a new amperometric enzyme electrode for alcohol determination
TRANSCRIPT
A new amperometric enzyme electrode for alcohol determination
H. Gulce a,*, A. Gulce a, M. Kavanoz a, H. Coskun a, A. Yıldız b
a Department of Chemistry, Suleyman Demirel University, Isparta 32260, Turkeyb Department of Chemistry, Hacettepe University, Beytepe, Ankara 06532, Turkey
Received 25 May 2001; received in revised form 23 October 2001; accepted 20 December 2001
Abstract
A new enzyme electrode for the determination of alcohols was developed by immobilizing alcohol oxidase in polvinylferrocenium
matrix coated on a Pt electrode surface. The amperometric response due to the electrooxidation of enzymatically generated H2O2
was measured at a constant potential of �/0.70 V versus SCE. The effects of substrate, buffer and enzyme concentrations, pH and
temperature on the response of the electrode were investigated. The optimum pH was found to be pH 8.0 at 30 8C. The steady-state
current of this enzyme electrode was reproducible within 9/5.0% of the relative error. The sensitivity of the enzyme electrode
decreased in the following order: methanol�/ethanol�/n -butanol�/benzyl alcohol. The linear response was observed up to 3.7 mM
for methanol, 3.0 mM for ethanol, 6.2 mM for n -butanol, and 5.2 mM for benzyl alcohol. The apparent Michaelis�/Menten
constant (KMapp) value and the activation energy, Ea, of this immobilized enzyme system were found to be 5.78 mM and 38.07 kJ/
mol for methanol, respectively. # 2002 Elsevier Science B.V. All rights reserved.
Keywords: Amperometric; Enzyme electrode; Alcohol determination
1. Introduction
Quantitative determination of alcohols is important infood, fermentation and wine industries and in clinical
chemistry. The classical methods such as refractometry,
chromatography and steam distillation which require
expensive equipment and are time consuming are being
replaced by inexpensive, rapid and reliable methods
based on immobilized enzyme electrodes. The ampero-
metric response of these electrodes is either due to the
electrooxidation of H2O2 generated from the analyte inthe presence of the enzyme, alcohol oxidase (AOx),
R � CH2OH�O2 0AOx
R � CHO�H2O2
or the electrooxidation of the reduced form (NADH) of
the coenzyme b-nicotinamide adenine dinucleotide
(NAD�) at the applied potentials when alcohol dehy-
drogenase (ADH) is used as an enzyme.
CH3CH2OH�NAD� 0ADH
CH3CHO�NADH�H�
NADH 0 NAD��2e�H�
AOx catalyzes other alcohols as well and is thereforenot selective. ADH is selective for ethanol but the
electrode is unstable (Rebelo et al., 1994). The enzymes
immobilized in various matrixes catalyze one of the
above reactions generating the electroactive species.
Several amperometric enzyme electrodes using AOx
(Mason, 1983; Lubrano et al., 1991) and ADH/NAD�
system (Wangsa and Danielson, 1991; Lobo et al., 1996;
Castanon et al., 1997; Gotoh and Karube, 1994; Zhaoand Buck, 1991) were reported. Microbial alcohol
sensor using the yeast candida vini which was immobi-
lized in the pores of acetylcellulose filter was also
reported (Mascini et al., 1989). Bienzyme electrode
which works with immobilized alcohol oxidase�/catalase
system measures the electroreduction current of O2
molecule that is generated from H2O2 by the catalytic
action of catalase (Verduyn et al., 1983).
CH3CH2OH�O2 0AOx
CH3CHO�H2O2
H2O2 0catalase
O2�2H��2e
O2�4H��4e 0 2H2O
Alcohol oxidase�/peroxidase system which requires an* Corresponding author. Fax: �90-246-237-1106.
E-mail address: [email protected] (H. Gulce).
