agilent 6400 series qqq method development triple ......agilent 6400 series triple quadrupole qqq...

46
Agilent 6400 Series Triple Quadrupole QQQ Method Development and Optimization Triple Quadrupole LC/MS/MS Users Session and Optimization MassHunter Quant: Method setup Peak detection optimization Q Quant troubleshooting David Presser David Presser Application Scientist Fall 2009 MH User workshop QQQ fall2009.pptx 1

Upload: others

Post on 09-Mar-2020

51 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Agilent 6400 Series Triple Quadrupole

QQQ Method Development and OptimizationTriple Quadrupole

LC/MS/MS Users Session

and Optimization

MassHunter Quant:Method setupPeak detection optimizationQQuant troubleshooting

David PresserDavid PresserApplication Scientist

Fall 2009MH User workshop QQQ fall2009.pptx

1

Page 2: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Agenda

•QQQ method development and optimizationSource optimization for standard and Agilent Jetstream sources– Source optimization for standard and Agilent Jetstream sources

– Compound optimization using flow injection or Optimizer software• MassHunter Quant software

– Current revision B.03.02: What's New– Review of Quant methods: creation from acquired data, existing batch

Global and Advanced parameters– Global and Advanced parameters– Optimizing peak identification and rejection– Quant method troubleshooting: identification and integration

Fall 2009MH User workshop QQQ fall2009.pptx

2

Page 3: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Agilent ESI and APCI sources:for polar to non-polar compoundsfor polar to non polar compounds

Nebulizer Pressure

NebulizerPressure

Pressure

Corona current

Vaporizer

VcapVcap

Fragmentor V

Drying Gas Temperature and Flow Fragmentor V

Drying gasTemperature

and Flow

Fall 2009MH User workshop QQQ fall2009.pptx

3

Page 4: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Agilent Multimode source for flexibilityESI or APCI or Mixed mode (simultaneous ESI/APCI)( )

HPLC inlet

Nebulizer

ESI Zone: Charging V

APCI ZoneIR emitters:"Vaporizer"

Fragmentor VCorona needle

Vaporizer

Drying gas

needle

Capillary V

Fall 2009MH User workshop QQQ fall2009.pptx

4

Page 5: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Agilent Jetstream Technology (AJT)Th h t d h thThe super-heated sheath gas collimates the nebulizer spray and presents more ions to the MS inlet.

N b li iNebulizing gas: pressure

*Sh th fl d

Enhanced efficiency nebulizer

*Sheath gas: flow and temperature

*Nozzle voltageDrying gas:flow and temperature

Resistive samplingcapillary exit: FragV

Resistive samplingcapillary entrance: Capillary V

Fall 20095MH User workshop QQQ fall2009.pptx

*New parameters unique to AJT source

Page 6: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Source optimization for Agilent LC/MS systemsESI, APCI, MM and Agilent Jetstream Technology (AJT), , g gy ( )

Flow dependent parameters:Compound dependent parameters:

•Nebulizer pressure

•Drying gas temperature and flow

•Capillary voltage

•Fragmentor voltage

•Vaporizer temperature (MM)

•Sheath gas flow (AJT)

•Collision Energy (QQQ, QTOF)

•Vaporizer temp (APCI, MM, AJT)

S ( )•Sheath gas temperature (AJT)

•Nozzle voltage (AJT)

Fall 2009MH User workshop QQQ fall2009.pptx

6

Page 7: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Starting parameters for Agilent sources(small molecules)(small molecules)

Parameter ESI source APCI sourceNebulizer pressure 25 psi 0 2 mL/min 60 psiNebulizer pressure 25 psi 0.2 mL/min

50 psi 1 mL/min60 psi

Drying gas flow 10 Lpm < 0.2 mL/min12 Lpm > 0.4 mL/min

5 Lpm12 Lpm 0.4 mL/min

Drying gas temp 325-350°C 350°CCapillary voltage 3500V 3500VF t lt 70 100 V Si l d 70 100 V Si l dFragmentor voltagecompound dependent

70-100 V Single quads120-140V QQQ, TOF140-170V 6230, 6530

70-100 V Single quads120-140V QQQ, TOF140-170V 6230, 6530

Fall 2009MH User workshop QQQ fall2009.pptx

7

Page 8: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Starting parameters for Agilent sources(small molecules)(small molecules)

Parameter Multimode source Agilent JetstreamNebulizer pressure 60 psi ESI 45 psiNebulizer pressure 60 psi ESI

