chapter 2 characterization...
TRANSCRIPT
35
CHAPTER 2
CHARACTERIZATION TECHNIQUES
2.1 INTRODUCTION
A complete description of the physical and chemical properties of a
material of interest is termed as characterization of that material. Characterization of a
crystal essentially consists of determination of chemical composition, structure,
defects, study of their optical properties etc [99].
The study of growth defects includes the defects such as inclusions,
mechanical stress etc., that result due to poor control of crystal growth parameters.
Standard techniques such as etching, chemical analysis, X-ray topography, scanning
electron microscopy etc., can be used for defect characterization. The measurement of
optical properties includes the study of optical transmission and absorption of the
crystal, SHG conversion efficiency, NLO coefficients, electro-optical coefficients and
structural dependence of these properties. Characterization of NLO crystals can
generally be divided into three types:
i) Crystal structural analysis
ii) Investigation of growth defects and
iii) Measurement of optical properties
The single crystals chosen for the present investigation were subjected to
the following studies:
Collection of XRD data by single crystal X-ray diffraction analysis
and Powder X-ray diffraction.
36
FTIR spectral analysis to confirm the various functional groups
present in the compound.
CHNS analyses to identify the percentage of carbon, hydrogen,
nitrogen and sulphur in the sample.
UV-Vis-NIR analysis for the detection of the transparency region
and cutoff wavelength.
Thermal analysis to study the thermal stability.
Hardness studies to estimate the hardness.
Dielectric measurements and AC conductivity studies to investigate
the dielectric response of the crystal with varying frequency.
Etching studies to study the behavior of the crystals under different
solvents.
SHG test to check the nonlinear response of the crystal to the
incident coherent light.
Laser damage threshold studies to determine the resistance provided
by the crystal to the laser beam.
2.2 SINGLE CRYSTAL X-RAY DIFFRACTION ANALYSIS
Single crystal X-ray diffraction (X-ray crystallography) is an analytical
technique in which X-rays are employed to determine the actual arrangement of atoms
within a crystalline specimen. Single crystal X-ray diffraction is a non-destructive tool
to analyze crystal structure of compounds, which can be grown as single crystals. The
molecular structure, atomic coordinates, bond lengths, bond angles, molecular
orientation and packing of molecules in single crystals can be determined by X-ray
crystallography. Single crystal X-ray diffractometer collects intensity data required
for structure determination.
37
Accurate measurements of intensities of reflections of all Miller indices
within a specified reciprocal radius (usually 25 for MoK and 68 for CuK ) is
needed to find the structure, while unit cell parameters depend only on direction of
reflections. As the name implies, a crystalline sample is required. For single-crystal
work, the specimen should be smaller than cross section diameter of the beam. Larger
crystals can be cut down to proper size and smaller crystals may be suitable if they
contain strongly diffracting elements.
The monochromatic X-rays incident on a plane of single crystal at an angle
theta are diffracted according to Bragg's relation, 2d sin = n where ‘d’ is the
interplanar spacing of the incident plane, ‘ ’ is the wavelength of X-rays and ‘n’ is a
positive integer. The intensity of the diffracted rays depends on the arrangement and
nature of atoms in the crystal. Collection of intensities of a full set of planes in the
crystal contains the complete structural information about the molecule. Fourier
transformation techniques are used to determine the exact coordinates of atoms in the
unit cell from this data.
With the set of X-ray diffraction data collected, unit cell parameters, space
groups, molecular structure of the crystalline solids and Miller indexing the different
faces of the crystal are possible. Unit cell parameter is simply the dimension of the
basic molecular brick with which the crystal is built. Space group tells us the
symmetry with which the molecules are arranged within the unit cell. All the
geometrical features of molecules (bond distance, bond angles, torsion angles between
bonds, dihedral angles between planes etc.,) may be obtained from coordinates.
In the present study, the single crystal X-ray diffraction analysis was
performed using an ENRAF NONIUS CAD4-F single crystal X-ray diffractometer.
The shield was equipped with graphite monochromated MoK radiation. Since the
crystal was transparent, the single crystallinity was studied with Leica polarizing
microscope. Single crystal of suitable size was cut and mounted on the X-ray
goniometer. The sample is mounted on a thin glass fiber that is attached to a brass pin
and mounted on a goniometer head. Adjustment of the goniometer head in the X, Y
and Z orthogonal directions allows centering of the crystal in the X-ray beam. The
38
unit cell dimensions (a, b, c, , , ) can be determined from the accurately measured
values (the angular deviation from the direct undeviated beam of about 25
reflections. The reflections were indexed using method of short vectors followed by
least square refinements. The unit cell parameters thus obtained were transformed to
correct Bravais cell.