Biosensors and Bioelectronics 17 (2002) 517�/521
www.elsevier.com/locate/bios
0956-5663/02/$ - see front matter # 2002 Elsevier Science B.V. All rights reserved.
PII: S 0 9 5 6 - 5 6 6 3 ( 0 2 ) 0 0 0 0 8 - 8
addition electron transfer mediator such as 1,1-di-
methylferrocene (DMFc) for the reduction of H2O2
measures the current due to the reduction of the
oxidized form of the mediator (Kulys and Schmid,1991).
CH3CH2OH�O2 0AOx
CH3CHO�H2O2
2DMFc�H2O2�2H� 0peroxidase
2DMFc��2H2O
DMFc��/e0/DMFc
The use of a redox polymer, polyvinylferroceniumperchlorate (PVF�ClO4
�) as an immobilization matrix
allowed the development of novel amperometric sensors
for glucose, galactose and sucrose and lactose. It is
known that the reduced form of this redox polymer film,
PVF, is a homogeneous compact film whereas the
oxidized form, PVF�, is inhomogeneous film (Inzelt
and Bacskai, 1992). The pores or pinholes exist in PVF�
film through which dissolved reactants could diffuse tothe underlying metal surface (Peerce and Bard, 1980). It
was already reported in previous reports related to the
enzymatic sensors for glucose (Gulce et al., 1995a),
galactose (Gulce et al., 2002a), sucrose (Gulce et al.,
1995b) and lactose (Gulce et al., 2002b) that the
covalently bonded ferrocenium centers (PVF�) of this
redox polymer act as chemical oxidants for H2O2
oxidation and the reduced forms of the redox couple(PVF) get oxidized at the applied potential. This
catalytic property of the polymer can be used as a basis
for the development of a rapidly responding sensitive
alcohol sensor.
This work describes a new amperometric enzyme
electrode in polyvinylferrocenium perchlorate
(PVF�ClO4�) matrix. Electrooxidation current of the
enzymatically produced H2O2 was measured and theeffects of enzyme concentration, buffer concentration,
the amount of the free enzyme in solution during
immobilization, pH, temperature and long term stability
were determined.
2. Experimental
PVF�ClO4� modified Pt surface was prepared by
electrooxidizing polyvinylferrocene (PVF) at �/0.70 Vversus Ag/AgCl in a methylene chloride solution con-
taining 0.1 M tetrabutylammonium percholarate
(TBAP). PVF was prepared using a method of chemical
polymerization (Smith et al., 1976) of vinylferrocene
(Alfa products). The electroprecipitation of PVF�ClO4�
was carried out under nitrogen atmosphere. Methylene
chloride was purificated before use. For this aim,
methylene chloride was shaken with portions of conc.H2SO4 until the acid layer remained colorless. It was
washed with water, aqueous 5% Na2CO3, and then
water again. Then it was predried with CaCl2 and
distillated from P2O5 (Perrin and Armorego, 1980).
TBAP was prepared by the reaction of tetrabutylam-
monium hydroxide (40% aqueous solution) (Merck)
with HClO4 (Merck), crystallized from an ethylalcohol�/water mixture (9:1) several times and kept
under nitrogen atmosphere after vacuum drying at 120
8C. The buffer solutions were prepared using NaH2PO4
(AnalaR BDH) and NaOH (Merck). Alcohol (Merck)solutions were prepared in a 0.10 M phosphate buffersolution of pH 8.0. Alcohol oxidase (E.C. 1.1.3.13Sigma A6941) solution was prepared in 0.01 M phos-phate buffer solution of pH 8.0. At pH values above 5the enzyme is in the form of anion and can be ion-exchanged and electrostatically held by the polymer.The isoelectric point of the enzyme is between pH 4.0and pH 5.0 (Mizutani et al., 1997). Enzyme wasincorporated into the polymer matrix by immersingPVF�ClO4
� coated Pt electrode in enzyme solution for30 min according to the ion exchange process
PVF�ClO�4 �AOx�0 PVF�AOx��ClO�
4
The enzyme is held electrostatically in the polymeric
structure. The enzyme electrode was then rinsed with the
buffer solution of working pH to remove the excess
enzyme which was not held electrostatically. We know
quite well that enzyme molecules are immobilized
through the ion exchange process. In a previous study,
the amount of enzyme incorporated in the modified
electrode was determined by following the decrease ofthe absorbance of the enzyme solution at 277 nm during
the surface immobilization procedure. The amount of
enzyme adsorbed onto the glass surface that is lost by
washing was measured spectrophotometrically and
found to be negligible. Furthermore, the amount of
the enzyme immobilized within the polymer could then
be determined using the same absorption peak after
desorbing the enzyme in a solution that has a pH valueless than the isoelectric point of the corresponding
enzyme (Gulce et al., 1995a). The activity of the enzyme
electrode was determined with a jacketed electrochemi-
cal cell which kept the solution at a desired temperature.
