Before disconnecting a column, make certain that the hydrogen has been turned OFF.

If two FIDs or FPDs are installed and only one has a column attached to it, make certain that you cap off the unused detector inlet fitting with a 1/8-inch stainless steel plug (Part No. N930-0061).

Page 34: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-20

WARNING

Contact the gas supplier for a material safety data sheet (MSDS) containing detailed information on the potential hazards associated with the gas. Carefully use, store, and handle compressed gases in cylinders. Gas cylinders can be hazardous if they are mishandled.

NOTE: The permanent installation of gas supplies is the responsibility of the user and should conform to local safety and building codes.

If liquid nitrogen is used, the gas cylinder must be fitted with an over-pressure regulator which will vent the cylinder as necessary to prevent it from becoming a safety hazard.

Consult the following references for more detailed information and additional guidelines about gas cyclinders.

• Compressed Gas Association (USA), "Safe Handling of Compressed Gases in Containers," pamphlet no. P-1, 1984.

• Compressed Gas Association (USA), "The Inert Gases – Argon, Nitrogen and Helium," pamphlet no. P-9, 1992.

Identification of Gas Cylinders

• Legibly mark cylinders to identify their contents. Use the chemical name or commercially accepted name for the gas.

Storing Gas Cylinders

Review the following precautions with the customer to ensure the safe use and storage of gas cylinders.

• Cylinders should be stored in accordance with the regulations and standards applicable to the customer’s locality, state, and country.

Page 35: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-21

• When cylinders are stored indoors in storage rooms, the storage room should be well ventilated and dry. Ensure that the ventilation is adequate to prevent the formation of dangerous accumulations of gas. This is particularly important in small or confined areas.

• Do not store cylinders near elevators, gangways, or in locations where heavy moving objects may strike or fall against them.

• Use and store cylinders away from exits and exit routes.

• Locate cylinders away from heat sources, including heat lamps. Compressed gas cylinders should not be subjected to temperatures above 52 °C (126 °F).

• It is recommended that gas cylinders be stored and placed outside the laboratory and connected to the instrument through copper lines.

Handling of Gas Cylinders

• Do not allow ignition sources in the storage area and keep cylinders away from readily ignitable substances such as gasoline or waste, or combustibles in bulk, including oil.

• Store cylinders standing upright, fastened securely to an immovable bulkhead or permanent wall.

• When storing cylinders outdoors, they should be stored above ground on a suitable floor and protected against temperature extremes (including the direct rays of the sun).

• Arrange gas hoses where they will not be damaged or stepped on and where things will not be dropped on them.

• Take care not to kink or stress the gas lines. For safety, cylinders should be firmly clamped in position.

Page 36: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-22

• If it becomes necessary to move the cylinders, do so with a suitable hand truck after insuring that the container cap is secured and the cylinder is properly fastened to the hand truck.

• Use only regulators, tubing and hose connectors approved by an appropriate regulatory agency.

• Do not refill cylinders.

• Check the condition of pipes, hoses and connectors regularly. Perform gas leak tests at all joints and seals of the gas system regularly, using an approved gas leak detection system.

• When the equipment is turned off, close all gas cylinder valves tightly at the cylinder. Bleed the remainder of the line before turning the exhaust vent off.

Page 37: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

2-23

Hazardous Chemicals

Before using samples, you should be thoroughly familiar with all hazards and safe handling practices. Observe the manufacturer’s recommendations for use, storage and disposal. These recommendations are normally provided in the Material Safety Data Sheets (MSDS) supplied with the solvents.

Be aware that the chemicals that you use in conjunction with the GC may be hazardous. DO NOT store, handle, or work with any chemicals or hazardous materials unless you have received appropriate safety training and have read and understood all related Material Safety Data Sheets (MSDS). MSDS’s provide information on physical characteristics, precautions, first aid, spill clean up and disposal procedures. Familiarize yourself with the information and precautions contained in these documents before attempting to store, use or dispose of the reagents. Comply with all federal, state, and local laws related to chemical storage, handling, and disposal.

You must work under a suitable hood when handling and mixing certain chemicals. The room in which you work must have proper ventilation and a waste collection system. Always wear appropriate safety attire (full-length laboratory coat, protective glasses, gloves, etc.), as indicated on Material Safety Data Sheets.

WARNING

Some chemicals used with this GC may be hazardous or may become hazardous after completion of an analysis. The responsible body (for example, the Lab Manager) must take the necessary precautions to ensure that the GC operators and the surrounding workplace are not exposed to hazardous levels of toxic substances (chemical or biological) as defined in the applicable Material Safety Data Sheets (MSDS) or OSHA, ACGIH, or COSHH documents. Venting for fumes and disposal of waste must be in accordance with all national, state and local health and safety regulations and laws.

Page 38: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Safety Practices

2-24

Definitions in Warning for Hazardous Chemicals

Responsible body. “Individual or group responsible for the use and maintenance of equipment, and for ensuring that operators are adequately trained.” [per IEC 61010-1].

Operator. “Person operating equipment for its intended purpose.” [per IEC 61010-1].

OSHA: Occupational Safety and Health Administration (United States)

ACGIH: American Conference of Governmental Industrial Hygienists

COSHH: Control of Substances Hazardous to Health (United Kingdom)

Page 39: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

System Description 3

Page 40: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

System Description

3-2

Page 41: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

3-3

The Clarus 600 Gas Chromatograph is a programmable pneumatic control (PPC™), dual-channel, temperature-programmable stand-alone gas chromatograph (GC). It is available in many configurations, such as with or without, an autosampler, and a variety of injector/detector combinations to provide you with total GC flexibility. The Clarus GC is microprocessor controlled. You use the touch screen to operate this instrument.

Figure 3-1. The Clarus 600 GC.

Page 42: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

System Description

3-4

Overview of the Clarus 600 GC

Your Clarus GC may have none, one, or two of the following detectors installed:

Flame Ionization (FID) Photoionization (PID)

Electron Capture (ECD) Nitrogen Phosphorus (NPD)

Thermal Conductivity (TCD) Flame Photometric (FPD)

The FID, ECD, TCD, NPD, or FPD may be installed in either the front or the rear detector position. But only one PID can be installed and it must be installed only in the rear position (Channel 2).

Each installed detector has one analog output which may be attached to either an integrator or recorder. Signals may be routed under instrument control.

Either none, one, or two packed column injectors; none, one, or two capillary column injectors; or one of each injector type may be installed. Capillary column injectors consist of the conventional split/splitless injector (CAP), a temperature-programmed split/splitless injector (PSS), and a temperature programmed on-column injector (POC).

Up to two gas sampling valves may be installed. In programmable pneumatic control (PPC) the carrier gas and detector gases are monitored and controlled by the microprocessor, thereby producing a fully automated system that is capable of managing all pneumatic functions within the gas chromatograph. The Clarus GC can store up to five GC methods. Methods can be generated, copied, deleted, edited, set up, and printed. The autosampler can run up to 15 injections per vial from as many as 82 vials and one priority vial using one or two autosampler programs. In the latter case, a different GC method can be used by each program if desired. The autosampler operates in one of two program modes, single or multi.

Page 43: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

3-5

Real-time digital readouts are provided to simplify setting carrier gas pressures and flows, if flow readout or PPC is installed. An additional digital readout can be displayed to show the detector output (in mV).

Page 44: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

System Description

3-6

About the Touch Screen

You control the Clarus GC by touch screens. The touch screens are your link to the software. is the screens are designed to provide the easiest access to the most frequently-used features. The touch screen allows quick access to settings and functions, such as the Oven program and Ignite button.

The System Status screen provides access to all heated zones. The icons represent configured components. These icons will guide you through the application.

The System Status screen provides you with one-click access to the oven settings. For detailed information on the touch screen see the Clarus 600 User Guide (P/N 0993- 6780).

Figure 3-3. The Clarus GC System Status screen.

Page 45: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Before You Install a Column 4

Page 46: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Before You Install a Column

4-2

Page 47: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

4-3

WARNING

At the moment you turn on the GC, the oven, injector(s), and detector(s) begin to heat up rapidly. To avoid injury while installing a column, all heaters should be turned off and their respective zones allowed to cool before touching the injector septum caps or any of the fittings inside the oven.

This chapter contains general column installation information, and the following procedures:

• Installing your column

• Protecting your column

Page 48: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Before You Install a Column

4-4

Column Installation Information

Injector and Detector Fittings

Columns are installed inside the oven. The injector fittings are on the left side and the detector fittings are on the right side of the oven ceiling. Figure 4-1 shows a capillary injector fitting in the front position and a packed injector fitting in the rear position.

WARNING

Before installing a column, make certain the oven is OFF (by opening the oven door), the oven fan has stopped, and the oven is cool.

Figure 4-1. Injector and detector fittings location in the oven.

Page 49: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

4-5

Capillary Column Hangers

Capillary columns are supported on column hangers. The left and right sides of the oven walls each have two rectangular slots into which column hangers are inserted. The two rear slots are used to install a column hanger in the rear position. The two front slots are used to install a column in the front position.

To install a column hanger, simply insert one end into the left slot and the other end into the right slot. If you are installing two capillary columns, install the rear hanger and the rear column before installing the front hanger and the front column.

Figure 4-2. A column hanger installed in the rear oven position.

Page 50: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Before You Install a Column

4-6

Protecting Your Column

The Clarus 600 GC provides a means for protecting your column(s) from overheating. It does this by not allowing the oven to heat up beyond the Oven Maximum Temperature Limit (OMTL), a value that you set in the active method.

You should set the OMTL equal to or less than the maximum permissible operating temperature recommended in the specifications for your column. If you are installing two columns, use the lower of the two permissible maximum operating temperatures.

To protect the column, the OMTL value (that you enter) works in conjunction with the Oven Temperature specified in the Active Method. Should you (or someone else) attempt to set an Oven Temperature in the Active Method to a value greater than the OMTL, the system displays an error message and will not allow you to continue until an appropriate new oven temperature is set.

To set the oven temperature:

1. Go to the System Status screen and touch the Oven icon:

2. Touch the field you need to update. It will appear black and then using the plus and minus buttons input the correct oven temperatures and time.

Page 51: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

4-7

For information on turning the heaters and detectors on and off see the Clarus 600 Users Guide (P/N 0993-6780) chapter on the Active Method.

Page 52: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Before You Install a Column

4-8

Page 53: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column 5

Page 54: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-2

Page 55: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-3

This chapter describes procedures for installing a packed column and setting the carrier gas flow using programmed pressure control (PPC). After installing a packed column as described in this chapter, you should set up the detector(s) to be used with this injector.

Page 56: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-4

Summary

The following steps summarize how to install a new packed column and get it ready for use:

• Step 1: Turn off the heaters.

• Step 2: Set the carrier gas flow.

• Step 3: Attach one end of the column to the packed injector.

• Step 4: Leak test.

• Step 5: Condition the column

• Step 6: Attach the other end of the column to the detector.

• Step 7: Leak test the column/detector connection.

NOTE: A new packed column should not be used until it is conditioned overnight with carrier gas flowing through the column. Do not connect the column to the detector during the conditioning process.

Page 57: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-5

Materials and Tools Required

• Two 7/16-inch open end wrenches when a normal 1/8-inch column is being installed. Two 9/16-inch open end wrenches will be required in addition if a 1/4-inch column is being installed.

• Two 1/8-inch to 1/4-inch adapters (P/N 0008-0100) if a 1/4-inch column is being installed.

• A flow meter or the optional flow readout.

• A packed column of your choice.

• A leak test solution or an electronic leak tester.

Page 58: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-6

Packed Column Injector Overview

The packed column injector consists of a septum cap, needle guide, quartz injector liner, and the injector body. This injector is used with 1/8-inch or 1/4-inch glass or metal packed columns. In addition, by installing the 530 Micron Wide-Bore Adapter Kit (P/N N612-0001) you can convert the injector to accept wide-bore capillary columns.

Figure 5-1. Packed Column Injector.

Page 59: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-7

About the Wide-Bore Adapter

If you are operating in the off-column mode at above optimum flow rates (>10 mL/min), you may not need to install the wide-bore quartz injector liner. Depending on your sample or solvent, the solvent profile (tail) may be acceptable for your application with the standard liner (the illustration at the left in Figure 5-2) and the addition of the adapter fitting (P/N N610-0083). However, if the solvent profile is not acceptable, install the wide-bore quartz injector liner.

The off-column or on-column flash vaporization mode of operation is determined by the position of the hourglass portion of the wide-bore quartz injector liner in the packed column injector. When installed correctly, this liner produces improved solvent profiles, especially at optimum flow rates. For complete installation instructions, refer to the Installation Instructions: 530 Micron Wide-Bore Adapter Kit for the AutoSystem GC and Clarus GC (P/N 0993-8661).

Insert the wide-bore quartz injector liner (P/N N612-1003) into the packed column injector with the hourglass portion in the correct position for your desired mode of operation. Figure 5-2 shows a cross section of a packed column injector containing a standard liner and a cross section of a packed column injector containing a wide-bore quartz injector liner installed in the off-column position and the on-column position.

For off-column flash vaporization (hourglass end first): To avoid contaminating the quartz wool, wear vinyl, powder-free disposable gloves (the same type used to perform maintenance on TurboMass). Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch rod (P/N N610-T100), push the quartz wool into the liner. Loosely pack some quartz wool in the top portion of the liner to wipe the syringe needle upon injection. Insert the wide-bore quartz injector liner into the packed column injector with the hourglass end first.

Or

For on-column flash vaporization (hourglass end last): Insert the wide-bore quartz injector liner into the packed column injector with the hourglass end of the liner last. Do not pack the wide-bore quartz injector liner with silanized quartz wool. You must use a 0.47-mm O.D. syringe in this mode.

Page 60: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-8

If you are using the autosampler, install a 0.47-mm O.D. syringe and use the “SLOW" injection mode. If the Clarus GC software is Rev. 1.6 or lower, use the “ON-COL" injection mode.

41.275 mm(1 5/8-inches)

AdapterFitting

AdapterFitting

AdapterFitting

Hourglass

Hourglass

SilanizedGlass Wool

SilanizedGlass Wool

Standard Liner

Wide-BoreGlass InjectorLiner

Wide-BoreGlassInjectorLiner

530 MicronWide-BoreColumn

530 MicronWide-BoreColumn

530 MicronWide-BoreColumnMark

Standard PackedColumn

Off-ColumnPosition

On-ColumnPosition

Figure 5-2. Cross sections of three packed injector configurations with a wide-bore column.

Page 61: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-9

Step 1: Turn off the Heaters

WARNING

The moment the GC is turned on, the oven, injector(s), and detector(s) begin to heat up rapidly. To avoid burns and injury while installing a column, all heaters should be turned off and their respective zones allowed to cool before touching the injector septum caps or any of the fittings inside the oven.

NOTE: See the Clarus 600 GC Users Guide for detailed procedures for turning heaters off and on.

NOTE: It is recommended that you remove the injector liner shipped with the packed injector and pack it with a small amount of silanized glass wool before performing analyses. Please refer to the Maintenance chapter in this manual.

Step 2: Set the Carrier Gas Flow

To set the carrier gas using PPC pneumatics modules:

1. Turn on the carrier gas at the tank.

2. Adjust the line pressure between 60 to 90 psig.

3. From the System Status screen select either A or B injector (the example below shows the B injector ready for setting the carrier gas).

Page 62: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-10

4. Touch the Pkd button and the following screen appears:

5. Touch the Program button in the Carrier Gas field.

Page 63: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-11

6. Touch a field to select it and using the up and down arrow buttons or the keypad button to enter the desired values.

Page 64: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-12

Setting the Carrier Gas Flow Using a Soap Bubble or Electronic Flowmeter The procedure below assumes that you know how to measure carrier gas flow using a soap bubble or electronic flowmeter and the built-in stopwatch. If you need instructions, please read “Using Tools,” in the Clarus 600 GC Users Guide (P/N 0993-6780) before proceeding.

1. Locate the packed injector fitting inside the oven.

Figure 5-4. A packed injector fitting.

2. Attach a soap bubble flowmeter to the packed injector fitting.

3. Turn on the carrier gas at the tank and adjust the line pressure to 90 psig.

4. From the System Status Screen touch the Tools button and then select Utilities from the drop down menu.

Page 65: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-13

5. In the Utilities window select the Stopwatch button.

6. Using the up and down arrow buttons or the keypad button enter the appropriate values and press Start.

Page 66: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-14

7. Start the carrier gas flowing by turning the flow controller knob counterclockwise and measure the flow.

NOTE: For best accuracy, use a soap bubble flowmeter volume that gives a reading of at least 30 seconds.

8. Adjust the flow to the desired set point by repeatedly measuring the flow and turning the flow controller knob counterclockwise to increase the flow, clockwise to decrease the flow, until the desired flow is obtained.

9. Disconnect the soap bubble flow meter before proceeding to the next step.

Step 3: Connect One End of the Column to the Packed Injector

NOTE: If you are installing a 1/4-inch column, attach a 1/8-inch to 1/4-inch adapter to the packed injector fitting before continuing. Finger tighten the adapter, then while holding the packed injector fitting steady with a 7/16-inch wrench, tighten the adapter with a 9/16-inch wrench.

1. Insert one end of the column into the packed injector fitting until it bottoms, then finger tighten the column nut onto the packed injector fitting (see below).

Figure 5-3. Packed column connected to a packed injector fitting.

2. While holding the packed injector fitting with one 7/16-inch wrench, tighten the column nut an additional 1/8 to 1/4 turn with the other wrench.

Page 67: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-15

NOTE: Do not overtighten column nuts. Overtightening causes permanent damage to the fittings.

Step 4: Leak Test Test the connection to the packed injector fitting for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To avoid contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections.

Step 5: Condition the Column This section contains a suggested temperature program for conditioning a column. The program starts off by holding the oven temperature at a medium value for 10 minutes, gradually increasing the oven temperature at a fixed rate (5 °C/min) to the column operating temperature, then holding that temperature overnight with the carrier gas flowing.

CAUTION

The temperatures shown in the following examples should only be used as guidelines. Please refer to the column manufacturer's operating instructions for specific temperature recommendations.

To condition the column:

1. Close the oven door, go to the System Status screen, and touch the oven icon.

Page 68: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-16

2. Select the Init Temp field. Then using the up and down arrow or keypad

buttons enter the initial (Init) oven temperature to a set point of 50.

3. Enter an Init Time of 10.

4. To add another program step, enter an Init Rate of 5 (°C/min).

Page 69: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-17

5. Using the up and down arrow or keypad buttons enter for

Temp 2, a set point 25 °C to 50 °C above your planned analytical operating temperature. For example, enter a set point of 150.

6. Set Time 2 to 999. This displays HOLD in the field.

CAUTION

To avoid damaging the column, do not enter a temperature higher than the maximum operating temperature specified by the column manufacturer.

7. Select the B injector by clicking the B-Pkd tab.

Page 70: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-18

8. In the above screen touch the temp field, then using the up and down arrow or keypad buttons set an Injector Temp about 50 °C higher than the Oven Temp 2 setting. See the chapter “Active Method” in the Clarus 600 GC User’s Guide for information on setting Temp 2.

9. Select the detector B-PID tab, then turn the Detector Temperature off and run the system overnight.

Page 71: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-19

10. The next morning go to the System Status screen and select the Oven button.

11. Reset the Oven temperature set point to that specified for Temp 1 at the beginning of the temperature program.

12. Open the oven door and allow the oven to cool until the oven fan goes off. This occurs when the oven cools down to 40 °C.

NOTE: Condition a new column before using it in an analysis. Once it is conditioned, you will not need to recondition it.

Step 6: Attach the Other End of the Column to the Detector

1. Insert the free end of the column into the detector fitting, then finger tighten the column nut onto the detector fitting.

Figure 5-5. Packed column attached to the rear detector fitting.

2. While holding the detector fitting with one of the 7/16-inch wrenches, tighten the column nut an additional 1/8 to 1/4 turn with the other wrench.

Page 72: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-20

CAUTION

Make certain that no part of the column touches the bottom or sides of the oven once it is installed.

NOTE: If you are installing a 1/4-inch column, attach a 1/8-inch to 1/4-inch adapter to the detector fitting before continuing. Finger-tighten the adapter, then while holding the detector fitting steady with a 7/16-inch wrench, tighten the adapter with a 9/16-inch wrench.

Step 7: Leak Test the Column/Detector Connection

The following procedures describe leak testing the column to detector connections.

1. From the System Status screen select either A or B injector (the example here shows the B packed injector ready for setting the carrier gas).

In this example, touch the Channel B packed injector button .

Page 73: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

5-21

2. In the following window touch the Program button in the Carrier Gas field.

3. Using the up and down arrow or keypad buttons enter pressure 0.1 to 100 psig for minimum and maximum.

Page 74: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Packed Column

5-22

The Clarus 600 GC monitors the column head pressure with a carrier-gas mass-flow controller in use. After selecting the carrier-gas Flow mode in configuration, you enter the minimum and maximum values. If the inlet pressure drops below or rises above the set limits for more than one consecutive minute, an error message will appear: Carrier (with specific carrier number) unable to maintain pressure.

NOTE: Refer to the Clarus 600 GC User’s Guide (0993-6780) for minimal and maximum values.

You should correct the leak; the most common area of a leak is the injector septum. Clear the message by touching the OK button before continuing to use the GC. This will stop an autosampler program from continuing.

If you do not make any entries, leak checking will not be done. If you are temperature programming, then the maximum value you enter must be above the column head pressure at the highest oven temperature. You can just enter a value for the minimum and leave the maximum blank. See the Clarus 600 User’s Guide (0995-6780) for more information.

Page 75: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column 6

Page 76: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-2

Page 77: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-3

This chapter describes how to install a capillary column in the following three injectors:

• Capillary Split/Splitless (CAP)

• Programmed Split/Splitless (PSS)

• Programmed On-Column (POC)

The information in this chapter is presented as one sequential procedure (Steps A through I) for all three injectors (CAP, PSS, and POC) with the following procedural steps common to all three injectors:

• Setting carrier gas flow using PPC modules

NOTE: For a thorough understanding of PPC, refer to Chapter 8, “PPC Fundamentals.”

• Leak testing

• Conditioning the column

• Attaching the column to the detector and leak checking

NOTE: If you are analyzing reactive compounds, appropriately deactivate injector liners and wool for your sample type.

CAUTION

All three capillary injectors; CAP, PSS, and POC use a 1/16-inch fitting for the column connection. This fitting is fragile. To preserve the integrity of the fitting, carefully connect the column nut to prevent cross-threading the fitting and/or overtightening the nut on the fitting. You can also preserve the integrity of the fitting by allowing the injector to cool before connecting a nut.

Page 78: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-4

Summary

The following steps summarize how to install a capillary column and get it ready for use:

A. Turn the heaters off.

B. Connect the column to the CAP, PSS, or POC injector.

1. Split/Splitless (CAP) injector 2. Programmed Split/Splitless (PSS) injector 3. Programmed On-Column (POC) injector

C. Set the carrier gas. (Set the pressure for the CAP and the PSS, or the flow for the POC, using the optional flow readout or a flowmeter.)

Setting carrier gas for PPC module

D. Leak test all new connections.

E. Condition the column (to the manufacturers specifications) and the mechanical joint between the column and pre-column.

F. Connect the column to the detector.

G. Leak test all new connections.

Page 79: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-5

Materials and Tools Required

• 1/8-inch x 1.0-mm graphite ferrules (P/N 0990-3394) for 0.53-mm i.d. columns

• 1/16-inch x 0.8-mm graphite ferrules (P/N 0992-0141)1 for 0.53-mm i.d. columns

• Two 7/16-inch open end wrenches

• Two 1/4-inch open end wrenches

• One 1/8-inch graphite ferrule (P/N 0990-3981)1 for 0.32/0.25-mm i.d. columns

• One 1/8-inch column nut (P/N 0990-3453)

• One 1/16-inch graphite ferrule (P/N 0990-3700)1 for 0.32/0.25-mm i.d. columns

• One 1/16-inch column nut (P/N 0990-3392)1

• One screwdriver (P/N 0990-7273)1

• Deactivated 0.53-mm i.d. fused silica (P/N N610-1724)

• Fused-silica universal connector (P/N N930-2149)

• Capillary column of your choice

• White-out or felt-tip marker

• Scribe for cutting columns (P/N N930-1376) (Pointed scribes are not recommended.)

• Leak-test solution or electronic leak tester

1 Shipped in the Clarus 600 GC Shipping Kit.

Page 80: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-6

Figure 6-1. Examples of required fittings.

Page 81: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-7

Step A: Turn the Heaters Off:

CAUTION

The moment the GC is turned on, the oven, injector(s), and detector(s) begin to heat up rapidly. To avoid injury while installing a column, all heaters should be turned off and their respective zones allowed too cool before touching the injector septum caps or any of the fittings inside the oven.

CAUTION

Before you install a column, follow the detailed procedures for turning heaters off and on, in Chapter 4 of this manual “Before You Install a Column.” If you have not read this chapter, please do so before proceeding

Page 82: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-8

Step B: Connect the Column to the Injector:

This step contains a separate procedure that describes how to connect a column to each of the following three Clarus 600 GC injectors:

Step B1: Connect the Column to the Split/Splitless (CAP) Injector.

Step B2: Connect the Column to the Programmed Split/Splitless (PSS) Injector.

Step B3: Connect the Column to the Programmed On-Column (POC) Injector.

Proceed to the page that contains the procedure for your injector.

Step B1: Connect the Column to the Split/Splitless (CAP) Injector

Overview

The Split/Splitless injector (CAP) consists of a septum purge assembly and the injector body. Carrier gas enters the injector body at the point just above the O-ring and flows through the quartz liner past the column tip.

Page 83: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-9

Figure 6-2. Cutaway view of the Split/Splitless injector (CAP).

About the Injector Liners

The CAP injector uses the following two quartz liners:

• Narrow-bore (2-mm i.d.) liner (P/N N612-1002).

• Wide-bore (4-mm i.d.) liner (P/N N612-1001).

Page 84: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-10

The narrow-bore liner is generally used for splitless injections and the wide-bore liner is generally used for split injections. Due to its small internal volume (0.3 mL), the amount of sample injected into the narrow-bore liner should be limited to about 0.5 μL. This prevents the solvent expansion upon injection from overfilling the liner with vapor.

To wipe the syringe needle, we recommend packing a small amount of quartz wool in the top portion of all liner types or injection modes (for example, split or splitless). Each liner should be packed with the quartz wool as described later in this chapter.

Splitless Injections

In the splitless injection mode, the narrow-bore quartz liner is typically used without quartz wool. The narrow-bore decreases the sample residence time in the liner, making it useful for trace analysis with smaller sample volumes (0.5 μL or less). By closing the split vent, most of the sample mixture enters the column. Then, opening the split vent clears the inlet of residual solvent.

For splitless injection volumes over 0.5 μL, the wide-bore liner with an internal volume of 1.25 mL should be used. However, the amount of sample should be limited to a maximum of 2 μL for hydrocarbon solvents and less than that for high-expansion solvents such as water or CH2Cl2. Refer to Table 6-1 for examples of gas volumes formed upon sample injection for selected solvents.

If the wide-bore liner is used for splitless injection, the splitless sampling time (vent-on time) should be at least one minute or more. Also, lower initial oven temperatures may be required to produce good solute resolution in the first few minutes after the solvent peak. The wide-bore liner should be used with columns having an i.d. of 0.32 mm or greater.

Page 85: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-11

Table 6-1. Gas Volumes Formed Upon Sample Injection (Injector 250 °C, Inlet Pressure 10 psig)

Solvent Volume Injected (µL) Gas Volume Generated (µL)

Methylene Chloride 1 401

2 802

Methanol 1 632

2 1264

Water 1 1420

2 2840

Split Injections

In the split injection mode, the wide-bore quartz liner is packed with quartz wool to ensure thorough mixing of the sample and carrier gas before they encounter the column tip. The split vent is open at the time of injection so that a fraction of the sample mixture enters the column while the remainder is routed out through the split vent.

The PPC pneumatics consists of PPC modules that regulate the inlet flow and pressure of the gases using software controlled by an Clarus GC method.

CAUTION

The CAP injector is shipped with the wide-bore liner installed without quartz wool packing. Before using the injector, remove the liner and pack it with quartz wool. If you are using the injector in the splitless mode, you may want to install the narrow-bore liner.

Connecting a Column to the Cap Injector

The following five steps describe how to connect a column to the CAP injector:

Step 1. Remove the CAP injector liner.

Step 2. Select an appropriate CAP injector liner.

Page 86: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-12

Step 3. Pack the CAP injector liner with quartz wool.

Step 4. Reinstall the liner in the CAP injector.

Step 5. Connect a column to the CAP injector.

Step 1. Remove the CAP Injector Liner.

To remove a CAP injector liner:

1. Ensure that the injector heater has been turned off.

Allow the injector to cool until it is slightly warm to the touch. Cooling the injector to too-low a temperature (less than 80 °C) will make it difficult to remove the injector liner.

2. Remove the septum cap.

Figure 6-3. Removing the septum cap.

3. Remove the injector cover.

Page 87: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-13

Figure 6-4. Removing the injector cover.

4. Loosen the threaded collar by using the spanner (P/N N610-1359) provided, then remove the threaded collar.

Figure 6-5. Loosening the threaded collar.

5. Replace the septum cap on the injector.

6. Pull the septum cap upwards to remove the septum purge assembly.

Page 88: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-14

Figure 6-6. Removing the septum purge assembly.

The carrier gas inlet line is coiled to allow you to pull the septum purge assembly over to the side and gain access to the liner.

7. Ensure that the CAP injector liner is cool, then twist the CAP injector liner-removal tool (P/N 0250-6534, see Figure 6-8) onto the injector liner. Remove the injector liner by lifting it up and out of the injector.

The CAP liner must be cool (no hotter than 100 °C) or the liner-removal tool will melt! The end of the CAP liner-removal tool may flare out with use. If this happens, cut off the flared end with a razor blade or scissors.

Figure 6-7. CAP injector liner-removal tool (P/N 0250-6534).

Page 89: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-15

Figure 6-8. Removing a capillary injector liner.

CAUTION

If the O-ring adheres to the injector body, use a small screwdriver to loosen the O-ring so that you can remove the liner and O-ring. Be careful not to scratch the barrel where the O-ring seals. Discard this O-ring and install a new O-ring.

NOTE: If the liner breaks inside the CAP injector, it can be removed by first removing the column. Then using a 9/16-inch wrench, remove the 1/4-inch injector fitting inside the oven. The liner should fall out. If the liner is stuck, you can push it out from the top or bottom.

Step 2. Select an Appropriate CAP Injector Liner.

Select the appropriate injector liner for your application. The following two injector liners are available for the CAP injector:

• 4-mm i.d. and 6-mm o.d. CAP injector wide-bore liner (P/N N612-1001)

• 2-mm i.d. and 6-mm o.d. CAP injector narrow-bore liner (P/N N612-1002)

The narrow-bore liner is generally used for a splitless injection, and the wide-bore liner is generally used for a split injection. Due to the small internal volume

Page 90: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-16

(0.3 mL) of the narrow-bore liner, you can prevent overfilling the liner with vapor (caused by solvent expansion upon injection) by limiting the amount of sample injected to 0.5 μL.

The wide-bore liner is used for splitless injection volumes over 0.5 μl since its internal volume is 1.25 mL. The sample size should be limited to a maximum of 2 μL for hydrocarbon solvents and less than that for high-expansion solvents, such as water or CH2Cl2. Refer to Table 6-1.

If the wide-bore liner is used for splitless injection, the splitless sampling time (vent-on time) should be more than one minute. Also, lower initial oven temperatures may be required to give good resolution in the first few minutes after the solvent peak. The wide-bore liner should be used with columns having an i.d. of 0.32 mm or greater.

Step 3. Pack the CAP Injector Liner with Quartz Wool.

To wipe the syringe needle, we recommend packing a small amount of quartz wool in the top portion of the liner regardless of the liner type or injector mode used (for example, split or splitless). This packing assures that reproducible volumes are injected because it wipes the syringe needle every time the needle is inserted.

Remove the liner and replace the quartz wool packing on a regular basis, particularly if your samples contain nonvolatile components that could build up on the wool. This could cause adsorption of peaks of interest, tailing, and loss of sensitivity. Remove the wool with a small hook on the end of a thin wire, or blow it out using compressed air.

NOTE: To avoid contaminating the quartz wool when packing the injection liner, wear vinyl, powder-free, disposable.

Packing CAP Injector Liner for Split Operation

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch rod (P/N N610-T100), push the

Page 91: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-17

quartz wool into the liner. Pack the wool tightly2 from the dimple upwards (about one inch [2.5 cm]). Loosely pack quartz wool in the top portion of the liner to wipe the syringe needle upon injection.

Packing a CAP Injector Liner for Splitless Operation

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch rod (P/N N610-T100), push the quartz wool into the liner. Pack a one-inch piece (2.5 cm) of quartz wool loosely below the top ground portion of the liner (see Figure 6-11). The sample is then injected into the wool, thereby preventing the delivery of sample beyond the column. The wool also wipes the syringe needle upon injection.

Figure 6-9. CAP injector liners packed with quartz wool.

Step 4. Reinstall the Liner in the CAP Injector.

To reinstall the liner:

1. Install a new O-ring near the ground portion of the liner.

2. Insert the liner in the injector body. 2 The recovery of high molecular weight components (e.g., C40) may be improved if the liner is loosely packed.

Page 92: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-18

3. Place the septum purge assembly over the liner.

4. Press the septum purge assembly down to correctly position the liner in the injector.

5. Replace the threaded collar and tighten the assembly using the spanner (P/N N610-1359).

Step 5. Connect a Column to the CAP Injector.

CAUTION

This injector terminates in a 1/16-inch fitting. This fitting is fragile. To preserve the integrity of the fitting, carefully connect the nut to prevent cross-threading the fitting and/or overtightening the nut on the fitting. You can also preserve the integrity of the fitting by allowing the injector to cool before connecting a nut.

To connect a column:

1. Insert a 1/16-inch column nut (P/N 0990-3392) and 1/16-inch graphite ferrule (0.8 mm i.d., P/N 0992-0141 or 0.5 mm i.d., P/N 0990-3700) over one end of the column as shown below:

Figure 6-10. Narrow-bore capillary column, nut, and ferrule on the injector end of a column.

NOTE: Verify that the tapered end of the ferrule is facing towards the nut as shown above.

2. Cut off about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N930-1376, pkg. of 10 scribes). Break off the tubing at the score mark

Page 93: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-19

so that the break is clean and square. Examine the cut with a magnifying glass and compare it to the examples shown in the following figure:

Figure 6-11. Example of a good column cut and bad cuts.

3. To automatically calculate the position of the column nut on the column so that the back of the nut is the correct distance from the end of the column, go to the system Status Screen and touch the Tools button.

4. Touch Utilities from the drop down menu. The Utilities screen appears.

Page 94: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-20

5. In the Utilities screen touch the Column Length Calc button .

6. Using the drop down menu touch the type of Detector and Injector you are

using. Select whether you want the calculation in mm or inches. The proper distance will be calculated automatically.

Page 95: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-21

7. Using correction fluid (white-out) or a felt-tipped pen, make a mark on the column just beyond the back edge of the column nut (see Figure 6-10).

CAUTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column.

8. Locate the capillary injector fitting inside the GC oven.

Page 96: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-22

CapilaryInjectorFitting

Figure 6-12. Capillary injector fitting inside the oven.

9. Insert the column into the capillary injector fitting. Then hand-tighten the column nut onto the capillary injector fitting. Insert the column into the capillary injector fitting until the mark is aligned with the back of the nut.

10. Using two 1/4-inch wrenches, tighten the column nut only until the column cannot be pulled out of the fitting.

CAUTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column.

Page 97: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-23

Figure 6-13. Capillary column attached to capillary injector fitting.

Step B2 Connect the Column to the Programmed Split/Splitless (PSS) Injector:

Overview

The Programmed Split/Splitless injector (PSS) consists of a septum purge assembly and the injector body. Carrier gas enters the injector at a point just above the O-ring and flows through the quartz liner past the column tip.

Page 98: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-24

Figure 6-14. Cutaway view of the Programmed Split/Splitless injector (PSS).

About the Injector Liners

The PSS injector uses the following three quartz liners:

• 2.0-mm i.d. (wide-bore) liner (P/N N612-1004)

• 1-mm i.d. (narrow-bore) liner (P/N N612-1006)

• hourglass liner (N610-1539)

In general, for split or splitless injections, use the 2-mm or 1-mm i.d. liner and operate the PSS in the inlet-programmed mode. For on-column operation, use the

Page 99: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-25

hourglass liner and the oven program mode. The 2-mm i.d. liner should be packed with quartz wool as described in this chapter and used for either split or splitless operation. The 1-mm i.d. liner may give better resolution of the early-eluting peaks in the split or splitless mode and it is better for labile compounds; however, it should be used for those samples with early-eluting peaks for which additional solute trapping focusing cannot be obtained by lowering the initial oven temperature or by using a column with a thicker stationary phase film.

To wipe the syringe needle, in all liner types (wide-bore or narrow-bore) or injection modes (split or splitless), we recommend packing a small amount of quartz wool in the top portion of the liner.

The sample is injected into the PSS injector at a "cool" temperature. The injector temperature is then programmed to increase. This is useful for samples that are thermally labile and/or have a wide molecular weight range. The PSS injector can also be used in a programmed on-column mode by replacing the quartz liner with the hourglass liner and closing the split vent flow.

CAUTION

The PSS can be used in the "hot" split or splitless mode. This, however, is not recommended for use with the 1 mm i.d. liner; it could cause solvent flashback in the injector. This mode should be used with caution depending upon the solvent and temperatures you choose. See Table 6-1.

CAUTION

When using the PSS in the on-column mode with the autosampler, you must use a special syringe that has a needle o.d. of 0.47 mm (P/N N610-1253 or N610-1380). Refer the Active Method Chapter in the Clarus 500 Users Guide, “Controlling the Autosampler,” for more detail. You must use only the "Norm" injection speed with this syringe in the on-column mode. The "Fast" injection speed will bend this thin needle; the "Slow" injection speed may produce double peaks due to the momentary stoppage of column flow during the longer injection. You can achieve better precision in the on-column mode when sample volumes of 1.0 µL or greater are injected.

Page 100: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-26

Split Injections

In the split mode, the wide-bore quartz liner is packed with quartz wool to insure thorough mixing of the sample and carrier gas before they reach the column tip. The split vent is open at the time of injection so that a fraction of the sample mixture enters the column while the remainder is routed out through the split vent.

Splitless Injections

In the splitless mode, the narrow-bore quartz liner is typically used without tightly packed quartz wool for mixing. Instead, a small amount of quartz wool in the center is recommended. This is useful for trace analysis with smaller sample volumes (less than 1 μL). By closing the split vent, most of the sample mixture enters the column, then the split vent is opened to clear the injector inlet of residual solvent.

Solvent Purge Injections

In applications where the sample is in a trace concentration and has a high molecular weight, this sample can be injected into the PSS with the vent open and the starting temperature near the boiling point of the solvent. In this way, only the solvent is purged out through the vent, then the vent is closed and the injector temperature is programmed up to elute the peaks of interest onto the column. This technique allows larger quantities of sample to enter the column without large solvent effects.

About the Pneumatics Control

The pneumatics consists of PPC modules that regulate the inlet flow and pressure of the gases using software controlled by a Clarus 600 GC method.

CAUTION

The PSS injector is shipped with the wide-bore liner installed but without the quartz wool packing. Before using the injector, remove the liner and pack it with quartz wool.

Page 101: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-27

Connect a column to the programmed split/splitless injector in either one of the following two modes:

• Split/Splitless Mode

• On-Column Mode using the hourglass liner

Connect the Column to the PSS Injector

The following five steps summarize how to connect a column to the PSS injector in the split/splitless mode:

Step 1. Remove the PSS injector liner.

Step 2. Select an appropriate PSS injector liner.

Step 3. Pack the PSS injector liner with quartz wool.

Step 4. Reinstall the liner in the PSS injector.

Step 5. Connect a column to the PSS injector.

Step 1. Remove the PSS Injector Liner.

To remove the PSS injector liner:

1. Ensure that the injector heater has been turned off.

Allow the injector to cool until it is slightly warm to the touch. Cooling the injector to too low a temperature (less than 80 °C) will make it difficult to remove the injector liner.

2. Remove the septum cap.

Page 102: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-28

Figure 6-15. Removing the septum cap.

3. Remove the injector cover.

Figure 6-16. Removing the injector cover.

4. Loosen the threaded collar by using the spanner (P/N N610-1359) provided, then remove the threaded collar.

Page 103: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-29

Figure 6-17. Loosening the threaded collar.

5. Replace the septum cap on the injector.

6. Pull the septum cap upwards to remove the septum purge assembly.

Figure 6-18. Removing the septum purge assembly.

The carrier gas inlet line is coiled to allow you to pull the septum purge assembly over to the side and gain access to the liner.

Page 104: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-30

7. Ensure that the PSS injector liner is cool, then insert the end of the PSS liner-removal tool (P/N 0250-6247) over the end of the injector liner that is shipped with PSS injector. Remove the liner by lifting it up and out of the injector.

