best practice nurse

Upload: iwan-panji-widayana

Post on 04-Apr-2018

221 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/29/2019 best practice nurse

    1/50

    Best Practices forElectrosurgery

  • 7/29/2019 best practice nurse

    2/50

    3

    Potential Adverse Outcomes

  • 7/29/2019 best practice nurse

    3/50

    Potential Electrosurgery Injuries

    Laparoscopic Injuries

    Alternate Site Injuries Return Electrode Site Injuries

    Fire Injuries

  • 7/29/2019 best practice nurse

    4/50

    5

    Basic Principles of Electrosurgery

  • 7/29/2019 best practice nurse

    5/50

    Basic Principles of Electricity

    Electricity always . . .

    Seeks ground (its source)

    Seeks the path of least

    resistance

  • 7/29/2019 best practice nurse

    6/50

    Properties of Electricity

    Current Flow of electrons during a

    period of time, measured in

    amperes

    Circuit Pathway for the

    uninterrupted flow of

    electrons (must be

    complete/ closed to flow)

    Impedance Obstacle to the flow of

    current measured in ohms

    Voltage Force pushing current

    through the resistance,

    measured in volts

  • 7/29/2019 best practice nurse

    7/50

    Frequency Spectrum

  • 7/29/2019 best practice nurse

    8/50

    The more concentrated the current,

    the greater the potential for a burn.

    Current Concentration/Dispersal

  • 7/29/2019 best practice nurse

    9/50

    Monopolar

    Active electrode at surgical site Return electrode at another site Current flows through the body

    between the electrodes High voltage

    Coag 3000 9000

    Cut 1350 - 4000

  • 7/29/2019 best practice nurse

    10/50

    Bipolar

    Active and return electrodes in theinstrument

    Current flow confined to tissuebetween electrodes

    Low Voltage (320 - 1200 volts)

  • 7/29/2019 best practice nurse

    11/50

    Electrosurgical Technologies

  • 7/29/2019 best practice nurse

    12/50

    Patient Burns

    Pad site burn

  • 7/29/2019 best practice nurse

    13/50

    Therapeutic current

    Interrogation current

    Contact Quality Monitoring

  • 7/29/2019 best practice nurse

    14/50

    Contact Quality Monitoring

    The most frequently reportedpatient injury has been a skin injury

    at the patient return electrode site

    The risk of this type of injury has

    been minimized through advancesin patient return electrode design

    Nondrying conductive material

    Use of return electrode contact

    quality monitoring

    AORN Recommended Practices, 2009

  • 7/29/2019 best practice nurse

    15/50

    20

    Best Practices to Achieve Favorable PatientOutcomes Related to Electrosurgery

  • 7/29/2019 best practice nurse

    16/50

    Follow Manufacturers Written Instructions

    The ESU and accessories shouldbe used according to the

    manufacturers written instructions.

    Each type of ESU has specificmanufacturers written operating

    instructions to be followed for safeoperation of the unit

    Accessories should be used,handled, cleaned, and processed

    according to manufacturers

    instructions

    AORN Recommended Practices, 2009

  • 7/29/2019 best practice nurse

    17/50

    General Safety Precautions

    Test alarm systems Set activation tone to audible level

  • 7/29/2019 best practice nurse

    18/50

    General Safety Precautions

    Confirm Power SettingsPlug Accessories into Correct Receptacles

  • 7/29/2019 best practice nurse

    19/50

    Selecting the Patient Return Electrode

    Patient return electrodes should becompatible with the ESU

    Do not attempt to use return

    electrodes that disable the contact

    quality monitoring system

    Verify that the: Electrode is intact

    Conductive gel, if present, is moist

    Manufacturers expiration date has

    not been reach

    Do not reuse patient returnelectrode

  • 7/29/2019 best practice nurse

    20/50

    Selecting the Patient Return Electrode Site

    Follow manufacturers writteninstructions

    The conductive and adhesivesurfaces of the electrode should be

    placed on:

    Clean, dry skin

    Over a large, well-perfused

    muscle mass of the surgical

    side

    Close to the surgical site

    Convex area

  • 7/29/2019 best practice nurse

    21/50

    Avoid Patient Return Electrode Sites High

    in Resistance

    Prostheses

    Hair

    Bony prominences

    Scar tissue

  • 7/29/2019 best practice nurse

    22/50

    Preparing the Patient Return Electrode Site

    Follow manufacturers

    instructions for hairremoval, cleaning, and

    drying site

    Protect return

    electrode from fluidinvasion

    Do not use flammable

    agents for PRE sitepreparation

  • 7/29/2019 best practice nurse

    23/50

    Patient Return Electrode Guidelines

    Do not overlap edges

    of return electrodeDo not cut or modify

    return electrode

  • 7/29/2019 best practice nurse

    24/50

    Patient Return Electrode Guidelines

    Do not warm return electrode Apply return electrode after

    positioning

    Place return electrode proximal totourniquet, preferably on operative

    side

  • 7/29/2019 best practice nurse

    25/50

    Follow Manufacturer Instructions When

    Using Different Sizes of Patient Electrodes

    Select patient electrode accordingto patient weight

    Infant patient return electrode

    Do not exceed 120 watts of power Use short activations

    Neonatal patient return electrode

    Do not exceed 300 milliamps of

    power

    Do not activate active electrode

    longer than 30 seconds

    For neonates < .45 kg, use of

    bipolar electrosurgery

  • 7/29/2019 best practice nurse

    26/50

    Compression Stockings and Patient Return

    Electrodes

    Compression stocking maycause

    Increased temperature at the

    pad site

    Increased impedance at padsite

    Current concentration with

    resultant heat build up

    Do not apply the patient returnelectrode to areas wherepressure is applied to the

