Universitat de Valencia

Master Thesis

A full-brain PET scanner based on theAX-PET concept: Monte Carlo

performance study

Author:

Gabriel Reynes-Llompart

Principal Supervisor:

Paola Solevi, PhD

Co-Supervisor:

Josep F. Oliver, PhD

Master Universitari en Fısica Medica

September 2014

Page 2: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Contents

Contents i

Abstract iii

Abbreviations iv

Rationale v

1 Introduction 1

1.1 PET imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

1.1.1 Working principles of PET imaging . . . . . . . . . . . . . . . . . . 1

1.1.2 Resolution limitations . . . . . . . . . . . . . . . . . . . . . . . . . 3

1.1.3 Types of coincidences . . . . . . . . . . . . . . . . . . . . . . . . . 4

1.1.4 Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

1.2 PET radiopharmaceuticals . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

1.3 Brain Imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

2 Methods 9

2.1 AX-PET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

2.1.1 AX-PET proof of concept . . . . . . . . . . . . . . . . . . . . . . . 10

2.2 Monte Carlo simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

2.3 Reconstruction algorithm . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

3 Materials 18

3.1 Full Ring Brain Scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

3.2 Phantoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

3.2.1 Cologne High Resolution Phantom . . . . . . . . . . . . . . . . . . 20

3.2.2 NEMA HOT-COLD . . . . . . . . . . . . . . . . . . . . . . . . . . 20

3.3 Simulation setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

3.4 Image Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

4 Results and discussion 25

4.1 Cologne Phantom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

4.2 NEMA HOT-COLD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

4.2.1 Reconstructed image analysis . . . . . . . . . . . . . . . . . . . . . 27

4.3 Comparison of AX-PET geometries . . . . . . . . . . . . . . . . . . . . . . 32

4.4 Inclusion of ICS events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

i

Page 3: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

ii

Conclusion and future outlook 35

A Supplementary Data 37

Bibliography 41

Page 4: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

UNIVERSITAT DE VALENCIA

Abstract

Facultat de Fısica

Grupo de Fısica en Imagen Medica

Master Universitari en Fısica Medica

A full-brain PET scanner based on the AX-PET concept: Monte Carlo

performance study

by Gabriel Reynes-Llompart1

AX-PET is a novel PET detector conceived in order to reduce the parallax error and simultaneously

improve spatial resolution and sensitivity. Instead of the radial orientation of the scintillating crystals of

conventional scanners, AX-PET design is based on long axially arranged crystals. The discrete geometry

of AX-PET permits to detect and use the inter-crystal scattering (ICS) events to increase the system

sensitivity. The aim of the present work is to test the capability of the AX-PET concept for brain

imaging.

Two full ring AX-PET geometries are compared, the standard, based on an AX-PET demonstrator setup,

and a novel system with an optimized crystal arrangement; for both geometries, Monte Carlo simulations

of different sources have been done: the Cologne phantom, for resolution studies, and a NEMA hot and

cold phantom at different lesion-to-background ratios for image quality assessment. Both phantoms were

reconstructed using only golden events or golden events plus ICS events, and several figures of merit

were analysed. Novel geometry presents a slightly higher and more homogeneous sensitivity, thanks to

its compact design.

Both systems were capable to reconstruct and detect the 1 mm spheres; however, the novel system

presents a better contrast. On the other hand, the standard geometry shows a better performance on

all image quality figures of merit. The inclusion of ICS increase the contrast-to-noise ratio by 20% but

also increases the spill-over ratio in 6%.

Although the Novel geometry improves sensitivity and resolution, more efforts need to be done in terms

of image quality. Furthermore, the incorporation of ICS data acquired simultaneously with golden

coincidence measurements increases image quality and has potential to help identify lesions in images

with a low LBR. AX-PET based brain scanner shows a potential in terms of spatial resolution, however

more efforts shall be invested in the calculation of the system response matrix and sensitivity in order

to achieve a better image quality when dealing with extended sources where the lack of an appropriate

sensitivity correction can indeed compromise the reconstructed image.

[email protected]

Page 5: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Abbreviations

CT Computed Tomography

DOI Depth of Interaction

FDG 2-deoxy-2-Fluoro-d-Glucose

ICS Iinter-Crystal Scattering

LBR Lesion-to-Background Ratio

LOR Line of Response

MC Monte Carlo

MLEM Maximum-Likelihood Expectation-Maximization

MMPC Multi Pixel Photon Counters

NEMA National Electrical Manufacturers Association

PET Positron Emission Tomography

ROI Region Of Interest

TSL Threshold Single Level

SOPL Simulated One Pass List-mode

WLS WaveLength Shifter

iv

Page 6: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Rationale

The international AX-PET collaboration2 is trying to improve the design of the current

state positron emission tomography (PET) imaging by changing the typical alignment of

the detectors. The collaboration showed the capabilities of the AX-PET demonstrator,

based on two detector modules, in terms of both resolution and sensitivity.

Results have been reported in different publications since 2008. However, the AX-PET

system was originally conceived for brain imaging, and the performance of an AX-PET

based full-ring system for brain applications is for the first time studied in the present

work. Inside the AX-PET collaboration, the Image Reconstruction Instrumentation and

Simulation for medical imaging applications team (IRIS3), part of the Institut de Fısica

Corpuscular (IFIC), has developed simulation models for a full ring AX-PET prototype

and novel reconstruction techniques adapted to the unique features of the system.

The main objective of the present work is to test the capability of the AX-PET concept

for brain imaging through the simulation, reconstruction and analysis of several phan-

toms. The standard scanner geometry based on the AX-PET demonstrator design is

compared to a novel system with an optimized crystal arrangement; to achieve this goal,

the understanding and employments of several tools, such as the Monte Carlo GATE

and ML-EM reconstruction algorithms, were required.

The original contributions to the assessment of the performance of the AX-PET brain

scanner coming from this work are: the development and adaptation of phantoms to

the AX-PET Monte Carlo simulations, the image reconstruction through different event

selection (with and without Inter-Crystal Scatter events, with and without random and

scatter events) and, most important, the posterior analysis of the different reconstructed

images. The corresponding contributions are presented from chapter 3 -materials.

2AX-PET website at CERN.3IRIS website.

v

Page 7: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Chapter 1

Introduction

1.1 PET imaging

Positron Emission Tomography (PET) is a technique used in medicine to provide a 3D

image of metabolic body processes. This is why that PET images are called functional

images, in contrast to other image techniques as computed tomography (CT) which

provides anatomic images. PET has a role in the diagnosis of many different diseases,

such as cancer, heart stroke or Parkinson’s disease, conditions caused or accompanied

by local changes in metabolism. The specific functional information that is sought in a

PET scan depends on the injected tracer. In this sense, PET is an extremely flexible

imaging modality whose applications are limited only by the research advances in the

area of radioactive tracers and by the technological advances to improve the performance

of PET scanners.

1.1.1 Working principles of PET imaging

In PET imaging positron emitter radioisotopes are employed. The positron, emitted

by spontaneous decay, travels a certain distance depending on its energy and the tissue

density to finally interact with one electron of a surrounding atom, resulting in an

annihilation of the positron and the electron. The annihilation process is done when the

positron has lost almost all its kinetic energy.

The general equation of positron decay is described in Eq. 1.1.

1

Page 8: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

1. Introduction 2

AZX −→ A

Z−1Y +01 β + ν (1.1)

where A and Z are the mass number and atomic number of the decaying nucleus, e+ is

a positron and ν is an antineutrino.

Due to the energy and momentum conservation, two gamma rays are emitted, each

having an energy equal to the rest mass of the electron (511 keV), which propagates in

the opposite directions. The attenuation coefficients of the 511 keV gammas in water is

0.096 cm2/g, low enough to allow the two photons escaping the patient’s body.

Surrounding the patient there is a ring of detectors; if both emitted photons are detected,

the process is called a coincidence event and a line of response (LOR) is created joining

the points of the detection. The information recorded in every LOR is assembled and

employed to produce an image of the activity uptake in the patient’s body.