The PSS liner must be cool (no hotter than 100 °C) or the liner-removal tool will melt! The end of the PSS liner-removal tool may flare out with use. If this happens, cut off the flared end with a razor blade or scissors.

CAUTION

The PSS injector does not have a removable fitting at the bottom of the assembly. Be very careful when removing this liner to prevent breaking it. As the injector cools, the O-ring adheres to the metal base of the injector body. Use a small screwdriver to loosen the O-ring, then remove the liner and O-ring. Be careful not to scratch the barrel where the O-ring seals. Replace the O-ring with a new O-ring when you replace the liner.

Figure 6-19. Removing a PSS injector liner.

Page 105: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-31

Step 2. Select an Appropriate PSS Injector Liner.

The PSS injector uses the following three liners:

• 2-mm i.d. PSS injector wide-bore liner (P/N N612-1004)

• 1-mm i.d. PSS injector narrow-bore liner (P/N N612-1006)

• PSS injector on-column (hourglass) liner (P/N N610-1539)

The PSS injector is operated in the inlet-programmed mode for split or splitless injection with either the 2-mm or 1-mm i.d. liner. If used in the “hot” split or splitless mode, the 2-mm i.d. liner should be used. It is recommended that you do not inject more than approximately 0.5 μL in the “Hot” mode. Refer to Table 6-1. For PSS on-column operation, use the hourglass liner and the oven-program mode. The 2-mm i.d. PSS liner that is used for either split or splitless operation should be packed with quartz wool as described in this chapter. The 1-mm i.d. PSS liner may give better early-eluting peak resolution in the split or splitless mode. This liner should be used for samples with early-eluting peaks that cannot be resolved by additional solute trapping/focusing (by lowering the initial oven temperature or by using a column with a thicker stationary-phase film).

Step 3. Pack the PSS Injector Liner with Quartz Wool.

CAUTION Never pack the hourglass liner with wool.

To wipe the syringe needle, we recommend packing a small amount of quartz wool in the top portion of the liner regardless of the liner type (wide-bore or narrow-bore but never the hourglass) or injector mode (split or splitless). Quartz wool assures that reproducible volumes are injected by wiping the syringe needle every time it is inserted. Remove the liner and replace the quartz wool on a regular basis, particularly if your samples contain nonvolatile components that could build up on the wool. This could cause adsorption of peaks of interest,

Page 106: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-32

tailing, and loss of sensitivity. Remove the wool with a small hook on the end of a thin wire, or blow it out using compressed air.

Figure 6-20. PSS liners packed with quartz wool.

Packing a PSS Injector Liner for the Splitless Mode

CAUTION

To prevent contaminating the quartz wool when packing the injection liner, wear vinyl, powder-free, disposable gloves (for example, the same type of gloves used to perform maintenance on TurboMass).

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch o.d. rod (P/N N610-T100), push the quartz wool into the liner. Pack a 2.5-cm (one-inch)

Page 107: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-33

piece of quartz wool loosely below the top ground portion of the liner (see Figure 6-21). The sample is then injected into the wool, thereby preventing the delivery of sample beyond the column. The wool also wipes the syringe needle upon injection.

NOTE: The narrow-bore liner is more difficult to pack because of its small i.d. However, there is a dimple in the middle of the liner to hold the wool in place. Do not pack the wool too tightly!

When the narrow-bore liner is installed, use a 5-cm needle-length syringe when making manual injections.

Packing PSS Injector Liner for the Split Mode

CAUTION

To prevent contaminating the quartz wool when packing the injection liner, wear vinyl, powder-free, disposable gloves.

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch o.d. rod (P/N N610-T100), push the quartz wool into the liner. Pack the wool tightly3 from the dimple upwards (about one inch [2.5 cm]). Loosely pack quartz wool in the top portion of the liner to wipe the syringe needle upon injection.

NOTE: As you can see in Figure 6-20, each liner has an O-ring installed on the part furthest away from the dimple on the PSS injector liner. If the O-ring has adhered to the liner, you may not be able to easily remove the liner. Use a small screwdriver to dislodge the O-ring before removing the liner and O-ring. Be careful not to scratch the barrel where the O-ring seals.

Step 4: Reinstall the Liner in the PSS Injector.

To reinstall the liner:

3 The recovery of high molecular weight components (e.g., C40) may be improved if the liner is loosely packed.

Page 108: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-34

1. Install a new O-ring on the top portion of the liner.

2. Insert the liner in the injector body.

3. Place the septum purge assembly over the liner.

4. Press the septum purge assembly down to correctly position the liner in the injector.

Make sure that you tightly secure the septum purge assembly to the injector base with the spanner.

Step 5. Connect a Column to the PSS Injector.

To connect a column:

1. Insert a 1/16-inch column nut (P/N 0990-3392) and 1/16-inch graphite ferrule (0.8 mm i.d., P/N 0992-0141 or 0.5 mm i.d., P/N 0990-3700) over one end of the column as shown below:

CAUTION

This injector terminates in a 1/16-inch fitting. This fitting is fragile. To preserve the integrity of the fitting, carefully connect the nut to prevent cross-threading the fitting and/or overtightening the nut on the fitting. You can also preserve the integrity of the fitting by allowing the injector to cool before connecting a nut.

Page 109: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-35

Figure 6-21. Column, nut, and ferrule on the injector end of a narrow-

bore capillary column.

NOTE: Make certain that the tapered end of the ferrule (in Figure 6-21) is facing towards the nut.

2. Cut off about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N930-1376, pkg. of 10 scribes). Break off the tubing at the score mark so that the break is clean and square. Examine the cut with a magnifying glass and compare it to Figure 6-22.

Figure 6-22. Example of a good cut and bad cuts.

3. Position the column nut on the column so that the back of the nut is 3.8 cm to 4.4 cm (1 1/2 to 1 3/4 inches) from the end of the column.

4. Using typewriter "white-out" or a felt tipped pen, make a mark on the column just beyond the back edge of the column nut (see Figure 6-23).

Page 110: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-36

CAUTION

To avoid contaminating the system, make certain that the nut and ferrule do not come into contact with the mark on the column.

PSSInjectorFitting

1/16" ColumnNut

Capillary Column

Figure 6-23. Capillary column attached to PSS injector fitting.

Connect the Column to the PSS Injector in the On-Column Mode Using the Hourglass Liner

The on-column mode is not recommended for use with columns having an internal diameter of 0.1 mm. This is due to the flow differences between the pre-column (0.53 mm i.d.) and the column.

The following three steps summarize how to connect a column to the PSS injector in the on-column mode:

Step 1. Remove the PSS split/splitless injector liner.

Step 2. Install the hourglass liner in the PSS injector.

Step 3. Connect a column to the PSS injector.

Page 111: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-37

Step 1. Remove the PSS Split/Splitless Injector Liner.

To remove the PSS injector liner:

1. Ensure that the injector heater has been turned off.

Allow the injector to cool until it is slightly warm to the touch. Cooling the injector to too low a temperature (less than 80 °C) will make it difficult to remove the injector liner.

2. Remove the septum cap.

Figure 6-24. Removing the septum cap.

3. Remove the injector cover.

Page 112: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-38

Figure 6-25. Removing the injector cover.

4. Loosen the threaded collar using the spanner (P/N N610-1359) provided, then remove the threaded collar.

Figure 6-26. Loosening the threaded collar.

5. Replace the septum cap on the injector.

6. Pull the septum cap upwards to remove the septum purge assembly.

Page 113: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-39

Figure 6-27. Removing the septum purge assembly.

The carrier gas inlet line is coiled to allow you to pull the septum purge assembly over to the side and gain access to the liner.

7. Ensure that the liner is cool, then insert the end of the PSS liner-removal tool (P/N 0250-6247) over the end of the wide-bore or narrow-bore PSS injector liner. Remove the liner by lifting it up. Gently probe the O-ring if it sticks to the injector body.

The PSS liner must be cool (no hotter than 100 °C) or the tubing will melt! The end of the PSS liner-removal tool may flare out with use. If this happens, cut off the flared end with a razor blade or scissors.

Page 114: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-40

Figure 6-28. Removing an injector liner.

Step 2. Install the Hourglass Liner in the PSS Injector.

To install the hourglass liner:

1. Replace the existing liner with the hourglass liner. Install the liner with the hourglass end closest to the septum and use a new O-ring.

Figure 6-29. Hourglass liner (P/N N610-1539).

2. Replace the septum purge assembly removed previously. DO NOT secure it with the collar at this time.

Page 115: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-41

We recommend the following procedure to connect a column and pre-column to the universal connector. You may find that changing the order of the steps is more convenient for you. However, the critical concerns are:

• Making straight, clean, even cuts on the column/pre-column.

• Wetting the ends of the column/pre-column.

• Wetting the fused-silica universal connector/hourglass needle guide.

• GENTLY twisting the column/pre-column into the hourglass fused-silica universal connector.

• Conditioning the mechanical joint (this is critical).

In the on-column mode, you will install a column (0.53 mm i.d.) or a deactivated fused silica pre-column (0.53 mm i.d.) in the injector body. This will enable the sample to be deposited directly into the column or the pre-column.

Step 3. Connect a Column to the PSS Injector

To connect a column to the PSS in the on-column mode using the hourglass liner:

1. Insert a 1/16-inch column nut (P/N 0990-3392) and 1/16-inch graphite ferrule (P/N 0992-0141) over the other end of the 0.53-mm i.d. column or pre-column as shown below:

CAUTION

If you are using a 0.53-mm i.d. column, insert it directly into the injector. If you are using a column with an i.d. less than 0.53 mm, insert a one-meter piece of deactivated 0.53-mm i.d. fused silica (pre-column) into the injector. This pre-column will also serve as a retention gap. In addition, the long length of this tubing will enable you to change columns and still have enough of the pre-column to make numerous new column connections. Connect a column to the end of this pre-column (in the oven) with a universal connector (P/N N930-2149).

Page 116: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-42

Figure 6-30. Column, Nut, and Ferrule on the injector end of a narrow-bore capillary column.

CAUTION

Ensure that the tapered end of the ferrule is facing towards the nut as shown above. We have found that graphite/Vespel ferrules used to connect the column to the injector fitting loosen after the injector temperature cycles several times. Use graphite ferrules to eliminate this problem.

2. Cut off about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N930-1376, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark so that the break is clean and square. If you are using a 0.53-mm i.d. pre-column, make a clean square cut on both ends. Examine the cut with a magnifying glass and compare it to the following figure:

CAUTION

A clean, square cut is especially critical to produce a leak-free seal in the fused-silica universal connector.

Page 117: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-43

Figure 6-31. Example of a good cut and bad cuts.

3. Locate the PSS injector fitting inside the oven.

PSSInjectorFitting

1/16" ColumnNut

Pre-Column(0.53mm o.d.)

Figure 6-32. PSS injector fitting in the column oven.

4. Insert the 0.53-mm i.d. column or pre-column into the PSS injector fitting. Tighten the column nut so the column/pre-column can still be moved but with some resistance.

5. Push the column/pre-column up into the injector until it can not go any further.

You may notice the septum purge assembly move upward.

6. Mark the column with typewriter "white out" just below the column nut.

Page 118: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-44

7. Push the septum purge assembly down into position, place the threaded collar on the assembly, and tighten the threaded collar with the spanner provided.

Figure 6-33. Replacing the septum purge assembly.

8. Using a 1/4-inch wrench, tighten the fitting until the column cannot be pulled out of the fitting.

CAUTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column.

9. Replace the septum cap but do not tighten it. Manually insert the

0.47-mm o.d. syringe needle into the 0.53-mm i.d. column/pre-column. With the syringe still inserted in the injector, tighten the septum cap completely. This aligns the column and septum cap to ensure a smooth movement of the syringe through the septum cap and into the 0.53-mm i.d. column/pre-column.

Page 119: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-45

10. If your column i.d. is smaller than 0.53 mm, connect your column to the 0.53-mm i.d. pre-column as follows:

a) Make a clean, even cut on the end of your column and examine it with a magnifying glass.

b) Drop some solvent (for example, methanol) into both sides of the fused-silica universal connector (P/N N930-2149, package of 5).

c) Wet a tissue with solvent and wipe the end of the column.

d) Insert the column end into the universal connector and gently twist it. The column should seal in the universal connector.

11. Wet a tissue with solvent (for example, methanol) and wipe the end of the pre-column.

12. Carefully push the universal connector, with the attached column, up into the end of the pre-column. Gently twist the universal connector to seal the connection between pre-column and universal connector. This connection is called the mechanical joint.

POCInjectorFitting

1/16-inch ColumnNut

Pre-Column(0.53mm o.d.)

UniversalConnector

(P/N N930-2149)

Column

Figure 6-34. Connecting a column and universal connector to the pre-column to create the mechanical joint.

Page 120: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-46

CAUTION

Typically the split vent is closed in this mode of operation. Condition the mechanical joint between the pre-column and column prior to making any analytical runs. Condition the joint by slowly temperature programming the oven up to 200 °C and holding it at 200 °C for one hour. Do not exceed the recommended temperature of your column. This can be incorporated when you condition the column, see Step E in this Chapter.

If you replace the column, do not reuse the universal connector. Cut the column and pre-column from the universal connector. Replace the universal connector with a new one. The pre-column ought to be long enough so that you can reuse it. If you replace the pre-column, it will also require some conditioning. If the entire installation is new, the column conditioning procedure will condition the pre-column, the column, and the mechanical joint. If you are just replacing the pre-column, you can condition it and the mechanical joint by the above procedure. If a high background persists with a fresh pre-column, condition the pre-column by only increasing the injector temperature for one hour.

CAUTION

When using the PSS injector in the on-column mode or with the autosampler, you must use a special syringe that has a needle o.d. of 0.47 mm (P/N N610-1253 or N610-1380). Refer to Chapter 9 for more details. You must use only the "Norm" injection speed with this syringe when you are in the on-column mode. The "Fast" injection speed will bend this thin needle and the "Slow" injection speed may produce peak break up or distorted peaks. You can achieve better precision in the on-column mode when sample volumes of 1.0 μL or greater are injected.

Page 121: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-47

Step B3 Connect the Column to the Programmed On-Column Injector (POC):

Overview

The Programmed On-Column injector (POC) consists of an hourglass adapter, a deactivated 0.53-mm i.d. fused-silica pre-column, and a fused-silica universal connector. The sample is injected into the POC at a “cool" temperature; the injector temperature is then programmed to increase. This is helpful for samples that are thermally labile and/or of a wide molecular weight range. This injector is used only for trace analysis or diluted solutions.

CAUTION

Only use a syringe that has a 0.47-mm o.d. needle (P/N N610-1380) with this injector.

Page 122: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-48

Figure 6-35. Cutaway view of a Programmed On-Column injector (POC).

About the Pneumatics Control

The POC pneumatics consist of a PPC (programmable pneumatics controlled) modules that regulate the inlet flow and pressure of the gases using software controlled by a Clarus 600 GC method.

Page 123: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-49

CAUTION

This injector terminates in a 1/16-inch fitting. When connecting a nut to this fitting, take special care not to crossthread the fitting or overtighten the nut..

On-column injection is not recommended for columns with an internal diameter of 0.1 mm. This is because of flow differences between the pre-column (0.53 mm) and the column.

CAUTION

If you have ordered the POC without the flow readout option, measure the carrier gas flow before connecting the column to the POC. Proceed to: C2 Setting the Carrier Gas for manual pneumatics and follow the procedure "Set the Carrier Gas Flow for a POC injector using the Soap Bubble Flowmeter" in this Chapter now!

We recommend that you use the following procedure to connect a column and pre-column with the universal connector. You may find that changing the order of the steps is more convenient for you. However, the critical concerns are:

• Making straight, clean, even cuts on the column/pre-column.

• Wetting the ends of the column/pre-column with solvent.

• Wetting the universal connector/hourglass needle guide with solvent.

• GENTLY twisting the column/pre-column into the hourglass universal connector.

• Conditioning the mechanical joint (this is critical).

Page 124: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-50

To connect a column to the programmed on-column injector (POC):

1. Remove the septum cap.

Figure 6-36. Removing a septum cap.

2. Remove the septum shield (P/N N610-1702) with the large end of the liner-removal tool (P/N N610-0102).

Figure 6-37. Liner-removal tool (P/N N610-0102).

3. Remove the hourglass needle guide (P/N N610-1703) with a pair of small pliers or tweezers.

Figure 6-38. Hourglass needle guide.

Page 125: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-51

CAUTION

If you are using a 0.53-mm i.d. column, insert it directly into the injector. If you are using a column with an i.d. of less than 0.53 mm, insert a one-meter piece of deactivated 0.53-mm i.d. fused silica into the injector to act as a pre-column. This pre-column will also serve as a retention gap. In addition, the long length of this tubing will enable you to change columns and still have enough of the pre-column to make numerous new column connections. Connect a column to the end of this pre-column (in the oven) with a universal connector (P/N N930-2149).

4. Insert the column nut (P/N 0990-3392) and ferrule (P/N 0992-0141) on

the end of the column/pre-column.

Figure 6-39. Column, nut, and ferrule on the injector end of a narrow-bore capillary column.

CAUTION

We have found that graphite/Vespel ferrules used to connect the column to the injector fitting loosen after the injector temperature cycles several times. Use graphite ferrules to eliminate this problem. Make certain that the tapered end of the ferrule is facing towards the nut as shown above.

Page 126: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-52

5. Make a clean, square cut on the end of the 0.53-mm column or on both ends of the 0.53-mm pre-column. Wipe the end with a tissue soaked in methanol to remove fragments of polyimide or silica.

CAUTION

A clean, square cut is especially critical to obtain a leak-free seal in the hourglass needle guide and fused-silica universal connector.

Figure 6-40. Example of a clean column cut and bad cuts.

6. Locate the injector fitting in the oven. Then insert the column/pre-column until it protrudes out the septum end of the injector.

7. Thread the nut onto the injector fitting. Then tighten it so that the column/pre-column can still be moved, but you feel some resistance.

8. Push the column/pre-column into the hourglass needle guide and twist it slightly.

9. Place the metal septum shield over the hourglass needle guide and push it into the injector until it is completely seated. Replace the septum cap.

10. Tighten the column nut inside the oven until the column cannot be pulled out of the fitting.

Page 127: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-53

CAUTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column.

You may want to mark the column, just behind the nut, with typewriter "white-out." The mark will act as a guide when you test to see if the fitting is tight enough.

CAUTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column.

11. Loosen the septum cap and manually insert the 0.47-mm o.d. syringe needle into the column/pre-column. With the syringe in the injector, tighten the septum cap completely. This ensures smooth movement of the syringe through the septum cap and into the column/pre-column.

12. If your column i.d. is smaller than 0.53 mm, connect your column to the 0.53-mm i.d. pre-column as follows:

a) Make a clean, even cut on the end of your column and examine it with a magnifying glass.

b) Drop some solvent (for example, methanol) into both sides of the fused-silica universal connector (P/N N930-2149, package of 5).

c) Wet a tissue with solvent and wipe the end of the column.

d) Insert the column end into the universal connector and gently twist it. The column should seal in the universal connector.

13. Wet a tissue with solvent (for example, methanol) and wipe the end of the pre-column.

Page 128: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-54

14. Carefully push the universal connector, with the attached column, up into the end of the pre-column. Gently twist the universal connector to seal the connection between the pre-column and universal connector. This connection is called the mechanical joint.

POCInjectorFitting

1/16-inch ColumnNut

Pre-Column(0.53mm o.d.)

UniversalConnector

(P/N N930-2149)

Column

Figure 6-41. Connecting the column and universal connector to the pre-

column to create the mechanical joint.

Page 129: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-55

CAUTION

Condition the mechanical joint between the pre-column and column prior to making any analytical runs. Condition the joint by slowly temperature programming the oven up to 200 °C and holding it at 200 °C for one hour. Do not exceed the recommended temperature of your column. This can be incorporated when you condition the column, see Step E in this Chapter.

If you replace the column, do not reuse the universal connector. Cut the column and pre-column from the universal connector. Replace the universal connector with a new one. The pre-column ought to be long enough so that you can reuse it.

If you replace the pre-column, it will also require some conditioning. If the entire installation is new, the column conditioning procedure will condition the pre-column, the column, and the mechanical joint. If you are just replacing the pre-column, you can condition it and the mechanical joint by the above procedure. If a high background persists with a fresh pre-column, condition the pre-column by only increasing the injector temperature for one hour

CAUTION

When using the POC with the autosampler, you must use a special syringe that has a needle o.d. of 0.47 mm (P/N N610-1253 or N610-1380). Refer to Chapter 9 for more detail. You must only use the "Norm" injection speed with this syringe in the on-column mode. The "Fast" injection speed will bend this thin needle and the "Slow" injection speed may produce peak break up or distorted peaks. You can achieve better precision in the on-column mode when injecting sample volumes of 1.0 mL or greater.

Page 130: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-56

Step C: Set the Carrier Gas

This step describes how to set the carrier gas for PPC modules

Setting the Carrier Gas for PPC Modules

The following procedure describes how to:

• Set the Carrier Gas Pressure for the Split/Splitless Injector (CAP) and Programmed Split/Splitless Injector (PSS)

• Set the Carrier Gas Flow for the Programmed On-Column Injector (POC)

Setting the Carrier Gas Pressure for the Split/Splitless Injector (CAP) and Programmed Split/Splitless Injector (PSS)

NOTE: The line pressure should be at least 15 psi greater than the highest carrier gas pressure setting, up to a maximum of 100 psig.

To set the carrier gas pressure:

1. Turn on the carrier gas at the tank. Then adjust the line pressure to 60 - 90 psig.

2. Since a capillary injector is in position A, and Capillary Control is selected the Carrier Control drop down menu in the Configuration screen allows you to select Pressure, Flow, or Velocity.

The following screen show Pressure selected.

Page 131: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-57

3. From the System Status screen, select either the A or B injector (the example here shows the A - PSS injector ready for setting the carrier gas). In this example, touch the Channel A PSS injector button

.

Page 132: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-58

4. In the following window touch the Program button in the Carrier Gas field.

5. Using the up and down arrow or keypad buttons enter an Initial pressure (0.1 to 75 psig). The following screen shows 2.0 psi.

Page 133: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-59

6. Using the up and down arrow or keypad buttons enter the appropriate values. Once entered select Close to exit.

NOTE: The pressure readout display is factory configured to display the actual pressure as psig.

Page 134: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-60

7. Enter a pressure set point. (See Suggested Capillary Column Pressures on Table 6-5.).

Setting the Carrier Gas Flow for the Programmed On-Column Injector (POC)

To set the carrier gas flow:

1. Turn on the carrier gas at the tank. Adjust the line pressure to 60 - 90 psig.

2. From the System Status screen select either A or B injector (the example here shows the B injector ready for setting the carrier gas).

3. In the following window touch the Program button in the Carrier Gas field.

Page 135: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-61

The Carrier Pressure Program screen appears.

4. Using the up and down arrow or keypad buttons enter a pressure (0.1 to 100 psig).

NOTE: If you wish to generate a flow program method, refer to the Clarus 600 User’s Guide (P/N 0993-6780).

Page 136: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-62

Suggested Capillary Column Pressures:

Table 6-2. Calculated Pressure Drops (psig) for 10m Column4 Column I.D. (μm)

_ u 5 320 250 100 10 1.0 2.4 10.0 20 2.1 4.9 21.2 30 3.1 7.3 31.8 40 4.1 9.8 42.3 60 6.2 14.6 63.5 80 8.3 19.5 84.7

Table 6-3. Calculated Pressure Drops (psig) for 25m Columns4

Column I.D. (μm) _ u 5 320 250 100 10 2.6 6.1 26.5 20 5.2 12.2 52.9 30 7.8 18.3 79.4 40 10.3 24.4 - 60 15.5 36.6 - 80 20.7 48.8 -

4 In psig, using helium as a carrier gas at 100 °C. 5 Average linear velocity (cm/sec).

Page 137: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-63

Table 6-4. Calculated Pressure Drops (psig) for 50m Columns4 Column I.D. (μm)

_ u 5 320 250 100 10 5.2 12.2 52.9 20 10.3 24.4 - 30 15.5 36.6 - 40 20.7 48.8 - 60 31.0 73.2 - 80 41.3 - -

Setting the Carrier Gas Flow for the Programmed On-Column Injector (POC) Using the Optional Flow Readout To set the carrier gas flow with the optional flow readout:

1. Turn on the carrier gas at the tank. Adjust the line pressure to 90 psig.

2. From the System Status screen select either A or B injector (the example here shows the B injector ready for setting the carrier gas).

Page 138: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-64

3. In the following window touch the Program button in the Carrier Gas field.

4. Using the up and down arrow or keypad buttons enter the pressure (0.1 to 100 psig) values.

4. Type the desired flow set-point value.

Page 139: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-65

Set the Carrier Gas Flow for the Programmed On-Column Injector (POC) Using the Soap Bubble Flowmeter

To perform this procedure, you must know how to measure carrier gas flow using a soap bubble flowmeter and the Clarus 500 GC built-in stopwatch. If you are not familiar with this measurement, please read "Using the Built-in Stopwatch" in Chapter 4 before you install a column.

To set the carrier gas flow using the soap bubble flowmeter:

1. Locate the POC injector fitting inside the column oven. The POC injector fitting is shown below:

POCInjectorFitting

Figure 6-44. POC injector fitting in the column oven.

2. Attach a soap bubble flowmeter to the POC injector fitting.

3. Turn on the carrier gas at the tank. Adjust the line pressure to 90 psig.

4. From the System Status Screen select the Tools button and then select Utilities.

Page 140: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-66

5. In the Utilities window select the Stopwatch icon.

6. Using the up and down arrow or keypad buttons enter the appropriate values and press Start.

Page 141: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-67

7. Start the carrier gas flowing by turning the flow controller knob

counterclockwise and measure the flow.

NOTE: For best accuracy, use a soap bubble flowmeter volume that gives a reading of at least 30 seconds.

8. Adjust the flow to the desired set point by repeatedly measuring the flow and turning the flow controller knob counterclockwise to increase the flow, clockwise to decrease the flow, until the desired flow is obtained.

9. Disconnect the soap bubble flowmeter before proceeding to the next step to Step D “Leak Test all New Connections.”

Table 6-5. Suggested Capillary Column Flow Rates Using Helium As Carrier

Column i.d. (µm) 250 320 530

Flow mL/min 0.6 - 0.8 1.0 - 1.5 2.5 - 4.0

Page 142: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-68

Step D Leak Test All New Connections:

PPC Pneumatics (POC Injector Only)

1. In the following example select the POC from the System Status screen.

2. In the next window select the Carrier Gas Program button.

Page 143: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-69

3. In the Program screen enter a minimum value between 0.1 and 100 psig. Enter a maximum value between 0.1 and 100 psig.

Page 144: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-70

The Clarus 600 GC monitors the inlet column head pressure with a carrier-gas mass-flow controller in use. You enter the minimum and maximum values after selecting the carrier-gas Flow mode in configuration. If the inlet column head pressure drops below or rises above the set limits for more than one minute, a warning message appears on the display. The following error message will appear: Carrier (with specific carrier number) unable to maintain pressure.

You should correct the leak; the most common area would be the injector septum. Then clear the message by touching OK before continuing to use the GC. This will stop an autosampler program from continuing. If you do not make any entries, leak checking will not be done. If you are temperature programming, then the value you enter for the maximum must be above the column head pressure at the highest oven temperature. You can just enter a value for the minimum and leave the maximum off (zero).

Page 145: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-71

Step E: Condition the Column and the Mechanical Joint Between the Pre-column and Column:

This section contains a suggested temperature program for conditioning a column. The program starts off by holding the oven temperature at a medium value for 10 minutes, gradually increasing the oven temperature at a fixed rate (5 °C/min) to the column operating temperature, then holding that temperature overnight with the carrier gas flowing.

CAUTION

The temperatures shown in the examples which follow should be used as guidelines. Please refer to the column manufacturer's operating instructions for specific temperature recommendations.

CAUTION

To keep the injector clean, open the split vent to direct more gas through the injector.

Page 146: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-72

To condition the column:

1. Close the oven door and from the System Status screen touch the Oven button.

2. In the oven screen enter an oven temperature set point of 50 °C and enter a (Hold) time of 10.

Page 147: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-73

3. To add another program step, enter a Rate of 5(ºC/min).

4. For Temp 2, enter a set point 25 to 50 ºC above your planned analytical operating temperature.

CAUTION

To avoid damaging the column, do not enter a temperature higher than the maximum recommended temperature specified by the column manufacturer.

5. Configure the injector for the Oven mode. From the System Status screen touch the Tools button and select the Configuration menu.

Page 148: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-74

6. In the Configuration menu select the Injector button. In the following example Injector B has been selected to be configured.

Page 149: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-75

7. Select the Program button.

8. Touch the OK button to close this screen then touch the Close button to close the Configuration screen.

Page 150: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-76

9. Touch the a detector button (for example FID) and turn the Detector Temperature off. Allow the system to run overnight.

10. In the morning, reset the oven the oven temperature set point to the initial Temp at the beginning of the temperature program. See step 4 earlier in this procedure.

11. Open the oven door. Allow the oven to cool until the oven fan turns off. This occurs when the oven cools down to 40 °C.

Page 151: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-77

Step F: Connect the Column to the Detector:

NOTE: Before connecting a wide-bore column (0.53-mm i.d.) to a Photoionization Detector (PID), remove the 1/8-inch diameter receiver from the PID base.

PID/ELCDSeries Tubing

PIDReceiver

Figure 6-45. PID receiver viewed inside the column oven

1. Place the column over the hanger so that no part of the column touches the bottom or sides of the oven.

2. Insert a 1/8-inch column nut and graphite ferrule over the free end of the column as shown below:

Figure 6-46. Nut and ferrule on the detector end of a narrow-bore

capillary column.

3. Cut about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N930-1376, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark so that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Page 152: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-78

Figure 6-47. Example of a good column cut and bad cuts.

4. Mark the column the following distances from the end using correction fluid (white-out) or a felt-tipped pen:

Column Diameter

Detector to which you are attaching the column

Distance from back of nut

�0.53 mm i.d. FID 2.75 inches (70 mm)

�0.53 mm i.d. ECD6 2.5 inches (64 mm)

�0.53 mm i.d. TCD 4 inches (103 mm)

�0.53 mm i.d. NPD 3 inches (77 mm)

0.53 mm i.d. FPD 2.5 inches (64 mm)

�0.32 mm i.d. FPD 4 inches (103 mm)

0.53 mm i.d. PID 4 inches (103 mm)

�0.32 mm i.d. PID 5.25 inches (135 mm)

�0.53 mm i.d. ElCD 3.5 inches (90 mm)

5. From the System Status screen select the Tools menu. From the tools menu select Utlitities.

6 A glass-lined receiver (P/N N600-0968) is available to reduce high background readings. For more details, see the Important on page 8-22.

Page 153: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-79

6. In the Utilities screen select the Column Length Calc icon.

7. Using the drop down menus you will be able to select the proper detector and injector for the calculator. The using the plus and minus buttons input the length from the back of the nut and the Column Length Calculator will automatically calculate the column length in millimeters or inches.

Page 154: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-80

CAUTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column.

8. Locate the detector fitting protruding from the right side of the oven roof.

9. Insert the column into the detector fitting, keeping the mark just behind the column nut.

10. While holding the column in position, hand-tighten the column nut.

11. Hold the detector fitting steady with one of the 7/16-inch wrenches as you gradually tighten the column nut with the other wrench. Tighten the nut only until you cannot pull the column out of the nut. DO NOT OVERTIGHTEN THE NUT!

Page 155: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-81

DetectorFitting

ColumnNut

Figure 6-48. Capillary column connected to the detector fitting.

CAUTION

Make certain that no part of the column touches the walls or bottom of the oven.

CAUTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column.

Step G: Leak Test All New Connections:

Test the detector connection for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To avoid contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections.

Page 156: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-82

Step H: Set up the Split Mode for a CAP or PSS Injector:

The Split Mode is used to analyze concentrated samples. In this mode only part of the sample enters the column; the remainder is split and vented through a charcoal filter to atmosphere. This step describes how to set the split mode for PPC modules.

NOTE: In the Split Mode, the split vent is always open.

NOTE: The injector is shipped with an unpacked wide-bore injector liner installed. Remove the liner and pack it with quartz wool before running your analysis. Refer to Step B, Connect the Column to the Injector, in this chapter

The following procedure assumes that the carrier gas pressure has been set (see Step C in this chapter).

For information on Setting the Split Mode Using PPC, Setting up the Splitless Mode for a CAP or PSS Injector, Setting the Splitless Mode Using PPC Modules, see the Clarus 600 GC Users Guide (P/N 0993-6780) Using the Active Method chapter.

Page 157: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-83

PSS and POC Operating Hints:

The Programmed Split/Splitless (PSS) and Programmed On-Column (POC) injectors can be operated in either the oven-programming mode or the inlet-programming mode. For specific instructions, refer to Chapter 9 of this manual.

CAUTION

If you have the subambient option, the POC and PSS injectors are linked to the oven subambient option; therefore, you cannot operate the injectors below the oven subambient temperature.

Oven Programming Mode

This is the default mode for both the POC and the PSS injectors. This mode is the easiest to operate since only the oven temperature program needs to be entered into the method. In this mode, the injector will follow the oven temperature profile plus five degrees Celsius. In this mode, the sample is introduced into the injector when the temperature of the inlet is at the boiling point of the solvent. Then start the injector and oven temperature program.

If the initial temperature of the oven is above the boiling point of the solvent you are using, then it would be better to modify the oven program to start at a lower temperature or to configure the injector for the Inlet mode and set a temperature for the injector separate from the oven temperature.

Inlet Programming Mode

This mode permits the use of independent injector temperatures and rates that you define in the method. The injector will be programmed for injector temperature 1, injector time 1, injector rate 1, injector temperature 2, injector time 2, etc. You can program up to three temperatures and two ramps for each PSS or POC configured in the inlet mode. It is important to set the initial injector temperature to approximately the boiling point of the solvent you are using.

Page 158: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-84

CAUTION

The PSS can be used in the "hot" split or splitless mode. This, however, is not recommended for use with the 1 mm i.d. liner; it could cause solvent flashback in the injector. This mode should be used with caution depending upon the solvent and temperatures you choose. Always use the 2-mm i.d. injector liner. See Table 6-1at the beginning of this chapter.

CAUTION

When using the PSS in the on-column mode, or the POC with the autosampler, you must use a special syringe that has a needle o.d. of 0.47 mm (P/N N610-1253 or N610-1380). You must only use the "Norm" injection speed with this syringe in the on-column mode. The "Fast" injection speed will bend this thin needle and the "Slow" injection speed may produce peak break up or distorted peaks. You can achieve better precision in the on-column mode when sample volumes of 1.0 uL or greater are injected.

If a column is used extensively at high temperatures (350 °C or greater), the polyimide may become very brittle. This brittleness will cause the column to fracture when you try to seal it in the universal adapter. If you wish to continue using a brittle column, use a low dead-volume union instead of the universal adapter.

Solvent Purge Mode

The PSS injector can be operated in the solvent purge mode without any modification. This technique is an enhanced mode of the splitless injection and may use either the 0.5-μL or 5.0-μL syringe. To obtain better detection limits, you may also use a 50-μL syringe to make larger volume injections. In the solvent purge mode, the split vent is open at the time of injection and the PSS injector temperature is held below the boiling point of the solvent. Once most of the solvent has vented, the split vent closes and the PSS injector is heated so that it can transfer the analytes onto the column.

Page 159: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

6-85

You must select the starting temperature of the PSS injector based on the solvent you are using. In addition, to eliminate most of the solvent without losing any of the compounds of interest, you must experimentally determine the minimum split flow and vent time.

The initial value of the split vent can be ON or OFF, depending upon your solvent volatility. When you make your injection, the split vent OPENS (turns ON) and stays open for the time you have selected. After that time, the split vent CLOSES (turns OFF via a timed event) and the PSS injector heats to the selected temperature. To help maintain good peak shape and quantitative recovery, we recommend holding the Clarus GC oven temperature at its starting value until the PSS injector temperature reaches its final temperature and all of the sample has been transferred onto the column. To minimize tailing of the residual solvent and ensure that the system is completely cleaned for the next injection, start the oven temperature program and re-open the split vent.

The charcoal trap's capacity is too small to handle the large volumes of solvent and it will saturate quickly. When making large volume injections, we recommend disconnecting the charcoal trap from the PSS injector. For details, refer to “Replacing the Charcoal Trap on the Split/Splitless, CAP, and PSS Injectors” in Chapter 9, Maintenance. To retain the vented solvent, we strongly recommend a larger capacity charcoal trap after (downstream from) the vent on the AutoSystem.

NOTE: When using the PSS injector in the split mode, the charcoal trap is required to maintain the split ratio throughout the injection.

Page 160: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Installing a Capillary Column

6-86

Calculating a Capillary Column Split Ratio

The following procedure and examples show how to calculate the capillary column split ratio using PPC.

1. Enter the column length, I.D., and vacuum compensation values.

2. Measure the unretained peak time and correct the column length, see Chapter 8.

3. Read the column and split flows from the screen on the Clarus 600 GC.

4. Calculate the Split Ratio:

Flow rate from the split vent in mL/min + Flow rate of the column Split Rate = _________________________________________________________

Flow rate of the column

Page 161: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent 7

Page 162: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-2

Page 163: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-3

Overview

PreVent™ is an enhanced capillary inlet system for the Clarus 600 GC that uses columns with an inside diameter (i.d.) between 0.25 mm and 0.53 mm. You can use the pre-column venting technique to:

• Manage difficult samples

• Protect the column and gas chromatograph

• Lower detection limits

• Increase analytical throughput

PreVent is used in the following five Modes of Operation:

1. Column Isolation

2. Solvent Purge (PSS injector only)

3. Large Volume Injection (PSS injector only)

4. Sample Residue Purge (PSS injector only)

5. Time Saver

This chapter describes how to install the PreVent adapter and restrictor on an injector or detector. Based on your mode of operation, the PreVent adapter is installed on either the injector or detector (including TurboMass MS). Turn to the procedure in this chapter that describes how to install PreVent:

Installing PreVent on an Injector

Installing PreVent on a Detector

Installing PreVent on a TurboMass MS Detector

Page 164: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-4

NOTE: The instructions in this chapter assume that the PreVent option has been installed in the front injector position (Channel A). The mid-point pressure for Channel A is controlled by valve 3, and the mid-point pressure for Channel B is controlled by valve 4.

NOTE: The PreVent option contains a bleed flow of nominally 15 mL/min of carrier gas that constantly flows and vents to atmosphere when you set the pressure on the PreVent mid-point module. If your carrier gas is hydrogen, you should connect tubing to the barbed fitting on the bleed controller (on the underside of the GC top cover), and direct it to a venting hood.

Page 165: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-5

Installing PreVent on an Injector

The following procedure describes how to install the PreVent adapter on a capillary inlet system (CAP and/or PSS) and how to make it ready for an analysis.

Summary

The following steps summarize how to install an enhanced capillary inlet system to an injector and get it ready for use:

A. Install the restrictor in the PreVent adapter.

B. Connect the PreVent adapter to the injector.

C. Connect the column to the PreVent adapter.

D. Leak-check the system.

E. Connect the column to the detector.

F. Set the initial pressures.

G. Condition the system.

A. Install the Restrictor in the PreVent Adapter

The following restrictors are available for the injector:

• 0.06 mm i.d. restrictor (P/N N610-3080)

• 0.06 mm i.d. to 0.250 mm i.d. restrictor (P/N N610-0342)

1. Select the appropriate restrictor for your analysis and cut it to the proper length. Refer to the following table:

Page 166: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-6

Table 7-1 Recommended Restrictors

Injector Injection Mode

Restrictor Part Number Length (mm)

Split/Splitless

Split 0.06 mm i.d. N610-3080 108

Split/Splitless

Splitless 0.06 mm i.d. N610-3080 108

PSS Split 0.06 mm i.d. to 0.250 mm i.d1.

N610-0342 104

PSS Splitless 0.06 mm i.d. or N610-3080 104

0.06 mm i.d. to 0.250 mm i.d.

N610-0342

Thread the selected restrictor first through a 0.5 mm graphite ferrule (P/N 0990-3700) and then through the PreVent adapter (P/N N610-0328) until 5 mm protrudes through the column connection on the bottom. When using the 0.06 mm i.d. to 250 mm i.d. restrictor, the narrow end should protrude through the bottom.

CAUTION

To prevent contamination problems and the need for extensive conditioning procedures, ensure that all PreVent components are thoroughly clean. Baking them in an oven prior to assembly is recommended or wiping with a tissue dampened with alcohol. Always wear clean gloves when assembling the PreVent components.

1 This restrictor is pre-cut to the correct length for PSS.

Page 167: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-7

Figure 7-1. Installing the restrictor in the PreVent adapter.

B. Connect the PreVent Adapter to the Injector

1. Verify that the injector liner is correct for your application. If necessary, pack the injector liner with quartz wool. For details, refer to packing a PSS or split/splitless injector liner in Chapter 6.

2. Carefully connect the PreVent adapter by screwing the injector end of the adapter into the capillary injector fitting until it is finger tight.

3. Using a 1/4-inch wrench, tighten the PreVent adapter until the graphite ferrule just starts to tighten on the restrictor.