2.3 POWDER X-RAY DIFFRACTION ANALYSIS
The powder method of diffraction was devised independently by Debye
and Scherrer. It is the most useful of all diffraction methods and when properly
employed, can yield a great deal of structural information about the material under
investigation. Powder diffraction method involves the diffraction of monochromatic
X-rays by a powder specimen. Monochromatic usually means a strong K
characteristic component of the filtered radiation from an X-ray tube operated above
the K excitation potential of the target material.
Selection of K renders the incident beam to be a highly monochromatised
one. The focusing monochromatic geometry results in narrower diffracted peaks and
low background at low angles. The sample is mounted vertically to the
Seemann-Bohlin focusing circle with the scintillation counter tube moving along the
circumference of it. It is possible to record the diffracted beam from 2 to 160 degrees.
The diffractometer is connected to a computer for data collection and analysis. The
scintillation counter tube can be moved in step of 0.01 degree by means of a stepper
motor and any diffracted beam can be closely scanned to study the peak profile.
High Resolution GUINIER Powder X-ray diffractometer (SEIFERT,
Germany) has been used for phase analysis as well as for calculation of the lattice
parameters. This diffractometer can record diffraction pattern at lower angles (starting
from 1°) with very little background and good resolution. The fine line focused beam
(0.4 mm wide and 0.8 mm in height) generated from the Cu-target, is used at the focal
plane in the present study. A bent primary quartz monochromatic situated close to the
target is used to select the CuK radiation.
39
The powder diffraction of a substance is characteristic of the substance and
forms a sort of fingerprint of the substance to be identified. The peaks of the X-ray
diffraction pattern can be compared with the standard available data for the
confirmation of the structure. For the purpose of comparison, many standards are
available, some of which are, Willars Hand book, Joint Committee on Powder
Diffraction Standards (JCPDS) Pcpdfwin and National Bureau of Standards.
2.4 FOURIER TRANSFORM INFRARED (FTIR) SPECTRAL
ANALYSIS
Fourier Transform Spectroscopy is a simple mathematical technique to
resolve a complex wave into its frequency components. The conventional IR
spectrometers are not of much use for the far IR region, as the sources are weak and
the detectors are insensitive. FTIR has made this energy-limited region more
accessible. It has also made the mid infrared more useful. Conventional spectroscopy,
called the frequency domain spectroscopy, records the radiant power as a function of
frequency. In the time domain spectroscopy, the change in radiant power is recorded
as a function of time. In the Fourier Transform Spectrometer, a time domain plot is
converted into a frequency domain spectrum. The actual calculation of the Fourier
transform of such systems is done by means of high – speed computers.
The FTIR spectrometer consists of an infrared source, a sample chamber
with a provision for holding solids, liquids and gases, monochromator, a detector and
a recorder, which are integrated with a computer. At present, all commercially
available infrared spectrophotometers employ reflection gratings rather than prisms as
dispersing elements. Interferometric multiplex instruments employing the Fourier
transform are now finding more general applications in both qualitative and
quantitative infrared measurements. The interference pattern is obtained from a two
beam interferometer, as the path difference between the two beams is altered, then
Fourier transformed output gives rise to the spectrum. The transformation of the
interferogram into spectrum is carried out mathematically with a dedicated online
computer. The spectrometer consists of globar and mercury vapour lamp as sources.
An interferometer chamber comprising of KBr and mylar beam splitters is followed
40
by a sample chamber and detector. The schematic diagram of a FTIR spectrometer is
shown in Figure 2.1. The spectrometer works under vacuum condition. Solid samples
are dispersed in KBr or polyethylene pellets depending on the region of interest. This
instrument has resolution of 0.1 cm–1. Signal averaging, signal enhancement, baseline
correction and other spectra manipulations are possible with multitasking OPUS
software on the dedicated PC/AT 486. Spectra are plotted on a HP plotter and data
can be printed.
Figure 2.1 Schematic diagram of a FTIR spectrometer
Recording of IR spectra of solid sample is more difficult because the
particles reflect and scatter the incident radiation and therefore transmittance is
always low. Three different techniques are employed commonly in recording such
spectra. For solid compounds, that are insoluble in the usual solvents, a convenient
sampling method is the Pressed Pellet Technique. A few milligrams of the sample are
ground together in an agate or mullite mortar with about 100 times the quantity of a
material (the matrix) transparent to the infrared. The usual material is KBr, although
Collimated lamp for visual alignment
Oscillating mirror ‘C’
Oscillating mirror ‘A’
Oscillating mirror ‘B’
Interferometer
IR Source
Sample component
Reference holder
Sample holder
TGS Detector
Microsampling position
41
other compounds such as CsI, TlBr and Polyethylene are used in special
circumstances. The ground powder is finally introduced into a mini pressing
arrangement made from two half-inch diameter stainless steel bolts and a stainless
steel nut. The ends of the bolts must be polished, flat and parallel. One bolt is
inserted about half way into the nut and the KBr plus sample mixture. The second bolt
is then screwed into the nut and pressure is applied by tightening the bolts
together. When the bolts are carefully withdrawn, a pellet suitable for infrared
transmission work, remains. The pellet is not removed from the nut, which acts as a
holder in the spectrometer [100].
