aerodynamics report

Upload: edk5035

Post on 05-Apr-2018

233 views

Category:

Documents


2 download

TRANSCRIPT

  • 8/2/2019 Aerodynamics Report

    1/14

    AERSP 305W

    Aerospace Technology Laboratory

    Laboratory Section 05

    Laboratory Experiment Number 2Hot-Wire Calibration and Validation

    March 28, 2012Performed in Room 8 Hammond Building

    Eric Kachel

    Lab Partners Names:Marcia CastilloToby FabisiakHsiaoting Ko

    Kimberly Norvina

    Lab TA: Kylie Flickinger

    Course Instructor: Richard Auhl

  • 8/2/2019 Aerodynamics Report

    2/14

    Abstract

    The objective of this lab was to calibrate and confirm the measurement capabilities of a hot

    wire anemometer. A secondary objective was to compare the hot-wires performance to readings from

    the Pitot-static probe. This information is needed for future experiments so it is known whether it is

    more appropriate to use a hot-wire anemometer or the Pitot-static probe. After a fourth order

    polynomial was used to calibrate the hot-wire it was used to try and detect turbulence in expected

    regions, and detect shedding vortices. Through this experiment it was found that the hot-wire

    anemometer performed well in detecting rapid fluctuations in the flow velocity. Although the Pitot-

    static probe works well for constant velocities it was unable to detect the rapid changes in the shear

    flow regime.

  • 8/2/2019 Aerodynamics Report

    3/14

    Introduction

    The objective of this lab was to calibrate and confirm the measurement capabilities of a hot

    wire anemometer. It is important to understand the capabilities of the hot-wire so that it can be used in

    later experiments.

    In order to calibrate the hot-wire the velocity of the flow needed to be determined. In this

    experiment this measurement was performed using the Pitot-static probe. The Pitot-static tube

    measured dynamic pressure (q), using this pressure, and the density of the air in the lab ( ), the

    velocity of the flow could be determined using Equation 1 below.

    )(2 qV (1)

    Using this velocity a fourth order polynomial can be used to calibrate the hot-wire. Once the

    hot-wire is calibrated its capabilities can be further examined. The first measure of hot-wire

    performance was through the measurement of turbulence intensity. In order to do this we look at the

    instantaneous velocity equation, Equation 2 below. Here this equation states that the instantaneous

    deviation of the velocity is equal to the total velocity minus the average velocity.

    ii uUu or Uuu ii (2)

    Figure 1, in the Appendix, is an example of the type of data seen later in the experiment. The

    components of the instantaneous velocity equation have been labeled on the figure to help enhance

    clarity of what each of these terms mean. By taking how much the velocity is fluctuating from the

    mean (u ) and taking the root mean square of that value the amount of fluctuation in the system can be

    quantified. If the root mean square velocity is then divided by the average velocity turbulence intensity

    is found. Notice how the v and w terms were neglected. This is because flow from the get is in only

    one direction and the instruments can generally only accurately measure velocity in one direction. This

    allows the unidirectional assumption to be made. Equation 3 below is for the turbulence intensity

    calculation.

    U

    u

    U

    u

    U

    wvu

    Ti rms

    2

    222

    3

    )(

    (3)

    Another way to validate the use of the hot-wire is through measuring the Karman vortex street.

    The Karman vortex street is when flow passes by a cylinder at a particular Reynolds number (80-200)

    it sheds off periodic vortices at a set frequency. By using the hot-wire to try and measure the frequency

  • 8/2/2019 Aerodynamics Report

    4/14

    of these vortices we can gain a better understanding of its capabilities. Figure 2 shows a flow past a

    cylinder at varying Reynolds numbers. The Karman vortex street occurs between eighty and two-

    hundred.

    The non-dimensional parameter that describes vortex shedding is the Strouhald number. The

    Strouhal number relates the frequency of the vortex shedding to the diameter of the cylinder, and to the

    flow velocity. The Strouhal number can also be approximated using the Reynolds number. Both the

    equation for exact and approximate Strouhal number can be seen below in Equation Set 4.