4. Locate the stainless steel tubing (that supplies the mid-point pressure) inside the oven.

5. Insert a 1/16-inch nut (P/N 0990-3392) and a graphite/Vespel ferrule (P/N 0992-0107) over the end of this tubing.

Page 168: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-8

6. Measure and mark the tubing 1/2-inch from the back of the nut.

7. Insert the tubing into side arm on the PreVent adapter, keeping the mark just behind the nut. Then tighten the nut and ferrule fingertight.

8. Make a leak-free seal by tightening the PreVent adapter, on the capillary injector fitting, another 1/4 to 1/2 turn.

Mid-Point Pressure Tubing

Graphite Vespel Ferrule (P/N 0992-0107)

PreVent Adapter(P/N N610-0329)

DetectorFitting

1/16-inch Nut (P/N 0990-3392) Figure 7-2. Capillary injector fitting inside the oven.

C. Connect the Column to the PreVent Adapter

1. Place the column on the hanger so that no part of the column touches the bottom or sides of the oven.

2. Insert a 1/16-inch extended nut (P/N 0990-3392, pkg. of 5), a graphite/Vespel ferrule (1/16-inch x 0.5 mm P/N 0992-0105, pkg. of 10 for 0.25/0.32 i.d. column, or a 1/16-inch x 0.8 mm P/N 0992-0107, pkg of 10 for a 0.53 mm i.d. column), and a spacer (P/N N610-3079), over one end of the column.

CAUTION

Do not use a graphite ferrule since it will extrude up into the body of the PreVent adapter.

Page 169: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-9

Figure 7-3. Extended nut, graphite/Vespel ferrule, and spacer.

3. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N930-1386, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 7-4. Example of a good column cut and bad cuts.

4. Insert the column end (as shown in Figure 7-5, Step 1) into the hourglass guide (P/N N610-3082).

5. Align the end of the restrictor tubing that protrudes from the PreVent adapter with the open end of the hourglass guide. Then, carefully push the hourglass guide, with the attached column, up into the PreVent adapter as shown in Figure 7-5, Step 2. Notice that the restrictor tubing fits inside the column.

Page 170: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-10

NOTE: Slightly twist the column as you push the hourglass guide into the PreVent adapter.

6. Loosely connect the column nut to the PreVent adapter.

7. Carefully push the column up and into the PreVent adapter until it bottoms in the adapter. Then pull the column back 3 mm as shown in the following figure, Step 3.

8. Use two 1/4-inch wrenches to tighten the nut. DO NOT OVERTIGHTEN THE NUT!

Figure 7-5. Connecting a column to the PreVent injector adapter.

D. Leak-Check the System

1. Locate the detector end of the column.

Page 171: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-11

2. Seal the column end by firmly inserting the detector end of the column into the side of a septum to about the mid-point as shown in the following figure.

Figure 7-6. Column inserted into a septum.

3. Turn on the GC. When the GC is turned on, it is set to the initial default settings.

4. From the System Status screen touch the injector to get to the following screen. Touch the numeric field until it turns black and use the up and down arrow or keypad buttons to get the proper split flow of 1.5 and a ratio of 25.

Page 172: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-12

5. From the Events tab turn off the mid-point carrier gas supply by closing

Valve 3.

6. Allow the system to equilibrate for several minutes, then view the Aux 1

pressure, view the Aux Gas button and view the current psig.

Page 173: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-13

7. Go back to the inject screen to view the pressure read out on the carrier gas which should be at 0 psig.

8. Set the injector pressure to 75 psig.

Page 174: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-14

9. After approximately 5 minutes, if no leaks exist, the Aux 1 reading should be stable at approximately 75 psi. If it is low, then leak-check the fitting connections.

10. If you suspect a leak, test the connections for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To prevent contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections.

If you detect a leak, check and tighten the fittings on the mid-point pressure tee and/or the connections on the rear of the Clarus 600 GC.

CAUTION

Do not overtighten the column nut. If this fitting has a persistent leak, replace the graphite/Vespel ferrule.

11. Turn off the two gas pressures and remove the septum from the end of the column.

E. Connect the Column to the Detector

CAUTION

Make certain that no part of the column touches the walls or bottom of the oven.

NOTE: To connect a column to the TurboMass, refer to the instructions in the TurboMass Software Guide (Part No. 0993-6691).

1. Insert a 1/8-inch column nut and graphite ferrule over the free end of the column as shown in the following figure.

Page 175: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-15

Figure 7-7. Nut and ferrule on the detector end of a narrow-bore

capillary column.

2. Cut about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N930-1376, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark so that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 7-8. Example of a good cut and bad cuts.

3. Mark the column using typewriter "white-out" or a felt-tipped pen. The distances are calculated using Column Length Calc procedure that follows.

Page 176: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-16

4. From the System Status screen select the Tools menu. From the tools menu select Utlitities.

5. In the Utilities screen select the Column Length Calc icon.

6. Using the drop down menus you will be able to select the proper detector and injector for the calculator. The using the plus and minus buttons input the

Page 177: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-17

length from the back of the nut and the Column Length Calculator will automatically calculate the column length in millimeters or inches.

CAUTION

To prevent contaminating the system, ensure that the nut and ferrule do not contact the mark on the column.

7. Locate the detector fitting that protrudes through the right side of the oven roof.

8. Insert the column into the detector fitting, keeping the mark just behind the column nut.

9. While holding the column in position, hand-tighten the column nut.

10. Holding the detector fitting steady with one of the 7/16-inch wrenches, gradually tighten the column nut with the other wrench only until you cannot pull the column out of the nut. DO NOT OVERTIGHTEN THE NUT!

Page 178: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-18

DetectorFitting

ColumnNut

Figure 7-9. Capillary column connected to the detector fitting.

CAUTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column.

F. Set the Initial Pressures

To set the mid-point pressure:

1. From the System Status screen select either A or B injector (the example here shows the B injector ready for setting the carrier gas).

Page 179: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-19

2. In the following window touch the Program graph screen under the Carrier Gas field.

Page 180: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-20

3. Using the up and down arrow or keypad buttons, enter the initial value of 30 psig for the Carrier gas. Press Close to go back to the previous screen (as seen in step 2).

4. In the Events tab touch the Split event and then press Edit. Set Split 1 to 100 mL/min.

Page 181: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-21

5. From the oven tab set the oven temperature to its highest programmed temperature.

6. From the Events tab turn off the mid-point carrier gas supply by closing

Valve 3.

Page 182: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-22

7. Allow the system to equilibrate for several minutes, then view the Aux 1 pressure, view the Aux Gas button and view the current psig.

Note: If the Aux 1 pressure is very low, then the graphite ferrule that holds the restrictor inside the PreVent adapter may be leaking. Remedy this by tightening the adapter on the injector and observe if there is an improvement.

8. Adjust the pressure until the Aux pressure is maintained at about 1.5 psig below the required column inlet pressure. Wait a few minutes for the system to stabilize to ensure that no further changes are required.

9. Open Valve 3 and set the Aux 1 pressure to the required column inlet pressure; for example, 1.5 psig higher than the Aux 1 reading established in step 9 above. This ensures a positive gas flow through the Aux 1 supply to the mid-point tee.

CAUTION

Valve 3 MUST remain open (ON) during PreVent operation.

Condition the system to ready it for an analysis.

Page 183: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-23

G. Condition the System

1. Write down the pressures established for Press 1 and Aux 1 in step F.

2. Set Press 1 to 2 psig, Split 1 to 100 mL/min, and Aux 1 to 80 psig.

• If you are using a PSS, set its temperature to the maximum programmed temperature.

• Set the oven to the maximum programmed temperature.

• Allow the system to bake at this temperature for several hours ⎯ overnight if possible.

• Shorter conditioning times may be possible, but that depends on the condition of the PreVent components and condition of the column.

3. Cool the oven (and if you are using a PSS injector, cool the injector).

4. Reset the initial settings of Press 1 and Aux 1.

The system is now ready to run an analysis.

Page 184: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-24

Installing PreVent on a Detector

This section describes how to install PreVent on a detector (except for TurboMass MS). The next section describes how to install PreVent on TurboMass MS.

Summary

The following steps summarize how to install a PreVent adapter to a detector and make it ready for an analysis:

A. Install the restrictor in the PreVent adapter.

B. Connect the PreVent adapter to the detector.

C. Connect the column to the PreVent adapter to the detector.

D. Leak-check the system.

E. Connect the column to the injector.

F. Set the initial pressures.

G. Condition the system.

A. Install the Restrictor in the PreVent Adapter

1. Cut the 0.075 mm i.d. restrictor (P/N N610-3081) to the proper length for your detector. Refer to the following table:

Page 185: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-25

Table 7-1. Recommended Restrictor Lengths for Detectors

Detector to which you are attaching the adapter

Restrictor Length

PID 195 mm (7.66 inches)

TCD 163 mm (6.40 inches)

FID 137 mm (5.38 inches)

ECD2 124 mm (4.87 inches)

ElCD 150 mm (5.89 inches)

NPD 137 mm (5.38 inches)

FPD 163 mm (6.40 inches)

2. Insert a 1/8-inch graphite ferrule over one end of the restrictor tubing.

3. Insert the end of the restrictor tubing into the top of the PreVent detector adapter (P/N N610-0329).

4. Feed the restrictor tubing through the PreVent adapter until 5 mm protrudes through the column connection on the bottom.

2 A glass-lined receiver (P/N N600-0968) is available to reduce high background readings.

Page 186: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-26

Figure 7-10. Installing the restrictor in the PreVent detector adapter.

B. Connect the PreVent Adapter to the Detector 1. Carefully connect the PreVent adapter by screwing the detector end of

the adapter into the detector fitting until it is finger tight.

2. Using a 7/16-inch wrench, tighten the PreVent adapter until the graphite ferrule just starts to tighten on the restrictor.

3. Locate the stainless steel tubing (that supplies the mid-point pressure) inside the oven.

4. Insert a 1/16-inch nut (P/N 0990-3392) and a graphite/Vespel ferrule (P/N 0992-0107) over the end of this tubing.

5. Measure and mark the tubing 1/2-inch from the back of the nut.

6. Insert the tubing into side arm on the PreVent adapter, keeping the mark just behind the nut. Then tighten the nut and ferrule fingertight.

Page 187: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-27

7. Make a leak-free seal by tightening the PreVent adapter on the detector fitting another 1/4 to 1/2 turn.

Mid-Point Pressure Tubing

Graphite Vespel Ferrule (P/N 0992-0107)

PreVent Adapter(P/N N610-0329)

DetectorFitting

1/16-inch Nut (P/N 0990-3392)

Figure 7-11. Location of the detector fitting.

C. Connect the Column to the PreVent Adapter 1. Place the column on the hanger so that no part of the column touches the

bottom or sides of the oven.

2. Insert a 1/16-inch extended nut (P/N 0990-3392, pkg. of 5), a graphite/Vespel ferrule (1/16-inch x 0.5 mm P/N 0992-0105, pkg. of 10 for 0.25/0.32 i.d. column or a 1/16-inch x 0.8 mm P/N 0992-0107, pkg of 10 for a 0.53 mm i.d. column), and a spacer (P/N N610-3079) over one end of the column.

CAUTION

Do not use a graphite ferrule since it will extrude up into the body of the PreVent adapter.

Page 188: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-28

Figure 7-12. Extended nut, graphite/Vespel ferrule, and spacer.

3. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N930-1386, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 7-13. Example of a good tubing cut and bad cuts.

4. Insert the column end into the hourglass guide (P/N N610-3082).

5. Align the end of the restrictor tubing that protrudes from the PreVent adapter with the open end of the hourglass guide. Then carefully push the hourglass guide, with the attached column, up into the PreVent adapter as shown in the following figure. Notice that the restrictor tubing fits inside the column.

Page 189: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-29

NOTE: Slightly twist the column as you push the hourglass guide into the PreVent adapter.

6. Loosely tighten the column nut.

7. Carefully push the column up into the PreVent adapter until it bottoms in the detector. Then pull the column back 3 mm.

8. Use two 1/4-inch wrenches to tighten the nut. DO NOT OVERTIGHTEN THE NUT!

Figure 7-14. Connecting a column to the PreVent detector adapter.

Page 190: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-30

D. Leak-Check the System

1. Turn on the GC. When the Clarus 600 GC is turned on, it is set to the initial default settings.

2. From the System Status screen touch the injector to get to the following screen. Turn the detector temperature off.

3. From the Events tab turn off the mid-point carrier gas supply by opening

Valve 3.

Page 191: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-31

4. Allow the system to equilibrate for several minutes, then view the Aux 1 pressure, view the Aux Gas button and view the current psig.

Page 192: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-32

5. Go back to the inject screen to view the pressure read out on the carrier gas which should be at 0 psig.

6. After approximately 5 minutes, if no leaks exist, the Aux 1 reading should be stable at approximately 75 psi. If it is low, then leak-check the fitting connections.

7. If you suspect a leak, test the connections for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To prevent contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections.

If you detect a leak, check and tighten the fittings on the mid-point pressure tee and/or the connections on the rear of the GC.

CAUTION

Do not overtighten the column nut. If this fitting has a persistent leak, replace the graphite/Vespel ferrule.

Page 193: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-33

E. Connect the Column to the Injector

CAUTION

Make certain that no part of the column touches the walls or bottom of the oven.

1. Insert a 1/16-inch extended nut (P/N 0990-3392, pkg. of 5) and graphite/Vespel ferrule (1/16-inch x 0.5 mm P/N 0992-0105, pkg. of 10 for 0.25/0.32 i.d. column or a 1/16-inch x 0.8 mm P/N 0992-0107, pkg of 10 for a 0.53 mm i.d. column) over one end of the column.

Figure 7-15. Extended nut and graphite/Vespel ferrule.

2. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N930-1386, pkg of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Page 194: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-34

Figure 7-16. Example of a good column cut and bad cuts.

3. Position the column nut on the column so that the back of the nut is 4.4 cm to 5.1 cm (1 3/4 inches to 2 inches) from the end of the column for the CAP injector.

or

Position the column nut on the column so that the back of the nut is 1 1/2 inches to 1 3/4 inches from the end of the column for the PSS injector.

4. Using typewriter "white out" or a felt-tipped pen, make a mark on the column just beyond the back edge of the column nut (see Figure 6-10).

CAUTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column.

5. Locate the capillary injector fitting inside the oven.

Refer to the following figures to connect the column to the Cap or PSS injector.

Page 195: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-35

CAUTION

The injector terminates in a 1/16-inch fitting. This fitting is fragile. To preserve the integrity of the fitting, carefully connect the nut to prevent cross-threading the fitting and/or overtightening the nut on the fitting. You can also preserve the integrity of the fitting by allowing the injector to cool before connecting a nut.

CAUTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column.

Capillary Injector

Fitting

CapillaryColumn

1/16-inchColumn Nut

Figure 7-17. Capillary column attached to a CAP injector fitting.

Page 196: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-36

PSS Injector

Fitting

CapillaryColumn

1/16-inchColumn Nut

Figure 7-18. Capillary column attached to PSS injector fitting.

F. Set the Initial Pressures

To set the mid-point pressure:

1. From the System Status screen select either A or B injector (the example here shows the B cap injector ready for setting the carrier gas).

Page 197: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-37

2. In the following window touch the Program Graph screen under the Carrier Gas field.

3. Using the up and down arrow or keypad buttons enter the initial value of 30 psig for the Carrier gas.

Page 198: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-38

4. Press Close to go back to the previous screen (as seen in step 2).

5. In the Events tab touch the Split event and then press Edit. Set Split B to 100 mL/min.

6. From the Oven tab set the oven temperature to its highest programmed temperature.

Page 199: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-39

7. From the Events tab turn off the mid-point carrier gas supply by closing

Valve 3.

8. Allow the system to equilibrate for several minutes, then view the Aux 1 pressure, view the Aux Gas button and view the current psig.

Page 200: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-40

Note: If the Aux 1 pressure is very low, then the graphite ferrule that holds the restrictor inside the PreVent adapter may be leaking. Remedy this by tightening the adapter on the injector and observe if there is an improvement.

8. Adjust the pressure until the Aux pressure is maintained at about 1.5 psig below the required column inlet pressure. Wait a few minutes for the system to stabilize to ensure that no further changes are required.

9. Open Valve 3 and set the Aux 1 pressure to the required column inlet pressure; for example, 1.5 psig higher than the Aux 1 reading established in step 9 above. This ensures a positive gas flow through the Aux 1 supply to the mid-point tee.

CAUTION

Valve 3 MUST remain open (ON) during PreVent operation.

Condition the system to ready it for an analysis.

Page 201: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-41

G. Condition the System

1. Write down the pressures established in step F for Press 1 and Aux 1.

2. Set Press 1 to 2 psig, Split 1 to 100 mL/min, and Aux 1 to 80 psig.

3. If you are using a PSS, set its temperature to the maximum programmed temperature. Set the oven to the maximum programmed temperature.

4. Allow the system to bake for several hours ⎯ overnight if possible.

Shorter conditioning times may be possible, but that depends on the condition of the PreVent components and condition of the column. Cool the oven (and if you are using a PSS injector, cool the injector.)

5. Reset the initial settings of Press 1 and Aux 1.

The system is now ready to use for an analysis.

Page 202: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-42

Installing PreVent on a Clarus MS Detector

This procedure describes how to install PreVent on a Clarus MS detector.

Summary

The following steps summarize how to install a PreVent injector adapter to a Clarus MS detector and make it ready for an analysis:

A. Prepare the Clarus MS

C. Install the restrictor in the Clarus MS Transfer Line.

D. Install the PreVent injector adapter.

E. Connect the column to the PreVent injector adapter.

F. Connect the Transfer Line to the Clarus MS

G. Leak-check the system.

H. Connect the column to the injector.

I. Set the initial pressures.

J. Condition the system

A. Prepare the Clarus 600 MS 1. Install the Clarus 600 MS and Clarus 600 GC according to the

instructions given in the Clarus 600 Mass Spectrometer and Clarus 600 GC installation manuals.

2. If a column is installed, cool the Gas Chromatograph oven, the Clarus MS ionizer and the Clarus MS transfer line to ambient temperature. Vent the detector and remove the column from the transfer line.

Page 203: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-43

B. Install the PreVent Injector Adapter in the TMMS Transfer Line

1. Cut a 40cm length of the 0.075mm i.d. fused silica tubing (P/N 610-3081). Make sure that the two ends are cut clean and square. Wipe the outside of the tubing with a tissue dampened with methanol.

2. Feed the fused silica tubing through the PreVent injector adapter (P/N N610-0328) and a graphitized Vespule ferrule (P/N 0992-0105) as shown in the following figure.

Figure 7-19. Feeding the fused silica tubing throught the injector adapter.

3. Introduce the fused silica tubing into the Clarus MS transfer line until it reaches a stop. Loosely screw the PreVent adapter onto the 1/16” threaded union at the end of the Clarus MS transfer line as shown in the following figure.

PreVent Injector Adapter(P/N N610-0328)

Ferrule(P/N 0992-0105)

Fuse silica tubing(P/N 610-3081)

Page 204: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-44

Figure 7-20. Normal connection of column to Clarus MS transfer line.

4. Withdraw the fused silica tubing by about 5mm and using two 1/4 inch wrenches tighten up the PreVent adapter on the union at the end of the Clarus MS transfer line.

5. Cut the fused silica tubing so that about 7mm is exposed beyond the PreVent adapter as shown in the following figure. The cut should be clean and square.

CAUTION

Take care when working near the exposed end of the fused silica tubing. It can be easily broken off.

Page 205: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-45

Figure 7-21. Example of a good column cut and bad cuts.

Refer to the following figure as you perform the following procedure.

6. Locate the stainless steel tubing (that supplies the mid-point pressure) inside the oven.

7. Insert a 1/16-inch nut (P/N 0990-3392) and a graphite/Vespel ferrule (P/N 0992-0107) over the end of this tubing.

8. Measure and mark the tubing 1/2-inch from the back of the nut.

9. Insert the tubing into side arm on the PreVent adapter, keeping the mark just behind the nut. Then tighten the nut and ferrule fingertight.

10. Make a leak-free seal by using a two 1/4-inch wrenches to tighten the PreVent adapter on to the union another 1/4 to 1/2 turn.

C. Connect the Column to the PreVent Injector Adapter

1. Place the column on the hanger so that no part of the column touches the bottom or sides of the oven.

2. Insert a 1/16-inch extended nut (P/N 0990-3392, pkg. of 5) a graphite/ Vespel ferrule (P/N 0992-0105, pkg. of 10), and a spacer (P/N N610-3079) over one end of the column.

Page 206: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-46

CAUTION

Do not use a graphite ferrule since it will extrude up into the body of the PreVent adapter.

Figure 7-22. Extended nut, graphite/Vespel ferrule, and spacer.

3. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N930-1386, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 7-23. Example of a good tubing cut and bad cuts.

4. Insert the column end into the hourglass guide (P/N N610-3082).

Page 207: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-47

5. Align the end of the restrictor tubing that protrudes from the PreVent adapter with the open end of the hourglass guide. Then carefully push the hourglass guide, with the attached column, into the PreVent adapter as shown in the following figure, step 2. Notice that the restrictor tubing fits inside the column.

NOTE: Slightly twist the column as you push the hourglass guide into the PreVent adapter.

6. Loosely connect the column nut to the PreVent adapter.

7. Carefully push the column up and into the PreVent adapter until it bottoms in the adapter. Then pull the column back 3 mm.

8. Use two 1/4-inch wrenches to tighten the nut. DO NOT OVERTIGHTEN THE NUT!

Page 208: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-48

Figure 7-24. Connecting a column to the PreVent adapter in a Clarus MS system.

D. Connect the Transfer Line to the Clarus MS

Install the PreVent device into the Clarus MS.as shown in Figure 7-25.

Take the PreVent ‘injector’ T-piece and screw it onto the 1/16” fitting at the end of the standard Clarus MS transfer line. Be sure the fused silica restrictor is inserted through it, through the heated transfer line and directly into the Clarus MS ion source.

Page 209: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-49

Ensure that the end of the restrictor is exposed beyond the T-piece to enable the column to be threaded over it with the aid of an hourglass guide. Figures 7-26 through 7-28 show details of the installation.

Figure 7-30 shows a detail of the column-restrictor interface. Because the restrictor is inserted directly into the end of the capillary column, there are no issues with active sites or dead volumes. Therefore an optimum chromatographic performance is assured.

Figure 7-25. Diagram showing PreVent device installed into the Clarus MS detector (The Clarus 600 MS is the figure on the left and the Clarus 600 GC is the figure on the right).

Page 210: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-50

Figure 7-26. Column is detached and PreVent adapter is attached to Clarus MS transfer line with fused silica restrictor pushed through into ionizer.

Page 211: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-51

Figure 7-27. Restrictor is sealed within adapter with end protruding. Midpoint gas supply is connected to port on adapter.

Page 212: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-52

Figure 7-28. Hourglass guides restrictor into capillary column

Page 213: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-53

Figure 7-29. Installation completed.

Page 214: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-54

midpoint gas

fused silicarestrictor

‘hour-glass’insert

column

PreVentadapter

Clarus MStransfer line

Figure 7-30. Detail of column-restrictor interface at end of Clarus MS transfer line.

• Once installed, this PreVent configuration will allow the capillary column to be back-flushed and so offers the following benefits:

• Allows injector maintenance without having to cool or vent the vacuum on the mass spectrometer.

• Allows a degraded section of column to be removed from its inlet without the need to cool or pump-down the detector.

• Facilitates easier column replacement.

• Reduces risk of air entering the detector in the event of a leak or column breakage.

• Prevents column bleed from entering the detector while conditioning or when the system is idle.

Page 215: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-55

• Allows the injector liner and quartz wool packing to be deactivated in-situ.

• Reduces analysis times by eliminating the need for extensive temperature programming to elute unwanted less volatile sample residue from the column.

• Increases the life of a column by eliminating temperature programming completely.

• Removes the potential of contaminants before they reach the detector.

It will also have the added benefit of maintaining a constant flow of carrier gas into the MS detector in both forward flow and backflush modes during temperature programming.

E. Leak-Check the System

1 Turn on the GC. When the Clarus 600 GC is turned on, it is set to the initial default settings.

2 From the System Status screen touch the injector to get to the following screen. Touch the numeric field until it turns black and use the plus minus buttons to get the proper split flow of 100 and a ratio of 100.

Page 216: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-56

3. Select Valve 3 (or valve 4 if PreVent is installed in channel 2) from the Event drop down menu.

Page 217: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-57

4. Allow the system to equilibrate for several minutes, then view the Aux 1 pressure, view the Aux Gas button and view the current psig.

5. Go back to the inject screen to view the pressure read out on the carrier gas which should be at 0 psig.

Page 218: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-58

6. Set the injector pressure to 75 psig.

7. After approximately 5 minutes, if no leaks exist, the Aux 1 reading should be stable at approximately 75 psi. f it is low, then leak-check the fitting connections.

8. If you suspect a leak, test the connections for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To prevent contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections.

If you detect a leak, check and tighten the fittings on the mid-point pressure tee and/or the connections on the rear of the GC.

CAUTION

Do not overtighten the column nut. If this fitting has a persistent leak, replace the graphite/Vespel ferrule.

F. Connect the Column to the Injector

CAUTION

Make certain that no part of the column touches the walls or bottom of the oven.

1. Insert a 1/16-inch extended nut (P/N 0990-3392, pkg of 5) and graphite/ Vespel ferrule (P/N 0992-0105, pkg of 10) over one end of the column.

Page 219: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-59

Figure 7-31. Extended nut and graphite/Vespel ferrule.

2. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N930-1386, package of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 7-32. Example of a good column cut and bad cuts.

3. Position the column nut on the column so that the back of the nut is 4.4 cm to 5.1 cm (1 3/4 inches to 2 inches) from the end of the column for the CAP injector.

or

Position the column nut on the column so that the back of the nut is 1 1/2 inches to 1 3/4 inches from the end of the column for the PSS injector.

4. Using typewriter "white out" or a felt-tipped pen, make a mark on the column just beyond the back edge of the column nut.

Page 220: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-60

CAUTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column.

5. Locate the capillary injector fitting inside the oven. Refer to the following figures to connect the column to the Cap or PSS injector.

CAUTION

The injector terminates in a 1/16-inch fitting. This fitting is fragile. To preserve the integrity of the fitting, carefully connect the nut to prevent cross-threading the fitting and/or overtightening the nut on the fitting. You can also preserve the integrity of the fitting by allowing the injector to cool before connecting a nut.

CAUTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column.

Page 221: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-61

Capillary Injector

Fitting

CapillaryColumn

1/16-inchColumn Nut

Figure 7-33. Capillary column attached to a CAP injector fitting.

PSS Injector

Fitting

CapillaryColumn

1/16-inchColumn Nut

Figure 7-34. Capillary column attached to PSS injector fitting.

G. Set the Initial Pressures

1. From the System Status screen select the oven icon.

Page 222: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-62

2. Set the oven to the lowest programmed temperature.

3. Set the Clarus MS transfer line to its operating temperature.

4. From the Events tab turn off the mid-point carrier gas supply by closing Valve 3.

Page 223: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-63

5. Start the Clarus MS vacuum

6. See the gas pressure to 30 psi.

Page 224: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-64

Note: If the aux 1 pressure is very low, then the graphite ferrule that holds the restrictor inside the PreVent adapter may be leaking. Remedy this by tightening the adapter on the injector and observe if there is an improvement.

7. Adjust the pressure until the Aux pressure is maintained at about 1.5 psig below the required column inlet pressure. Wait a few minutes for the system to stabilize to ensure that no further changes are required.

8. Open Valve 3 and set the Aux 1 pressure to the required column inlet pressure; for example, 1.5 psig higher than the Aux 1 reading established in step 9 above. This ensures a positive gas flow through the Aux 1 supply to the mid-point tee.

CAUTION

Valve 3 MUST remain open (ON) during PreVent operation.

Condition the system to ready it for an analysis.

Page 225: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-65

H. Condition the System

1. Write down the pressures established for Press 1 and Aux 1 in step I.

2. Set Press 1 to 2 psig, Split 1 to 100 mL/min, and Aux 1 to 80 psig.

3. Set oven to maximum programmed temperature.

If you are using a PSS, set its temperature to the maximum programmed temperature.

4. Allow the system to bake for several hours ⎯ overnight if possible.

Shorter conditioning times may be possible, but that depends on the condition of the PreVent components and condition of the column.

5. Cool the oven (and if you are using a PSS injector, cool the injector.)

6. Reset the initial settings of Press 1 and Aux 1.

The system is now ready to use for an analysis.

I. Replacing the Restrictor

To replace a restrictor in the PreVent injector or detector adapter:

1. Turn off the GC oven and detector and wait until they are cool to the touch.

2. Turn off all gases.

3. Using a 1/4-inch wrench, remove the 1/16-inch column nut and the column from the PreVent adapter.

4. Using two 1/4-inch wrenches, remove the nut and mid-point pressure line from the side arm of the PreVent adapter.

5. Using the appropriate wrenches, remove the PreVent adapter from either the injector or detector.

Page 226: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-66

• Use two 1/4-inch wrenches to remove the PreVent adapter from the injector fitting.

• Use two 7/16-inch wrenches to remove the PreVent adapter from the detector fitting.

6. Insert a 1/16-inch rod (P/N N610-T100) in the column connection end of the PreVent adapter and push the restrictor tubing and ferrule of the injector end.

7. Install a new restrictor by referring to the appropriate procedure in this chapter.

Figure 7-35. Removing a restrictor from a PreVent adapter.

Page 227: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-67

PreVent Operating Techniques

Summary of the PreVent Techniques

Table 7-3 summarizes the various PreVent techniques and the hardware configurations that support them.

• Column Isolation

• Solvent Purge

• Large Volume Injection

• Sample Residue Purge

• Time Saver

Table 7-3. Summary of PreVent Techniques and Supporting Configurations

Technique Split/Splitless PSS Restrictor in Injector

Restrictor in Detector

Column Isolation Yes Yes Yes Yes* Solvent Purge No Yes Yes Yes

Large Volume Injection No Yes Yes Yes Sample Residue

Purge No Yes Yes** Yes

Time Saver Yes Yes No Yes * Cannot be used while chromatography is in progress ** Not recommended

Table 7-3 demonstrates the advantage of the PSS injector over the conventional Split/Splitless injector in that all the PreVent techniques are supported by this one injector. Several of these techniques are possible with a single restrictor configuration. If a different configuration is required, changing the restrictor type is a simple operation requiring only a few minutes to perform.

Page 228: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-68

Column Isolation Technique

This technique uses the Injector Restrictor Configuration to fully isolate the chromatographic column from actions occurring within the injector. Figures 7-36 and 7-37 illustrate the principle.

For sample introduction, the inlet pressure (P1) is set up to be higher than the mid-point pressure (P2). P1 and P2 are set as described in “Installing PreVent on the Injector” in this chapter. P2 represents the pressure at the column inlet. Carrier gas will now flow from the injector, through the restrictor and into the column, carrying the vaporized sample with it. The sample may be injected manually using a standard syringe or by autosampler. The injection temperature and split vent operation are set as for normal split or splitless injection techniques. The pressures are not adjusted until the injector needs to be isolated. To isolate the column, Press 1 (P1) is simply reduced to below that of Aux 1 (P2) causing the carrier gas to flow backwards from the mid-point ‘T’, through the restrictor, into the injector and out through the split vent. It is important to ensure that the split vent is open (for example, 50 mL/min) during this step or the technique will not work.

P1 P2

F

Press 1 Aux 1

Split Vent Control (Open or Closed)

Injector Restrictor

Chromatographic Column

Detector

(20 psig) (14 psig)

FLOW

Figure 7-36. Normal operation.

Page 229: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-69

P1 P2

F

Press 1 Aux 1

Split Vent Control (50 ml/min)

InjectorRestrictor

Chromatographic Column

Detector

(1 psig) (14 psig)

FLOW

Figure 7-37. Column Isolation Mode.

P1 should not be turned off completely, but to a level (for example, 1.0 psig) that maintains forward flow of carrier gas through the liner to prevent the build-up of contamination within the liner while the restrictor is being backflushed and also to maintain a purge flow out of the injector inlet while it is being serviced with the septum cap removed. The reduction in P1 is not normally included as a timed event in the GC method as this will not be routinely adjusted during each run.

The Column Isolation Mode can be used to support the following actions:

• Septum exchange and conditioning even during chromatography

• Liner exchange and conditioning even during chromatography

• In situ chemical deactivation of the liner even during chromatography

• Column and detector protection while the instrument is idle

Column Isolation Mode is also possible with the restrictor fitted to the detector as shown in Figures 7-38 and 7-39. In this instance, however, chromatography is

Page 230: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-70

not possible during column isolation although it now would be possible to condition a column without the effluent reaching the detector.

P1 P2

F

Aux 1

Split Vent Control (Open or Closed)

Injector

Restrictor

Chromatographic Column Detector

Press 1(20 psig) (6 psig)

FLOW

Figure 7-38. Normal operation.

P1 P2

F

Press 1 Aux 1

Split Vent Control (50mL/min)

Injector

Restrictor

Chromatographic ColumnDetector

(1 psig) (6 psig)

FLOW

Figure 7-39. Alternative Column Isolation Mode.

Page 231: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-71

Solvent Purge Technique This technique uses the ability of the PSS injector to rapidly change its liner temperature according to a user-defined program. Figure 7-40 shows how this ability can be used to remove the solvent from the liner and selectively transfer the less-volatile sample analytes to the column. This technique is best suited to splitless injections of low volatility analytes. While the solvent is being purged from the liner, there is still forward flow into the column and so some of the solvent will still enter the column (and hence the detector). The PreVent system, configured as for the Column Isolation Mode may be used to prevent all traces of the solvent entering the column during the purge process by reversing the flow of carrier gas at the injector column port.

123

Step 1 : Sample injected into cool liner

Step 2 : Solvent and other volatile components swept from cool liner and out of split vent

Step 3 : Liner heated and the vaporized residue swept into column

Figure 7-40. Using the PSS to eliminate solvent from the column.

Table 7-4 shows typical entries that would be made in a method to enable a 1-minute solvent purge at 100 mL/min and 50 °C. Actual settings will depend upon the type and volume of solvent involved. The initial settings for P1 and P2

Page 232: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-72

are established as described in this chapter. The column oven is programmed as for normal splitless injection except that the temperature is not raised until after the solvent purge step is complete.

Table 7-4. Typical Method for Solvent Purge Technique

Time Event Comment INITIAL SPLIT 1 (F) = 100 mL/min Purge flow rate. INITIAL PSS = 50 °C Purge temperature. INITIAL AUX 1 (P2) = 14 psig Mid-point pressure at column inlet. INITIAL PRESS 1 (P1) = 2 psig Pressure in injector lower than mid-

point to keep solvent out of column. 1.00 SPLIT 1 (F) = OFF Close split vent to transfer everything

to column. 1.01 PRESS 1 (P1) = 20 psig Raise injector pressure so carrier gas

flows from liner to column. 1.02 PSS = 350 °C Heat liner to vaporize sample residue. 2.00 SPLIT = 50 mL/min Finally open split vent to clean liner for

next run.

The Solvent Purge Technique may be used to support the following actions:

• Elimination of peak distortion caused by solvent flooding effects as a result of excess liquid solvent entering the column and becoming fractionated by the carrier gas flow

• Elimination of hostile solvents from sensitive columns and detectors

• Large volume injections - (See Large Volume Injections below.)

Large Volume Injections (LVI)

The Solvent Purge Technique can also be used to support LVI applications of the PSS injector. The PSS must be used with a 2-mm i.d. liner which is firmly packed with glass wool. A large volume of liquid sample (up to 100 µl - a 50-µl syringe is available for the autosampler) may be injected into the liner which must be kept at a low temperature to prevent (very) explosive vaporization. The solvent is purged out of the split vent as for the Solvent Purge Technique. The purge time and flow rate need to be sufficient to remove all of the solvent (for example, about 100-200 mL of purge gas is needed to remove 50 µl of n-hexane

Page 233: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-73

at 50 °C). Extended purge times may result in the loss of some of the analytes and so this technique is only suitable for low volatility compounds (for example, lower than n-decane). Typical method entries are given in Table 7-5.

Table 7-5. Typical Method for LVI Technique

Time Event Comment INITIAL SPLIT 1 (F) = 100 mL/min Purge flow rate. INITIAL PSS = 50 °C Purge temperature. INITIAL AUX 1 (P2) = 14 psig Mid-point pressure at column inlet. INITIAL PRESS 1 (P1) = 2 psig Pressure in injector lower than mid-point

to keep solvent out of column. 2.00 SPLIT 1 (F) = OFF Close split vent to transfer everything to

column. 2.01 PRESS 1 (P1) = 20 psig Raise injector pressure so carrier gas

flows from liner to column. 2.02 PSS = 350 °C Heat liner to vaporize sample residue. 3.00 SPLIT = 50 mL/min Finally open split vent to clean liner for

next run.

Sample Residue Purge Technique

The separation described under the Solvent Purge Mode can be reversed so that the volatile sample content enters the column and heavy (unwanted) low volatility sample residue left in the liner can be purged out through the split vent. This technique works best with the detector restrictor PreVent configuration so that the traces of low volatility material that do leave the liner can be backflushed from the column (see next Section). Use of the injector restrictor configuration will not prevent some of the low-volatility material from entering the column during the purge process. The figure below illustrates the separation process inside the PSS liner.

Page 234: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-74

123

Step 1 : Sample injected into cool liner

Step 2 : Volatile components swept from cool liner into column

Step 3 : Liner heated and most of the residue swept out of split vent

Figure 7-41. Using the PSS to eliminate low volatility sample material from the column.

The figures below illustrate how the PreVent system backflushes the heavier sample from the column. For further details on column backflushing refer to the “Time Saver Technique” Table 7-6 lists some typical entries in a method for the Sample Residue Purge technique. The initial settings for P1 and P2 are established as described in this chapter. Split or splitless injection with either a split/splitless or PSS injector is possible and the oven temperature program is set as for normal chromatography.

This technique is useful for the determination of volatile components in heavy oils, natural extracts, etc.

Page 235: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-75

P1 P2

F

Press 1 Aux 1

Split Vent Control (Open or Closed)

Injector

Restrictor

Chromatographic ColumnDetector

(20 psig) (6 psig)

FLOW

Figure 7-42. Injection and chromatography.

P1 P2

F

Press 1 Aux 1

Split Vent Control (100 ml/min)

Injector

Restrictor

Chromatographic ColumnDetector

(1 psig) (80 psig)

FLOW

Figure 7-43. Column backflush.

Page 236: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-76

Table 7-6. Typical Method for Sample Residue Purge Technique

Time Event Comment INITIAL SPLIT 1 (F) = 100 mL/min Injection split flow (can use splitless). INITIAL PSS = 200 °C Injection temperature. INITIAL AUX 1 (P2) = 14 psig Mid-point pressure at column inlet. INITIAL PRESS 1 (P1) = 20 psig Pressure in injector higher than mid-point

to allow sample into column. 2.00 SPLIT 1 (F) = 100 mL/min Open split vent (if not already) to reduce

the amount of further sample entering column.

20.00 PRESS 1 (P1) = 5 psig When last peak of interest has eluted, reduce inlet pressure...

20.01 AUX 1 (P2) = 80 psig and increase mid-point pressure to backflush any sample material left in column.

20.02 PSS = 450 °C Heat liner to bake out any sample residue and purge it out of the split vent.

Time Saver Technique

The PreVent system is able to support the classical, single-column, backflush technique. The use of PPC pneumatics enables the backflush conditions to be optimized for rapid backflush of unwanted sample material after the last peak of interest has eluted. The initial settings for P1 and P2 are established as described in this chapter. Split or splitless injection with either a split/splitless or PSS injector is possible and the oven temperature program is set as for normal chromatography. When the last peak of interest has eluted, the split vent is opened, P1 is reduced and P2 is increased. Any sample material left in the column is rapidly backflushed even at low oven temperatures eliminating the need, in many instances, for temperature programming.

This technique is suitable for the determination of components in any sample where there is unwanted low-volatility material entering the column. By reducing the need for temperature programming to remove such material, it reduces the analysis time and protects the column by not exposing it to high temperatures.

Page 237: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

7-77

Table 7-7. Typical Method for Column Backflush Technique

Time Event Comment INITIAL SPLIT 1 (F) = 100 mL/min Injection split flow (can use splitless). INITIAL AUX 1 (P2) = 14 psig Mid-point pressure at column inlet. INITIAL PRESS 1 (P1) = 20 psig Pressure in injector higher than mid-point

to allow sample into column. 1.00 SPLIT 1 (F) = 100 mL/min Open split vent (if not already) to allow

backflushed components to escape to vent and help clean liner.

20.00 PRESS 1 (P1) = 5 psig When last peak of interest has eluted, reduce inlet pressure...

20.01 AUX 1 (P2) = 80 psig ... and increase mid-point pressure to backflush any sample material left in column.

Page 238: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Prevent

7-78

Page 239: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals 8

Page 240: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-2

Page 241: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-3

Introduction

Programmed Pneumatic Control (PPC) is the electronic control of pressures and flows for inlet, detector, and auxiliary gases. The PPC control modules regulate pressures and flows using electronically driven variable flow restrictors. The control modules also contain pressure and flow transducers to provide feedback for complete monitoring. A PPC controller board drives the variable restrictors on the control modules by comparing actual pressures and flows with setpoints determined from user-entered values.

