best practice nurse
TRANSCRIPT
-
7/29/2019 best practice nurse
1/50
Best Practices forElectrosurgery
-
7/29/2019 best practice nurse
2/50
3
Potential Adverse Outcomes
-
7/29/2019 best practice nurse
3/50
Potential Electrosurgery Injuries
Laparoscopic Injuries
Alternate Site Injuries Return Electrode Site Injuries
Fire Injuries
-
7/29/2019 best practice nurse
4/50
5
Basic Principles of Electrosurgery
-
7/29/2019 best practice nurse
5/50
Basic Principles of Electricity
Electricity always . . .
Seeks ground (its source)
Seeks the path of least
resistance
-
7/29/2019 best practice nurse
6/50
Properties of Electricity
Current Flow of electrons during a
period of time, measured in
amperes
Circuit Pathway for the
uninterrupted flow of
electrons (must be
complete/ closed to flow)
Impedance Obstacle to the flow of
current measured in ohms
Voltage Force pushing current
through the resistance,
measured in volts
-
7/29/2019 best practice nurse
7/50
Frequency Spectrum
-
7/29/2019 best practice nurse
8/50
The more concentrated the current,
the greater the potential for a burn.
Current Concentration/Dispersal
-
7/29/2019 best practice nurse
9/50
Monopolar
Active electrode at surgical site Return electrode at another site Current flows through the body
between the electrodes High voltage
Coag 3000 9000
Cut 1350 - 4000
-
7/29/2019 best practice nurse
10/50
Bipolar
Active and return electrodes in theinstrument
Current flow confined to tissuebetween electrodes
Low Voltage (320 - 1200 volts)
-
7/29/2019 best practice nurse
11/50
Electrosurgical Technologies
-
7/29/2019 best practice nurse
12/50
Patient Burns
Pad site burn
-
7/29/2019 best practice nurse
13/50
Therapeutic current
Interrogation current
Contact Quality Monitoring
-
7/29/2019 best practice nurse
14/50
Contact Quality Monitoring
The most frequently reportedpatient injury has been a skin injury
at the patient return electrode site
The risk of this type of injury has
been minimized through advancesin patient return electrode design
Nondrying conductive material
Use of return electrode contact
quality monitoring
AORN Recommended Practices, 2009
-
7/29/2019 best practice nurse
15/50
20
Best Practices to Achieve Favorable PatientOutcomes Related to Electrosurgery
-
7/29/2019 best practice nurse
16/50
Follow Manufacturers Written Instructions
The ESU and accessories shouldbe used according to the
manufacturers written instructions.
Each type of ESU has specificmanufacturers written operating
instructions to be followed for safeoperation of the unit
Accessories should be used,handled, cleaned, and processed
according to manufacturers
instructions
AORN Recommended Practices, 2009
-
7/29/2019 best practice nurse
17/50
General Safety Precautions
Test alarm systems Set activation tone to audible level
-
7/29/2019 best practice nurse
18/50
General Safety Precautions
Confirm Power SettingsPlug Accessories into Correct Receptacles
-
7/29/2019 best practice nurse
19/50
Selecting the Patient Return Electrode
Patient return electrodes should becompatible with the ESU
Do not attempt to use return
electrodes that disable the contact
quality monitoring system
Verify that the: Electrode is intact
Conductive gel, if present, is moist
Manufacturers expiration date has
not been reach
Do not reuse patient returnelectrode
-
7/29/2019 best practice nurse
20/50
Selecting the Patient Return Electrode Site
Follow manufacturers writteninstructions
The conductive and adhesivesurfaces of the electrode should be
placed on:
Clean, dry skin
Over a large, well-perfused
muscle mass of the surgical
side
Close to the surgical site
Convex area
-
7/29/2019 best practice nurse
21/50
Avoid Patient Return Electrode Sites High
in Resistance
Prostheses
Hair
Bony prominences
Scar tissue
-
7/29/2019 best practice nurse
22/50
Preparing the Patient Return Electrode Site
Follow manufacturers
instructions for hairremoval, cleaning, and
drying site
Protect return
electrode from fluidinvasion
Do not use flammable
agents for PRE sitepreparation
-
7/29/2019 best practice nurse
23/50
Patient Return Electrode Guidelines
Do not overlap edges
of return electrodeDo not cut or modify
return electrode
-
7/29/2019 best practice nurse
24/50
Patient Return Electrode Guidelines
Do not warm return electrode Apply return electrode after
positioning
Place return electrode proximal totourniquet, preferably on operative
side
-
7/29/2019 best practice nurse
25/50
Follow Manufacturer Instructions When
Using Different Sizes of Patient Electrodes
Select patient electrode accordingto patient weight
Infant patient return electrode
Do not exceed 120 watts of power Use short activations
Neonatal patient return electrode
Do not exceed 300 milliamps of
power
Do not activate active electrode
longer than 30 seconds
For neonates < .45 kg, use of
bipolar electrosurgery
-
7/29/2019 best practice nurse
26/50
Compression Stockings and Patient Return
Electrodes
Compression stocking maycause
Increased temperature at the
pad site
Increased impedance at padsite
Current concentration with
resultant heat build up
Do not apply the patient returnelectrode to areas wherepressure is applied to the
patient
ECRI
-
7/29/2019 best practice nurse
27/50
Warming Devices and the Patient Return
Electrode
Always follow manufacturersinstructions when using a warming
device
Avoid close contact with warmingblanket
