Calibration of the CMS Electromagnetic Calorimeter

with first LHC data Maria Margherita Obertino

(Universita’ del Piemonte Orientale – INFN Torino)On behalf of the CMS Collaboration

IPRD10Siena, June 7-10 2010

1

Page 2: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

2

OUTLINE

The CMS Electromagnetic Calorimeter (ECAL)

Pre-calibration: crystal intercalibration before the LHC startup

In situ calibration: crystal intercalibration with collision data

Results with very first collisions

Calibration data streams

2010 plans

M.M.Obertino – IPRD10

Page 3: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

ECAL

SupermoduleES

EE

EE

Hermetic homogeneous calorimeter made of PbWO4 Crystals

Barrel (EB) [|h|<1.48 ] 36 Supermodules (SM) with 1700 crystals each Photodetectors: APD

Endcaps (EE) [1.48<|h|<3.0 ] 2 Dees each side, 14648 crystals in total Photodetectors: VPT

Preshower (ES) [1.65<|h|<2.6 ] Sampling calorimeter (lead, silicon strips)

3

M.M.Obertino – IPRD10

Page 4: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

4

ECAL ENERGY RESOLUTIONExcellent energy (and position) resolution for photons and electrons crucial for studying interesting physics channel (Hgg, HZZ4e, …)

σ E

E=a

E⊕b

E⊕c

Constant term depends on:• channel intercalibration accuracy • non uniformity of the longitudinal light collection• rear leakageTarget value: 0.5 % set by H channel

M.M.Obertino – IPRD10

Hγγ width

High precision in the channel to channel intercalibration crucial.

Zee width

Page 5: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

5

PRE-CALIBRATION

10 EB SMs

460/14648 EE crystals

Test Beam at CERN

90-120 GeV electrons

(2004-2007)

Beam

Extensive10 years long pre-calibration program carried out before ECAL installation in CMS.

[ Intercalibration of the barrel electromagnetic calorimeter of the CMS experiment at start-up - 2008 JINST 3 P10007 ]

iIntercalibration precision ~0.3% Nu

mb

er

of

cry

sta

ls

2(cA − cB ) /(cA + cB )

dominated by statistical uncertainty, determined comparing two independent data samples (A and B) collected in the same conditions.

M.M.Obertino – IPRD10

Intercalibration reproducibility

From repeated measurements:

Page 6: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

6

PRE-CALIBRATION

EB: LY Measurements Intercalibration precision ~4.5-6.0%

EE: LY + VPT gain Measurements Intercalibration precision ~8%

Light yield (LY) measurementswith 60Co (1MeV g) during crystal qualification (2000-2006)

Page 7: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

7

PRE-CALIBRATION

EB: LY Measurements Intercalibration precision ~4.5-6.0%

EE: LY + VPT gain Measurements Intercalibration precision ~8%

36 EB SMs operated on a cosmic ray stand for a period of ~1 week/SM

Deposited energy per crystal ~ 250 MeV

Mean intercalibration precision ~1.5%

Channel intercalibration with cosmic muons (2006-2007)

Cosmic muons

Light yield (LY) measurementswith 60Co (1MeV g) during crystal qualification (2000-2006)

EB SM h=0 h=1.48

Page 8: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

PRE-CALIBRATION

Intercalibration precision ~6%

8

CLOSED COLLIMATOR

All Endcap crystals

Beam splash: in September 2008 and November 2009 beam was circulated in LHC, stopped in collimators 150 m away from CMS

BEAM

Combination of LY and beam splash gives ~5% intercalibration precision over the entire EE

Page 9: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

9

EB

LY, cosmic, test beam.

Combination strategy: Select best calibration available Combine when comparable

precision from two sources

EE

Test beam (460 crytals) Combination of LY and beam splash

M.M.Obertino – IPRD10

PRE-CALIBRATION CONSTANTS

Intercalibration precision achieved with these constants: 0.3-2.2%

Intercalibration precision achieved with these constants: 5%

Page 10: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

10

ENERGY SCALE: ADC to GeV

Distribution of 5x5 amplitude sum (S25) with and without intercalibration for EE crystals exposed to 90 GeV electron beam .

(*) presented by F.De Guio

Confirmed later in dE/dx analysis with cosmics(*) and in collisions with p0 and h mass peak.

Waiting for higher mass resonances to check the scale in-situ at high energy.

ECAL energy scale fixed by Test Beam data in Barrel and Endcap separately

~63 MeV/ADC in EE

~39 MeV/ADC for EB

INTERCALIBRATION

90 GeV

Page 11: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

11

IN-SITU CALIBRATIONSeveral methods to calibrate in-situ:

f-symmetry method: intercalibration for crystals at the same pseudorapidity. Based on invariance around the beam axis of the energy flow in minimum bias events.