This chapter presents an overview of the concepts and use of Programmable Pneumatic Control (PPC) of the Clarus 600 GC. In particular, it describes the following microprocessor-controlled pneumatic functions that comprise PPC:

• Carrier-gas flow, pressure, and linear-velocity programming

• Split-flow and split-ratio control

• Detector-gas flow control

• Auxiliary-gas pressure and flow control

The following unique pneumatic devices make these control functions possible:

• Carrier-gas mass-flow controller

• Split-pneumatic controller

• Detector-gas flow controller

• Auxiliary pressure controller

Pressure-readout module

You can combine one or more of these modules to create the control functions required for a specific pneumatic configuration in the GC.

Page 242: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-4

A dedicated microprocessor control system reads all operating parameters from — and sends control signals to — the pneumatic devices. This PPC controller receives pneumatic setpoints, the ambient temperature, pressure, the oven temperature, and GC status information from the main Clarus 600 GC processor. The PPC controller sends back the pneumatic operating parameters and PPC status information to the main GC processor.

The PPC controller accommodates up to eight flow controllers and/or twelve pressure controllers, for a maximum of twelve installed PPC devices.

The main Clarus 600 GC processor accepts and stores method setpoints and timed events for the configured PPC devices. In addition, the main Clarus 600 GC processor contains PPC functions that include calibration and setup utilities, diagnostics, and fault-condition monitoring.

Page 243: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-5

Carrier Gas Control

The type of primary PPC carrier gas modules and their operation depends on the inlet (injector) option to which they are connected. For example, the packed-column inlet and the programmed-temperature on-column inlet (POC) use a single carrier gas mass-flow controller plus a pressure-readout module. The split/splitless (capillary) and programmed-temperature split/splitless (PSS) inlets use a combination of a carrier gas mass-flow controller and a split-pneumatic pressure controller. Other carrier-gas controllers may use auxiliary PPC zones (see Auxiliary Carrier Gas Control).

Packed Column and Programmed-Temperature On-Column Inlets

The PPC carrier gas mass-flow controller delivers a constant flow rate into a column. The controller operates by increasing or decreasing the column head pressure to maintain a constant mass-flow as the column temperature changes, or if you use different columns. Its internal operation is similar to a conventional mass-flow controller, but instead of utilizing a spring-and-diaphragm mechanism for control, the PPC controller reads and sets the flow rate electronically. Unlike a conventional controller, it constantly monitors the ambient room pressure, the tank pressure, and the temperature of each flow module, and it compensates as required to maintain a constant mass-flow output. The schematic in the following figure illustrates the PPC mass-flow controller configuration for packed column and POC inlets.

Page 244: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-6

Figure 8-1 Schematic of packed column and POC pneumatics

Carrier Gas Mass-Flow Controller Operating Range

The carrier gas mass-flow controller for packed-column inlets has a nominal flow range of 0 – 30 mL/min for helium carrier gas at 90 psig tank pressure. The controller for the POC inlet has a range of 0 – 10 mL/min helium for capillary column applications. The two controllers differ only in the installed flow-range element (restrictor). By changing the flow-range element, you can use the same controller in different GC configurations. The adapter kit for packed-column installation of 0.53-mm i.d. capillary columns, for example, contains the appropriate 0 – 10 mL/min element. After changing the flow-range element, you must always recalibrate the flow controller.

Mass-Flow Controller Setup

The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery backed-up random access memory (BRAM). This information is written into the BRAM during manufacturing. Normally, there is no need to access or modify the type or location of the PPC controllers. If the PPC hardware configuration is changed — by installing or removing PPC devices — the operator must access the PPC connection menus on the GC touch screen and modify the type and location of the PPC devices.

Page 245: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-7

If the installed hardware does not match the stored configuration, the Clarus 600 GC issues a warning message.

CAUTION

Do not modify the PPC connection information if the PPC hardware configuration has not changed.

Mass-Flow Controller Calibration

The carrier gas mass-flow controller is factory calibrated to a National Institute of Standards and Technology (NIST) traceable standard with helium carrier gas; however, it requires periodic calibration. Recalibration is required after changing the flow-range element; otherwise the calibration frequency is up to the individual operator. The controller accommodates changing the carrier gas (to other than helium) by approximating a new set of calibration parameters based on the new gas. The approximation is good but controller recalibration is recommended for the best accuracy. It is good laboratory practice to recalibrate all measuring and control devices periodically.

The mass-flow controller calibration procedure consists of specifying the reference temperature and pressure; zeroing the internal sensors in the flow controller with the tank pressure off; specifying the correct carrier gas type and installing the correct flow-range element; measuring the background flow rate at zero-flow with the tank pressure on; and measuring the flow rate at 60% of full-scale flow. This procedure is built into the GC firmware.

Once the PPC is set up and calibrated you must configure it

Carrier Gas Mass-Flow Controller Configuration

Once the carrier gas mass-flow controller has been set up and calibrated for the PPC system, you must configure its operating mode on the GC keypad. For packed-column and POC inlets the selections are Zero, Flow, or Pressure.

1. From the System Status Screen touch the Tools button.

Page 246: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-8

2. From the tools drop down menu select Configuration.

3. In the Configuration touch screen touch the Pneumatics button .

Page 247: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-9

4. In the Configuration PPC touch screen touch PPC Configure bar.

Page 248: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-10

5. A popup warning appears, select OK.

6. On the Configure PPC Devices screen touch the Zero button then touch OK to set the gas flow to zero.

Page 249: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-11

Headspace Analysis

Headspace analysis is the analysis of the vapor lying in equilibrium over a solid/liquid sample in a sealed vial.

For practical headspace analysis the sample is sealed in a vapor tight vial, placed in an thermostatted oven and heated to a pre-determined temperature.

The sample vial contains the volatile material in equilibrium between the solid/liquid sample and the vapor lying over it.

After equilibrium is reached between solid/liquid phase and the vapor phase, a defined amount of the vapor is taken and carried to the column in the gas chromatograph for analysis. With this technique only highly volatile substances reach the column, the non-volatile substances remain in the sample vial.

Using this technique, samples containing constituents which are unsuitable for injection with a syringe can be analyzed (e.g. polymers, highly viscous liquids).

Suitable fields of application are in the analysis of polymers, certifying of the volatile components in drinks and foodstuffs, blood alcohol levels, water and environmental analysis.

For information on vacuum see the section on Vacuum Compensation later in this chapter.

Capillary and Programmed Split/Splitless Inlets A packed-column or POC inlet requires only a source of flow-controlled carrier gas; however, split-type inlet systems are more complex. The capillary and programmed-temperature split/splitless (PSS) inlets have identical split pneumatics. The split pneumatics occupy two PPC control zones, one for column inlet pressure control and one for split flow rate control.

Split Pneumatic Setup The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery-backed-up random access memory (BRAM). This information is written into the BRAM during manufacturing. Normally, there is

Page 250: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-12

no need to access or modify the type or location of the PPC controllers. If the PPC hardware configuration is changed — by installing or removing PPC devices — you must access the PPC connection menus on the GC keypad and modify the type and location of the PPC devices. If the installed hardware does not match the stored configuration, the GC issues a warning message.

CAUTION

Do not modify the PPC setup information if the PPC hardware configuration has not changed.

Split Pneumatic Control A split system such as the capillary or PSS inlet pneumatics controls three parameters: the split flow, the inlet pressure, and the septum purge flow. See the following figure that shows two diagrams showing the forward pressure mode (splitless control) and the backpressure mode (split control) of the PPC split pneumatics in the Clarus 600 GC.

Page 251: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-13

Septum Purge

Charcoal Trap

SolenoidValve

(Closed)

PressureRegulator Restrictor

Δp

ControllerTemperatureΔp

Δp

Ambient Pressure

Δp

FlowMonitor

Pressure Control

Forward Pressure Mode (Splitless Control)

Septum Purge

Ambient Pressure

Pressure Control

Total Split Flow Control

Charcoal Trap

SolenoidValve

(Open)

PressureRegulator Restrictor

Δp

ControllerTemperatureΔp

Δp

Δp

Backpressure Mode (Split Control) Figure 8-2 Schematic of the split PPC pneumatics

Page 252: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-14

Split Flow ─ A mass-flow controller supplies carrier gas to the split or PSS inlet system. This is the same mass-flow controller used for packed or POC inlets, but here it is used with a 0 – 300 mL/min flow-range element (fixed restrictor). Carrier gas passes through the flow controller and into the inlet (injector) at the connection labeled Carrier In. Some of the gas flows across the septum and out the septum purge, while the rest flows down through the inlet liner. At the bottom of the liner, a fraction of the carrier gas enters the column, and the balance flows back up the outside of the liner and out the split vent. An electrical solenoid valve provides positive split flow shutoff when required.

Inlet Pressure ─ A pressure transducer connected at the split vent (before the charcoal trap) measures the inlet pressure at the head of the column. The inlet pressure may range from 0 – 100 psig.

Septum Purge ─ A constant-flow device (fixed pressure regulator) regulates the septum purge flow at approximately 3 mL/min, independently of the inlet pressure. The regulator is preset in the factory; it is not adjustable.

Split Pneumatic Calibration

The split flow controller requires periodic calibration. The controller is factory calibrated to a National Institute of Standards and Technology (NIST) traceable standard with helium carrier gas. Recalibration is required upon changing the flow-range element; otherwise, the calibration frequency is up to the individual operator. The PPC controller accommodates changing the carrier gas to other than helium by approximating a new set of calibration parameters based on the new gas. The approximation is good but controller recalibration is recommended for the best accuracy. It is good laboratory practice to recalibrate all measuring and control devices periodically.

The calibration procedure consists of specifying the reference temperature and pressure; zeroing the internal sensors in the flow controller with the tank pressure off; specifying the flow-range element and carrier gas type; measuring the background flow rate at zero-flow with the tank pressure on; and measuring the flow rate at 60% of full-scale flow. This procedure is built into the GC firmware.

Page 253: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-15

Split Pneumatic Carrier Gas Configuration

Once the split pneumatic system has been set up and calibrated on the PPC system you must configure its operating mode by using the GC touch screen. There are configuration choices for column carrier-gas control and split flow or ratio control.

The PPC controller includes a mathematical model of a capillary column. This model is enabled by setting Capillary Control Mode ON (see Capillary Control Mode, below). The PPC controller uses this model to calculate the column exit flow at ambient pressure and temperature, the column average linear gas velocity, and the inlet split ratio in real time. You may choose to control the column flow or velocity, in which case the PPC controller calculates the pressure required to maintain the flow or velocity setpoint. You may also choose to control the split ratio; in this case the PPC controller calculates the split flow rate required to obtain the setpoint split ratio.

The PPC controller microprocessor carries out the method setpoint program for both carrier-gas channels by taking the setpoint pressure, flow, or velocity value; the column dimensions; the carrier-gas type; the reference temperature and pressure; the oven actual temperature; and the ambient pressure. It calculates the other two parameters in real time. For pressure control, it calculates the flow and velocity; for flow control, it calculates the pressure and velocity; for velocity control, it calculates the pressure and flow. These calculations track ambient pressure changes, changing oven temperatures, and the execution of the multi-ramp carrier-gas program. In each case, the actual pressure in the inlet is controlled by the calculated or method pressure setpoint.

For more information on these internal calculations, see the section on “Theory of Capillary Column Control,” in this chapter.

Capillary Control Mode

The first choice for split carrier-gas control is the Capillary Control Mode. This parameter may be set On or Off. Each carrier gas channel may be setup with or without capillary control.

Off ─ uses direct pressure control of the column through a multi-step pressure program and carrier gas timed events. You can directly control the split flow

Page 254: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-16

control through the initial setpoint and timed-event control of the split flow controller. Split ratio control, capillary column flow, and capillary column velocity control and information are not available. This mode of operation does not require knowledge of the capillary column dimensions.

On ─ enables indirect control of capillary column flow or velocity in addition to direct pressure control, each through a multi-step program and timed events. This mode also enables direct split flow or indirect split ratio control of the split flow controller. The display reports actual pressure, flow, velocity, split flow, and split ratio. In this mode you must enter accurate column length and diameter information (except for pressure-controlled operation with oven tracking).

Carrier Gas Configuration without Capillary Column Control

When capillary control is turned off for a split PPC controller, the choices for carrier-gas control are None and Pressure.

None ─ removes the split pneumatics from the GC method and turns off the split flow and pressure.

Pressure ─ enables direct pressure control of the split inlet. The PPC controller uses a 5 psig initial setpoint with an infinite hold time when Pressure is first selected. The user may add up to three additional setpoints, ramps, and hold times. This mode also enables direct split flow control (see below).

Carrier Gas Configuration with Capillary Column Control ON

In this mode the choices for carrier-gas control are None, Pressure, Flow, and Velocity. With pressure control you may select either Pressure: Program or Pressure: Oventrack.

None ─ removes the split pneumatics from the GC method and turns off the split flow and pressure.

Pressure: Program ─ enables direct pressure control of the column pressure. The split carrier-gas pressure program mode gives a multi-ramp program with real-time calculation and display of flow and velocity, plus flow control or ratio control of the split flow. The PPC controller uses a 5 psig initial setpoint with an infinite hold time when Pressure: Program is first selected. The pressure range

Page 255: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-17

is 0 – 100 psig. You may add up to three additional setpoints, ramps, and hold times.

Pressure: Oventrack ─ also enables direct carrier-gas pressure control. The PPC controller automatically varies the initial inlet pressure so that the capillary column flow rate remains constant as the oven temperature changes away from its initial setpoint value. The PPC controller uses a 5 psig initial setpoint when Pressure: Oventrack is first selected. Multi-ramp programming is not available. This mode allows only direct flow control of split flow; however, this mode can be used for constant column flow control when you do not know your column dimensions.

Flow ─ enables indirect capillary column mass-flow control. The PPC controller calculates the inlet pressure required to maintain the method flow program setpoints from the column dimensions, carrier-gas type, oven temperature, and ambient pressure. The display shows the flow, pressure, and calculated average linear gas velocity. The PPC controller uses a 2 mL/min initial setpoint with an infinite hold time when Flow is first selected. The column flow has a total flow range of 0 – 500 mL/min (including the split flow). You may add up to three additional setpoints, ramps, and hold times. This mode enables flow control or split ratio control of the split flow.

Velocity ─ enables indirect capillary column average linear gas velocity control. The PPC controller calculates the inlet pressure required to maintain the method velocity program setpoints from the column dimensions, carrier-gas type, oven temperature, and ambient pressure. The display shows the velocity, pressure, and calculated flow rate. The PPC controller uses a 30 cm/s initial setpoint with an infinite hold time when Velocity is first selected. You may add up to three additional setpoints, ramps, and hold times. The carrier-gas velocity has a range of 0–200 cm/s. This mode enables flow control or split ratio control of the split flow.

Column Dimensions

After selecting any of the capillary control modes (except Pressure: Oventrack) the GC prompts you for the column dimensions. These parameters control the relationships between the pressure, flow, and velocity as calculated by the PPC controller. If the column dimensions are inaccurate, the PPC controller will

Page 256: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-18

produce incorrect readings and will incorrectly control the column flow, velocity, and split ratio.

Column Length

The column length defaults to 25.0 meters. It has a range from 1.0 – 200.0 meters. Small errors in the column length affect the average linear velocity and the flow rate only slightly. See the section “PPC Tips and Techniques” at the end of this chapter for procedures to correct errors in column length.

Column Inner Diameter

The column inner diameter defaults to 250 μm. It has a range from 50 –1000 μm. Small errors in the column inner diameter strongly affect the average linear velocity and the flow rate.

Vacuum Compensation

The next Clarus GC split pneumatic configuration parameter in the capillary control mode is vacuum compensation. The PPC controller uses a mathematical model of a capillary column to determine pressure – flow – velocity relationships. The column outlet pressure plays an important role in these calculations. The PPC controller continually monitors the barometric pressure because normal daily atmospheric pressure fluctuations affect the column average linear gas velocity—and thus retention times—if the controlling system does not compensate. Vacuum compensation controls the role of the outlet pressure in the PPC controller. Vacuum compensation may be set On or Off.

Off ─ causes the PPC controller to assume that the column outlet is at atmospheric pressure.

On ─ causes the PPC controller to assume that the column outlet is at vacuum. This setting is appropriate for GC-MS systems with a direct interface where the column enters the mass-spectrometer source. Vacuum compensation should be turned off when an open-split interface is used since it keeps the GC column outlet close to atmospheric pressure.

Page 257: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-19

Split Control Mode Configuration

The last Clarus GC carrier-gas configuration parameter is the split control mode. The ratio of the total flow through the inlet liner to the column flow rate — the fraction of sample vapor entering the column, or the split ratio — controls the quantitative transfer of sample into the column. For split ratio control the PPC system must know both the column flow rate and the total gas flow through the inlet liner.

When capillary control is off, the PPC system does not compute the capillary column flow rate or the split ratio; only split flow control is possible. With capillary control on (except for Pressure: Oventrack mode) the PPC controller calculates the column flow rate and the split ratio.

The choices for split-flow control are Flow or Ratio. There is no choice for None, since the split pneumatics will not operate without incoming carrier-gas flow. Setting the carrier-gas configuration to None automatically sets the split-flow controller configuration to none.

Flow ─ enables direct split flow control of the carrier gas mass-flow controller. You enter a split vent flow rate in the method. The PPC controller adds an optional offset to obtain the total split flow (see “Split Flow Offset” below) and sets that amount at the flow controller. The GC also displays the calculated split ratio if capillary control is on.

Ratio ─ enables indirect split ratio control of the carrier gas mass-flow controller. The user enters a split ratio in the method. The PPC controller calculates the required split vent flow from the column flow rate and adds an optional offset to obtain the total split flow (see “Split Flow Offset” below) and sets that amount at the flow controller. The GC also displays the calculated split flow rate defined by the following equation:

total split flow = split flow + septum purge flow + column flow

NOTE: Any carrier system leaks will add to the observed total split flow.

Page 258: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-20

Split Flow Offset

The PPC controller sets and monitors the total split flow through the inlet system. The split ratio is based on the flow out the split vent. Since the septum purge flow is taken from the total split flow rate before passing through the inlet liner, across the column entrance, and out the split vent, split ratio calculations must account for the flow reduction. The Clarus 600 GC includes a split flow offset parameter that is accessed from the CONFIG CARRIER menus when a split controller is setup. This parameter may be set to Fixed or Auto.

Fixed ─ prompts the user to enter the septum purge offset value. This offset is the septum purge flow plus the column flow. The initial column flow is displayed on the [Carrier Prog] key when the GC is READY at its initial setpoints with capillary control ON. The PPC controller adds this offset to the method setpoint split flow rate to obtain the actual flow delivered to the split inlet. The septum purge flow is factory set at 3 mL/min. To measure the actual septum purge offset, set the carrier pressure to 5 psig or greater and turn off the split flow (set to 0). The remaining split vent flow displayed is the septum purge flow, plus the column flow. The actual septum purge flow can be measured from the split vent bulkhead fitting.

Auto ─ measures the offset flow rate before each run by momentarily setting the split vent flow to 0, waiting for the remaining flow to stabilize, and then recording its magnitude. During this time the GC status is PRE-RUN. This operation occurs after equilibration, and before executing any Pre-Run timed events (events with negative times). The Auto measurement includes both column and septum purge flows.

NOTE: The split flow screen (shown below) displays the flow from the injector vent as the setpoint and the total split system flow as the actual values.

The value in the upper right carrier is the total split flow. The value in the lower right corner is the split vent setpoint.

The flow measured at the split vent fitting on top of the GC is the vent flow plus the septum purge flow.

Page 259: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-21

Split Pneumatic Operation

Once the split system has been configured, you may modify the default setpoints in the active GC method, edit the setpoints in stored methods, or generate a new method and enter split pneumatic setpoints. The following parameters control split pneumatic operation.

Initial Carrier Setpoint

The PPC controller takes setpoints of flow, velocity or pressure depending on the configured split pneumatic carrier gas mode. It receives its initial setpoint from the active GC method. The initial setpoint is established upon setting up a method, ending a run, or otherwise resetting the instrument. The PPC controller holds the setpoint constant at the initial value if no additional carrier-gas entries are made in the GC method.

Carrier Gas Program

The GC method accommodates up to three additional carrier-gas program steps consisting of a hold time, ramp rate, and plateau setpoint. Sequential setpoints may be greater than or less than the preceding setpoint; the PPC controller will program the split pneumatics with a positive or negative ramp as required. The carrier gas resets to the initial setpoint value in the method at the end of the GC oven program, even if a carrier-gas program is in progress. The carrier gas also resets if the end of the carrier-gas program occurs before the end of the oven program.

Carrier Gas Timed Events

You may also program carrier-gas timed events to produce step changes in the setpoint. Carrier-gas timed events can occur before or after a run starts. A carrier-gas timed event supersedes the method setpoint value at the time the event occurs. However, if the carrier-gas program in the method advances to a new step after a carrier-gas event, the new program step or ramp takes control.

Split Flow or Ratio Setpoint

The GC method includes an initial split flow or ratio setpoint. This setpoint is established upon setting up a method, ending a run, or otherwise resetting the

Page 260: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-22

instrument. The PPC controller holds the setpoint constant at the initial value if no additional carrier-gas entries are made in the GC method.

Split Flow or Ratio Timed Events

The split flow or ratio is changed to a new setpoint by a split timed event. Split flow or ratio events can occur before or after a run starts. A split flow or ratio event supersedes the method setpoint value at the time the event occurs.

Theory of Capillary Column Control

This section gives an overview of the theory behind PPC. For more detailed information please refer to Reference [1].

The Clarus 600 GC Programmable Pneumatic Control system can program capillary column pressure, flow, or velocity. It accomplishes this by calculating and maintaining the column pressure drop required by the carrier-gas program and column temperature. The PPC controller then reads the actual pressure and uses that value to calculate the actual column flow rate and average linear gas velocity. The column dimensions, ambient pressure, and carrier-gas type also enter into the equations.

Page 261: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-23

Column Temperature Effects

Figure 8-3 Carrier-gas viscosity–temperature relationships1

The viscosity of the carrier gas changes with the column temperature. As the column temperature increases, so does the carrier-gas viscosity. The increasing viscosity raises the resistance of the carrier gas to being pushed through the column. Thus, higher column pressure drops are required to maintain column flow or average linear velocity at higher temperatures. Conversely, keeping the pressure drop across the column constant while increasing the column temperature causes the flow rate and linear velocity to drop off.

Figure 8-3 illustrates the effect of column temperature on carrier-gas viscosity. Notice that the viscosity is different for each carrier gas, and that the “slopes” of the curves are also different in each case. For each carrier gas, the Clarus 600 GC includes an accurate model of this viscosity-temperature relationship as represented by the lines in Figure 8-3.

1 J. V. Hinshaw and L.S. Ettre, “Introduction to Open-Tubular Column Gas

Chromatography,” (Advanstar Communications, 1994), p. 25. (P/N N930-6007).

Page 262: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-24

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

-100 -50 0 50 100 150 200 250 300 350 400 450

Column Temperature (°C)

Flow

(mL/

min

)

(a)

0

5

10

15

20

25

30

35

40

45

50

-100 -50 0 50 100 150 200 250 300 350 400 450

Column Temperature (°C)

Velo

city

(cm

/s)

(b)

Figure 8-4 Effect of column temperature on (a) flow and (b) velocity Column: 25-m x 250-μm i.d. Helium carrier at 12 psig constant pressure drop.

The viscosity changes caused by increasing temperature affect the flow and velocity through a column with a constant pressure drop. As shown in Figure 8-4, both flow and velocity decrease across the column temperature range. This decreasing flow and velocity can cause loss of column efficiency as well as

Page 263: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-25

changes in flow-sensitive detector operation during programmed-temperature analyses. In this case, the column pressure drop was selected so that the average linear velocity was close to 30 cm/s at 50 °C; at 350 °C the velocity drops to around 20 cm/s. The effect on flow is more severe, dropping from 1.19 mL/min at 50 °C to 0.4 mL/min at 350 °C.

Without PPC, compensation for this effect is difficult and limited. Although it is possible to program the PPC controller pressure to crudely compensate, it is much easier and more accurate to use either the flow-programmed or velocity-programmed modes instead.

Flow Programmed Operation

The PPC controller operates in a pressure-, flow-, or velocity-programmed mode. The flow-programmed mode is useful in applications where the column flow rate must be held constant, or where the operator wishes to express the column carrier program in flow terms. Figure 8-5 shows the pressure program and velocity that result when a 50-m x 530 μm i.d. column operates with nitrogen carrier at a constant 4 mL/min flow across a range of temperatures.

Page 264: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-26

0

2

4

6

8

10

12

14

-100 -50 0 50 100 150 200 250 300 350 400 450

Column Temperature (°C)

Pres

sure

dro

p (p

sig)

(a)

0

5

10

15

20

25

30

35

40

45

50

-100 -50 0 50 100 150 200 250 300 350 400 450

Column Temperature (°C)

Velo

city

(cm

/s)

(b)

Figure 8-5 Effect of column temperature on (a) pressure and (b) velocity. Column: 50-m x 530-μm i.d. Nitrogen carrier at 4 mL/min constant mass flow

Velocity Programmed Operation

Programming the PPC controller in average linear gas velocity mode adjusts the column pressure drop to maintain the setpoint velocity as the oven temperature

Page 265: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-27

and ambient pressure change. This helps ensure better repeatability of chromatograms from one instrument to another — assuming the ovens are calibrated to a common temperature reference standard—by eliminating effects due to differences in altitude or weather. Differences in the columns, however, are not accounted for.

0

5

10

15

20

25

30

35

40

-100 -50 0 50 100 150 200 250 300 350 400 450

Column Temperature (°C)

Pres

sure

dro

p (p

sig)

(a) (b)

Figure 8-6 Effect of altitude on pressure programs. Column: 15-m x 100-μm i.d. Hydrogen carrier at 30 cm/s constant average linear gas velocity. (a): Sea level; (b): 3000 m altitude

Figure 8-6 shows two pressure-temperature curves, one for a constant velocity of 30 cm/s on a 15-m x 100 μm i.d column with hydrogen carrier with an atmospheric pressure equivalent to sea level, and the other for the same column at the equivalent of 3000 m altitude. When operating the oven at a constant temperature with identical columns, all solute retention times will be the same in both cases because the velocities are unchanged.

With oven temperature programming, the retention times will be very similar but not quite the same due to slight differences in the velocity profiles across the two columns at different altitudes.

Page 266: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-28

Pressure / Oventrack Mode

0

10

20

30

40

50

60

-100 -50 0 50 100 150 200 250 300 350 400 450

Column Temperature (°C)

Pres

sure

dro

p (p

sig)

Figure 8-7 Pressure program in oventrack mode. Helium carrier at 25 psig, 50 °C initial conditions

The Pressure: Oventrack operating mode maintains a constant column mass flow as the column temperature changes by adjusting the inlet pressure as required. No knowledge of column dimensions is required in this mode: only the initial pressure and temperature are specified. Figure 8-7 shows the pressure-temperature profile for Pressure: Oventrack mode with an initial pressure of 25 psig at 50 °C with helium carrier gas. This operating mode is convenient for mass-spectrometric or thermal-conductivity detectors where column dimensions are not important.

Page 267: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-29

Vacuum Compensation

0

5

10

15

20

25

30

35

40

45

-100 -50 0 50 100 150 200 250 300 350 400 450

Column Temperature (°C)

Pres

sure

dro

p (p

sig)

(a) (b)

Figure 8-8 Pressure program (a) without and (b) with vacuum compensation. Column: 60-m x 320 μm i.d. Helium carrier at 2 mL/min constant flow

When using a mass-spectrometric detector with the column exit under vacuum, it is necessary to reduce the column pressure drop to account for the reduced exit pressure. The effect of putting the column exit under vacuum is shown in Figure 8-8 for a 60-m x 320 μm i.d. column with helium carrier at a 2 mL/min constant flow rate.

Page 268: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-30

Detector Gas Flow Control

The Clarus 600 GC can control up to two detectors, each with up to two primary detector gas flow controllers. The GC determines the type of controllers required for each detector by sensing the presence of the detector amplifier board. Certain detectors, such as the ElCD/PID combination, use additional auxiliary detector gas control zones (see “Auxiliary Detector Gas Control,” below).

The Clarus 600 GC PPC detector gas flow controller delivers a set flow rate of gas into a device (detector) at atmospheric pressure. Unlike the carrier gas controller, the detector gas controller does not compensate for significant changes in its outlet pressure. It is not suitable for providing column flow. The detector gas controller is not affected by normal changes in the incoming gas supply pressure. The PPC controller compensates for changes in each detector gas controller’s internal temperature. Figure 8-9 illustrates the detector flow control module.

Figure 8-9 Schematic of a detector gas flow control module

Page 269: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-31

Detector Gas Flow Controller Operating Range

The detector gas flow controller operating range depends on the detector in use. Each flow controller uses a flow-range element (fixed restrictor) that determines the available flow range for a particular gas. The nominal detector gas flow lies within 20–50% of the full-scale controller range for the factory-installed flow-range elements. The flow-range elements should not be changed unless the controller is to be used with a different detector. Please refer to Chapter 9, “Maintenance,” the Practical Hints section for information on individual flow-range elements for specific detectors.

NOTE: Only the fixed restrictor changes for various requirements.

Detector Gas Flow Controller Setup

The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery-backed-up random access memory (BRAM). This information is written into the BRAM during manufacturing; normally there is no need to access or modify the type or location of the PPC controllers. If the PPC hardware configuration is changed — by installing or removing PPC devices — you must access the PPC connection menus on the GC keypad and modify the type and location of the PPC devices. See “Connecting the PPC Modules” in Chapter 13, “System Utilities,” for detailed instructions.

CAUTION

Do not modify the PPC connection information if the PPC hardware configuration has not changed.

Detector Gas Flow Controller Calibration

The detector gas mass-flow controller requires periodic calibration and the calibration is gas specific. The controller is factory calibrated to a National Institutes of Standards Technology (NIST) traceable standard with air, hydrogen, or helium gas, depending on the detector type. Recalibration is required upon

Page 270: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-32

changing the gas or the flow-range element; otherwise, the calibration frequency is up to the individual operator. It is good laboratory practice to recalibrate all measuring and control devices periodically.

The detector gas flow controller calibration procedure consists of specifying the reference temperature and pressure; zeroing the internal sensors in the flow controller with the tank pressure off; specifying the flow-range element; selecting the gas type (for reference or makeup only); then measuring the flow rate at 20% and 40% of full-scale flow. This procedure is built into the Clarus 600 GC firmware. See Chapter 13, “System Utilities,” for detailed instructions.

Detector Gas Mass-Flow Controller Operation

For flame-type detectors, setting up the detector gas flow controllers sets a default flow rate equal to the nominal flow specified for the detectors. For detectors with makeup or reference gases the flow rate defaults to 30 mL/min. Once the detector gas flow controllers have been set up and calibrated on the PPC system, the operator must enter their flow rates on the GC touch screen.

Unlike carrier gas flows, detector gas flow rates do not appear in the GC method. Instead, they are set in the GC configuration. There are no timed events that modify detector flows. Detector gas flow controllers may be set up as Auxiliary gas flow zones if it is necessary to incorporate them into the GC method or to modify them with timed events. See “Auxiliary Detector Gas Control” below for further information.

Flame Gases

The Clarus 600 GC automatically determines gas types for flame detectors, including the flame-ionization (FID), flame-photometric (FPD), and nitrogen-phosphorous (NPD).

Once the flame gas controllers are set up, you may modify the default flow rates as required by entering new values under the in the GC configuration.

Flame Ignition

The PPC version of the FID has flameout detection and an auto-ignite feature. Flameout detection occurs 0.5 minutes into the PRE-RUN time and it measures

Page 271: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-33

the user-entered baseline threshold level in mV. If the threshold level is exceeded, the flame is considered lit; otherwise, the GC detects that the flame is out, and auto-ignite automatically lights the flame before the start of a run. You can also manually light the flame at any time by pressing the ignite button on the touch screen.

Makeup and Reference Gases

The Clarus 600 GC automatically determines the presence of reference and makeup gases for detectors. The thermal-conductivity detector (TCD) uses reference gas. The electron-capture detector (ECD), photoionization detector (PID), and, optionally, the thermal-conductivity detector (TCD) use makeup gas. Makeup and reference gases operate in a fashion similar to flame gases, except that you may change the gas type in the PPC controller setup.

Once the makeup or reference gas controllers are set up, you may modify the default flow rates as required by entering new values in the active.

Detector makeup gas flow controllers may be set up as Auxiliary gas flow zones if it is necessary to incorporate them into the GC method or to modify them with timed events. See “Auxiliary Detector Gas Control” below for further information.

Page 272: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-34

Auxiliary Pressure and Flow Control

In addition to carrier and detector gas control, the Clarus 600 GC accommodates up to four auxiliary PPC controllers. The auxiliary zones can control carrier gas mass-flow controllers, pressure controllers, pressure readouts, and detector gas flow controllers. These zones all have an initial setpoint, and incorporate timed-event changes.

Auxiliary Carrier Gas Control

The Clarus 600 GC controls up to two primary carrier-gas zones and up to four primary detector gas zones. In addition, up to four carrier gas mass flow controllers, pressure controllers, or pressure readouts may be set up as auxiliary PPC carrier zones, subject to a maximum of eight flow controllers (carrier plus detector flow) and twelve zones total in the PPC system. The auxiliary flow control zones do not accommodate split pneumatics, which must be configured as the primary carrier control zones on the GC.

Auxiliary Carrier Gas Setup

The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery-backed-up random access memory (BRAM). This information is written into the BRAM during manufacturing; normally there is no need to access or modify the type or location of the PPC controllers. Some factory installed options use auxiliary carrier-gas zones. In order to use other auxiliary zones the operator must access the PPC connection menus on the GC touch screen and modify the type and location of the PPC devices. For auxiliary carrier-gas control you may choose either a carrier gas mass-flow controller, a pressure controller, or a pressure readout.

CAUTION

Do not modify the PPC connection information if the PPC hardware configuration has not changed.

Page 273: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-35

Auxiliary Carrier Gas Mass-Flow Controller Calibration

The calibration procedure for mass-flow controllers designated as auxiliary carrier gas zones is the same as for controllers designated as primary carrier zones. Please refer to the section “Mass-Flow Controller Calibration” in this chapter.

Auxiliary Carrier Gas Zone Operation

Once the auxiliary carrier gas zone has been set up, you may modify the default setpoints in the active GC method, edit the carrier-gas setpoints in stored methods, or generate a new method and enter carrier-gas setpoints. The following parameters control the auxiliary carrier gas.

Initial Auxiliary Carrier Gas Setpoint

The auxiliary carrier gas zone takes setpoints of flow or pressure depending on the configured controller. It receives its initial setpoint from the active GC method. The initial setpoint is established upon setting up a method, ending a run, or resetting the instrument. The setpoint is held constant at the initial value if no additional auxiliary carrier entries are made in the GC method. The initial auxiliary carrier gas setpoint appears after touching Carrier Program button in the GC method.

Auxiliary Carrier Gas Timed Events

You may program auxiliary carrier-gas timed events to produce step changes in the setpoint. Auxiliary carrier-gas events can occur before or after a run starts. An auxiliary carrier-gas event supersedes the method setpoint value at the time the event occurs.

Auxiliary Detector Gas Control

The Clarus 600 GC controls up to four primary detector gas zones and up to two primary carrier-gas zones. In addition, up to four detector gas flow controllers may be set up as auxiliary PPC detector zones, subject to a maximum of eight flow controllers (carrier plus detector flow) and twelve zones total in the PPC system.

Page 274: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-36

Auxiliary Detector Gas Setup

The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery-backed-up random access memory (BRAM). This information is written into the BRAM during manufacturing: normally there is no need to access or modify the type or location of the PPC controllers. Some factory installed detectors use auxiliary detector gas zones. In order to use other auxiliary zones the operator must access the PPC CONNECTION menus on the GC touch screen and modify the type and location of the PPC devices.

CAUTION

Do not modify the PPC connection information if the PPC hardware configuration has not changed.

Auxiliary Detector Gas Flow Controller Calibration

The calibration procedure for detector gas flow controllers designated as auxiliary detector gas zones is the same as for controllers designated as primary detector zones. Please refer to the section “Detector Gas Flow Controller Calibration” in this chapter.

Auxiliary Detector Gas Zone Operation

Once the auxiliary detector gas zone has been set up, you may modify the default setpoints in the active GC method, edit the detector gas setpoints in stored methods, or generate a new method and enter detector gas setpoints. The following parameters control the auxiliary detector gas.

Initial Auxiliary Detector Gas Setpoint

The auxiliary detector gas zone takes flow setpoints. It receives its initial setpoint from the active GC method. The initial setpoint is established upon setting up a method, ending a run, or resetting the instrument. The setpoint is held constant at the initial value if no additional auxiliary detector gas entries are made in the GC method.

Page 275: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-37

Auxiliary Detector Gas Timed Events

You may program auxiliary detector gas timed events to produce step changes in the setpoint. Auxiliary detector gas events can occur before or after a run starts. An auxiliary detector gas event supersedes the method setpoint value at the time the event occurs.

Page 276: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-38

PPC Tips and Techniques

This section contains tips and techniques that you will find useful when using PPC.

Effect of Flow on Pressurization Rate

When first turned on, when a higher setpoint pressure is programmed, the PPC carrier-gas zones require some time to come up to pressure. The pressurization rate is a function of the type of inlet, the incoming carrier-gas flow rate and the outgoing column and septum purge flow rate (if applicable).

For typical packed columns the inlet will pressurize within one minute. For a POC inlet where the column flow rate is low it may take longer to reach a steady-state inlet pressure. The PPC gas-leak monitor will not trigger until five minutes have elapsed, which should be enough time in all cases.

For capillary or PSS inlets, setting a high inlet pressure and a low split flow rate (but not a zero flow) can result in the inlet never coming to pressure because the incoming split flow is less than the column plus septum purge rates. A zero split-flow rate is treated as a special condition for splitless injection and will not cause this problem. In general, use a minimum split flow rate that is the greater of 20 mL/min or four times the sum of the column and septum-purge flows.

NOTE: If the column pressurization rate (psig/min) is too low, increase the split flow rate during this time. If you want to quickly pressurize the column in the split mode, temporarily increase the split vent flow while pressurizing the column. You can set the split vent flow up to a maximum of 500 mL/min.

Correcting the Column Dimensions

The PPC controller relies on operator input of column dimensions to calculate pressure – flow and pressure – velocity relationships. If the column dimensions are incorrect, then the calculations will be inaccurate. In general, actual column diameters are within ±2 μm of the nominal specified value. Column lengths may vary for used columns because the ends may have been cut off.

Page 277: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-39

Alternatively, the column dimensions may be unknown. In such cases set the Carrier mode to pressure and oventrack, if you need to maintain a constant column flow rate.

Correcting the Column Length

The average linear gas velocity is the most accurate measure of capillary column conditions, because the low flow rates are difficult to measure. By measuring the average velocity you can compare it to the displayed velocity on the GC, and then make a correction to the column length. This procedure assumes that the column inner diameter entered into the GC is correct.

Set Up the Clarus 600 GC

Install a column and turn on the GC. In Configuration set the CAPILLARY CONTROL MODE ON, and choose the VELOCITY mode. Enter the nominal column length and exact diameter. Select the split FLOW mode. In the Method, accept the default velocity of 30 cm/s, or choose another value. Set the oven temperature to at least 100 °C, isothermal.

Determine the Unretained Peak Time

Inject 1–5 μL of methane (natural gas) and time how long it takes for the methane peak to elute from the column. This is best done by recording the methane chromatogram with a data handling system and reading the methane peak time off the chromatogram or report. You may use methylene chloride (CH2Cl2) vapor for an ECD.

Calculate the Actual Average Linear Gas Velocity

Calculate the average linear gas velocity ( u ) from this formula:

u LtPPC

M= X 10

where LPPC is the column length (in meters) entered in the Clarus 600 GC and tM is the measured methane peak time in seconds.

Page 278: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-40

Compute the Correction Factor

Compare the measured velocity ( u ) to the displayed velocity ( uPPC ) and compute the column length correction factor (f):

f uu

PPC=

If the correction factor is less than 0.8 or greater than 1.2, then you should double check the column diameter and nominal length for errors.

Enter the Corrected Column Length

Now, go back to the GC configuration and enter the corrected column length (Lnew):

L f Lnew PPC= ⋅

Check the New Length

Upon entering the new length the PPC controller will change the inlet pressure up or down to reflect the new conditions. If the measured velocity was less than the displayed value, then the PPC controller will reduce the pressure so that the measured velocity is now equal to the displayed value. If the measured velocity was higher, then the PPC controller will increase the inlet pressure. Check the new value by measuring the methane peak time again. Your calculated linear velocity should now be very close to the displayed value.

Splitless Injection

Splitless injection refers to a technique that has a zero split flow rate during injection with an increased split flow rate after injection is complete. There are two ways to accomplish splitless injection on the Clarus 600 GC.