20 psi APCI40 psi Mixed mode

45 psi

Drying gas flow 5 Lpm 7 LpmDrying gas flow 5 Lpm 7 LpmDrying gas temp 250°C ESI, Mixed

300°C APCI300°C

Capillary voltage 2000V 3500VCapillary voltage 2000V 3500VCharging voltage (MM) 2000VNozzle voltage (AJT) 500VFragmentor voltageCompound dependent

70-100 V Single quads120-140V QQQ, TOF140-170V 6230, 6530

70-100 V Single quads120-140V QQQ, TOF140-170V 6230, 6530

Fall 2009MH User workshop QQQ fall2009.pptx

8

Page 9: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Tuning and Calibration for Agilent QQQ systems•Agilent QQQ systems are very stable and require infrequent Autotuning (if notAgilent QQQ systems are very stable and require infrequent Autotuning (if not, place a service call!).

•Current software uses "ESI-Low" tune mix for all QQQs.

•Routine use only requires resolution and mass axis verification for polarity and•Routine use only requires resolution and mass axis verification for polarity and resolution modes to be used.

• Run performance check sample (lab SOP)• Do Checktune (checks all three resolution modes)Do Checktune (checks all three resolution modes)• In Manual Tune, turn on calibrant and observe MS1 and MS2 profiles. Use Adjust

Gain and Offset button if width or mass axis requires adjustment.•If using negative ion, Autotuning should be done monthly or less to minimize source exposure to TFAsource exposure to TFA.

• Flush source extensively with LC flow after 30-minute Autotune before running samples in negative ion.

• Autotune at least quarterly to optimize ion transmission and update EM utotu e at east qua te y to opt e o t a s ss o a d updatevoltage.

MH Users Meeting Morning Session Fall2009.pptx

Fall 2009

Page 10: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Optimizing the AJT sourcebeginning with recommended starting values ( )beginning with recommended starting values ( )

Order of effect on sensitivity Things to note

Sheath gas temperature (250°C)

Sheath gas flow (8 Lpm)

Requires time to stabilize

Generally 10-12 Lpm for 0.2-0.7mL

Nebulizer pressure (45 psi)

Capillary voltage (3500V)

LC flow- and mobile phase- dependent

Somewhat MW dependentCapillary voltage (3500V)

Nozzle voltage (0V)

Somewhat MW dependent

Very compound- and polarity-dependent

Drying gas temperature (250°C)

Drying gas flow (5 Lpm)

Interaction with flow and sheath parms

Generally 5-7 Lpm, higher when in doubt

Fall 2009MH User workshop QQQ fall2009.pptx

10

Page 11: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Optimizing the AJT sourceTypical values ( ) and IncrementsTypical values ( ) and Increments

Order of effect on sensitivity Increments and ranges to test

Sheath gas temperature (325°C)

Sheath gas flow (10 Lpm)

50°C, 250-400°C

2 Lpm, 8 -12 Lpm

Nebulizer pressure (40 psi)

Capillary voltage (3500V)

5 psi, 25 – 50 psi

500V 2500-4500VCapillary voltage (3500V)

Nozzle voltage (0V)

500V, 2500-4500V

500V, 0 - 2000V

Drying gas temperature (300°C)

Drying gas flow (7 Lpm)

50°C, 250-350°C

2 Lpm, 5-11 Lpm

Fall 2009MH User workshop QQQ fall2009.pptx

11

Page 12: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Optimization techniques

•Fragmentor V and Collision Energy (for MS/MS systems) should be optimized first using recommended starting parameters for source.•Voltages, gas flows change quickly and can be optimized with series of flow injections using method with injector program andseries of flow injections using method with injector program and time segments.

•Temperatures (sheath gas, drying gas, vaporizers) require p ( g y g g p ) qequilibration time between injections; perhaps best done with final chromatography conditions.• Template flow injection methods also come with Optimizer• Template flow injection methods also come with Optimizer software

Fall 2009MH User workshop QQQ fall2009.pptx

12

Page 13: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Injector program for series of flow injectionsInjector program steps:Injector program steps:Valve to BypassRemote StartpulseRepeat n times [n = number of time segments]Draw default amount from sample

Check to use injector program

pValve to MainpassWait 0.25 min [adjust to sync with acq time segments]End repeat

Example Injector p jProgram with 2-min spacing and explicit repeats

Fall 2009MH User workshop QQQ fall2009.pptx

13

Note: Inject and Remote Startpulse commands work differently on QQQ and TOF/QTOF.

