spo2 for neonates tl-535u, tl-273t3 - nihon kohden · pdf filereused for length of...
Post on 19-Feb-2018
216 Views
Preview:
TRANSCRIPT
Proven Accuracy 1Nihon Kohden commissioned two studies to evaluate pulse oximetry probes at the esteemed Hypoxia Research Laboratory at The University of California, San Francisco in 2008 and 2010.1 All tested probes meet the accuracy specifications and exceed the ISO 80601-2-61: 2011 specification for SpO2 probe accuracy.1
ACCURACY CRITERIA FOR SpO2 MEASUREMENT
SpO2 for Neonates TL-535U, TL-273T3
Care for Your TiniestPatients
us.nihonkohden.com Different Thinking for Better Healthcare.®
Probe TL-535U
Result for TL-535U Bias (SpO2-SaO2) Analysis
Hemoximeter SaO2 Range 70 – 80% 80 – 100%
Count 162 356
Mean 0.72% 0.03%
Standard Deviation 1.88% 1.37%
Root Mean Square (RMS) 2.01% 1.37%
ISO 80601-2-61: 2011 3% or less 2% or less
SpO2 for NeonatesTL-535U, TL-273T3
References 1 INTERNAL: SpO2 Monitoring Pulse Oximeter Accuracy Study, 2008 study by John R. Feiner, M.D., Phillip E. Bickler, Ph.D.M.D., and John W. Severinghaus, M.D. For the 2010 study, authors are Phillip E. Bickler, Ph.D.M.D. and John R. Feiner, M.D. at the Hypoxia Research Laboratory at the University of California, San Francisco.
THIS DATA SHEET MAY BE REVISED OR REPLACED BY NIHON KOHDEN AT ANY TIME WITHOUT NOTICE.
us.nihonkohden.com
MMLB 149[A]-CO-1253
NIHON KOHDEN AMERICA, INC. | 15353 BARRANCA PARKWAY, IRVINE, CA 92618
800.325.0283 | 949.580.1555 | FAX: 949.580.1550 | EMAIL: INFO@NIHONKOHDEN.COM
Comfortable Patients Simplifies Clinical WorkflowTL-535U:
• Attachment wrap uses soft foam where it touches baby’s skin
• No Adhesive – Attachment wrap uses supremely soft Velcro®-style closure
• Sensor feels like soft felt, designed so sensor cord doesn’t leave indentation on baby’s delicate skin
• TL-535U Semi-Disposable Sensor can be reused for length of baby’s stay in NICU, simply change the wrap to freshen
• TL-273T3 uses traditional soft adhesive
• Neonate SpO2 options depending on patient needs, choose traditional adhesive or non-adhesive sensors for more sensitive patients
• Neonatal Dual-SpO2 for detecting duct dependent congenital heart diseases in the newborn – dual SpO2 is simplified with the Nihon Kohden Smart Cable™ – just plug and measure
Velcro is a registered trademark of Velcro BVBA.
Smart Cable is a trademark of Nihon Kohden Coorporation.
Different Thinking for Better Healthcare is a registered trademark of Nihon Kohden.
5 SpO2 sensors, each has 3 size S attachment wraps
24
24
1
1
1
24
1
1
Neonate Soft Sensor
Attachment Wrap Size S
Attachment Wrap Size L
Connection Cable 0.3m for TL-535U Sensor
Connection Cable 1m for TL-535U Sensor
SpO2 Cable 3m for TL-535U
Neonate Adhesive SpO2
SpO2 for TL-273T3 or Connection Cord for
TL-535U
Dual SpO2 for TL-273T3 or Connection Cord
for TL-535U
TL-535U
YS-102P0
YS-102P1
JL-030U1
JL-030U2
JL-930P
TL-273T3
JL-900P
JL-500P1
ITEM PART NUMBER QUANTITY
① YS-102P0: Small Attachment Wrap② YS-102P1: Large Attachment Wrap③ TL-535U Sensor: Neonate Semi-Disposable Non-Adhesive④ TL-273T3: Neonate Adhesive-Style SpO2 Probe ⑤ JL-930P: SpO2 Cable for TL-535U, 3m⑥ JL-030U1: Connection Cable for TL-535U, 0.3m⑦ JL-900P: SpO2 Cable for TL-273T3 or Connection Cable for TL-535U⑧ JL-500P1: Dual SpO2 Cable for TL-273T3 or Connection Cable for TL-535U
①
③
④ ⑤
⑦
⑥
⑧
②
SpO2 for Neonates Product Line
top related