Hot plate (LAMDA-P)

[1]

Page 2: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

After uniform thin films coated, the liquid solution is necessary to be solidified.

At this moment, hot plate is used to solidify the solution.

The temperature is accurately controlled by a temperature controller that uses digital

PID auto-tuning function, setting and real temperature can be simultaneously con-

firmed.

Temperature controller built-in timer function.

Ceramic coated top plate has rapid thermal conductivity and very uniformly maintain

the surface temperature.

Stirring speed is maintained constant speed by feedback control.

What is the hot plate and features?

[2]

Page 3: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

Specifications (HSD 180)

Type Hot plate & Magnetic Stirrer

Temperature

Range Max 380℃

Accuracy ±1℃Controller Digital PID auto-tuning

Timer 99 hr. 59 min.

Heater power 680 W

Stirring

Speed 100 ~ 1500 rpm

Speed display Digital

Capacity (H2O) up to 5L

Speed Controller Feedback controlTop Plate Ceramic coated top plate

DimensionsPlate area (mm) 180 × 180

Overall (W×D×H) (mm) 200 × 310 × 107

Electrical supply 220 V 50/60 Hz or 110 V 50/60 Hz

Weight 3.8 Kg

[3]

Page 4: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

1. Measurement unit display (PV): The current temperature during control and the various parameters

during setting are displayed respectively.

2. Setting unit display (SV): The setting temperature during control and each parameter value during

parameters setting are displayed respectively. While the timer is operated by pressing 9 ( ) but-

ton for a few sec., remaining time is displayed with blinking.

3. Control output indicator lamp (OUT): The light is on, after starting control output.

4. Upper and lower limit alarm indicator lamp (ALM): When set upper and lower limit alarm is acti-

vated, the lamp is turned on.

Each part designation and description

[4]

Page 5: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

5. Auto-tuning operation indicator lamp (AT): When PID auto-tuning (automatic operation) is started

by pressing 11 ( ) button, the current operation state is displayed with turning on the lamp. After

finishing operation, the lamp is turned off.

6. Timer operation lamp: When the timer is on by pressing 10 ( ) button, the operating timer is dis-

played with blinking. After the set time is over, the lamp is turned off and “tENd” is displayed on

setting unit display (SV).

7. LBA operation lamp: When the control loop break alarm is occurred, the lamp is turned on.

8. Mode setting button: If pressing button for more than 2 sec., PV and SV are changed parameter

setting mode. This button is also used to switch each parameter.

Each part designation and description cont.

[5]

Page 6: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

9. Decreasing setting value button (remaining time display button): The setting value is decreased one

by one with pressing button. If pressing button for more than 2 sec., the value is continu-

ously decreased. While the timer is operated by pressing button for a few sec. in controlling

state, remaining time is displayed with blinking.

10. Increasing setting value button (timer on/off button): The setting value is increased one by one with

pressing button. If pressing button for more than 2 sec., the value is continuously increased.

If pressing button for a few sec. during controlling state, the timer can be on/off. After the timer

is over, “tENd” is displayed on setting unit display (SV). If pressing button for a few sec. again,

returning to initial state.

Each part designation and description cont.

[6]

Page 7: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

11. Auto-tuning button: After setting parameters, if pressing button for more than 2 sec., operating

the optimum PID integer automatically with turning on auto-tuning operation indicator lamp (AT).

Each part designation and description cont.

[7]

Page 8: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

Parameters setting methodOrder Symbol Designation Hot plate & Magnetic Stirrer Initial value

1 SuSetting

temperatureSet the target temperature value 30

2 tlmControl end time setting

*H.m display unit: 00 hr. 01 min. ~ 99 hr. 59 min.After reaching setting temperature,the time is counted timer is over.

※ If setting “0”, the timer is not operated.

0

3 ALSAlarm mode

setting

Alarm modes of all nine kinds are displayed by figure. Proper kind of alarm can be set using up and down button.

----

4 ALHUpper limit alarm value

When operating alarm, set upper limit alarm value.Setting value is absolute value.

0

5 ALLLower limit alarm value

When operating alarm, set lower limit alarm value.Setting value is absolute value.

0

6 PProportional

bandRange of 0.1 ~ 999.8% can be set in 0.1% unit.

Set automatically during auto-

tuning

7 I Integration time Range of 5 ~ 9998 sec. can be set in 1 sec. unit.

8 dDifferentiation

timeRange of 0 ~ 2500 sec. can be set in 1 sec. unit.