Figure 1.1: Schema of the PET acquisition sequence: the back-to-back detection, theprocessing of the coincidences and the reconstruction procedures. Image belongs to

Jens Maus, released to public domain.1

The detection of coincidence events is performed by an electronic coincidence sorting

unit through the application of a coincidence time window to detected photons. If two

events are measured within the time window, it is interpreted as a coincidence event.

The time window τ applied to score coincidences depends on the time resolution δt of

the device.

If data are stored in the list-mode each individual LOR is registered, preserving the

1http://jens-maus.de/home

Page 9: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

1. Introduction 3

spatial resolution of the detector. In this format, usually the time stamp of the LOR is

not recorded.

When dealing with analytic reconstruction and/or poor data statistics other formats

such as the sinogram histogram format might be more convenient. LORs are binned

thus the spatial resolution is decreased to the size of the bin.

A schema of the PET acquisition sequence can be seen in Figure 1.1.

1.1.2 Resolution limitations

In PET there are some intrinsic limitations to the best spatial resolution that can be

achieved in the final image. Most of these limitations arise from detector itself or the

physics associated to the positron annihilation [1, 2].

• Detector resolution:

The most important factor that degrades the spatial resolution is the scintilla-

tion detector intrinsic resolution of the PET scanner, which is due to solid angle

coverage and the underdetermination of the exact position of the interaction in-

side the detector crystal. In particular, in order to achieve a higher sensitivity

given the stopping efficiency of 511 keV photons in inorganic scintillators, longer

radial detection units are employed. The larger radial thickness implies a larger

uncertainty in the determination of the Depth Of Interaction (DOI) of the gamma

photon within the crystal. The resolution loss due to DOI uncertainty increases by

increasing the radial distance from the scanner centre. Thus the intrinsic resolution

(Ri ) primarily depends on the detector size.

• Positron range:

The positron range is the distance from the point of emission to the point of

annihilation; it depends on the kinetic energy with which the positron is emitted

by the nuclide. Positron range limits the maximum resolution achievable by a

PET scanner, it adds an intrinsic error Rp in the determination of the position

where annihilation occurred. With increasing scanner resolution, the trend is to

incorporate the positron range into the imaging model. Table 1.1 presents the

range of some commonly used radionuclides.

Page 10: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

1. Introduction 4

Table 1.1: Half life, maximum positron energy, and average positron range in waterof some nuclides commonly used in PET

Isotope Half-life (min)Maximum PositronEnergy (MeV)

Average PositronRange (mm)

11C 20.3 0.96 1.5213N 9.97 1.19 2.0515O 2.03 1.7 3.2818F 109.8 0.64 0.8382Rb 1.26 3.15 7.02

• A-collinearity:

The positron and the electron are not completely at rest when annihilating. As a

consequence, the angle between both gamma rays is not 180 degree and is instead

better described as a distribution around this value. This angular distribution can

be modeled as a gaussian function with a full width at half maximum (FWHM) of

approximately 0.5 degree. The a-collinearity error (Ra) increases with the distance

between the two detectors.

• Reconstruction technique:

A factor on the total error (Kr) due to some reconstruction techniques is intro-

duced by some filters in the filtered back-projection reconstruction method. Filters

applied to suppress noise in the reconstructed image employ a cut-off frequency in

the reconstruction resulting in a loss of spatial resolution.

• Detector localization:

The use of block detectors instead of single detectors causes an error Rl in the

event localization which deteriorates the spatial resolution.

Assuming that almost all the effects above add in quadrature, the total intrinsic recon-

structed spatial resolution is described by Equation 1.2.

RT = Kr

√R2i +R2

p +R2a +R2

l (1.2)

1.1.3 Types of coincidences

When both photons from the same annihilation are detected in coincidence without any

kind of interaction prior to the detection, it is called a true coincidence.

Page 11: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

1. Introduction 5

If at least one of the photons interacts within the patient’s body by Compton scattering

the original direction of the photon is changed, being then assigned a wrong LOR to

the true event. This type of events are known as scattering events, and result in a

noisy background to the true coincidence distribution, decreasing the image contrast

and increases the estimation of the activity.

Random coincidences occur when two photons from different annihilation events are

detected within the coincidence time window of the system.

When three or more photons are detected within the same time window then a multiple

coincidence is scored. Depending on the coincidence policy of the scanner, multiple

coincidences can be either included in the reconstruction or neglected.

The fraction of random and multiple coincidence events increases with the activity in

the imaged volume and may result in a loss of image contrast.

Figure 1.2: Illustration of the main coincidence event types: a) true; b) multiple; c)single; d) random and e) scattered. (Adapted from Simon R. Cherry)

Depending on the detectors employed in a PET scanner, a photon may undergo multiple

interactions in different detectors (e.g. Compton + Compton, Compton+Photoelectric)

As shown in Figure 1.3, these events, named Inter-Crytal Scatter events (ICS), can yield

two possible LORs but only one will be true.

Some PET scanners are not able to distinguish multiple hits, whereas modular detectors

are usually capable of it, and the interaction sequence cannot be stated. Then, the

position of the different interactions is stored in an arbitrary order. If this type of events

are included in the image reconstruction they will produce an increase of the sensitivity,

but in contrast, the spatial resolution of the image may be jeopardized.

Page 12: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

1. Introduction 6

Figure 1.3: Inter-crystal scatter event of one photon produced in the annihilationpair. In this example, one photon suffers Compton scattering inside one crystal andphotoelectric absorption in the next crystal. Two possible photon sequences resultingin two different LORs can be reconstructed: the wrong one represented in dashed blue,instead of the true one, represented in red, which is the one crossing the annihilation

point.

1.1.4 Sensitivity

PET sensitivity is determined by the intrinsic efficiency of the detectors employed and

by the geometry of the PET scanner that determines the angular coverage of the imaged

object. Some additionally parameters such as the energy thresholds applied at detector

level and coincidence policy also affect the sensitivity of the system.

The closer a detector is positioned to the source, the larger the solid angle it can cover.

The intrinsic efficiency of a detector is the probability to detect a gamma photon once

it enters a detection element, and it depends on the material composition and thickness

of the detector. When the detector thickness increases, the intrinsic efficiency increases

as well, but the depth of interaction (DOI) introduces an uncertainty on the origin of

the gamma rays, known as parallax error, that decreases the spatial resolution.

1.2 PET radiopharmaceuticals

The most common radiopharmaceutical used in PET is [18F]FDG2 for oncological imag-

ing and several different pathologies. The success of PET imaging in clinical practice is

associated to [18F]FDG and its widespread availability, the relatively long half-life of the

nuclide and the fact that many pathologies are associated with changes in the metabolic

rate of glucose. For example, many tumors present a high uptake of glucose due to an

22-deoxy-2-(18F)fluoro-d-glucose

Page 13: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

1. Introduction 7

accelerated metabolism required by tumoral cells to replicate faster.

However, many PET compounds have been synthesized with other β+ emitters as 11C,

13N, 15O. The short half-life of these radionuclides makes it necessary to produce them

in a biomedical cyclotron in-situ for a fast transfer and synthesis of the radioisotopes.

Figure 1.4: PET image of a normal brain using [18F]FDG.

1.3 Brain Imaging

The present work is focused on the performance estimation of a brain PET scanner.

Brain imaging is covering an increasing importance in clinical practice for mostly two

scenarios: oncological and neurological studies. Brain tumors are not very common, with

less than 0.1% prevalence in western population. However, they are among the most

fatal cancers [3], being gliomas the most frequent primary brain tumors, with an inci-

dence of 70%. With respect to neurological studies, with the increase in life expectancy

cognitive impairment has become a critical health issue. Projection of WHO expect

more than 48 million people in 2040 affected by dementia, with a dominant incidence of

Alzheimer’s disease (AD) [4].

There are two ways of extract the metabolic information of the central nervous system[5].

A first possibility is to extract information of brain functional activity and metabolism,

such as blood flow, rates of glucose and oxygen metabolism. This first approach allows

us to look for an specific abnormal function of the brain. An example of that is the use

of [18F]FDG for the diagnosis of several different pathologies as dementias, epilepsia,

brain tumors, etc. A second approach is based on the measurement of neurotransmitter

synthesis or enzyme activity. These are variables related to the function of the neuronal

populations that compose the central nervous system. An example of this is the radio-

pharmaceutical [18F]Fluoro-L-dopa used in the diagnosis and staging of the Parkinson

disease. Table 1.2 presents a list of some common brain radiopharmaceuticals.