Molecular vibrations that oscillate with the same frequency absorb IR
light. The frequency of the vibration and the probability of absorption are influenced
by intra and intermolecular effects. Thus, information about structure and
environment can be deduced from the spectral parameters, bandwidth and absorption
coefficient. There are no rigid rules for interpreting the vibration spectrum. Certain
requirements, however, must be met before an attempt is made to interpret a
spectrum :
i) The spectrum must be adequately resolved and of adequate
intensity.
ii) The spectrum should be that of a reasonably pure compound.
iii) The spectrometer should be calibrated.
iv) The method of sample handling must be specified. If a solvent is
employed, the solvent, concentration and the cell thickness should be
indicated.
A precise treatment of the vibrations of a complex molecule is not feasible;
thus, the spectrum must be interpreted from empirical comparison of spectra and
extrapolation of studies of similar molecules. Many of the group frequencies of
organic compounds vary over a wide range. Because the bands arise from complex
interacting vibrations within the molecule, absorption bands may, however represent
predominantly a single vibration mode. Important details of structure may be revealed
42
by the exact position of an absorption band within a narrow region. Shifts in
absorption position and changes in band contours, accompanying changes in
molecular environment, may also suggest important structural details
Some of the general uses of FTIR spectra are :
i) Identification of all types of organic and many types of
inorganic compounds
ii) Determination of functional groups in organic materials
iii) Determination of the molecular composition of surfaces
iv) Identification of chromatographic effluents
v) Quantitative determination of compounds in mixtures
vi) Determination of molecular conformation (structural isomers) and
stereochemistry (geometrical isomers)
vii) Determination of molecular orientation (polymers and solutions)
2.5 CHNS ANALYZER
Elemental analyzer technique determines the presence of the elements like
carbon, nitrogen, hydrogen and sulphur in a given sample and gives the result as
percentage amount of these against the total weight. This technique specially
determines these three elements and is also called “CHNS analyzer”. Most of the
organic compounds are made up of these three elements and oxygen; hence after
determining these three elements the percentage weight of oxygen can be easily
calculated [101].
This is an important study when it comes to a newly synthesized
compound to prove its exact composition. Solid samples weighing from less than
1 mg up to 200 mg are packed in tin boats. Samples are dropped into the combustion
tube automatically at user selected temperature i.e. 1000 ºC by combustion to tin
boats. The principle of the instrumentation is that the substance under study is
combusted under oxygen stream in a furnace at high temperatures (1000 °C). In the
43
combustion process, carbon is converted to carbon dioxide; hydrogen to water;
nitrogen to nitrogen gas/oxides of nitrogen and sulphur to sulphur dioxide. That is, the
end product of the combustion would be mostly the oxides of the concerned elements
in the form of gases. These are then separated and carried to the detector using inert
gases helium or argon. The Vario EL III-Germany CHNS analyser was used in the
present study.
It gives a clear quantitative measurement of the carbon, hydrogen and
nitrogen. It finds applications in almost every field of chemistry like the analysis of
organics, polymers, pharmaceuticals, energy fuels, environmental studies etc.
2.6 UV - VISIBLE SPECTROSCOPY
Ultraviolet-visible spectroscopy (UV/ VIS) is also known as electronic
spectroscopy. Ultraviolet (200 - 400 nm) and visible (400 - 800 nm) absorption
spectroscopy is the measurement of the attenuation of a beam of light after it passes
through a sample or after reflection from a sample surface. The schematic
representation of a UV-Vis spectrophotometer is shown in Figure 2.2. It uses light in
the visible and adjacent near ultraviolet (UV) and near infrared (NIR) ranges. In this
region of energy space molecules undergo electronic transitions.
Ultraviolet and visible light are energetic enough to promote outer
(valence) electrons to higher energy levels. Valence electrons are found in three types
of electron orbitals namely bonding orbitals, bonding orbitals and non-bonding
orbitals (n-lone pair electrons). Sigma ( ) bonding orbitals tend to be lower in energy
than bonding orbitals, which in turn are lower in energy than non-bonding orbitals.
The unoccupied or anti bonding orbitals ( * and *) are the orbitals of highest energy.
An energy level diagram showing electronic transitions is depicted in Figure 2.3. Of
the six transitions outlined, only the two lowest energy ones (left-most, coloured blue)
are achieved by the energies available in the 200 to 800 nm spectrum. As a rule,
energetically favored electron promotion will be from the highest occupied molecular
orbital (HOMO) to the lowest unoccupied molecular orbital (LUMO) and the
resulting species is called an excited state.
