1-s2.0-s0958694601000395-main
TRANSCRIPT
-
7/31/2019 1-s2.0-S0958694601000395-main
1/6
International Dairy Journal 11 (2001) 9398
Moisture, solids-non-fat and fat analysis in butter
by near infrared spectroscopy
Mar!a Hermidaa, Jose M. Gonzaleza, Mar Sanchezb, Jose L. Rodriguez-Oterob,*aLaboratorio de Mouriscade, Diputaci!on Provincial de Pontevedra, 36500 Lal n (Galicia), Spain
bFacultad de Veterinaria, Instituto de Investigaci!on y An !alisis de Alimentos, Universidad de Santiago, 27002 Lugo (Galicia), Spain
Received 10 July 2000; accepted 25 February 2001
Abstract
Moisture, solids-non-fat and fat were analysed in butter by near infrared (NIR) spectroscopy without any previous sample
treatment. A set of 102 samples was used to calibrate the instrument by modified partial least-squares regression. The following
statistical results were achieved: standard error of calibration (SEC)=0.192 and square correlation coefficient R2 0:90 formoisture, SEC=0.086 and R2 0:93 for solids-non-fat, and SEC=0.168 and R2 0:94 for fat. To validate the calibrationperformed, a set of 24 butter samples was used. Standard errors of validation were 0.255, 0.071 and 0.377 for moisture, solids-non-
fat and fat, respectively, and R2 for the regression of measurements by reference method versus NIR analysis were 0.83, 0.94 and
0.72 for moisture, solids-non-fat and fat, respectively. To compare, for all three components of the validation set, the results
obtained by NIR spectroscopy with those obtained by the reference methods, linear regression and paired t-test were applied
and the methods did not give significantly different results for p 0:05. When mean square prediction error analysis was applied, itcould be concluded that a strong calibration model was obtained and that for all three components the main error was unexplained.
# 2001 Elsevier Science Ltd. All rights reserved.
Keywords: NIR spectroscopy; Butter analysis
1. Introduction
At present, with the great increase in food manufac-
ture and trade, a large number of analyses of both raw
materials and finished products are in demand. The
classical analytical methods are slow, expensive and
need highly qualified staff; therefore these methods are
not sufficiently effective in satisfying the current needs of
the food industry.
The main advantages of near infrared (NIR) spectro-
scopy for food analysis lies in its speed, no or little
sample pre-treatment and the avoidance of the use of
chemicals (Osborne, Fearn, & Hindle, 1993). NIR
spectroscopic analysis of milk (Albanell, C !aceres, Caja,
Molina, & Gargouri, 1999; Laporte & Paquin, 1999),
cheese (Rodr!guez-Otero, Hermida, & Cepeda, 1995;
Sorensen & Jepsen, 1998a, b; Wittrup & Norgaard,
1998) and other dairy products (Rodr!guez-Otero &
Hermida, 1996; Rodr!guez-Otero, Hermida, & Centeno,
1997; Laporte & Paquin, 1998) is a very useful technique
for the determination of major components in milk and
dairy products. However, little work has been done in
butter analyses; in the literature reviewed only one
article was found (Weaver, 1984). Measurements at 20
different wavelengths with equipment fitted with optical
filters were described in this article, and calibration was
performed by multiple linear regression (MLR) for
moisture. Evaluation of data on fat and solids-non-fat
was not, however, included.
Moisture determination is the most important control
in butter production, especially in modern factories
where hundreds of tons may be produced daily. The
economy of manufacture demands that the final
products be as near as possible to the target value. But
in addition to moisture, solids-non-fat and fat limits are
specified in butter regulations; therefore these two
parameters must be analysed routinely.
Since studies on butter analysis with modern full
scanning instruments have not been found in the
literature, the aim of this work is the application of
NIR spectroscopy to the analysis of all three major*Corresponding author. Tel.: +34-982-252231.