From the technological point of view, brain imaging has different needs than whole-body

Page 14: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

1. Introduction 8

Table 1.2: Positron Emission Tomography radiopharmaceuticals.[6]

Compound Application

15O2 Oxygen metabolism and flow.11C methionine Amino acid metabolism.11C methylpiperone Dopamine receptor activity.18F FDG Glucose metabolism.18F fluoro-L-dopa Neurotransmitter.18F MISO Tumor hypoxia.

imaging, then it is possible to create dedicated brain PET scanners with a small detec-

tor ring diameter, which provide higher sensitivity when compared with a multipurpose

whole-body PET scanner. At the same time, the increased sensitivity achieved by such

scanners comes with decreased of the a-collinearity degradation of the spatial resolution

but usually increasing parallax and solid-angle errors. So the optimization of both sen-

sitivity and spatial resolution is crucial in brain imaging given the variety of lesion and

contrast magnitudes to be detected combined. In fact dealing with brain tumors requires

the detection of small lesions with poor contrast activity given the high overall uptake

of brain with [18F]FDG as well as large necrotic regions. In degenerative dementia small

hot and cold regions may be detected, with uptake and then contrast that may vary

depending on the injected radio-tracer and on the stage of the dementia. Sensitivity is

crucial in particular for early stage AD detection , when the neurodegeneration is less

and treatment may still be administered.

Page 15: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Chapter 2

Methods

2.1 AX-PET

Most PET systems are based on a radial arrangement of scintillating crystals [7]; such a

geometry usually requires to optimize either the spatial resolution or the sensitivity, due

to the lack of information about the DoI, which implies a parallax error. This problem

can be partially solved with the use of smaller crystals, but then the sensitivity of the

system is compromised. Therefore, a trade between both features is needed.

AX-PET is a novel PET detector conceived in order to reduce the parallax error and

simultaneously improve spatial resolution and sensitivity. Instead of the radial orienta-

tion of the scintillating crystals of conventional scanners, AX-PET design is based on

long axially arranged crystals and orthogonal Wavelength shifter (WLS) strips, both

individually read-out.

This configuration allows obtaining the 3D coordinates of the gamma interaction point

within the crystal. Similarly to common PET systems, the interaction of the photon

with the crystal gives the 2D coordinates of the detections (X,Y) and the photon energy.

However, to get the axial coordinate (Z), there are WLS plastic strips underneath each

crystal layer, placed perpendicularly to the crystals, with a small gap between them.

The process to detect each coordinate is described as follows: after the interaction of

a photon within the detector the crystal emits light isotropically, which is partially

absorbed by an MPPC1 (Multi Pixel Photon Counters) to get the coordinates and the

1Geiger Mode Avalanche Photodiode (G-APD) also known as SiPM and MPPC.

9

Page 16: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

2. Methods 10

(a) (b)

Figure 2.1: A) Drawing of a prototype of AX-PET module geometry. It is composedof six layers of eight crystals each interleaved with six layers of WLS. The MPPC aremounted in order to avoid dead areas. The crystals in adjacent layers are staggered byhalf a pitch size. B) Scheme of the method used for light propagation in the scintilla-tion crystal and the WLS strips. Both pictures belong to the AX-PET collaboration,

Beltrame et al.

energy of the photon. A portion of the light escaping the LYSO 2 crystal will be absorbed

by the WLS strips. These strips will absorb the blue light and re-emit it in the green

band. Optical photons will travel inside the strips towards their own WLS MPPC.

Moreover, not all light is absorbed by a given WLS3, and more than one WLS can detect

the light emitted by a given crystal (Si), making possible to get the Z position with a

higher resolution than the one obtained by a single WLS:

zevent =

∑∀i Si · zWLS,i

Stotal(2.1)

This design allows AX-PET to perform a 3D photon tracking and to identify the Comp-

ton interactions in the crystal matrix as ICS.

2.1.1 AX-PET proof of concept

An AX-PET demonstrator has been built at CERN and tested with different phantoms

and small animals at different radio-pharmaceutical facilities. The demonstrator con-

sists of two identical modules, each AX-PET module is composed of six layers of eight

crystals (LYSO), each one with an hodoscope of 26 WLS strips, hence a total of 48 scin-

tillator crystals and 156 WLS strips. Crystals and strips are read-out by two different

2Cerium-doped Lutetium Yttrium Orthosilicate. Composition: Lutetium (72%), Yttrium (4%), Sili-con (6%), Oxygen (18%). Density: 7.1 g/cm3, refractive index: ∼ 1.8.

3The light is detected by about 3 different WLS.

Page 17: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

2. Methods 11

types of MPPCs.

From the two modules fully assembled and characterized, Table 2.1 presents some num-

bers of its performance. Is important to mind the substantial fraction of Compton

scattered events which could be included in the reconstruction algorithm to enhance the

sensitivity.

Table 2.1: Some AX-PET demonstrator performance characteristics. Energy resolu-tion measured at 511 keV. Data extracted from Bolle et al.

Resolution FWHM Compton eventsx,y z Energy

2.03 mm 1.79 mm 11.7% 25%

The potential of the AX-PET demonstrator to image small animals was successfully as-

sessed; the measurements were performed at the ETH4 radio-pharmaceuticals laboratory

where a mouse and a rat were injected with [18F]FDG and imaged by the two-modules

AX-PET system. The image of a rat injected with the [18F]FDG is shown in Fig 2.2.

Figure 2.2: Maximum intensity projections of the rat imaged using [18F]FDG, recon-structed with golden events and ICS events. Adapted from Gillam et al.

4http://www.radiochem.pharma.ethz.ch/

Page 18: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

2. Methods 12

2.2 Monte Carlo simulation

The aim of the Monte Carlo (MC) method when used in radiation physics is to simulate

the radiation transport in matter, by numerically sampling random variables associ-

ated to probability distributions of each kind of interaction that can occur. Thus, MC

simulations are a common tool to address various issues related to PET imaging, from

designing and optimizing imaging systems to developing and assessing correction meth-

ods or tomographic reconstruction algorithms for improved image quality.

Within the context of the AX-PET system, the development of an accurate MC based

model is an important tool for a better understanding of the device, and to exploit its

specific system characteristics.

GATE5 is a Monte Carlo simulation tool based on Geant4 libraries, an object oriented

multi-purpose Monte Carlo toolkit developed at CERN [11]. It offers easy tools to

simulate SPECT and PET systems plus the description of source decay phenomena,

modelling the signal processing or the inclusion of the time component like real acquisi-

tions.

AX-PET non-conventional design, which relies on the WLS strips signals to retrieve

the interaction coordinate of the photon along the crystal, makes it necessary to de-

velop a computational model for the propagation and some major modifications to the

GATE source code. The adaptation of the AX-PET system to the MC simulation can

be divided in three steps:

• Simulation of the optical phenomena:

WLS strips behavior has to be included in the MC simulation to provide a realistic

axial resolution and a good event selection. Since Monte Carlo simulations are

time consuming, the optical photon transport in the crystal and in the WLS strips

geometry is not simulated but computed through an analytical model.

Light distribution on the WLS strips depends on the energy deposited in the LYSO

En, on the axial position of the interaction zn and on the depth of interaction in

the crystal xn. This is modeled as a Gaussian distribution of the number of

photoelectrons Npe, with Npe(i) photoelectrons on the ith strip:

Npe(i) =

N∑n=1

xiA(xn, En)e(zn−zi)

2

2σ2xn . (2.2)

5http://www.opengatecollaboration.org/

Page 19: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

2. Methods 13

In this expression indices i and n label the WLS strip and the interaction in

the crystal, respectively. A represents the signal amplitude associated to the nth

interaction and it varies with the deposited energy, En, and the depth of the hit

within the crystal, xn.