    patient

    ECRI

  • 7/29/2019 best practice nurse

    27/50

    Warming Devices and the Patient Return

    Electrode

    Always follow manufacturersinstructions when using a warming

    device

    Avoid close contact with warmingblanket

    A towel, blanket, or folded sheetmay be placed between the patient

    return electrode and the warming

    blanket to distance the heating

    source from the patient return

    electrode

    Covidien Clinical Information Hotline

  • 7/29/2019 best practice nurse

    28/50

    Reducing the Potential for Alternate Site

    Burns

    Avoid skin-to-skin contact, such asfingers touching leg or knee

    touching knee when positioning the

    patient

    Use non-conductive dry padding toprevent skin-to-skin contact

    Position the return electrode toprovide a direct current route

  • 7/29/2019 best practice nurse

    29/50

    When to Use Two Patient Return

    Electrodes

    Placement of additional PREs

    increases the dispersion of

    electrosurgical current and heat at

    the pad-to-patient interface

    Obesity

    Emaciation

    Fluid environment

    Long duty cycles

    AORN Recommended Practices, 2009

  • 7/29/2019 best practice nurse

    30/50

    Removing the Patient Return Electrode

    Remove return electrode slowly,supporting the tissue underneath

    Observe skin under pad Chart appearance of site

  • 7/29/2019 best practice nurse

    31/50

    Active Electrodes

    When not in use, always placeactive electrodes in a non-

    conductive holster

    Electrodes that do not fit in theholster should be placed in a

    designated location with tips away

    from flammable material

    Active electrode tips should besecurely seated into the hand

    piece

    AORN Recommended Practices, 2009

  • 7/29/2019 best practice nurse

    32/50

    Active Electrodes

    Do not use suctioncoagulator as a retractor

    Avoid tissue contact withshaft during activation

    Ensure that the outside ofthe suction coagulatorremains free of fluids and

    mucus which may conduct

    electrical current

    Unintended activation mayresult if the suction

    coagulator handswitch isimmersed in saline solution

    or other conductive fluids

    Covidien Instructions for Use, Suction Coagulator

  • 7/29/2019 best practice nurse

    33/50

    Active Electrodes

    Build-up of eschar increasesresistance at the active electrode

    and may require the use of higher

    power setting

    Eschar debris may ignite

    Use a coated electrode to easilyremove eschar buildup onelectrode tip

  • 7/29/2019 best practice nurse

    34/50

    Active Electrodes

    Do not use red rubber catheters or

    other materials as a sheath

    Red rubber or plastic materials

    may ignite

    Use manufacturer approvedinsulated tips

    ERCI, 2006

  • 7/29/2019 best practice nurse

    35/50

    Radio Frequency Current Leakage

    Active electrode cords should notbe wrapped around metal

    instruments

    Active electrode and otherelectrical device cords should not

    be bundled together

    Phippen, Ulmer, & Wells, 2009

  • 7/29/2019 best practice nurse

    36/50

    Avoiding Hemostat Burns

    Do not buzz the hemostat with a needle

    electrode

    Do not lean on the patient, the table, or the

    retractors while buzzing the hemostat

    Activate cut rather than coag

    Firmly grasp as much of the hemostat as

    possible before activating

    Buzz the hemostat below hand level

    Activate the generator after the instrument

    makes contact with the hemostat

    When using a coated or nonstick blade

    electrode, place the edge of the electrode

    against the hemostat Note: Surgical gloves do not insulate

    against RF current

    Ulmer, 2008

  • 7/29/2019 best practice nurse

    37/50

    Pacemakers

    Consult pacemaker manufacturer Use

    electrosurgery with care

    Use bipolar

    Use low power setting

    Avoid current flow through heart and

    pacemaker

    Keep cords away from pacemaker

    and leads

    Postoperatively have the device

    checked by a trained individual

    Phippen, Ulmer, & Wells, 2009

  • 7/29/2019 best practice nurse

    38/50

    Internal Cardioverter-Defibrillators

    Obtain preoperative cardiology consult toevaluate correct functioning and determine

    risks

    Have defibrillator immediately available

    Deactivation of ICD by trained personnel

    before ESU use

    Continuous ECG and peripheral pulse

    monitoring

    Use bipolar

    If monopolar used ensure distance between

    the active and patient return electrode is as

    short as possible

    Avoid current flow through heart and ICD

    Postoperatively have the device checked by

    a trained individual

    AORN Recommended Practices, 2009

  • 7/29/2019 best practice nurse

    39/50

    Hearing Aids

    Best practice is to remove hearingaids prior to surgery

    RF leakage may causeinterference to the hearing aid

    reception

    RF current may damage two piecewireless hearing aids, thus theyshould not be worn when