Page 14: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Injector program for series of flow injections Synchronize with MS time segmentsSynchronize with MS time segments

Fall 2009MH User workshop QQQ fall2009.pptx

14

Page 15: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Fragmentor and Collision Energy optimizationusing flow injection with injector programusing flow injection with injector programMethylmalonic acid, dibutyl ester

Maximize MH+ ion transmission, minimize CID with Fragmentor voltageMaximize product ion signal(s) with Collision EnergyMaximize product ion signal(s) with Collision EnergyCan fine tune each with smaller steps after initial experiments

Fragmentor 60-160V 20V steps

Collision Energy 5-30V 5V steps

Fall 2009MH User workshop QQQ fall2009.pptx

15

Page 16: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Fragmentor and Collision Energy optimizationusing flow injection with injector programusing flow injection with injector programAndrostenedione

Maximize MH+ ion transmission, minimize CID with Fragmentor voltageMaximize product ion signal(s) with Collision Energy

Fragmentor 100-260V

Maximize product ion signal(s) with Collision EnergyCan fine tune each with smaller steps after initial experiments

ag e o 00 6020V steps

Collision Energy 5-45V 5V steps5V steps

Fall 2009MH User workshop QQQ fall2009.pptx

16

Page 17: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Fragmentor and Collision Energy Determination with Optimizer softwarewith Optimizer software•Optional add-on to MassHunter QQQ acquisition software

Current version: B 02 01 (requires B 02 01 Acquisition)•Current version: B.02.01 (requires B.02.01 Acquisition)

•Useful for frequent method development [new compound each week] or for multi-compound methods [e g toxicologyweek] or for multi-compound methods [e.g. toxicology, environmental applications]

•Many toxicology and pesticide transitions are available in newMany toxicology and pesticide transitions are available in new Agilent application kits

•Can use Flow Injection before LC separation is developedj p p

•Can use LC method to optimize multiple compounds in a few injections

Fall 2009MH User workshop QQQ fall2009.pptx

17

Page 18: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter Optimizer user interface: Compound list - can import from Excel-created csv filep p[screen captures courtesy of Ken Peck, Texas Veterinary Diagnostic Lab]

Optimizer will calculate precursor ion m/z from molecular formulaeOptimizer will calculate precursor ion m/z from molecular formulaeCan assign compounds to Groups for later browsing

Fall 2009MH User workshop QQQ fall2009.pptx

18

Page 19: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter Optimizer – precursor ion selectionUser-selectable adducts (H, Na, K, OAc- , HCOO-, etc.)Can run both positive and negative ion methods in same Project

Fall 2009MH User workshop QQQ fall2009.pptx

19

Page 20: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Optimizer – product ion selection and rejection(set mass range, minimum abundance)(set mass range, minimum abundance)

Fall 2009MH User workshop QQQ fall2009.pptx

20

Page 21: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Optimizer – method type, parameter step selection

Fall 2009MH User workshop QQQ fall2009.pptx

21

Page 22: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Optimizer resultsAutomatically added to MRM databaseAutomatically added to MRM database

Fall 2009MH User workshop QQQ fall2009.pptx

22

Page 23: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Optimizer – interactive parameter selection

Fall 2009MH User workshop QQQ fall2009.pptx

23

Page 24: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter OptimizerResults for the project can be viewed immediately after run:p j yprecursor and product ions, Fragmentor and Collision Energies

Fall 200924MH User workshop QQQ fall2009.pptx

Page 25: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

OptimizerBrowse compounds in Acquisition for importBrowse compounds in Acquisition for importCompounds can be

filtered by project, group polarity or date

Compounds can be searched for by name or formulagroup, polarity or date name or formula

Fall 200925MH User workshop QQQ fall2009.pptx

Page 26: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter OptimizerBuild LC/MS/MS method from compound databaseBuild LC/MS/MS method from compound database

• Select compounds in database after filtering• Select transitions for each compound (all or individual)• Easily import into template LC/MS method• Easily import into template LC/MS method

Fall 200926MH User workshop QQQ fall2009.pptx

Page 27: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Dynamic MRM

Fall 200927MH User workshop QQQ fall2009.pptx

Page 28: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Optimization of MRM Acquisition

•Good quantitation requires adequate number of data points across each peak (ideally 10-20)p ( y )

•High confidence or regulated identification requires > 1 MRM per compound.

•Monitoring many compounds simultaneously lowers dwell time per MRM.

•Therefore for best sensitivity, only monitor compounds in retention time window where they elute.

•Traditional approach of time segments has limitations and is tedious to setup up and maintain.