Zero Initial Split Flow

For this method, you set the initial split flow to zero and enter a timed event after injection (typically at 1–2 minutes) that sets the split vent flow to 50–100

Page 279: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

8-41

mL/min. The Clarus 600 GC will automatically turn off the split vent flow at the end of the run.

Zero Split Flow by Pre-Run Event

In some cases — with a PSS inlet in particular — it may not be desirable to have zero split vent flow while the instrument is cooling and recycling for another run. Set the initial split vent flow to 50–100 mL/min. Then enter a timed event at -2 minutes to set the split vent flow to 0, and enter another event after injection (typically at 1–2 minutes) that sets the split vent flow to 50–100 mL/min.

Pressure-Pulse Injection

Pressure-pulse injection refers to a technique that facilitates sample transfer from inlet to column by raising the inlet pressure (column flow and velocity) above the normal level for injection, and then lowering it back to a normal setting for peak elution.

To perform pressure-pulse injection, enter a Pre-Run timed event at −1 minute that raises the pressure, flow, or velocity — depending on the operating mode — to a level 2–20 times the method’s initial setpoint value. Then enter another timed event 1–2 minutes into the run that reestablishes the initial method setpoint. If you are using a carrier-gas program, then set a 1–2 minute step 1 hold time in the program. The carrier-gas program will “take over” the PPC system at the end of the hold period. To discharge the pressure-pulse, the split flow must have a non-zero volume at the time of discharge. To rapidly change the pressure pulse, the split flow should be zero or as large as possible during pressurization.

Reducing Carrier-Gas Consumption You can significantly reduce carrier-gas consumption with a split or PSS inlet by setting the initial split flow to 20 mL/min, using a Pre-Run timed event at −2 minutes to increase the flow for split injection or to set the flow to 0 for splitless injection, and then reducing the split flow back to 20 mL/min after completion of the injection. You should delay the flow reduction until at least 5 minutes or longer after injection, or after a PSS inlet has reached its final temperature.

Don’t forget to include a timed event at 1–2 minutes that sets the flow to 50–100 mL/min for splitless injection.

Page 280: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

PPC Fundamentals

8-42

Page 281: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance 9

Page 282: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-2

Page 283: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-3

This chapter contains procedures for:

• Autosampler Maintenance — changing a syringe and replacing a vial-locator mechanism.

• Syringe Maintenance — cleaning the 5-μL and 50-μL syringe plungers.

• Injector Maintenance — changing septa, changing and repacking packed-injector liners, changing and repacking injector liners on the capillary (CAP), programmed split/splitless (PSS), and programmed on-column (POC) injectors, and changing the charcoal trap on the split/splitless injector.

• ECD Maintenance — baking out ECD cells, cleaning the ECD anode, and wipe testing an ECD cell.

• FID Maintenance — replacing the FID jet, cleaning the FID jet, replacing an O-ring in the FID collector, and cleaning the FID collector and cap.

• PID Maintenance — changing PID lamps, cleaning PID lamp windows, changing PID lamp window seals, and positioning disks.

• NPD Maintenance — changing and conditioning the NPD bead and replacing an NPD jet.

• FPD Maintenance — cleaning or replacing an FPD optical filter assembly, cleaning the detector liner, cleaning/replacing the detector window, replacing the photomultiplier tube, and cleaning and replacing the FPD jet.

• PPC Maintenance — replacing PPC module restrictors.

Page 284: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-4

Autosampler Maintenance

Autosampler maintenance consists of changing a syringe and replacing a vial locator mechanism.

Changing a Syringe

1. From the System Status screen touch the Run button to display the run screen.

2. On the Run screen touch the Autosampler radio button and the Tasks tab as shown in the following screen.

Page 285: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-5

3. Touch the Park button and the autosampler tower moves to the park position (facing the front of the Clarus 600 GC).

4. Open the tower door on the autosampler tower cover.

Figure 9-1 Autosampler tower in the park position

Page 286: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-6

Removing a Syringe 1. Locate the plunger assembly shown in the previous illustration.

Then, refer to the following figure, as you lift up the plunger cap handle and rotate it until it rests on the collar. Then release the plunger cap handle.

Figure 9-2 Plunger assembly

2. Hold the syringe by the barrel or syringe nut (see the following figure) and turn the carriage thumbscrew clockwise until the syringe is free.

3. Gently pull the top of the syringe forward until it just clears the carriage assembly.

4. Gently lift the syringe out of the carriage assembly.

Page 287: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-7

Figure 9-3 Removing a syringe

Page 288: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-8

Installing a Syringe

Please refer to Figures 9-2 through 9-4 as you follow these steps.

1. Guide the needle through the hole in the carriage thumbscrew, then thread the needle through the needle guide. Use your fingers as a guide.

2. Rest the top of the plunger on the plunger cap slide, which is a shelf located on the underside of the plunger assembly.

3. While holding the syringe nut, engage the carriage thumbscrew on the threaded part of the syringe by turning the carriage thumbscrew counterclockwise.

4. Continue turning the thumbscrew counterclockwise. This slowly lowers the needle. Carefully guide the needle through the needle guide into the vial locator.

5. Tighten the carriage thumbscrew.

Figure 9-4 Installing a syringe

Page 289: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-9

Replacing the Vial-Locator Mechanism

The vial-locator mechanism will wear out with extended use and require replacement. If the autosampler begins missing vials, or if the hole for the syringe needle begins to plug, it is an indication that you should replace the vial-locator mechanism

To replace a vial-locator mechanism:

1. Remove the two shoulder screws that secure the locator to the autosampler tower frame. Remove the two springs, then remove the vial locator. Discard the vial locator.

2. Mount the new vial locator (P/N N610-1182) on the autosampler tower frame.

3. Install the two shoulder screws through the two springs and into the vial locator. This secures the vial locator to the autosampler tower frame.

WARNING

When securing the vial-locator molding, be sure that the flag is centered (not touching either side) in the sensor. If it touches a side, adjust the flag by loosening and then tightening the screws. DO NOT ADJUST THE SENSOR.

Page 290: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-10

Figure 9-5 Exploded view of the vial locator

Page 291: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-11

Syringe Maintenance

Syringe maintenance consists of cleaning the 5-μL and 50-μL syringe plungers and servicing idle syringes.

Cleaning the 5-μL and 50-μL Syringe Plungers

The 5-μL and 50-μL syringe plungers should be cleaned regularly, after approximately 500 injections, since insolubles can build up and cause friction.

To clean the syringe plunger:

1. Remove the syringe using the procedure described in the preceding section.

2. Remove the plunger from the syringe barrel.

3. Wipe the plunger with a tissue soaked in an appropriate solvent.

4. Replace the plunger.

5. Pull and expel the same solvent through the barrel several times.

6. Replace the syringe using the procedure descried in the preceding section.

NOTE: Only syringes distributed by PerkinElmer should be used with the Clarus GC. Plungers are not interchangeable from syringe to syringe.

Servicing Idle Syringes

Syringes that are not used for several hours could "freeze," i.e., the syringe plunger will not move. To avoid this condition, PARK the tower, then remove and clean the syringe plunger as described above.

NOTE: If you notice the Clarus GC precision degrading, replace the syringe. The Clarus GC syringe is a consumable part. After extended use, you will need to replace it.

Page 292: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-12

Injector Maintenance

CAUTION

If you are analyzing reactive compounds, you should use deactivated liners and wool which are appropriate for your sample type.

Injector maintenance consists of changing septa, changing and repacking injector liners, changing the hourglass needle guide on the POC injector, changing and repacking CAP and PSS injector liners, removing a broken liner from the PSS injector body, changing the charcoal trap or replacing charcoal on the split/splitless CAP and PSS injectors.

Changing Septa

Septa should be replaced on a regular basis. How often depends on the type of septa used, the temperature of the injection port, and the number of injections made.

The septum shipped with your instrument is a Thermogreen LB-2 Septa (P/N N662-1028, package of 50). This septum can handle over 200 injections at moderate temperatures.

To change a septum:

1. Turn off the injector heater and allow the injector to cool.

2. Remove the septum cap.

3. Pry the old septum from the septum cap with a screwdriver.

4. Insert a new septum in the septum cap.

5. Replace the septum cap.

Page 293: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-13

NOTE: To minimize the possibility of contamination, avoid unnecessary handling of septa.

Figure 9-6 Changing a septum

Changing and Repacking Packed Column Injector Liners

To improve the performance of the injector used with packed columns, insert a small amount of quartz wool (P/N N610-2354) into the top portion of the injector liner (P/N N610-1048). The quartz wool accomplishes two things: (1) it wipes the end of the syringe needle to insure that reproducible sample volumes are injected, and (2) it retains any nonvolatile components present in your sample, making cleaning the liner easier.

Page 294: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-14

The injector liner should be removed and the wool packing replaced on a regular basis, particularly if your samples contain nonvolatile components that could build up on the wool. This could cause adsorption of peaks of interest, tailing, and loss of sensitivity.

You can remove the wool with a small hook on the end of a thin wire, or blow it out using compressed air.

To remove a packed injector liner and install new wool:

1. Turn off the injector heater.

Allow the injector to cool until it is slightly warm to the touch. Cooling the injector to a temperature that is too low (<100 °C) will make it difficult to remove the injector liner.

2. Remove the septum cap (see Figure 9-6).

3. Remove the septum shield (P/N N610-1050) with the large end of the liner-removal tool (P/N N610-0102).

Figure 9-7 Liner-Removal Tool (P/N N610-0102)

4. Press the small end of the liner-removal tool into the injector liner, then pull the injector liner out.

Page 295: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-15

Figure 9-8 Removing the packed column injector liner

NOTE: To avoid contaminating the quartz wool when packing the injector liner, wear vinyl, powder-free, disposable gloves (for example, the same type of gloves used to perform maintenance on a mass spectrometer).

5. Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the injector liner. Then, using the supplied 1/16-inch rod (P/N N610-T100), push the quartz wool into the injector liner. Loosely pack a 1-inch (2.5 cm) piece of quartz wool into the top portion of the liner (see the following figure).

6. Replace the injector liner, septum shield, and septum cap.

Page 296: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-16

Figure 9-9 Packed column injector liner (P/N N610-1048) packed with wool

Changing the Hourglass Needle Guide on the Programmed On-Column (POC) Injector To change the hourglass needle guide:

1. Turn off the injector heater. Allow the injector to cool until it is slightly warm to the touch.

2. Remove the septum cap.

Figure 9-10 Removing a septum cap

Page 297: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-17

3. Remove the septum shield (P/N N610-1702) with the large end of the liner-removal tool (P/N N610-0102).

4. Remove the hourglass needle guide (see the following figure) with a pair of tweezers or small pliers.

Figure 9-11 Cutaway view of the POC injector showing the location of the hourglass needle guide

Page 298: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-18

5. Install the new hourglass needle guide (P/N N610-1703).

6. Replace the septum shield.

7. Replace the septum cap.

8. Reinstall your column.

Refer to Chapter 6, Installing a Capillary Column, for the proper procedures.

Changing and Repacking Capillary Split/Splitless (CAP) and Programmed Split/Splitless (PSS) Injector Liners

The procedure below is applicable to the following injector liners:

Injector Liner Size Part Number CAP wide-bore liner 4.0-mm i.d. and a 6.0-mm o.d. N612-1001 CAP narrow-bore liner 2.0-mm i.d. and a 6.0-mm o.d. N612-1002 PSS wide-bore liner 2.0-mm i.d. and a 4.0-mm o.d. N612-1004 PSS narrow-bore liner 1.0-mm i.d. and a 4.0-mm o.d. N612-1006 PSS on-column liner -- N610-1539

Removing a CAP or PSS Injector Liner

The liner-removal procedure is similar for CAP and PSS wide-bore and narrow-bore liners. To remove the liners, you need a CAP liner-removal tool (P/N 0250-6534) or a PSS liner-removal tool (P/N 0250-6247) as shown below.

Figure 9-12 CAP and PSS liner-removal tools

Page 299: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-19

To remove a capillary injector liner:

1. Turn off the injector heater.

Allow the injector to cool until it is slightly warm to the touch. Cooling the injector to a temperature that is too low (<80 °C) will make it difficult to remove the injector liner.

2. Remove the septum cap.

Figure 9-13 Removing the septum cap

3. Remove the injector cover.

Injector Cover (P/N N610-1762)

Page 300: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-20

Figure 9-14 Removing the injector cover

4. Loosen the threaded collar using the spanner (P/N N610-1359) provided, then remove the threaded collar.

Figure 9-15 Loosening the threaded collar

5. Replace the septum cap on the injector.

6. Pull the septum cap upwards to remove the septum purge assembly.

Page 301: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-21

Figure 9-16 Removing the septum purge assembly

The carrier gas inlet line is coiled. This allows you to pull the septum purge assembly over to the side and gain access to the injector liner.

NOTE: The inlet line used in PPC capillary injectors (CAP and PSS) is coiled but the septum purge assembly does not terminate in a snubber as shown. Instead it is connected to a PPC module. The coil of the inlet line is long enough so that you can pull the septum purge assembly out of the opening in the top cover and gain access to the injector liner.

7. Insert the CAP liner-removal tool (P/N 0250-6534) over the end of the CAP liner and lift the liner out of the injector.

OR

Insert the PSS liner-removal tool (P/N 0250-6247) over the end of the PSS liner and lift liner out of the injector.

CAUTION

The liner must be cool (no hotter than 100 °C) or the liner-removal tool will melt! The end of the liner-removal tool may flare out with use. If this happens,

Page 302: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-22

Figure 9-17 Removing an injector liner

NOTE: If the quartz liner breaks inside the CAP injector, it can be removed by first removing the column, then removing with a 9/16-inch wrench the 1/4-inch injector fitting that is inside the oven. The liner should fall out of the injector with the fitting. If the liner is stuck, you can push it out from the top or bottom of the injector.

CAUTION

The PSS injector liner does not have a 1/4-inch fitting like the CAP injector. Be very careful when removing this liner to prevent breaking it. Do not cool the injector below 80 °C. This will make it easier to remove the liner and O-ring. As the injector cools, the O-ring adheres to the metal base.

NOTE: Each capillary liner has an O-ring installed on the frosted portion of the CAP liner and on the part furthest away from the dimple on the PSS injector. If the O-ring has adhered to the injector, you may not be able to easily remove the liner (step 7 above). If this is the case, use a small screwdriver to dislodge the O-ring before removing the liner and O-ring.

About O-Rings

CAUTION

Each time a capillary injector liner is removed, you should replace the O-ring, especially if the O-ring adheres to the injector body and you had to pry it loose with a screwdriver. This action may damage the O-ring thereby causing a bad seal if the damaged O-ring is reinstalled.

If your results produce background contamination when a new O-ring is first installed, condition the injector at the maximum temperature of the O-ring (listed below). Depending on the type of column used, you may first want to remove the column before baking it out at a high temperature.

NOTE: High-temperature seals should be used at temperatures of 300 °C or higher. These seals are available in Kalrez or graphite from PerkinElmer’s catalog service (in the U.S. dial: 1-800-762-4000) or on-line at www. perkinelmer.com. Viton maximum temperature of 250 °C is recommended for the mass spectrometer.

Page 303: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-23

Injector O-Rings Recommended Maximum Temperature

CAP Injector N610-1374 Silicone (pkg .of 10)

250 °C

N610-1378 Graphite (pkg. of 5)

450 °C

N930-2782 Kalrez (pkg. of 1)

450 °C

N930-2783 Viton (pkg. of 1)

250 °C (not recommended for use with ECD)

PSS Injector N610-1751 Graphite (pkg. of 5)

450 °C

0992-1004 Kalrez (pkg. of 1)

450 °C

N610-1747 Viton (pkg. of 10)

250 °C (not recommended for use with ECD)

Selecting an Appropriate CAP Injector Liner

Select the correct CAP liner for your application and pack it with quartz wool. The CAP injector uses the following two liners:

• CAP wide-bore liner (P/N N612-1001); 4.0-mm i.d. and 6.0-mm o.d.

• CAP narrow-bore liner (P/N N612-1002); 2.0-mm i.d. and 6.0-mm o.d.

The narrow-bore liner is generally used for a splitless injection, and the wide-bore liner is generally used for a split injection. Due to the small internal volume (0.3 mL) of the narrow-bore liner, prevent overfilling the liner with vapor (caused by solvent expansion upon injection) by limiting the amount of sample injected to 0.5 μL. The wide-bore liner is used for splitless injection volumes over 0.5 μL, since its internal volume is 1.25 mL. The sample size should be

Page 304: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-24

limited to a maximum of 2 μL for hydrocarbon solvents, and less than that for high-expansion solvents such as water or CH2Cl2.

If the wide-bore liner is used for splitless injection, the splitless sampling time (the vent-on time) should be more than one minute. Also, lower initial oven temperatures may be required to give good resolution in the first few minutes after the solvent peak elutes. The wide-bore liner should be used with columns having an i.d. of 0.32 mm or greater.

Packing the CAP Injector Liner with Quartz Wool

We recommend packing a small amount of quartz wool in the top portion of the liner to wipe the syringe needle regardless of the liner type or injector mode (for example, split or splitless). This packing assures that reproducible volumes are injected by wiping the syringe needle every time it is inserted.

Remove the liner and replace the quartz wool packing on a regular basis, particularly if your samples contain nonvolatile components that could build up on the wool. This buildup could cause adsorption of peaks of interest, tailing, and loss of sensitivity.

Remove the wool by making a small hook on the end of a thin wire and using that to pull it out, or blow it out using compressed air.

NOTE: To avoid contaminating the quartz wool when packing the injection liner, wear vinyl, powder-free, disposable gloves (for example, the same type of gloves used to perform maintenance on a mass spectrometer).

Packing a CAP Injector Liner for Split Operation

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using the supplied 1/16-inch rod (P/N N610-T100), push the quartz wool into the liner. Pack the wool snuggly∗ from the dimple upwards [about 1 in. (2.5 cm)]. Loosely pack quartz wool in the top portion of the liner to wipe the syringe needle upon injection.

∗ The recovery of high-molecular-weight components (e.g., C40 and higher) may be improved if the

liner is packed loosely.

Page 305: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-25

Packing a CAP Injector Liner for Splitless Operation

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using the supplied 1/16-inch rod (P/N N610-T100), push the quartz wool into the liner. Pack a 1-inch (2.5 cm) piece of quartz wool loosely below the top ground portion of the liner (see the following figure). The sample is then injected into the wool, thereby preventing the delivery of sample beyond the column. The wool also wipes the syringe needle upon injection.

Figure 9-18 CAP injector liners packed with quartz wool

Page 306: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-26

NOTE: As you can see in Figure 9-18, each CAP injector liner has an O-ring installed on the ground portion. If the O-ring has adhered to the liner, it may not be easy to remove the liner. (If this is the case, use a small screwdriver to dislodge the O-ring before removing the liner and O-ring).

Reinstalling the Liner in the CAP Injector

1. Install a new O-ring near the ground portion of the liner.

2. Insert the liner in the injector body.

3. Place the septum purge assembly over the liner.

4. Press the septum purge assembly down to correctly position the liner in the injector.

Selecting an Appropriate PSS Injector Liner

Select the correct PSS liner for your application and pack it with quartz wool. The PSS injector uses the following three liners:

• PSS injector wide-bore liner (P/N N612-1004); 2.0-mm i.d.

• PSS injector narrow-bore liner (P/N N612-1006); 1.0-mm i.d.

• PSS injector on-column (hourglass) liner (P/N N610-1539)

In general, operate the PSS in the inlet-programmed mode with the 2-mm or 1-mm-i.d. liner for PSS split or splitless injection. For PSS on-column operation, use the hourglass liner and the oven-program mode.

The 2-mm-i.d. PSS liner that is used for either split or splitless operation should be packed with quartz wool as described in this chapter. The 1-mm-i.d. PSS liner may give better early-eluting peak resolution in the split or splitless mode. It should be used for samples with early-eluting peaks for which additional solute trapping/focusing cannot be obtained by lowering the initial oven temperature or by using a column with thicker stationary-phase film.

Page 307: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-27

Packing the PSS Injector Liner with Quartz Wool

CAUTION Never pack the hourglass liner with quartz wool.

Pack a small amount of quartz wool in the top portion of the liner to wipe the syringe needle regardless of the liner type or injector mode (for example, split or splitless). Packing assures that reproducible volumes are injected by wiping the syringe needle every time it is inserted.

Remove the liner and replace the quartz wool packing on a regular basis, particularly if your samples contain nonvolatile components that could build up on the wool. This could cause adsorption of peaks of interest, tailing, and loss of sensitivity. You can remove the wool by making a small hook on the end of a thin wire and using that to pull it out, or blow it out using compressed air.

NOTE: To avoid contaminating the quartz wool when packing the injection liner, wear vinyl, powder-free, disposable gloves (for example, the same type of gloves worn when performing maintenance on a mass spectrometer).

Packing a CAP Injector Liner for Split Operation

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then, using the supplied 1/16-inch rod (P/N N610-T100), push the quartz wool into the liner. Pack the wool tightly∗ from the dimple upwards [about 1 in. (2.5 cm)]. Loosely pack quartz wool in the top portion of the liner to wipe the syringe needle upon injection.

* The recovery of high-molecular-weight components (e.g., C40 and higher) may be improved if the

liner is packed loosely.

Page 308: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-28

Packing a CAP Injector Liner for Splitless Operation

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then, using the supplied 1/16-inch rod (P/N N610-T100), push the quartz wool into the liner. Pack a 1-inch (2.5 cm) piece of quartz wool loosely below the top ground portion of the liner (see the following figure). The sample is then injected into the wool, thereby preventing the delivery of sample beyond the column. The wool also wipes the syringe needle upon injection.

NOTE: The narrow-bore liner is more difficult to pack because of its small inner diameter. However, there is a dimple in the middle of the liner to hold the wool in place. Do not pack the wool too tightly!

Figure 9-19 PSS injector liners packed with quartz wool

NOTE: As you can see in the preceding figure, each PSS injector liner has an O-ring installed near the end furthest away from the dimple. If the O-ring has adhered to

Page 309: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-29

the liner, it may not be easy to remove the liner. If this is the case, use a small screwdriver to dislodge the O-ring before removing the liner and O-ring.

Reinstalling the Liner in the PSS Injector

1. Install a new O-ring on the top portion of the liner.

2. Insert the liner in the injector body.

3. Place the septum purge assembly over the liner.

4. Press the septum purge assembly down to correctly position the liner in the injector.

Make sure that you secure the septum purge assembly tightly to the injector base with the 1/4-inch spanner.

Removing a Broken Liner from the PSS Injector Body If a liner breaks in the PSS injector body, the best way to remove all pieces of quartz is to remove the injector from the Clarus GC.

To remove a broken PSS liner from the injector:

1. Turn off the Clarus GC. Allow the injector to cool until it is slightly warm to the touch.

2. Disconnect and remove the column from the injector.

3. Loosen the two screws the secure the Clarus GC top cover and raise the top cover until it locks in the raised position.

4. Remove the top cover from the PSS injector. Mark the position of the injector on the metal deck with a pencil.

5. Remove the screw that secures the fan assembly to the PSS injector, then remove the fan assembly (see the following figure).

6. Remove the two screws that secure the PSS injector to the metal deck.

Page 310: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-30

7. Remove the cable clamps that hold the heater and sensor wires and the gas tubing (see the following figure).

8. Carefully lift the PSS injector out of the Clarus GC. Invert the PSS injector and remove all of the broken quartz liner from it.

9. Reinstall the PSS injector to the Clarus GC and reinstall the fan assembly.

10. Lower the Clarus GC top cover and properly align the PSS injector with the top cover before you completely tighten the PSS injector mounting screws.

Figure 9-20 Removing the fan assembly from the PSS injector

CAUTION

To prevent autosampler needle damage after the Clarus GC top cover has been opened and closed, verify that the autosampler tower is aligned with both injectors.

Do this by manually rotating the autosampler tower and stopping over injector 1 and injector 2 to check that the vial locator is in the center of the septum cap. If the vial locator does not align with the center of the septum cap, loosen the two hold-down screws that secure the top cover (see Figure 9-22). Then move the top cover so that the vial locator is aligned with the center of the septum cap. Secure the Clarus GC top cover in this position by tightening the two screws.

Page 311: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-31

Figure 9-21 Location of the cable clamps that secure the wires and tubing

Page 312: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-32

Changing the Charcoal Trap or Replacing Charcoal on the Split/Splitless CAP and PSS Injectors

The charcoal trap will eventually become saturated. When this occurs, ghost peaks and changes in split ratio will be observed.

Removing a Charcoal Trap

1. Turn off the Clarus GC. Allow the injectors/detectors to become cool to the touch.

2. Loosen the two hold-down screws on the top cover of the Clarus GC (see following figure) and raise the top cover until the cover locks in the raised position.

DetectorCover

Electronics Cover

Hold-Down Screw

AutosamplerTower

Hold-Down Under Cover

Screws

Figure 9-22 Location of the top cover hold-down screws

3. Remove the septum cap, then remove the top cover from the injector.

4. Loosen the threaded collar using the 1/4-inch spanner (P/N N610-1359) provided, then remove the threaded collar.

5. Replace the septum cap on the injector.

Page 313: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-33

6. Pull the septum cap upwards to remove the septum purge assembly.

7. Using an 1/8-inch wrench, loosen the fittings that are connected to the charcoal trap and remove the charcoal trap (see the following figure). If you have a PPC charcoal trap, use a 1/4-inch wrench to loosen the fitting that connects the trap to the transducer.

Installing a New Charcoal Trap

1. Install a new PPC version charcoal trap (P/N N610-0331) or just replace the charcoal in your current trap.

Replace the charcoal by removing the glass wool plug from the 1/4-inch tubing end. Empty the old charcoal and the glass wool plug at the 1/8-inch end. Push a small glass wool plug from the 1/4-inch end to the end of the 1/4-inch tube (near the 1/8-inch end) then fill with charcoal (P/N 0330-0904) and plug the end with a small piece of silanized glass wool.

Figure 9-23 Charcoal Trap on an injector controlled by PPC

2. Replace the septum purge assembly and remove the septum cap.

3. Replace the threaded collar and tighten it using the spanner.

4. Replace the injector cover then replace the septum cap.

5. Lower the Clarus GC top cover and tighten the two hold-down screws.

Page 314: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-34

CAUTION

To prevent autosampler needle damage after the Clarus GC top cover has been opened and closed, verify that the autosampler tower is aligned with both injectors.

Do this by manually rotating the autosampler tower and stopping over injector 1 and injector 2 to check that the vial locator is in the center of the septum cap. If the vial locator does not align with the center of the septum cap, loosen the two hold-down screws that secure the top cover (see Figure 9-22). Then move the top cover so that the vial locator is aligned with the center of the septum cap. Secure the Clarus GC top cover in this position by tightening the two screws.

Page 315: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-35

ECD Maintenance

If you observe that the ECD background is higher than normal for your operating conditions, the cell could be contaminated.

You can view the ECD background reading from the Clarus GC touch screen by selecting the signal tab in the ECD tab. Also select the signal tab to display the detector background and the AutoZero button. Press the AutoZero button to have the system on autozero.

Under normal operating conditions, the ECD background will be up to 7 mV.

If you suspect cell contamination, first eliminate column bleed by lowering the oven temperature to ambient. If bleed is not the problem and the high background coincided with changing the carrier gas tank, the carrier gas may be contaminated. To check for this condition:

1. Cool the ECD to <100 ºC.

2. Remove the column from the ECD, then cap the ECD with a 1/8 inch steel plug (P/N 0990-3098).

3. Increase the make-up flow. (The ECD is a concentration-sensitive detector. Increasing the make-up gas flow would normally dilute the contamination and cause a decrease in the background.)

If the background remains the same or increases as the make-up flow increases, the carrier gas could be contaminated. If bleed or carrier gas contamination is not the problem, bake the ECD using the following procedure.

Baking the ECD

1. Remove the column, then cap the ECD with a plug (P/N 0990-3098).

2. Increase the flow of make-up gas to 60 mL/min and raise the detector temperature to 450 °C.

Page 316: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-36

3. Bake the system until the background returns to normal levels. This could take from hours to several days.

NOTE: It may also help to remove the column and increase the oven temperature to 450 °C to bake out the lower portion of the ECD body.

Changing the Charcoal Filters

The ECD is shipped with charcoal filters (P/N N660-0037) installed in the make-up and injector pneumatics lines to remove contamination from the needle valve, flow controller, or pressure regulator. The filters should be replaced periodically.

To change charcoal filters:

1. Turn off the Clarus GC and allow the injectors/detectors to cool.

2. Loosen the two hold-down screws on the GC top cover (see Figure 9-22) and raise the top cover until it locks in the raised position.

3. Disconnect the charcoal filters from the make-up gas and injector lines.

4. Install new charcoal filter filters.

��� 1

��� 2

PPCInternal Bulkhead

Graphite/Vespel Ferrules(P/N 0992-0301)

HYD or Make Up

1/16-inch x 9-inch Long Tubing (P/N N610-1267)

Charcoal Filter

ECD

Page 317: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-37

Cleaning the ECD Anode

WARNING

THE FOLLOWING PROCEDURE SHOULD BE PERFORMED ONLY ON ECD’S THAT HAVE A VALID ACCEPTABLE WIPE TEST RESULT WITHIN THE LAST SIX MONTHS.

PERFORMING UNAUTHORIZED CLEANING PROCEDURES SUCH AS STEAM CLEANING OR FLUSHING SOLVENTS THROUGH THE ECD ARE STRICTLY FORBIDDEN AND ARE KNOWN TO DISCHARGE RADIOACTIVE CONTAMINATION FROM THE SOURCE. SOURCES WHICH ARE CLEANED IN THIS FASHION SHOULD BE CONSIDERED HOT.

NOTE: If a dirty or contaminated ECD is suspected, try baking out the detector before using this procedure.

NOTE: Wear plastic or rubber gloves when cleaning the ECD anode.

To clean the ECD anode:

1. Turn off the ECD heater and allow the system to cool to room temperature.

2. Unscrew the knurled ring and lift out the anode assembly (see the following figure).

3. Place the collector assembly on top of a beaker of hexane with the anode tube submerged and soak for several minutes. DO NOT submerge the side arm in the hexane; submerge only the anode.

4. Remove the anode assembly and wipe it dry with a tissue.

5. Replace the anode assembly, then tighten the knurled ring.

6. Turn on the detector temperature and observe that the background signal has returned down to a normal level.

Page 318: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-38

Figure 9-24 Isometric view and cross section view of the ECD

Wipe Testing an ECD Cell

CAUTION

Until the results of the wipe test are known, use caution and suitable protection when handling the cell and equipment in contact with it. Wear disposable plastic or rubber gloves when performing this test.

It is strongly recommended that you become familiar with the NRC regulation covering the use of Nickel-63, as well as any other national, state, or local requirements.

To perform the wipe test:

1. Turn the instrument off and allow the detector to become cool to the touch.

Page 319: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-39

2. Gain access to the detector by lifting the detector cover (see the following figure).

Detector Cover

Figure 9-25 Detector cover location

3. Remove the two screws holding down the ECD insulating cover, then remove the insulating cover (see the following figure). (Removing the insulating cover exposes the knurled ring and detector outlet.)

WARNING

DO NOT DISMANTLE THE ECD CELL!

Page 320: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-40

Figure 9-26 ECD insulating cover

Figure 9-27 ECD surfaces to wipe

4. Using the instructions included with the wipe-test kit (P/N 0009-1667) supplied with the detector, wipe the external surfaces of the items shown in the previous figure with the “low Activity Source” filter papers or stick swab:

Page 321: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-41

• Detector outlet

• Knurled ring

• Detector fitting

CAUTION

Do not remoisten the wipe-test paper once it has been moistened or any part of the ECD has been wiped. Do not allow any of the wipe-test solution to enter the cell.

5. Place the wipe-test paper in the container provided in your wipe-test kit. Include a data sheet stating that the wipe test was performed on a PerkinElmer electron capture detector cell (P/N N610-0063, 120 volts or P/N N610-0134, 230 volts) and the date of the test.

6. Request a new wipe-test kit to be sent with the test results.

7. Return the envelope to:

National Leak Test Center P.O. Box 486 North Tonawanda, New York 14120 Tel: 716-693-0550

or

NRD P.O. Box 310 Grand Island, New York 14072 Tel: 800-525-8076 716-773-7634

NOTE: The sensitivity of the wipe test is 0.0001 μCi.

Page 322: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-42

Disposal of an ECD Cell

If it is necessary to dispose of an ECD cell, contact: Nuclear Radiation Development Corp. 2937 Alt. Blvd. North Grand Island, NY 14072 Tel: (716) 773-7634 Fax: (716) 773-7744

for disposal instructions and current fees.

In addition, report the ECD cell disposal to: PerkinElmer, Inc. Radiation Safety Officer 710 Bridgeport Ave. Shelton CT 06484

and Nuclear Material Safety and Safeguard U.S. Nuclear Regulatory Commission Washington, DC 20251

and/or your local agency, if applicable.

Refurbish/Refoil of an ECD Cell

If it is necessary to refurbish/refoil an ECD cell, contact: Nuclear Radiation Development Corp. 2937 Alt. Blvd. North Grand Island, NY 14072 Tel: (716) 773-7634 Fax: (716) 773-7744

for instructions and current fees.

Page 323: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-43

FID Maintenance

FID maintenance consists of replacing the FID jet, cleaning the FID jet, replacing an o-ring in the collector, and cleaning the FID collector and cap.

Replacing a FID Jet

NOTE: The FID jet rarely becomes plugged. However, if plugging occurs, it is usually sample dependent. It is recommended that you replace a plugged jet rather than clean it.

To replace the FID jet:

WARNING

Before you begin, extinguish the flame via the keyboard by setting the hydrogen flow to “0” or if you have manual pneumatics, turn the outer knob on the hydrogen needle valve completely clockwise to off.

1. Turn off the Clarus GC power.

WARNING

The FID is hot and can cause serious burns! To prevent injury, allow the detector to become cool to the touch.

2. Open the detector cover (see Figure 9-25).

3. Remove the polarizing cable from the pin on the polarizing filter assembly.

Page 324: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-44

Figure 9-28 FID polarizing voltage wire

4. Loosen the knurled ring, then lift the FID collector off of the FID base and put it out of the way.

5. Insert the nozzle removal tool (P/N N610-3188) into the FID base and lift the nozzle out of the FID base. Do not unscrew it.

Page 325: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-45

Nozzle

FID Base

Nozzle RemovalTool (P/N N610-3188)

Figure 9-29 Removing the nozzle assembly from the FID base

6. Insert a 1/4-inch nutdriver into the FID base to engage the 1/4-inch nut on the FID jet assembly.

7. Loosen the FID jet assembly (turn the 1/4-inch nut counterclockwise) and pull it out of the FID. You should be able to pull out the FID jet assembly with the nutdriver. If not, then pull out the FID jet assembly with a pair of forceps or needle nose pliers.

Page 326: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-46

Figure 9-30 Cross section view of the FID

8. Insert a new FID jet assembly (P/N N610-0361) and secure it in place with the 1/4-inch nut driver.

9. Insert the nozzle assembly into the FID base until you feel it bottom.

10. Insert the FID collector back on the FID base and tighten the knurled ring.

11. Reconnect the polarizing wire to the polarizing pin on the FID collector.

12. Turn on the FID heater and allow it to return to the temperature setting.

13. Re-ignite the flame.

Page 327: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-47

Cleaning a FID Jet

Although it is not recommended, you may try to clean the FID jet as a last resort. Use one or both of the following techniques:

• Based on your analytical application, wash the jet with an appropriate solvent.

• Dislodge the plug with a fine wire such as a syringe needle, then blow out loosened debris using compressed air.

Replacing the O-Ring in the FID Collector

Since the O-ring in the FID collector is in contact with the heated surface of the FID base, you will notice over time that it has become brittle or broken and must be replaced.

WARNING

The FID is hot and can cause serious burns! To prevent injury, extinguish the FID flame, turn off the FID heater, and allow the detector to become cool to the touch.

To replace the O-ring in the FID collector:

1. Remove the polarizing voltage wire from the polarizing pin (Figure 28).

2. Loosen the knurled ring, then lift the FID collector off of the FID base.

3. Remove the old O-ring (see the following figure) from the FID collector and insert a new O-ring (P/N 0990-2143).

4. Insert the FID collector back on the FID base and tighten the knurled ring.

5. Connect the polarizing wire to the polarizing pin on the FID collector.

Page 328: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-48

6. Turn on the FID heater and allow it to return to the temperature setting.

7. Re-ignite the flame.

Cleaning the FID Collector and Cap

Occasionally clean the collector and cap if you are running samples that may generate soot, for example, carbon disulfide.

WARNING

The FID is hot and can cause serious burns! To prevent injury, extinguish the FID flame, turn off the FID heater, and allow the detector to become cool to the touch.

To clean the FID collector:

1. If necessary, disconnect the amplifier coaxial cable, and other wires from the FID collector.

2. Loosen the knurled ring on the collector and remove the collector from the FID base.

3. Using a pipe cleaner, wipe the inside of the collector and then the outside of the collector near the top.

4. Wash the collector with a laboratory soap such as Alconox. Try to keep the side-arm dry.

5. Air dry the collector replace it on the FID base, and tighten the knurled ring to secure the collector in place.

6. If you disconnected the amplifier coaxial cable and any wires from the FID collector, reconnect them to the FID collector.

Page 329: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-49

Figure 9-31 View of the FID collector and O-ring

Page 330: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-50

PID Maintenance

Routine PID maintenance consists of the following:

• Changing PID lamps

• Cleaning PID lamp windows

• Changing PID lamp window seals and positioning disks

Procedures for performing these tasks follow.

WARNING

The PID operates at a high temperature and voltage. To avoid injury, before attempting maintenance procedures, disconnect the PID from line power and allow it to cool to room temperature.

CAUTION

To avoid contamination, wear rubber gloves or use tweezers when disassembling the PID.

Changing a PID Lamp

To change a PID lamp:

1. Disconnect the instrument from line power.

2. Allow the system to cool to room temperature.

3. Place a piece of paper towel to the right of the detector area (see shaded area in the following figure) to cover the space between the oven skin

Page 331: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-51

and the electronics compartment of the Clarus GC. This will prevent the lamp housing fasteners (see next step) from falling into the instrument when they are loosened.

Figure 9-32 Placing a towel in the space between detector and electronics compartment

4. Please examine the previous figure. While pressing down on the top of the PID (lamp housing), remove the four fasteners securing the lamp housing to the detector and place them in a safe place.

5. Slowly ease up on the lamp housing. As you do, the spring inside the housing (see the following figure) will push the housing up. Pull off the lamp housing and reveal the spring. The spring is quite powerful and it may shoot off if the lamp housing is removed too quickly.

6. Remove the spring.

Page 332: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-52

7. Replace the worn out or damaged lamp, window seal, and positioning disk.

8. Replace the spring and housing.

9. While holding the lamp housing down tightly, replace the four fasteners previously removed.

10. Remove the paper inserted in step 3.

Cleaning PID Lamp Windows

CAUTION Wear rubber gloves to avoid contaminating the lamp window.

To clean PID lamp windows:

1. Remove the lamp using steps 1 through 7 in the previous procedure.

Page 333: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-53

Figure 9-33 Exploded view of the PID

2. First clean the window with a moist, clean Kimwipe or other lint-free tissue. Use a gentle, circular motion.

3. Complete the cleaning by using a tissue moistened with cleaning compound supplied (P/N 0330-2775).

4. Rinse the window using a warm (30 °C) solution of a mild dishwashing detergent in water.

Page 334: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-54

5. Rinse with warm (30 °C) distilled water.

6. Dry with air or a lint-free tissue.

7. Ensure that a new lamp window seal and a new positioning disk are in place (see Figure 9-33). Then position the lamp over the window seal, place the spring over the lamp, and place the housing over the lamp and spring.

8. While holding the lamp housing down tightly, replace the four fasteners previously removed.

Changing PID Lamp Window Seals and Positioning Disks

CAUTION

New seals and positioning disks must be baked (preferably in a vacuum oven) before use for one to two hours at 240 °C.

To change PID lamp window seals and positioning disks, refer to the previous figure and follow this procedure:

1. Wearing rubber gloves to avoid contaminating the lamp window, remove the lamp using the appropriate steps in the previous procedure.

2. The lamp window seal will either be in the center of the positioning disk or stuck to the lamp window. Using a pair of tweezers, remove the lamp window seal.

3. Lift out the positioning disk.

4. Replace the positioning disk with a new one, making certain that you do not crease or bend it.

5. Place a new lamp window seal in the center of the positioning disk. Viewing the detector from the end, make certain that the opening in the seal is lined up with the ion chamber opening. The whole window seal must rest on the gold-plated ceramic portion of the ionization chamber.

Page 335: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-55

6. Position the lamp over the seal, place the spring over the lamp, and place the housing over the lamp and spring.

7. Secure the lamp housing with the four fasteners.

Leak-Test

You can check for leaks by turning on the make-up gas and checking the flow rates at the detector inlet and outlet to make certain they are equal.

Page 336: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-56

NPD Maintenance

NPD maintenance consists of changing and conditioning the NPD bead and replacing a NPD jet.

