chapter 4 1 atomic spectrocopy

Upload: lindaamharis

Post on 03-Apr-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    1/34

    Chapter 4

    Atomic Spectroscopy

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    2/34

    Optical spectroscopic methods

    In Optical spectrometry, elements in thesample are converted to gaseous atoms orelementary ions by a process calledatomization. The absorption, emission, or

    fluorescence of the atomic species in thevapor is then measured.

    Determination of analytical sample

    concentration Linear relationship between absorbance and

    concentration

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    3/34

    Optical spectroscopic methods

    Atomic Absorption Spectroscopy

    Atomic Emission Spectroscopy

    Atomic Fluorescence Spectroscopy

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    4/34

    Atomic Absorption Spectroscopy

    Light Source Monochromator Detector Amplifier

    E.g. Hollow

    cathode lamp

    Analyte solution

    Atomiser Fuel (e.g. acetylene)Air

    I0 It

    Nebuliser, spray

    chamber, and burner

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    5/34

    How does it work?

    The solution is aspirated into the flame as a fine spray

    The solvent evaporates, leaving the dehydrated salt

    The salt is dissociated into free gaseous atoms in the ground state

    A certain fraction of these atoms absorbed energy from the flame

    (some of them are collided each other) and be raised to the excited

    electronic state

    The excited state have a short lifetime and returned to the ground

    state by emitting photon with characteristic wavelength,

    The intensity of emission/absorption is directly proportional to theconc. of analyte in the solution, therefore a calibration curve of

    emission/absorption intensity vs conc. is prepared

    However, side reactions in the flame may decrease the population

    of free atom and hence reduced the emission/absorption signal

    hcE

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    6/34

    Atomization

    Desolvation and vaporization of ions or atoms

    in a sample: high-temperature source such as

    flame or graphite furnace

    Flame atomic absorption spectroscopy

    Graphite furnace atomic absorption spectroscopy

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    7/34

    Flame atomic absorption spectroscopy

    A flame provides a high-temperature source for desolvating and

    vaporizing a sample to obtain free atoms for spectroscopic

    analysis. In atomic absorption spectroscopy ground state atoms

    are desired. For atomic emission spectroscopy the flame must

    also excite the atoms to higher energy levels. The table lists

    temperatures that can be achieved in some commonly usedflames.

    Temperatures of some common flames

    Fuel Oxidant Temperature (K)H2 Air 2000-2100

    C2H2 Air 2100-2400

    H2 O2 2600-2700

    C2H2 N2O 2600-2800

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    8/34

    Processes

    In

    Flame

    M*excited

    salt vapourised

    *ATOMISED

    *

    Solution MXLiquid aerosol droplets

    Salt mist of MXMolecules of MX

    M

    M+

    ionised

    MX

    compound

    formed

    nebulisation

    solvent evaporation

    Dissociation

    Thermal and chemical

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    9/34

    Effect of temperature-boltzmann

    distribution

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    10/34

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    11/34

    Source of radiation

    Hollow cathode lamps or electrode lessdischarge lamps

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    12/34

    Hollow cathode lamp

    Sources: hollow-cathode lamp (HCL), used in atomic absorbtion

    Tube filled with inert gas (Ne or Ar)

    Hollow cathode (negative) made with metal we want to detect

    Run a high voltage between anode and cathode

    This makes Ne or Ar ionize

    Ne+ or Ar+ attracted to hollow metal cathode

    As these ions hit the metal, atoms of metal are ejected into the gas

    As metal atom interact with energetic electrons atoms are excited,

    so generate light at that metals wavelength

    since excitation is not by flame the linewidth is extra sharp

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    13/34

    Monochromator

    Isolation of the absorption line from

    background light and from molecular

    emissions originating in the flame (tuned to a

    specific wavelength)

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    14/34

    Detection

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    15/34

    Calibration curve

    Standards containing known concentrations of the analyte areintroduced into the instrument

    Response is recorded

    Response is corrected for instrument output obtained with a

    blank Blank contains all of the components of the original sample except for

    the analyte

    Resulting data are then plotted to give a graph of correctedinstrument response vs. analyte concentration

    An equation is developed for the calibration curve by a least-squares technique so that sample concentrations can becomputed directly

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    16/34

    Detector

    A photomultiplier measures the intensity ofthe incident light and generate an electrical

    signal proportional to the intensity

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    17/34

    Interference

    Lowers the signal Chemical interference: formation of stable or

    refractory compounds (not atomized at certain T)

    Use higher temperature

    Releasing agents, EDTA

    Chelating agents, LaCl3

    Ionization interference

    Alkali metals easily ionized (low ionization energy) Energy level of ion lower than parent

    Suppressed by adding elements that are easily

    analyzed, CsCl when using Na or K

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    18/34

    Disadvantages:

    only solutions can be analysed

    relatively large sample quantities required (1 2 mL)

    problems with refractory elements

    Advantages:

    inexpensive (equipment, day-to-day running)

    high sample throughput easy to use

    high precision

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    19/34

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    20/34

    Atomic Emission Spectroscopy

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    21/34

    How it works?

    A plasma source is used to dissociate the

    sample into its constituent atoms or ions,

    exciting them to a higher energy level.

    They return to their ground state by emitting

    photons of a characteristic wavelength

    depending on the element present.

    This light is recorded by an optical

    spectrometer.

