1 marlap science advisory board chapters 6 & 7 stan morton doe-resl april 2002

40
1 MARLAP SCIENCE ADVISORY BOARD CHAPTERS 6 & 7 Stan Morton DOE-RESL April 2002

Upload: betty-carpenter

Post on 28-Dec-2015

214 views

Category:

Documents


0 download

TRANSCRIPT

1

MARLAPSCIENCE ADVISORY BOARD

CHAPTERS 6 & 7

Stan Morton

DOE-RESL

April 2002

2

Multi-Agency Radiation Laboratory Protocols Manual

Purpose:Provide guidance and a framework for project planners, managers, and laboratory personnel to ensure that radioanalytical laboratory data will meet a project’s or program’s data requirements and needs.

3

Chapter 6 - Selection and Application of an Analytical Method

• MARLAP recommends the performance-based approach to method selection

– Lab selects and proposes a method in response to the APSs and method validation level specified in the SOW

• method selection is a complex process that must consider:

– APSs - MQOs, method validation status, qualified staff availability, production schedule, radiological and sample TATs, equipment calibration / availability, etc.,

– Project manager approves use of selected method• evaluates submitted method validation documentation or

• evaluates performance of lab re analysis of method validation PT samples

4

Measurement Quality Objectives

• MQOs - analytical portion of the DQOs

– statement of performance or requirement for a particular method performance characteristic - can be quantitative or qualitative

• method uncertainty, uMR, at a specified concentration (expressed as an estimated standard deviation)

• quantification capability (expressed as the minimum quantifiable concentration - MQC)

• detection capability (expressed as the minimum detectable concentration - MDC)

• applicable analyte concentration range - method’s ability to measure analyte over some specified range

5

Measurement Quality Objectives

• MQOs - analytical portion of the DQOs

– statement of performance or requirement for a particular method performance characteristic - can be quantitative or qualitative

• method specificity - ability of method to measure the analyte in the presence of interferences

• method ruggedness - relative stability of method performance for small variations in method parameter values

6

Chapter Six: Selection and Application of an Analytical Method

6.1 Introduction6.2 Method Definition

6.2.1 Performanced-Based Approach and Application6.3 Life Cycle of Method Application6.4 Generic Considerations for Method Devel/Selection6.5 Project Specific Considerations for Method Selection

6.5.1 Matrix and Analyte Identification6.5.2 Process Knowledge6.5.3 Radiological Holding and Turn Around Times6.5.4 Unique Processing Specifications

7

6.5.5 Measurement Quality Objectives6.5.5.1 Method Uncertainty6.5.5.2 Quantification Capability6.5.5.3 Detection Capability6.5.5.4 Applicable Analyte

Concentration Range6.5.5.5 Method Specificity6.5.5.6 Method Ruggedness6.5.5.7 Method Bias

Considerations

8

6.6 Method Validation6.6.1 Purpose of Method Validation6.6.2 Laboratory’s Method Validation

Protocol6.6.3 Tiered Approach to Validation

6.6.3.1 Existing Methods6.6.3.2 Validation for Similar Matrices6.6.3.3 New Application of a Validated

Method6.6.3.4 Newly Developed or Adapted

Methods6.6.4 Method Validation Documentation

9

6.7 Analyst Qualifications and Proficiency6.8 Method Control6.9 Continued Performance Assessment6.10 Documentation Sent to Project Manager

Summary of Recommendations

6.11 References

10

F ie ld S am p le P rep aratio nan d P reserv atio n

S am p le D isso lu tio n

C h em ical S ep aratio n

S am p le P rep aratio nfo r In stru m en t M easu rem en t

In stru m en t M easu rem en t o fR ad io n u cl id es

A n aly t ica l C alcu latio n s an dD ata R ed u ctio n

S am p le R eceip tan d T rackin g

L ab o rato ry S am p leP rep aratio n

D ata V eri f icatio n , V al id atio nan d R ep o rtin g

These Steps AreTypically C onsideredto be "The Method"

M ay be Included

11

12

MQO - Method Uncertainty Requirement, uMR

• The recommended value of uMR is based on the assumption that any known bias in the measurement process has been corrected and any remaining bias is much smaller than the shift, , when a concentration near the gray region is measured.

– MAPEP & QAP PE programs should measure laboratory bias as a testing parameter - bias is determined through multiple analyses, not a single measurement

• If decisions are to be made about the mean of a sampled population, estimate the uMR of the analytical method at the UBGR (action level) and require that the uMR be the width of the gray region divided by 10.

• If this requirement can not be met, then require at least that the uMR be the width of the gray region divided by 3.

