application of reference materials for quality assessment in neutron activation analysis-use of...
TRANSCRIPT
Journal of Radioanalytical and Nuclear Chemisoy, Articles, Vol. 169, No. 2 (1993) 347-361
APPLICATION OF REFERENCE MATERIALS FOR QUALITY ASSESSMENT IN NEUTRON ACTIVATION
ANALYSIS - USE OF INFORMATION THEORY
1. OBRUSNIK,* K. ECKSCHLAGER**
*Nuclear Physics hlstitute, Czechoslovak A cademy of Sciences, 250 68 ~e~ (Czechoslovakia) **Faculty of Natural Sciences, Charles University, 128 40 Prague (Czechoslovakia)
(Received November 16, 1992)
It is generally accepted that an analytical procedure can be regarded as an information production system yielding information on the composition of the analyzed sample. Thus, information theory can be useful and the quantities characterizing the information properties of an analytical method may be applied not only as evaluation criteria but also as objective functions in the optimization. The usability of information theory is demonstrated on the example of neutron activation analysis. Both precision and biasof NAA results are taken into account together with the possible use of reference materials for quality assessment. The influence of the above-mentioned parameters on information properties such as informatiori gain and profitability of NAA results is discussed in detail. It has been proved that information theory is especially useful in choosing suitable reference materials for the quality assessment of routine analytical procedures not only with respect to matrix and analyte concentration in the sample but also to concentrations and uncertainties of certified values in the CRM used. In the extreme trace analysis, CRMs with relatively large uncertainties and very low certified concentrations can still yield rather high information gain of results.
Introduction
The use of information theo ry in analytical chemistry can yield rather
useful and interesting results. Both principles and applications of
information theory to analytical problems have been described in several
review articles and monographs. 1-4 As a rule, analytical information serves as
a basis for decision making in ecology, hygiene, medicine, economy, etc. As
the importance of such decisions has rapidly increased, the demands on the
quality of analytical results and their information content have also
increased.
The amount of information obtained from the analysis (Information Gain,
IG) is determined by metrological properties (precision, accuracy, detection
limit, etc.) of the analytical method used. Information Profitability (IP)
introduces the time or economic factoi" and the' relevamSe of results into
decision making. The information properties (gain, profitability) can be used
for comparison and optimization, o f various analytical metl{ods and procedures.
Elsevier Sequoia S. A., Lausanne, A kaddmiai Kiad& Budapest
I. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
Up to now information theory has been applied to the evaluation and
optimization of various analytical methods including Neutron Activation
Analysis (NAA). s Instrumental (lNAA) procedures have been optimized by means
of information gain and profitability with respect to irradiation, decay and
counting times, precision, relevance and costs. In another work, e a possible
bias of ~t-ray spectrometric results has also been included in the optimization
procedure.
In this work, we have tried to investigate the use of information theory
for the optimization and evaluation of selected cases of INAA procedures
yielding biased results. Moreover, the use of Reference Materials (RMs) for
Quality Assessment (QA) has also been discussed. As the relations used for
optimization are derived for quantitative analysis we assume that the value of
a (method precision) usually does not exceed 10% relative.
We can suppose for simplicity that a can be estimated by the counting
(statistical) error. The value of this error can be controlled to some extent
by altering the decay, irradiation and especially the counting times, by
lowering of spectral background or by including a radiochemical separation
into the NAA procedure in extreme cases.
Theory
In general, the Information Gain IG is given by the extended divergence
measure as
x ~ - x I 1 , . ,, (3 z
E -~2~e "~ 2
We assume that the value of the elemental concentration x is known before
the measurement (a priori) to be only within the broad interval (xl, x ) with
the same probability at any x (rectangular distribution) and that after the
measurement (a posteriori), the value has been .determined with a standard
deviation a (normal distribution). The exponent k = (Or~a)" takes into account
by means of a the uncertainty of an elemental concentration in an RM used for r
QA, and 6 in the second term on the right hand side of Eq. (I) denotes a mean
error (bias). 4'6'7 For accurate results (6 ~ 0) this term vanishes.
In instrumental analytical technique, bias rather arises from improper
calibration or from the calibration procedure or standards available being
inadequate for a perfect elimination of errors appearing during the analysis.
Other sources of bias are inirerferences and contamination or losses in the
348
I, OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
elements determined. Eq. (1) can be applied in the cases where the bias 6 is
either known (or estimated) from the theory or can be established �9 6
experimentally.