E-mail address: [email protected] (J.L. Rodriguez-Otero).
0958-6946/01/$ - see front matter # 2001 Elsevier Science Ltd. All rights reserved.
PII: S 0 9 5 8 - 6 9 4 6 ( 0 1 ) 0 0 0 3 9 - 5
-
7/31/2019 1-s2.0-S0958694601000395-main
2/6
components of butter: moisture, solids-non-fat and fat,
without any previous sample treatment. We also study
the effectiveness of different multivariate analysis
techniques (Martens & Naes, 1993).
2. Materials and methods
2.1. Samples
A total of 126 butter samples were divided into two
groups: 102 samples for calibration and 24 samples for
validation of the calibration performed. Samples of
salted as well as unsalted butter were included in both
groups. Samples were supplied by the principal factories
in, and purchased from the markets of Galicia (NW
Spain) and Portugal.
2.2. Reference analysis
Water, solids-non-fat and fat contents were deter-
mined in accordance with International Dairy Federa-
tion (IDF) Standard 80 (1977). The water content was
determined by drying a known mass of butter at
102 28C and weighing it afterwards to determine the
loss of mass. The solids-non-fat content was determined
after extracting the fat from the dried butter with light
petroleum (boiling range between 308C and 608C) and
weighing the residue. The fat content was calculated as a
percentage by subtracting the percentages of water
content and solids-non-fat content from 100.
2.3. NIR analysis
A wavelength scanning instrument, NIRSystems 6500
(NIRSystems, Silver Spring, MD), with a scanning
range from 400 to 2500 nm and wavelength increments
of 2 nm was used. Instrument checks recommended by
the manufacturer were performed daily prior to use.
Samples were analysed at room temperature (approxi-
mately 208C) in a closed 30 mm thick cuvette, using the
transport module. Reflectance measurements of mono-
chromatic light were made from 1108 to 2492 nm. The
means of 25 spectra scans were taken for each sample;data was recorded as log 1=R, where R is the reflectanceenergy.
2.4. NIR calibration
ISI software was used (ISI software, 1992). Scatter
correction was performed by standard normal variate
transformation (SNV) and detrend method (Barnes,
Dhanoa, & Lister, 1989) and by multiplicative scatter
correction (MSC) (Geladi, McDougall, & Martens,
1985).
A general Mahalanobis distance (H statistic) was
calculated from principal components analysis (PCA)
scores. The H values were standardized by dividing
them by the average H value for the calibration file. If
a new spectrum sample differed by more than 3.0
standardized units from the mean spectrum of the
calibration file, the sample was defined as a global Houtlier, liable to give inaccurate predictions.
The calibrations were performed by MLR, principal
components regression (PCR) and modified partial
leasts-square (MPLS) regression (Martens & Naes,
1993) using first and second derivatives of the spectra
(Meuret, Dardenne, Biston, & Poty, 1993).
* The first derivative of the spectra was calculated
using a subtraction gap and smoothing segment of 4
data points (1, 4, 4).* The second derivative of the spectra was calculated
using a subtraction gap and smoothing segment of 6
data points (2, 6, 6).
The optimal number of terms for the calibration
minimising overfitting was based on the standard error
of cross-validation (SECV). The approach used was as
follows: 80% of the samples from the calibration set
were used for calibration, and for the remaining 20%
the standard error of prediction (SEP) was calculated.
This operation was carried out a total of 5 times, each
time using a different group for calibration and
prediction. The SECV was calculated as the square root
of the average of the squares of the 5 SEP. The final
calibration equation was developed with the totalsamples of the calibration set using the number of
factors with the lowest SECV.
The standard error of calibration (SEC) was calcu-
lated, and the critical T value for eliminating outliers
was fixed at 2.5 (T=residual/SEC).
To check the calibration performed the validation set
was used. The standard error of validation (SEV) and
square correlation coefficient (R2) of reference analyses
versus NIR values were calculated. Statistical errors
were calculated in accordance with Workman (1992).