A towel, blanket, or folded sheetmay be placed between the patient
return electrode and the warming
blanket to distance the heating
source from the patient return
electrode
Covidien Clinical Information Hotline
-
7/29/2019 best practice nurse
28/50
Reducing the Potential for Alternate Site
Burns
Avoid skin-to-skin contact, such asfingers touching leg or knee
touching knee when positioning the
patient
Use non-conductive dry padding toprevent skin-to-skin contact
Position the return electrode toprovide a direct current route
-
7/29/2019 best practice nurse
29/50
When to Use Two Patient Return
Electrodes
Placement of additional PREs
increases the dispersion of
electrosurgical current and heat at
the pad-to-patient interface
Obesity
Emaciation
Fluid environment
Long duty cycles
AORN Recommended Practices, 2009
-
7/29/2019 best practice nurse
30/50
Removing the Patient Return Electrode
Remove return electrode slowly,supporting the tissue underneath
Observe skin under pad Chart appearance of site
-
7/29/2019 best practice nurse
31/50
Active Electrodes
When not in use, always placeactive electrodes in a non-
conductive holster
Electrodes that do not fit in theholster should be placed in a
designated location with tips away
from flammable material
Active electrode tips should besecurely seated into the hand
piece
AORN Recommended Practices, 2009
-
7/29/2019 best practice nurse
32/50
Active Electrodes
Do not use suctioncoagulator as a retractor
Avoid tissue contact withshaft during activation
Ensure that the outside ofthe suction coagulatorremains free of fluids and
mucus which may conduct
electrical current
Unintended activation mayresult if the suction
coagulator handswitch isimmersed in saline solution
or other conductive fluids
Covidien Instructions for Use, Suction Coagulator
-
7/29/2019 best practice nurse
33/50
Active Electrodes
Build-up of eschar increasesresistance at the active electrode
and may require the use of higher
power setting
Eschar debris may ignite
Use a coated electrode to easilyremove eschar buildup onelectrode tip
-
7/29/2019 best practice nurse
34/50
Active Electrodes
Do not use red rubber catheters or
other materials as a sheath
Red rubber or plastic materials
may ignite
Use manufacturer approvedinsulated tips
ERCI, 2006
-
7/29/2019 best practice nurse
35/50
Radio Frequency Current Leakage
Active electrode cords should notbe wrapped around metal
instruments
Active electrode and otherelectrical device cords should not
be bundled together
Phippen, Ulmer, & Wells, 2009
-
7/29/2019 best practice nurse
36/50
Avoiding Hemostat Burns
Do not buzz the hemostat with a needle
electrode
Do not lean on the patient, the table, or the
retractors while buzzing the hemostat
Activate cut rather than coag
Firmly grasp as much of the hemostat as
possible before activating
Buzz the hemostat below hand level
Activate the generator after the instrument
makes contact with the hemostat
When using a coated or nonstick blade
electrode, place the edge of the electrode
against the hemostat Note: Surgical gloves do not insulate
against RF current
Ulmer, 2008
-
7/29/2019 best practice nurse
37/50
Pacemakers
Consult pacemaker manufacturer Use
electrosurgery with care
Use bipolar
Use low power setting
Avoid current flow through heart and
pacemaker
Keep cords away from pacemaker
and leads
Postoperatively have the device
checked by a trained individual
Phippen, Ulmer, & Wells, 2009
-
7/29/2019 best practice nurse
38/50
Internal Cardioverter-Defibrillators
Obtain preoperative cardiology consult toevaluate correct functioning and determine
risks
Have defibrillator immediately available
Deactivation of ICD by trained personnel
before ESU use
Continuous ECG and peripheral pulse
monitoring
Use bipolar
If monopolar used ensure distance between
the active and patient return electrode is as
short as possible
Avoid current flow through heart and ICD
Postoperatively have the device checked by
a trained individual
AORN Recommended Practices, 2009
-
7/29/2019 best practice nurse
39/50
Hearing Aids
Best practice is to remove hearingaids prior to surgery
RF leakage may causeinterference to the hearing aid
reception
RF current may damage two piecewireless hearing aids, thus theyshould not be worn when
electrosurgery is used
AORN Recommended Practices, 2009
-
7/29/2019 best practice nurse
40/50
Cochlear Implant
Do not use monopolar electrosurgery on
the head or neck of a cochlear implant
patient
Use bipolar electrosurgery on the head
and neck of patients with implants that do
not have extracochlear reference
electrodes
The bipolar electrode must not contact the
implant
If the implant has extracochlear reference
electrodes, bipolar may be used if the
electrode is kept more than 10 cm (4
inches) from the extracochlear electrodes
Postoperatively have the device checked
by a trained individual
AORN Recommended Practices, 2009; Hotline News, Vol 6, Number 4
-
7/29/2019 best practice nurse
41/50
Dental Appliances
Avoid direct contact with any metaldental appliances
Cover braces with teeth guards(silicone or rubber) or use bipolar
energy when possible
Remove palate spreaders prior tosurgery if electrosurgery will beused in immediate area of the
device
Covidien Clinical Information Hotline
-
7/29/2019 best practice nurse
42/50
Jewelry
Jewelry should be removed if it iswithin the activation range of the
active electrode.
