alignment tests review of clic two beam module lab program 06/11/2013 on behalf of clic...
TRANSCRIPT
ALIGNMENT TESTS
Review of CLIC Two Beam Module lab program
06/11/2013
On behalf of CLIC pre-alignment team
2
Summary
4 main tasks/objectives of the alignment tests on TBTM:
Validation of measurement methods for the alignment tests
Validation of the pre-alignment strategy on short range
Inter-comparison between alignment systems on short range
Study of the alignment of supports and components when the conditions change:
Additional constraints like waveguides, connection to vacuum pipes, vacuum
Thermal tests
Conclusion : resources needed and next tasks
Initially foreseen on an independent mock-up
3
Validation of measurements methods for the alignment tests
Objectives:
To have a range of tools:• Allowing precise and accurate measurements• Allowing cross check of measurements• Taking into consideration the small space available around the module
Several instruments qualified using CMM measurements as reference (Leitz Infinity: 0.3 μm + 1 ppm)
Romer arm AT401 Micro-triangulation
Performances(over 2 m)
~ 10 µm ~ 5 µm ~ 5 µm
Drawbacks Limited range Displacement of the prism, contact with the object
Needs permanent stations
4
Validation of measurements methods for the alignment tests
Inter-comparison between micro-triangulation and AT401:
Two MB girders equipped with both types of fiducials and measured on CMM
Alignment of girders measured by the two instruments and compared
5
Validation of measurements methods for the alignment tests
Other methods to be considered for the alignment tests:
Photogrammetry• This could be very interesting for thermal tests as it is very
quick: a series of pictures is needed on site, then analysis can be performed far from the module.
• Not ready for such an accuracy, targets need to be adapted (R&D needed)
Other methods to be considered for integration purposes:
3D scans• To solve the problems of integration that were met (3D models
did not correspond to what was installed a lot of time lost)
Postponed due to lack
of time
Postponed due to lack
of time
6
Fiducialisation of components
Fiducialisation of their common support
Alignment on a common support
Whole assembly ready to be aligned
Validation of the pre-alignment strategy on short range
Strategy of pre-alignment:
7
Budget of alignment errors
Requirements:
Budget of error:
The zero of each component will be included in a cylinder with a radius of a few microns:
14 µm (RF structures & MB quad BPM)
17 µm (MB quad)
20 µm (DB quad)
Steps AS MB quad
Achieved
Zero of components to fiducials 5 µm 10 µm ????
Fiducials to sensor interface on support 5 µm 5 µm Yes
Sensor interface on support 5 µm 5 µm Yes
Sensor measurement w.r.t straight reference
5 µm 5 µm Yes
Stability knowledge of the straight reference
10 µm 10 µm Yes, on short range
Total error budget 14 µm 17 µm
The combination of the 3 first steps is the object of PACMAN
Validation of the pre-alignment strategy on short range
8
Determination of the position
Reference network: layout and configuration of sensors
Comparison of the alignment of the mean axis of the Ves by AT401 and alignment sensors
X (µm) Z (µm)
MC1-E -5 12
MC1-S 10 -5
MC2-E -1 -4
MC2-S -11 -7
Difference between coordinates of mean axis extremities calculated by 2 different methods
Validation of the pre-alignment strategy on short range
9
Re-adjustment
Two solutions studied in parallel:
Algorithms of re-positioning: relative ok, absolute to be tested
Linear actuators
Cam movers
Ok when no constraints
Not ready
Validation of the pre-alignment strategy on short range
10
Case of the supports:
Girder:• Mean axis of the V-shaped supports:
Boostec: radius of the cylinder containing the center of the V-shaped support : 6μm and 4μm
Micro-Contrôle: radius of the cylinder containing the center of the V-shaped support: 7.5 μm and 5.5 μm
Girder + cradle: • Measurements out of the range of the CMM: accuracy ~ 15 µm, some
faults detected.• Total length above 2 m• Different types of fiducials implied different types of measuring devices
sometimes outside the range of measurement.