3. Results and discussion
3.1. Spectral information and butter samples composition
Fig. 1 shows the mean spectra of butter in the
calibration set. The following bands can be observed:
at 1210 nm the second overtone of the CH stretch of
CH2 groups; at 1450 nm the first overtone of the OH
stretch; at 1728 and 1764 nm the first overtone of the
CH stretch of CH2 groups; at 1940 nm a combination
of the second overtone of the OH bend and stretching
band of water; at 2310 nm the second overtone of the
M. Hermida et al. / International Dairy Journal 11 (2001) 939894
-
7/31/2019 1-s2.0-S0958694601000395-main
3/6
CH bend of CH2 groups; at 2350 nm a stretch and bend
combination band of CO (Osborne et al., 1993).
Table 1 shows the range of chemical composition and
standard deviation (Sd) of butter samples in the
calibration and validation sets.
3.2. Repeatability
Repeatability standard deviation (Sr) was calculatedfor reference methods and for NIR spectroscopy in
accordance with IDF Standard 128 (1985) by analysing
in duplicate 12 butter samples. Table 2 shows that for
the repeatability standard deviation better values were
achieved in the NIR spectroscopy analysis than by using
the reference methods. This is certainly due to the
simplicity of the instrumental method in comparison
with the reference methods.
3.3. Calibration
The calibration set was selected with the aim of
achieving a strong calibration for butter with and
without added salt by maximising variability among
samples composition and obtaining a wide range of
spectra to avoid H outliers in the validation set.
Therefore, the mean value for H statistic for the
validation set of samples was 1.15 with respect to the
mean value of the calibration set and no H outliers
were found. Consequently, no statistical differences were
found between the spectra of the validation and
calibration sets, indicating that the calibration set was
wide enough.
In order to evaluate the different calibrations, the
SEC, SECV and R2 of the calibration set, and the bias,
SEV, R2 and mean square prediction error (MSPE) of
the validation set were evaluated. Although 102 samples
were used for calibration, T outliers were eliminated
and the final number of samples used to calibrate each
parameter is shown in Table 3.
MPLS regression achieved the best calibration equa-
tions in all cases. Relatively small differences were found
for the statistical results of calibrations, when compar-
ing the use of SNV and detrend method with MSC for
scatter correction of radiation and when using first or
second derivative of the spectra. Therefore SNV and
detrend and first derivative were chosen for all three
calibrations, with the aim of simplifying the discussion.
Fig. 1. Mean near infrared spectrum of butter samples.
Table 1
Range of chemical composition (% w/w) and Sd of butter in the
calibration and validation sets
Set Moisture Solids-non-fat Fat
Range Sd Range Sd Range Sd
Calibration 13.0117.05 0.675 1.143.12 0.321 81.0985.52 0.761
Validation 13.8616.44 0.604 1.292.60 0.282 81.3184.34 0.617
Table 2
Repeatability standard deviation (% w/w) of determination of
moisture, solids-non-fat and fat by reference and NIR methods
Method Moisture Solids-non-fat Fat
Reference 0.052 0.060 0.085
NIR 0.028 0.012 0.023
M. Hermida et al. / International Dairy Journal 11 (2001) 9398 95
-
7/31/2019 1-s2.0-S0958694601000395-main
4/6
Table 3 shows the number of PLS terms, SEC, SECV
and R2 values for moisture, solids-non-fat and fat for
the calibration set. The number of PLS terms is not too
high, less than 1 per 10 samples of the calibration set.
Therefore, no overfitting should be expected, the SEC
values are only about 20% lower than those of the
SECV for moisture and solids-non-fat and 30% for fatR2 values were between 0.90 and 0.94. These figures can
be considered acceptable, bearing in mind that butter is
a product with a very narrow range of composition
variation, owing to the high level of standardisation in
butter factories.
