The Anesthesia Machine
Dr. Smriti Gupta
![Page 2: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/2.jpg)
Introduction
• The anesthesia gas machine is a device which delivers a precisely-known but variable gas mixture, including anesthetizing and life-sustaining gases.
• Original Boyle was made by the firm COXTERS.
• There are several differences between newer and older anesthesia machines.
• Advanced ventilators are the biggest difference between newer and older gas machines.
![Page 3: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/3.jpg)
• The differences between older gas machines -such as the Ohmeda modulus, Excel, or Aestiva and the Dräger Narkomed GS, Mobile, MRI, 2B, 2C, 3 or 4- are less than their similarities.
• While the differences between new models from GE Healthcare (ADU, Aisys, Aespire, Avance) and Dräger (Fabius GS, Narkomed 6000/6400, Apollo) are more than their similarities.
![Page 4: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/4.jpg)
Types of anesthesia machine• Intermittent-Gas flows only during inspiration• Egs: Entonox apparatus ,Mackessons
apparatus• Continuous-Gas flows both during inspiration
and expiration. Egs :• Boyle Machine• Forregar • Dragger
![Page 5: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/5.jpg)
History
![Page 6: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/6.jpg)
![Page 7: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/7.jpg)
Anesthesia work station
• Integrates most of the components necessary for administration of anesthesia into single unit.
• Standard guidelines have been given to manufacturers for minimum performance, design, characteristics and safety requirements of machine.
• The current standard for anesthesia workstation as( promulgated by American society for testing and materials) (ASTM) is F1850. European standard is EN740
![Page 8: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/8.jpg)
Components of anesthesia machine
![Page 9: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/9.jpg)
Pressure units to remember
1 atm pressure = 1 bar = 760 mm Hg =14.7 psi = 100 kilopascals
![Page 10: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/10.jpg)
Components of pressure systems
![Page 11: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/11.jpg)
![Page 12: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/12.jpg)
![Page 13: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/13.jpg)
![Page 14: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/14.jpg)
![Page 15: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/15.jpg)
![Page 16: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/16.jpg)
Limitations of check valve assembly
• The check valves are not designed to act as
permanent seals for empty yokes.
• Small amounts of gases can escape if the yoke is
empty or an empty cylinder (or cylinder with low
pressure) and valve open is present in the yoke.
![Page 17: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/17.jpg)
• In order to minimize such losses –
• Yokes should not be left vacant for extended periods
• An empty cylinder should be replaced as soon as possible , if
not then,
• An yoke plug can be used to prevent gas leak or
• An empty cylinder can be left behind after closing the valve
![Page 18: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/18.jpg)
Yoke Block(Dummy Cylinder plug)
• Yoke block is a solid piece of metal that has a conical depression on one side to fit into the retaining screw and a hollow area on the other side to fit over the nipple.
• They are pin indexed.• Uses • Prevents gas leak from the machine when placed in an
empty yoke.• Connect cylinders larger than size E to the machine.• Connect pipeline supply to the machine that does not
have pipeline inlet connections.
![Page 19: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/19.jpg)
Limitations of yoke block
• Can be responsible for crossover of gases in the machine.
• Pressure regulators in the machine are designed to accept gases at cylinder pressure may not function properly when supplied with gases at reduced pressure(pipeline supply through yoke block)
![Page 20: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/20.jpg)
![Page 21: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/21.jpg)
Limitations of pin index safety system
• A wrong cylinder can be placed in yoke if-• 2 washers are placed on the port.• Pins on the yoke are broken.• Holes on the cylinder valve are too deep.
![Page 22: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/22.jpg)
![Page 23: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/23.jpg)
![Page 24: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/24.jpg)
Pin index safety system
![Page 25: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/25.jpg)
![Page 26: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/26.jpg)
Bourdon pressure gauge
![Page 27: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/27.jpg)
![Page 28: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/28.jpg)
Safety features in cylinder pressure indicator
• Gauge is usually color coded.• Name and symbol of gas are written over dial.• If bourdon tube ruptures gas is vented from
back side • Gauges are angled and placed in such a way
that it can be easily read by anesthetist.• Instructions like “use no oil’’ “open the valve
slowly’’ are written on the gauge.