44
Figure 2.2 Schematic representation of a UV-Vis-NIR spectrophotometer
Figure 2.3 Energy level diagram with electronic transitions
EN
ER
GY
n * *
n *
*
* (anti – bonding)
* (anti – bonding)
n (non – bonding)
( bonding)
(bonding)
45
In particular, absorption bands results from transitions ( *) and n *
involving -orbitals and lone pairs (n = non-bonding) are important and so
UV-Vis spectroscopy is of most use for identifying conjugated systems which tend to
have stronger absorptions. Absorption bands can also arise from certain molecules
where the energy required for removing an electron from one atom and placing it on
another falls within the UV/Vis region. This process is known as a charge transfer
excitation. Molecules with the ability to exhibit the above types of electronic
transitions are said to possess chromophores. An isolated functional group not in
conjugation with any other group is said to be a chromophore if it exhibits absorption
of a characteristic nature in the ultraviolet or visible region. The most common are
(C=C) and (C=O) which exhibits * and n * transitions respectively.
If a series of compounds have the same functional group and no
complicating factors are present, all of them will generally absorb at very nearly the
same wavelength. Thus, it is readily seen that the spectrum of a compound, when
correlated with data from the literature for known compounds, can be a very valuable
aid in determining the functional groups present in the molecule.
Samples are typically placed in a transparent cell, known as a cuvette. The
sample holders (cuvettes) are the rectangular shaped quartz or glass cells of about
10 mm path length. The transmitted light radiation is received at the photomultiplier
tube alternately from the reference and the sample beams. A photoelectric signal
timing system is synchronized with the alternate pulses which permits the comparison
of signals from the two beams. The difference between the two signals is recorded
with the help of a motor driven pen or is interfaced with a PC-XT and stored there for
easy reference. Samples in solid form, powder, pellets are dissolved in suitable
solvents to form the contents of the sample cell and the solvents are taken in the
reference cell. In the present work, the UV-Visible spectra was recorded using Perkin
Elmer Lamda Instrument.
46
2.7 THERMAL STUDIES
Thermal analysis is a very essential method to study the thermal behaviour of materials and finds widespread applications in diverse industrial and research fields. It is a general term, which covers a group of related techniques in which the temperature dependence of the parameters of any physical property of a substance is measured. In addition to providing valuable information on the thermal stability of the compounds and the decomposition products, these studies often provide an insight into their mode of decomposition.
Thermoanalytical methods involve the measurement of various properties of materials subjected to dynamically changing environments under predetermined conditions of heating rate, temperature range and gaseous atmosphere or vacuum. In many cases, the use of a single thermoanalytical technique may not provide sufficient information to solve the problem on hand and hence, the use of other thermal techniques, either independently or simultaneously for complementary information becomes necessary.
Thermoanalytical analysis incorporates the following three closely related techniques:
i) Thermogravimetric analysis (TGA), which involves monitoring weight while varying temperature.
ii) Differential thermal analysis (DTA), which involves comparing the precise temperature difference between a sample and an inert reference material, while heating both.
iii) Differential scanning calorimetry (DSC), similar to DTA except that electrical energy is used to restore the cooler of the two materials to the same temperature as the other. This allows direct measurement of energy changes.
These techniques are useful for determining glass points, phase changes, water of crystallization and mixtures where the components have different melting or decomposition points. Among the thermal methods, the most widely used techniques
47
are TG, DTA and DSC, which find extensive use in all fields of inorganic and organic chemistry, metallurgy, mineralogy and many other areas. In the present work, thermal behaviour of the grown crystals has been investigated using DTA and TG-DTG techniques.
Thermogravimetry is a technique in which the mass of a substance is
measured as a function of temperature or time, while the substance is subjected to a
controlled temperature program. The curve obtained in a thermogravimetric analysis
is called a thermogram (TG) and its first derivative is called derivative thermogram
(DTG). The inflexion point in the program corresponds to the peak point in the
derivative thermogram. Modern commercial TG instruments consist of the following:
i) A sensitive analytical balance
ii) A temperature programmable furnace
iii) A pure gas system for providing suitable gas atmosphere and
iv) A microprocessor for instrument control, data acquisition and display
Even though different types of balance mechanism are available today,
those employing null-point-weighing mechanism are favoured as the sample remains
in the same zone of furnace irrespective of changes in mass. The furnace is normally
an electrical resistive heater and the temperature range for most of the furnace is from
ambient to 1000 - 2000 ºC. Thermogravimetry is widely used to determine the
thermal stability, decomposition temperature, temperature of desorption and drying,
oxidative stability, etc.