Analytical information serves as a basis for making decisions on some
often non-chemical hypothesis. Information enabling a correct decision is
called relevant, s In analytical practice we should often take into account the
cost of an analysis. For this purpose another information property -
Information Profitability IP can be advantageously used and for multicomponent
analysis obtained as
~. I G t . k t I P = ( ,2)
"17
where IG i is the information gain and k i the relevance coefficient for the
i- th element, r denotes the cost of an analysis. In quantitative
multicomponent analysis the coefficients k i are often calculated as a function
of IG i (dynamic model). In some simple cases a static model can also be
untilized. 3'8
R e s u l t s a n d d i s c u s s i o n
Influence of bias 6
Though INAA can often produce rather accurate results (with low or
practically no bias), biased results cannot I~e quite avoided. Therefore, for
application of information theory, the general Eq. (!) for IG should be used.
It enables a comparison of analytical procedures yielding accurate and
inaccurate (biased) results directly from the value of the information gain.
Fig. 1 shows the dependence of IG on the value of a for three different values
of bias 6 (0%, 3% and 10% relative), calculated from Eq. (1).
It can be seen from the figure that 1G for unbiased results increases with
decreasing cr even for very low a values (see the curve for (S = 0%). However,
for biased results (3% or 10%), the IG curves decrease rather rapidly in the
region of highly precise (low a) results. The higher the level of bias the
lower value of IG is obtained. In general, the cases with IG < 0 can be 9
interpreted as a situation where incorrect results misinform us.
Fig. J can easily be applied to investigate the influence of the
ca l ib ra t ion procedure on IG of INAA results. The calibration by means of
synthetic standards prepared from pure elements and compounds can produce an
error (bias of the r e s u l t s ) u p to about 3% relative, Then, IG obtainable by
INAA with this kind of calibration is depicted by the area between IG curves
for 0% and 3% bias. This information gain is reasonably high. Only for highly
349
I. OBRUSNIIC K. ECKSCHLAGER: APPLICATION OF REFERENCE
6
0 5 10 15 20
sigma (~) Fig. i. Dependence os IG on 6" according to Eq. (1)
6 = constant (0%, 3% and 10%)
x e - x I = I 0 0 0 ppm, x = I 0 0 ppm, k = 1
precise results (below e = 3%) IG decreases as the 6 value becomes
s ta t is t ical ly s ignif icant . On the other hand, by using cer t i f ied reference
mater ia ls (CRMs) for cal ibrat ion with c, up to 20% relat ive (it corresponds to r
an uncer ta in ty of about 10% relat ive) for some elements , a rather wide range
(and mostly lower values) of IG can be obtained. Thus, informat ion theory
shows very clearly the disadvantages of using CRMs for cal ibrat ion in INAA, as
pointed out by B E C K E R 1~ and HEYDORN. n
hl/luence of the cost and relevance
It can be der ived from the Eqs (1) and (2) that the cost of an analysis
usually grows more rapidly than the informat ion gain of results. In the
analyses involving t h e measurement of act ivi ty , the value of ~r is inversely
proport ional to the square root of the count ing t ime te, and thus of the cost
of measurement . In this work, we assume for s impl ic i ty that the cost of
measurement is g iven only by the price of count ing t ime t . Fig. 2 shows e
di f fe ren t shapes of the dependence of IG and costs, respect ively, on the a
value.
3 5 0
I. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
Fig. 2.
O �9
~.4
.c �9
0 2 t4---
.c
0- 0 4 8 12
sigma (s~) D e p e n d e n c e os I G and t h e C O s t os a n a l y s t s on i f '
( i n a r b i t r a r y u n i t s ) f o r a n a l y s e s i n v o l v i n g t h e mea-
s u r e m e n t os act iv i ty , other p a r a ~ t e r s " l t k e t n F i g . 1
However, for calculation of information profitability (see Eq. (2)), not
only IG and the cost r but also the relevance of information 3,s should be
taken into account. This case is shown in Figs. 3a and 3b. The IP curve in
Fig. 3a exhibits a maximum while the IG curve does not (for unbiased results).
We applied the dynamic model for calculation of the relevance
coefficient, s with the assumption that we need the results having a between 1%
and 10% relative for our decision making. It can be seen from the figure that
highly precise results (low a) are rarely obtained as the cost of an analysis
grows too rapidly.
