  • STLCC-CPLS:Morrison 6/26/2013 Page 1

    BioAnalyzer; Agilent 2100 (Detect Instrument Picture to Jump to that section of SOP)

    Prepared by: Bob Morrison

    STLCC CPLC, Instrumentation Specialist

    Initial Feb 2008, Latest Rev June 2013

    IKA MS2 Minishaker

    Vortexer for use with

    Agilent 2100Bioanlyzer

  • STLCC-CPLS:Morrison 6/26/2013 Page 2

    Agilent BioAnalyzer: Connections and Startup

    The Agilent 2100 Bioanalyzer is very useful tool in analyzing RNA

    quality for real-time PCR and microarray experiments. It combines

    a miniature of gel electrophoresis and scanning spectrophotometry

    to allow analysis of quality of RNA, DNA and protein samples using

    very small amounts of material, i.e., nanograms or even picograms

    of total RNA

    2100 Bioanalyzer Specifications

    Weight 10 kg (22 lbs)

    Dimensions (w × d × h) 162 × 412 × 290 mm (6.4 × 16.2 × 11.4 in)

    Line voltage 100-240 VAC

    Line frequency 50-60 Hz

    Power consumption 60 VA

    Ambient operating temperature5-40 °C (41-104 °F)

    Software: V B.02.08.SI48 SR1

    1. Connect device(s) to host CPU via

    RS232 directly or via RS232/USB

    adapter

    2. On/Off toggle (rear panel, extreme lower

    center by power plug)

    Electrode Cartridge Type Stamped:

    (1) Electrophoresis

    (2) Flow cytometry

    3. Wait for light to reach steady green

    Light: Green (ready), Blinking Green

    (busy), Orange (diagnostics running),

    Red (not ready)

    4. Logon: Adminstrator/3000hanover

    5. Start Expert 2100 Software after

    green lights on all instruments

    Serial Number of Instrument

  • Bioanalyzer: Agilent 2100, Bayonet Cartridge

    STLCC-CPLS:Morrison 6/26/2013 Page 3

    Link to Agilent Bayonet Removal

    guide (pdf)

    Users.stlcc.edu/Departments/fvbio/Bioanalyzer_Agilent_Changing_16-Pin_Bayonet_Cartridge.pdf

  • STLCC-CPLS:Morrison 6/26/2013 Page 4

    BioAnalyzer: Agilent 2100 Expert Software Startup Screen

    1. Wait for Expert2100 software to

    detect each attached instrument

    located at respective Com ports.

    (will be Com1 if RS232 direct, or

    Com >4 if RS232/USB is used)

    2. Note Serial Numbers of instruments

    detected and checklist on each.

    DE 72901434

    DE 54700853

  • STLCC-CPLS:Morrison 6/26/2013 Page 5

    BioAnalyzer : Agilent 2100 Manuals, Reference Materials

    Link to Agilent Protein 80 Kit Quick Start Guide (pdf)

    Link to Agilent 2100 Expert Users Guide (pdf)

    Link to U of Virginia Lab: Scanning Gels using GS-800 (pdf)

    Link to Agilent Protein 230 Kit Quick Start Guide (pdf)

    Link to Agilent Protein 80 Kit Ebook (pdf)

    Link to Agilent 2100 Troubleshooting Guide (pdf)

    Link to Agilent RNA 6000 Nano Kit (w BB notes) Quick Start Guide (pdf)

    http://www.chem.agilent.com/Library/usermanuals/Public/G2938-90063_Protein80Assay.pdfUsers.stlcc.edu/Departments/fvbio/BioAnalyzer_Agilent2100_Expert_User_Manual.pdfhttp://www.itc.virginia.edu/achs/documents/ACHS-103.pdfUsers.stlcc.edu/Departments/fvbio/Bioanalyzer_Agilent2100_Protein230Kit_Analysis.pdfUsers.stlcc.edu/fvbio/Bioanalyzer_Protein80_ebook.pdfUsers.stlcc.edu/Departments/fvbio/BioAnalyzer_Agilent2100_Troubleshooting_Guide.pdfUsers.stlcc.edu/Departments/fvbio/BioAnalyzer_Agilent_RNA6000_Nanokit.pdf

  • BioAnalyzer: Agilent, Recent (June 2011) Error Message

    STLCC-CPLS:Morrison 6/26/2013 Page 6

    With the dual Bioanalyzers running on

    USB/RS232 adapters, this error message

    has been encountered. See the next slide

    for a method to correct the problem.