Differential scanning calorimetry (DSC) is similar to DTA except that
electrical energy is used to restore the cooler of the two materials to the same
temperature as the other. This allows direct measurement of energy changes. DSC
deals with recording of the energy necessary to establish the zero temperature
difference between a sample and a reference material against time or temperature
when both are heated or cooled at a predetermined rate. DSC monitors the heat effects
associated with phase transitions and chemical reactions as a function of temperature.
48
It is developed from DTA. In a power compensated DSC, the heat to be measured is
compensated with electrical energy, by increasing or decreasing adjustable heat
energy. The measuring system consists of two micro furnaces of the same type made
of a Platinum-Iridium alloy, each of which contains a platinum resistance
thermometer as a temperature sensor and a heating resistor made of platinum wire.
Both micro furnaces-separated from each other, are positioned in an aluminium block
of constant temperature. The maximum heating power of a micro furnace is about
14 W; the maximum heating rate is 500 K/min. The measuring range extends from
-175 ºC to 725 ºC.
During the programmed heating, some heating powder is supplied to both
micro furnaces through a control circuit in order to change with their mean
temperature in accordance with the preset heating rate. If there is ideal thermal
symmetry, the temperature of both the furnaces is always the same. Whenever
asymmetry occurs, for example as a result of a reaction in the sample, a temperature
difference results between the furnaces and the system adjusts the power supplied to
both the furnaces in such a way that the temperature difference is maintained at zero.
Two control circuits employed to obtain differential thermogram are:
i) Differential temperature control and
ii) Average temperature control
In the average temperature control circuit, the programmer generates an
electrical signal proportionate to the desired temperature of the signal and reference
holders, which is recorded on a chart. The programmed signal is then compared with
the average signal from the temperature sensors in two holders and the resultant signal
difference is amplified by the average temperature amplifier.
If the programmed temperature (Tp) is greater than the average temperature
(Tp + Tr) / 2, more power will be fed to the two furnaces. If it is lower, the power to
the furnaces will be reduced. In this manner, the average temperature of the furnaces
is made to follow closely the programmer command and the same power is fed to the
two micro furnaces.
49
In the differential temperature control circuit, signals representing the
temperatures of the sample and reference holders are compared and are fed to the
amplifier. The amplifier output adjusts the power input to the two furnaces in such a
way that their temperatures are kept identical. In the absence of any reaction taking
place in the sample, the differential power input to the sample and reference furnaces
is almost zero. When a reaction takes place in the sample, a differential power input is
fed to the furnaces. Depending upon whether the reaction is exothermic or
endothermic, the differential power input increases the power to one furnace while
decreasing it to other furnaces. This differential power usually in mW is plotted as a
function of temperature. In the present work, Netzch Thermal TGA/DSC equipment
was used with alumina as reference material.
2.8 MICROHARDNESS STUDY
Hardness is a physicochemical property that characterizes the state of the
material under test and gives information on some specific features of the material
such as the character of the chemical bonding. It is the resistance which the material
offers to indentation by a much harder body and may be termed as a measure of the
resistance against lattice destruction or permanent deformation or damage. As the
hardness properties are basically related to the crystal structure of the material and the
bond strength, microhardness studies have been applied to understand the plasticity of
the crystals. Hardness tests are commonly carried out to determine the mechanical
strength of materials and it correlates with other mechanical properties like elastic
constants and yield stress [102]. Hardness measurements can be defined as macro,
micro and nano according to the forces applied and displacement obtained [103].
The various methods using which hardness measurement can be carried
out are classified as follows:
i) Static indentation test
ii) Dynamic indentation test
iii) Scratch test
50
iv) Rebound test and
v) Abrasion test
The most popular and simplest test is the static indentation test, wherein an
indenter of specific geometry is pressed into the surface of a test specimen under a
known load. The indenter may be a ball or diamond cone or diamond pyramid. A
permanent impression is retained in the specimen after removal of the indenter. The
hardness is calculated from the area or the depth of indentation produced. The
variables are the type of indenter or load. The indenter is made up of a very hard
material to prevent its deformation by the test piece, so that it can cover materials over
a wide range of hardness. For this reason, either a hardness steel sphere or a diamond
pyramid or cone is employed. A pyramid indenter is preferred as geometrically
similar impressions are obtained at different loads. In this static indentation test, the
indenter is pressed perpendicularly to the surface of the sample by means of an
applied load. By measuring the cross sectional area or depth of the indentation and
knowing the applied load, empirical hardness number may be calculated. This method
is followed by Brinell, Meyer, Vickers, Knoop and Rockwell tests [104].
In the dynamic indentation test, a ball or a cone (or a number of small
spheres) is allowed to fall from a definite height and the hardness number is obtained
from the dimensions of the indentation and the energy of impact.
The scratch test can be classified into two types:
i) Comparison test, in which one material is said to be harder than
another, if the second material is scratched by the first
ii) A scratch test is made with a diamond or steel indenter on the surface
at a steady rate and under a definite load. The hardness number is
expressed in terms of the width or depth of the groove formed.