13

14

ExampleThe project planners have identified that the action level for the analyte is 0.10 Bq/g and the lower boundary of the gray region is 0.02 Bq/g. If decisions are to be made about the contaminated areas based on samples, then the uMR at 0.10 Bq/g would be:

uMR = / 10 = (0.10 - 0.02) / 10 = 0.008 Bq/g (8%)

If this uMR cannot be achieved, then a method uncertainty as large as

/ 3 = 0.027 Bq/g (27%) may be allowed if more samples are taken per contaminated survey area.

In terms of method selection, this MQO calls for a method that can ordinarily produce measured results with an expected combined standard uncertainty of 8% at the action level

15

is defined as DCGL - LBGR

is the standard deviation of the measured analyte distribution for the area being remediated

2 is mathematically defined as (S2 + M

2) where S2 is the variance

of sampled population (analyte) and M2 is average analytical

method (measurement) variance

M is effected by laboratory sample preparation, subsampling, radiochemical and radiometric processes

/ is defined as the relative shift - number of standard deviations separating the DCGL and the LBGR

Non-parametric tests are used to determine if the null hypothesis (survey unit does not meet the release criterion) can be rejected; Sign and Wilcoxon Rank Sum Tests.

16

17

MQO - Method Uncertainty Requirement

If decisions are to be made about individual samples in a survey, MARLAP recommends the use of:

uMR = / ( z1- + z1- )

Example: UBGR = 1 Bq/L; LBGR = 0.5 Bq/L; = 0.05; = 0.10

uMR = 1.0 - 0.5 / ( 1.645 + 1.282 ) = 0.17 Bq/L

or 17% at the action level.

If = 0.05; = 0.05, then the following simplification can be used

uMR = 0.3

18

MQO - Method Uncertainty RequirementFlexibility in Data Review

• When Project Planners establish the MQO for method uncertainty for method selection and development, the maximum allowable standard deviation, uMR at the UBGR is specified.

• During data evaluation, the measurement uncertainty at any sample analyte concentration UBGR should not exceed uMR

• During data evaluation, the measurement uncertainty at any sample analyte concentration > UBGR should not exceed uMR /

UBGR or MR

19

MQO - Detection Capability

• If the lower bound of the gray region is at or near zero and decisions are to be made about individual samples, choose an analytical method whose MDC is no greater than the upper bound of the gray region (action level).

– LBGR = 0; = UBGR

– uMR UBGR / ( z1- + z1- )

– uMR * (z1- + z1- ) UBGR

– Form of: MDC UBGR

20

TABLE 6.1 Tiered Method Validation Approach.

Validation Level Application SampleType

AcceptanceCriteria*

Levels(Concentrations)** Replicates # of Analyses

A Without

AdditionalValidation

ExistingValidatedMethod

-

Method Previously Validated(By One of the Validation

LevelsB though H)

- - -

B Similar Matrix Internal QC 3 3 9

C Similar Matrix

/ NewApplication

External PE 3 7 21

D ASTM D2777

NewApplication

InternalQC

Three to Five Groups of Two SamplesWith Concentrations Within 20% of Each

Other6-10

E ASTM D2777

NewApplication External PE

Three to Five Groups of Two SamplesWith Concentrations Within 20% of Each

Other6-10

F EPA Equivalency

NewApplication /

NewlyDeveloped or

AdaptedMethod

MVRMSamples 3 7 21

G ASTM D2777

MVRMSamples

Measured Value Within 3uMR of Known Value

Three to Five Groups of Two SamplesWith Concentrations Within 20% of Each

Other6-10

Measured Value Within 3uMR of Known Value

Three to Five Groups of Two SamplesWith Concentrations Within 20% of Each

Other6-10H

ASTM D2777Involves the TwoTesting Protocols

Stated to the Right

NewlyDeveloped

orAdaptedMethod

MVRMSamples for

BothProtocols

Each Measured Value 30% ofKnown

at 5 times MDC

Three to Five Groups of Two SamplesWith Concentrations Within 20% of Each

Other Bracketing 5 Times the MDC6-10

* Assumes that each sample is counted to have a Poisson counting uncertainty of < 3% (sigma) when the analyte concentration is near the actionlevel or MQC. This criterion is applied to each analysis in the method validation, not to the mean of the analyses. uMR is the required methoduncertainty at the action level or required concentration. uMR is an absolute value for concentrations less than the action level and a relative (%)value for concentrations greater than the action level. In the absence of a specified value, the default of 3 uMR acceptance criterion is: eachmeasured value at the action level or other specified concentration must be within 30% of known value. See references for ASTM D2777.**Concentration levels should cover the expected analyte concentration range for a project including the action level concentration. A set of threeblanks (not considered a level) should be analyzed during the method validation process.