    U

    DfSt

    and

    Re

    St7.19

    1198.0 (4)

    Figure 2: Flow Past a Cylinder for Varying Reynolds Numbers

  • 8/2/2019 Aerodynamics Report

    5/14

    Experimental Procedure

    The experimental configuration for consisted of mainly the jet and the probes (Pitot-static and

    hot-wire). These probes were located apart from one another by 0.2 to avoid interference between the

    probes themselves. The instrumentation was also placed on a multi-axis track so that its position could

    be managed precisely. Figure 3 shows the experimental jet, and how the instruments are oriented.

    For the data processing the Pitot-static tube voltage was sent to Channel 0 in the data

    acquisition system (DAQ), and then sent straight to LabView where it could be processed. The hot-

    wire voltage took a more complicated path. The voltage was first sent to the digital readout so it could

    be viewed real time to help locate the shear region of the flow. Then the unfiltered portion of the

    hotwire data entered Channel 1 of the DAQ. The hot-wire data was also sent through an oscilloscope,

    and high pass filter (sampling at 20,000 Hz) before returning to the DAQ through Channel 2. The hot-

    wire data was also then accessible through LabView. Figure 4 shows a diagram of this process.

    HOT-WIREANEMOMETER

    HIGH PASSFILTER

    DATA ACQUISITIONSYSTEM

    LABVIEW

    DIGITALREADOUT

    OSCILLOSCOPE

    PITOT-STATICPROBE

    CH 1

    CH 0 CH 2

    Figure 4: Block Diagram of Data Processing

    Hot-WirePitot-StaticProbeFlow

    Direction

    0.2 Offset

    Figure 3: Jet Calibration Facility

  • 8/2/2019 Aerodynamics Report

    6/14

    Hot-Wire Calibration

    In order to use the hot-wire anemometer in later portions of the experiment it first needed to be

    calibrated. While both the Pitot-static and hot-wire probes were 1 away from the jets exit nozzle the

    jet was turned to multiple dial settings. Using the known calibration factor of 5.27 volts/psf for the

    Pitot-static probes transducer the dynamic pressure was found at each dial setting. Using the dynamicpressure from the Pitot-static probe and the atmospheric data collected in the lab the velocity data was

    able to be determined. This velocity was then used to calibrate the hot-wire anemometer.

    Turbulence Intensity

    Once the turbulence hot-wire is calibrated the next part of the experiment was measuring the

    turbulence intensity of different regions in the flow. For this portion of the experiment the dial setting

    for jet velocity was set at a constant 77%. First the hot-wire was placed in the core (center) of the jets

    flow field. Here 2000 samples of velocity were taken. Then the probe was moved toward the edge of

    the jets flow field in the shear region. The location of this region was found by observing the hot-wire

    voltage displayed on the oscilloscope and adjusting the position of the probe accordingly. These same

    steps were also performed with the Pitot-static tube.

    Karman Vortex Street

    In order to examine the frequency of shedding vortices this experiment studied a 0.032 and a

    0.062 cylinder. Only the hot-wire was used for this portion of the experiment. As with the hot-wire

    calibration a variety of jet velocity settings were used in order to capture the shedding characteristics.

    In order to have the best chance of finding the vortices the hot-wire was originally placed five diameter

    lengths of the cylinder downstream, and two diameters below the cylinders center line. A diagram

    showing this set-up can be seen below in Figure 5.

    D

    5 D

    2 D

    Hot-Wire

    Flow

    Figure 5: Hot-Wire Location for Measuring Vortex Shedding

  • 8/2/2019 Aerodynamics Report

    7/14

    A spectrum analyzer was then used to examine the shedding frequency. If there was not a

    strong response where the hot-wire was located the position was altered until the spectrum showed a

    distinct peak, which was then recorded.

  • 8/2/2019 Aerodynamics Report

    8/14

    Results and Discussion

    In order to perform the necessary calculations atmospheric data in the lab was measured before

    the experiment. The temperature and pressure of the air were measured directly. From these values the

    air density and dynamic viscosity were also calculated. Since the jet is an open system, unlike the

    closed wind tunnel, there was no temperature variation through this experiment. The atmospheric

    values are shown below in Table 1.