(a) (b)

Figure 2.3: (A) Experimental and simulated setup for the characterization of theWLS response. Two WLS strips were adjusted underneath a LYSO crystal excitedby a collimated electron beam. Figure adapted from Solevi et al. (B) Monte Carlosimulation of the z resolution. Both figures adapted from P. Solevi at CHIPP meeting.

• GATE simulation adapted to the set up of AX-PET:

The Monte Carlo model of the AX-PET uses a PET system available in GATE

as a reference, which includes the staggering and layered structure of the device.

GATE also processes the total energy deposited in the crystal which is used to

model the electronics of the system.

Given the photon interactions in the crystals -known as hits-, GATE is modified

with the inclusion of the WLS model, described earlier, in order to produce the

final output, named singles. The simulation uses a new digitizer module to operate

on the generated WLS signals from each strip of the output Single data, deter-

mining the new axial coordinate through determination of the center of gravity.

Moreover, the simulation output is adapted to the potential of the AX-PET with

the processing of the ICS events. It includes information that can be used to recon-

struct the true kinematics of the gamma-rays in the detector. The system stores

the track ID describing the inter-crystal scatter and the number of photoelectric

interactions occurring in the scintillation elements.

Page 20: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

2. Methods 14

• Electronics modeling via a dedicated sorter, performing an off-line processing of

the GATE output:

A threshold at single level (TSL) is applied to each single detected. For each single

passing the threshold, the pile-up effect 6 is modeled as a 250 ns window, where

all singles belonging to the same module and within this time window form an

event. As described below, all the singles recorded in the LYSO crystals belonging

to one module were then summed up and stored if the final energy is between the

thresholds of 400 keV and 600 keV. If two such events are found in coincidence -that

is, with a time difference below the 5 ns coincidence window applied- a coincidence

is formed.

Next, the coincidence is processed to introduce some dead-time. The dead time is

parametrized as follows:

τdt = Nwc · τwc + τ0(µs). (2.3)

Where Nwc is total number of hit channels (world count), τwc is the mean dead-

time per readout channel, and τ0 is the intrinsic dead-time of the system.

2.3 Reconstruction algorithm

AX-PET is a novel PET concept, and to exploit at best its potential, a huge effort in

terms of dedicated reconstruction software development is required.

In image reconstruction, there are two big families of algorithms: analytical and sta-

tistical iterative. The second group of algorithms can offer a higher image quality, but

needs an accurate model of the system and the physics involved.

An iterative reconstruction algorithm solves the problem defined by

y = T · x (2.4)

where y is the measured data, x the voxelized image and T is the system matrix, see

Figure 2.4.

There is a large set of different approaches to solve this problem; all results from this doc-

ument are obtained with the iterative Maximum-Likelihood Expectation-Maximization

6Pile-up effect: dead time due to signal pulses arriving closer in time than the pulse resolution timeof the system.

Page 21: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

2. Methods 15

(MLEM) algorithm in the standard list-mode implementation. MLEM algorithm pro-

vides an estimation of the mean intensity from voxel j, λj :

λk+1j =

λkjsj

∑i∈M

aij∑Jj=0 aijλ

kj

where sj =

I∑i=0

aij (2.5)

where nkj denotes the value of the voxel j at iteration k and aij the probability that

an emitted photon from the voxel j is detected in a measurement of i, and sj is the

sensitivity matrix. For the list mode, M refers to the set of measurements.

Computing aij is one of the big challenges of the ML-EM algorithm. The matrix could

be determined by calculations, simulations, or a combination of both; the more accurate

way to calculate it is to position a point source at all locations within the imaged

volume and record the counts in all elements of all possible projection profiles. Indeed,

this is a very time consuming procedure and it is usually simplified using symmetry

considerations.

(a) (b)

Figure 2.4: Reconstruction of an AX-PET image. A) The system matrix element aijis defined as the probability to detect the activity in voxel j in LOR yi, adapted fromP. Solevi at CHIPP meeting. B) Example of an inter-crystal event, from Gillam et al.

As mentioned above, one of the main features of the AX-PET is the possibility to include

the ICS events in the reconstruction procedure.

Consider an ICS consisting of a triple event detection, figure 2.4. This means that

there are two possible LOR but only one is correct, see figure 2.5. To include the ICS

events into the MLEM algorithm one should maximize the probability to include the

correct LOR. Given the i′ measurement of an ICS event there are usually two common

approaches to include it in Equation 2.5,

Page 22: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

2. Methods 16

• Separation: each possible LOR is separated and added with an appropriate weight

(η).ai′j∑J

j=0 ai′jnkj

≡ η1ai1j∑Jj=0 ai1jn

kj

+η2ai2j∑Jj=0 ai2jn

kj

(2.6)

• Selection: only the most probable LOR is considered, or in case of ignorance the

LOR will be selected randomly.

ai′j∑Jj=0 ai′jn

kj

≡ η1aitj∑Jj=0 aitjn

kj

(2.7)

The weights ηt for t = 1, 2 are the probability that each LOR is the correct one. There

are several ways to set these probabilities. The algorithm described by Gillam et al. uses

the differential Klein-Nishina cross-section, computed using the geometrical scattering

angle taken from the interaction location. If no extra information is given, a uniform

η = 0.5 is assumed.

Figure 2.5: Three different slices from reconstructed images of a NEMA phantomusing either Golden or ICS events. Adapted from Gillam et al.

For the AX-PET reconstruction, Gillam et al., propose an alternative approach to in-

clude the ICS events.

Page 23: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

2. Methods 17

• Inclusion or v-projection: it incorporates the full probability function associated

to the measurement.

ai′j ≡ η1ai1j + η2ai2j (2.8)

Using this method both LOR are kept, forming the v-shape, but with appropriate weights

assigned. In this study, η = 0.5 is used, which is equivalent to a randomized selection.

Fig 2.5 presents a comparison of the three different approaches to include ICS events on

AX-PET.

Data were reconstructed using simulated one pass list-mode (SOPL) [13]. SOPL is es-

sentially a multi-ray algorithm; the method models the full detection response with less

computational burden by spreading calculations over multiple iterations and with some

approximations.

The matrix elements are calculated on-the-fly (only during the reconstruction process)

and events are handled in list-mode. The algorithm uses a bootstrap Monte-Carlo sam-

pling to detect uncertainty functions and generate new rays at each iteration, the SOPL

adds a blurring over the detector comparing to single one ray. For the present work,

each measurement is modeled using five rays per LOR candidate. Each ray is simulated

using different distributions (random flat in y, exponential in x and gaussian in z) with

(dx, dy, dz) = [2.5, 2.5, 0.5] for the standard geometry, and (dx, dy, dz) = [2.5, 0, 0.5] for

the novel geometry. The parameters used are optimized based on the cross-section of

the LYSO crystals, 3×3 mm2 and on the axial resolution.

With this approach to image reconstruction, the treatment of ICS measurements inter-

acts with the computational burden as both separation and inclusion incur a computa-

tional penalty.

Figure 2.6: Schematic draw of a detector crystal and the SOPL algorithm. For eachray, five more rays are generated following a certain distribution; in the case of the z

coordinate a gaussian distribution around the center of the crystal.

Page 24: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Chapter 3

Materials

3.1 Full Ring Brain Scanner

AX-PET was initially conceived for brain imaging. After successful tests with small

animals, we are now interested in studying and evaluating the performance of a brain

scanner based on the AX-PET concept.

Two scanner designs are proposed: a conservative one having a full ring of AX-PET

modules, and an optimized design based on a slanted arrangement of the crystal layers.

The latter has been designed at CERN in order to reduce the gaps between modules.

In this configuration each 8 crystal layer is slanted 20 degree along the Z axis. Lower

gaps will be translated in a better homogeneity of the field of view (FOV) which has the

potential to improve the sensitivity performance of the system. The full ring AX-PET

scanner based on the original demonstrator design contains 48 modules arranged over

a ring with a diameter of 468 mm. In the novel slanted design the ring contains 300

layer blocks with a diameter of 476 mm. Since in the standard configuration one module

consists of 48 crystals to which the trigger for coincidence scoring is applied, the same

read-out logics is applied in the case of the novel geometry. In the latter, one module is

defined as the sum of 6 consecutive layers, thus yielding a 50 module full-ring scanner.