Fall 2009MH User workshop QQQ fall2009.pptx

28

Page 29: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Peak Capacities are very high with UHPLC 40 MRM Transitions in 1 minute time window40 MRM Transitions in 1 minute time window

Page 30 Fall 2009MH User workshop QQQ fall2009.pptx

Page 30: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

The solution: MassHunter Dynamic MRMIncluded in QQQ Acquisition B 02 01Included in QQQ Acquisition B.02.01

For applications requiring quantitation of 100 – 1000 compounds in one run; some examples:co pou ds o e u ; so e e a p es• Food and environmental analysis (e.g. pesticides)• Targeted quantitation of proteins via peptides (proteomics)

WITHOUT Dynamic MRM:• Need to manually set up multiple time segments to maximize dwell times

T di bl i if h i i i• Tedious to set up; problematic if changes in retention times

WITH Dynamic MRM:A t ti t f l i ti t ith t i t ti• Automatic setup of overlapping time segments without user intervention

• Fewer MRMs per unit time results in longer dwell time => incr sensitivity• Unaffected by minor chromatographic time shiftsy g p

Page 31: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Dynamic MRM for 6400 Series Triple Quads monitors transitions only when compounds elute y p

Cycle Time:Concurrent MRMsDwell Time:

1. # Concurrent MRMs fewer than with time segments → more data points across each peak2 All l d ll ti b tt iti it S/N

Page 34

2. Allows longer dwell times → better sensitivity, S/N

Fall 2009MH User workshop QQQ fall2009.pptx

Page 32: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Acquisition setup with Dynamic MRM:No time segments, insteadg ,

Retention Time and "Delta Ret Time" [MRM time range]

Fall 2009MH User workshop QQQ fall2009.pptx

35

Page 33: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

8 Minute Dynamic MRM Analysis with 6460 QQQ- 250-compound Pesticide Screen.250 compound Pesticide Screen.

2x10

8.4

8.6

8.8

9

9.2

9.4

9.6 1 1

500ppt, dynamic MRMRetention Time Window = 12 sec

6.4

6.6

6.8

7

7.2

7.4

7.6

7.8

8

8.2

Peak widths ~ 1 second

4.4

4.6

4.8

5

5.2

5.4

5.6

5.8

6

6.2

Zorbax Eclipse Plus C182.1 x 100mm (1.8μm)

2 4

2.6

2.8

3

3.2

3.4

3.6

3.8

4

4.2

4.4 ( μ )

0 4

0.6

0.8

1

1.2

1.4

1.6

1.8

2

2.2

2.4

3.193427

0.4

Counts vs. Acquisition Time (min)0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4 5.6 5.8 6 6.2 6.4 6.6 6.8 7 7.2 7.4

Fall 200939MH User workshop QQQ fall2009.pptx

Page 34: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Summary – Optimization of QQQ Acquisition

•Source optimization: flow and compound dependent parameters

Agilent Jetstream Technology source: increased sensitivity vs•Agilent Jetstream Technology source: increased sensitivity vsregular ESI source if parameters optimized correctly:

– Sheath gas temperature and flow are most importantSheath gas temperature and flow are most important– Other parameters have less effect on response– Need less drying gas flow and temperature, but keep capillary clean

•Dynamic MRM is available for all 6400 models– More data points across peak– Longer dwell times for better sensitivity, S/N– Easier to set up and maintain than time segment methods

Fall 2009MH User workshop QQQ fall2009.pptx

40

Page 35: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter Quantitative SoftwareReview and Quant Method OptimizationReview and Quant Method OptimizationTopics • Method setup from acquired data• Method setup from acquired data• Method re-use and updating• Peak detection optimization• Peak identification troubleshooting• What's new in MH Quant B.03.02

Fall 2009MH User workshop QQQ fall2009.pptx

41

Page 36: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Important MassHunter Quant concepts and rules:BatchesBatches•A Batch is a file which contains all the Quant results from a set of data files AND the Quant method used. Very convenient for ybackup and moving data around.

•All the data files in a Batch must reside in a single directory, so put them all together before creating New Batch.

•Select the data directory BEFORE naming the new Batch

•Using the Browse to Copy Samples button when creating the Batch can be dangerous: you will have two copies of the same files (one with Quant results and one without)!files (one with Quant results and one without)!

Fall 2009MH User workshop QQQ fall2009.pptx

42

Page 37: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Important MassHunter Quant concepts and rules:Quant methodsQuant methods•THE most common method problem: Calibration/QC level names in Batch and Quant method do not match, e.g.