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    22/34

    Plasma: atomization and excitation source A plasma is an electrically neutral,

    highly ionized gas that consists of ions,

    electrons, and atoms

    A high-frequency current of 27.120MHz

    is sent through a high-frequency coil,

    generating a magnetic field. Plasma is then

    formed from the gas (argon) flowing through

    the coil. In the plasma, there are the samenumber of Ar+ and electrons, maintaining an

    electrical balance.

    Temperature reaching 10000K

    Advantages over flame:

    Lower inter-element interference (higher temperature With a single set of conditions signals for dozens of elements can be

    recorded simultaneously

    Monochromator

    Detector

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    23/34

    Interferences

    two or more elements in the matrix emitting radiationat the same wavelength (e.g., Cu at 515.323 nm and Arat 515.139 nm). These spectral interferences can beminimized by using a high resolution system by using

    several analytical lines for the detection of a singleelement

    interference involving the formation of undesiredspecies (e.g., ions). It is important to note that an atom

    of a specific element (e.g., Fe) has a different emissionspectra than one of its ions (e.g., Fe+, Fe+2, etc.)

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    24/34

    Atomic Fluorescence Spectroscopy

    Determination of single elements in analyticalsample

    Determination of analytical sample concentration

    Monitor the fluorescence emission from theexcited state

    Atomic fluorescence spectra:

    Resonance fluorescence

    Radiationless transition + transition to ground state

    Transition to lower state + radiationless transition

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    25/34

    Atomic Fluorescence Spectroscopy

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    26/34

    How it works?

    the gaseous atoms obtained by flame orelectrothermal atomisation are excited to higherenergy levels by absorption of the

    electromagnetic radiation and the fluorescenceemission from these excited atoms is measured.

    This technique incorporates aspects of bothabsorption and emission.

    Measure the fluorescence emission resultingfrom the relaxation of the excited atoms.

    However, measurement is made 90o from source

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    27/34

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    28/34

    Advantages and application

    The main advantage of fluorescence detectioncompared to absorption measurements is the greatersensitivity achievable because the fluorescence signalhas a much lower background as compared to the one

    observed in atomic absorption method Except for some favorable metals and metalloids (like

    Pb, Cd, Tl, Se, Te, As, Sb, etc.) there is no specialadvantage over more established AAS. Thus, not

    widespread use due to overwhelming success of AAS Applications: analysis of metals in lubricating oil,

    seawater, biological substance, agriculture samples

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    29/34

    Sample Preparation

    The sample must be in the diluted form and

    filtered for particulates Type of sample: blood, urine, tissues,

    cerebral spinal fluid and other biologicalfluids by direct aspiration of the sample,

    usually dilution with water is required toprevent clogging of the burner

    In the preparation of standards, the matrix

    of the analyte must always be matched, Eg.Analysing Zn in waste water, standardsolution is made up from ZnCl2

    d d f h l ( b

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    30/34

    Case study: determination of heavy metals (Pb, Cr,

    As and K) in Selom plant

    Raw Materials : Selom plant

    Chemicals: nitric acid and hydrochloric acid with ratio of 3:1and distilled water

    Procedure: dry the fresh Selom in the oven before ashingprocess to avoid unnecessary explosion

    Burn the dry selom in the furnace at 500C for one hour

    About 5 g of Selom ash is requiredbefore acid digestionprocess

    The solution is then filtered and diluted to the requiredconcentration for AAS and/or ICP-OES analysis

    Prepared the standard solution for each metal and obtained

    their calibration curve From the regresion equation the concentration of each

    element can be calculated

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    31/34

    Ashing method

    placingthe sample in an open inert vesseland destroying the combustible (organic)

    portion of the sample by thermal

    decomposition using a muffle furnace.Typical ashing temperatures are 450 to 550

    C. Magnesium nitrate is commonly used as

    an ashing aid. Charring the sample prior to

    muffling is preferred. Charring is

    accomplished using an open flame.

    A id di ti /

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    32/34

    Acid digestion/pressure

    digestion/microwave digestion

    Acid digestion process are employed for thedetermination of elements in solidsubsequent to sampling and mechanical

    sample preparation in order to completelytransfer the analytes into solution so that itcan be introduced to analysis instrument- ICP-OES, AAS and ICP-MS

    Common acid used, mineral acids (HCl, HNO3,HF, H2SO4 etc)

    Representatives Detectin Limit by AAS and Flame

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    33/34

    Representatives Detectin Limit by AAS and Flame

    Emission Spectroscopy, FES/ICP

    Element Wavelength (nm)Detection Limit (ppm)

    AAS FES (ICP)

    Ag 328.1 0.001(A) 0.01

    Al 309.3 0.1 (N)

    Au396.2 0.08

    242.8 0.03 (N)

    Ca267.7 3

    422.7 0.003 (A) 0.0003

    Cu 324.8 0.006 (A) 0.01

    Eu 459.4 0.06 (N) 0.0008Hg 253.6 0.8 (A) 15

    K 766.5 0.004 (A) 0.00008

    Mg 285.2 0.004 (A) 0.1

    Na 589 0.001 (A) 0.0008

    Tl276.8 0.03 (A)

    535 0.03

    Zn 213.9 0.001(A) 15

    Detection Limit (ppm): The conc required to give a signal equal to three times the

    standard deviation of base line (Blank)

    < 300 nm: AAS shows superior detectability because high thermal energy required

    to excite the atom for emission at these wavelength

    300 < < 400 nm: either method exhibit comparable detectability

  • 7/28/2019 Chapter 4 1 atomic spectrocopy

    34/34