Articulation point:• Not so bad at the beginning• Degradation along time (shocks, loads, constraints)
Validation of the pre-alignment strategy on short range
11
Case of the components:
2 steps:• Determination of the position within a few microns• Alignment on the support
PETS:• Assembly ok• No problem on alignment (on V-shaped supports)
DB quad:• System of adjustment not ok, no stability of the position (offset of 200 μm
w.r.t. theoretical position) => new system under design
AS:• Assembly not ok
Validation of the pre-alignment strategy on short range
12
Inter-comparison between alignment systems
oWPS versus cWPS:
Summary of sensor performances
oWPS (μm) cWPS (μm)
Noise 2 8
Repeatability < 1.5 < 2
Reproducibility < 2 < 3
Interchangeability 6 - 8 3 - 4
Resolution < 0.5 < 0.5
Accuracy of measurement of oWPS is ~ 20 µm and needs to be improved Accuracy of measurement of cWPS is ~ 5 µm thanks to new benches and new
procedures of calibration.
Noise of cWPS is a serious drawback for active alignment and needs to be understood.
13
Inter-comparison between alignment systems
cWPS versus NIKHEF alignment systems:
2 main systems installed by NIKHEF (+1 in longitudinal)
RasDif RasNik
Development of sensors
Qualification (@ NIKHEF)
Integration in 3D models
Installation & analysis of data
• Length between girders not the same
• Use of oWPS interface• Choice of the components• Software, preparation of
database
• Influence of T°• Necessity of thermal
shielding
• Exchange of 3D models
• Longitudinal position of cradles not good
• Interferences with other systems
• One cradle with problem
14
Inter-comparison between alignment systems
Need to develop an inclinometer that is absolute:
To avoid 2 wires per beam, 4 wires per module, as in lab and CLEX
Difficulty: absolute measurement combined with kinematic interface
Development of a special measurement bench and special tool, to be tested on TBTM
Next step: development of a rad hard version (manufacturers are not interested to do this in-house development)
15
Alignment of components & supports when conditions change
Test of DOF of girders along 3 steps of installation:
Step 0: components installed Step 1: connection of the bellows of the vacuum tank with PETS and AS Step 2: connection of the waveguides between AS and PETS Step 3: connection of the vacuum network between TANK and AS.
16
Alignment of components & supports when conditions change
Thermal tests: introduction
First tests performed between 20°C and 40°C to check that the performed measurements are correct:
• Network all around the room on the concrete beams of the ceiling and walls
• Fiducials of the girders considered as reference of measurement (low thermal expansion of the girder): best fits were performed with measurements performed at 20°C by CMM, to check the coefficient of thermal expansion.)
• Redundancy of measurements and study of the residuals
Special care for all the measurements:• Nobody else inside• Station < 45’• Use of a heavy tripod• Warm-up of instruments
Cross-check with other methods (photogrammetry, micro-triangulation under study)
Nb of days: 26 Stations: 165 Measurements: > 18 000
17
Alignment of components & supports when conditions change
Thermal tests: some particular cases
Repeatable measurements
Warm-up of DB components has no impact on MB components
The initial misalignment of components that is important in some cases makes the displacements more difficult to be understood
18
Alignment of components & supports when conditions change
Vacuum tests
Displacement of girders
Roll:• DB: ~ 1mrad (T0-1), ~ 0.1 mrad (T0-2)• MB: 0.327 mrad (T0-1), 0 (T0-2)
Displacement of cradles
19
Alignment of components & supports when conditions change
Vacuum tests
Displacement of cradles versus girders
Non repeatability
Consequences:• Fiducialisation lost: no possibility to perform again absolute
measurements re-measure on CMM needed• ZTS vvu Kosice has copied this solution for CLEX same problem for CLEX
no possibility to align the components in an absolute way• Articulation point lost
20
Alignment of components & supports when conditions change
Vacuum tests
Independence of girders
Impact on components:
• Less than 10 μm for PETS and DBQ1• Longitudinal displacements of 678 μm for DBQ2• Displacements of 115 μm in radial for AS1
21
Next tasks
Validation of measurements methods: Measurements of 1 module take ~ 1 day: to improve the speed if needed up to
30’, R&D needed [0.2 FTE] Development of photogrammetry and micro-triangulation (if possible to have
permanent stations) [0.2 FTE] Implementation of 3D scans [0.2 FTE]
Validation of the pre-alignment strategy: Validation of absolute repositioning algorithm (if module re-fiducialised) DB quad support:
• Validation of the prototype• Design of the support• Qualification on DB type 1
Inter-comparison between alignment systems: Cross-check measurements of RasNik, RasDif and cWPS Re-installation of oWPS once recalibrated and comparison between cWPS and
oWPS Impact of temperature on sensors.
