problems the ratios e/m of 4 he + and d 2 + ions are very close: m d 2 = 4.0282 amu m he = 4.0026...
Post on 29-Mar-2015
220 Views
Preview:
TRANSCRIPT
ProblemsProblems The ratios e/m of 4He+ and D2
+ ions are very close:
mD2 = 4.0282 amu
mHe = 4.0026 amum = 0.0256 amu
A high resolution mass spectrometer should be required in order to separate the two peaks.
D2 concentration » expected 4He concentration
Capability of on line analysis during the electrolytic experiment
Conventional techniquesConventional techniques Dynamic analysis technique (differential
pumping or diffusion through large impedance circuits):
alter the original composition of the gas mixture; exhibit low sensitivity; do not allow for on line analysis.
Static analysis technique (Cryosorption pumps):
may erratically trap helium in the condensates; require a long cycle time and are not suitable for
on line analysis. Unable to pump hydrogen isotopes to high degree
Innovative solutionsInnovative solutions Non evaporable getter (NEG) pumps:
remove to a high degree all non inert components of the gas mixture (especially hydrogen isotopes); all inert gases are obviously preserved;
exhibit very high pumping speeds (especially for hydrogen);
the pumping speed for other active gases depends on the operating temperature of the getter alloy;
do not require frequent regeneration, allowing periodic analysis of gas samples during the experiment (roughly every 40 minutes).
A storage circuit has been purposely designed, which:
allows to periodically sample the gas mixture without influence on the electrolytic process;
eliminates any alteration of the gas mixture composition; allows to easily compare the results of the analysis of
each sample.
Main featuresMain features High resolution Mass
Spectrometer (Balzers QMA 410).
Two NEG pumps are used: SAES Capacitorr B 1300 (600 g of
ST185 TiV alloy) operating at high temperature (300 - 400°C)
SAES GP200 MK4 W (170 g of ST707 alloy) operating at RT.
Dynamic pumping: Main pump: Pfeiffer TMU 261
turbo molecular drag pump Backing pump: Pfeiffer TMU 071
+ MD4
No cryosorption pumps are used.
Automatic operation (Field-Point + LabView).
Pressure stabilized storage circuit.
Schematic layout of the gas Schematic layout of the gas analysis systemanalysis system
QMAUHV
chamber
Thermostatic Box
Cell
UHVcontainment
vessel
Storagecircuit
Samplevolume
HT NEGpump
n.o.
n.o.
n.c.
n.o. = normally open valve
n.c. = normally closed valve
Gas mixture inert gases
RT NEGpump
GV1
GV3GV5
High resolution spectra of a 4He-D2 mixture.
dwell time 1 s, scan width 10 amu, scan time for each cycle 10s.
Analog spectrum of an air sampleAnalog spectrum of an air sample
0 15 20 355 10 25 30 40 45 50
Mass (amu)
Ion c
urr
ent
(x1
0-8
A)
0
1
2
3
4
5
640Ar+
40Ar++