acoustic focusing cytometry: sensitivity and...

1
Acoustic cytometry is a new technology that utilizes forces derived from acoustic radiation pressure to focus particles in flow cytometry. This technology exploits the physical differences between particles relative to the background medium, allowing cells to be tightly focused. The ability to focus cells without relying on hydrodynamic forces allows many possibilities outside the scope of conventional flow cytometry. The use of low flow rates are required for sensitive measurements with conventional hydrodynamic focusing, while high flow rates are reserved for less critical measurements. This has created a trade-off of sensitivity and throughput historically accepted with conventional flow cytometry. In contrast, acoustic focusing of particles separates the alignment of cells from a particular flow rate allowing for sensitivity at collection rates much higher than conventional cytometry. In this study we examined changes in sensitivity with sample flow rate using acoustic focusing cytometry. We performed typical assays using the Applied Biosystems ® Attune ® Acoustic Focusing Cytometer, such as DNA content cell cycle analysis, immunophenotyping, apoptosis, intracellular antibody labeling, phagocytosis, white blood cell light scatter, and microshpere analysis. Data was collected at all seven collections rates (25μl/min-1000μl/min) and results compared across collection rates for percent positive, percent coefficient of variation, signal to background ratio and/or staining index. In the DNA content cell cycle testing, conventional hydrodynamic focusing demonstrated a loss of focusing at higher sample rates resulting in higher %CVs and reduction in data quality. In comparison, consistent results and tighter %CVs were obtained across all flow rates using the Attune™ acoustic focusing cytometer. Distinct separation of positive and negative cell populations was demonstrated with the other assays, although a slight drop in staining index was noted at two flow rates. In this study we demonstrate that collection rates up to 1000μl/minute are possible without reduction in sensitivity when using acoustic focusing cytometry. For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use J06 New and Emerging Technologies for Cell Biology B634 Barbara Seredick, Chris Langsdorf, April Anderson, Yu-Zhong Zhang, Brad Dubbels, Kathy Kihn, and Jolene Bradford Life Technologies, Eugene, Oregon 97402 USA Acoustic Focusing Cytometry: Sensitivity and Throughput Life Technologies • 5791 Van Allen Way • Carlsbad, CA 92008 • www.lifetechnologies.com D. Phagocytosis: pHrodo™ E.coli BioParticles ® Phagocytosis Kit for Flow Cytometry E. Apoptosis: Annexin V, Pacific Blue™ conjugate and SYTOX ® AADvanced™ dead cell stain G 0 G 1 : 39.80 % CV: 4.21% G 2 M: 17.89 % G 2 /G 1 : 1.96 S-Phase: 42.31 % G 0 G1: 40.13 % CV: 4.16% G 2 M: 18.58 % G 2 /G 1 : 1.96 S-Phase: 41.29 % G 0 G 1 : 40.70 % CV: 3.17% G 2 M: 18.55 % G 2 /G 1 : 1.96 S-Phase: 40.75 % G 0 G 1 : 47.84 % CV: 4.00% G 2 M: 17.64 % G 2 /G 1 : 1.94 S-Phase: 34.52 % G 0 G 1 : 43.53 % CV: 4.41% G 2 M: 16.71 % G 2 /G 1 : 1.96 S-Phase: 39.75 % G 0 G 1 : 44.32 % CV: 3.22% G 2 M: 17.46 % G 2 /G 1 : 1.96 S-Phase: 38.22 % G 0 G 1 : 42.20 % CV: 3.16% G 2 M: 19.13 % G 2 /G 1 : 1.96 S-Phase: 38.68 % Standard 1000μl/min Standard 500 μl/min Standard 200 μl/min Standard 100 μl/min Standard 25μl/min Sensitive 100μl/min Sensitive 25μl/min Sensitive 25μl/min 4.6% 68.2% 27.2% Sensitive 100μl/min 6.3% 65.2% 28.5% Standard 25μl/min 4.5% 70.4% 25.1% Standard 100 μl/min 4.0% 71.3% 24.7% Standard 200 μl/min 5.2% 69.5% 25.3% Standard 500 μl/min 5.0% 69.3% 25.7% Standard 1000μl/min 6.6% 67.4% 26.0% Sensitive 25μl/min Sensitive 100μl/min Standard 25μl/min Standard 100 μl/min Standard 200 μl/min Standard 500 μl/min 25.8% 24.2% 0.9% 49.1% 26.4% 0.3% 25.8% 47.6% 27.4% 0.3% 26.6% 45.8% 25.3% 0.3% 27.3% 47.2% 25.9% 0.3% 27.4% 46.4% 25.1% 0.5% 27.2% 47.2% Standard 1000μl/min 26.6% 0.3% 25.1% 47.9% Standard 200 μl/min Standard 100 μl/min Standard 25μl/min Sensitive 100μl/min Sensitive 25μl/min Ammonium chloride was used with human whole blood to lyse RBCs and labeled with anti-human CD4-Pacific Blueand anti-human CD8-Pacific Orange™ antibodies. Samples were analyzed using the Attune ® Acoustic Focusing Cytometer collecting 5,000 lymphocyte gated events. A dual parameter plot was analyzed for comparison of populations of CD4+CD8-, CD4-CD8+, CD4-CD8-, and CD4+CD8+ cell populations across all collection rates. C. Immunophenotyping: CD-4 Pacific Blue™ and CD-8 Pacific Orange™ staining of human lymphocytes