In the rebound test, an object of standard mass and dimension is bounced
from the test surface and the height of rebound is taken as the measure of hardness.
51
In the abrasion test, a specimen is loaded against a rotating disk and the
rate of wear is taken as a measure of hardness. Generally good sized which are
polished are used for the hardness studies.
2.8.1 Vickers Hardness and Anisotropy Studies
Vickers hardness method is the reliable and most common among the various methods of hardness measurement discussed above. In this method, microindentation is made on the surface of a specimen with the help of diamond pyramidal indenter. Smith et al., [105] have proposed that a pyramid be substituted for a ball in order to provide geometrical similitude under different values of load.
The Vickers pyramid indenter where opposite faces contain an angle ( = 136 ) is the
most widely accepted pyramid indenter. A pyramid indenter is suited for hardness tests due to the following reasons [106].
i) The contact pressure for a pyramid indenter is independent of indent size and
ii) Pyramid indenters are less affected by elastic release than other indenters
The base of the Vickers pyramid is a square and the depth of indentation corresponds to (1/7)th of the indentation diagonal. Hardness is generally defined as the ratio of the load applied to the surface area of the indentation. The Vickers hardness number Hv of Diamond Pyramid Number (DPN) is defined as
22 2
vp sin ( / )H
d (2.1)
where ‘ ’ is the apex angle of the indenter ( = 136 ). The Vickers hardness number
is thus calculated using the relation,
22
1.8544v
pH kg/mm d
(2.2)
52
where ‘p’ is the applied load in ‘kg’ and ‘d’ is the diagonal length of the indentation mark in ‘mm’. Hardness values are always measured from the observed size of the impression remaining after a loaded indenter has penetrated and has been removed from the surface.
Thus, the observed hardness behaviour is the summation of a number of
effects involved in the materials response to the indentation pressure during loading,
in the final measurement of the residual impression.
The importance of microhardness study lies in the possibilities of making
an indirect estimate of mechanical characteristics of materials such as yield strength
and toughness having a specific correlation with the hardness.
The hardness measurements may depend upon the orientation of the
indented crystals. To study the hardness anisotropy present in crystals, the crystals are
initially mounted on the stage of microscope properly and indented. The initial
position is 0 degree. The stage of the microscope was then rotated keeping the
indenter fixed and ‘Hv’ was measured at every 30 degrees interval till the original
position. No distortion in shape of indenter will be observed with crystal orientation.
When the variation of ‘Hv’ with angular displacement is periodic, then it brings the
anisotropic nature of crystals.
2.8.2 Knoops Hardness Studies
Knoop hardness can be treated as an alternative to the Vickers test,
particularly for very thin layers, Fredrick Knoop developed a low-load test with a
rhombohedral-shaped diamond indenter. The long diagonal is seven times
(7.114 actually) as long as the short diagonal. With this indenter shape, elastic
recovery can be held to a minimum. Knoop tests are mainly done at test forces from
10 g to 1000 g, so a high powered microscope is necessary to measure the indent size.
Because of this, Knoop tests have mainly been known as microhardness tests. The
magnifications required to measure Knoop indents dictate a highly polished test
surface. To achieve this surface, the samples are normally mounted and
metallurgically polished, therefore Knoop is almost always a destructive test.
53
The Knoop indented impressions are approximately rhombohedral in
shape. The average diagonal length (d) was considered for the calculation of the
Knoop hardness number (Hk) using the relation,
214.229
k PH
d (2.3)
where P is the applied load in kg, d in mm and Hk is in kg/mm2. In the present work,
MUTUTOYO M112 Japan, Instrument was used.
2.9 DIELECTRIC STUDIES
Dielectric measurement is one of the useful characterizations of electrical
response of solids. A study of the dielectric properties of solids gives information
about the electric field distribution within the solid. The frequency dependence of
these properties gives a great insight into the materials applications. The different
polarization mechanisms in solids can be understood from the study of dielectric
constant as a function of frequency and temperature.
Polarization ‘P’ of a dielectric is the numerical magnitude, which
describes the phenomenon of polarization of a dielectric in an external electric field.
In the absence of an external electric field, each element in the volume of a dielectric
has no electric moment. The action of an electric field brings the charges of the
molecules of the dielectric into a certain ordered arrangement in space. The study of
dielectric constant of a material gives an insight into the nature of bonding in the
material. The study of the electrical and other properties of dielectrics in relation to
their chemical composition and structure will lay the basis for obtaining new materials
with new properties. A lot of work had been carried out on dielectric measurement for
a variety of materials including ceramics and single crystals by many authors yielding
valuable information [107 - 109].
54
The relative dielectric constant ( r) is defined as,
0
r (2.4)
It is known that, CdA
(2.5)
Hence, 0r ACd
(2.6)
where, ‘A’ is the area of the sample and ‘d’ is the thickness of the sample.