The slanted structure is based on the geometry of the high-energy physics detectors,

where more weight is given to improve detection rather than the final product cost.

Relevant parameters of both scanners for the coincidence scoring are:

• A 5 ns coincidence window is applied to select coincidences.

18

Page 25: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

3. Materials 19

• A threshold of [400,650] keV is applied at module level.

• Events with no associated WLS signal (e.g. interactions occurring at the edges of

the crystals where no WLS strips are placed) are neglected.

Additionally, each coincidence is stored together with a flag to identify the event type

(true, random or phantom scatter). The LORs to be reconstructed are assigned to the

center of the crystal in order to mimic the real acquisition conditions in PET scanners.

(a) (b)

Figure 3.1: Monte Carlo model of the conventional geometry (A) and the novelgeometry (B). One can see the slanted configuration of the novel geometry which implies

a better fit of the crystals and WLS.

The effect of the slanted geometry can be noted in the sensitivity maps presented in

Figure 3.2. For the standard geometry, the effect of the air gaps between the crystals

can be observed as inhomogeneities on the sensitivity matrix.

(a) (b)

Figure 3.2: Comparison of sensitivity map for the conventional geometry (A), andthe novel geometry (B). The homogeneity of the slanted geometry in comparison to the

standard geometry can be appreciated.

Page 26: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

3. Materials 20

3.2 Phantoms

In order to study the performance of the two full ring systems previously described,

different phantoms have been simulated: a Cologne phantom for resolution studies and

a NEMA phantom for image quality assessment.

3.2.1 Cologne High Resolution Phantom

The idea behind the use of this phantom is to test the resolution of the system[14]. It

consists of a container and two different inserts, which can be filled with radioactivity

diluted in water. The ’hot’ insert of the Cologne High Resolution Phantom consists of

a 28 mm thick lucite disk with a diameter of 219 mm. It is divided into areas with

holes that have different diameters and center-to-center spacings, figure 3.3. The holes’

diameters are 2 mm with a spacing of 4 mm, 3 mm with a spacing of 6 mm, and 4 mm

with spacing of 8 mm or 10 mm.

(a) (b)

Figure 3.3: Cologne High Resolution Phantom, A) schematic and B) real image. Hotinserts can be appreciated.

3.2.2 NEMA HOT-COLD

The evaluation of the image quality using the National Electrical Manufacturers Associ-

ation (NEMA) NU 2-2007 [15] requires a test phantom only suitable for a scanner with

a ring diameter of at least 350 mm. Then, the image quality test is designed to emulate

Page 27: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

3. Materials 21

a whole-body imaging performance, and therefore is not appropriate for a brain-only

tomograph. The same standard recommends the previous version, the NU 2-1994 [16],

as a better choice for brain-only tomographs, although is not either designed for such

equipments. The proposed phantom -usually refereed as Hot-Cold phantom- consists of

a 20 cm diameter and 20 cm long cylinder filled with water and with two cylindrical

inserts of 5 cm, usually one filled with activity (hot insert) and the other without (cold-

insert).

For the AX-PET performance analysis the phantom was slightly modified, 3.4a. The

length of the phantom was reduced to 6 cm to adapt to the AX-PET FOV -which is

not yet optimized for brain applications- and two more inserts of 4 mm diameter were

added, one filled with activity and the other one not.

This was performed in order to study the detection capability of small and big regions.

(a) (b)

Figure 3.4: A) Schematic draw of the NEMA modified phantom. B) Regions ofinterest analyzed.

3.3 Simulation setup

The total activity in each phantom was set according to the computed noise equivalent

count rate curve (NEC). NEC is a common metric used to describe the effective number

of counts measured by a PET scanner as a function of the activity in the FOV. As

commented in section 1, the main sources of statistical error in a PET system are

random and scatter events, but only true events are what we want. The NEC is defined

to show this effect as:

NEC =T 2

T + S +R, (3.1)

Page 28: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

3. Materials 22

where T , S and R are the true, scatter and random coincidences, respectively.

For both AX-PET geometries, previous studies have shown that the NEC curve presents

a maximum at about 60 MBq. The studies of the NEC were done with a uniform

phantom 20 cm long.

In case of the Cologne Phantom the total activity was set to the maximum of the NEC,

this is 60 MBq. The activity of each sphere can be seen in Table 3.1.

Table 3.1: Source setup for the Cologne phantom. The phantom is composed of 320spheres of different radii and activity with a total activity of 60 MBq.

NºRadii(mm)

Concentration(kBq/mm3)

120 1 69.6100 1.5 105.050 2 139.150 2 217.4

In the case of the NEMA phantom the total activity was set at 24 MBq, lower than

the NEC peak in order to work in a fiducial region. Each hot region of the phantom

simulates possible brain lesions; in brain imaging sometimes lesions are difficult to dis-

tinguish from background activity, depending in the radiopharmaceutical employed, and

tumor size at a certain stage. Afterwards, in order to simulate all possibilities in clinical

practice, we used three different ratios between the hot inserts and the background,

known as lesion-to-background ratio (LBR). The three LBR chosen were 20:1, 5:1 and

1.2:1.

At the moment, the AX-PET reconstruction method does not use any attenuation cor-

rection. Then, in order to be confident with the results, in the presented study instead

of water the phantom was filled with air.

Similarly, in an initial attempt simulations were reconstructed only with true events,

this is without any scatter or random events.

Either for the Cologne Phantom and the NEMA Hot-Cold, the time of simulation was

set to achieve at least one million golden events.

3.4 Image Analysis

We are interested in quantifying the capability of the AX-PET system to perform brain

imaging, considering different features of the AX-PET. Therefore, to estimate which

Page 29: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

3. Materials 23

provides better image quality, we need to use the figures of merits (FOM) according to

the tasks brain PET was designed for.

A series of regions of interest (ROI) were extracted from the phantom data, figure 3.4b,

and the following FOM for the NEMA Hot-Cold phantom were computed:

• Root-Mean-Square Error (RMSE): the discrepancy between the known emitted

source and the MC simulation were computed using the RMSE.

RMSE =

√∑j(nj − Sj)2

N(3.2)

where nj and Sj are the pixel values of the reconstructed image and the simulation

respectively and N is the total number of voxels. Thus FOM is used also to study

the global convergence of the image with iterations. It can be applied to all the

reconstructed images or to a certain structure, as the hot insert.

• Signal-to-Noise Ratio: it compares the level of a desired signal to the level of

background noise. It is defined as the ratio of signal to noise,

SNR =µUσU

, (3.3)

where µU and σU refer to the mean and standard deviation of a uniform background

region of the Hot-Cold phantom.

• Contrast: it refers to differences in intensity in parts of the image corresponding

to different levels of radioactive uptake. Thus the contrast of the lesion is defined

as

CX =|µX − µU |µX + µU

, (3.4)

where µX can be any of the two hot regions of the Hot-Cold phantom which

simulates a lesion.

• Contrast-to-Noise Ratio (CNR): even when the size of an object is substantially

larger than the limiting spatial resolution of the image, noise can impair detectabil-

ity, especially if the object has low contrast. A better definition of image contrast

for noisy images is the CNR,

CNRX =CXσU

(3.5)

Page 30: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

3. Materials 24

where, as contrast, X can be any of the two hot regions that simulate a lesion.

• Spill-over ratio: the ratio of the mean in each cold region to the mean of the

background uniform area was reported as spill-over ratio,

SOR =µCµu, (3.6)

where C can be any of the two cold regions of the Hot-Cold phantom. This param-

eter can give an idea about the inclusion of incorrect LORs in the reconstruction

process due to scatter, random or ICS events. For an ideal image the value ex-

pected for a cold region is 0.

• Recovery coefficient (RC): this parameter indicates how similar is the activity

concentration of a ROI to the expected value.

RCX =µX

µxexpected. (3.7)

For an ideal reconstructed image RC is equal to 1. The RC value is affected

by the resolution of the system and the consequent partial volume effect, by the

sensitivity correction applied to the image, as well as by the cleanness of the data

since random and scatter events can spread activity.