Changing the NPD Bead

The Nitrogen Phosphorus Detector utilizes a glass bead containing alkali metal (single bead P/N N612-0092 or package of five P/N N612-0093) to detect organically bound nitrogen and phosphorus compounds. In time, the bead will not respond, or the wire on which the bead is placed may break. The bead is considered a consumable part.

If you cannot achieve a response at your normal operating background (0.25 mV or greater with the detector range set to x1), increase the potentiometer setting (try higher settings). If you cannot achieve a response at higher settings, the bead must be replaced.

NOTE: An indication of a broken bead wire is that the bead does not glow when you increase the bead potentiometer setting (by turning it clockwise).

To change the NPD bead:

1. Lift open the detector cover.

2. Locate the bead potentiometer.

If the NPD is installed in the front detector position, the bead potentiometer is located on the left side of the detector panel. If the NPD is installed in the rear detector position, the bead potentiometer is located on the right side of the detector panel.

3. Turn the NPD bead off by turning the potentiometer counterclockwise.

NOTE: If two NPDs are installed, turn both beads off, even if you are only replacing one bead.

4. Turn off the Clarus GC.

Page 337: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-57

WARNING

The gases can remain on during this procedure, but the detector should be cool to the touch to protect you from getting burned.

5. Remove the NPD collector assembly (see the following figure) by loosening its knurled ring and lifting the collector assembly upward.

CAUTION

Lift the collector assembly straight up so that it does not chip the ceramic header of the bead assembly. You may find it easier to remove the coaxial cable from the collector assembly before you remove the collector assembly from the detector body (see the following figure).

6. Remove the screw that secures the bead transformer assembly to the top of the Clarus GC oven.

7. Carefully remove the bead portion from the detector body by lifting the bead transformer assembly straight up and out of the detector body.

8. Remove the bead assembly from the transformer assembly by unplugging it from the connector (see the following figure).

9. Plug a new bead assembly (P/N N612-0092) into the connector on the bead transformer. The connector is keyed so that the bead assembly can only be inserted one way.

10. Carefully insert the bead portion of the bead assembly in the detector body as shown in (see the following figure).

11. Secure the bead transformer to the top of the oven with the screw removed in step 6 of this procedure.

12. Replace the collector assembly on the detector body, and secure it by tightening the knurled ring. If the coaxial cable was removed, connect it to the collector assembly.

Page 338: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-58

NOTE: Check that the polarizing wire has not fallen off the detector. If it has, replace it (Figure 9-37).

13. Turn on the Clarus GC.

Figure 9-37 Exploded view of the NPD bead assembly

Page 339: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-59

Conditioning a New NPD Bead for Use with a Packed Column

CAUTION

Never condition a NPD bead when a column is connected to the detector fitting. To properly condition the NPD bead, the column must be removed from the detector fitting and the detector fitting must be capped.

1. Open the oven door and allow the oven to cool.

2. Loosen and remove the column nut from the detector fitting using a 7/16-inch wrench. Remove the column from the detector fitting.

3. Install a 1/8-inch Swagelok plug (P/N N930-0061) on the detector fitting. Provide a leak-free seal by tightening the plug with a 7/16-inch wrench.

4. Close the oven door and continue to flow carrier gas through the column.

5. Set the hydrogen flow, air flow, and all temperatures to operating conditions.

6. Set the detector range to x1.

7. Select the signal tab to display the detector background and the AutoZero button. Press the AutoZero button to have the system on autozero.

Page 340: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-60

When the bead is turned off, the background value should be 0 ±0.25 mV. Write down the background value with the bead off.

8. Open the detector cover and locate the bead potentiometer dial (see Figure 9-33).

If the NPD is detector 1 (front position), the potentiometer is located on the left side of the panel. If the NPD is detector 2 (rear position), the potentiometer is located on the right side of the panel.

9. Turn the bead potentiometer dial clockwise to apply current to the bead. Slowly increase the setting to 700 on the dial.

The millivolt reading on the autozero display will begin to increase. A maximum reading of 997.61 mV at a range setting of x1 is required to condition the bead.

Page 341: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-61

10. Monitor the display for about one minute at this setting. If the millivolt reading is less than 997.61, increase the bead potentiometer setting by an increment of 10 (to 710) and wait about one minute.

Repeat this process until the bead potentiometer setting produces a maximum signal of 997.61 mV.

11. Allow the bead to "condition" at this setting for one hour. The background readout will drift downward during the process.

12. After conditioning the bead, lower the bead potentiometer dial setting to 600. Remove the Swagelok plug from the detector fitting and reinstall the packed column.

13. Make sure the column flow is properly set and select the signal tab to display the detector background produced by carrier gas flow and the AutoZero button. Press the AutoZero button to have the system on autozero.

The bead background will be less due to cooling effects.

Page 342: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-62

14. Turn the bead potentiometer dial clockwise until you achieve a reading of 0.25 mV above the "off" reading, with the detector range set to x1 as noted in step 7 of this procedure.

15. Once the system has stabilized, it is ready for operation.

If you are unable to achieve adequate sensitivity for your standard, increase the background of the NPD. Increasing intervals of 0.25 mV are recommended.

NOTES:

The bead will be stable for several hours of operation but will drift in time. At the start of each day, adjust the background reading to the setting you are using and allow a few minutes for it to stabilize.

Due to the loss of alkali metal with use, the nature of the bead is to drift over time. We strongly recommend using an internal standard for quantitative analysis.

The bead can operate at higher background settings. The higher the setting the greater the signal and noise. Therefore, the signal to noise ratio will not increase dramatically. Operate at the lowest possible setting to achieve the required sensitivity. This will also prolong the life of the bead.

CAUTION

Before you turn off the Clarus GC, turn off The bead by turning the bead potentiometer dial fully counterclockwise . Remember that the bead has a finite life. You can extend the bead life by turning the bead off when it is not in use for an extended period of time (for example, over the weekend).

Page 343: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-63

Conditioning a New NPD Bead for Use with a Capillary Column (0.53 mm i.d. and smaller)

CAUTION

Never condition a NPD bead when a column is connected to the detector fitting. To properly condition the NPD bead, the column must be removed from the detector fitting and the detector fitting must be capped.

1. Open the oven door and allow the oven to cool.

2. Loosen and remove the column nut from the detector fitting using a 7/16-inch wrench. Remove the column from the detector fitting.

3. Install a 1/16-inch Swagelok plug (P/N N930-0061) on the detector fitting. Provide a leak-free seal by tightening the plug with a 7/16-inch wrench.

4. Close the oven door and continue to flow carrier gas through the column.

5. Set the hydrogen flow, air flow, and all temperatures to operating conditions.

6. Set the hydrogen flow, air flow, and all temperatures to operating conditions.

7. Set the detector range to x1.

8. Select the signal tab to display the detector background and the AutoZero button. Press the AutoZero button to have the system on autozero.

Page 344: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-64

When the bead is off, the background value should be 0 ±0.25 mV. Write down the background value with the bead off.

9. Open the detector cover and locate the bead potentiometer dial.

If the NPD is detector 1 (front position), the potentiometer is located on the left side of the panel. If the NPD is detector 2 (rear position), the potentiometer is located on the right side of the panel.

10. Turn the bead potentiometer dial clockwise to apply current to the bead. Slowly increase the setting to 700 on the dial.

The millivolt reading on the autozero display will begin to increase. A maximum reading of 997.61 mV at a range setting of x1 is required to condition the bead.

11. Monitor the display for about one minute at this setting. If the millivolt reading is less than 997.61, increase the bead potentiometer setting by an increment of 10 (to 710), and wait about one minute.

Page 345: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-65

.

Repeat this process until the bead potentiometer setting produces a maximum signal of 997.61 mV.

12. Allow the bead to "condition" at this setting for one hour. The background readout will drift downward during this process.

13. After conditioning the bead, readjust the column flow if necessary.

14. Select the signal tab to display the detector background and the AutoZero button. Press the AutoZero button to have the system on autozero.

15. Turn the bead potentiometer dial clockwise or counterclockwise until you achieve a reading of 0.25 mV above the "off" reading with the detector range set to x1 (see step 4 of this procedure).

16. Once the system has stabilized, it is ready for operation.

Page 346: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-66

If you are unable to achieve adequate sensitivity for your standard, increase the background of the NPD. Increasing the background in intervals of 0.25 mV is recommended.

NOTE: 1. The bead will be stable for several hours of operation but will drift in time. At the start of each day, adjust the background reading to the setting you are using and allow a few minutes for it to stabilize. 2. Due to the loss of alkali metal with use, the nature of the bead is to drift over time. We strongly recommend using an internal standard for quantitative analysis. 3. The bead can operate at higher background settings. The higher the setting the greater the signal and noise. Therefore, the signal to noise ratio will not increase dramatically. Operate at the lowest possible setting to achieve the required sensitivity. This will also prolong the life of the bead.

CAUTION

Before you turn off the Clarus GC, turn off the bead by turning the bead potentiometer dial fully counterclockwise. Remember that the bead has a finite life. You can extend the bead life by turning the bead off when it is not in use for long periods of time (for example, over the weekend).

Replacing an NPD Jet

NOTE: The NPD jet rarely becomes plugged. However, if the jet does become plugged, it is usually because of the type of sample used. We recommend replacing a plugged NPD jet.

To replace an NPD jet:

1. Turn off the bead by turning the potentiometer dial fully counterclockwise.

2. Turn off the Clarus GC.

3. Turn off the hydrogen and air flows.

4. Open the detector cover.

Page 347: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-67

WARNING

Wait until the detector is cool to the touch to protect you from getting burned.

5. Loosen the knurled ring on the collector assembly, then remove the collector assembly (see the following figure).

Figure 9-38 Location of the NPD jet assembly

Page 348: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-68

CAUTION

Lift the collector straight up so that it does not chip the ceramic header of the bead assembly. You may find it easier to remove the coaxial cable from the collector before you remove the collector (see the previous figure).

6. Remove the screw that secures the bead transformer assembly to the top of the Clarus GC oven.

7. Remove the bead assembly from the detector body by lifting the entire bead transformer assembly straight up and out of the detector body (see Figure 9-38).

NOTE: Carefully place the bead transformer assembly out of the way so that the bead is not damaged. You may want to remove (unplug) the bead assembly from the transformer in order to protect the bead.

8. Remove the polarizing wire. This exposes a spring-loaded polarizing pin, which is a piece of wire about 3/8-inch long (see the following figure).

Page 349: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-69

Polarizing Pin

Polarizing Cable

NPD Chimney

Nozzle Assembly

Liner Removal Tool(P/N N610-0102)

Figure 9-39 Removing the NPD nozzle assembly and polarizing wire

9. Grasp and pull the spring-loaded polarizing pin using a pair of needle nose pliers with your left hand. Maintain a steady pull on the spring-loaded pin.

10. With your right hand, insert the large end of the liner removal tool (P/N N610-0102) into the NPD chimney so that it engages the nozzle assembly. Then remove the nozzle assembly by lifting it out.

Page 350: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-70

11. Still maintaining a steady pull on the spring-loaded polarizing pin, use your right hand to hold a pair of curved pliers or forceps and remove the ceramic insulator (not shown) from the detector body.

12. Still maintaining a steady pull on the spring-loaded polarizing pin, use your right hand to insert a 1/4-inch nutdriver (P/N N610-1297 provided in shipping kit) into the NPD chimney (see Figure 9-39) and engage the nut on top of the NPD jet assembly.

13. Loosen the NPD jet assembly by turning the nut counterclockwise. Then pull the jet assembly out of the NPD with the nutdriver.

If you cannot pull the jet assembly out with the nutdriver, use a pair of forceps or pliers.

14. Install a new jet assembly (P/N N610-0038) by reversing steps 5 through 13.

Page 351: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-71

FPD Maintenance

The most common causes of FPD performance deterioration which results in a decrease of the signal-to-noise ratio are as follows:

• Dirty optical filter

• Dirty detector liner

• Dirty window

• Defective photomultiplier tube

• Dirty jet

Cleaning/Replacing an Optical Filter Assembly

Obtain one of the following filters before you begin:

• Sulfur Filter Assembly, P/N N600-0637 (standard shipping filter) (Purple)

• Phosphorus Filter Assembly, P/N N600-0981 (Yellow)

• Tin Filter Assembly, P/N L413-5472 (Orange)

To clean or replace an optical filter assembly:

1. Turn off the photomultiplier voltage by selecting the following screen. Locate PMT Voltage on the screen and select OFF.

Page 352: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-72

2. Extinguish the flame by turning off the hydrogen.

CAUTION

The photomultiplier tube can be damaged if exposed to light while voltage is ON. Never remove the detector cap or Photomultiplier tube WITH the voltage on.

3. Turn off the Clarus GC and allow the detector to cool.

4. Loosen the knurled nut that secures the photomultiplier tube assembly to the detector body.

5. Slide the photomultiplier tube assembly away from the detector. If the FPD is detector 1 (front position), slide the photomultiplier tube assembly to the left. If the FPD is detector 2 (rear position), slide the photomultiplier tube assembly back.

6. Remove the filter assembly by pulling it out of the photomultiplier tube assembly. Clean the filter assembly with lens paper.

Page 353: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-73

7. Insert the cleaned filter or a replacement filter with the colored side facing the photomultiplier tube assembly. The sulfur filter is purple, the phosphorus filter is yellow, the tin filter is orange.

8. Insert the filter end of the photomultiplier tube assembly into the detector body and secure it with the knurled nut.

Cleaning/Replacing the Detector Liner

To clean or replace the detector liner:

1. Turn off the photomultiplier voltage by selecting the following screen. Locate PMT Voltage on the screen and select OFF.

2. Extinguish the flame by turning off the hydrogen needle valve. (Turn the outer knob fully clockwise.)

Page 354: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-74

CAUTION

The photomultiplier tube can be damaged if exposed to light while voltage is ON. Never remove the detector cap or Photomultiplier tube WITH the voltage on.

3. Turn off the Clarus GC and allow the detector to cool.

4. Remove the cap (see the following figure).

5. Lift out the glass liner and O-ring. Clean the liner by washing it in a soap solution and/or solvent.

Figure 9-40 Removing a glass liner from the FPD

Page 355: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-75

6. Insert the glass liner and O-ring into the detector body, then replace the cap.

NOTE: Replace the glass liner (P/N N600-3057) if it is chipped or if it cannot be cleaned. Replace a worn O-ring (P/N 0990-2247).

Cleaning/Replacing the Detector Window

To clean or replace the detector window:

1. Turn off the photomultiplier voltage by selecting the following screen. Locate PMT Voltage on the screen and select OFF.

2. Extinguish the flame by turning off the hydrogen.

CAUTION

The photomultiplier tube can be damaged if exposed to light while voltage is ON. Never remove the detector cap or Photomultiplier tube WITH the voltage on.

Page 356: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-76

3. Turn off the Clarus GC and allow the detector to cool.

Removing the Window

1. Loosen the knurled nut that secures the photomultiplier tube assembly to the detector body.

2. Slide the photomultiplier tube assembly away from the detector body. If the FPD is detector 1 (front position), slide it to the left. If the FPD is detector 2 (rear position), slide it back.

3. Remove the cap (see the following figure).

4. Lift out the glass liner and O-ring from the detector body.

5. Remove the two hex head screws from the base of the FPD by using a 3/32-inch hex wrench. Then lift the detector body off the base.

6. Remove the spacer from the light pipe on the detector body.

7. Push the detector window out of the light pipe on the detector body (see the following figure). You can do this by inserting the end of a small pair of curved pliers into the top opening on the detector body. Then push the window out of the light pipe.

8. Remove the seal from the window.

Page 357: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-77

Figure 9-41 Removing a FPD detector window

Cleaning and Replacing the Window

1. Clean the window with soap and water. If necessary, replace the window and seal (P/N N930-0096).

2. Install the seal on the clean window.

3. Place the detector window on the window holder so that the spring portion of the seal faces the beveled edge in the window holder (see the following figure).

4. Use the spacer to insert the window into the light pipe on the detector body. Push the window and holder completely into the light pipe.

Page 358: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-78

Figure 9-42. Inserting a seal and window assembly into the FPD body

5. Insert the detector body on the detector base. Secure the detector body with the hex screws.

6. Replace the detector liner, O-ring, and detector cap on the detector body.

7. Connect the photomultiplier tube assembly to the light pipe on the detector body.

8. Turn on the detector heater. Then turn on the photomultiplier tube and light the flame.

Replacing the Photomultiplier Tube

To replace the photomultiplier tube:

1. Turn off the photomultiplier voltage by selecting the following screen. Locate PMT Voltage on the screen and select OFF.

Page 359: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-79

2. Extinguish the flame by turning off the hydrogen.

CAUTION

The photomultiplier tube can be damaged if exposed to light while voltage is ON. Never remove the detector cap or Photomultiplier tube WITH the voltage on.

3. Turn off the Clarus GC and allow the detector to cool.

4. Remove the two screws from the base of the photomultiplier assembly using a 7/64-inch hex wrench (see the following figure).

Page 360: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-80

Figure 9-43 Exploded view of a FPD photomultiplier tube

5. Pull the photomultiplier tube and base out of the housing.

6. Carefully unplug the photomultiplier tube from the base. Be careful not to dislodge the O-ring and plastic seal.

7. Align the key in the new photomultiplier tube with the key hole in the socket.

8. Plug the new photomultiplier tube (P/N 0997-2321) into the socket in the base.

Page 361: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-81

CAUTION

Do not force the key into an incorrect position. Do not touch the glass. Do not expose the photomultiplier tube to any bright, direct light

9. Position the photomultiplier tube in the housing so that the cathode (the horizontal filament grating; not visible in Figure 9-43) faces the detector body. Align the holes in the plastic seal with the holes in the base.

CAUTION

Make certain that the photomultiplier tube cathode is facing the detector body

10. Secure the base to the photomultiplier housing with the two screws removed in step 4 of this procedure.

11. Turn on the Clarus GC. Then turn on the hydrogen flow and the photomultiplier tube, and light the flame.

NOTE: A new photomultiplier tube may require several hours to stabilize.

CAUTION

When a replacement tube is installed, the noise should be about 2% at attenuation 8 with the linearizer off. If it is not, modify the voltage setting until this level of noise is achieved. Adjust the voltage as follows:

a. In the System Status screen select Tools for the configuration menu.

Page 362: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-82

b. In the Configuration screen select the FPD photomultiplier to view the PMT voltage.

Page 363: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-83

c. Using the touch pad increase the percent (%) value or to increase the voltage in 0.1% increments or decrease the voltage in 0.1% increments, by using the up or down arrows.

NOTE: The millivolt setting is displayed in the upper-right corner. This number will be close to zero and will only change slightly as the photomultiplier tube voltage is modified.

Cleaning/Replacing the FPD Jet

To clean or replace the FPD jet:

1. Turn off the photomultiplier voltage by selecting the following screen. Locate PMT Voltage on the screen and select OFF.

2. Extinguish the flame by turning off the hydrogen.

Page 364: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-84

CAUTION

The photomultiplier tube can be damaged if exposed to light while voltage is ON. Never remove the detector cap or Photomultiplier tube WITH the voltage on.

3. Turn off the Clarus GC and allow the detector to cool.

4. Remove the two screws from the base of the FPD by using a 3/32-inch hex wrench. Lift the detector body and the attached photomultiplier tube assembly off the detector base. This exposes the FPD jet (see the following figure).

5. Insert a 1/4-inch nutdriver (P/N N610-1297) over the FPD jet.

6. Turn the nutdriver counterclockwise until the jet assembly is completely loosened. Then remove the jet assembly from the detector base.

7. Clean the jet in a suitable solvent (such as methanol or acetone) and scrape off any deposits with a cotton swab on a piece of soft wood. DO NOT INSERT WIRE OR HARD TOOLS INTO THE JET. Replacement jets are available from PerkinElmer by ordering P/N N610-0245.

8. Insert the jet in the detector base and secure it with the 1/4-inch nutdriver.

9. Replace the detector body and the attached photomultiplier tube assembly.

10. Turn on the Clarus GC. Then turn on the hydrogen flow and the photomultiplier voltage, and light the flame.

Page 365: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-85

Figure 9-44 Cutaway view of a FPD showing the jet assembly

Page 366: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-86

PPC Maintenance

PPC maintenance consists of replacing a plugged or restricted PPC module restrictor.

Replacing a Restrictor

To replace a restrictor:

1. Locate the PPC module on the rear of the Clarus GC.

RestrictorAIR OR

TCD REF.

DET.2

HYD ORMAKE UP

CARR.2 CARR.1

Figure 9-45 Location of a PPC module restrictor

2. Turn off all gases to the module on which you are replacing the restrictor.

3. Using a 1/2-inch socket or open-end wrench, loosen and remove the restrictor and install a new restrictor.

4. Turn on the gas and check for leaks.

5. Calibrate the PPC module.

Zeroing the Pressure

The Clarus GC PPC modules are zeroed at the factory, prior to shipping, but it should be periodically checked to identify any change. This is particularly important if there have been large changes in the ambient temperature from

Page 367: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-87

22 ºC. For information about zeroing the pressure, refer to the “Clarus Users Guide.”

PPC Restrictor Information

Available Restrictors

Restrictor Number Color Helium Flow at 30 PSIG, vent to atmosphere

1 Silver 1 mL/min

2 Red 3 mL/min

3 Yellow 10 mL/min

4 Black 30 mL/min

5 Green 100 mL/min

6 Blue 300 mL/min

Page 368: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-88

Restrictors for Carrier Gas Control

Injector Restrictor Number

Helium Flow (mL/min) (with a 90 psig inlet pressure and a 10 psig drop across the restrictor)

Packed 4 (std) 30

Split/Splitless or PSS 6 (std) 300

Split/Splitless or PSS 4 (acc) 30

POC 3 (std) 10

POC 2 (acc) 3

Page 369: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-89

Restrictors for Detector Gas Control

NOTE: Maintain an inlet pressure between 60 and 90 psig.

Detector Gas Restrictor Number

Nominal Flow Required (mL/min)

FID Air 6 450

FID Hydrogen 4 45

NPD Air 5 100

NPD Hydrogen 1 2

FPD Air 5 90

FPD Hydrogen 4 75

ECD Argon/Methane 4 30

ECD Nitrogen 4 30

TCD Helium 4 30

TCD Hydrogen 3 30

Page 370: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-90

Practical Hints

Reversing TCD Polarity

The following examples indicate when you may want to change the TCD polarity.

• When one of the components being analyzed has a higher thermal conductivity than the carrier gas.

• For example, if hydrogen is a sample component and helium the carrier gas, set up a timed event to reverse the polarity of the TCD prior to elution of the hydrogen (to generate a positive peak for the hydrogen). Then change the polarity back for the remaining components.

• If negative peaks are produced when two packed injectors are installed, each has a different column attached, and you are running two different analyses. Change the polarity to produce positive peaks.

• When two packed injectors are installed with two identical columns and the TCD is being operated at maximum sensitivity. In this case, alternate the column into which the sample is being injected in order to expose both sets of filaments (reference and sample) to sample, thus keeping the filaments more electrically balanced.

• To change the TCD polarity, enter a negative detector range. For example, TCD ranges of +2 and –2 have opposite polarities.

Optimizing FID Performance

FID sensitivity is affected primarily by the hydrogen flow. The optimum hydrogen flow varies slightly if the column flow changes dramatically. For example, if you go from a packed column with a flow rate of 30 mL/min or higher to a capillary column with flows of 2 mL/min or less, the optimum hydrogen flow will be a different value.

Page 371: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-91

The hydrogen flows recommended in this manual assume packed column flow rates. If you switch from a packed to a capillary column, re-optimizing the hydrogen flow will help to improve the FID sensitivity.

The following is the suggested FID optimization procedure after you have switched from a packed to a capillary column.

1. Prepare a one component standard.

2. Set up the carrier gas flow.

3. Set up the hydrogen and air flows, then ignite the flame.

4. Make 2 to 3 injections at varying hydrogen flows.

The optimum hydrogen flow is that which produces the maximum area counts.

Filtering Detector Output

You can select one of three software filters from the Filter Menu. If your column delivers a peak width at half height of ≥1 s, select 200. If your column delivers a peak width at half height of <1 s, select 50.

A value of 800 should be used only with caution to minimize the signal-to-noise ratio. Try 800 with your application. If the peak height and area are not affected but the noise is decreased, then 800 will improve the signal-to-noise ratio.

NOTE: The ECD has values of 200 and 800 only.

Page 372: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-92

Autozero Display Sensitivity The maximum detector signals for various detectors that produce a 1-V reading on the Autozero Display are as follows: FID, NPD, or PID (nA, Range 1) – 21.3 FID, NPD, or PID (nA, Range 20) – 426 TCD (mV) – 510 ECD (KHz) – 150

Page 373: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-93

Attenuation vs. Detector Output The following table lists attenuation vs. the maximum detector signal producing 100% deflection on a 1-mV recorder.

Atten

NPD, PID and FID (fA)

Range 1

NPD, PID and FID (fA)

Range 20

TCD (µV)

ECD (Hz)

1 3.3x102 6.7x103 8.0 2.3

2 6.7x102 1.3x104 1.6x10 4.7

4 1.3x103 2.7x104 3.2x10 9.4

8 2.7x103 5.3x104 6.4x10 19.0

16 5.3x103 1.1x105 1.3x102 38.0

32 1.1x104 2.1x105 2.6x102 75.0

64 2.1x104 4.3x105 5.1x102 150.0

128 4.3x104 8.5x105 1x103 3.0x102

256 8.5x104 1.7x106 2x103 6.0x102

512 1.7x105 3.4x106 4.1x103 1.2x103

1024 3.4x105 6.8x106 8.2x103 2.4x103

2048 6.8x105 1.4x107 1.6x104 4.8x103

4096 1.4x106 2.7x107 3.3x104 9.6x103

8192 2.7x106 5.5x107 6.6x104 1.9x104

16384 5.4x106 1.1x108 1.3x105 3.8x104

32768 1.1x107 2.2x108 2.6x105 7.7x104

65536 2.2x107 4.4x108 5.2x105 1.5x105

Page 374: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-94

The following table lists attenuation vs. the maximum detector signal producing a 1-V output to an integrator.

Atten

NPD, PID and FID (pA)

Range 1

NPD, PID and FID (pA)

Range 20

TCD (mV)

ECD (kHz)

1 3.3x102 6.7x103 8.0 2.3

2 6.7x102 1.3x104 1.6x10 4.7

4 1.3x103 2.7x104 3.2x10 9.4

8 2.7x103 5.3x104 6.4x10 19.0

16 5.3x103 1.1x105 1.3x102 38.0

32 1.1x104 2.1x105 2.6x102 75.0

64 2.1x104 4.3x105 5.1x102 150.0

Prolonging the Life of a PID UV Lamp

Lamp life decreases at elevated temperatures. (For example, when conditioning the detector above 250 °C.) To increase lamp life, substitute a disc (P/N N0330-2989) for the UV lamp during such procedures.

Optimizing ECD Performance

ECDs are extremely sensitive. Therefore, care should be taken to avoid contamination from any part of the system (for example, pneumatics, injector, column, gases, etc.). To help assure a clean system: condition the column, bake out the injector and detector, use clean tubing, and use pure filtered gases.

CAUTION

To minimize detector contamination, run the ECD hot, at a temperature of at least 375 °C.

Page 375: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-95

Optimizing FPD Performance

About Ionization Gas Flows

FPD sensitivity is affected by both the hydrogen flow and the air flow. Hydrogen has a true optimum flow; that is, above or below a certain flow there is less response. Air, on the other hand, is more critical. That is, below a certain flow there is no response at all. The flows provided in the Clarus Users Guide are in the optimum range, but there may be slight variation from detector to detector.

Optimizing FPD Gas Flows

To optimize gas flows:

1. Set the hydrogen flow to 65 mL/min.

2. Set the air flow to 90 mL/min.

3. Set the detector temperature to 250 °C, and wait for the detector to reach that temperature.

4. Light the flame.

5. Prepare a one-component standard (sulfur, phosphorus, or tin, depending on the filter installed).

6. Make two or three injections.

7. Increase the hydrogen flow by 5 mL/min and repeat the injections.

8. Continue increasing the hydrogen, and running the sample, until a maximum signal is obtained. Set the hydrogen at that flow.

9. Repeat the above procedure, varying the air flow by 5 mL/min (both below and above 90 mL/min).

10. Set the air flow to at least 10 mL/min above the flow that previously produced no response.

Page 376: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-96

Quenching

When hydrocarbons co-elute with the sulfur compound (or phosphorus or tin compound depending on which filter is being used), they will “quench” the flame, thereby causing a diminished response, or no response at all, to the peak of interest. To avoid this problem, complete separation and elution of all peaks is necessary. Because the FPD is a selective detector, hydrocarbons may not show as significant peaks. We recommend that you first profile your sample by using the same column with either an FID or TCD. All peaks will then be seen and you can determine the necessary conditions to completely separate and elute all compounds.

PSS and POC Operating Hints

Both of the Programmed Split/Splitless (PSS) and Programmed On-Column (POC) injectors can be operated in either the oven programming mode or the inlet programming mode. For specific operation, refer to the Clarus 600 Users Guide.

Oven Programming Mode

This is the default mode for both the POC and the PSS injectors. This mode is the easiest to use, since only the oven temperature program needs to be entered into the method. In this mode, the injector will follow the oven temperature profile plus 5 °C. In this mode, the sample should be introduced into the injector when the temperature of the inlet is at the boiling point of the solvent. Then temperature program the injector and oven.

If the initial temperature of the oven is above the boiling point of the solvent you are using, then it would be better to modify the oven program to start at a lower temperature or to configure the injector for the Inlet mode and set a temperature for the injector separate from the oven temperature.

Inlet Programming Mode

This mode permits the use of independent injector temperatures and rates that you define in the method. The injector will be programmed for injector temperature 1, injector time 1, injector rate 1, injector temperature 2, injector

Page 377: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-97

time 2, etc. You can program up to three temperatures and two ramps for each PSS or POC configured in the inlet mode.

It is important to set the initial injector temperature to about the boiling point of the solvent you are using.

CAUTION

The PSS injector can be used in the "hot" split or splitless mode. However, this is not recommended due to liner size and could cause solvent flashback in the injector. This mode should be used with caution, depending upon the solvent and temperatures you choose.

CAUTION

When using the PSS injector in the on-column mode or the POC injector with the autosampler, you must use a special syringe that has a needle o.d. of 0.47 mm (P/N N610-1253 or N610-1380). Refer to the Clarus 600 User’s Guide for more detail. You must use only the "Norm" injection speed with this syringe in the on-column mode. The "Fast" injection speed will bend this thin needle and the "Slow" injection speed may produce peak break-up or distorted peaks. You can achieve better precision in the on-column mode when sample volumes of 1.0 µL or greater are injected.

Page 378: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-98

CAUTION

If you have the subambient option, the POC and PSS injectors are linked to the oven subambient option; therefore, you cannot operate the injectors below the oven subambient temperature.

If a column is used extensively at high temperatures (350 °C or greater), the polyimide may become very brittle. This brittleness will cause the column to fracture when you try to seal it in the universal adapter. If you wish to continue to use the brittle column, a low dead volume union may be a helpful alternative.

Page 379: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

9-99

PPC Restrictor Information

Available Restrictors

Restrictor Number Color Helium Flow at 30 PSIG, vent to atmosphere

1 Silver 1 mL/min

2 Red 3 mL/min

3 Yellow 10 mL/min

4 Black 30 mL/min

5 Green 100 mL/min

6 Blue 300 mL/min

Restrictors for Carrier Gas Control

Injector Restrictor Number/Color

Helium Flow (mL/min) (with a 90 psig inlet pressure and a 10 psig drop across the restrictor)

Packed 4 (std)/Black 30

Split/Splitless or PSS 6 (std)/Blue 300

Split/Splitless or PSS 4 (acc)/Black 30

POC 3 (std)/Yellow 10

POC 2 (acc)/Red 3

Page 380: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Maintenance

9-100

Restrictors for Detector Gas Control

NOTE: Maintain an inlet pressure between 60 and 90 psig.

Detector Gas Restrictor Number/Color

Nominal Flow Required (mL/min)

FID Air 6/Blue 450

FID Hydrogen 4/Black 45

NPD Air 5/Green 100

NPD Hydrogen 1/Silver 2

FPD Air 5/Green 90

FPD Hydrogen 4/Black 75

ECD Argon/Methane 4/Black 30

ECD Nitrogen 4/Black 30

TCD Helium 4/Black 30

TCD Hydrogen 3/Yellow 30

Page 381: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Troubleshooting 10

Page 382: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Troubleshooting

10-2

Page 383: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

10-3

This chapter contains the following sections:

• Instrument Error Messages

• Autosampler Error Messages

• Data Manager (DM) Error Messages

• GC Troubleshooting

Page 384: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Troubleshooting

10-4

Error Messages

The error messages are shown on the touch screen display. On a Warning message the user can touch OK and continue. On a Fatal Error there is no OK and the user must turn off the instrument and resolve the issue.

Table 10-1 Instrument Error Messages Error Messages

Timed Events Error Events table is full. 32 events maximum. Timed Events Error Duplicate event time. Enter a value at least 0.1 minute different. RS 232 Timeout Error. Device failed to respond within specified time frame. Oven Control Error No Coolant available. Check coolant supply. PPC Hardware Fault Power instrument OFF, wait 10 seconds and then power ON. If problem persists Call your PerkinElmer Service representative. Autosampler Error No vial found at location xx. Autosampler Error No autosampler programs active. Background Calibration Error Isothermal Method selected. Background calibration required temperature program method. Background Calibration Error Negative temperature ramp specified. Background calibration requires positive temperature ramps. Background Calibration Error Run Time is less than 1.0 minute.

Page 385: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

10-5

Background Calibration Error Run Time is greater than 999 minutes. Background Calibration Error HOLD Event is present in temperature program ramp. This is not permitted. No Inlet Configured Select Configuration on the Tools menu to access configuration settings. No Detector Configured Select Configuration on the Tools menu to access configuration settings. Flow Calibration Error Flow too high for ‘low’ calibration. Check flow and repeat calibration. Flow Calibration Error Flow too low for ‘hi’ calibration. Please check flow and repeat calibration. Autosampler Error Access door open. Close door. Autosampler Error No post wash specified. Set post-inject field in configuration to a minimum of 1. Relay Fault Turn power OFF for 10 seconds and then ON. If problem persists Call your PerkinElmer Service representative. Temperature Error Unable to reach specified temperature. Check coolant. PPC Error PPC Carrier xx not connected. Check PPC Carrier module connections.

Page 386: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Troubleshooting

10-6

PPC Error PPC Detector xx not connected. Check PPC Detector module connections. PPC Error The AUX flow PPC module being calibrated is not connected. Connect the AUX PPC flow module and calibrate. PPC Error Selected connection is currently assigned. Select another connection. PPC Error Warning: the PPC Alarm is OFF! If this is correct, press OK. To enable the PPC Alarm, select PPC Alarm ON in configuration setup. Ignite Error The detector xx failed to ignite. Check the hydrogen and air flows. Ignite Error The detector xx failed to ignite. The hydrogen and air supplies are OFF. Turn the gases ON.

Page 387: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

10-7

Table 10-2 Oven Error Messages. Oven Error Messages

INSTRUMENT SHUTDOWN! "CHECK OVEN FAN MOTOR" Actual oven cooling fan speed is out of tolerance (+/- 20% of set speed) for 20 seconds. Fatal error, call your PerkinElmer Service Center. INSTRUMENT SHUTDOWN! "CHECK STEPPER MOTOR" One or both of the 2 stepper motors for the oven doors (inlet/exhaust) stalled. This is determined by the firmware monitoring a PRT. Fatal error, call your PerkinElmer Service Center. INSTRUMENT SHUTDOWN! "CHECK OVEN FAN DRIVE" Oven cooling fan motor reports a generic fault. Fatal error, call your PerkinElmer Service Center. Temperature sensor fault OPEN. Cycle the power OFF for 10 seconds then ON. If the error persists call the PerkinElmer Service Center. Temperature sensor fault SHORT. Cycle the power OFF for 10 seconds then ON. If the error persists call the PerkinElmer Service Center. No Heat. Cycle the power OFF for 10 seconds then ON. If the error persists call the PerkinElmer Service Center. Temperature exceeds setpoint! Instrument has shut down. Cycle the power OFF for 10 seconds then ON. If the error persists call your PerkinElmer Service representative. OVEN NOT CONFIGURED "Press OK to continue" Oven is not configured (std/high power heater and line voltage). All instruments should have the oven configured before being shipped. Customer should not see this. Resolution must be by service. OVEN CAL FAILURE "Press OK to continue" Startup oven calibration procedure of inlet/exhaust doors failed. Warning A/S control error "Pre-wash Timeout" A sample pre-wash has taken place (reduced injection wait time) and system has not come ready within 30 minutes to inject. Warning

Page 388: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Troubleshooting

10-8

Table 10-3 Autosampler error messages. Autosampler Error Messages

Bad Sensor state

Bad motor mode

Bad motor speed

Elevator not initializing

Syringe error. Input value exceeds maximum volume of installed syringe.

Carousel not initializing

Tower not initializing

No W/W vial present

Invalid vial number

Invalid injection port

Slot not found firmware to clarify Tower location sensor not finding slot.

No vial at position xx.

Invalid syringe size

Encoder read error (reserved)

Encoder position err

Bar Code error (reserved)

Encoder position err

Bar Code error (reserved)

Vial Sensor not initializing

Phase error

Reserved for barcode reader

Bad slot firmware to clarify Tower slot is wrong width.

Bad volume firmware to clarify Incorrect volume for syringe.

Bad injection firmware to clarify Injection failed.

Page 389: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

10-9

Bad motor id firmware to clarify Tried to move invalid motor.

Illegal command firmware to clarify Illegal command to Autosampler

Table 10-4 –Formatted error messages.

Error Messages

Cannot delete oven program step <step number> Cannot delete inlet program step <step number> Invalid Method<method number> in sequence Carrier <carrier number> unable to maintain pressure Split leaking flow above max FID flame out Touch OK to continue Warning: Car<inlet number> Inlet < 50 Psi, Touch OK to continue

Page 390: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Troubleshooting

10-10

Table 4 Data Manager (DM) error messages.

The DM Errors consist of a DM number and an associated index. The following table lists the errors, indices and gives a brief explanation of the error. Number Index Explanation

1 1 Autosampler Resume Validation error 2 1 Autosampler Program Validation error 3 1 Background Compensation Validation error 4 PPC Channel

0 - 11 PPC Calibration Mode Validation error

5 PPC Channel 0 - 11

PPC Calibration Maximum Validation error

6 PPC Channel 0 - 11

PPC Calibration Minimum Validation error

7 11 Configuration Validation error 8 1 Cryo Configuration Validation error 9 1 Extended Configuration Validation error

Valves 5 and 6, FPD PMT % voltages 10 1 PPC Configuration Validation error 11 1 Time Configuration Validation error 12 1 Autosampler Control Validation error 13 1 Flame Off Validation error 14 1 Flow Calibration Validation error 15 1 GC Serial Number Validation error 16 1 GC Tune Validation error 17 1 GC Tune Flag Validation error 18 1 GC Tune Mode Validation error 19 1 GC Tune Valid Flag Validation error 20 1 Heat Delay Validation error

Page 391: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

10-11

21 1 HS 40 Mode Validation error 22 1 PPC Hardware Configuration Validation error 23 1 PPC Map Validation error 24 1 PPC Limits Validation error 25 1 Maxget Validation error 26 Method index

0 - 4 Method Extension Validation error Valves 5 and 6, PSS and POC Inlet 1 and 2 programs

27 Method index 0 - 4

Method Validation error

28 Method index 0 - 4

PPC Method Validation error

29 1 Oven Calibration Validation error 30 1 Priority Vial Cycles Validation error 31 1 Data Collection Rate Validation error 32 1 Sleep Validation error 33 1 Split Autozero Validation error 34 1 Start Structure Validation error 35 1 PPC References Validation error 36 1 TCD Protection Validation error 37 1 Valve Validation error 38 1 PPC Zero Offset Validation error

Page 392: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Troubleshooting

10-12

GC Troubleshooting

The cardinal rule in troubleshooting your gas chromatograph is "If it ain't broke, don't fix it." When things are working fine, leave well enough alone, but when problems occur, this section will help you identify what could be wrong and how to solve your problem.

There are several sources of problems in gas chromatography:

The operator: When the operator is new to chromatography and/or a new instrument, problems can be introduced during the learning curve. Once the operator becomes familiar with both the technique and the instrument, this problem source diminishes greatly.

The sample: Unlike clean standards, real world samples such as environmental samples, can introduce problems because they are difficult to handle, have complicated matrices, contain unknown constituents, etc..

The column: The column is most often the major factor contributing to poor analyses. The more a column is used, the greater the possibility of contamination, loss of substrate, etc. Columns do not last forever and should be changed when results become suspect.

The gas flow system: Leaks are a major concern in gas chromatography and can lead to many problems.

The electronics: The problem must be identified as either chromatographic or hardware. Electronics used in the system can malfunction.

Data handling: Today, most chromatographers rely on sophisticated data handling systems to integrate their results. Some problems can be related to the incorrect setting of data handling parameters.

Page 393: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

10-13

Spare Components.

Following is a list of items you should have on hand to help solve problems.

