NeuroEngineeringandmedicAl Robotics

LabBioengineering

Department

A-mode

ultrasound acquisition system for computer assisted

orthopaedic surgery…and other

Surgical Robotics - 4th summer school - Montpellier 9-16 September 2009

Danilo De Lorenzo

[email protected]

Page 2: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

Politecnico di Milano Politecnico di Milano –– Milan, ItalyMilan, Italy

2Danilo De Lorenzo, [email protected]

Neuroengineering and medical robotics laboratory

www.biomed.polimi.it/nearlab

Page 3: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

ROBOt and sensors integration for Computer Assisted Surgery and Therapy

www.robocast.eu

2 main projects

BONUS system

Computer assisted orthopaedic surgery

ROBOCAST

Keyhole neurosurgery

3Danilo De Lorenzo, [email protected]

Page 4: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

ROBOCAST ProjectROBOCAST Project�Bioengineering Department Politecnico di Milano, Italy

�Department of Neurological and Vision Sciences - University of Verona, Italy

� SIRSLab - Robotics and System Lab - University of Siena, Italy

� Department of Mechanical Engineering and the Institute of Biomedical Engineering - Imperial College, London, UK

� Prosurgics Ltd, UK

� Laboratory for CAS and Medical Image Processing - The Hebrew University of Jerusalem, Israel

4Danilo De Lorenzo, [email protected]

� Faculty of Mechanical Engineering - Technion, Israel Institute of Technology, Haifa, Israel

� Mazor Surgical Technologies Ltd, Israel

� Lehrstuhl für Computeranwendungen in der Medizin, Institut für Informatik - Technische Universität München, Germany

� Institute for Process Control and Robotics - Universität Karlsruhe (TH), Germany

� Finanziamenti Unione Europea S.r.l, Milan, Italy

Page 5: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

Closed loop control with

external sensors (optical

electromagnetic)

Use a Gross Positioner to

hold the Fine positioner in

place (close to the patient’s

head)

TheThe ConceptConcept

��tumour biopsies, tumour biopsies,

��treatment of hydrocephalus, treatment of hydrocephalus,

��aspiration and evacuation of deep brain hematomas, aspiration and evacuation of deep brain hematomas,

��‘‘OmmayaOmmaya’ catheter insertion, ’ catheter insertion,

��deep brain stimulation, deep brain stimulation,

��and minimal access craniotomiesand minimal access craniotomies

KEYHOLE NEUROSURGERYKEYHOLE NEUROSURGERY

5

head)

Develop an intelligent

automatic planner based

on risks atlas and

updateable with new

information

��and minimal access craniotomiesand minimal access craniotomies

Surgeon is handling an

haptic interface in order to

perform the needle

insertion

Page 6: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

ForceForce sensingsensing

52mm

6

Credit: Ilya Dyagilev, MRL laboratory, Technion University, Haifa (Israel)

10 20 30 40 50 60 70

-0.6

-0.5

-0.4

-0.3

-0.2

-0.1

0

0.1

Time (s)

Fo

rce

(N

)

Page 7: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

AimAim

•• ValidationValidation ofof anan ultrasoundultrasound systemsystem forfor bonebone surfacesurface acquisitionacquisition……..

8

•• …… inin computercomputer assistedassisted orthopaedicorthopaedic sugerysugery..

Danilo De Lorenzo, [email protected]

Page 8: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

Knee prosthesisKnee prosthesis

Femural componentDistal cut

Anterior cut

Posterior cut

arthrosisarthrosis70% aged (>65 years)

population affected

9

Liner

Tibial component

• Requirements on cuts:

• Ex. The plane of the distal

cut has to be perpendicular

to mechanical axe of the

femur

Posterior cut

Thanks to Lima L.to, Udine Italy

Computer assisted orthopaedic surgery

Danilo De Lorenzo, [email protected]

Page 9: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

Surgical navigation systemSurgical navigation system

Optical localization systemVicra NDI, Canada

Reference frames

10

Visual feedback

Localization error up to 7mm

(Yau et al. 2005)

Affecting correct prosthesis insertion

Registration

Page 10: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

Why ultrasound system?Why ultrasound system?

Optical localizationsystem

Opticalmarker

11

bone

Soft tissue

skin

marker

• No bone surface palpation

• Invasiveness reduction

handle

Danilo De Lorenzo, [email protected]

Page 11: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

The developed systemThe developed system

Pulser/Receiver TGCAntialiasing &

ADC

Analog Signal

FPGA Board(Elaboration, control and communication)

16 bit distancevia USBBlanking Switches

driverPulse

&reception

SyncSync

12

US Probe(5MHz transducer)

Sync signal

Localization systemPassive marker localizer

Sync signal interception board

Start sync Burst

Illuminating flash

Danilo De Lorenzo, [email protected]

Page 12: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

MethodsMethods -- silicone silicone layerlayer thickessthickess validationvalidation

MRI & CT

17

• After calibration position of the probe

tip and beam direction in the imaging

space is known (after ICP).

• It’s possible to measure thickness of

each silicone layer with oblique slices

and compare those thickness with the

ones calculated with US (velocity of

sound is known from characterization)

Danilo De Lorenzo, [email protected]

Page 13: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

ResultsResults –– OverallOverall US system vs US system vs PointerPointer

• On exposed surfaces (Bone Interface), the tracked pointer shows a median

distance value of 0.14mm which is significantly better than the US probe

value (0.21mm median value), but which is still acceptable.

• In the phantom regions covered by silicone, the US probe behaves

significantly better (max median value 1.2 mm) compared to the pointer.

• US

18

• US

• Pointer

Danilo De Lorenzo, [email protected]

Page 14: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

ConlusionsConlusions

• A-mode signal detection allows for real-time surface points detection.

• Accuracy tests of the whole system showed that the US probe acquisition

behaves significantly better (max median value 1.2 mm vs. 4.2mm) compared

to the acquisition with pointer.

• The tested system can considerably increase accuracy of registration and

biomechanical point computation

• Perpendicularity problem � further development

19Danilo De Lorenzo, [email protected]

• Perpendicularity problem � further development

• Problem: A-mode acquisition is slower than standard pointer points

acquisition

Page 15: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department

ConlusionsConlusions

Future developments:

• system testing in vivo during surgical procedure

• Perpendicularity check / feedback

• US real time velocity correction for each soft tissue, to improve distance

detection

22Danilo De Lorenzo, [email protected]

Page 16: A-mode ultrasound acquisition system for computer assisted orthopaedic surgery presentations... · 2009-09-25 · Neuro Engineering and medic Al Robotics Lab Bioengineering Department