9 LdAControl loop break alarm

Range of 0 ~ 9998 sec. can be set in 1 sec. unit.

[8]

Page 9: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

Parameters setting method cont.Order Symbol Designation Hot plate & Magnetic Stirrer Initial value

9 LdAControl loop break alarm

Range of 0 ~ 9998 sec. can be set in 1 sec. unit.

Set automatically during auto-

tuning

10 InSInput value correction

Range of -100.0 ~ 100.0 can be set℃ in 0.1 unit.℃ 0.0 or 1.0

11 LOCSetting value

lock

Lock: prevent to change setting valueOn: impossible to change parametersOff: possible to change parameters

Off

12 PASS Producer side mode

[9]

A. After all parameters are set, button is pressed for more than 2 sec. to escape the parameter set-

ting state. At this moment, the current and setting temperature are displayed at measurement unit

display (PV) and setting unit display (SV) respectively and the control is started.

B. If pressing button for more than 2 sec., the controller runs optimum control to operate automati-

cally the most proper PID integer to the thermal properties of the actuator controller.

C. If there is no any button operation for more than about 60 sec., returning to the initial state. At

this moment, the previous parameter value is stored, not set parameter value.

Page 10: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

1. ① After designating external sensor operation by # 16 button, insert sensor connector at # 17.

(When using external sensor)

② Designate internal sensor operation. (When using internal sensor)

2. After connecting cable to # 15, turn on the power switch of # 14.

3. If pressing # 8 button for about 2 sec., the parameter of temperature setting is displayed. The expec-

tative temperature is defined by pressing # 9 and 10 buttons.

4. If pressing # 8 button for a few sec. after setting the temperature, the temperature controller is oper-

ated. (initial state)

Instruction to control the temperature

[10]

Page 11: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

1. The parameter of temperature setting is displayed with pressing # 8 button for a few sec.. If press-

ing # 8 button again, “tlm” is displayed. The expectative time is defined by pressing # 9 and 10 but-

tons. Minimum 1 min. (0001) ~ maximum 99 hr. 59 min. (9959)

2. If pressing # 8 button for a few sec. after timer setting, returning to initial state.

3. When operating the timer, press # 10 button for about 2 sec.. At this moment, the lamp of # 6 is

blinked. If the timer is not set, not operating.

4. If pressing # 10 button for about 2 sec. while the timer is operating, operation is stopped.

5. If pressing # 9 button for about 2 sec. while the timer is operating, remaining time is displayed. At

this moment, if pressing # 9 button for a few sec. again, the setting temperature is displayed.

6. When the timer is over, “tENd” is displayed. If returning to initial state, press # 10 button for a few

sec.

Instruction to operate the timer

[11]

Page 12: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

1. Because this PID Temp. Controller have already calculated PID values by experiment, it can be

used without additional auto-tuning. If the temperature difference occurs a lot, auto-tuning is rec-

ommended.

2. Auto-tuning is started by pressing # 11 button and the lamp of # 5 is blinked.

3. Although there are some temperature errors during auto-tuning, the exact temperature is activated

after auto-tuning is over.

4. When using the thermostat, the temperature is firstly increased more than the set temperature but

returned to normal state after a while. (Overshoot phenomenon)

Instruction to operate the auto-tuning

[12]

Page 13: 3D Micromanufacturing Lab. School of Mechatronics Gwangju Institute of Science and Technology 3D Micromanufacturing Lab. School of Mechatronics Gwangju

3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology 3D Micromanufacturing Lab.School of MechatronicsGwangju Institute of Science and Technology

The expectant speed (# 12 display) is defined by rotating # 13 volume.

1. The speed can be controlled from 100 to 1500 rpm by adjusting volume. (The speed is in-

creased by clockwise rotating # 13 volume or decreased by anticlockwise rotating # 13 vol-

ume in 10 rpm step.)

2. If anticlockwise rotating # 13 volume fully, the operation is stopped.

3. If rpm is set by rotating # 13 volume, the speed is reached to the setting rpm within a few

sec.. The speed is maintained by the operation from the arrival time.

4. If the motor is not immediately rotated due to the influence of the load at first operation, the

output should be quickly increased up to above 10 rpm to rotate the motor within 3 sec..

Instruction to operate the motor

[13]