Fig. 3b shows the dependence of IP on a computed for several bias levels
and for the same way of relevance coefficient calculation. The higher 6 value
the lower level of IP is obtained. Moreover, the maximum : o n the IP curve
shifts in direction to higher cr for highly biased results. In our case, the
information profitabili ty will reach a maximum for a between 5 and 7% relative
as a further decrease of ~, by prolongation of t , is too expensive.
In the case of multielement INAA, where a group of several elements should
be determined with a good information gain (Mgb relevance), the counting time
35 t
I. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
6
(13
C 5
b4
_I~ 3 LP
0
2 m
(_3
0 i i i i i " i
2 4 6 8 10 12
sigma (%)
Fig. 3a. Dependence os IG and IP on ~ ( i n a r b i t r a r y u n i t s ) DYnamic
~ d e l aS r e l e v a n c e c ~ s 1 6 3 c a l c u l a t i o n s r e s u l t m With ~r
b e t ~ e n I% and IO~ - see 3,s d = 0%; o t h e r p a r a m t e r s l i k e in F i g . I
2.C j
1.8
, , ~ ,
1.5 - - / / ~ \ ,~ 1.3
_.Q
;4-~ 1.0
// �9
(3 ,-~ 0 .5
0.3
0 , 0 i J i J i w r ~ J I ~ i i i i F I I I I 1
0 4 8 2
sigma (~) Fig. 3b. D e p e n d e n c e 0 s I P on ~ ( i n a r b i t r a r y u n i t s ) r e l e v a n c e c o e s 1 6 3 c a l c u l a t i o n l i k e i n F i g . 3 a 3 l e v e l s os 6 (0%, 5%
and 10%) ; o ~ h e r p a r a m e t e r s 1 i k e i n F i g . 1 352
I. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
should be chosen with respect to achieving a satisfactory a level for elements
having the highest detection limits in the group. Otherwise, we will obtain
zero values for some IG i and, consequently, rather low values of IP (Eq. (2)).
In general, an optimization of irradiation , decay and counting times should
be carried out (seeS).
Moreover, in the cases, where we need highly accurate results (precise and
practically without bias) to make a decision, we have to use a different model
for relevance coefficient calculation to obtain maximum information
profitability IP for such highly precise (and expensive) results.
Figs 3a and 3b show quite clearly that it is often not effective to
measure with extremely high precision especially if a relatively high nonzero
bias of results can be present.
Eq. (1) should be discussed in more detail. Any value 6 > 0 should be
substituted regardless of its statistical significance. However, in analytical
practice, when testing by means of reference materials (quality assessment),
we can find out that: 7
a) 5 _< at (m,c0/~n, where t(m,c~) is the critical value of the Student
distribution with m = n-I degrees of freedom, i.e., the bias is not
statistically significant on the level c~; then we only know about the true
information gain of the results that
x ~ - - x I 1 x 2 - - x t I n t 2 ( m , ~ ) ~ I G , ~ I n ( 3 )
6" ,~2ne ~ 2 6" ~2n'e ~
b) 6 > at (m,~)/,] n, i.e., bias is statistically significant and the 6
value found by means of RM should be substituted in Eq. (1).
Fig. 4 shows an example where 5% bias has been found and a CRM with a = r
5% (uncertainty about 2.5%) has been applied for QA. One or two samples of CRM
have been analyzed with each batch of samples. When we apply three measure-
ments (n=3), the area between the curves for 0 and 3% bias depicting the
obtainable information gain (see Eq. (3) for o > 2)) is rather wide and IG can
reach low levels. Below cr = 2% Eq. (1) has to be applied. The information gain
can be improved by using more, e.g. n=5, measurements. Then Eq. (3) should
be applied for o > 4, and the area depicting a possible IG is much narrower
and closer to higher levels of IG. Fig. 4 clearly shows that a simple increase
of the number of measurements will significantly improve the information gain
of the results.
353
I. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
Fig. 4.