  • STLCC-CPLS:Morrison 6/26/2013 Page 7

    BioAnalyzer: Agilent, Error Message Potential Fix

    Some errors have been corrected by

    selecting this button and typing in the

    Device ID name in the pop-up window

  • STLCC-CPLS:Morrison 6/26/2013 Page 8

    BioAnalyzer: Agilent2100, Insert Sample Chip

  • STLCC-CPLS:Morrison 6/26/2013 Page 9

    BioAnalyzer: Agilent Expert-Assay-Electrophoresis-Protein Options

  • STLCC-CPLS:Morrison 6/26/2013 Page 10

    BioAnalyzer: Agilent-Expert2100-Proteinchip-Run Checklist completed

    Select Start Run after the

    Checklist is complete

    Select Start Run after the

    Checklist is complete

  • STLCC-CPLS:Morrison 6/26/2013 Page 11

    BioAnalyzer: Agilent2100- Observations during the Chip Run

  • STLCC-CPLS:Morrison 6/26/2013 Page 12

    BioAnalyzer: Agilent2100, Data context, Chip Summary

  • STLCC-CPLS:Morrison 6/26/2013

    BioAnalyzer: Agilent2100 During the Run -Measurement

    Page 13

  • STLCC-CPLS:Morrison 6/26/2013

    BioAnalyzer: Agilent2100, Electropherogram tab

    Page 14

  • STLCC-CPLS:Morrison 6/26/2013

    BioAnalyzer: AgilentExpert 2100,Example-Protein 80 Assay Results

    Page 15

  • STLCC-CPLS:Morrison 6/26/2013 Page 16

    BioAnalyzer: Agilent2100, Options after the Run

  • STLCC-CPLS:Morrison 6/26/2013 Page 17

    BioAnalyzer: Agilent Expert-INSTRUCTOR Instrument-Diagnostics-Selectall-Start

    Diagnostics Test requires two special

    chips: Electrode Chip and Optics

    Chip. Locate these before beginning

    diagnostics. Also have a

    blank/empty chip handy.

    Follow software instructions during

    test.

  • STLCC-CPLS:Morrison 6/26/2013 Page 18

    BioAnalyzer: Agilent2100 INSTRUCTOR

    Expert-Instrument-Diagnostics-Follow Instructions

    Diagnostics Test requires two special

    chips: Electrode Chip and Optics

    Chip. Locate these before beginning

    diagnostics. Also have a

    blank/empty chip handy.

    Follow software instructions during

    test.

  • STLCC-CPLS:Morrison 6/26/2013 Page 19

    BioAnalyzer: Agilent2100 Expert INSTRUCTOR

    -Instrument-Diagnostics- Report

  • STLCC-CPLS:Morrison 6/26/2013 Page 20

    BioAnalyzer: Agilent2100 Expert-Instrument-Diagnostics-LogBook-Print-pdf

  • STLCC-CPLS:Morrison 6/26/2013 Page 21

    BioAnalyzer: Agilent2100 Expert-Assay-Electrophoresis-DNA Options

  • STLCC-CPLS:Morrison 6/26/2013 Page 22

    BioAnalyzer: Agilent2100 Expert-Assay-Flow Cytometry

  • BioAnalyzer: Vortexer; IKA Works-MS2 Minishaker,

    (use with Agilent 2100 Bioanalyzer)

    STLCC_CPLS;Morrison 6/26/2013 Page 23

    Universal, small-sized shaker. With an orbit

    of 4.5 mm and a speed range of 0 - 2,500

    rpm, the MS 2 is suitable for all small

    vessels. - Electronic speed control -

    On/Off

    Set rpm using

    analog dial

    Fitting (not shown)

    for Agilent 2100

    Bioanalyzer Plates

  • STLCC-CPLS:Morrison 6/26/2013 Page 24

    BioAnalyzer : Other Notes and Options

    • Future home of other detailed procedures or printout options ….