Previous to the image analysis, in order to be able to compute some quantitative param-

eters, it is important to normalize the values of the image to some unit. In the present

work we chose to normalize the image to the known background concentration, then

each voxel has the value of Bq/ml (the voxel size chosen for the image reconstruction is

1×1×1 mm3).

Page 31: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Chapter 4

Results and discussion

4.1 Cologne Phantom

The results for the Cologne phantom are presented in figures 4.1 and 4.2. Only golden

events are selected, with a filter on scatter and random events. Images are reconstructed

with 10·105 golden and 8·105 ICS counts for the standard geometry, and 11·105 golden

and 8·105 ICS counts for the novel geometry. Both geometries present a good resolution

performance. One can appreciate all sizes of spheres and different spacing, including the

1 mm spheres. From the plots of figures 4.2, an increase of the resolution in the central

smallest spheres for the novel geometry can be observed, in comparison to the standard.

Presented images correspond to iteration 10.

0

2

4

6

8

10x 10

8

(a)

0

2

4

6

8

10x 10

8

(b)

Figure 4.1: Reconstruction of the Cologne phantom for the A) standard geometry,and B) novel geometry.

25

Page 32: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

4. Results and discussion 26

(a)

50 100 150 200

0

0.2

0.4

0.6

0.8

1

1.2

mm

Co

un

ts (

Arb

. U

nits)

Standard

Novel

(b)

(c)

50 100 150 200

0

0.2

0.4

0.6

0.8

1

1.2

mm

Co

un

ts (

Arb

. U

nits)

Standard

Novel

(d)

Figure 4.2: Plot profiles for the Cologne phantom for two different sets of spheres.

4.2 NEMA HOT-COLD

Table 4.1 summarizes a list of all reconstructions performed and event selection. Most

of the images have been reconstructed for both golden and golden with ICS events, to

estimate the impact of ICS inclusion on the image. A more comprehensive study is

performed for the 5:1 LBR, where images were reconstructed with and without scatter

and random event rejection. Table 4.3 presents the scatter and golden events of all

Table 4.1: List of all simulations and reconstructions performed for the NEMA phan-tom. It includes both geometries (standard and novel), if the reconstruction uses onlygolden events or golden and ICS, and if the simulations includes scatter and random

events (s1) or not (s0).

Standard NovelGolden Golden

& ICSGolden Golden

& ICSs0 s1 s0 s1 s0 s1 s0 s1

1:1.2 x√

x√

x√

x√

1:5√ √ √ √ √ √ √ √

1:20 x√

x√

x√

x√

Page 33: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

4. Results and discussion 27

simulations. One can notice the higher sensitivity of the novel geometry, with more

golden events and a higher ratio of golden events per ICS events.

Table 4.2: Counts of golden events and ICS events depending on the lesion-to-background ratio.

LBR Standard Novel

Ratio Golden Events ICS Ratio Golden Events ICS Ratio

1:1.2 2.69E6 2.09E6 1.28 2.72E6 2.38E6 1.141:5 2.68E6 2.07E6 1.30 2.74E6 2.37E6 1.151:20 2.67E6 2.02E6 1.32 2.76E6 2.35E6 1.32

Table 4.3: Counts of golden events, scatter in phantom and random events dependingon the lesion-to-background ratio.

LBR Standard Novel

Ratio Scatter Random Scatter Random

1:1.2 78 136584 72 1325211:5 73 134837 68 1307791:20 71 130719 68 126906

For an LBR of 5:1, Figure 4.3 presents the central slice of the NEMA phantom. At first

sight, both geometries present similar results, although it can be observed that in the

slanted geometry appears a central artifact. In none of the two geometries the 4 mm

lesion can be appreciated, requiring a higher LBR, see Appendix A. Moreover, in the

reconstructions using golden and ICS events, Figure (4.3d and 4.3d) the image gains

definition despite the cold region increases activity.

Figure 4.4 presents a comparison between images obtained for both geometries with an

LBR of 1.2:1 and 20:1. The small hot lesion can be clearly observed for the LBR of 20:1,

in contrast to lower LBR; see also Figure 4.3. The novel geometry presents a slightly

higher definition of the small lesion in agreement with the Cologne phantom, Figure 4.1

4.2.1 Reconstructed image analysis

Figure 4.5 presents the variation of the RMSE as a function of the iteration. Recon-

structed data from Table 4.1 include scatter and randoms. For all different concentra-

tions the minimum occurs between iterations 10 and 15, where minimum RMSE value

is achieved by the standard geometry. The novel geometry presents a faster RMSE in-

crease with iteration. In both geometries the inclusion of ICS events reduces the slope of

Page 34: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

4. Results and discussion 28

0

10

20

30

40

50

60

70

80

90

100

(a)

0

10

20

30

40

50

60

70

80

90

100

(b)

0

10

20

30

40

50

60

70

80

90

100

(c)

0

10

20

30

40

50

60

70

80

90

100

(d)

Figure 4.3: Central slice of the NEMA Hot-Cold phantom reconstruction for a lesion-to-background ratio 5:1. A) standard geometry using golden events, B) novel geometryusing golden events, C) standard geometry using golden and ICS events, D) novelgeometry using golden and ICS events. Random and scatter events are included. Images

refer to iteration 10.

the increase. In all simulated sets the standard geometry behaves better than the novel

one, but once ICS events are included, the RMSE of the novel geometry gains over the

standard with only golden events. The best RMSE is obtained for an LBR of 5:1, still

comparable to the value achieved in the case of LBR 1.2:1. Larger RMSE values are

computed for the highest 20:1 LBR, for both geometries.

Figure 4.6 presents the RMSE of the hot 50 mm diameter insert as a function of the

standard deviation of the uniform background ROI. The minimum value is between

iterations 10 and 15 for both geometries. One can see a faster increases in the standard

deviation and RMSE of the insert for the novel geometry, and for both geometries with

the inclusion of ICS events. Nevertheless, for LBR of 1.2:1 and 5:1 all curves present the

same tendency. An LBR of 20:1 novel geometry with the inclusion of ICS events shows

a lower slope, Figure 4.6c.

Figure 4.7 presents the value of RC for the 5 mm hot insert as a function of the number

of iterations. The highest RC is obtained for the lowest LBR 1.2:1, although the conver-

gence to the stability is lower for this ratio. Standard geometry achieves values closer to

Page 35: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

4. Results and discussion 29

0

10

20

30

40

50

60

(a)

0

10

20

30

40

50

60

(b)

0

50

100

150

(c)

0

50

100

150

(d)

Figure 4.4: Central slice of the NEMA Hot-Cold phantom reconstruction using onlygolden events for a lesion-to-background ratio of A) 1.2:1 and standard geometry, B)1.2:1 and novel geometry, A) 20:1 and standard geometry, B) 20:1 and novel geometry.

The reconstruction was stopped at 10th iteration.

0 20 40 60 80 1002

4

6

8

10x 10

−3

Number of iterations

RM

SE

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

0 20 40 60 80 1000

0.005

0.01

0.015

0.02

Number of iterations

RM

SE

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

0 20 40 60 80 1000

0.005

0.01

0.015

0.02

Number of iterations

RM

SE

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(a)

Figure 4.5: RMSE as a function of the number of iterations, for the novel and standardgeometries and using only golden events or golden+ICS events. LBR ratios of A) 1.2:1,

B) 5:1, and C) 20:1.

unity compared to the novel geometry. On the other hand, the inclusion of ICS events

reduces the RC.

For the 4 mm hot insert, the RC values as a function of the iterations are shown in

Figure 4.7. The three figures present a slower convergence of the values compared to

the previous figures. The RC stabilizes at iteration 40 in the case of 20:1 LBR, while no

regular trend is seen for lower LBR 1.2:1.The size of the lesion can lead to partial volume

Page 36: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

4. Results and discussion 30

0 5 10 15 20 25 30 350

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

50 m

m in

se

rt R

MSE

Uniform ROI standard deviation

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(a)

0 5 10 15 20 250

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

50 m

m in

se

rt R

MSE

Uniform ROI standard deviation

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(b)

0 5 10 15 200

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

50 m

m in

se

rt R

MSE

Uniform ROI standard deviation

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(c)

Figure 4.6: RMSE of the 50 mm diameter hot insert vs the standard deviation of auniform background ROI, for the novel and standard geometry and using only goldenevents or golden+ICS events. LBG ratios of A) 1.2:1, B) 5:1, and C) 20:1. Each point

represents an increase of 5 iterations starting from 0.