Ethanol-fixed Jurkat cells were stained with propidium iodide + RNase. Samples were analyzed on the Attune ® Acoustic Focusing Cytometer and on a conventional cytometer, collecting 20,000 total events. Singlet cells were used for analysis of cell cycle using ModFit LT modeling software (Verity House) for unbiased data interpretation. Whole blood was processed using the pHrodo™ E.coli BioParticles ® Phagocytosis Kit for Flow Cytometry. Blood was treated with pHrodo™BioParticles ® and either incubated on ice or at 37 o C for 15 minutes. Following RBC lysis and wash, samples were analyzed on the Attune ® Acoustic Focusing Cytometer at all collection rates. The granulocyte population was gated and analyzed for fluorescence. Phagocytosing neutrophils were identified as the population of cells with increased fluorescence upon incubation at 37 o C due to phagocytosis of E.coli BioParticles ® into the acidic phagosome. Phagocytosis was inhibited by incubating other samples on ice. The median fluorescence of the positive population was divided by the median fluorsecence of the negative population to obtain a Signal/ Noise (S/N) ratio, which was consistent at all collection rates. Jurkat cells were treated with 10µM camptothecin for 4 hours at 37 0 C to induce apoptosis, prior to labeling with Annexin V Pacific Blue™ conjugate and SYTOX ® AAdvanced™ dead cell stain. Samples were analyzed on the Attune ® Acoustic Focusing Cytometer at all collection rates. Live, Apoptotic, and Dead cells were identified and percent of each population was compared across collection rates. F. Cell Cycle Analysis: No loss of Focusing at Higher Collection Rates Using Acoustic Focusing B. Lysed Whole Blood: Resolution of Cell Populations using 405 nm Forward and Side Light Scatter Standard 1000μl/min Sensitive 25μl/min Sensitive 100μl/min Standard 25μl/min Standard 100 μl/min Standard 200 μl/min Standard 500 μl/min S/N = 19 S/N = 17 S/N = 17 S/N = 19 S/N = 19 S/N = 18 S/N = 17 Consistent scatter Clear resolution of three blood populations (lymphocytes, monocytes, neutrophils) is possible without change in settings at all collection rates. G 0 G 1 : 41.73% CV: 4.83% G 2 M: 20.44% G 2 /G 1 : 1.96 S-Phase: 37.83% G 0 G 1 : 40.16% CV: 6.12% G 2 M: 20.89% G 2 /G 1 : 1.96 S-Phase: 38.95% G 0 G 1 : 44.60% CV: 7.76% G 2 M: 29.23% G 2 /G 1 : 1.90 S-Phase: 26.17% Acoustic Focusing Cytometry Hydrodynamic Focusing Cytometry Low 12μl/min Medium 35μl/min High 60μl/min Acoustic Focusing: lower CV & consistent results across collection rates: Acoustic CV: 3.17% 4.21% Across all collection rates Hydrodynamic Focusing causes a loss of focus at higher flow rates Conventional CV: 4.83% 6.12% 7.76% low medium high flow rate Consistent staining ratio (S/N) over all collection rates. Ammonium chloride was used with human whole blood to lyse RBCs, and Forward and Side Scatter using 405nm laser light was used to look at resolution of cell populations at all collection rates on the Attune ® Acoustic Focusing Cytometer. A consistent pattern was observed at all collection rates. Consistent cell populations identified by immunophenotyping over all collection rates. Fixed and permeabilized lymphocytes from human whole blood were stained for the intracellular antigen ZAP-70. Sample were labeled with surface CD3-Pacific Blue™ antibody and intracellular ZAP-70-rPE antibody, and data was collected across all collection rates on the Attune ® Acoustic Cytometer. CD3-negative lymphocytes were identified for ZAP-70 expression. Median fluorescence of positive and negative ZAP-70 cells were used to determine signal:background (S/B) ratio. G. Intracellular Staining: ZAP-70 S/B=39 S/B=27 S/B=27 S/B=36 S/B=28 S/B=18 S/B=18 Sensitive 25μl/min Sensitive 100μl/min Standard 25μl/min Standard 100 μl/min Standard 200 μl/min Standard 500 μl/min Standard 1000μl/min Clear distinction of positive and negative populations Consistent cell populations identified across all collection rates. A. Fluorescent Microspheres, 8-peak Standard 25μl/min Standard 100 μl/min Sensitive 100μl/min Sensitive 25μl/min 2.6% 3.0% 2.3% 2.3% Standard 500 μl/min Standard 1000μl/min Standard 200 μl/min 2.2% 2.3% 3.4% Eight peak fluorescent microspheres (Spherotech 8-peak Rainbow beads) were analyzed at all collection rates on the Attune ® Acoustic Cytometer using one fluorescent channel with 405 nm excitation and 450/40 bandpass emission collection optics. Coefficients of variation (CV) for the brightest peak are reported and show little variation across collection rates. Tight CVs are seen across all collection rates Standard 500 μl/min Standard 1000μl/min