22
Next tasks
Alignment and fiducialisation of components
Type 0-1:• Transport tests (and re-alignment of all components and girders if needed)• Any additional tests• Refiducialisation of the module
Type 0-2:• Control of assembly and fiducialisation of components AS and PETS• Control of their alignment on girders• Alignment of the 2 modules type 0, tests of actuators, tests of articulation
point, etc.• Tests with constraints, T°, vacuum
Type 1:• Control of assembly of PETS, AS• Fiducialisation of DB quad, PETS, AS• Design, order, assembly of the cradle linking MC girder to Boostec girder (MB)• Assembly of articulation + cradles on Epucret girder• Alignment cradle versus girder on MB and DB side• Fiducialisation of girders + cradles• Fiducialisation of supports + stabilization system + MB quad
23
Next tasks
Alignment and fiducialisation of components
Type 1:• Assembly of cam movers• Installation and validation of cam movers (+ control/command system)• Installation and validation of alignment sensors (on cradles and MB quad) (+
acquisition system + software + database)• Aligment of all the supports• Tests of actuators and cam movers• Tests of absolute repositioning• Tests with constraints?
Transfer Type 0- Type 1:• Study of the new configuration (longitudinal problem to be solved)• Design of new parts, procurement, assembly,…• Fiducialisation of new cradles• Dismounting, marking on the floor, drilling, reinstallation of the new solution• Alignment of the new configuration
24
Next tasks
Alignment and fiducialisation of components
Type 4:• Assembly, fiducialisation of DB girder• Manufacturing (redesign?) of articulation point• Fiducialisation of components: DB quad• Design of supporting system (and sensor interfaces), procurement,
fiducialisation• Control of assembly of MB quad, MB quad + stabilization system, MB quad +
stabilization system + supporting system• Preparation of the algorithms of repositioning• Installation and validation of cam movers (+ control/command system)• Installation and validation of alignment sensors (on cradles and MB quad) (+
acquisition system + software + database)• Alignment of type 4• Tests of actuators and cam movers• Tests of relative, absolute repositioning• Tests with constraints?
25
Resources linked to the TBTM in lab
Study of cam movers1 FTE (PhD student) 0.4 FTE in 2014 ?
Mechatronics0.7 FTE (PJAS student) 0.3 FTE in 2012
0.3 FTE in 20130.3 FTE in 2014
Fiducialisation, alignment1 FTE (fellow) 0.8 FTE in 2013
0.8 FTE in 2014Sensors, actuators
1 FTE (fellow –PJAS?) 0.6 FTE in 20120.6 FTE in 20130.4 FTE in 2014
Mechanics, prototypes0.5 FTE (FSU) 0.3 FTE in 2013
0.3 FTE in 2014
SupervisionM. Sosin: 0.3H. Mainaud Durand: 0.6
Help from ABP/SU (oWPS, photogrammetry, scans, second operator)
CLIC TBTM in lab
26
Summary of the situation
Alignment tests on TBTM
Development and qualification of sensors
Development and qualification of
actuators
Integration of alignment systems
Fiducialisation with the metrology lab
Mechanical designs
Design of articulation points & cradles
Study of new methods of measurements
Development of acquisition system,
databases, analysis scripts
Implementation of a measurement lab
Alignment tests on CLEX
27
List of publications
• IPAC 2011: Theoretical and practical feasibility demonstration of a micrometric remotely
controlled pre-alignment system for the CLIC linear collider, H. Mainaud Durand et al.