0 20 40 60 80 1000

0.2

0.4

0.6

0.8

1

Number of iterations

Re

co

ve

ry C

oe

ffic

ien

t

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(a)

0 20 40 60 80 1000

0.2

0.4

0.6

0.8

1

Number of iterations

Re

co

ve

ry C

oe

ffic

ien

t

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(b)

0 20 40 60 80 1000

0.2

0.4

0.6

0.8

1

Number of iterations

Re

co

ve

ry C

oe

ffic

ien

t

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(c)

Figure 4.7: Recovery coefficient as a function of the number of iterations, for thenovel and standard geometry and using only golden events or golden+ICS events. LBG

ratios of A) 1.2:1, B) 5:1, and C) 20:1.

0 20 40 60 80 1000

0.2

0.4

0.6

0.8

1

Number of iterations

Re

co

ve

ry C

oe

ffic

ien

t

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(a)

0 20 40 60 80 1000

0.2

0.4

0.6

0.8

Number of iterations

Re

co

ve

ry C

oe

ffic

ien

t

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(b)

0 20 40 60 80 1000

0.2

0.4

0.6

0.8

Number of iterations

Re

co

ve

ry C

oe

ffic

ien

t

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(c)

Figure 4.8: Recovery coefficient for the 4 mm lesion. Recovery coefficient as a functionof the number of iterations, for the novel and standard geometry and using only golden

events or golden+ICS events. LBG ratios of A) 1.2:1, B) 5:1, and C) 20:1.

Page 37: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

4. Results and discussion 31

effects, with consequent spreading of the activity to nearby voxels, probably more visible

in the case of the lowest contrast scenario.

According to previous results, all parameters for the analysis were computed at iteration

10 of the reconstruction.

The difference with the expected contrast is presented in figure . Values are similar

for both geometries, although the novel geometry presents slightly better results at low

ratios. Contrast for both systems converges to a similar lower value at higher contrast.

The difference with the expected contrast increases dramatically for low ratios and with

the size of the lesion.

0 5 10 15 200

20

40

60

Ab

solu

t d

iffe

ren

ce

with

exp

ec

ted

co

ntr

ast

(%

)

Ratios

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(a)

0 5 10 15 200

50

100

150

200

250

300

350

400

450

Ab

solu

t d

iffe

ren

ce

with

exp

ec

ted

co

ntr

ast

(%

)

Ratios

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(b)

Figure 4.9: Difference with the expected contrast vs the LBR, for A) 50 mm diameterhot insert, and B) 4 mm diameter hot insert.

The SOR is represented in figure 4.10. As expected, the inclusion of ICS events slightly

increases the reconstruction of false LOR, which is translated into activity in cold zones.

The SOR increases as LBR increases or as lesion size decreases. At the same time, the

standard geometry presents a better SOR curve.

The SNR for the background ROI is presented in figure 4.11a; it decreases as the LBR

increases, which is due to less activity in the background ROI and consequent noise

increase. At the same time, it increases with the inclusion of ICS events, which could

be related to an enhancement of data statistics, thus improving image noise. The novel

geometry presents a better SNR, which can be related to a more uniform image back-

ground.

As previously mentioned, for noisy images CNR can be a better figure of merit than pure

contrast. Figure presents the CNR for both lesions, showing a comparable trend among

Page 38: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

4. Results and discussion 32

0 5 10 15 200

0.1

0.2

0.3

0.4

Sp

ill−

Ove

r R

atio

Ratios

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(a)

0 5 10 15 200

0.2

0.4

0.6

0.8

1

Sp

ill−

Ove

r R

atio

Ratios

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(b)

Figure 4.10: Spill-over ratio vs LBR, for the A) 50 mm diameter hot insert, and B)4 mm diameter hot insert.

0 5 10 15 200

0.2

0.4

0.6

0.8

1

1.2

1.4

Sig

na

l−to

−N

ois

e R

atio

Ratios

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(a)

Figure 4.11: Signal-to-noise ratio vs LBR.

the different systems. The standard geometry presents, for both Golden and Golden

with ICS events, better values. At same time the inclusion of ICS events increases the

CNR. That means that ICS inclusion on one side deteriorates the image contrast while

improving the image noise by improving the statistics. The trend among the two figures

usually results in better images when ICS events are reconstructed. As it can be seen,

the detection of the small lesion can be only appreciated with higher LBR where CNR

values are comparable with the largest hot lesion.

4.3 Comparison of AX-PET geometries

From the results of the Cologne phantom, in terms of resolution both geometries present

excellent results. At naked eye the smallest spheres of 1 mm radium can be appreciated,

whose size is equal to the matrix resolution.

Page 39: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

4. Results and discussion 33

0 5 10 15 200

0.05

0.1

0.15

0.2

0.25

Co

ntr

ast−

to−

No

ise

Ra

tio

Ratios

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(a)

0 5 10 15 20−0.1

−0.05

0

0.05

0.1

0.15

Co

ntr

ast−

to−

No

ise

Ra

tio

Ratios

Standard Golden

Novel Golden

Standard Golden+ICS

Novel Golden+ICS

(b)

Figure 4.12: Contrast-to-noise ratio vs LBR for the A) 50 mm diameter hot insert,and B) 4 mm diameter hot insert.

From image quality assessment, the novel geometry presents a promising higher unifor-

mity in the FOV, which implies a more uniform sensitivity matrix, it does not present

an enhancement in image quality but, on the contrary, the standard geometry seems to

be superior to the slanted one.

Appearance of an artifact at the centre of the image, specially for the novel geometry,

suggests a problem with the employed sensitivity matrix. The method used to compute

the sensitivity is less time consuming but compromises accuracy. At the same time, the

uniformity of the sensitivity matrix implies an increase in detection sensitivity due to

less gaps between detectors.

Another point which could affect the image quality is the calculation of the system

response elements using the SOPL algorithm. This method generates new rays for a

given crystal associated to a LOR using different distributions, but the method was

not optimized for a slanted geometry, what could implie that some rays are generated

outside the crystal.

From the point of view of the capability to perform a quantitative analysis of the activity,

both geometries present comparable results, as the differences with the expected contrast

are under 20% for normal LBR and for the big insert. For the small insert, errors on

the contrast are above the 50%. In the case of the LBR of 1:1.2, the error associated to

statistics makes it difficult to extract conclusions; a study with more counts needs to be

performed.

Page 40: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

4. Results and discussion 34

4.4 Inclusion of ICS events

As can be appreciated at naked eye, the inclusion of ICS events reduces noise in the

images. This can be seen observing that there is a small decrease of the contrast but an

increases of 20% on the CNR, in comparison to using only golden events. This effect is

the same for both geometries.

On the other hand, the only concern of using ICS events is the increase in the reconstruc-

tion of wrong LOR. This has a clear effect on the SOR parameter, which has maximum

increase of 6%. There is also a decrease on the recovery coefficient for the LBR of 5:1

and 20:1, which could affect a quantitative analysis of activity. Nevertheless, an increase

of the RC appears for the lower LBR, which shows the potential of ICS in cases where

it is important to get more counts for an accurate image.

Page 41: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Conclusion and future outlook

In this work we have studied the performance of a full ring scanner based on the AX-

PET concept for brain imaging. AX-PET is a novel detector concept based on axially

oriented crystals. The discrete geometry provides a 3D reconstruction of the gamma

interaction point, and additionally yields a large fraction of ICS events that can be used

to enhance the sensitivity of the system.

The AX-PET detector was originally conceived for brain imaging, with the choice of 3×3

mm2 crystal cross section. However, while the demonstrator was widely tested for small

animals imaging, no previous study has been performed with respect to brain imaging.