New syringes – a syringe can break, become plugged or begin leaking. Always have spares available.

Duplicate columns – a column does not last forever; therefore, a duplicate column should be on hand in the event that your separation begins to degrade. Also, capillary columns can be damaged if oxygen is introduced at high temperatures. A duplicate column will allow you to identify if the column is the cause of the problem.

Septa – this is the one area of the gas chromatograph which requires routine maintenance. Always have spare septa available.

Leak detector – the gas flow system can be a problem as fittings wear with age and can begin to leak. A leak detector should be available to help find and fix leaks.

Injector liners – are made of glass and can be easily broken when removed. A supply of spare liners should therefore be kept on hand. Please remember that you cannot run satisfactory analyses without an injector liner.

Logical Troubleshooting Steps.

There are some simple steps that should be taken when trying to locate the problem. Use the following guide to troubleshoot your GC.

Note the symptoms - define the problem. Compare your runs with good analysis, that is, with the results normally obtained.

Systematically eliminate possible causes.

The first rule here is, "What did you change last?" Many times a problem arises when a change is made to the system, such as changing a gas tank, septum, or glass liner. If the problem occurred after such a change, then the change is the most likely cause of the problem.

Page 394: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Troubleshooting

10-14

Change the simplest thing first. For example, if you suspect a gas leak, the easiest change to make is the GC septum instead of replumbing the internal pneumatics.

Change only one GC parameter at a time and check for its effect. If you change three items at once and your problem goes away, you may not know which of the three moves or combination of moves corrected the problem. This way, if the problem happens again, you will know exactly what corrective action to take.

Dual Identical Channels Only

If your GC is a dual-channel system (dual identical detectors and dual identical injectors):

1. Try switching the column to the second channel. If the problem is corrected, then the problem was caused by the detector, injector, amplifier, or the pneumatics.

2. Replace each of the above components one at a time to identify which one is defective. If the problem is the same as before you switched the column, you should suspect the column, syringe, standard or sample, electronics, or data handling device.

Page 395: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I U.S. Nuclear Regulations I

NOTE: All USNRC regulations can be obtained through the internet at www.nrc.gov/reading-rm/

Page 396: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-2

Page 397: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-3

Appendix I - SUPPLEMENT 2 Agreement States Alabama Kirksey E. Whatley, Director Office of Radiation Control The Alabama Department of Public Health The RSA Tower, Suite 700 P.O. Box 303017 Montgomery, AL 36130-3017 PH (334)206-5391 FX (334)206-5387 INET: [email protected] Arizona Aubrey V. Godwin, Director Arizona Radiation Regulatory Agency 4814 South 40th Street Phoenix, AZ 85040 PH (602)255-4845 ext. 222 FX (602)437-0705 INET: [email protected] Arkansas Jared W. Thompson, Program Leader Division of Radiation Control & Emergency Mgmt Radioactive Materials Program, Department of Health Freeway Medical, Suite 100 5800 West 10th Street Little Rock, AR 72204-1755 PH (501)661-2108 FX (501)661-2468 INET: [email protected] California Edgar D. Bailey, C.H.P., Chief Radiologic Health Branch Division of Food, Drugs, and Radiation Safety California Department of Health Services P.O. Box 942732 Sacramento, CA 94234-7320

PH (916)322-3482 FX (916)324-3610 INET: [email protected] Colorado Warren E. (Jake) Jacobi, Program Manager Laboratory & Radiation Services Division Colorado Department of Public Health & Environment 8100 Lowry Boulevard Denver, CO 80230-6928 PH (303)692-3036 FX (303)692-3692 INET: INET: [email protected] Florida William A. Passetti, Chief Bureau of Radiation Control Florida Department of Health 4052 Bald Cypress Way, SE, Bin C21 Tallahassee, FL 32399-1741 PH (850)245-4266 FX (850)487-0435 INET: [email protected] Georgia Thomas E. Hill, Manager Radioactive Materials Program Department of Natural Resources 4244 International Parkway, Suite 114 Atlanta, GA 30354 PH (404)362-2675 FX (404)362-2653 INET: [email protected] Illinois Thomas W. Ortciger, Director Illinois Department of Nuclear Safety 1035 Outer Park Drive Springfield, IL 62704 PH (217)785-9868 FX (217)524-4724 INET: [email protected] Iowa Donald A. Flater, Chief Bureau of Radiological Health Iowa Department of Public Health

Page 398: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-4

401 SW 7th Street, Suite D Des Moines, IA 50309 PH (515)281-3478 FX (515)725-0318 INET: [email protected] Kansas Victor L. Cooper, Section Chief Air Operating Permit & Compliance Section Bureau of Air & Radiation Division of Environment Kansas Department of Health & Environment 1000 SW Jackson, Suite 310 Topeka, KS 66612-1366 PH (785)296-1561 FX (785)291-3953 INET: [email protected] Kentucky John A. Volpe, Ph.D., Manager Radiation Health & Toxic Agents Branch Cabinet for Health Services 275 East Main Street Frankfort, KY 40621-0001 PH (502)564-7818 ext 3692 FX (502)564-1492 INET: [email protected] Louisiana Michael Henry, Senior Environmental Scientist, Permitting Division Department of Environmental Quality Office of Environmental Services Permits Division 7290 Bluebonnet Road Baton Rouge, LA 70884-2135 PH (225)765-0892 FX (225)765-0222 INET: [email protected] Maine

Jay Hyland, Program Manager Radiation Control Program Division of Health Engineering 10 State House Station Augusta, ME 04333

PH (207)287-5677 FX (207)287-3059 INET: [email protected]

Maryland Roland G. Fletcher, Manager Radiological Health Program Air and Radiation Management Administration Maryland Department of the Environment 2500 Broening Highway Baltimore, MD 21224 PH (410)631-3300 FX (410)631-3198 INET: [email protected] Massachusetts Robert M. Hallisey, Director Radiation Control Program Department of Public Health 174 Portland Street, 5th Floor Boston, MA 02114 PH (617)727-6214 FX (617)727-2098 INET: [email protected] Minnesota Linda Bruemmer, Manager Section of Asbestos, Indoor Air, Lead and Radiation Division of Environmental Health Department of Health 121 E. Seventh Place, Suite 220 P.O. Box 64975 St. Paul, MN 55164-0975 PH (651)215-0945 FX (651)215-0975 INET: [email protected] Mississippi Robert W. Goff, Director Division of Radiological Health State Department of Health 3150 Lawson Street, P.O. Box 1700 Jackson, MS 39215-1700 PH (601)987-6893 FX (601)987-6887 INET: [email protected]

Page 399: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-5

Nebraska Dick Nelson, Director Department of Regulation and Licensure Nebraska Health and Human Services System P.O. Box 95007 Lincoln, NE 68590-5007 PH (402)471-8566 FX (402)471-9449 INET: [email protected] New Hampshire Diane E. Tefft, Administrator Radiological Health Bureau Division of Public Health Services Health and Welfare Building 6 Hazen Drive Concord, NH 03301-6527 PH (603)271-4588 FX (603)225-2325 INET: [email protected] Nevada Stanley R. Marshall, Supervisor Radiological Health Section Health Division Department of Human Resources 1179 Fairview Drive, Suite 102 Carson City, NV 89701-5405 PH (775)687-5394 ext. 276, FX (775)687-5751 INET: [email protected] New Mexico William Floyd, Manager Radiation Control Bureau Field Operations Division Environment Department 1190 St. Francis Drive P.O. Box 26110 Santa Fe, NM 87502 PH (505)476-3236 FX (505)476-3232 INET: [email protected]

New York Clayton Bradt, Principal Radiophysicist New York State Dept. of Labor Radiological Health Unit Building 12, Room 169 State Office Building Campus Albany, NY 12240 PH (518)457-1202 FX (518)485-7406 INET: [email protected]

John P. Spath, Director Radioactive Waste Policy and Nuclear Coordination New York State Energy Research & Development Authority Corporate Plaza West 286 Washington Avenue Extension Albany, NY 12203-6399 PH (518)862-1090 ext.3302 FX (518)862-1091 INET: [email protected]

Paul J. Merges, Ph.D., Director, Bureau of Radiation and Hazardous Site Management New York State Department of Environmental Conservation 625 Broadway Albany, NY 12233-7255 PH (518)402-8605 FX (518)402-9025 INET: [email protected]

Karim Rimawi, Ph.D., Director Bureau of Environmental Radiation Protection New York State Department of Health 547 River Street Troy, NY 12203 PH (518)402-7590 FX (518)402-7554 INET: [email protected]

Gene Miskin, Director Bureau of Radiological Health New York City Department of Health Two Lafayette Street, 11th Floor New York, NY 10007 PH (212)676-1556 FX (212)676-1548 INET: [email protected]

Page 400: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-6

North Carolina Beverly O. Hall, Acting Director Division of Radiation Protection Department of Environment & Natural Resources 3825 Barrett Drive Raleigh, NC 27609-7221 PH (919)571-4141 FX (919)571-4148 INET: [email protected] North Dakota Terry L. O'Clair, Director Division of Air Quality North Dakota Department of Health 1200 Missouri Avenue P.O. Box 5520 Bismarck, ND 58506-5520 PH (701)328-5188 FX (701)328-5200 INET: [email protected] Ohio Roger L. Suppes, Chief Bureau of Radiation Protection Ohio Department of Health 35 East Chestnut Street Columbus, OH 43266 PH (614)644-7860 FX (614)466-0381 INET: [email protected] Oklahoma Mike Broderick, Environmental Program Administrator Radiation Management Section Oklahoma Department of Environmental Quality P.O. Box 1677 Oklahoma City, OK 73101-1677 PH (405)702-5155 FX (405)702-5101 INET: [email protected] Oregon Terry D. Lindsey, Acting Manager Oregon Radiation Protection Services

Section 800 N.E. Oregon Street, Suite 260 Portland, OR 97232 PH (503)731-4014 ext. 660 FX (503)731-4081 INET: [email protected] Pennsylvania David Allard, CHP, Director Bureau of Radiation Protection Department of Environmental Protection Rachel Carson State Office Building P.O. Box 8469 Harrisburg, PA 17105-8469 PH (717)787-2480 FX (717)783-8965 INET: [email protected] Rhode Island Marie Stoeckel, Chief Division of Occupational & Radiological Health Department of Health 3 Capitol Hill, Room 206 Providence, RI 02908-5097 PH (401)222-2438 FX (401)222-2456 INET: [email protected] South Carolina T. Pearce O'Kelley, Chief Bureau of Radiological Health Department of Health & Environmental Control 2600 Bull Street Columbia, SC 29201 PH (803)545-4400 FX (803)545-4412 INET: [email protected] Henry Porter, Assistant Director Division of Waste Management Bureau of Land and Waste Management Department of Health & Environmental Control 2600 Bull Street Columbia, SC 29201

Page 401: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-7

PH (803)896-4245 FX (803)896-4242 INET: [email protected] Tennessee L. Edward Nanney, Director Division of Radiological Health Tennessee Department of Environment and Conservation L&C Annex, Third Floor 401 Church Street Nashville, TN 37243-1532 PH (615)532-0360 FX (615)532-7938 INET: [email protected] Texas Richard A. Ratliff, P.E., L.M.P. Chief Bureau of Radiation Control Texas Department of Health 1100 West 49th Street Austin, TX 78756-3189 PH (512)834-6679 FX (512)834-6708 INET: [email protected]

Susan Jablonski Health Physicist and Technical Advisor Office of Permitting, Remediation & Registration Texas Natural Resource Conservation Commission P.O. Box 13087, MS 122 Austin, TX 78711-3087 PH (512)239-6731 FX (512)239-5151 INET: [email protected]

Utah William J. Sinclair, Director Division of Radiation Control Department of Environmental Quality 168 North 1950 West P.O. Box 144850 Salt Lake City, UT 84114-4850 PH (801)536-4250 FX (801)533-4097 INET: [email protected]

Washington John L. Erickson, Director Division of Radiation Protection Department of Health Building #5 P.O. Box 47827 7171 Cleanwater Lane Olympia, WA 98504-7827 PH (360)236-3210 FX (360)236-2255 INET: [email protected] Wisconsin Paul Schmidt, Manager Radiation Protection Unit Bureau of Public Health Department of Health and Family Services P.O. Box 309 Madison, WI 53701-0309 PH (608)267-4792 FX (608)267-4799 INET: [email protected]

Page 402: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-8

U.S. Nuclear Regulatory Commission Regional Offices

REGION ADDRESS TELEPHONE

I WPI's Region

U.S. Nuclear Regulatory Commission, Region I

475 Allendale Road King of Prussia, PA 19406-1415

(800) 432-1156

II U.S. Nuclear Regulatory Commission, Region II

101 Marietta St., N.W., Suite 2900 Atlanta, GA 30323-0199

(800) 577-8510

III U.S. Nuclear Regulatory Commission, Region III 801 Warrenville Road Lisle, IL 60137-5927

(800) 522-3025

IV U.S. Nuclear Regulatory Commission, Region IV

611 Ryan Plaza Drive, Suite 400 Arlington, TX 76011-8064

(800) 952-9677

Walnut Creek Field Office

U.S. Nuclear Regulatory Commission 1450 Maria Lane

Walnut Creek, CA 94596-5368

(800) 882-4672

Page 403: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-9

Nuclear Regulatory Commission Regulations

The following NRC regulations are from Title 10 Energy in the Code of Federal Regulations revised as of June 30, 1996. Subpart M-Reports Source: 56 FR 23406, May 21, 1991, unless otherwise noted. § 20.2201 Reports of theft or loss of licensed material. (a) Telephone reports. (1) Each licensee shall report by telephone as follows: (i) Immediately after its occurrence becomes known to the licensee, any lost, stolen, or missing licensed material in an aggregate quantity equal to or greater than 1,000 times the quantity specified in Appendix C to Part 20 under such circumstances that it appears to the licensee that an exposure could result to persons in unrestricted areas; or (ii) Within 30 days after the occurrence of any lost, stolen, or missing licensed material becomes known to the licensee, all licensed material in a quantity greater than 10 times the quantity specified in Appendix C to Part 20 that is still missing at this time. (2) Reports must be made as follows: (i) Licensees having an installed Emergency Notification System shall make the reports to the NRC Operations Center in accordance with § 50.72 of this chapter, and (ii) All other licensees shall make reports by telephone to the NRC Operations Center (301-951-0550). (b) Written reports. (1) Each licensee required to make a report under paragraph (a) of this section shall, within 30 days after making the telephone report, make a written report setting forth the following information: (i) A description of the licensed material involved, including kind, quantity, and chemical and physical form; and (ii) A description of the circumstances under which the loss or theft occurred; and (iii) A statement of disposition, or probable disposition, of the licensed material involved; and (iv) Exposures of individuals to radiation, circumstances under which the exposures occurred, and the possible total effective dose equivalent to persons in unrestricted areas; and (v) Actions that have been taken, or will be taken, to recover the material; and (vi) Procedures or measures that have been, or will be, adopted to ensure against a recurrence of the loss or theft of licensed material. (2) Reports must be made as follows: (i) For holders of an operating license for a nuclear power plant, the events included in paragraph (b) of this section must be reported in accordance with the procedures

Page 404: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-10

described in § 50.73(b), (c), (d), (e), and (g) of this chapter and must include the information required in paragraph (b)(1) of this section, and (ii) All other licensees shall make reports to the Administrator of the appropriate NRC Regional Office listed in appendix D to Part 20. (c) A duplicate report is not required under paragraph (b) of this section if the licensee is also required to submit a report pursuant to §§ 30.55(c), 40.64(c), 50.72, 50.73, 70.52, 73.27(b), 73.67(e)(3)(vi), 73.67(g)(3)(iii), 73.71, or 150.19(c) of this chapter. (d) Subsequent to filing the written report, the licensee shall also report any additional substantive information on the loss or theft within 30 days after the licensee learns of such information. (e) The licensee shall prepare any report filed with the Commission pursuant to this section so that names of individuals who may have received exposure to radiation are stated in a separate and detachable part of the report. [56 FR 23406, May 21, 1991, as amended at 58 FR 69220, Dec. 30, 1993; 60 FR 20183, Apr. 25, 1995]

§ 20.2202 Notification of incidents. (a) Immediate notification. Notwithstanding any other requirements for notification, each licensee shall immediately report any event involving byproduct, source, or special nuclear material possessed by the licensee that may have caused or threatens to cause any of the following conditions- (1) An individual to receive- (i) A total effective dose equivalent of 25 rems (0.25 Sv) or more; or (ii) An eye dose equivalent of 75 rems (0.75 Sv) or more; or (iii) A shallow-dose equivalent to the skin or extremities of 250 rads (2.5 Gy) or more; or (2) The release of radioactive material, inside or outside of a restricted area, so that, had an individual been present for 24 hours, the individual could have received an intake five times the annual limit on intake (the provisions of this paragraph do not apply to locations where personnel are not normally stationed during routine operations, such as hot-cells or process enclosures). (b) Twenty-four hour notification. Each licensee shall, within 24 hours of discovery of the event, report any event involving loss of control of licensed material possessed by the licensee that may have caused, or threatens to cause, any of the following conditions: (1) An individual to receive, in a period of 24 hours- (i) A total effective dose equivalent exceeding 5 rems (0.05 Sv); or (ii) An eye dose equivalent exceeding 15 rems (0.15 Sv); or

Page 405: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-11

(iii) A shallow-dose equivalent to the skin or extremities exceeding 50 rems (0.5 Sv); or (2) The release of radioactive material, inside or outside of a restricted area, so that, had an individual been present for 24 hours, the individual could have received an intake in excess of one occupational annual limit on intake (the provisions of this paragraph do not apply to locations where personnel are not normally stationed during routine operations, such as hot-cells or process enclosures). (c) The licensee shall prepare any report filed with the Commission pursuant to this section so that names of individuals who have received exposure to radiation or radioactive material are stated in a separate and detachable part of the report. (d) Reports made by licensees in response to the requirements of this section must be made as follows: (1) Licensees having an installed Emergency Notification System shall make the reports required by paragraphs (a) and (b) of this section to the NRC Operations Center in accordance with 10 CFR 50.72; and (2) All other licensees shall make the reports required by paragraph (a) and (b) of this section by telephone to the NRC Operations Center (301) 816-5100 and by telegram, mailgram, or facsimile to the Administrator of the appropriate NRC Regional Office listed in appendix D to 10 CFR part 20. (e) The provisions of this section do not include doses that result from planned special exposures, that are within the limits for planned special exposures, and that are reported under § 20.2204. [56 FR 23406, May 21, 1991, as amended at 56 FR 40766, Aug. 16, 1991; 57 FR 57879, Dec. 8, 1992; 59 FR 14086, Mar. 25, 1994] § 20.2203 Reports of exposures, radiation levels, and concentrations of radioactive material exceeding the limits. a) Reportable events. In addition to the notification required by §20.2202, each licensee shall submit a written report within 30 days after learning of any of the following occurrences: (1) Any incident for which notification is required by §20.2202; or (2) Doses in excess of any of the following:

(i) The occupational dose limits for adults in §20.1201; or (ii) The occupational dose limits for a minor in §20.1207; or (iii) The limits for an embryo/fetus of a declared pregnant woman in §20.1208; or (iv) The limits for an individual member of the public in §20.1301; or (v) Any applicable limit in the license; or

Page 406: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-12

(vi) The ALARA constraints for air emissions established under §20.1101(d); or

(3) Levels of radiation or concentrations of radioactive material in -- (i) A restricted area in excess of any applicable limit in the license; or (ii) An unrestricted area in excess of 10 times any applicable limit set forth in this part or in the license (whether or not involving exposure of any individual in excess of the limits in §20.1301); or

(4) For licensees subject to the provisions of EPA's generally applicable environmental radiation standards in 40 CFR part 190, levels of radiation or releases of radioactive material in excess of those standards, or of license conditions related to those standards. (b) Contents of reports. (1) Each report required by paragraph (a) of this section must describe the extent of exposure of individuals to radiation and radioactive material, including, as appropriate:

(i) Estimates of each individual's dose; and (ii) The levels of radiation and concentrations of radioactive material involved; and (iii) The cause of the elevated exposures, dose rates, or concentrations; and (iv) Corrective steps taken or planned to ensure against a recurrence, including the schedule for achieving conformance with applicable limits, ALARA constraints, generally applicable environmental standards, and associated license conditions.

(2) Each report filed pursuant to paragraph (a) of this section must include for each occupationally overexposed(7) individual: the name, Social Security account number, and date of birth. The report must be prepared so that this information is stated in a separate and detachable part of the report. (c) For holders of an operating license for a nuclear power plant, the occurrences included in paragraph (a) of this section must be reported in accordance with the procedures described in §50.73(b), (c), (d), (e), and (g) of this chapter and must also include the information required by paragraph (b) of this section. Occurrences reported in accordance with §50.73 of this chapter need not be reported by a duplicate report under paragraph (a) of this section. (d) All licensees, other than those holding an operating license for a nuclear power plant, who make reports under paragraph (a) of this section shall submit the report in writing to the U.S. Nuclear Regulatory Commission, Document Control Desk, Washington, DC 20555, with a copy to the appropriate NRC Regional Office listed in appendix D to part 20.

Page 407: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-13

[56 FR 23406, May 21, 1991, as amended at 60 FR 20186, Apr. 25, 1995; 61 FR 65127, Dec. 10, 1996] 7 With respect to the limit for the embryo-fetus (§20.1208), the identifiers should be those of the declared pregnant woman. §20.2204 Reports of planned special exposures. The licensee shall submit a written report to the Administrator of the appropriate NRC Regional Office listed in appendix D to Part 20 within 30 days following any planned special exposure conducted in accordance with § 20.1206, informing the Commission that a planned special exposure was conducted and indicating the date the planned special exposure occurred and the information required by § 20.2105. [Amended 60 FR 20183, Apr. 25, 1995] § 20.2205 Reports to individuals of exceeding dose limits. When a licensee is required, pursuant to the provisions of §§20.2203, 20.2204, or 20.2206, to report to the Commission any exposure of an identified occupationally exposed individual, or an identified member of the public, to radiation or radioactive material, the licensee shall also provide a copy of the report submitted to the Commission to the individual. This report must be transmitted at a time no later than the transmittal to the Commission. [60 FR 36043, July 13, 1995] §20.2206 Reports of individual monitoring. (a) This section applies to each person licensed by the Commission to- (1) Operate a nuclear reactor designed to produce electrical or heat energy pursuant to §50.21(b) or §50.22 of this chapter or a testing facility as defined in §50.2 of this chapter; or (2) Possess or use byproduct material for purposes of radiography pursuant to Parts 30 and 34 of this chapter; or (3) Possess or use at any one time, for purposes of fuel processing, fabricating, or reprocessing, special nuclear material in a quantity exceeding 5,000 grams of contained uranium-235, uranium-233, or plutonium, or any combination thereof pursuant to part 70 of this chapter; or (4) Possess high-level radioactive waste at a geologic repository operations area pursuant to part 60 of this chapter; or (5) Possess spent fuel in an independent spent fuel storage installation (ISFSI) pursuant to part 72 of this chapter; or

Page 408: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-14

(6) Receive radioactive waste from other persons for disposal under part 61 of this chapter; or (7) Possess or use at any time, for processing or manufacturing for distribution pursuant to parts 30, 32, 33 or 35 of this chapter, byproduct material in quantities exceeding any one of the following quantities:

Radionuclide Quantity of radionuclide1 in curies Cesium-137 1 Cobalt-60 1 Gold-198 100 Iodine-131 1 Iridium-192 10 Krypton-85 1,000 Promethium-147 10 Techetium-99m 1,000

1 The Commission may require as a license condition, or by rule, regulation, or order pursuant to § 20.2302, reports from licensees who are licensed to use radionuclides not on this list, in quantities sufficient to cause comparable radiation levels. (b) Each licensee in a category listed in paragraph (a) of this section shall submit an annual report of the results of individual monitoring carried out by the licensee for each individual for whom monitoring was required by § 20.1502 during that year. The licensee may include additional data for individuals for whom monitoring was provided but not required. The licensee shall use Form NRC 5 or electronic media containing all the information required by Form NRC 5. (c) The licensee shall file the report required by § 20.2206(b), covering the preceding year, on or before April 30 of each year. The licensee shall submit the report to the REIRS Project Manager, Office of Nuclear Regulatory Research, U.S. Nuclear Regulatory Commission, Washington, DC 20555. [56 FR 23406, May 21, 1991, as amended at 56 FR 32072, July 15, 1991] § 30.34 Terms and conditions of licenses. (a) Each license issued pursuant to the regulations in this part and the regulations in parts 31 through 36 and 39 of this chapter shall be subject to all the provisions of the Act, now or hereafter in effect, and to all valid rules, regulations and orders of the Commission. (b) No license issued or granted pursuant to the regulations in this part and parts 31 through 36, and 39 nor any right under a license shall be transferred, assigned or in any manner disposed of, either voluntarily or involuntarily, directly or indirectly, through transfer of control of any license to any person, unless the Commission shall, after

Page 409: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-15

securing full information, find that the transfer is in accordance with the provisions of the Act and shall give its consent in writing. (c) Each person licensed by the Commission pursuant to the regulations in this part and parts 31 through 36 and 39 shall confine his possession and use of the byproduct material to the locations and purposes authorized in the license. Except as otherwise provided in the license, a license issued pursuant to the regulations in this part and parts 31 through 36 and 39 of this chapter shall carry with it the right to receive, acquire, own, and possess byproduct material. Preparation for shipment and transport of byproduct material shall be in accordance with the provisions of part 71 of this chapter. (d) Each license issued pursuant to the regulations in this part and parts 31 through 36 and 39 shall be deemed to contain the provisions set forth in section 183b. - d., inclusive, of the Act, whether or not these provisions are expressly set forth in the license. (e) The Commission may incorporate, in any license issued pursuant to the regulations in this part and parts 31 through 36 and 39, at the time of issuance, or thereafter by appropriate rule, regulation or order, such additional requirements and conditions with respect to the licensee's receipt, possession, use and transfer of byproduct material as it deems appropriate or necessary in order to: (1) Promote the common defense and security; (2) Protect health or to minimize danger to life or property; (3) Protect restricted data; (4) Require such reports and the keeping of such records, and to provide for such inspections of activities under the license as may be necessary or appropriate to effectuate the purposes of the Act and regulations thereunder. (f) Licensees required to submit emergency plans by §30.32(i) shall follow the emergency plan approved by the Commission. The licensee may change the approved without Commission approval only if the changes do not decrease the effectiveness of the plan. The licensee shall furnish the change to the appropriate NRC Regional Office specified in §30.6 and to affected offsite response organizations within six months after the change is made. Proposed changes that decrease, or potentially decrease, the effectiveness of the approved emergency plan may not be implemented without prior application to and prior approval by the Commission. (g) Each licensee preparing technetium-99m radiopharmaceuticals from molybdenum-99/technetium-99m generators shall test the generator eluates for molybdenum-99 breakthrough in accordance with §35.204 of this chapter. The licensee shall record the results of each test and retain each record for three years after the record is made. (h)(1) Each general licensee that is required to register by Sec. 31.5(c)(13) of this chapter and each specific licensee shall notify the appropriate NRC Regional Administrator, in writing, immediately following the filing of a voluntary or involuntary petition for bankruptcy under any chapter of title 11 (Bankruptcy) of the United States Code by or against: (i) The licensee; (ii) An entity (as that term is defined in 11 U.S.C. 101(14)) controlling the licensee or listing the license or licensee as property of the estate; or

Page 410: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-16

(iii) An affiliate (as that term is defined in 11 U.S.C. 101(2)) of the licensee. (2) This notification must indicate: (i) The bankruptcy court in which the petition for bankruptcy was filed; and (ii) The date of the filing of the petition. § 30.35 Financial assurance and record keeping for decommissioning. a) Each applicant for a specific license authorizing the possession and use of unsealed byproduct material of half-life greater than 120 days and in quantities exceeding 105 times the applicable quantities set forth in appendix B to part 30 shall submit a decommissioning funding plan as described in paragraph (e) of this section. The decommissioning funding plan must also be submitted when a combination of isotopes is involved if R divided by 105 is greater than 1 (unity rule), where R is defined here as the sum of the ratios of the quantity of each isotope to the applicable value in appendix B to part 30. (b) Each applicant for a specific license authorizing possession and use of byproduct material of half-life greater than 120 days and in quantities specified in paragraph (d) of this section shall either -- (1) Submit a decommissioning funding plan as described in paragraph (e) of this section; or (2) Submit a certification that financial assurance for decommissioning has been provided in the amount prescribed by paragraph (d) of this section using one of the methods described in paragraph (f) of this section. For an applicant, this certification may state that the appropriate assurance will be obtained after the application has been approved and the license issued but before the receipt of licensed material. If the applicant defers execution of the financial instrument until after the license has been issued, a signed original of the financial instrument obtained to satisfy the requirements of paragraph (f) of this section must be submitted to NRC before receipt of licensed material. If the applicant does not defer execution of the financial instrument, the applicant shall submit to NRC, as part of the certification, a signed original of the financial instrument obtained to satisfy the requirements of paragraph (f) of this section. (c)(1) Each holder of a specific license issued on or after July 27, 1990, which is of a type described in paragraph (a) or (b) of this section, shall provide financial assurance for decommissioning in accordance with the criteria set forth in this section. (2) Each holder of a specific license issued before July 27, 1990, and of a type described in paragraph (a) of this section shall submit, on or before July 27, 1990, a decommissioning funding plan as described in paragraph (e) of this section or a certification of financial assurance for decommissioning in an amount at least equal to $750,000 in accordance with the criteria set forth in this section. If the licensee submits the certification of financial assurance rather than a decommissioning funding plan, the licensee shall include a decommissioning funding plan in any application for license renewal.

Page 411: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-17

(3) Each holder of a specific license issued before July 27, 1990, and of a type described in paragraph (b) of this section shall submit, on or before July 27, 1990, a decommissioning funding plan as described, in paragraph (e) of this section, or a certification of financial assurance for decommissioning in accordance with the criteria set forth in this section. (4) Any licensee who has submitted an application before July 27, 1990, for renewal of license in accordance with §30.37 shall provide financial assurance for decommissioning in accordance with paragraphs (a) and (b) of this section. This assurance must be submitted when this rule becomes effective November 24, 1995. (d) Table of required amounts of financial assurance for decommissioning by quantity of material. greater than 104 but less than or equal to 105 times the applicable quantities of appendix B to part 30 in unsealed form. (For a combination of isotopes, if R, as defined in §30.35(a), divided by 104 is greater than 1 but R divided by 105 is less than or equal to 1.) .... $750,000 greater than 103 but less than or equal to 104 times the applicable quantities of appendix B to part 30 in unsealed form. (For a combination of isotopes, if R, as defined in §30.35(a), divided by 103 is greater than 1 but R divided by 104 is less than or equal to 1.) .... $150,000 greater than 1010 times the applicable quantities of appendix B to part 30 in sealed sources or plated foils. (For a combination of isotopes, if R, as defined in §30.35(a), divided by 1010 is greater than 1). .... $75,000 (e) Each decommissioning funding plan must contain a cost estimate for decommissioning and a description of the method of assuring funds for decommissioning from paragraph (f) of this section, including means for adjusting cost estimates and associated funding levels periodically over the life of the facility. The decommissioning funding plan must also contain a certification by the licensee that financial assurance for decommissioning has been provided in the amount of the cost estimate for decommissioning and a signed original of the financial instrument obtained to satisfy the requirements of paragraph (f) of this section. (f) Financial assurance for decommissioning must be provided by one or more of the following methods: (1) Prepayment. Prepayment is the deposit prior to the start of operation into an account segregated from licensee assets and outside the licensee's administrative control of cash or liquid assets such that the amount of funds would be sufficient to pay decommissioning costs. Prepayment may be in the form of a trust, escrow account, government fund, certificate of deposit, or deposit of government securities. (2) A surety method, insurance, or other guarantee method. These methods guarantee that decommissioning costs will be paid. A surety method may be in the form of a surety bond, letter of credit, or line of credit. A parent company guarantee of funds for decommissioning costs based on a financial test may be used if the guarantee and test are as contained in appendix A to this part. A parent company guarantee may not be used in combination with other financial methods to satisfy the requirements of this section. For

Page 412: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-18

commercial corporations that issue bonds, a guarantee of funds by the applicant or licensee for decommissioning costs based on a financial test may be used if the guarantee and test are as contained in appendix C to this part. For commercial companies that do not issue bonds, a guarantee of funds by the applicant or licensee for decommissioning costs may be used if the guarantee and test are as contained in appendix D to this part. For nonprofit entities, such as colleges, universities, and nonprofit hospitals, a guarantee of funds by the applicant or licensee may be used if the guarantee and test are as contained in appendix E to this part. A guarantee by the applicant or licensee may not be used in combination with any other financial methods used to satisfy the requirements of this section or in any situation where the applicant or licensee has a parent company holding majority control of the voting stock of the company. Any surety method or insurance used to provide financial assurance for decommissioning must contain the following conditions: (i) The surety method or insurance must be open-ended or, if written for a specified term, such as five years, must be renewed automatically unless 90 days or more prior to the renewal date, the issuer notifies the Commission, the beneficiary, and the licensee of its intention not to renew. The surety method or insurance must also provide that the full face amount be paid to the beneficiary automatically prior to the expiration without proof of forfeiture if the licensee fails to provide a replacement acceptable to the Commission within 30 days after receipt of notification of cancellation. (ii) The surety method or insurance must be payable to a trust established for decommissioning costs. The trustee and trust must be acceptable to the Commission. An acceptable trustee includes an appropriate State or Federal government agency or an entity which has the authority to act as a trustee and whose trust operations are regulated and examined by a Federal or State agency. (iii) The surety method or insurance must remain in effect until the Commission has terminated the license. (3) An external sinking fund in which deposits are made at least annually, coupled with a surety method or insurance, the value of which may decrease by the amount being accumulated in the sinking fund. An external sinking fund is a fund established and maintained by setting aside funds periodically in an account segregated from licensee assets and outside the licensee's administrative control in which the total amount of funds would be sufficient to pay decommissioning costs at the time termination of operation is expected. An external sinking fund may be in the form of a trust, escrow account, government fund, certificate of deposit, or deposit of government securities. The surety or insurance provisions must be as stated in paragraph (f)(2) of this section. (4) In the case of Federal, State, or local government licensees, a statement of intent containing a cost estimate for decommissioning or an amount based on the Table in paragraph (d) of this section, and indicating that funds for decommissioning will be obtained when necessary. (5) When a governmental entity is assuming custody and ownership of a site, an arrangement that is deemed acceptable by such governmental entity.

Page 413: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-19

(g) Each person licensed under this part or parts 32 through 36 and 39 of this chapter shall keep records of information important to the decommissioning of a facility in an identified location until the site is released for unrestricted use. Before licensed activities are transferred or assigned in accordance with §30.34(b), licensees shall transfer all records described in this paragraph to the new licensee. In this case, the new licensee will be responsible for maintaining these records until the license is terminated. If records important to the decommissioning of a facility are kept for other purposes, reference to these records and their locations may be used. Information the Commission considers important to decommissioning consists of -- (1) Records of spills or other unusual occurrences involving the spread of contamination in and around the facility, equipment, or site. These records may be limited to instances when contamination remains after any cleanup procedures or when there is reasonable likelihood that contaminants may have spread to inaccessible areas as in the case of possible seepage into porous materials such as concrete. These records must include any known information on identification of involved nuclides, quantities, forms, and concentrations. (2) As-built drawings and modifications of structures and equipment in restricted areas where radioactive materials are used and/or stored, and of locations of possible inaccessible contamination such as buried pipes which may be subject to contamination. If required drawings are referenced, each relevant document need not be indexed individually. If drawings are not available, the licensee shall substitute appropriate records of available information concerning these areas and locations. (3) Except for areas containing only sealed sources (provided the sources have not leaked or no contamination remains after any leak) or byproduct materials having only half-lives of less than 65 days, a list contained in a single document and updated every 2 years, of the following: (i) All areas designated and formerly designated restricted areas as defined in 10 CFR 20.1003 (For requirements prior to January 1, 1994, see 10 CFR 20.3 as contained in the CFR edition revised as of January 1, 1993.); (ii) All areas outside of restricted areas that require documentation under §30.35(g)(1). (iii) All areas outside of restricted areas where current and previous wastes have been buried as documented under 10 CFR 20.2108; and (iv) All areas outside of restricted areas that contain material such that, if the license expired, the licensee would be required to either decontaminate the area to meet the criteria for decommissioning in 10 CFR part 20, subpart E, or apply for approval for disposal under 10 CFR 20.2002. (4) Records of the cost estimate performed for the decommissioning funding plan or of the amount certified for decommissioning, and records of the funding method used for assuring funds if either a funding plan or certification is used. [53 FR 24044, June 27, 1988, as amended at 56 FR 23471, May 21, 1991; 58 FR 39633, July 26, 1993; 58 FR 67659, Dec. 22, 1993; 58 FR 68730, Dec. 29, 1993; 59 FR 1618, Jan. 12, 1994; 60 FR 38238, July 26, 1995; 61 FR 24673, May 16, 1996; 62 FR 39090, July 21, 1997]

Page 414: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-20

§ 30.41 Transfer of byproduct material. (a) No licensee shall transfer byproduct material except as authorized pursuant to this section. (b) Except as otherwise provided in his license and subject to the provisions of paragraphs (c) and (d) of this section, any licensee may transfer byproduct material: (1) To the Department; (2) To the agency in any Agreement State which regulates radioactive material pursuant to an agreement under section 274 of the Act; (3) To any person exempt from the licensing requirements of the Act and regulations in this part, to the extent permitted under such exemption; (4) To any person in an Agreement State, subject to the jurisdiction of that State, who has been exempted from the licensing requirements and regulations of that State, to the extent permitted under such exemption; (5) To any person authorized to receive such byproduct material under terms of a specific license or a general license or their equivalents issued by the Atomic Energy Commission, the Commission, or an Agreement State; (6) To a person abroad pursuant to an export license issued under part 110 of this chapter; or (7) As otherwise authorized by the Commission in writing. (c) Before transferring byproduct material to a specific licensee of the Commission or an Agreement State or to a general licensee who is required to register with the Commission or with an Agreement State prior to receipt of the byproduct material, the licensee transferring the material shall verify that the transferee's license authorizes the receipt of the type, form, and quantity of byproduct material to be transferred. (d) The following methods for the verification required by paragraph (c) of this section are acceptable: (1) The transferor may have in his possession, and read, a current copy of the transferee's specific license or registration certificate; (2) The transferor may have in his possession a written certification by the transferee that he is authorized by license or registration certificate to receive the type, form, and quantity of byproduct material to be transferred, specifying the license or registration certificate number, issuing agency and expiration date; (3) For emergency shipments the transferor may accept oral certification by the transferee that he is authorized by license or registration certificate to receive the type, form, and quantity of byproduct material to be transferred, specifying the license or registration certificate number, issuing agency and expiration date: Provided, That the oral certification is confirmed in writing within 10 days; (4) The transferor may obtain other sources of information compiled by a reporting service from official records of the Commission or the licensing agency of an Agreement State as to the identity of licensees and the scope and expiration dates of licenses and registration; or

Page 415: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-21

(5) When none of the methods of verification described in paragraphs (d)(1) to (4) of this section are readily available or when a transferor desires to verify that information received by one of such methods is correct or up-to-date, the transferor may obtain and record confirmation from the Commission or the licensing agency of an Agreement State that the transferee is licensed to receive the byproduct material. [38 FR 33969, Dec. 10, 1973, as amended at 40 FR 8785, Mar. 3, 1975; 43 FR 6922, Feb. 17, 1978] RECORDS, INSPECTIONS, TESTS, AND REPORTS § 30.50 Reporting requirements. (a) Immediate report. Each licensee shall notify the NRC as soon as possible but not later than 4 hours after the discovery of an event that prevents immediate protective actions necessary to avoid exposures to radiation or radioactive materials that could exceed regulatory limits or releases of licensed material that could exceed regulatory limits (events may include fires, explosions, toxic gas releases, etc.). (b) Twenty-four hour report. Each licensee shall notify the NRC within 24 hours after the discovery of any of the following events involving licensed material: (1) An unplanned contamination event that: (i) Requires access to the contaminated area, by workers or the public, to be restricted for more than 24 hours by imposing additional radiological controls or by prohibiting entry into the area; (ii) Involves a quantity of material greater than five times the lowest annual limit on intake specified in appendix B of §§ 20.1001-20.2401 of 10 CFR part 20 for the material; and (iii) Has access to the area restricted for a reason other than to allow isotopes with a half-life of less than 24 hours to decay prior to decontamination. (2) An event in which equipment is disabled or fails to function as designed when: (i) The equipment is required by regulation or license condition to prevent releases exceeding regulatory limits, to prevent exposures to radiation and radioactive materials exceeding regulatory limits, or to mitigate the consequences of an accident; (ii) The equipment is required to be available and operable when it is disabled or fails to function; and (iii) No redundant equipment is available and operable to perform the required safety function. (3) An event that requires unplanned medical treatment at a medical facility of an individual with spreadable radioactive contamination on the individual's clothing or body. (4) An unplanned fire or explosion damaging any licensed material or any device, container, or equipment containing licensed material when: (i) The quantity of material involved is greater than five times the lowest annual limit on intake specified in appendix B of §§ 20.1001-20.2401 of 10 CFR part 20 for the material; and

Page 416: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-22

(ii) The damage affects the integrity of the licensed material or its container. (c) Preparation and submission of reports. Reports made by licensees in response to the requirements of this section must be made as follows: (1) Licensees shall make reports required by paragraphs (a) and (b) of this section by telephone to the NRC Operations Center.{1} To the extent that the information is available at the time of notification, the information provided in these reports must include: |{1} The commercial telephone number for the NRC Operations |Center is (301) 816-5100. (i) The caller's name and call back telephone number; (ii) A description of the event, including date and time; (iii) The exact location of the event; (iv) The isotopes, quantities, and chemical and physical form of the licensed material involved; and (v) Any personnel radiation exposure data available. (2) Written report. Each licensee who makes a report required by paragraph (a) or (b) of this section shall submit a written follow-up report within 30 days of the initial report. Written reports prepared pursuant to other regulations may be submitted to fulfill this requirement if the reports contain all of the necessary information and the appropriate distribution is made. These written reports must be sent to the U.S. Nuclear Regulatory Commission, Document Control Desk, Washington, DC 20555, with a copy to the appropriate NRC Regional office listed in appendix D of 10 CFR part 20. The reports must include the following: (i) A description of the event, including the probable cause and the manufacturer and model number (if applicable) of any equipment that failed or malfunctioned; (ii) The exact location of the event; (iii) The isotopes, quantities, and chemical and physical form of the licensed material involved; (iv) Date and time of the event; (v) Corrective actions taken or planned and the results of any evaluations or assessments; and (vi) The extent of exposure of individuals to radiation or to radioactive materials without identification of individuals by name. (3) The provisions of § 30.50 do not apply to licensees subject to the notification requirements in § 50.72. They do apply to those part 50 licensees possessing material licensed under part 30, who are not subject to the notification requirements in § 50.72. [56 FR 40767, Aug. 16, 1991, as amended at 59 FR 14086, Mar. 25, 1994] § 30.51 Records.

