calibration point selection

35
Calibration Point Selection Who makes the call? Presented by: Joe Petersen Metrology Engineer Abbott Laboratories

Upload: trinhtuyen

Post on 01-Jan-2017

231 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Calibration Point Selection

Calibration Point SelectionWho makes the call?

Presented by: Joe PetersenMetrology EngineerAbbott Laboratories

Page 2: Calibration Point Selection

Calibration Point Selection08/01/2012

2© 2012 Abbott

Common Questions

• Do calibration points have to cover the range of the instrument?

• How many calibration points are needed?

• Are 10, 50, 90 % of full scale appropriate?

• Should zero be used? Is it valid?

Page 3: Calibration Point Selection

Calibration Point Selection08/01/2012

3© 2012 Abbott

Introduction

• Calibration points effect measurement reliability– ISO 17025 addresses many quality issues;

• calibration point selection is not directly addressed• perhaps indirectly addressed by:

– 5.4 Methods, or– 5.7 Sampling

Page 4: Calibration Point Selection

Calibration Point Selection08/01/2012

4© 2012 Abbott

Agenda

• General principles

• Who makes the call?

• A form of sampling?

Page 5: Calibration Point Selection

Calibration Point Selection08/01/2012

5© 2012 Abbott

Learning Objectives

After this session, the participant should be able to:• List key considerations for selecting calibration points• Promote best practices for using a team approach to define

calibration points

Page 6: Calibration Point Selection

Calibration Point Selection08/01/2012

6© 2012 Abbott

Where are calibration point selection requirements defined?

QM

Policy

General Operating Procedure

Specific Calibration Procedure(Work Instruction)

Page 7: Calibration Point Selection

Calibration Point Selection08/01/2012

7© 2012 Abbott

Example of Policy / General Procedure

• “The selection of calibration points will vary with the normal use of the meter ....”– Usage

• “For example, a meter which monitors a fixed value can be spot calibrated at the critical value.”– Single Point Calibration

• “Meters which may indicate any value within their ranges will require a linearity test at full scale and at least three equally spaced divisions on one range, and a full-scale test on each remaining range.”– Full Calibration

• “Other meters, such as those used over a narrow range, may require calibration over a restricted portion of the scale.”– Limited Calibration

– CALIBRATION OR QUALIFICATION OF PANEL METERS, Metrology Bulletin, Oct '64--USN Metrology Engineering Center, Pomona, Calif. ; as published in National Conference of Standards Laboratories, NCSL NEWSLETTER 67-4, 1967

Page 8: Calibration Point Selection

Calibration Point Selection08/01/2012

8© 2012 Abbott

Example of Policy / General Procedure• “In general, for each TI characteristic being calibrated, the smallest number of test

points consistent with obtaining verification of the TI performance over its specified range should be used.”– Risk based approach

• “The number and choice of required test points will vary, depending on the characteristics of the TI.”– Consider design and technology

• “Guidance concerning choice of test points for selected generic classes of equipment is contained in Calibration Requirements Documents (CRDs).”– Specific requirements are in individual calibration procedures

• “In some cases, TI application/use requirements will dictate test point selection.”– Consider use

– REQUIREMENTS FOR PREPARATION OF INSTRUMENT CALIBRATION PROCEDURES• NAVAIR 17-35TR-4• NAVY METROLOGY ENGINEERING CENTER• 15 AUGUST 1990• 3.7.2.10 Calibration Test Points.

Page 9: Calibration Point Selection

Calibration Point Selection08/01/2012

9© 2012 Abbott

Key Considerations for Point Selection

• Measurement Technology

• External Guidance Documents

• Calibrated Range

– full or limited

• Assigned Accuracy

– (maximum permissible error)

• # of adjustment points

• User requirements

– critical usage points may be preferred calibration points

Page 10: Calibration Point Selection

Calibration Point Selection08/01/2012

10© 2012 Abbott

Are 3 points sufficient?Balance calibration guidelines

5 or moreEA-10/18 Guidelines on the calibration of non-automatic weighing instruments

10ASTM E898

typically 4Sartorius Tolerances for Testing Metrological Specifications

typically 4Mettler Balance Tolerances

5SANAS R15-03 (South African National Accreditation System)

5Code of Practice for Industrial Weighing Machines, United Kingdom Weighing Federation

10UKAS LAB145OIML R76-1

Minimum # of Accuracy/Linearity Test Points (excluding zero)

Publication

Page 11: Calibration Point Selection

Calibration Point Selection08/01/2012

11© 2012 Abbott

EURAMET Guidelines on the Calibration of Digital Multimeters

• cg-15

– Previously EA-10/15

Page 12: Calibration Point Selection

Calibration Point Selection08/01/2012

12© 2012 Abbott

EURAMET Guidelines on the Calibration of Digital Multimeters

– 3.4.1 “… consider the working principles of the instrument ...”– “[obtain guidance] from the manufacturer’s instructions,

although the points listed therein should not necessarily be considered exhaustive.”

