advantages of polymer- based hplc columns – an … denko europe gmbh 1 advantages of polymer-based...
TRANSCRIPT
-
Showa Denko Europe GmbH 1
Advantages of polymer-based HPLC columns
an alternative for ODS
Yukiko Higai
Shodex, Showa Denko Europe GmbH
-
Showa Denko Europe GmbH
Content
(1) Comparison between Silica and Polymer-based
RP columns
(2) Characteristics of Polymer RP columns
- an example: a new column type
(3) Characteristics of Polymer HILIC columns
2
-
Showa Denko Europe GmbH
HPLC
60-70% of samples
can be applied
Reversed Phase (RP)
(mainly ODS)
Size ExclusionIon Exchange
(Ion Chromatography)Normal PhaseLigand Exchangeetc.
Use of HPLC in general
3
-
Showa Denko Europe GmbH 4
Silica-based exODS Polymer-based exODP
Higher resolution Relatively lower resolution
Reasonable price Relatively higher price
Narrow pH range pH2-7 Wide pH range pH2-12
Impossible to use underthe low salt concentration
Possible to use underthe low salt concentration
Short life time Long life time
Impossible to injectlarge concentration sample
Possible to inject large concentration sample
Not enough separation of hydrophilic substances
Good separation ofhydrophilic substances
Need ion-pair reagent to separate basic substances
Possible to separate basic substances without any reagent
(1) Comparison between Silica and Polymer-based RP
-
Showa Denko Europe GmbH
Problems of Silica-based RP columns
O
O
O
Si OH
Si OH
O
Si OH
O
O
OHSi
O
OHSi
Si O Si
Si O Si
OSi Si
Residual silanol capping
difficult to remove ionic adsorption
Lot-to-lot differences
effect of metal coodination, difficult to control residual silanols
Low durability against solvents
Si
O
Si
OHSi
OH
Si
HO
O Si C18
OH
OH
5
-
Showa Denko Europe GmbH
Sample suitable for ODS analysis
Suitable for ODS Unsuitable for ODS
Molecularsize
LowHigh also sample including
high MW impurities
PolarityCompounds
with different polarityCompounds
with similar polarity
Low to Moderate High or Ionic
Solution Acidic or Neutral Basic
6
other choice Polymer
-
Showa Denko Europe GmbH 7
(2) Characteristics of polymer-based RP columns
a) Separation Almost the same
b) Analysis of basic sample Good
c) pH range in eluent Wide
d) Effect of salt concentration in eluent Low
e) Reproducibility Good
X=OR, COOR, etc
Si
O
Si
O
O
O
H
SiC
CH
X
-
Showa Denko Europe GmbH 8
a) Separation performance (1)
Polymer-based Silica-based
Eluent : CH3CN / H2O = 50 / 50
Flow rate : 1.0mL/min
Detector : UV (262nm)
Column temp. : 40oC
Almost the same separation
Sample : 5L1. 0.2% Methylethyl ketone
2. 0.2% Diethyl ketone
3. 0.4% Di-n-propyl ketone
4. 0.6% Di-n-butyl ketone
-
Showa Denko Europe GmbH
a) Separation performance (2)
Column : Shodex RSpak DE-413
ODS column from other manufacturer
Eluent : 10mM H3PO4 aq.
Flow rate : 1.0mL/min
Detector : RI
Column temp. : 50 oC
Sample : 0.2% each,
5L
1. Lactitol
2. Erythritol
Better separation
of hydrophilic samples
9
-
Showa Denko Europe GmbH
Amine
compounds
R-N+
-O-SiResidual
Silanol
adsorption
Polymer-based column Silica-based column
Tends to cause tailing( ion-pair reagent needed)
Less Tailing
b) Analysis of basic compounds (1)
10
-
Showa Denko Europe GmbH
b) Analysis of basic compounds (2)
(ODS)
Eluent : 0.1% 1-Pentanesulfonic acid
sodium salt aq./CH3CN=80/20
Flow rate : 1.4mL/min
Column temp. : 60oC
ODP-50no need
ion-pair reagent
(ODP-50 4D)
Eluent : 50mM Phosphate buffer (pH10)
/CH3CN=80/20
Flow rate : 0.6mL/min
Column temp. : 30
ODSwith
ion-pair reagent
no need of ion-pair reagent
11
-
Showa Denko Europe GmbH
c) Better resolution in high pH (1)
Durability in Alkaline condition
better selectivity
higher TPN
for example
ODS
1 7 14
ODP
2 13
Available pH
Good for the separation
of basic sample under
alkaline conditions
12
-
Showa Denko Europe GmbH
pH 11pH 7
Basic sample (local anesthetic drug)
Column: Shodex Asahipak ODP-50 4DEluent: 25mM Phosphate buffer/CH3CN (60/40)
Flow Rate: 0.6mL/min
Detector: UV(254nm)
Column temp.: 30oC
pH 3
Working condition of
Polymer column!