    electrosurgery is used

    AORN Recommended Practices, 2009

  • 7/29/2019 best practice nurse

    40/50

    Cochlear Implant

    Do not use monopolar electrosurgery on

    the head or neck of a cochlear implant

    patient

    Use bipolar electrosurgery on the head

    and neck of patients with implants that do

    not have extracochlear reference

    electrodes

    The bipolar electrode must not contact the

    implant

    If the implant has extracochlear reference

    electrodes, bipolar may be used if the

    electrode is kept more than 10 cm (4

    inches) from the extracochlear electrodes

    Postoperatively have the device checked

    by a trained individual

    AORN Recommended Practices, 2009; Hotline News, Vol 6, Number 4

  • 7/29/2019 best practice nurse

    41/50

    Dental Appliances

    Avoid direct contact with any metaldental appliances

    Cover braces with teeth guards(silicone or rubber) or use bipolar

    energy when possible

    Remove palate spreaders prior tosurgery if electrosurgery will beused in immediate area of the

    device

    Covidien Clinical Information Hotline

  • 7/29/2019 best practice nurse

    42/50

    Jewelry

    Jewelry should be removed if it iswithin the activation range of the

    active electrode.

    When using a reusable, capacitive-coupled return electrode, all of the

    patients metal jewelry should be

    removed.

    AORN Recommended Practices, 2009

  • 7/29/2019 best practice nurse

    43/50

    Tattoos

    Avoid placing the patient returnelectrode over tattoos

    Inks (red in particular) containmetals which could serve as a heat

    or electrical conductor

    Valleylab Clinical Hotline

    AORN Recommended Practices, 2009

  • 7/29/2019 best practice nurse

    44/50

    Electrosurgery During Pregnancy

    No evidence to contraindicate the use ofelectrosurgery during pregnancy

    Amniotic fluid protects fetus from anyconcentration of electrical current owing

    to the dispersion effect

    Just as high frequency of electrosurgerycurrent above 100 kHz prevents

    neuromuscular stimulation in adults,

    children, and infants, the same is true

    for the fetus Leonardo da Vincic. 1510 - 1512

    Te Linde's Operative Gynecology Text, 8th Edition

  • 7/29/2019 best practice nurse

    45/50

    Electrosurgery During Cesarean Sections

    If flammable prepping solutionused follow manufacturers

    instructions

    Apply the patient return electrodeto a site that will not be exposed to

    pooling of amniotic fluid Activate electrosurgery pencil only

    when in contact with intended

    tissue

    Place electrosurgery pencil in non-conductive safety holster when not

    in use

  • 7/29/2019 best practice nurse

    46/50

    Pediatric Applications and Small Anatomic

    Structures

    Pediatric applications and/or proceduresperformed on small anatomic structures

    may require reduced power settings

    Potential for unintended thermal damageto tissue with high current flow and

    longer activation

    For procedures where the current couldflow though delicate parts of the body,

    use of bipolar technique may be

    desirable in order to avoid unwanted

    coagulation

    Covidien Electrosurgery Users Guide

    Best Practices to Achieve Favorable Patient

  • 7/29/2019 best practice nurse

    47/50

    Best Practices to Achieve Favorable PatientOutcomes Related to Electrosurgery

    1. Manufacturers Instructions

    2. Safety Precautions (2)

    3. Selecting the Patient ReturnElectrode

    4. Selecting the Patient ReturnElectrode Site (2)

    5. Preparing the Patient ReturnElectrode Site

    6. Patient Return Electrode Guidelines

    7. Different Sizes of Patient Electrodes

    8. Compression Stockings and PatientReturn Electrodes

    9. Warming Devices and the PatientReturn Electrode

    10.Alternate Site Burns

    11. Use Two Patient ReturnElectrodes

    12.Active Electrodes (4)

    13. RF Current Leakage

    14.Avoiding Hemostat Burns

    15. Pacemakers

    16. Internal Cardioverter-Defibrillators

    17. Hearing Aids

    18. Cochlear Implant

    19. Dental Appliances & Jewellery (2)

    20. Tattoos

    21. Electrosurgery During Pregnancy /Cesarean Sections (2)

    22. Pediatric Applications and SmallAnatomic Structures

  • 7/29/2019 best practice nurse

    48/50

    Evaluate Patient After Surgery

    Electrosurgery InjuryPressure InjuryChemical Injury

  • 7/29/2019 best practice nurse

    49/50

    Questions ???

    Thank you !

  • 7/29/2019 best practice nurse

    50/50

    Program Summary

    Potential adverse outcomes related to electrosurgery

    Basic principles of electrosurgery

    Best practices to achieve favorable patient outcomes related toelectrosurgery

    Prior to use, clinicians should review the Instructions for Use product

    inserts and/or Users Guide provided by the manufacturer to ensure

    safe and appropriate utilization of electrosurgical devices.