‘trust but verify’

Page 13: Calibration Point Selection

Calibration Point Selection08/01/2012

13© 2012 Abbott

EURAMET Guidelines on the Calibration of Digital Multimeters

* Based on five parameters and five ranges per parameter.

• Conclusion: The extent of testing should be based on instrument accuracy.

100 to 1145 1/2 digits or more

(e.g. 200,000 counts)2

58 to 704 1/2 digits or less

(e.g. 20,000 counts)1

Recommended # of calibration Points*

Accuracy Class (Resolution)Table

Page 14: Calibration Point Selection

Calibration Point Selection08/01/2012

14© 2012 Abbott

# of calibration Points

• Factors to Consider

– Calibrated Span– Assigned Accuracy– Measurement Technology

• First principles• Empirical

– Cost– Risk

Relative Humidity Performance

-4

-2

0

2

4

0 20 40 60 80 100

% RH

Erro

r, %

RH

Relative Humidity Performance

-4

-2

0

2

4

0 20 40 60 80 100% RH

Erro

r, %

RH

Assume ± 2 %RH

Page 15: Calibration Point Selection

Calibration Point Selection08/01/2012

15© 2012 Abbott

Hind Sight is 20/20

• Given an excessive # of calibration points, it is easy to see which points are unnecessary

• You generally can’t tell you have enough points until you’ve tested more than enough

– Type testing– Retrospective review of historical results

Page 16: Calibration Point Selection

Calibration Point Selection08/01/2012

16© 2012 Abbott

Hind Sight is 20/20Stir Plate Calibrations

-4

-3

-2

-1

0

1

2

3

4

0 200 400 600 800 1000 1200

Speed, rpm

Aver

age

Erro

r, %

of r

eadi

ng .

Average Error, % of reading -- 1 standard deviation Error BarsAverage Error, % of reading -- 1 standard deviation Error Bars

Page 17: Calibration Point Selection

Calibration Point Selection08/01/2012

17© 2012 Abbott

Build on Experience

“Test results showed that measurements made at 50 test points were sufficient to allow accurate predictions of the instrument's performance at all 255 test points specified by the manufacturer.”

A. D. Koffman, H. L. Stott, Modeling and Test Point Selection for a Thermal Transfer Standard [AC Voltage], NCSL Workshop & Symposium, 1993

Page 18: Calibration Point Selection

Calibration Point Selection08/01/2012

18© 2012 Abbott

Points associated with secondary parameters

temperaturerelative humiditywavelengthoptical transmittancefrequencyAC voltage

.. depends upon..

Secondary ParameterPrimary Parameter

Page 19: Calibration Point Selection

Calibration Point Selection08/01/2012

19© 2012 Abbott

Full matrixSpectrophotometer Calibration Points – Commercially Available Filters– Visible Light Region

0.00.20.40.60.81.01.21.41.61.82.0

380 480 580 680 780wavelength, nm

optic

al d

ensi

ty o

r abs

orba

nce

Page 20: Calibration Point Selection

Calibration Point Selection08/01/2012

20© 2012 Abbott

Risk based approach

AC Voltage Calibration PointsEURAMET cg-15/v.01

0102030405060708090

100

10 100 1000 10000 100000 1000000frequency, Hz

mea

sure

men

t poi

nt,

% o

f ful

l sca

leOne Intermediate RangeOthers

Page 21: Calibration Point Selection

Calibration Point Selection08/01/2012

21© 2012 Abbott

Uncertainty increases as calibration points decrease

0.01 - 0.03Multipoint (5-point) calibration

0.02 - 0.03Calibration (2-point) by bracketing

0.3Calibration (1-point), ΔpH = |pH(X) – pH(S)| = 3 and assumed slope

Target uncertainty, U(pH)

Procedure

Source: R.P. Buck, et al., Measurement of pH. Definition, standards, and procedures (IUPAC Recommendations 2002)International Union of Pure and Applied Chemistry

Page 22: Calibration Point Selection

Calibration Point Selection08/01/2012

22© 2012 Abbott

# of calibration Points

• Summary of General Requirements

– Typically three or more to verify linearity– One is suitable for devices which are inherently linear (e.g. flow

totalizers, stopwatches, frequency counters)– Two are suitable for very narrow portions of a range– Five or more may be appropriate for some technologies

• Often recommended in guidance documents

Page 23: Calibration Point Selection

Calibration Point Selection08/01/2012

23© 2012 Abbott

Specific Principles

• Primarily related to technology

– Analog v. Digital– First Principles v. Empirical– Mechanical v. Electronic

Page 24: Calibration Point Selection

Calibration Point Selection08/01/2012

24© 2012 Abbott

Who makes the call?