13
c) Better resolution in high pH (2)
-
Showa Denko Europe GmbH
Sample : 1mL/mL Scopolamine50LODS-A ODS-C
Low concentration for ionic drugs
excellent for LC/MS
ODP-50
25m
M10m
M1m
M
Column : Shodex Asahipak ODP-50 4D
ODS Columns from other manufacturer
Eluent : Phosphate buffer(pH7.0)/CH3CN=60/40
Flow rate : 0.6mL/min
Detector : UV (254nm)
Column temp. : 30 oC
14
d) Effect of salt concentration in analysis of ionic drugs
-
Showa Denko Europe GmbH
e) Reproducibility
Rt
(min)
Lot A Lot B Lot C CV (%)
1 2.504 2.505 2.482 0.425
2 3.148 3.152 3.123 0.409
3 4.878 4.894 4.847 0.400
4 7.961 8.001 7.920 0.415
Peak
area
(UV x
sec)
Lot A Lot B Lot C CV (%)
1 36616 37614 36414 1.422
2 32168 33166 31934 1.648
3 48560 50090 48075 1.756
4 64814 66686 64276 1.583
Peak
area
(%)
Lot A Lot B Lot C CV (%)
1 20.102 20.055 20.151 0.195
2 17.659 17.683 17.673 0.056
3 26.658 26.706 26.605 0.155
4 35.581 35.555 35.571 0.030
Control of batch differences
15
-
Showa Denko Europe GmbH
When ODS is not suitable
Problem Solution
ODP2 HPHigh MW impurityHigh polarity (ionic)
16
- an example: a new column type
ex) direct analysis of drugs in biological fluids
-
Showa Denko Europe GmbH
Characteristics of polymer-based RP ODP2 column
proteindrug with low Mw
ODSODP2 HP
Polyhydroxymethacrylate gel Silica gel
ca. 100-120ca. 40
More adsorption of proteinsLess adsorption of proteins
Advantages:Hydrophilic surface of packing
material
Small pore size
Proteins will be removed without sample pre-treatment
less time-consuming
17
-
Showa Denko Europe GmbH 18
a) Analysis of high polarity substances
-1.00
-0.50
0.00
0.50
1.00
-1 0 1 2
log P
log
k'
ODP2 HP
ODP-50
ODS(B)ODS(A)
High
Lo
wH
igh
Ret
enti
on
Polarity Low
Column : Shodex ODP2 HP-4D (4.6mmID x 150mm)
Eluent : H2O/CH3CN=75/25
Flow rate : 1.0mL/min
Column temp. : 40oC
Strong retention of high polarity substances
-
Showa Denko Europe GmbH 19
ODP2 HP-4D Polymer-based RP column (Shodex Existing column)
3123
4
15
45
6
N: 13,200
7
5
6N: 5,700
7
1 2
0 5 10 0 5 1015 min min
Sample : 5L
1. Uracil 30mg/L2. Theobromine 75mg/L
3. Caffeine 130mg/L
4. Phenol 300mg/L
5. Methyl benzoate 350mg/L
6. Toluene 1000mg/L
7. Naphthalene 150mg/L
Column : Shodex ODP2 HP-4D(4.6mmID x 150mm)
Eluent : H2O/CH3CN=55/45Flow rate : 0.6mL/min
Detector : UV(254nm)Column temp. : 40oC
Column : Shodex Asahipak ODP-50 4D(4.6mmID x 150mm)
Eluent : H2O/CH3CN=35/65Flow rate : 0.6mL/minDetector : UV(254nm)Column temp. : 40oC
hydrophilic
Better separation for hydrophilic materials
b) Analysis of hydrophilic compound
-
Showa Denko Europe GmbH 20
90909090
100100100100
110110110110
120120120120
130130130130
140140140140
150150150150
160160160160
0000 20202020 40404040 60606060 80808080 100100100100 120120120120 140140140140
Injection number
Pre
ssu
re c
han
ge
rati
o [
%]
Column : ODP2 HP-2B (2.0mmID x 50mm), ODS(A) (2.0mmID x 50mm)Eluent : 1mM CH3COONH4 aq./CH3CN=90/10Flow rate : 0.2mL/minDetector : UV (220nm)Column temp. : 30oC
c) Removing proteins
ODP2 HP-2B
ODP2 HP-2BODS(A)
Sample : 5L
BSA 7.0mg/mL
Proteins are not adsorbed in ODP2 HP and the pressure did not increase.
No adsorption
no C18 end-capping
Hydrophilic surface
minminminmin0000 0.50.50.50.5 1.01.01.01.0 1.51.51.51.5 2.02.02.02.0
Longer column life
-
Showa Denko Europe GmbH 21
d) LC/MS
Column : ODP2 HP-2B (2.0mmID x 50mm) Eluent : 10mM CH3COONH4 aq./CH3CN=70/30Flow rate : 0.2mL/minDetector : UV (220nm), ESI-MS Column temp. : 30oC
Sample :
1. Barbital 500g/L
2. BSA 7.0mg/L
There is no ion suppression by protein.
