analysis of benzene

Upload: ecrd

Post on 05-Apr-2018

219 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/2/2019 Analysis of Benzene

    1/10

    Analysis of Benzene

    Envirotech Center for Research & DevelopmentNew Delhi

    Harsh K. Dixit

  • 8/2/2019 Analysis of Benzene

    2/10

    Principle

    The ambient air is sucked through the activatedcharcoal tube using a low flow sampler used forcollection of Banzene in a way that results in anenrichment of the relevant substances in the activated

    charcoal.Desorption of the adsorbed benzene is done by using

    carbon disulphide (CS2)

    The substances desorbed in the CS2 are analyzed by

    capillary gas chromatography. A flame ionizationdetector (FID) is used for analysis while quantification isperformed using the standard.

  • 8/2/2019 Analysis of Benzene

    3/10

    Requirements

    Low flow sampler

    Activated charcoal tube

    Gas chromatography with FID

    Fused silica capillary columns

    Carbon disulphide (CS2) Purity >99.9% (GLC)

    UltrasonicatorCarrier Gas - Helium or Nitrogen Purity >99.9%

  • 8/2/2019 Analysis of Benzene

    4/10

    Sampling Procedure

    A sample is collected by opening a tube at two ends,connecting it to a sampler Airborne chemicals are trapped ontothe surface of the sorbent.

    Two tubes are used in series to take care of breakthroughcompatible to the thermal desorber.

    Keep the tube in a vertical position during sampling toprevent the possibility of channeling that can lead to undersampling.

    Sampling flow rate is around 50ml/m.

    A sampling component may breakthrough from the backend of tube, if excessive flow rates are used. Sample is tobe discarded, if the breakthrough is observed more than10%.

    If analyzed concentration in backup section is more than

    10% of front section, sample needs to be discarded.

  • 8/2/2019 Analysis of Benzene

    5/10

    Analytical Procedure

    Sample collected through sorbent tubes are extracted ordesorbed by 5ml of CS2 using ultrasonication for 15 min.

  • 8/2/2019 Analysis of Benzene

    6/10

    Gas flow

    Nitrogen: 30ml/min (FID make up + column)

    (Column flow ~1ml/min)

    Hydrogen: 30ml/min

    Air: 300ml/min

    Capillary Column 624Coating: cyanopropyl phenyl polysiloxane

    Length: 30 meter

    Internal Diameter: 0.25mm

    Film thickness: 1.4m

  • 8/2/2019 Analysis of Benzene

    7/10

    Temperature programming

    Injection port: 250C

    FID: 300C

    Column/Oven: 50C (hold for 3 min)

    Ramp @ 10C/min

    140C (for 1 min)

    Ramp @ 20C/min

    240C (for 1 min)

  • 8/2/2019 Analysis of Benzene

    8/10

    Retention Time

    Injection volume: 5 l

    Total run time: 19.5 min

    Split: 10

    Benzene: 6.8 min, Search window: 1 sec, 3%

    Toluene: 9.18min, Search Window: 1 sec, 3%

    Xylene: 11.37 min, Search window: 1 sec, 3%

  • 8/2/2019 Analysis of Benzene

    9/10

    Calculation

    Concentration (g/m3) =

    at ambient condition

    WhereC = amount of benzene found injection sample volumefrom standard curve, in g/l

    VT = total volume of the sample extracted in ml

    V2 = volume of sample extract injected in GC, in lV3 = volume of air sucked through the tube, in m3

    Blank value is to be subtracted from the amount ofbenzene found in the sample.

    C x VT x 103V2 x V3

  • 8/2/2019 Analysis of Benzene

    10/10

    Calculation

    Concentration (g/m3) =

    at STP

    Where

    C = concentration at ambient condition, in g/m3

    T = temperature of the ambient air, in C

    P = atmospheric pressure, in kPa

    Volume of air sampled (m3) =

    S = sampling rate, in ml/min

    t = sampling time, in min

    C x 101.3 (273 +T)273 x P

    S x t106