Oxygen was introduced into the solution in this cell at a
constant flow rate to obtain an oxygen saturated
solution. Oxygen flow was continued above the solution
to keep it saturated with oxygen during the measure-ments.
Constant potential of �/0.70 V versus SCE was
applied to the enzyme electrode to measure the ampero-
metric response due to the electrooxidation of H2O2
produced enzymatically. Steady state background cur-
rent was first measured at this potential with a blank
buffer solution of working pH. After the steady state
background current value was reached, certain volumesof alcohol solution of known concentration were added
and the currents for each added amount of substrate
were recorded.
H. Gulce et al. / Biosensors and Bioelectronics 17 (2002) 517�/521518
A three electrode system was used as an electroche-
mical cell with separate compartments for the counter
and reference electrodes. SCE was used in aqueous
solution, Ag/AgCl electrode immersed in 0.10 M TBAPsolution that contained saturated amount of AgCl was
used in methylene chloride as reference electrodes. Pt
foil electrode (A�/0.5 cm2) was a working electrode.
The electrochemical instrumentation consisted of
PAR Model 362 Potentiostat�/Galvanostat. Current�/
time curves were recorded on a model 16100-II Linseis
recorder.
3. Results and discussion
When a constant potential of �/0.70 V versus SCE
was applied to the enzyme electrode and alcohol was
added to the electrochemical cell while stirring, a sharp
current response was obtained. No response was ob-
tained when the polymer coated Pt electrode did not
contain any immobilized enzyme. Fig. 1 shows the
responses of the enzyme electrode to methanol. Theresponse reached a steady state value within 30�/50 s.
Steady state current values were used to construct the
calibration plots.
The amount of the enzyme present in solution during
the immobilization process effects the current response
value as seen in Fig. 2. The response of the electrode did
not change significantly after about 6.4 units/ml solu-
tion. The following measurements were therefore taken
with an enzyme electrode prepared with a solution that
contained 10.24 units/ml.
An increase in the buffer concentration caused some
decrease in the response of the enzyme electrode after
about 0.10 M buffer concentration. The concentration
of buffer solution was kept at 0.10 M for all measure-
ments.
The pH dependence of the response of the enzyme
electrode was determined using a buffer solution con-
taining concentrated methanol solution (9.0 mM). The
current values increased up to a pH value of 8.0 and
decreased thereafter (Fig. 3). For each point in Fig. 3 a
new electrode was prepared in order to eliminate the
errors that might arise from re-use of the enzyme loaded
polymer surface. The data points are the averages of
three measurements. The pH of the solution was kept at
8.0 for all measurements. The maximum current values
obtained for the free enzyme using a bare Pt electrode
under the same conditions were with a solution of pH
8.0 and at 80 8C as well as seen in Fig. 3.The effect of temperature on the activity of the
enzyme electrode was determined at pH 8.0. The
temperature was increased from 17 to 90 8C. The
temperature at which the enzyme electrode yielded a
Fig. 1. The response of the alcohol oxidase electrode to methanol
additions (0.10 M buffer concentration, pH 8.0, 30 8C).