Upload: others

Post on 18-Feb-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

  • Acoustic cytometry is a new technology that utilizes forces derived from acoustic radiation pressure to focus particles in flow cytometry. This technology exploits the physical differences between particles relative to the background medium, allowing cells to be tightly focused. The ability to focus cells without relying on hydrodynamic forces allows many possibilities outside the scope of conventional flow cytometry. The use of low flow rates are required for sensitive measurements with conventional hydrodynamic focusing, while high flow rates are reserved for less critical measurements. This has created a trade-off of sensitivity and throughput historically accepted with conventional flow cytometry. In contrast, acoustic focusing of particles separates the alignment of cells from a particular flow rate allowing for sensitivity at collection rates much higher than conventional cytometry.

    In this study we examined changes in sensitivity with sample flow rate using acoustic focusing cytometry. We performed typical assays using the Applied Biosystems® Attune® Acoustic Focusing Cytometer, such as DNA content cell cycle analysis, immunophenotyping, apoptosis, intracellular antibody labeling, phagocytosis, white blood cell light scatter, and microshpere analysis. Data was collected at all seven collections rates (25µl/min-1000µl/min) and results compared across collection rates for percent positive, percent coefficient of variation, signal to background ratio and/or staining index.

    In the DNA content cell cycle testing, conventional hydrodynamic focusing demonstrated a loss of focusing at higher sample rates resulting in higher %CVs and reduction in data quality. In comparison, consistent results and tighter %CVs were obtained across all flow rates using the Attune™ acoustic focusing cytometer. Distinct separation of positive and negative cell populations was demonstrated with the other assays, although a slight drop in staining index was noted at two flow rates. In this study we demonstrate that collection rates up to 1000µl/minute are possible without reduction in sensitivity when using acoustic focusing cytometry.