• Instrument Manufacturer

• Calibration Supplier

• Metrological Confirmation Group

• Instrument Owner / User

Page 25: Calibration Point Selection

Calibration Point Selection08/01/2012

25© 2012 Abbott

Who makes the call?

– Instrument Manufacturer

– Calibration Supplier

– Instrument Owner / User

Going back to key considerations

• Measurement technology – (perhaps biased view)– Perhaps the most expertise

• Likely failure modes

• Available calibration standards• Perhaps an unbiased view of measurement

technology based on experience with multiple manufacturers

• Critical Points

Page 26: Calibration Point Selection

Calibration Point Selection08/01/2012

26© 2012 Abbott

Best practice for using a team approach

1. Obtain input from stakeholders

2. Calibration laboratory proposes calibration points

3. Owner reviews and approves

• Aligns with ISO 17025, 4.4 Review of requests, tenders and contracts

Page 27: Calibration Point Selection

Calibration Point Selection08/01/2012

27© 2012 Abbott

Is calibration point selection a form of sampling?

Page 28: Calibration Point Selection

Calibration Point Selection08/01/2012

28© 2012 Abbott

Is calibration point selection a form of sampling?

• ISO 17025 - 5.7 Sampling

– 5.7.1 The laboratory shall have a sampling plan and procedures for sampling when it carries out sampling of substances, materials or products for subsequent testing or calibration. …

– NOTE 1: Sampling is a defined procedure whereby a part of a substance, material or product is taken to provide for testing or calibration of a representative sample of the whole.

• How many organizations have a defined procedure for selection of calibration points?

Page 29: Calibration Point Selection

Calibration Point Selection08/01/2012

29© 2012 Abbott

Is calibration point selection a form of sampling?

5.7 Sampling – Not Applicable for calibrations(How many have seen this or a similar statement in a quality manual?)

• “Sampling does not apply. This laboratory only performs calibration, not testing.”

• “The laboratory does not sample items submitted for calibration.”

• “Sampling is not a part of our calibration process.

Page 30: Calibration Point Selection

Calibration Point Selection08/01/2012

30© 2012 Abbott

Is calibration point selection a form of sampling?

• Three interpretations of ISO 17025 for point selection

1. a form of sampling covered by 5.7, “Sampling”2. not sampling but is covered by 5.4, “Test and calibration

methods and method validation”3. not covered by 17025

Page 31: Calibration Point Selection

Calibration Point Selection08/01/2012

31© 2012 Abbott

Inadequate Calibration Points

XYZ Calibration Company ISO 17025 Accredited Calibration

Accreditations Universal

Certificate #: 01.02.03 Item: Humidity Meter Manufacturer: Meter Company Model: 87Y21

Temperature Standard Unit Under

Test Error Uncertainty

°C °C °C °C 23.04 23.1 +0.06 0.05

Relative Humidity

Standard Unit Under Test

Error Uncertainty

% RH % RH % RH % RH 50.3 52.8 +2.5 1.4

Approved by: J. Jones 03/30/2013

Device Range:

-10 to 70 °C

0 to 100 %RH

Page 32: Calibration Point Selection

Calibration Point Selection08/01/2012

32© 2012 Abbott

Between Point Uncertainty

• If user specifies range

– Reliability and uncertainty between points is lab’s responsibility

• If user specifies points

– Reliability and uncertainty between points is user’s responsibility

– Mark Kuster, Enough Is Enough: Selecting Points for Range Calibration, Measure Journal, Sep. 2011

Page 33: Calibration Point Selection

Calibration Point Selection08/01/2012

33© 2012 Abbott

Between Point Uncertainty

Sample certification statement:

• If user specifies range

– “Instrument meets manufacturer’s specifications throughout all functions and ranges.”

• If user specifies points

– “Instrument meets manufacturer’s specifications at all points tested.”

Page 34: Calibration Point Selection

Calibration Point Selection08/01/2012

34© 2012 Abbott

Appeal to Accreditation Bodies

• Whether point selection is an element of calibration methods or sampling:– Establish directives to assure adequate calibration points

• Eliminate the quality system gap• Level the playing field

– Prohibit single point certificates that imply a fully calibratedinstrument • (except as appropriate)• Some instruments, purchased new, come with an accredited calibration

Page 35: Calibration Point Selection

Calibration Point Selection08/01/2012

35© 2012 Abbott

Conclusion

• Risk of too few calibration points

– Untested regions– Undetected linearity errors

• Calibration point selection is technology dependent

• Each laboratory should establish a general policy / procedure

• Guidelines should also be defined in specific calibration procedures

• Team approach is desirable

• Accreditation should address point selection