The best values for the SEV and R2 of the validation
set (Table 4) were those obtained for solids-non-fat,
perhaps because of its higher variation range due to the
inclusion of samples of salted butter, which helps to
broaden the range of this parameter. On the other hand,
the worst SEV and R2 values were those of fat, in our
opinion, due to the fact that in the fat determination by
the reference method two errors are accumulated, as a
consequence of subtracting the percentages of water and
solids-non-fat from 100. The bias value is low for
moisture and zero for solids-non-fat and fat.
To compare the results obtained by NIR spectroscopyfor all three components of the validation set with those
obtained by the reference methods, linear regression and
paired t-test were applied (IDF, Standard 128, 1985):
(i) When calculating the slope and intercept of NIR
values versus reference values, no statistical differ-
ences (p 0:05) were found from the theoreticalvalues 1.00 and 0.00, respectively.
(ii) The calculated t values were lower than the
theoretical t values (p 0:05).
Therefore the null hypothesis was retained: the twomethods did not give significantly different results.
Graphic comparisons between reference values and
NIR predicted values of the global validation set are
shown in Figs. 24.
The MSPE is the sum of three types of errors
(Dhanoa, Lister, France, & Barnes, 1999): errors in
central tendency, errors due to regression and errors
due to uncontrolled disturbance or unexplained errors
(Table 5). Errors in central tendency are also known
as mean bias; this error is very small in percentage
terms for all three components analysed; the highest
value was for moisture, where the bias value was also the
highest.
The errors linked to regression will be equal to zero
when the slope of regression is unity. This error also
accounts for a small proportion of the MSPE and in the
Table 3
Statistical data for the calibration set (% w/w)
Component N PLS terms SEC R2 SECV
Moisture 96 6 0.192 0.90 0.242
Solids-non-fat 99 8 0.086 0.93 0.103
Fat 93 8 0.168 0.94 0.240
Table 4
Statistical data for the validation set (% w/w)
Component N Bias SEV R2
Moisture 24 0.04 0.255 0.83
Solids-non-fat 24 0.00 0.071 0.94
Fat 24 0.00 0.377 0.72
Fig. 2. Validation set: moisture content. Reference method versus NIR.
M. Hermida et al. / International Dairy Journal 11 (2001) 939896
-
7/31/2019 1-s2.0-S0958694601000395-main
5/6
case of fat, where the slope is not far from unity, for
only 0.40%. Therefore it can be stated that a strong
calibration model was obtained.
The unexplained error accounts for a high percentage
of the MSPE, more than 87% for all three components
and for fat even nearly 100%.
Fig. 3. Validation set: solids-non-fat content. Reference method versus NIR.
Fig. 4. Validation set: fat content. Reference method versus NIR.
Table 5
Mean square prediction error and its components (percentage in parenthesis) calculated in accordance with Dhanoa et al. (1999)
Component MSPE Errors in central tendency Errors due to regression Unexplained error
Moisture 0.0680 1.610-4 0.0070 0.0600
(2.35%) (10.31%) (87.34%)
Solids-non-fat 0.0053 6.2510-6 0.0007 0.0046
(0.12%) (12.71%) (87.17%)
Fat 0.1300 2.110-5 0.0005 0.1300
(0.02%) (0.40%) (99.58%)
M. Hermida et al. / International Dairy Journal 11 (2001) 9398 97
-
7/31/2019 1-s2.0-S0958694601000395-main
6/6
4. Conclusions
NIR spectroscopy is an adequate technique for the
analysis of moisture, solids-non-fat and fat in butter
without any previous sample treatment. Owing to the
simplicity of the instrumental method, better values of
repeatability were achieved by NIR spectroscopy thanby the reference methods. The best calibrations were
performed through MPLS regression. After MSPE
analysis it can be concluded that a strong calibration
model was obtained and that the main source of error
was an unexplained error, accounting for more than
87% for all three components.
References
Albanell, E., C!aceres, P., Caja, G., Molina, E., & Gargouri, A. (1999).