Calibration performed comparing the total energy deposited in each crystal with the mean of the distribution of the total energies for all crystals at the same pseudorapidity

Page 12: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

12

IN-SITU CALIBRATIONSeveral methods to calibrate in-situ:

f-symmetry method: intercalibration for crystals at the same pseudorapidity. Based on invariance around the beam axis of the energy flow in minimum bias events.

Calibration performed comparing the total energy deposited in each crystal with the mean of the distribution of the total energies for all crystals at the same pseudorapidity

p0 and h method: single crystal intercalibration Constants derived using unconverted g’s from p0 and h decay reconstructed in 3x3 matrices

The reconstructed mass is used to determine the photon energy

Page 13: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

13

IN-SITU CALIBRATION Several methods to calibrate in-situ:

f-symmetry method: intercalibration for crystals at the same pseudorapidity. Based on invariance around the beam axis of the energy flow in minimum bias events.

Calibration performed comparing the total energy deposited in each crystal with the mean of the distribution of the total energies for all crystals at the same pseudorapidity

p0 and h method: single crystal intercalibration Constants derived using unconverted g’s from p0 and h decay reconstructed in 3x3 matrices

The reconstructed mass is used to determine the photon energy

High energy electrons from Wen and Zee decays (J/y under study) Higher integrated luminosity required. Helpful at the startup only for energy scale.

Page 14: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

SUPERMODULE RELATIVE SCALEFirst calibration with collisions of the barrel. First results in 2009 with ~0.01 nb-1 collected at

Scale derived in 2010 with ~0.4 nb-1 collected at Systematic error: 0.5% (see next slide) Statistical error negligible w.r.t. systematic one

Tested with beam

s = 900GeV

s = 7TeV

CMS Preliminary CMS Preliminary

s = 7TeV

• f-symmetry data

• p0 data

14

M.M.Obertino – IPRD10

Page 15: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

15

SUPERMODULE RELATIVE SCALE

Distribution of the relative scale of 10 SMs calibrated with test beam electrons.

Distribution of the relative scale of 26 SMs calibrated only with cosmic rays.

The RMS of 0.5%±0.1% estimates the current precision of the combined in-situ SM relative scale calibration.

The RMS is 1.2%±0.2% is consistent with the expected precision of cosmic ray scale.

In-situ SM relative scale defined as the weighted average of p0 and f-symmetry results.

CMS Preliminary CMS Preliminary

Page 16: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

16

RECONSTRUCTED p0 gg PEAK L = 0.43 nb-1

PT (g) > 0.4 GeV, PT(ggpair) > 1.0 GeV (pure ECAL selection)

Good agreement between data and MC. Number of p0 from the fit: 1.46 106

DATA MC

The fitted mass agrees with the expectation to within 1% (first in situ validation of absolute energy scale)

M.M.Obertino – IPRD10

Page 17: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

17

RECONSTRUCTED h gg PEAK L = 0.43 nb-1

PT (g) > 0.5 GeV, PT(ggpair) > 2.5 GeV (pure ECAL selection)

Good agreement between data and MC Number of h from the fit: 2.55 104

DATAMC

Page 18: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

18

CALIBRATION DATA STREAMIn CMS dedicated paths of data acquisition (streams) are implemented for calibration.

In p0 and f-symmetry streams, only few tens of “useful”crystal hits are stored for each accepted event (output limited to 1kHz per stream).

Calibration stream commissioned with collision data.

p0 peak from dedicated calibration stream

(better S/B because of higher transverse energy cuts in calibration stream)

Page 19: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

19

PLANS FOR 2010

improve crystal calibration down to 1% in the whole EB with few pb-1.

reach the goal of 0.5% precision in EB with ~10pb-1, expected by the end of the year.

Improve Endcap crystal calibration from 5% to 1-2%.

Set the absolute scale to few permille.

Fluctuation due to material in front of ECAL

In the Barrel

Page 20: Calibration of the CMS Electromagnetic Calorimeter  with first LHC  data

M.M.Obertino – IPRD10

20

SUMMARY

Z width already almost insensitive to residual mis-calibration.

First calibration result with collision data: SM relative scale in EB. p0and f-symmetry in agreement.

Calibration streams commissioned.

Collected statistics (~17 nb-1 at 26th May) good for regional intercalibration (5x5 matrix).

2010 goals: 0.5% intercalibration precision in EB 1-2% intercalibration precision in EE

ECAL crystal intercalibration in very good shape.

Method Calibrated crystal

Precision

Lab. measurement All ~ 5% in EB ~ 8% in EE

Cosmic All EB ~1.5% in average

Test Beam 10 SMs in EB~400 xtal in EE

~0.3%

Beam splash All EE ~6%

0.3%-2% in EB

5% in EE