    For Research Use Only. Not intendedfor any animal or human therapeutic or diagnostic use

    J06 New and Emerging Technologies for Cell Biology B634

    Barbara Seredick, Chris Langsdorf, April Anderson, Yu-Zhong Zhang, Brad Dubbels, Kathy Kihn, and Jolene BradfordLife Technologies, Eugene, Oregon 97402 USA

    Acoustic Focusing Cytometry: Sensitivity and Throughput

    Life Technologies • 5791 Van Allen Way • Carlsbad, CA 92008 • www.lifetechnologies.com

    D. Phagocytosis: pHrodo™ E.coli BioParticles® Phagocytosis Kit for Flow Cytometry

    E. Apoptosis: Annexin V, Pacific Blue™ conjugate and SYTOX® AADvanced™ dead cell stain

    G0G1: 39.80 %CV: 4.21%G2M: 17.89 % G2/G1: 1.96S-Phase: 42.31 %

    G0G1: 40.13 %CV: 4.16%G2M: 18.58 %G2/G1: 1.96S-Phase: 41.29 %

    G0G1: 40.70 %CV: 3.17%G2M: 18.55 %G2/G1: 1.96S-Phase: 40.75 %

    G0G1: 47.84 %CV: 4.00%G2M: 17.64 % G2/G1: 1.94S-Phase: 34.52 %

    G0G1: 43.53 %CV: 4.41%G2M: 16.71 % G2/G1: 1.96S-Phase: 39.75 %

    G0G1: 44.32 %CV: 3.22%G2M: 17.46 %G2/G1: 1.96S-Phase: 38.22 %

    G0G1: 42.20 %CV: 3.16%G2M: 19.13 %G2/G1: 1.96S-Phase: 38.68 %

    Standard 1000µl/minStandard 500 µl/minStandard 200 µl/minStandard 100 µl/minStandard 25µl/minSensitive 100µl/minSensitive 25µl/min

    Sensitive 25µl/min

    4.6%

    68.2%

    27.2%

    Sensitive 100µl/min6.3%

    65.2%

    28.5%

    Standard 25µl/min4.5%

    70.4%

    25.1%

    Standard 100 µl/min4.0%

    71.3%

    24.7%

    Standard 200 µl/min5.2%

    69.5%

    25.3%

    Standard 500 µl/min5.0%

    69.3%

    25.7%

    Standard 1000µl/min6.6%

    67.4%

    26.0%

    Sensitive 25µl/min Sensitive 100µl/min Standard 25µl/min Standard 100 µl/min Standard 200 µl/min Standard 500 µl/min

    25.8%

    24.2%

    0.9%

    49.1%

    26.4% 0.3%

    25.8%

    47.6%

    27.4% 0.3%

    26.6%

    45.8%

    25.3% 0.3%

    27.3%

    47.2%

    25.9% 0.3%

    27.4%

    46.4%

    25.1% 0.5%

    27.2%

    47.2%

    Standard 1000µl/min

    26.6% 0.3%

    25.1%

    47.9%

    Standard 200 µl/minStandard 100 µl/minStandard 25µl/minSensitive 100µl/minSensitive 25µl/min

    Ammonium chloride was used with human whole blood to lyse RBCs and labeled with anti-human CD4-Pacific Blue™ and anti-human CD8-Pacific Orange™ antibodies. Samples were analyzed using the Attune® Acoustic Focusing Cytometer collecting 5,000 lymphocyte gated events. A dual parameter plot was analyzed for comparison of populations of CD4+CD8-, CD4-CD8+, CD4-CD8-, and CD4+CD8+ cell populations across all collection rates.

    C. Immunophenotyping: CD-4 Pacific Blue™ and CD-8 Pacific Orange™ staining of human lymphocytes

    Ethanol-fixed Jurkat cells were stained with propidium iodide + RNase. Samples were analyzed on the Attune® Acoustic Focusing Cytometer and on a conventional cytometer, collecting 20,000 total events. Singlet cells were used for analysis of cell cycle using ModFit LT modeling software (Verity House) for unbiased data interpretation.

    Whole blood was processed using the pHrodo™ E.coli BioParticles® Phagocytosis Kit for Flow Cytometry. Blood was treated with pHrodo™BioParticles® and either incubated on ice or at 37oC for 15 minutes. Following RBC lysis and wash, samples were analyzed on the Attune®Acoustic Focusing Cytometer at all collection rates. The granulocyte population was gated and analyzed for fluorescence. Phagocytosing neutrophils were identified as the population of cells with increased fluorescence upon incubation at 37oC due to phagocytosis of E.coli BioParticles® into the acidic phagosome. Phagocytosis was inhibited by incubating other samples on ice. The median fluorescence of the positive population was divided by the median fluorsecence of the negative population to obtain a Signal/ Noise (S/N) ratio, which was consistent at all collection rates.