Page 417: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-23

(a) Each person who receives byproduct material pursuant to a license issued pursuant to the regulations in this part and parts 31 through 36 of this chapter shall keep records showing the receipt, transfer, and disposal of the byproduct material as follows: (1) The licensee shall retain each record of receipt of byproduct material as long as the material is possessed and for three years following transfer or disposal of the material. (2) The licensee who transferred the material shall retain each record of transfer for three years after each transfer unless a specific requirement in another part of the regulations in this chapter dictates otherwise. (3) The licensee who disposed of the material shall retain each record of disposal of byproduct material until the Commission terminates each license that authorizes disposal of the material. (b) The licensee shall retain each record that is required by the regulations in this part and parts 31 through 36 of this chapter or by license condition for the period specified by the appropriate regulation or license condition. If a retention period is not otherwise specified by regulation or license condition, the record must be retained until the Commission terminates each license that authorizes the activity that is subject to the recordkeeping requirement. (c)(1) Records which must be maintained pursuant to this part and parts 31 through 36 of this chapter may be the original or a reproduced copy or microform if such reproduced copy or microform is duly authenticated by authorized personnel and the microform is capable of producing a clear and legible copy after storage for the period specified by Commission regulations. The record may also be stored in electronic media with the capability for producing legible, accurate, and complete records during the required retention period. Records such as letters, drawings, specifications, must include all pertinent information such as stamps, initials, and signatures. The licensee shall maintain adequate safeguards against tampering with and loss of records. (2) If there is a conflict between the Commission's regulations in this part and parts 31 through 36 and 39 of this chapter, license condition, or other written Commission approval or authorization pertaining to the retention period for the same type of record, the retention period specified in the regulations in this part and parts 31 through 36 and 39 of this chapter for such records shall apply unless the Commission, pursuant to § 30.11, has granted a specific exemption from the record retention requirements specified in the regulations in this part or parts 31 through 36 and 39 of this chapter. (d) Prior to license termination, each licensee authorized to possess radioactive material with a half-life greater than 120 days, in an unsealed form, shall forward the following records to the appropriate NRC Regional Office: (1) Records of disposal of licensed material made under §§ 20.2002 (including burials authorized before January 28, 1981 {1}), 20.2003, 20.2004, 20.2005; and | {1} A previous § 20.304 permitted burial of small |quantities of licensed materials in soil before January |28, 1981, without specific Commission authorization. |See § 20.304 contained in the 10 CFR, parts 0 to 199, |edition revised as of January 1, 1981. (2) Records required by § 20.2103(b)(4).

Page 418: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-24

(e) If licensed activities are transferred or assigned in accordance with § 30.34(b), each licensee authorized to possess radioactive material, with a half-life greater than 120 days, in an unsealed form, shall transfer the following records to the new licensee and the new licensee will be responsible for maintaining these records until the license is terminated: (1) Records of disposal of licensed material made under §§ 20.2002 (including burials authorized before January 28, 1981 {1}), 20.2003, 20.2004, 20.2005; and | {1} A previous § 20.304 permitted burial of small |quantities of licensed materials in soil before January |28, 1981, without specific Commission authorization. |See § 20.304 contained in the 10 CFR, parts 0 to 199, |edition revised as of January 1, 1981. (2) Records required by § 20.2103(b)(4). (f) Prior to license termination, each licensee shall forward the records required by § 30.35(g) to the appropriate NRC Regional Office. [41 FR 18301, May 5, 1976, as amended at 43 FR 6922, Feb. 17, 1978; 52 FR 8241, Mar. 17, 1987; 53 FR 19245, May 27, 1988; 58 FR 7736, Feb. 9, 1993; 61 FR 24669, May 16, 1996] § 30.52 Inspections. (a) Each licensee shall afford to the Commission at all reasonable times opportunity to inspect byproduct material and the premises and facilities wherein byproduct material is used or stored. (b) Each licensee shall make available to the Commission for inspection, upon reasonable notice, records kept by him pursuant to the regulations in this chapter. [30 FR 8185, June 26, 1965] § 30.53 Tests. Each licensee shall perform, or permit the Commission to perform, such tests as the Commission deems appropriate or necessary for the administration of the regulations in this part and parts 31 through 36 and 39 of this chapter, including tests of: (a) Byproduct material; (b) Facilities wherein byproduct material is utilized or stored; (c) Radiation detection and monitoring instruments; and (d) Other equipment and devices used in connection with the utilization or storage of byproduct material. [30 FR 8185, June 26, 1965, as amended by 43 FR 6922, Feb. 17, 1978; 52 FR 8241, Mar. 17, 1987; 58 FR 7736, Feb. 9, 1993]

Page 419: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-25

§ 30.55 Tritium reports. (a)-(b) [Reserved] (c) Except as specified in paragraph (d) of this section, each licensee who is authorized to possess tritium shall report promptly to the appropriate NRC Regional Office listed in appendix D of part 20 of this chapter by telephone and telegraph, mailgram, or facsimile any incident in which an attempt has been made or is believed to have been made to commit a theft or unlawful diversion of more than 10 curies of such material at any one time or more than 100 curies of such material in any one calendar year. The initial report shall be followed within a period of fifteen (15) days by a written report submitted to the appropriate NRC Regional Office which sets forth the details of the incident and its consequences. Copies of such written report shall be sent to the Director, Office of Nuclear Material Safety and Safeguards, U.S. Nuclear Regulatory Commission, Washington, DC 20555. Subsequent to the submission of the written report required by this paragraph, the licensee shall promptly inform the Office of Nuclear Material Safety and Safeguards by means of a written report of any substantive additional information, which becomes available to the licensee, concerning an attempted or apparent theft or unlawful diversion of tritium. (d) The reports described in this section are not required for tritium possessed pursuant to a general license provided in part 31 of this chapter or for tritium contained in spent fuel. [37 FR 9208, May 6, 1972, as amended at 38 FR 1271, Jan. 11, 1973; 38 FR 2330, Jan. 24, 1973; 41 FR 16446, Apr. 19, 1976; 43 FR 6922, Feb. 17, 1978; 46 FR 55085, Nov. 6, 1981; 49 FR 24707, June 15, 1984; 52 FR 31611, Aug. 21, 1987] ENFORCEMENT § 30.61 Modification and revocation of licenses. (a) The terms and conditions of each license issued pursuant to the regulations in this part and parts 31 through 35 of this chapter shall be subject to amendment, revision or modification by reason of amendments to the Act, or by reason of rules, regulations and orders issued in accordance with the terms of the Act. (b) Any license may be revoked, suspended or modified, in whole or in part, for any material false statement in the application or any statement of fact required under section 182 of the Act, or because of conditions revealed by such application or statement of fact or any report, record or inspection or other means which would warrant the Commission to refuse to grant a license on an original application, or for violation of, or failure to observe any of the terms and provisions of the Act or of any rule, regulation or order of the Commission. (c) Except in cases of willfulness or those in which the public health, interest or safety requires otherwise, no license shall be modified, suspended or revoked unless, prior to the

Page 420: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-26

institution of proceedings therefore, facts or conduct which may warrant such action shall have been called to the attention of the licensee in writing and the licensee shall have been accorded an opportunity to demonstrate or achieve compliance with all lawful requirements. [30 FR 8185, June 26, 1965, as amended at 35 FR 11460, July 17, 1970; 43 FR 6922, Feb. 17, 1978] § 30.62 Right to cause the withholding or recall of byproduct material. The Commission may cause the withholding or recall of byproduct material from any licensee who is not equipped to observe or fails to observe such safety standards to protect health as may be established by the Commission, or who uses such materials in violation of law or regulation of the Commission, or in a manner other than as disclosed in the application therefore or approved by the Commission. [30 FR 8185, June 26, 1965, as amended at 40 FR 8785, Mar. 3, 1975] § 30.63 Violations. (a) The Commission may obtain an injunction or other court order to prevent a violation of the provisions of- (1) The Atomic Energy Act of 1954, as amended; (2) Title II of the Energy Reorganization Act of 1974, as amended; or (3) A regulation or order issued pursuant to those Acts. (b) The Commission may obtain a court order for the payment of a civil penalty imposed under section 234 of the Atomic Energy Act: (1) For violations of- (i) Sections 53, 57, 62, 63, 81, 82, 101, 103, 104, 107, or 109 of the Atomic Energy Act of 1954, as amended; (ii) Section 206 of the Energy Reorganization Act; (iii) Any rule, regulation, or order issued pursuant to the sections specified in paragraph (b)(1)(i) of this section; (iv) Any term, condition, or limitation of any license issued under the sections specified in paragraph (b)(1)(i) of this section. (2) For any violation for which a license may be revoked under section 186 of the Atomic Energy Act of 1954, as amended. [57 FR 55072, Nov. 24, 1992] § 30.64 Criminal penalties. (a) Section 223 of the Atomic Energy Act of 1954, as amended, provides for criminal sanctions for willful violation of, attempted violation of, or conspiracy to violate, any regulation issued under sections 161b, 161i, or 161o of the Act. For purposes of section

Page 421: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-27

223, all the regulations in part 30 are issued under one or more of sections 161b, 161i, or 161o, except for the sections listed in paragraph (b) of this section. (b) The regulations in part 30 that are not issued under sections 161b, 161i, or 161o for the purposes of section 223 are as follows: §§ 30.1, 30.2, 30.4, 30.5, 30.6, 30.8, 30.11, 30.12, 30.13, 30.15, 30.16, 30.31, 30.32, 30.33, 30.37, 30.38, 30.39, 30.61, 30.62, 30.63, 30.64, 30.70, 30.71, and 30.72. [57 FR 55072, Nov. 24, 1992]

Schedules § 30.70 Schedule A-exempt concentrations.

[See footnotes at end of this table]

Element (atomic number) Isotope Col. I Col. II Gas concentration

μCi/ml {1} Liquid and solid concentration μCi/ml{2}

Antimony (51) Sb 122 3x10-4 Sb 124 2x10-4 Sb 125 1x10-3 Argon (18) A 37 1x10-3 A 41 4x10-7 Arsenic (33) As 73 5x10_-3 As 74 5x10_-4 As 76 2x10_-4 As 77 8x10_-4 Barium (56) Ba 131 2x10_-3 Ba 140 3x10_-4 Beryllium (4) Be 7 2x10_-2 Bismuth (83) Bi 206 4x10_-4 Bromine (35) Br 82 4x10_-7 3x10_-3 Cadmium (48) Cd 109 2x10_-3 Cd 115m 3x10_-4 Cd 115 3x10_-4 Calcium (20) Ca 45 9x10_-5 Ca 47 5x10_-4 Carbon (6) C 14 1x10_-6 8x10_-3 Cerium (58) Ce 141 9x10_-4 Ce 143 4x10_-4 Ce 144 1x10_-4 Cesium (55) Cs 131 2x10_-2 Cs 134m 6x10_-2 Cs 134 9x10_-5 Chlorine (17) Cl 38 9x10_-7 4x10_-3 Chromium (24) Cr 51 2x10_-2 Cobalt (27) Co 57 5x10_-3 Co 58 1x10_-3 Co 60 5x10_-4 Copper (29) Cu 64 3x10_-3 Dysprosium (66) Dy 165 4x10_-3 Dy 166 4x10_-4

Page 422: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-28

Erbium (68) Er 169 9x10_-4 Er 171 1x10_-3 Europium (63) Eu 152 6x10_-4 (T/2=9.2 Hrs) Eu 155 2x10_-3 Fluorine (9) F 18 2x10_-6 8x10_-3 Gadolinium (64) Gd 153 2x10_-3 Gd 159 8x10_-4 Element (atomic number) Isotope Col. I Col. II Gallium (31) Ga 72 4x10_-4 Germanium (32) Ge 71 2x10_-2 Gold (79) Au 196 2x10_-3 Au 198 5x10_-4 Au 199 2x10_-3 Hafnium (72) Hf 181 7x10_-4 Hydrogen (1) H 3. 5x10_-6 3x10_-2 Indium (49) In 113m 1x10_-2 In 114m 2x10_-4 Iodine (53) I 126 3x10_-9 2x10_-5 I 131 3x10_-9 2x10_-5 I 132 8x10_-8 6x10_-4 I 133 1x10_-8 7x10_-5 I 134 2x10_-7 1x10_-3 Iridium (77) Ir 190 2x10_-3 Ir 192 4x10_-4 Ir 194 3x10_-4 Iron (26) Fe 55 8x10_-3 Fe 59 6x10_-4 Krypton (36) Kr 85m 1x10_-6 Kr 85 3x10_-6 Lanthanum (57) La 140 2x10_-4 Lead (82) Pb 203 4x10_-3 Lutetium (71) Lu 177 1x10_-3 Manganese (25) Mn 52 3x10_-4 Mn 54 1x10_-3 Mn 56 1x10_-3 Mercury (80) Hg 197m 2x10_-3 Hg 197 3x10_-3 Hg 203 2x10_-4 Molybdenum (42) Mo 99 2x10_-3 Neodymium (60) Nd 147 6x10_-4 Nd 149 3x10_-3 Nickel (28) Ni 65 1x10_-3 Niobium (Columbium) Nb 95 1x10_-3 (41). Nb 97 9x10_-3 Osmium (76) Os 185 7x10_-4 Os 191m 3x10_-2 Os 191 2x10_-3 Os 193 6x10_-4 Palladium (46) Pd 103 3x10_-3 Pd 109 9x10_-4 Phosphorus (15) P 32 2x10_-4 Platinum (78) Pt 191 1x10_-3

Page 423: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-29

Pt 193m 1x10_-2 Pt 197m 1x10_-2 Pt 197 1x10_-3 Potassium (19) K 42 3x10_-3 Praseodymium (59) Pr 142 3x10_-4 Pr 143 5x10_-4 Element (atomic number) Isotope Col. I Col. II Promethium (61) Pm 147 2x10_-3 Pm 149 4x10_-4 Rhenium (75) Re 183 6x10_-3 Re 186 9x10_-4 Re 188 6x10_-4 Rhodium (45) Rh 103m 1x10_-1 Rh 105 1x10_-3 Rubidium (37) Rb 86 7x10_-4 Ruthenium (44) Ru 97 4x10_-4 Ru 103 8x10_-4 Ru 105 1x10_-3 Ru 106 1x10_-4 Samarium (62) Sm 153 8x10_-4 Scandium (21) Sc 46 4x10_-4 Sc 47 9x10_-4 Sc 48 3x10_-4 Selenium (34) Se 75 3x10_-3 Silicon (14) Si 31 9x10_-3 Silver (47) Ag 105 1x10_-3 Ag 110m 3x10_-4 Ag 111 4x10_-4 Sodium (11) Na 24 2x10_-3 Strontium (38) Sr 85 1x10_-4 Sr 89 1x10_-4 Sr 91 7x10_-4 Sr 92 7x10_-4 Sulfur (16) S 35 9x10_-8 6x10_-4 Tantalum (73) Ta 182 4x10_-4 Technetium (43) Tc 96m 1x10_-1 Tc 96 1x10_-3 Tellurium (52) Te 125m 2x10_-3 Te 127m 6x10_-4 Te 127 3x10_-3 Te 129m 3x10_-4 Te 131m 6x10_-4 Te 132 3x10_-4 Terbium (65) Tb 160 4x10_-4 Thallium (81) Tl 200 4x10_-3 Tl 201 3x10_-3 Tl 202 1x10_-3 Tl 204 1x10_-3 Thulium (69) Tm 170 5x10_-4 Tm 171 5x10_-3 Tin (50) Sn 113 9x10_-4 Sn 125 2x10_-4 Tungsten (Wolfram) (74) W 181 4x10_-3 W 187 7x10_-4

Page 424: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-30

Vanadium (23) V 48 3x10_-4 Xenon (54) Xe 131m 4x10_-6 Xe 133 3x10_-6 Xe 135 1x10_-6 Element (atomic number) Isotope Col. I Col. II Ytterbium (70) Yb 175 1x10_-3 Yttrium (39) Y 90 2x10_-4 Y 91m 3x10_-2 Y 91 3x10_-4 Y 92 6x10_-4 Y 93 3x10_-4 Zinc (30) Zn 65 1x10_-3 Zn 69m 7x10_-4 Zn 69 2x10_-2 Zirconium (40) Zr 95 6x10_-4 Zr 97 2x10_-4 Beta and/or gamma emitting byproduct material not listed above with half-life less than 3 years.

1x10_-10 1x10_-6

--------------------------------------------------------------------------- Footnotes to Schedule A: {1}Values are given only for those materials normally used as gases. {2}æCi/gm for solids. Note 1: Many radioisotopes disintegrate into isotopes which are also radioactive. In expressing the concentrations in Schedule A, the activity stated is that of the parent isotope and takes into account the daughters. Note 2: For purposes of § 30.14 where there is involved a combination of isotopes, the limit for the combination should be derived as follows: Determine for each isotope in the product the ratio between the concentration present in the product and the exempt concentration established in Schedule A for the specific isotope when not in combination. The sum of such ratios may not exceed ``1'' (i.e., unity). Example: Concentration of Isotope A in Product -------------------------------------- Exempt concentration of Isotope A Concentration of Isotope B in Product --------------------------------------<= 1 Exempt concentration of Isotope B

Page 425: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-31

[30 FR 8185, June 26, 1965, as amended at 35 FR 3982, Mar. 3, 1970; 38 FR 29314, Oct. 24, 1973; 59 FR 5520, Feb. 7, 1994] § 31.5 Certain detecting, measuring, gauging, or controlling devices and certain devices for producing light or an ionized atmosphere.(2)

(a) A general license is hereby issued to commercial and industrial firms and research, educational and medical institutions, individuals in the conduct of their business, and Federal, State or local government agencies to acquire, receive, possess, use or transfer, in accordance with the provisions of paragraphs (b), (c) and (d) of this section, byproduct material contained in devices designed and manufactured for the purpose of detecting, measuring, gauging or controlling thickness, density, level, interface location, radiation, leakage, or qualitative or quantitative chemical composition, or for producing light or an ionized atmosphere. (b)(1) The general license in paragraph (a) of this section applies only to byproduct material contained in devices which have been manufactured or initially transferred and labeled in accordance with the specifications contained in-- (i) A specific license issued under Sec. 32.51 of this chapter; or (ii) An equivalent specific license issued by an Agreement State. (2) The devices must have been received from one of the specific licensees described in paragraph (b)(1) of this section or through a transfer made under paragraph (c)(9) of this section. (c) Any person who acquires, receives, possesses, uses or transfers byproduct material in a device pursuant to the general license in paragraph (a) of this section: (1) Shall assure that all labels affixed to the device at the time of receipt and bearing a statement that removal of the label is prohibited are maintained thereon and shall comply with all instructions and precautions provided by such labels; (2) Shall assure that the device is tested for leakage of radioactive material and proper operation of the on-off mechanism and indicator, if any, at no longer than six-month intervals or at such other intervals as are specified in the label; however: (i) Devices containing only krypton need not be tested for leakage of radioactive material, and (ii) Devices containing only tritium or not more than 100 microcuries of other beta and/or gamma emitting material or 10 microcuries of alpha emitting material and devices held in storage in the original shipping container prior to initial installation need not be tested for any purpose; (3) Shall assure that the tests required by paragraph (c)(2) of this section and other testing, installation, servicing, and removal from installation involving the radioactive materials, its shielding or containment, are performed: (i) In accordance with the instructions provided by the labels; or (ii) By a person holding a specific license pursuant to parts 30 and 32 of this chapter or from an Agreement State to perform such activities; (4) Shall maintain records showing compliance with the requirements of paragraphs (c)(2) and (c)(3) of this section. The records must show the results of tests. The records

Page 426: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-32

also must show the dates of performance of, and the names of persons performing, testing, installing, servicing, and removing from the installation radioactive material and its shielding or containment. The licensee shall retain these records as follows: (i) Each record of a test for leakage or radioactive material required by paragraph (c)(2) of this section must be retained for three years after the next required leak test is performed or until the sealed source is transferred or disposed of. (ii) Each record of a test of the on-off mechanism and indicator required by paragraph (c)(2) of this section must be retained for three years after the next required test of the on-off mechanism and indicator is performed or until the sealed source is transferred or disposed of. (iii) Each record that is required by paragraph (c)(3) of this section must be retained for three years from the date of the recorded event or until the device is transferred or disposed of. (5) Shall immediately suspend operation of the device if there is a failure of, or damage to, or any indication of a possible failure of or damage to, the shielding of the radioactive material or the on-off mechanism or indicator, or upon the detection of 185 bequerel (0.005 microcurie) or more removable radioactive material. The device may not be operated until it has been repaired by the manufacturer or other person holding a specific license to repair such devices that was issued under parts 30 and 32 of this chapter or by an Agreement State. The device and any radioactive material from the device may only be disposed of by transfer to a person authorized by a specific license to receive the byproduct material in the device or as otherwise approved by the Commission. A report containing a brief description of the event and the remedial action taken; and, in the case of detection of 0.005 microcurie or more removable radioactive material or failure of or damage to a source likely to result in contamination of the premises or the environs, a plan for ensuring that the premises and environs are acceptable for unrestricted use, must be furnished to the Director of Nuclear Material Safety and Safeguards, ATTN: GLTS, U.S. Nuclear Regulatory Commission, Washington, DC 20555-0001 within 30 days. Under these circumstances, the criteria set out in Sec. 20.1402, ``Radiological criteria for unrestricted use,'' may be applicable, as determined by the Commission on a case-by-case basis; (6) Shall not abandon the device containing byproduct material; (7) Shall not export the device containing byproduct material except in accordance with part 110 of this chapter; (8)(i) Shall transfer or dispose of the device containing byproduct material only by export as provided by paragraph (c)(7) of this section, by transfer to another general licensee as authorized in paragraph (c)(9) of this section, or to a person authorized to receive the device by a specific license issued under parts 30 and 32 of this chapter, or part 30 of this chapter that authorizes waste collection, or equivalent regulations of an Agreement State, or as otherwise approved under paragraph (c)(8)(iii) of this section. (ii) Shall furnish a report to the Director of Nuclear Material Safety and Safeguards, ATTN: GLTS, U.S. Nuclear Regulatory Commission, Washington, DC 20555-0001

Page 427: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-33

within 30 days after the transfer of a device to a specific licensee or export. The report must contain-- (A) The identification of the device by manufacturer's (or initial transferor's) name, model number, and serial number; (B) The name, address, and license number of the person receiving the device (license number not applicable if exported); and (C) The date of the transfer. (iii) Shall obtain written NRC approval before transferring the device to any other specific licensee not specifically identified in paragraph (c)(8)(i) of this section. (9) Shall transfer the device to another general licensee only if-- (i) The device remains in use at a particular location. In this case, the transferor shall give the transferee a copy of this section, a copy of Secs. 31.2, 30.51, 20.2201, and 20.2202 of this chapter, and any safety documents identified in the label of the device. Within 30 days of the transfer, the transferor shall report to the Director of Nuclear Material Safety and Safeguards, ATTN: GLTS, U.S. Nuclear Regulatory Commission, Washington, DC 20555-0001-- (A) The manufacturer's (or initial transferor's) name; (B) The model number and the serial number of the device transferred; (C) The transferee's name and mailing address for the location of use; and (D) The name, title, and phone number of the responsible individual identified by the transferee in accordance with paragraph (c)(12) of this section to have knowledge of and authority to take actions to ensure compliance with the appropriate regulations and requirements; or (ii) The device is held in storage by an intermediate person in the original shipping container at its intended location of use prior to initial use by a general licensee. (10) Shall comply with the provisions of §§20.2201, and 20.2202 of this chapter for reporting radiation incidents, theft or loss of licensed material, but shall be exempt from the other requirements of parts 19, 20, and 21, of this chapter. (11) Shall respond to written requests from the Nuclear Regulatory Commission to provide information relating to the general license within 30 calendar days of the date of the request, or other time specified in the request. If the general licensee cannot provide the requested information within the allotted time, it shall, within that same time period, request a longer period to supply the information by submitting a letter to the Director, Office of Nuclear Material Safety and Safeguards, U.S. Nuclear Regulatory Commission, Washington, DC 20555-0001 and provide written justification as to why it cannot comply. (12) Shall appoint an individual responsible for having knowledge of the appropriate regulations and requirements and the authority for taking required actions to comply with appropriate regulations and requirements. The general licensee, through this individual, shall ensure the day-to-day compliance with appropriate regulations and requirements. This appointment does not relieve the general licensee of any of its responsibility in this regard.

Page 428: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-34

(13)(i) Shall register, in accordance with paragraphs (c)(13)(ii) and (iii) of this section, devices containing at least 370 MBq (10 mCi) of cesium-137, 3.7 MBq (0.1 mCi) of strontium-90, 37 MBq (1 mCi) of cobalt-60, or 37 MBq (1 mCi) of americium-241 or any other transuranic (i.e., element with atomic number greater than uranium (92)), based on the activity indicated on the label. Each address for a location of use, as described under paragraph (c)(13)(iii)(D) of this section, represents a separate general licensee and requires a separate registration and fee. (ii) If in possession of a device meeting the criteria of paragraph (c)(13)(i) of this section, shall register these devices annually with the Commission and shall pay the fee required by Sec. 170.31 of this chapter. Registration must be done by verifying, correcting, and/or adding to the information provided in a request for registration received from the Commission. The registration information must be submitted to the NRC within 30 days of the date of the request for registration or as otherwise indicated in the request. In addition, a general licensee holding devices meeting the criteria of paragraph (c)(13)(i) of this section is subject to the bankruptcy notification requirement in Sec. 30.34(h) of this chapter. (iii) In registering devices, the general licensee shall furnish the following information and any other information specifically requested by the Commission-- (A) Name and mailing address of the general licensee. (B) Information about each device: the manufacturer (or initial transferor), model number, serial number, the radioisotope and activity (as indicated on the label). (C) Name, title, and telephone number of the responsible person designated as a representative of the general licensee under paragraph (c)(12) of this section. (D) Address or location at which the device(s) are used and/or stored. For portable devices, the address of the primary place of storage. (E) Certification by the responsible representative of the general licensee that the information concerning the device(s) has been verified through a physical inventory and checking of label information. (F) Certification by the responsible representative of the general licensee that they are aware of the requirements of the general license. (iv) Persons generally licensed by an Agreement State with respect to devices meeting the criteria in paragraph (c)(13)(i) of this section are not subject to registration requirements if the devices are used in areas subject to NRC jurisdiction for a period less than 180 days in any calendar year. The Commission will not request registration information from such licensees. (14) Shall report changes to the mailing address for the location of use (including change in name of general licensee) to the Director of Nuclear Material Safety and Safeguards, ATTN: GLTS, U.S. Nuclear Regulatory Commission, Washington, DC 20555-0001 within 30 days of the effective date of the change. For a portable device, a report of address change is only required for a change in the device's primary place of storage. (15) May not hold devices that are not in use for longer than 2 years. If devices with shutters are not being used, the shutter must be locked in the closed position. The testing required by paragraph (c)(2) of this section need not be performed during the period of

Page 429: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix I-35

storage only. However, when devices are put back into service or transferred to another person, and have not been tested within the required test interval, they must be tested for leakage before use or transfer and the shutter tested before use. Devices kept in standby for future use are excluded from the two-year time limit if the general licensee performs quarterly physical inventories of these devices while they are in standby. (d) The general license in paragraph (a) of this section does not authorize the manufacture or import of devices containing byproduct material. 2 Persons possessing byproduct material in devices under a general license in Sec. 31.5 before January 15, 1975, may continue to possess, use, or transfer that material in accordance with the labeling requirements of Sec. 31.5 in effect on January 14, 1975.

Page 430: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix I

Appendix 1-36

`

Page 431: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix II

Ionization Potential II

Page 432: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix II-2

Page 433: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix II-3

For your convenience, we have reproduced ionization potentials for representative compounds in the following categories:

• Some atoms and simple molecules

• Paraffins and cycloparaffins

• Alkyl halides

• Aliphatic alcohols, ethers, thiols, and sulfides

Page 434: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix II-4

Page 435: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Clarus 600 GC Hardware Guide

Appendix II-5

Page 436: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

Appendix II-6

Page 437: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

I-1

Index

A About This Manual, 1-3 Attenuation vs. detector output, 9-93 Autosampler

changing a syringe, 4 installing a syringe, 8 maintenance, 4 replacing the vial-locator mechanism, 9

Autosampler Error Messages, 10-8, 10-10

B Bead, NPD, changing, 9-56 Bead, NPD, conditioning, 9-59, 9-63

C CAP Injector Liner, 9-23 Capillary column

attaching to detector, 6-77 attaching to injector, 6-34 calculating a capillary column split ratio, 6-86 condition the column, 6-71 connect the column to the detector, 6-77 connect the column to the injector, 6-8 connect the column to the POC, 6-47 connect the column to the PSS, 6-23 installing

materials required, 6-5

overview, 6-8 summary, 6-4 tools required, 6-5

leak test, 6-68, 6-81 PSS and POC operating hints, 6-83 set up for split mode, 6-82 turn the heaters off, 6-7

Capillary column pressures, 6-62 Capillary column pressures, suggested values, 6-62 Capillary injector

fitting, 6-21 liner packing, 6-16, 9-27

Carrier Gas PPC

Mass Flow Controller, 8-6 Carrier gas flow

setting using soap bubble flowmeter, 5-12

Changing charcoal trap, 9-32 septa, 12

Charcoal trap, changing, 9-32 Chemicals

Definitions of Warnings, 2-24 Hazardous, 2-23

Clean vs. bad cuts, 6-19, 6-35 Cleaning the syringe, 11 Column

before installing, 4-3 protecting, 4-6

Column hangers, installing, 4-5

Page 438: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

I-2

Compressed gases, safety practices, 2-19 Conventions

Notes, cautions and warnings, 1-5 text, 1-5

D Detector

output vs. attenuation, 9-93 Detector window, FPD, 9-75

E ECD

cell failure, 2-16 damage, 2-16 labels, 2-16 performance, optimizing, 9-94 purchasers, 2-15 reporting radioactive exposure, 2-17 reporting theft or losses, 2-17 testing for radioactive leaks, 2-16 U.K. regulations, 2-17 wipe testing, 9-38

ECD Disposal, 9-42 ECD insulating cover, 9-39 ECD Maintenance

cleaning the anode, 9-37 ECD Maintenance, 9-35 Electrical Safety, 2-8 Electrical, safety practices, 2-9 Electromagnetic Compatibility, 1-6 EMC see Electromagnetic Compatibility, 1-6 Environmental Conditions, 2-6 Error Messages, 10-4

F FID

cross section view, 9-47 jet cleaning, 9-47 jet replacing, 9-43 performance, optimizing, 9-90

FID Maintenance, 9-43 Fitting

capillary injector, 6-21 packed column injector, 5-12

FPD cleaning

detector liner, 9-73 detector window, 9-75 optical filter assembly, 9-71

performance, optimizing, 9-95

replacing detector liner, 9-73 detector window, 9-75 optical filter assembly, 9-71 photomultiplier tube, 9-78

FPD Maintenance, 9-71

G Gas Cylinders, 2-19 GC Troubleshooting, 10-12

H Hangers, column, 4-5 heated zones, safety practices, 2-4 hourglass needle guide, replacing, 16

I Injector fittings, 4-4 Installing

capillary column materials required, 6-5 overview, 6-8 summary, 6-4 tools required, 6-5

packed column materials required, 5-5 overview, 5-3 summary, 5-4 tools required, 5-5

Ionization Potential, 0-3

J Jet assembly

FID, replacing, 9-43, 9-47 FPD, cleaning, 9-83 FPD, replacing, 9-83 NPD, replacing, 9-66

L Labels

Warning Signs, 1-8 Leak test

packed column injector fitting, 5-15

M Maintenance, 3

Page 439: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

I-3

autosampler, 4 autosampler, changing a Syringe, 4 autosampler, installing a Syringe, 8 autosampler, replacing the vial-locator

mechanism, 9 CAP injector Liner, 9-23 ECD, 9-35 ECD disposal, 9-42 ECD, cleaning the anode, 9-37 FID, 9-43 FPD, 9-71 NPD, 9-56 PID, 9-50 PID, changing lamp window seals, 9-54 PID, cleaning the lamp, 9-52 PPC, 9-86

Midpoint pressure PreVent Injector, setting, 7-18, 7-36

N NPD

bead conditioning for a packed column, 9-59, 9-63

changing a bead, 9-56 cross section view, 9-66

NPD Maintenance, 9-56

O Optical filter assembly, FPD, 9-71 Optimizing FID performance, 9-91 Oven Error Messages, 10-7 Overview

Clarus 600 GC, 3-4 Clarus 500 GC, 3-3 installing a capillary column, 6-8 PPC fundamantals, 8-3 touch screen, 3-6

P Packed column

attaching to detector, 5-19 to packed injector, 5-14

installing materials required, 5-5 overview, 5-3 summary, 5-4 tools required, 5-5

Packed column injector fitting, 5-12

Packed column injector liner changing, 13 repacking, 13

Packed injector removing the liner, 14

Packed injector liner removing, 14

Photomultiplier tube, FPD, 9-78 PID

changing a lamp, 9-50 prolonging the life of a UV lamp, 9-94

PID Maintenance changing lamp window seals, 9-54 cleaning the lamp, 9-52

PID Maintenance, 9-50 POC

operating hints, 9-96 POC operating hints, 6-83 PPC

auxiliary pressure and flow control, 8-34 capillary and PSS inlets, 8-11 carrier gas configuration, 8-16 carrier gas control, 8-5 detector gas flow control, 8-30 headspace analysis, 8-11 introduction, 8-3 mass-flow controller, 8-6 restrictor, replacing, 9-86 schematic of detector gas, 8-30 schematic of split PPC, 8-12 split control mode configuration, 8-19 theory

column temperature effects, 8-22 flow programmed operation, 8-26 pressure/oventrack operation, 8-28 velocity programmed operation, 8-27

theory of capillary column control, 8-22 tips and techniques, 8-38

correcting column dimensions, 8-38 effect of flow on pressurization rate, 8-38 pressure-pulse injection, 8-41 reducing carrier gas consumption, 8-41

vacuum compensation, 8-18, 8-29 PPC Maintenance, 9-86 PPC Restrictor Information, 9-87, 9-99 Practical hints, 9-90 Precautions, 2-5 PreVent

Column Isolation Technique, 7-68 initial pressures, 7-18, 7-36 Large Volume Injections, 7-73 Operating Techniques, 7-67 Overview, 7-3 Solvent Purge Technique, 7-71

Page 440: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

I-4

Time Saver Technique, 7-76 TurboMass

connecting TurboMass transfer line, 7-48 PreVent adapter

detector installation summary, 7-24 installing a column, 7-27 installing a restrictor, 7-24

injector installation summary, 7-5 installing a column, 7-8 installing a restrictor, 7-5 leak checking, 7-10

restrictor replacement, 7-65 TurboMass

installation summary, 7-42 installing a column, 7-45 installing a restrictor, 7-43

Programmed Pneumatic Control definition, 8-3

Protecting columns, 4-6 PSS

operating hints, 9-96 PSS operating hints, 6-83

R Radioactive exposure, reporting, 2-17 Repacking

packed column injector liners, 13 Restrictor, replacing a PPC, 9-86 Reversing TCD polarity, 9-90

S Safety

chemical use, 2-23 compressed gases, 2-19 ECD Radioactive Hazards, 2-15 electrical, 2-9 generic warnings, 2-3 heated zones, 2-4 high voltage, 2-8 moving the Clarus GC, 2-11 thermal runaway protection, 2-4 ventilation, 2-19

Safety Information, 2-5 electrical safety, 2-8 environmental conditions, 2-6 gas cylinders, 2-19

Septa, changing, 12 Setting up

split mode, 6-82 Soap bubble flowmeter

using, 5-12 Spare components, 10-13 Spare Components, 10-13 Split liner, 6-17, 9-25 Split mode

setting up, 6-82 Splitless liner, 6-17, 9-25 Syringe

cleaning, 11 servicing an idle, 11

T TCD polarity, reversing, 9-90 Top cover hold down screws, 9-32 Touch Screen, 3-6 Troubleshooting

autosampler error messages, 10-8, 10-10 dual identical channels, 10-14 error messages, 10-4 formatted errors, 10-9 oven error messages, 10-7 spare components, 10-13 steps, 10-13

V Ventilation, safety practices, 2-19

W Warnings

Hazardous Chemical, 2-24 Wipe test

See Wipe testing ECD, 9-38 Wipe testing

ECD, 9-38

Page 441: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

WEEE Instructions for PerkinElmer Products

or

A label with a crossed-out wheeled bin symbol and a rectangular bar indicates that the product is covered by the Waste Electrical and Electronic Equipment (WEEE) Directive and is not to be disposed of as unsorted municipal waste. Any products marked with this symbol must be collected separately, according to the regulatory guidelines in your area. The objectives of this program are to preserve, protect and improve the quality of the environment, protect human health, and utilize natural resources prudently and rationally. Specific treatment of WEEE is indispensable in order to avoid the dispersion of pollutants into the recycled material or waste stream. Such treatment is the most effective means of protecting the customer’s environment. Requirements for waste collection, reuse, recycling, and recovery programs vary by regulatory authority at your location. Contact your local responsible body (e.g., your laboratory manager) or authorized representative for information regarding applicable disposal regulations. Contact PerkinElmer at the web site listed below for information specific to PerkinElmer products. Web address: http://las.perkinelmer.com/OneSource/Environmental-directives.htm For Customer Care telephone numbers select “Contact us” on the web page. Products from other manufacturers may also form a part of your PerkinElmer system. These other producers are directly responsible for the collection and processing of their own waste products under the terms of the WEEE Directive. Please contact these producers directly before discarding any of their products. Consult the PerkinElmer web site (above) for producer names and web addresses.

Page 442: Clarus 600 GC Hardware Guide - PerkinElmer€¦ · Clarus 600 GC Hardware Guide 1-5 Conventions Used in this Manual Normal text is used to provide information and instructions. Bold

WEEE Instructions for PerkinElmer Products

or

A label with a crossed-out wheeled bin symbol and a rectangular bar indicates that the product is covered by the Waste Electrical and Electronic Equipment (WEEE) Directive and is not to be disposed of as unsorted municipal waste. Any products marked with this symbol must be collected separately, according to the regulatory guidelines in your area. The objectives of this program are to preserve, protect and improve the quality of the environment, protect human health, and utilize natural resources prudently and rationally. Specific treatment of WEEE is indispensable in order to avoid the dispersion of pollutants into the recycled material or waste stream. Such treatment is the most effective means of protecting the customer’s environment. Requirements for waste collection, reuse, recycling, and recovery programs vary by regulatory authority at your location. Contact your local responsible body (e.g., your laboratory manager) or authorized representative for information regarding applicable disposal regulations. Contact PerkinElmer at the web site listed below for information specific to PerkinElmer products. Web address: http://las.perkinelmer.com/OneSource/Environmental-directives.htm For Customer Care telephone numbers select “Contact us” on the web page. Products from other manufacturers may also form a part of your PerkinElmer system. These other producers are directly responsible for the collection and processing of their own waste products under the terms of the WEEE Directive. Please contact these producers directly before discarding any of their products. Consult the PerkinElmer web site (above) for producer names and web addresses.