Fig. 2. The changes in Imax values of methanol with the enzyme
concentration in solution during immobilization (0.10 M buffer
concentration, pH 8.0, 30 8C).
Fig. 3. The effects of pH on the responses of the enzyme electrode and
free enzyme for methanol (0.10 M buffer concentration, 9.0 mM
methanol concentration, 30 8C).
H. Gulce et al. / Biosensors and Bioelectronics 17 (2002) 517�/521 519
maximum current was found to be 80 8C (Fig. 4a). The
activation energy of this immobilized enzyme system,
Ea, was found to be 38.07 kJ/mol from Arrhenius
plot. The response of the bare Pt electrode was similar
to that of the immobilized enzyme electrode at a
temperature range between 30 and 908C (Fig. 4b). The
response was found to reach a maximum at 808C as seen
in Fig. 4b.
Fig. 5 shows the changes in the response of the
enzyme electrode as a function of the concentration of
various alcohols. Each point in Fig. 5 is the average of at
lest three measurements and relative standard deviation
values were calculated to be less than 5%. The sensitivity
of the enzyme electrode decreased in the following
order: methanol�/ethanol�/n -butanol�/benzyl alco-
hol. The upper limit of the linear working portion in
the calibration plot was found to be 3.7 mM for
methanol, 3.0 mM for ethanol, 6.2 mM for n-butanol,
and 5.2 mM for benzyl alcohol. The apparent
Michaelis�/Menten constant (KMapp) value calculated
from the Lineweaver�/Burk plot data for methanol was
5.78 mM.
The Ea and KM values for the free enzyme system for
methanol were calculated to be 32.40 kJ/mol and 15.57
mM, respectively. The comparison of these values with
those of immobilized enzyme system gives evidence for
the fact that no substantial structural changes and
diffusional limitations occur in the immobilized state.
The electrode which was prepared optimum condi-
tions was tested at 30 8C with respect to its stability
using 9.0 mM methanol solution. A total of 132
measurements were made in 36 days. The electrode
stability was found to be satisfactory for a few days and
decreased considerably for the following periods as seen
in Fig. 6.The effects of several possible interferants such as
cholesterol, sucrose, D-glucose, D-fructose, D-galactose,
lactose, L-tyrosine and ascorbic acid were also investi-
gated. The most important interference was caused by
trozin and ascorbic acid which produced current values
of 1.5 and 2.4 mA when they are present in 5 mM
concentration in 0.10 M buffer solution of pH 8.0 at 30
8C.The alcohol sensor whose characteristics are outlined
above showed better value for the upper limit in the
linear portion of the calibration plot when compared to
the reported values in the literature. The response time
of the electrode was comparable to the published values
(Rebelo et al., 1994; Mizutani et al., 1997; Mascini et al.,
1989; Verduyn et al., 1983). The redox polymer, PVF�,
acts as a suitable immobilizing medium for alcohol
Fig. 4. (a) The changes in the activity of the enzyme electrode for
methanol with temperature. (b) The changes in the activity of free
enzyme for methanol with temperature (0.10 M buffer concentration,
9.0 mM methanol concentration, pH 8.0).
Fig. 5. The changes in the activity of the enzyme electrode for various
alcohols as a function of substrate concentration (0.10 M buffer
concentration, pH 8.0, 30 8C).
Fig. 6. Long-term stability of the enzyme electrode (0.10 M buffer
concentration, 9.0 mM methanol concentration, pH 8.0).
H. Gulce et al. / Biosensors and Bioelectronics 17 (2002) 517�/521520
oxidase and catalyzes the H2O2 oxidation that was
produced enzymatically.
References
Castanon, M.J., Ordieres, J.M., Blanco, P.T. 1997. Amperometric
detection of ethanol with poly-(o-phenylenediamine)-modified
enzyme electrodes. Biosens. Bioelectr. 12, 511�/520.