SIMSIMSIMSIMSIM
m/zm/zm/zm/zm/z ====183.2183.2183.2183.2=183.2
SIMSIMSIMSIMSIM
m/zm/zm/zm/zm/z ====183.2183.2183.2183.2=183.2
BarbitalBarbitalBarbitalBarbitalBarbital
Peak Area : Peak Area : Peak Area : Peak Area : 100100100100%%%%Peak Area : 100%
Peak Peak Peak Peak Peak HightHightHightHightHight : : : : 100100100100%%%%: 100%
BarbitalBarbitalBarbitalBarbitalBarbital
Peak Area : Peak Area : Peak Area : Peak Area : 107107107107%%%%Peak Area : 107%
Peak Peak Peak Peak Peak HightHightHightHightHight : : : : 102102102102%%%%: 102%
BarbitalBarbital
UV (220nm)
0 1.0 2.0 3.0 4.0
Time (min)
100000
200000
300000
Inte
nsi
ty
0 1.0 2.0 3.0 4.0
Time (min)
0
100000
200000
300000
Inte
nsi
ty
0 1.0 2.0 3.0 4.0
Time (min)
100000
200000
300000
400000
500000
600000
700000
800000
900000
1000000
1100000
uA
U Inject to MS
UV (220nm)
0 1.0 2.0 3.0 4.0
Time (min)
100000
200000
300000
400000
500000
600000
700000
800000
900000
1000000
1100000
uA
U
0.120.120.120.12
Inject to MS
Blank with BSA
BSA
Sample pre-treatment
Pore size: 40SEC effect to
remove protein
Bar
bit
al r
eco
very
rat
io [
%]
ODP2 HP Blank
ODP2 HP with BSA
BarbitalPeak Area :100%
0.0 1.0 2.0 3.0 4.0
Time (min)
Inte
nsi
tyIn
ten
sity
0.0 1.0 2.0 3.0 4.0
Time (min)
0
20000
40000
60000
80000
100000
Inte
nsi
ty
0
20000
40000
60000
80000
100000
Inte
nsi
ty
0
20000
40000
60000
80000
100000
0
20000
40000
60000
80000
100000
ODS(A)Blank
ODS(A) with BSA
BarbitalPeak Area :100%
BarbitalPeak Area :71%
BarbitalPeak Area :99%
TIC
m/Z 182.5-183.5
TIC
m/Z 182.5-183.5
-
Showa Denko Europe GmbH
(3) Characteristics of Polymer-based amino column (HILIC)
a) Enables sugar analysis at room temp.
(alkali atmosphere)
b) Long life time/stability.
Regeneration possible with alkali solvents
c) Reduced noise level because of less bleeding
22
-
Showa Denko Europe GmbH 23
Comparison with an amide column at low temp.
Column : Shodex Asahipak NH2P-50 4E (4.6mmID x 250mm), Amide column from another manufacturer (4.6mmID x 250mm)
Eluent : CH3CN / H2O = 75 / 25
Flow rate : 1.0mL/min Detector : Shodex RI Column temp. :30deg-C
a) Sugar analysis at room temp.
Amino columnAlkali atmosphere due to amino groups
No anomers even at room temp.
Amide column
Not alkali atmosphere
(acryl amide groups)
Need high temp. to avoid anomers
-
Showa Denko Europe GmbH 24
Silica
-based
Polymer-based
Comparison with a silica-column in Life timeb) Long life time/stability
-
Showa Denko Europe GmbH
c) Reduced noise level less bleeding
Silica Amino column
Polymer Amino ColumnAbsolute amount 500 ng 100 times better
Noise level: 30pA
Lower limit (S/N=3)
Absolute amount 100 ng
Noise level: 0.8pA
Lower limit (S/N=3)
Absolute amount 1 ng
Reduced noise level because of less bleeding
Source: LMS CO., LTD. Japan
25
Absolute amount 500 ng
-
Showa Denko Europe GmbH 26
Column : Shodex Asahipak NH2P-50 4E (250mmLx4.6mmID) Eluent : (A) CH3CN / H2O = 50 / 50
(B) 0 to 7min CH3CN=57/43 to 50/50 (Linear gradient) 7 to 16min CH3CN=50/50 Flow rate : 1.0mL/min
Detector : ELSD Column temp. : 30oC Sample : 0.5%, 20 L
Sample : 0.5% Hydrolyzed dextran, 20 L
Poly- and oligosaccharides
Gradient mobile phase enables a better separation of hydrolyzed dextran in Mw. ELSD detectionis effective.
-
Showa Denko Europe GmbH
Summary: Polymer vs ODS
1. Separation performance similar
Good separation for hydrophilic substances
2. Better separation of basic substances
Possible to analyze trace level of basic substances
No need of ion-pair reagent
3. Alkaline solvent can be used
Wide selection of solvent for analyses
4. Low influence of salt
Better peak shape under low concentration of salt
lower ion suppression
5. Good Reproducibility
27
-
Showa Denko Europe GmbH 28
for details >>>
www.shodex.de
www.bioanalytics.co.il
Thank you for your attention!