The presented work is the first attempt to investigate the potential of the AX-PET

detector for brain imaging. In addition to the conventional AX-PET design, a second

scanner geometry was also studied with tilted layers of crystals, expected to provide a

more homogeneous response in the FOV. The Cologne high resolution phantom confirms

the good spatial resolution of the system (in both geometrical configurations). Hot-Cold

NEMA phantoms with different contrast ratios and lesion sizes have been simulated and

reconstructed. With respect to extended sources, the conventional system provides

better images in terms of noise and image contrast than the novel geometry. Such a

result is against the expectations motivated by the more uniform sensitivity matrix of

the novel geometry.

However both scanners are unable to return with good contrast the smallest 4 mm

lesion, only visible at highest contrast value (20:1). In order to understand the observed

behavior several improvements are required. Among them we mention the need of a more

accurate calculation of the sensitivity matrix. A dedicated Monte Carlo calculation to get

a better sensitivity estimation may be required. Additionally, sensitivity changes when

dealing with pure golden events and ICS events, thus possibly a dedicated calculation of

the sensitivity matrix for both events selection may be of interest. The system matrix

35

Page 42: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Conclusion and future outlook 36

elements are computed on-the-fly. While the system response was widely tested for

the standard scanner design, improvements may be required in the case of the novel

configuration. Besides, the present study worked with a total number of counts lower

than real studies’ statistics. The incorporation, during image reconstruction, of ICS

data acquired simultaneously with golden measurements increases image quality and

has potential to help identifying lesions in images with a low LBR.

Since the last year the AX-PET collaboration started upgrading the detector, switching

to digitalSiPM, thus providing a better time resolution than conventional SiPM. The

good time resolution promotes AX-PET to a potential Time-Of-Flight (TOF) system,

with further chance to improve the image contrast. The benefits of TOF on brain images

however are expected to be limited, even if higher resolution can help in random rejection

whose fraction is not negligible for brain studies.

Page 43: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Appendix A

Supplementary Data

In this appendix we include a list of images of interest.

0

50

100

150

(a)

0

50

100

150

(b)

0

50

100

150

(c)

0

50

100

150

(d)

Figure A.1: Central slice of the NEMA Hot-Cold phantom reconstruction for a lesion-to-background ratio 20:1. A) standard geometry using golden events, B) novel geometryusing golden events, C) standard geometry using golden and ICS events, D) novel ge-ometry using golden and ICS events. The reconstruction was stopped at 10th iteration.

37

Page 44: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Appendix A. Supplementary Data 38

0

10

20

30

40

50

60

(a)

0

10

20

30

40

50

60

(b)

0

10

20

30

40

50

60

(c)

0

10

20

30

40

50

60

(d)

Figure A.2: Central slice of the NEMA Hot-Cold phantom reconstruction for a lesion-to-background ratio 1.2:1. A) standard geometry using golden events, B) novel geom-etry using golden events, C) standard geometry using golden and ICS events, D) novelgeometry using golden and ICS events. The reconstruction was stopped at 10th itera-

tion.

Page 45: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Appendix A. Supplementary Data 39

0

10

20

30

40

50

60

70

80

90

100

(a)

0

10

20

30

40

50

60

70

80

90

100

(b)

0

10

20

30

40

50

60

70

80

90

100

(c)

0

10

20

30

40

50

60

70

80

90

100

(d)

0

10

20

30

40

50

60

70

80

90

100

(e)

0

10

20

30

40

50

60

70

80

90

100

(f)

Figure A.3: Central slice of the NEMA Hot-Cold phantom reconstruction for a lesion-to-background ratio 1:5, only golden events, using a A) direct reconstruction of the trueevent localization (standard geometry), B) direct reconstruction of the true event lo-calization (standard geometry), C) non inclusion of scatter or random events (standardgeometry), D) non inclusion of scatter or random events (novel geometry), E) inclusionof scatter or random events, normal reconstruction, (standard geometry), F) inclusionof scatter or random events, normal reconstruction, (novel geometry). For the recon-

structions of true events it was stooped at iteration 0, for the rest at iteration 10.

Page 46: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Appendix A. Supplementary Data 40

0

10

20

30

40

50

60

70

80

90

100

(a)

0

10

20

30

40

50

60

70

80

90

100

(b)

0

10

20

30

40

50

60

70

80

90

100

(c)

0

10

20

30

40

50

60

70

80

90

100

(d)

0

10

20

30

40

50

60

70

80

90

100

(e)

0

10

20

30

40

50

60

70

80

90

100

(f)

Figure A.4: Central slice of the NEMA Hot-Cold phantom reconstruction for a lesion-to-background ratio 1:5, only golden events, stopped at A) 20 iterations (standardgeometry), A) 20 iterations (novel geometry), A) 50 iterations (standard geometry), A)50 iterations (novel geometry), A) 100 iterations (standard geometry), A) 100 iterations

(novel geometry).

Page 47: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Bibliography

[1] W. W. Moses, Nuclear Instruments and Methods in Physics Research Section A: Ac-

celerators, Spectrometers, Detectors and Associated Equipment 648, S236 (2011).

[2] M. E. P. Simon R. Cherry, James A. Sorenson, Physics in nuclear medicine, 4th

edition (Elsevier Sounders, 2012).

[3] S. Bauer, R. Wiest, L.-P. Nolte, and M. Reyes, Physics in medicine and biology 58,

R97 (2013).

[4] J. Jimenez Bonilla and J. Carril Carril, Revista Espanola de Medicina Nuclear e

Imagen Molecular (English Edition) 32, 301 (2013).

[5] C. Schiepers and M. Allen-Auerbach, Diagnostic nuclear medicine (Springer, 2006).

[6] H. A. Ziessman, J. P. O’Malley, and J. H. Thrall, Nuclear medicine: the requisites

(Elsevier Health Sciences, 2013).

[7] P. Beltrame, E. Bolle, a. Braem, C. Casella, E. Chesi, N. Clinthorne, R. De Leo,

G. Dissertori, L. Djambazov, V. Fanti, et al., Nuclear Instruments and Methods in

Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated

Equipment 654, 546 (2011), ISSN 01689002.

[8] P. Beltrame, E. Bolle, A. Braem, C. Casella, E. Chesi, N. Clinthorne, R. D. Leo,

G. Dissertori, L. Djambazov, C. Joram, et al. (????).

[9] E. Bolle, A. Braem, C. Casella, E. Chesi, N. Clinthorne, E. Cochran, R. De Leo,

G. Dissertori, G. Djambazov, V. Fanti, et al., Nuclear Instruments and Methods in

Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated

Equipment 617, 214 (2010).

41

Page 48: A full-brain PET scanner based on the AX-PET concept ... · A full-brain PET scanner based on the AX-PET concept: Monte Carlo performance study ... Inside the AX-PET collaboration,

Bibliography 42

[10] J. E. Gillam, P. Solevi, J. F. Oliver, C. Casella, M. Heller, C. Joram, and M. Rafecas,

Physics in medicine and biology 59, 4065 (2014).

[11] S. Jan, G. Santin, D. Strul, S. Staelens, K. Assie, D. Autret, S. Avner, R. Barbier,

M. Bardies, P. Bloomfield, et al., Physics in medicine and biology 49, 4543 (2004).

[12] P. Solevi, J. F. Oliver, J. E. Gillam, E. Bolle, C. Casella, E. Chesi, R. De Leo,

G. Dissertori, V. Fanti, M. Heller, et al., Physics in medicine and biology 58, 5495

(2013), ISSN 1361-6560.

[13] M. G. Bickell, L. Zhou, and J. Nuyts (????).

[14] M. Schmand, K. Wienhard, M. Casey, L. Eriksson, W. Jones, J. Reed, J. Treffert,

M. Lenox, P. Luk, J. Bao, et al., Nuclear Science Symposium, 1999. Conference

Record. 1999 IEEE 2, 1067 (1999).

[15] N. E. M. Association, NEMA Standards Publication NU 4-2008: Performance Mea-

surements of Positron Emission Tomographs (2008).

[16] N. E. M. Association, NEMA Standards Publication NU 4-2008: Performance Mea-

surements of Positron Emission Tomographs (1994).