3-
.c
. c 1
o
0
L
5 10 15 20
sigma D e p e n d e n c e o s IG on ~" ,i"or b i a s e d r e s u l t s a n d
d i s 1 6 3 n u m b e r o s m e a s u r e m e n t s
CRH w i t h ~ r = 6 % u s e d s QA, 6 = 6 % ; o t h e r p a r a m e L e r s
l i k e i n F i g . l
~ / z ~ IG F o r P e s u l t s w i t h s t a t i s t i c a l l y i n s i g n i s
b i a s ( s ~ 2: 2%) s n = 3
I ~ . IG s r e s u l t s w i t h s t a t i s t i c a l l y i n s i g n i s
b i a s ( s ~ ~ 4%) s n = 5
Influence of the quality of the reference material
Eq. (1) includes many parameters , among them the parameter k =
(~ /~)2 charac ter iz ing the re l iab i l i ty of a qual i ty assessment procedure. The
general Eq. (1) for 6 = 0 models a- case where the exis tence of a nonzero bias
is admi t t ed but it is proved exper imenta l ly that 6 = 0. Therefore , for k < 1
the IG according to Eq. (1) for ~ = 0 is h igher than that according to Eq. ( I )
wi thou t the second term on the r ight hand side (presence of bias is not
assumed). The d i f fe rence is a cont r ibut ion to the informat ion gain fol lowing
f rom the fact that a qual i ty assessment is used. This d i f fe rence depends on
the qua l i ty ( re l iabi l i ty) of the reference mater ia l used (at). It can be as
high as 0.5 natural uni ts (see Fig. 5).
3 5 4
5 -
w ithou 2 RM for QA_
5 .
' - 1-
-1
Fig. 5.
I. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
i i i i t i r i i I I r i ~ I i i f I I i i J i J i i J " l '
5 I0 15
sTgrno RM (~) Influence of RM on IG
= 6%, 6 = 0% ; other parameters like in FJg. I
The values of a and a substituted (by means of k) in Eq. (1) should be r
expressed in absolute concentration units both in the unknown and the
reference samples. The c, value may be a problem - some producers of CRMs r
indicate this value in the certificate. If it is not the case we can estimate
c~ r from the uncertainty (usually expressed as 95% confidence interval) by
using an estimation of the number N of determinations employed for f inding the
certified value as N ~ <15,20> (see12). We assume for simplicity that the
matrix of the CRM is very similar to that of the sample.
Practical aspects of the use of information theory f o r choosing the
optimum RM for QA can be demonstrated by means of Figs. 6-9. Fig. 6 shows a
choice of 5 CRMs availabe for the determination of arsenic in a f ly-ash type
of matrix. These CRMs differ by the producer, the certified concentrations as
well as by the uncertainties (and err) of these concentrations. We assume that
the concentration range of As in analyzed samples of f ly ash is known apriori
to be less than 1000 ppm and real As concentrations in most of the samples are
close to 100 ppm (x2-x 1 = 1000, x = 100 ppm in Eq. (1)).
The calculated dependence of IG on the value of a for the above mentioned
CRMs exhibit maxima for k=l (or r = or). It can be seen that, for given
355
I. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
Fig. 6.
4 - 1 As in Fly Ash
0
- - 1 " F T ' T " T " T ~ L ~ i i i 1 ~ i i i i i i i i r ~ 1 I n 1 . r l i i ~ i i ~ i i ~ -
0 5 10 15 20
sigma (%) Use os CRIdS for QA - determination os As in fly esh
6 = 0 * 4 , x 2 - x t = i O 0 0 p p m , x = t 0 0 p p m
I - BCR CRM 03B Fly Ash 48 ppm • 4.8%
- IR~T ECH Fly Ash 56.gppm • 7.6%
3 - IRANT E O P F l y A s h 79.1ppm • 0 . 4 %
4 - N I S T SRbl 1 6 4 8 U r b a n P a r t i c u l a ' t . e 1 1 5 p p m • e . 7 %
6 - NIST S R H 1 6 3 3 a F l y Ash 1 4 5 ppm • l O %
E
~3 2 tgn
�9 " 4 - -
c- 1
conditions, the BCR 038 Fly Ash yields the highest information gain when used
for QA purposes - other CRMs have higher certified 'values and uncertainty
ranges and, consequently, higher cr values. The disadvantage of using SRM r
1633a in this case is its relatively high a value. The use of this CRM for r
sample containing more than I50-200 ppm of As in quite adequate. Moreover,
uncertainties of CRMs produced in NIST are rather too conservative (high) in
comparison with comparable CRMs from other vendors.
The determination of As in fly ash is a relatively simple case of
practical analysis. The situation can be much more complicated, as in the case
of the determination of Cd traces in biological materials (of living origin).
The range of possible Cd concentrations can cover almost 6 orders of magnitude
from 0.0002 to 200 ppml Here, only the use of CRMs with concentrations similar
356
L OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
Fig. 7,
7-
5.
C
0 C~
0 3.