![Page 29: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/29.jpg)
Electronic cylinder pressure indicator
• Light emitting diodes(LED’S)in electronic pressure gauge indicate
• Cylinder valve is close –Dark color• Cylinder valve is open –• Pressure adequate –Green• Pressure inadequate-Red
![Page 30: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/30.jpg)
![Page 31: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/31.jpg)
![Page 32: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/32.jpg)
Types of pressure regulators
• Adam’s pressure regulator valve-• Used in past• Working principle is same differs in internal structure
from newer valves(toggle levers instead of push rod)• Fins on Adam’s regulator for N2O which are not present
in newer regulators.• Mackesson’s regulator• Medishield valve • Endurance valve• Modern preset type of reducing valve-Direct and Indirect.
![Page 33: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/33.jpg)
Safety features on pressure regulator
• Pressure regulators have safety relief valves• If due to any reason there is build up of
pressure in pressure regulator then the safety valve blow off at a set pressure of 525 k pa(70psi)
![Page 34: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/34.jpg)
![Page 35: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/35.jpg)
![Page 36: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/36.jpg)
![Page 37: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/37.jpg)
![Page 38: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/38.jpg)
![Page 39: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/39.jpg)
![Page 40: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/40.jpg)
Oxygen failure safety valve
Pressure sensor shut off valve• Based on threshold principle• Cuts off N2O supply when
O2 pressure falls below 25 psi.
• MOA-O2 pressurises and holds open shut off valve that interrupts the supply of N2O and other gases if O2 pressure<threshold setting.
oxygen failure protection device• Based on proportioning principle
• Gas loaded regulator• When O2 pressure
decreases there is a proportional decrease in N2O supply and complete cut off seen at<12psi.
• MOA-O2 pressure regulator(primary regulator) controls secondary(slave) regulator located in N2O line
![Page 41: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/41.jpg)
![Page 42: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/42.jpg)
![Page 43: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/43.jpg)
![Page 44: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/44.jpg)
![Page 45: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/45.jpg)
![Page 46: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/46.jpg)
![Page 47: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/47.jpg)
Master switch
• Turning the master switch to the ‘on’ position activates both pneumatic and electrical functions of the machine as well as certain alarms and safety devices.
![Page 48: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/48.jpg)
![Page 49: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/49.jpg)
![Page 50: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/50.jpg)
![Page 51: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/51.jpg)
![Page 52: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/52.jpg)
![Page 53: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/53.jpg)
![Page 54: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/54.jpg)
![Page 55: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/55.jpg)
![Page 56: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/56.jpg)
![Page 57: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/57.jpg)
![Page 58: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/58.jpg)
![Page 59: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/59.jpg)
![Page 60: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/60.jpg)
![Page 61: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/61.jpg)
Electronic flow meter
• Prefer digital system• Solenoid valves• Control flow on or off
valves• Computer controlled
• .
![Page 62: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/62.jpg)
![Page 63: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/63.jpg)
![Page 64: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/64.jpg)
![Page 65: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/65.jpg)
![Page 66: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/66.jpg)
![Page 67: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/67.jpg)
![Page 68: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/68.jpg)
![Page 69: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/69.jpg)
Link-25
• .
![Page 70: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/70.jpg)
Oxygen ratio monitor controllerSensitive oxygen ratio controller
• Linear resistors [3:1 ratio for N2O & O2)between O2 andN2O flow control valves.
• Ensure 25% O2 by limiting N2O flow.
• ORMC shuts off N2O if ratio of O2 flow falls below 30%
• S-ORC-newest hypoxic guard . installed in Fabius-GS by Drager.ensures aFiO2 of 23%. O2 flow <200ml/min.
• .
![Page 71: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/71.jpg)
Limitations of proportioning system
• Anti hypoxic device may also deliver hypoxic mixture under following conditions-
• Wrong supply of gas.• Defective pneumatics/mechanics.• Inert gas administration(3rd gas-He,N2,CO2).• Leaks downstream.
![Page 72: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/72.jpg)
![Page 73: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/73.jpg)
![Page 74: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/74.jpg)
![Page 75: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/75.jpg)
![Page 76: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/76.jpg)
![Page 77: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/77.jpg)
![Page 78: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/78.jpg)
![Page 79: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/79.jpg)
![Page 80: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/80.jpg)
![Page 81: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/81.jpg)
![Page 82: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/82.jpg)
![Page 83: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/83.jpg)
![Page 84: Anaesthesia machine](https://reader033.vdocuments.mx/reader033/viewer/2022061312/5551727db4c905723b8b4dc9/html5/thumbnails/84.jpg)
Thank You