– Batch: levels are 1,2,…5, QC-Lo, QC-Hi– Method: levels are L1, L2…L5, QC-Low, QC-High

2nd most common problem: a hidden column in the Method with•2nd most common problem: a hidden column in the Method with a key parameter, e.g. Criteria in MRM Compound Setup for peak selection, or Ion Polarity if method created manually.•Many Quant parameters can be copied between compounds with Apply to All button.

Wh i d bt Ri ht Cli k t l k f i t f t d•When in doubt, Right-Click to look for convenient features and shortcuts, like Fill Down or Fill Column.

Fall 2009MH User workshop QQQ fall2009.pptx

43

Page 38: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter Quant –optimizing target compound identificationoptimizing target compound identificationMake the Criteria column visible in MRM Compound Setup:

Using the default of Greatest Response may result in the wrong peak being chosen as the target compound:

Endogenous interference withwrong retention time incorrectly

identified as tramadolidentified as tramadol(no qualifier)

Fall 2009MH User workshop QQQ fall2009.pptx

44

Page 39: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter Quant –optimizing target compound identificationoptimizing target compound identificationChange the Criteria to be more specific:

This might still not be selective enough, so specify how close Now the offending peak with the

wrong RT is ignored:the retention time must match the calibrated RT with Non-reference Window parameter in Gl b l

wrong RT is ignored:

Globals:

Fall 2009MH User workshop QQQ fall2009.pptx

45

Page 40: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter Quant –optimizing target compound identificationoptimizing target compound identificationQualifying the Qualifier:For qualifiers, you can ensure that the qualifier and the quant ion have the same retention time (i e line up properly as they should if coming from thesame retention time (i.e. line up properly as they should if coming from the same chromatographic peak), by setting the Correlation Window in Globals correctly. The default of 2 minutes isWAY too largeWAY too large.

The Ignore Peaks Not Found checkbox will prevent reporting

Fall 2009MH User workshop QQQ fall2009.pptx

46

"Amount=0.00" for target compounds not present with your criteria.

Page 41: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter Quant –Setting a reporting threshold:Setting a reporting threshold:You can also make Quant ignore peaks that meet these criteria but are below your desired detection or reporting threshold, or that really are background (if you can’t use Qualifiers) Set a Peak Filter (area reject)background (if you can t use Qualifiers). Set a Peak Filter (area reject) value in Advanced Tasks …Integration Parameters.

You can have the same Peak Filter for all compounds with Apply to All, or specific values for each compound. You can also limit integration to the N largest peak(s), like only one in this

Fall 2009MH User workshop QQQ fall2009.pptx

47

p g g p ( ), yexample.

Page 42: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

MassHunter Quantitation Integrators:

General UniversalMS/MS requires 64 data points within windowMS/MS – requires 64 data points within window

Fall 2009MH User workshop QQQ fall2009.pptx

48

Page 43: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

What's new in MassHunter Quant B.03.02(just type 'What's new' in the Online Help!)(just type What s new in the Online Help!)

Fall 2009MH User workshop QQQ fall2009.pptx

49

Page 44: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

What's new in MassHunter Quant B.03.02Something for everybodySomething for everybody

Fall 2009MH User workshop QQQ fall2009.pptx

50

Page 45: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

What's new in MassHunter Quant B.03.02More legibly:More legibly:Software Enhancements (some are instrument-specific)

•Approximately 10 times faster batch analysis processingApproximately 10 times faster batch analysis processing•Compound Math ! •Metrics plot (e.g. for IS areas)•Support for QQQ dynamic MRM data•More integrator choices: “Universal" (ChemStation) and GC/MS/MS integratorsMore integrator choices: Universal (ChemStation) and GC/MS/MS integrators•Fixed graphics to lowest calibration level•Signal to noise setup•Continuing calibration•Dynamic background subtractionDynamic background subtraction•Standard addition calibration•Compound library setup and searching•Peak purity and deconvolution•More choices in the Generate Reports dialog (e g no graphics)More choices in the Generate Reports dialog (e.g. no graphics)

•Direct Link to Getting Started Videos and Familiarization Guide

Fall 2009MH User workshop QQQ fall2009.pptx

51

Page 46: Agilent 6400 Series QQQ Method Development Triple ......Agilent 6400 Series Triple Quadrupole QQQ Method Development Triple Quadrupole and Optimization LC/MS/MS Users Session and Optimization

Final recommendation for learning MassHunter Quant:Quant:

Uses data filesfiles included with the software

Fall 2009MH User workshop QQQ fall2009.pptx

52