    Jurkat cells were treated with 10µM camptothecin for 4 hours at 370C to induce apoptosis, prior to labeling with Annexin V Pacific Blue™ conjugate and SYTOX® AAdvanced™ dead cell stain. Samples were analyzed on the Attune® Acoustic Focusing Cytometer at all collection rates. Live, Apoptotic, and Dead cells were identified and percent of each population was compared across collection rates.

    F. Cell Cycle Analysis: No loss of Focusing at Higher Collection Rates Using Acoustic Focusing

    B. Lysed Whole Blood: Resolution of Cell Populations using 405 nm Forward and Side Light Scatter

    Standard 1000µl/minSensitive 25µl/min Sensitive 100µl/min Standard 25µl/min Standard 100 µl/min Standard 200 µl/min Standard 500 µl/min

    S/N = 19 S/N = 17 S/N = 17 S/N = 19 S/N = 19 S/N = 18 S/N = 17

    Consistent scatterClear resolution of three blood populations (lymphocytes, monocytes, neutrophils) is possible without change in settings at all collection rates.

    G0G1: 41.73% CV: 4.83%G2M: 20.44%G2/G1: 1.96S-Phase: 37.83%

    G0G1: 40.16%CV: 6.12%G2M: 20.89%G2/G1: 1.96S-Phase: 38.95%

    G0G1: 44.60%CV: 7.76%G2M: 29.23%G2/G1: 1.90S-Phase: 26.17%

    Acoustic Focusing Cytometry

    Hydrodynamic Focusing Cytometry

    Low 12µl/min Medium 35µl/min High 60µl/min

    Acoustic Focusing: lower CV & consistent results across collection rates: Acoustic CV:

    3.17% 4.21% Across all collection rates

    Hydrodynamic Focusing causes a loss of focus at higher flow ratesConventional CV:

    4.83% 6.12% 7.76%

    low medium high flow rate

    Consistent staining ratio (S/N) over all collection rates.

    Ammonium chloride was used with human whole blood to lyse RBCs, and Forward and Side Scatter using 405nm laser light was used to look at resolution of cell populations at all collection rates on the Attune®Acoustic Focusing Cytometer. A consistent pattern was observed at all collection rates.

    Consistent cell populations identified by immunophenotyping over all collection rates.

    Fixed and permeabilized lymphocytes from human whole blood were stained for the intracellular antigen ZAP-70. Sample were labeled with surface CD3-Pacific Blue™ antibody and intracellular ZAP-70-rPE antibody, and data was collected across all collection rates on the Attune® Acoustic Cytometer. CD3-negative lymphocytes were identified for ZAP-70 expression. Median fluorescence of positive and negative ZAP-70 cells were used to determine signal:background (S/B) ratio.

    G. Intracellular Staining: ZAP-70

    S/B=39 S/B=27 S/B=27 S/B=36 S/B=28 S/B=18 S/B=18

    Sensitive 25µl/min Sensitive 100µl/min Standard 25µl/min Standard 100 µl/min Standard 200 µl/min Standard 500 µl/min Standard 1000µl/min Clear distinction of positive and negative populations

    Consistent cell populations identified across all collection rates.

    A. Fluorescent Microspheres, 8-peak Standard 25µl/min Standard 100 µl/minSensitive 100µl/minSensitive 25µl/min

    2.6% 3.0% 2.3% 2.3%

    Standard 500 µl/min Standard 1000µl/minStandard 200 µl/min

    2.2% 2.3% 3.4%

    Eight peak fluorescent microspheres (Spherotech 8-peak Rainbow beads) were analyzed at all collection rates on the Attune® Acoustic Cytometer using one fluorescent channel with 405 nm excitation and 450/40 bandpass emission collection optics. Coefficients of variation (CV) for the brightest peak are reported and show little variation across collection rates.

    Tight CVs are seen across all collection rates

    Standard 500 µl/min Standard 1000µl/min