    Pressure (lb/ft2) Temperature (R) Density (slug/ft3) Dynamic Viscosity slug/(ft*s)

    2059.55 459.67 0.002262 3.8457E-07

    Once the atmospheric conditions in the lab were known the hot-wire anemometer could be

    calibrated. In order to do this the velocity of the jet must first be determined. The Pitot-static probe was

    used to determine the velocity of the flow. The conversion factor for the volts to dynamic pressure in

    psf was given to be 5.27 (volts/psf). This factor was applied at the various dial settings of the jet to

    convert the voltage reading to useful dynamic pressure data. Figure 6 shows the pressure transducer

    calibration.

    Once the voltage from the Pitot-static probe was converted in to dynamic pressure the velocity of

    the flow could be determined. This was done by also using the air density calculated from the

    atmospheric conditions in the lab. The velocity and voltage from the hot-wire anemometer were

    plotted together in Figure 7. A fourth order polynomial was then fitted to the plot and served as the

    conversion factor from volts to psf for the experiments hot-wire readings. This polynomial can also be

    found in Figure 7.

    Table 1: Laboratory Atmospheric Conditions

    Figure 6: Transducer Calibration Data

  • 8/2/2019 Aerodynamics Report

    9/14

    The limiting factor for the hot-wire calibration was the fact that the jet could not operate at

    lower velocities. The lowest data collected was at the 24 percent dial setting. The offset point at zero

    velocity could also be recorded to be 3.4 volts when there was no flow over the wire. By examining the

    fourth order trend line it can be seen that it matches the collected data points very closely. The R

    2

    valuefor the equation is 0.9999. If the trend line were to match the data exactly the R 2 value would equal

    one. This shows that using this equation for converting the hot-wires voltages to pressure will be

    accurate.

    The next part of the experiment was using the calibrated Pitot-static probe, and hot wire to examine

    the turbulence intensity of the flow. Measurements were taken in the core, and shear regions of the

    flow with both instruments as discussed in greater detail in the experimental procedure. During this

    section of the experiment the velocity was recorded over a time sample. From this time sample the

    mean velocity, and turbulence intensity were calculated. The time trace along with mean velocity and

    turbulence intensity can be seen in Figure 8 for the hot-wire anemometer.

    Figure 7: Hot-wire Calibration Data

  • 8/2/2019 Aerodynamics Report

    10/14

    By examine a smaller time slice of this data it may be easier to examine some of the features. A

    smaller time slice of the data represented in Figure 9 can be found in the Appendix in Figure 1. Figure

    1 shows how the velocity for both the core and shear values fluctuates about the average. In the shear

    region this fluctuation appears to be much greater. This can be represented quantitatively by

    calculating the turbulence intensity also provided for both regions in Figure 9. As expected the

    turbulence intensity is greater in the shear region where turbulent features such as eddies cause the

    large fluctuations in instantaneous velocity.

    The turbulence intensity in both the core and the shear region was also examined by using the

    Pitot-static probe. Similarly to the hot-wire method the data recorded from the Pitot-static tube can be

    seen below in Figure 10.

    Figure 8: Hot-wire Turbulence Analysis

  • 8/2/2019 Aerodynamics Report

    11/14

    From Figure 10 it can be seen that the turbulence intensity is very low for the Pitot-static probe.

    From the hot-wire analysis we know there is turbulence in the shear region that the Pitot-static probe is

    unable to record. The sampling mechanism is too low for this device to pick up the rapid fluctuations

    in velocity associated with turbulent flow characteristics. Using this information we know that in

    future experiments a Pitot-static tube will not be a good instrument to measure turbulence. Figure 11 in

    the Appendix shows a time slice of Figure 10. When viewed alongside Figure 1 showing the

    turbulence intensity from the hot-wire, the lack of fluctuation can be observed in the Pitot-static data.