Gulce, H., .Ozyoruk, H., Celebi, S.S., Yıldız, A. 1995a. Amperometric
enzyme electrode for aerobic glucose monitoring prepared by
glucose oxidase immobilized in poly(vinylferrocenium). J. Electro-
anal. Chem. 394, 63�/70.
Gulce, H., .Celebi, S.S., Ozyoruk, H., Yıldız, A. 1995b. Amperometric
enzyme electrode for sucrose determination prepared from glucose
oxidase and invertase co-immobilized in poly(vinylferrocenium). J.
Electroanal. Chem. 397, 217�/223.
Gulce, H., Ataman I., Gulce, A., Yıldız, A., 2002a. A new ampero-
metric enzyme electrode for galactose determination, Enzym.
Microbiol. Techn., 30, 41�/44.
Gulce, H., Gulce, A., Yıldız, A., 2002b. A novel two-enzyme
amperometric electrode for lactose determination, Anal. Sci.,
2002b, Vol. 18, in press.
Gotoh, M., Karube, I. 1994. Ethanol biosensor using immobilized
coenzyme. Anal. Lett. 27, 273�/284.
Inzelt, G., Bacskai, J. 1992. Electrochemical quartz crystal micro-
balance study of the swelling of poly(vinylferrocene) films.
Electrochem. Acta 37, 647�/654.
Kulys, J., Schmid, R.D. 1991. Bienzyme sensors based on chemically
modified electrodes. Biosens. Bioelectr. 6, 43�/48.
Lobo, M.J., Miranda, A.J., Tunon, P. 1996. A comparative study of
some phenoxazine and phenothiazine modified carbon paste
electrodes for ethanol determination. Electroanal. 8, 591�/596.
Lubrano, G.J., Faridnia, M.H., Dallecchi, G., Guilbault, G.G. 1991.
Amperometric alcohol electrode with extended linearity and
reduced interferences. Anal. Biochem. 198, 97�/103.
Mascini, M., Memoli, A., Olana, F. 1989. Microbial sensor for
alcohol. Enzym. Microbiol. Techn. 11, 297�/301.
Mason, M. 1983. Ethanol determination in wine with an immobilized
enzyme electrode. Am. J. Enol. Vitic. 34, 173�/175.
Mizutani, F., Yabuki, S., Iijima, S. 1997. Amperometric alcohol-
sensing electrode based on a polyion complex membrane contain-
ing alcohol oxidase. Anal. Sci. 13, 83�/87.
Peerce, P.J., Bard, A.J. 1980. Polymer films on electrodes Part II. film
structure and mechanism of electrone transfer with electrodepos-
ited poly(vinylferrocene). J. Electroanal. Chem. 112, 97�/115.
Perrin, D.D., Armorego, W.L.F. 1980. Purification of Laboratory
Chemicals, 2nd ed.. Pergamon Press, Oxford.
Rebelo, J.F., Compaguone, D., Guilbault, G.G. 1994. Alcohol
electrodes in beverage measurements. Anal. Lett. 27, 3027�/3037.
Smith, W.T., Kuder, J., Wychick, E. 1976. Voltammetric behaviour of
poly(vinylferrocene). J. Polym. Sci. 14, 2433�/2448.
Verduyn, C., Van Dijken, J.P., Scheffers, W.A. 1983. A simple,
sensitive, and accurate alcohol electrode. Biotechnol. Bioeng. 25,
1049�/1055.
Wangsa, J., Danielson, N. 1991. Enzymatic determination of ethanol
by flow injection analysis using a KEL-F wax carbon paste
electrode. Electroanal. 3, 625�/630.
Zhao, J., Buck, R.P. 1991. An all-solid-state amperometric ethanol
sensor. Biosens. Bioelectr. 6, 681�/687.
H. Gulce et al. / Biosensors and Bioelectronics 17 (2002) 517�/521 521