Validation of micrometric remotely controlled pre-alignment system for the CLIC test setup with 5 DOF, H. Mainaud Durand et al.
• MEDSI 2012: Issues & feasibility demonstration of positioning closed loop control for the
CLIC supporting system using a test mock-up with 5 DOF, M. Sosin et al. CLIC MB quadrupole active pre-alignment based on cam movers, J.
Kemppinen et al.• FIG 2012:
Augmentation of total stations by CDD sensors for automated contactless high precision metrology, S. Guillaume
• IWAA 2012: Validation of the CLIC alignment strategy, H. Mainaud Durand et al. oWPS versus cWPS, H. Mainaud Durand et al.
• IPAC 2012: Strategy and validation of fiducialisation for the pre-alignment of CLIC
components, S. Griffet et al.
28
List of reports
• 1096126: Evaluation du laser tracker AT401 par 1 CMM• 1096127: Simulation d’un réseau applicable à un module CLIC• 1096130: Fiducialisation & pre-alignment• 1096133: Pre-alignment solutions applied to girders• 1097661: Fiducialisation & dimensional control• 1098660: Poutres Boostec: tolérances à contrôler sur site• 1100438: Poutres Epucret: tolérances à contrôler sur site• 1103378: Contrôle des poutres Boostec sur site• 1106507: Evaluation des performances du prototype de micro-triangulation• 1108528: Evaluation des performances du bras de mesures Romer Multi Gage• 1108692: Contrôle des poutres Micro-Contrôle sur site• 1131579: Emplacement des fiducielles et interfaces capteur sur les poutres de la
maquette TM0• 1137443: Measurements of MB supporting systems, fiducialisation• 1141392: Qualification of linear actuators from ZTS vvu Kosice• 1142857: Contrôle de la position des poutres Micro-contrôle au bâtiment 169• 1146050: Evaluation du laser tracker AT401 par des mesures du banc de micro-
triangulation• 1155733: Inter-comparison of measurements performed on the micro-triangulation
bench• 1163017: Mesures laser tracker sur les poutres TM0 de la maquette CLIC• 1166274: Coordonnées des fiducielles des composants de la maquette CLIC• 1171946: Alignement des DBQ sur la maquette TM0• 1175924: Contrôle des PETS à l’aide du bras Romer Multi Gage, confrontation aux
mesures CMM• 1218458: Inter-comparaison par des mesures sur la maquette CLIC TM0: micro-
triangulation et laser tracker AT401
29
List of reports
• 1209967: Influence de l’installation des DB quad et PETS sur l’alignement des poutres• 1218458: Procédure de fiducialisation des 4 premières structures accélératrices et
calendrier associé• 1227067: Fiducialisation des 4 premières structures accélératrices: résultats et analyse• 1233948: Fiducialisation du 2ème stack TM0: résultats et analyse• 1242279: 1er et 2ème stack TM0:alignement avant EBW, contrôle sur poutres après• 1246581: Rattachement des plaques aux extrémités de la maquette CLIC• 1247059: Test T-Scan CS• 1257114: Mesures de la maquette avant RFN• 1273476: Rapport de test à réception du bras Romer Multi Gage 12/12/12• 1308072: Dimensional control and fiducialisation of DB girder (Epucret) for the TM1 of
the lab• 1308123: Influence of different factors on the mock-up (connection between the different
components and thermal test)• 1308128: Control of the position of the components during the assembly steps• 1308603: Tests des nouveaux supports photogrammétriques• 1309127: Tests des nouveaux supports 1.5’’ amagnétiques aux aimants amovibles• 1322106: ZTS linear actuators test report• 1325401: Historique des décalages des points d’articulation sur la maquette test module• 1325402: Variations des lectures des capteurs lors du changement de température de la
maquette du test module• 1325403: Impact du vide sur l’alignement de la maquette CLIC Test Module• 1325404: Test de contrainte lié aux connexions entre le MB et le DB de la maquette CLIC
test module