1 using in-situ cryogenic radiometers to measure the performance of a large thermal vacuum chamber...

12
1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1 , Edgar Canavan 1 , James Tuttle 1 , Keith Havey 2 , Jesse Huguet 2 , Wes Ousley 3 , Peter Shirron 1 , and Thomas Hait 1 Code 552, NASA/Goddard

Upload: adele-barber

Post on 19-Jan-2016

215 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

1

Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber

Michael DiPirro1, Edgar Canavan1, James Tuttle1, Keith Havey2, Jesse

Huguet2, Wes Ousley3, Peter Shirron1, and Thomas Hait1

1Code 552, NASA/Goddard Space Flight Center, Greenbelt MD 207712Exelis Corporation, Rochester, NY

3Genesis Engineering, Lanham, MD

Page 2: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

2

Background

JWST will test the cold section (combined instrument/telescope) in a very large thermal vacuum chamber at Johnson Space Center

Prior to that final test several preliminary tests will verify chamber and ground support equipment performance

Cold radiometers will be used to check for heat leaks and measure known heat sources

Shown is the first test, the Chamber A Commissioning Test Dummy mass

Auto-collimating Mirrors& Center of CurvatureOptical Assembly (CoCOA)

Page 3: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

3

Radiometer Operation

Winston cone collects thermal radiation from 11° half-angle

Copper disk coated with “Steelcast” absorbs radiation

Cernox thermometer mounted to back measures temperature increase

Cernox mounted to base measures background temperature

Flexible mount moved to point at potential heat source

Read out uses optical/electrical calibration and known temperature dependence

Output in mW/m2 of source

50m

m

Page 4: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

4

Sensitivity depends on radiometer isolation (calibrated for each radiometer), thermometry (dR/dT) and readout system

For 20 K heat sink typical readout resolution is better than 1 mK which gives:

– <1.6 mW/m2

– Ability to resolve 21 K black body– 20 mK temperature change in a 70 K black body

Sensitivity Expected

Page 5: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

5

Optical and Electrical Calibration

Optical calibration performed by inserting the input end of four radiometers into a black box which can be controlled at various temperatures

In the same configuration the radiometers can be calibrated by self heating the thermometers

Page 6: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

6

Measure performance of the CoCOA shutter and baffle Measure stray light at chamber doors and “gaps” Measure Photogrammetry Camera emission if possible Measure other potential heat sources – cabling, wall

penetrations, etc.

Objectives

Page 7: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

7

Typical Results

Typical output of 9 radiometers in Chamber A test Negative values can indicate that source is colder than radiometer or that the thermometer

calibrations do not match

Page 8: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

8

Results-Chamber Ceiling

R1 R2 R3

R1 shows output of CoCOA with shutter open and closedR2 does not show any particular heat loadR3 does not see open CoCOA and indicates that CoCOA baffle is working

Page 9: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

9

Results-CoCOA Shutter

CoCOA shutter works well and comes to thermal equilibrium quickly

Page 10: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

10

Results-Photogrammetry Cameras

R3 and R4 pick up changes due to PG Camera position changes – will investigate in next test

Page 11: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

11

Results-Stray Thermal Energy

Warm proof mass thermal signature seen in reflection– Periodic variation is proportional to temperature fluctuations

Page 12: 1 Using In-situ Cryogenic Radiometers to Measure the Performance of a Large Thermal Vacuum Chamber Michael DiPirro 1, Edgar Canavan 1, James Tuttle 1,

12

Future Tests

Optical GSE1completed

Optical GSE2to start Sept. 2015

Thermal Pathfinderearly 2016

OTISearly 2017

Pathfinder structure with 2 mirror segments,and secondary mirror deployed ready to be rolled into Chamber A