The relative permittivity ( r) is usually known as permittivity. It is always greater
than unity. The capacitance ‘C0’ of a parallel plate condenser having a capacitance ‘C’
in air can be given by,
C = r C0 (2.7)
Thus, 0
rC
C (2.8)
‘ r’ can be found by the measurement of capacitance. The dielectric
constant of a substance is a property of the constituent ions. Major contributions to the
dielectric constant are from,
i) The extrinsic nature of the material,
ii) The electronic polarizability,
iii) The ionic polarizability and
iv) The deformation of the ions
55
In the present work, HIOKI 3532-50 LCR HITESTER meter was used for
the dielectric study which is shown in Figure 2.4.
The A.C. Conductivity is given by:
ac = 2 f tan 0 r (2.9)
where f is the frequency.
Using the above equation (2.9), the A.C. Conductivity at different
temperatures were calculated and the graph between ac and 1/T are plotted.
Figure 2.4 Photograph of HIOKI 3532-50 LCR HITESTER
2.10 ETCHING STUDIES
Etching studies have been carried out to understand the growth mechanism
and to assess the perfection of the grown crystals. Chemical etching is the oldest,
easiest and most widely used method to evaluate the quality of the crystal. Etching is
selective dissolution of the crystal which is used to reveal the crystal symmetry and
lattice defects.
56
The establishment of the relation between etch pits and the dislocation
array present in imperfect crystals by employing the etching technique has turned out
to be a powerful tool for the preparation and characterization of single crystals for
electronic and optical devices [110, 111].
The rate of reaction of a solution with a solid surface can depend distinctly
on the crystallographic orientation on the reacting surfaces. This rate dependence is
the basis of etching and brought about intentionally by specific chemical attack on the
crystal surface. When a surface is etched, well defined etch patterns of negative
volume are produced at the dislocation sites according to Thomas and Renshaw [112].
Etching may be considered as the reversal of crystal growth; therefore, a surface step,
for example, a screw dislocation, will dissolve more easily than a flat surface as
reported by Burton et al., [113]. The relative rates of removal of atoms along different
directions determine the geometry of etch pits. The shape of the etch pits may be
changed by varying the solvent. Moreover, factors such as temperature of etching,
stirring and adsorption of impurities or reaction products, which alter the absolute
value of these rates, also lead to the change in the geometry of etch pits [114]. Since
the geometrical variation of etch pits strongly depends on the arrangement of ions or
atoms which comprises the crystal faces, the shapes of the etch pits also differ for
various crystal faces. The shape of the etch pits may be changed by varying the
amount of the solvent. The etch patterns obtained also depend upon the nature of the
etchant. In the present work, the etching studies were carried out using Carl Zeiss
optical microscope (Model Imager AIM).
2.11 REFRACTIVE INDEX MEASUREMENTS
The refractive index of the crystals can be determined by Brewster’s angle
method using He-Ne laser of wavelength 632.8 nm. A polished flattened single crystal
of is mounted on a rotating mount at an angle varying from 0 to 90 degrees. The
angular readings on the rotary stage was observed, when the crystal is perfectly
perpendicular to the intracavity beam. The crystal is rotated until the laser oscillates
and the angle has been set for maximum power output. Brewster’s angle ( p) for the
57
crystal is measured. The refractive index is calculated using the equation, n = tan p,
where p is the polarizing angle.
2.12 NLO TEST – KURTZ POWDER SHG METHOD
Growth of large single crystal is a slow and difficult process. Hence, it is
highly desirable to have some technique of screening crystal structures to determine
whether they are noncentrosymmetric and it is also equally important to know
whether they are better than those currently known. Such a preliminary test should
enable us to carry out the activity without requiring oriented samples. Kurtz and Perry
proposed a powder SHG method for comprehensive analysis of the second order
nonlinearity. Employing this technique, Kurtz surveyed a very large number of
compounds.
The nonlinear optical property of the grown single crystal is tested by
passing the output of Nd:YAG Quanta ray laser through the crystalline powder
sample. The schematic of the experimental setup used for SHG studies is shown in
Figure 2.5. A Q-switched, mode locked Nd:YAG laser was used to generate about
6 mJ/pulse at the 1064 nm fundamental radiation. This laser can be operated in two
modes. In the single shot mode, the laser emits a single 8 ns pulse. In the multishot
mode, the laser produces a continuous train of 8 ns pulses at a repetition rate of 10 Hz.
In the present study, a single shot mode of 8 ns laser pulse with a spot radius of 1mm
was used. This experimental setup used a mirror and a 50/50 beam splitter (BS) to
generate a beam with pulse energies about 6 mJ. The input laser beam was passed
through an IR reflector and then directed on the microcrystalline powdered sample
packed in a capillary tube of diameter 0.154 mm.