- 1
s Cd in Biological Material 2 ~ t r a t i o n ) , ~ ~ i
1
i i i i l [ r i F [ I F I J r i i i i j l l l t l r l l l [ ' T l r l l r i ~
5 ~0 15 20
sigma (~) Use 0s CRHs s Ok - determina%ion of Cd in
cal matenial
6 = 0 % , x 2 - x t = 3 0 0 p p m , x = 5 p p m
I - N I S T S R M 1 5 6 6 O y s t e r 3 . 5 p p m • 1 1 %
2 - NIST SRM 1577a B. Llver 0.44 ppm • 14%
3 - N I E S C R M 6 Mussel O. B2 p p m • 3.7%
blologi-
to or lower than the concentrations of Cd in real samples yields a reasonably
high information gain.
Figs 7,8 and 9 show this situation for three levels of Cd concentrations
in real samples: 5, 0.05 amd 0.01 ppm (and apriori known concentration
intervals of 300, 1 and 0.1 ppm, respectively). For the highest concentration
region (Fig. 7) NIES CRM-6 Mussel and NIST SRM 1577a Bovine Liver give the
best information gain while NIST SRM 1566 Oyster Tissue has a relatively high
absolute a value (higher concentration level and higher uncertainty). The use r
of the IAEA H-8 Horse Kidney CRM with certified Cd concentration 189 ppm -+
2.4% contributes no positive IG, as the certified level (and %) is too high
in comparison with the Cd concentrations in the analyzed samples.
For medium concentrations (0.05 ppm of Cd), 4 CRMs with a milk matrix can
be used (Fig. 8). The highest information gain can be achieved by using
uncontaminated milk CRMs with rather low Cd concentrations, like SRM 1549
357
1. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
Fig, 8.
5"
C
�9
0 3-
c
Cd in Biological Material 7- (lOW concentration)
1
/
0 5 10 15 20
sigma (~) U s e o s C R M s s Q A - d e t e r m i n a t i o n o s C d i n
c a l m a t e r i a l
6 = 0 % , x 2 - x I = 1 p p m , x = 0 . 0 5 p p m
- NIST SRM 1549 Milk
3 - BCR CRM OG3 M i l k
3 - BCR CRM 160 M l l k ( s p i k e d )
4 - BCR CRM 151 M i l k ( s p i k e d )
b i o l o g i -
0 . 0 0 0 5 p p m • 4 0 %
0 . 0 0 2 9 p p m • 4 1 %
0 . 0 2 1 8 p p m • 6 . 4 %
0 . 1 0 1 p p m • 7 . 9 %
(NIST) and CRM 063 (BCR) despite the uncertainties in these CRMs being at a
relative level of 40%. Quality assessment by means CRMs with higher Cd
concentrations like SRM 1566, CRM-6 or SRM I577a, cannot yield any information
gain as the certified concentrations are too high in comparison to the 0.05
ppm concentrations of Cd in the samples. It can easily be proved by using
Eq. 1). The CRMs with substantially higher certified concentrations and
low relative uncertainties can be used for the assessment of calibration
(standards, comparators).
Fig. 9 shows and extreme case of analysis of Cd in concentrations at the
0.01 ppm level. In this case, only 3 CRMs with the lowest certified
concentrations (SRM 1549, CRM 063 and CRM 150) can contribute a positive
information gain for quality assessment. The SRM 1549 Milk (NIST) is the best
CRM which could be found for this type of analysis even though the uncertainty
358
Fig. 9.
3
C
C3
O
1
I. OBRUSN1K, K. ECKSCHLAGER: APPLICATION OF REFERENCE
Cd in Biological Material (extra row concentration)
I 1 1 ; I I I T I ~ ] 1 1 1 1 1 I ~ ~ l f } 1 1 1 1 ] 1 1 ~ 1 f I I I I I I
0 5 10 15 2 0
s;gma (~;)
Use os CRMs s OA - d e t e r m i n a t i o n os Cd i n
c a l m a t e r i a l
5 = 0 % , x 2 - x i = O . t p p m , x = 0 . 0 1 ppm
1 - N I S T SRM 1 5 4 9 M i l k
2 - BCR CRM 063 M i l k
3 - BCI~ CRM 1S0 M i l k ( s p i k e d )
b i o l o g i -
0 . 0 0 0 5 ppm • 4 0 %
0 . 0 0 2 9 ppm • 4 1 %
0 . 0 2 1 8 ppm • 6 . 4 %
of the cer t i f ied concentra t ion is 40% relative! It can be expla ined by the
fact that more impor tan t than the relat ive value is a (or uncer ta in ty) r
expressed in concentra t ion units and its ratio to a of the analyt ical method
expressed in the same units. For instance for the last case (Fig. 9) these
values are: 0.00043, 0.0026 and 0.003 ppm for CRMs SRM 1549, C R M 063 and r
C R M 150, respect ively. Other pract ical examples can be explained in a s imi lar
way.
