chm 418, experiment 08 - design of experiments using a hewlett-packard 5890 series ii gas...
TRANSCRIPT
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 1/28
Design of Experiments using a Hewlett-Packard 5890 Series II Gas
Chromatography Machine
Tyler Hacker, W. Daniel Smith & Nathaniel Wise
Bob Jones University, Division of Science, Department of Chemistry 1700 Wade Hampton Blvd. Greenville, SC 29614
Abstract
Chromatography is an important analysis tool for the chemist. Learning to properly utilize
instruments available will aid tremendously in the chemist’s career. To that end, this lab group
performed a design of experiments lab, using software to predict the optimal conditions for
adequate resolution of peaks and minimum retention time. Thirty samples were run with different
values for several variables. These were then used to give conditions for a final set of three runs, allof which exhibited good resolution and tolerable retention times.
Introduction
Gas chromatography is one of the more
widely used types of chromatography. It is
used for the separation and analysis of
compounds that can be vaporized. Many
times, gas chromatography is used to test thepurity of a substance, or to help in the
identification of the solution analyzed. In
separation chromatography, there are two
phases. The first is the mobile phase, in which
the sample is dissolved. The mobile phase is
then forced through the stationary phase,
which is fixed in place on a column. The
mobile phase in gas chromatography is a
carrier gas, which can be either an inert gas or
an unreactive gas. For this experiment,nitrogen, and unreactive gas, was used as the
carrier gas. The stationary phase for this
experiment was inside a fused-silica capillary
column. Different compounds being analyzed
interact with the stationary phase on the walls
of the column, and this causes the
compounds to pass through the column, or to
be eluted, at different times which are called
retention times. Many different types of
detectors are used with gas chromatography.
The most common of these are flame
ionization detectors (FID), which are more
sensitive to hydrocarbons. Effluent from the
column is passed through a hydrogen flame, which will produce ions and electrons at the
temperature of the flame. The current
produced by these charges are monitored as
the ions and electrons move toward the
collector electrode, located above the flame.
The resulting current is measured, and the
chromatogram is acquired.
In this experiment, four fatty acid methyl
esters were analyzed. This was done by
varying the initial injection temperature, initial
column oven temperature, final column oven
temperature, and the temperature ramping
rate. Using these data, a design of experiment
(DOE) was conducted. From the acquired
data from the first part of the experiment,
calibration curves were generated. With these
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 2/28
curves, the ideal separation conditions for the
fatty acid methyl esters could be predicted.
Also, unknown samples could be assessed to
determine their identity as well.
Another example of the use of calibrationcurves is found in an article found in the
Journal of Analytical Chemistry. The authors
in the reported experiment needed to find the
best coating possible for the materials used to
make stents. In order to do this, calibration
curves were established from data acquired
from confocal Raman microscopy
measurements. From the values obtained
from the calibration curve, a coating with the
right percentages of the chemicals sirolimus,PBMA, and PEVA along with others, were
found. (Balss, K., et al. Multivariate Analysis
Applied to the Study of Spatial Distributions Found
in Drug-Eluting Stent Coatings by Confocal Raman
Microscopy . J. Anal. Chem. 2008, 80, 4853–
4859.)
Also, and example of the use of gas
chromatography is found in an article in the
Journal of Agriculture and Food Chemistry.For this experiment, the authors used gas
chromatography along with chemometric data
analysis and solid-phase microextraction to
determine whether or not apple puree had
been added to samples of strawberry. When
comparing the chromatograms of adulterated
strawberry and unadulterated strawberry
samples, the differences were considered
obvious. However, the authors concluded that
only the sample adulterated to about seventy percent puree was easily distinguishable.
(Reid, L., et al. Potential of SPME-GC and
Chemometrics To Detect Adulteration of Soft Fruit
Purees . J. Agric. Food Chem. 2004, 52, 421-
427.)
Methodsand
Materials
This experiment used a Hewlett-Packard 5890
Series II gas chromatography machine. After
prepping the GC instrument and allowing it
to warm up, we programmed the auto sampler
to the conditions specified. The initial time
was set to 1.00 min; final time to 25 min;
dectector temperature to 250° C; and split
ratio to 75/1. The instrument automatically
ran through all the samples, and we processed
the data on the computer. It is presentedbelow. Once the computer gave us the
optimal values for the final runs, we
programmed those in and completed the
experiment.