When using a reusable, capacitive-coupled return electrode, all of the
patients metal jewelry should be
removed.
AORN Recommended Practices, 2009
-
7/29/2019 best practice nurse
43/50
Tattoos
Avoid placing the patient returnelectrode over tattoos
Inks (red in particular) containmetals which could serve as a heat
or electrical conductor
Valleylab Clinical Hotline
AORN Recommended Practices, 2009
-
7/29/2019 best practice nurse
44/50
Electrosurgery During Pregnancy
No evidence to contraindicate the use ofelectrosurgery during pregnancy
Amniotic fluid protects fetus from anyconcentration of electrical current owing
to the dispersion effect
Just as high frequency of electrosurgerycurrent above 100 kHz prevents
neuromuscular stimulation in adults,
children, and infants, the same is true
for the fetus Leonardo da Vincic. 1510 - 1512
Te Linde's Operative Gynecology Text, 8th Edition
-
7/29/2019 best practice nurse
45/50
Electrosurgery During Cesarean Sections
If flammable prepping solutionused follow manufacturers
instructions
Apply the patient return electrodeto a site that will not be exposed to
pooling of amniotic fluid Activate electrosurgery pencil only
when in contact with intended
tissue
Place electrosurgery pencil in non-conductive safety holster when not
in use
-
7/29/2019 best practice nurse
46/50
Pediatric Applications and Small Anatomic
Structures
Pediatric applications and/or proceduresperformed on small anatomic structures
may require reduced power settings
Potential for unintended thermal damageto tissue with high current flow and
longer activation
For procedures where the current couldflow though delicate parts of the body,
use of bipolar technique may be
desirable in order to avoid unwanted
coagulation
Covidien Electrosurgery Users Guide
Best Practices to Achieve Favorable Patient
-
7/29/2019 best practice nurse
47/50
Best Practices to Achieve Favorable PatientOutcomes Related to Electrosurgery
1. Manufacturers Instructions
2. Safety Precautions (2)
3. Selecting the Patient ReturnElectrode
4. Selecting the Patient ReturnElectrode Site (2)
5. Preparing the Patient ReturnElectrode Site
6. Patient Return Electrode Guidelines
7. Different Sizes of Patient Electrodes
8. Compression Stockings and PatientReturn Electrodes
9. Warming Devices and the PatientReturn Electrode
10.Alternate Site Burns
11. Use Two Patient ReturnElectrodes
12.Active Electrodes (4)
13. RF Current Leakage
14.Avoiding Hemostat Burns
15. Pacemakers
16. Internal Cardioverter-Defibrillators
17. Hearing Aids
18. Cochlear Implant
19. Dental Appliances & Jewellery (2)
20. Tattoos
21. Electrosurgery During Pregnancy /Cesarean Sections (2)
22. Pediatric Applications and SmallAnatomic Structures
-
7/29/2019 best practice nurse
48/50
Evaluate Patient After Surgery
Electrosurgery InjuryPressure InjuryChemical Injury
-
7/29/2019 best practice nurse
49/50
Questions ???
Thank you !
-
7/29/2019 best practice nurse
50/50
Program Summary
Potential adverse outcomes related to electrosurgery
Basic principles of electrosurgery
Best practices to achieve favorable patient outcomes related toelectrosurgery
Prior to use, clinicians should review the Instructions for Use product
inserts and/or Users Guide provided by the manufacturer to ensure
safe and appropriate utilization of electrosurgical devices.