Determination of fat, protein and total solids in ovine milk bynear-infrared spectroscopy. Journal of AOAC International, 82,
753758.
Barnes, R. J., Dhanoa, M. S., & Lister, S. J. (1989). Standard normal
variate transformation and detrending of near infrared diffuse
reflectance spectra. Applied Spectroscopy, 43, 772777.
Dhanoa, M. S., Lister, S. J., France, J., & Barnes, R. J. (1999). Use of
mean square prediction error analysis and reproducibility measures
to study near infrared calibration equation performance. Journal of
Near Infrared Spectroscopy, 7, 133143.
Geladi, P., McDougall, D., & Martens, H. (1985). Linearization and
scatter correction for near infrared reflectance spectra of meat.
Applied Spectroscopy, 39, 491500.
IDF Standard 80. (1977). Butter}determination of water, solids-non-
fat and fat contents on the same test portion. Brussels, Belgium.
IDF Standard 128. (1985). Definition and evaluation of the overallaccuracy of indirect methods of milk analysis}application to
calibration procedure and quality control in dairy laboratory.
Brussels, Belgium.
ISI Software. (1992). ISI NIRS 2: Software for near infrared
instruments. Intrasoft International, Port Matilda, PA, USA.
Laporte, M. F., & Paquin, P. (1998). Near-infrared technology and
dairy food products analysis: A review. Seminars in Food Analysis,
3, 173190.
Laporte, M. F., & Paquin, P. (1999). Near-infrared analysis of fat,
protein and casein in cows milk. Journal of Agricultural and Food
Chemistry, 47, 26002605.Martens, H., & Naes, T. (1993). Multivariate calibration. Chichester,
UK: Wiley.
Meuret, M., Dardenne, P., Biston, R., & Poty, O. (1993). The use of
NIR in predicting nutritive value of Mediterranean tree and shrub
foliage. Journal of Near Infrared Spectroscopy, 1, 4554.
Osborne, B. G., Fearn, T., & Hindle, P. H. (1993). Practical near
infrared spectroscopy with applications in food and beverage analysis
(pp. 2833). Harlow, UK: Longman Scientific & Technical.
Rodr!guez-Otero, J. L., & Hermida, M. (1996). Analysis of fermented
milk products by near-infrared reflectance spectroscopy. Journal of
AOAC International, 79, 817821.
Rodr!guez-Otero, J. L., Hermida, M., & Centeno, J. (1997). Analysis
of dairy products by near infrared spectroscopy: A review. Journal
of Agricultural and Food Chemistry, 45, 28152820.
Rodr!
guez-Otero, J. L., Hermida, M., & Cepeda, C. (1995).Determination of fat, protein, and total solids in cheese by near-
infrared spectroscopy. Journal of AOAC International, 78, 802806.
Sorensen, L. K., & Jepsen, R. (1998a). Comparison of near infrared
spectroscopy techniques for determination of semi-hard cheese
constituents. Milchwissenschaft, 53, 263267.
Sorensen, L. K., & Jepsen, R. (1998b). Assessment of sensory
properties of cheese by near-infrared spectroscopy. International
Dairy Journal, 8, 863871.
Weaver, R. W. V. (1984). Near infrared reflectance analysis applied to
dairy products. In Abstract of papers; Challenge to contemporary
dairy analytical technique. Reading, UK: Royal Society of
Chemistry.
Wittrup, Ch., & Norgaard, L. (1998). Rapid near infrared spectro-
scopy screening of chemical parameters in semi-hard cheese using
chemometry. Journal of Dairy Science, 81, 18031809.Workman (1992). NIR spectroscopy calibration basis. In D. A. Burns,
E. W. Gurczak (Eds.), Handbook of near-infrared analysis
(pp. 270278). New York: Marcel Dekker Inc.
M. Hermida et al. / International Dairy Journal 11 (2001) 939898