    After the turbulence intensity was determined the Karman vortex street was also observed. The

    procedure for this was discussed earlier. For a 25 micron wire at maximum velocity the Reynolds

    number will be 87. The vortex street occurs when the Reynolds number is between 80 and 200, so it is

    likely that there will be vortex shedding. Since the sample rate of the hot-wire anemometer was 20,000

    Hz using the Nyquist sampling criteria we can only detect frequencies up to 10,000 Hz. Using the

    theoretical formula presented earlier, and the maximum Reynolds number for the experiment it was

    found that the smallest diameter cylinder that shedding could be detected from had a

    Figure 10: Pitot-static Probe Turbulence Analysis

  • 8/2/2019 Aerodynamics Report

    12/14

    0.000365diameter. This corresponds with a 9.271 micron cylinder. Therefore if the 25 micron hot-

    wire probe does produce a vortex it is likely to be detected.

    For the 0.0032 and 0.0062 diameter cylinders the Reynolds number and Strouhal number were

    calculated at each dial setting. This plot was again limited by the minimum velocity of the jet. The

    theoretical equation discussed earlier is shown along with the experimental data for both cylinders inFigure 12.

    The achievable Reynolds number was limited to be approximately 200 for this experiment. Figure

    12 helps to show that the experimental data followed the theoretical calculation closely.

    Figure 12: Vortex Shedding Analysis

  • 8/2/2019 Aerodynamics Report

    13/14

    Conclusions

    Overall this experiment yielded very useful results. The first portion of the experiment was

    calibrating the hot-wire anemometer. This was done by adding a fourth order trend line to the plot of

    velocity vs. hot-wire voltage. This fourth order polynomial was then used throughout the rest of the

    experiments in order to convert the voltage readings to velocity. Because of this it is important that the

    equation used closely represents the data. Calculating the R2 value of the trend line to the data gives a

    quantitative number of how well the trend line fits that data. If R 2 equals one the equation matches

    exactly. For the fourth order conversion equation the value was 0.9999. Since this value is extremely

    close to 1 the calibration used was a very accurate one.

    Ultimately the goal of calibrating the hot-wire was that we could confirm that it was working so

    that it could be used in further experiments. To do this the hot-wire was tested under several well

    documented flow conditions. By placing the anemometer in the core of the jet flow the instrumentsability to measure a constant velocity was examined. Another place the hot wire tested was in the shear

    region of the jets flow field. Here there will be turbulent eddies and the velocity at a particular set

    point would be expected to fluctuate. The hot-wire detected these turbulent properties. Lastly the hot-

    wire was used in order to measure the Karman vortex street. By using the Reynolds number to

    theoretically estimate the Strouhal number the frequency of the vortex shedding can be determined. By

    examining the data from this test it was seen that the hot-wire detected flow properties similar to the

    theoretical data. The fact that the hot-wire detected all of the expected flow characteristics shows that it

    can be accurately used to determine flow properties other than just pure constant velocity.

    Data was also collected using the Pitot-static probe for the core and shear regions. Although the

    Pitot-static tube performed was at the constant velocity setting at the core, it was unable to detect much

    turbulence intensity in the shear region. This is because the velocity was changing so rapidly that the

    transducer didnt have time to record the difference in pressures. So when designing an experiment if

    the turbulence of the flow is of interest the hot-wire should be used. However, if turbulence is not a

    parameter in the experiment a Pitot-tube can be used in order to reduce noise.

    One of the improvements that could be made to this experiment is with the jet itself. The

    current jet can only operate above twenty four percent power. It may be useful to try and make the

    flow through the jet slower. This way more data points could be taken to better characterize the results.

    Also a lower velocity may allow for the use of smoke to visualize the Karman vortex street. If the

    frequency shedding was visible the frequency could be observed visually and then compared to the

    spectrum frequency data to better enhance confidence in the results.

  • 8/2/2019 Aerodynamics Report

    14/14

    The Appendix

    iu

    iu

    U

    Figure 1: Hot-wire Turbulence Analysis Time Slice

    Figure 11: Pitot-Static Turbulence Analysis Time Slice