The photodiode detector and oscilloscope assembly measure the light
emitted by the sample. Microcrystalline powder of urea or KDP is taken in a similar
capillary tube sealed at one end for comparison. The intensity of the second harmonic
output from the sample is compared with that of either KDP or urea. Thus, the figure
of merit of SHG of the sample is estimated.
58
Figure 2.5 Schematic experimental setup for SHG efficiency measurement
2.13 LASER DAMAGE THRESHOLD STUDY
An important related property of NLO crystals is the threshold for catastrophic laser induced damage. Laser induced damage in optical materials is a phenomenon involving interaction of high power laser radiation with matter and various physical, chemical, mechanical, optical and other aspects of materials that come into play. It is evident that the harmonic conversion efficiency is proportional to the power density of the fundamental beam. Hence, a convenient way to increase the efficiency is to focus the beam into the crystal. But, this often leads to breakdown of the materials, catastrophically damaging the crystal. It is then useful to prescribe the maximum permissible power for a particular crystal, defined as damage threshold.
The minimum power level which causes damage to at least 50% of sites irradiated is defined as the single shot (1-on-1) damage threshold (P1). However, power levels much lower than this can cause damage to the material under continuous exposure. Hence, at n pulses, damage may occur to the material at power levels Pn < P1. Following the definition of Nakatani et al., [115], the multiple shot (n-on-1) damage threshold ( ,) is the maximum power level below which the crystal does not suffer damage even after 1800 pulses. The origin of laser induced damage is highly complex. However, there are some processes which are basic to the phenomenon of laser damage.
59
Laser damage threshold is an important material parameter, the knowledge of which is essential for using the crystal as an NLO element in various applications involving large laser input power like frequency doubling, optical parametric processes, etc. In fact, laser induced damage in optical materials remains the limiting factor in the development of high power laser systems and optoelectronic devices. Although the optical damage has been studied in materials ever since the advent of high power lasers, much still remains to be learnt about the interaction of a high intensity laser beam with materials [116]. It is believed that the formation of an ionized region of dense plasma is the first and the most important step in a damage event [117]. In recent times, there have been several effects to model the laser damage processes by taking the pulse shape and duration in to account [118]. Laser damage in general manifests itself as a localized microscopic entity. Microscopic examination of the damage sites created in different materials often suggests the mechanism responsible for such a process. The damage in NLO materials like KDP and Deuterated Potassium Phosphate (DKDP), manifests itself as micro-cavity whose sizes are insensitive to the wavelength of the laser beam, suggesting a wavelength independent absorption mechanism. However, the damage mechanisms vary from material to material and hence damage studies become important for newly discovered materials. Laser damage measurement was carried out on the crystal at 1064 nm. The single shot surface damage thresholds have been determined. The laser damage threshold depends on pulse duration, focal spot geometry, sample quality, previous history of the sample, experimental technique employed etc. It may be noted that in the present case 13 ns pulse was used for the laser damage threshold measurement. Recent investigations into laser damage in various optical materials by nanosecond pulses have shown that the temperature reached at the damage site could be as high as 12000 K. One can expect the damage to be of thermal origin. However, one cannot rule out other mechanisms being operative simultaneously, as the damage mechanism is quite complex and depends on the nature of the material and various experimental parameters.
The experimental set-up used for the measurement of laser damage of the samples is shown in Figure 2.6. A Q-switched Nd:YAG laser (Continuum USA, Model: Surelite-III) of wavelength 1064 nm and pulse width 13 ns was used. The energy of the laser pulses was controlled by an attenuator (combination of / 2 plate
60
and glan polarizer) and delivered to the test sample located near the focus of a plano-convex lens of focal length 30 cm. The occurrence of single pulse damage was observed by monitoring the fall of transmitted intensity as detected by a fast PIN type Si photodiode and traced in a digital storage oscilloscope (Tektronix:TDS 3054B). A reference pyro-electric energy meter records the energy of the input laser pulse for which the crystal gets damaged. The laser damage threshold was calculated using the relation,
P = (E / A) (2.10)
where P is the Laser damage threshold in GW/cm2, E is the energy required to break the crystal, is the response time in ns and A represents the beam area.
Figure 2.6 Functional Block Diagram of the Setup used for damage threshold measurement
2.14 CONCLUSION
Single crystals grown by the various methods need to be characterized to that assess the suitability of the crystal for various applications including NLO device application. In order to understand the behaviour of any solid material, firstly, the structural characterization was carried out. Secondly, the optical characterization to check the transparency window and cut-off frequency of the crystal was performed. Since the entire thrust is with respect to NLO application, the SHG property was estimated. In addition, the mechanical, thermal, dielectric, refractive index and laser damage threshold of the grown crystal were also investigated.