All three f igures show the dependence of IG on the ~ value over a wide
range. Usual ly , we do not expect to achieve bet ter precisions (a) for such
kinds of analysis than 5 - 10% relat ive, in the ext reme case even 10-20%
relat ive. A steep decrease of IG values for h ighly precise results,
p rac t ica l ly for all cases shown in Figs 7,8 and 9, is due to our inabi l i ty
assess the bias of so precise results by means of CRMs wi th much h igher
cr levels. r
359
I. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
All the above-mentioned examples of using information theory for the
evaluation of quality assessment by means of CRMs nicely show that the proper
choice of CRM is very important especially with respect to concentration
levels of the analyte. Moreover, an absolute value (in units of concentration)
of the CRM uncertainty has a crucial effect as it was pointed out by
RASBERRYJ a Information theory proves that for extreme trace analysis even the
use of reference materials with relatively wide uncertainty intervals, like
40% relative in the case of SRM 1549 Milk Powder, can increase the information
gain of the results. From this point of view, new CRMs with very low "natural
levels" of element concentrations - "second generation CRMs" like human
serum 14 will be very useful.
Influence of blank:
In analytical practice, especially in trace analysis, bias is frequently
determined by carrying out a blank experiment 6 o. By subtracting it, the bias
is eliminated or at least reduced; variance, s however, increases by the blank
experiment variance o 2, so that u
= ~(612 + ~02) (2 )
where (7 s is the measurement variance without blank subtraction. i
It has been proved that subtraction of the blank substantially increases
the information gain. z Therefore, blank estimation and subtraction should not
be omitted in NAA procedures used for the determination of extreme trace con-
centrations or in other cases influenced by the blank (INAA of aerosols on
filters, etc.).
Conclusions
This work has shown advantages of information theory for the evaluation
and optimization of some INAA procedures with respect to measurement
parameters like bias, profitability of the results, quality of reference
materials used for QA and blank. Many conclusions in this work could be
applied also to other modern analytical methods. Information theory
significantly helps to choose suitable reference materials for quality
assessment of routine analytical procedures not only with respect to matrix
and analyte concentrations in the sample but also to concentrations and
uncertainties of certified values in the CRM used. This work has proved that
in extreme trace analysis CRMs with relatively large uncertainties (but with
small absolute errors) can be very useful.
360
I. OBRUSNIK, K. ECKSCHLAGER: APPLICATION OF REFERENCE
References
1. K. ECKSCHLAGER, V. ST~P,~NEK, Anal. Chem., 54 (1982) ll15A. 2. K. ECKSCHLAGER, V. ~TI~PANEK, Information Theory as Applied to Chemical Analysis, Wiley, New
York, 1979. 3. K. ECKSCHLAGER, V. ~TEPfiaNEK, Analytical Measurement and Information, Research Studies Press,
Letchworth, 1989. 4. K. ECKSCHLAGER, Collect. Czech. Chem. Commun., 56 (1991) 505. 5. I. OBRUSN~K, K. ECKSCHLAGER, J. Radioanal. Nucl, Chem., 112 (1987) 233. 6. I. OBRUSN~K, K. ECKSCHLAGER, Anal. Chem., 62 (1990) 565. 7. K. ECKSCHLAGER, J. FUSEK, Collect. Czech. Chem. Commun., 53 (1988) 3021. 8. K. ECKSCHLAGER, V. ~TI~PANEK, Chemometr. Intel. Lab. Sys., 1 (1987) 273. 9. K. DANZER, K. ECKSCHLAGER, D. W/ENKE, Frasen. Z. Anal. Chem., 327 (1987) 312.
10. D. BECKER, J. Radioanal. Nucl. Chem., 13 (1987) 5. 11. K. HEYDORN, J. Res. NBS, 93 (1988) 479. 12. J. MUSIL, Chem. listy, 80 (1986) 1233. 13. S. D. RASBERRY, J. Res. NBS, 93 (1988) 213. 14. J. VERSIECK et al., Anal. Chim. Acta, 204 (1988) 63.
361