Results&
Discussion
In designing our experiment, we operated on
four variables that affect the elution times and
the resolution of the gas chromatography
instrument. The first variable was the
injection temperature with a range of 200-
300°C. The second was the initial oven
temperature with a range of 100-200°C. The
third was the oven ramp rate with ranges of
5°C/min-50°C/min. The fourth variable we
set limits for was the final oven temperatureof 180-220°C.
By modifying these variables, we had four
outcomes that we looked at: the last peak
elution time which we wanted to be
minimized, and the three resolution values
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 3/28
between the four fatty acid methyl ester peaks
which we wanted to be maximized.
The DOE took our limiting variables and
gave us a list of specific variables to test.
These specific runs we limited to 30 and there
are 6 duplicates to test for instrumental erroras well as several outliers to make sure our
confining limits are reasonably set. The
remaining of the 30 runs were selected
through a randomization process that the
DOE will analyze later. The following chart
gives the list of variables the DOE had us test
and their corresponding results.
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 4/28
To determine the peak resolution, we used the output from the GC in this format.
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 5/28
By using the peaks width listed here and the retention time, we can determine the resolution by
this formula.
(Note: Within
our
calculations,
we
did
not
need
to
multiply
by
2
in
the
numerator
because
the
width values given in the GC readout were ½ height widths.)
The following pictures are the DOE analysis of the data and its prediction:
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 6/28
Estimate of error in the model.
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 7/28
Last peak retention time analysis:
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 8/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 9/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 10/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 11/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 12/28
Resolution between peaks 1-2 analysis:
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 13/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 14/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 15/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 16/28
Resolution between peaks 2-3 analysis:
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 17/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 18/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 19/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 20/28
Resolution between peaks 3-4 analysis:
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 21/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 22/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 23/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 24/28
Optimum conditions and prediction:
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 25/28
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 26/28
To verify these predictions we did 3 runs. The results follow.
RUN 1 RetTime(min Width(min)
1 0.762 7.55E-03time last peak eluted 7.391
2 2.645 1.44E-02 rs. b/t 1-2 19.81707
3 3.295 0.0184 rs. b/t 2-3 19.757714 4.192 0.027 rs. b/t 3-4 38.77576
5 7.391 0.0555
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 27/28
RUN 2 RetTime(min Width(min)
1 0.765 7.67E-03time last peak eluted 7.403
2 2.648 1.43E-02 rs. b/t 1-2 19.96933
3 3.299 0.0183 rs. b/t 2-3 20.08949
4 4.197 0.0264 rs. b/t 3-4 37.19258
5 7.403 0.0598
8/3/2019 Chm 418, Experiment 08 - Design of Experiments Using a Hewlett-Packard 5890 Series II Gas Chromatography Ma…
http://slidepdf.com/reader/full/chm-418-experiment-08-design-of-experiments-using-a-hewlett-packard-5890 28/28
RUN 3 RetTime(min Width(min)
1 0.766 7.88E-03time last peak eluted 7.408
2 2.65 1.40E-02 rs. b/t 1-2 19.90826
3 3.301 0.0187 rs. b/t 2-3 19.97778
4 4.2 0.0263 rs. b/t 3-4 38.83777
5 7.408 0.0563
For each of these runs, the experiment
values of time of last peak and the resolution
values between the peaks were within 2
standard deviations of the predicted value of
the DOE.
The experimental data we obtained
was accurate and very usable; however, the
process can be sped up. During in-class
discussion we determined that a resolution
value over 1 is sufficient and that the process
can be done in much faster time with
acceptable results. After talking with Dr. Lee,
we determined that the reason the DOE did
not predict these results was because we did
not set up a full hierarchy in terms of
importance and value of results. This means
that we did not specify that a smaller last peak
elution time is much more important than a
large resolution value and that any value over
1 is quite sufficient. Because of this, the DOEsought to find the best medium between the
maximized resolution values and the
minimized last peak elution value.