21

Method Validation Process

• Parameters specified or ascertained from the analytical results generated

– Defined Method Validation Level (Table 6.1)

– Analytes

– Defined matrix for testing, including chemical and physical characteristics that approximate project samples or

– Selected project specific or appropriate alternative matrix PT samples, including known chemical or radionuclide interferences at appropriate levels

– Defined sample preservation

– Stated additional data testing criteria

– Establish acceptable chemical / radiotracer yield values

22

Method Validation Process

• Parameters specified or ascertained from the analytical results generated

• APSs including MQOs for each analyte / matrix

– Chemical or physical characteristics of analyte when appropriate

– Action level when applicable

– Applicable analyte concentration range including zero analyte (blanks)

– Method uncertainty at a specific concentration

– MDC or MQC

– Bias if applicable

– Other qualitative parameters to measure the degree of method ruggedness or specificity

23

Chapter 7 - Evaluating Analytical Methods and Laboratories

• First part of chapter discusses the evaluation of the documentation that the lab sends on the proposed method, method validation and performance in PE programs– Follow up on Chapter 6 on Selection and Application of an

Analytical Method

• Second part of the chapter discusses the initial and ongoing evaluation of lab services– Follow up on Chapter 5 and Appendix E on Obtaining

Laboratory Services

24

CHAPTER 7EVALUATING ANALYTICAL METHODS AND

LABORATORIES

• 7.1 Introduction

• 7.2 Evaluation of Proposed Analytical Methods– 7.2.1 Documentation of Required Method Performance

• 7.2.1.1 Method Validation Documentation

• 7.2.1.2 Method Experience, Previous Projects, and Clients

• 7.2.1.3 Internal and External Quality Assurance Assessments

– 7.2.2 Performance Requirements of the Sow – Analytical Protocol

• 7.2.2.1 Matrix and Analyte Identification

25

CHAPTER 7 CONTINUEDEVALUATING ANALYTICAL METHODS AND

LABORATORIES

• 7.2.2.2 Process Knowledge

• 7.2.2.3 Radiological Holding and Turnaround Times

• 7.2.2.4 Unique Processing Specifications

• 7.2.2.5 Measurement Quality Objectives

• 7.2.2.6 Bias Considerations

• 7.3 Initial Evaluation of the Laboratory– …

– 7.3.4 Review of Performance Indicators• 7.3.4.1 Review of Internal QC Results

• 7.3.4.2 External PE Program Results

26

Cross Reference of Information Available for Method Evaluation.

EvaluationElement

AddressedMethodValidation

Internal andExternal QCReports

External PEPrograms

Internal /External QAAssessments

Informationfrom RFP andOther Sources

Analyte / Matrix

ProcessKnowledge

PreviousExperience

Rad. HoldingTime

TurnaroundTime

denotes that the information relevant to method evaluation may be

27

Cross Reference of Information Available for Method Evaluation.

EvaluationElement

AddressedMethodValidation

Internal andExternal QCReports

External PEPrograms

Internal /External QAAssessments

Informationfrom RFP andOther Sources

Unique ProcessSpecifications

Bias

MethodUncertainty(MQC) / MDC

Analyte /InterferenceRange

MethodRuggedness

MethodSpecificity

denotes that the information relevant to method evaluation may bepresent.

28

GENERIC SPECIFICATIONSFOR LABORATORY

OPERATIONS

EXPERIENCE BASE FOR SOWANALYTICAL REQUIREMENTS

RADIOANALYTICAL METHODS APPLICABILITY / QUALITY

INITIAL EVALUATION OF LABORATORY

CAPABILITY TO MEET SAMPLEPROCESSING AND REPORTING

REQUIREMENTS

CAPABILITY TO MEET SOWSPECIFICATIONS

EXISTING PERFORMANCEINDICATOR REVIEW

ADEQUACY OF FACILITIES,INSTRUMENTATION AND STAFF

LEVELS

QUALITY SYSTEMDOCUMENTATION

QUALITY MANUAL

Organization and ManagementQuality Sytem - Establishment, Audits,

Essential Quality Controls andEvaluations, and Data Verification

PersonnelPhysical Facilities - Accomodations and

EnvironmentEquipment and Reference Materials

Measurement Traceability and CalibrationsMethod and Standard Operating

ProceduresSample Handling, Sample Acceptance

Policy and Sample ReceiptRecords

Subcontracting Analytical SamplesOutside Support Services and Supplies

Complaints

PROFICIENCY TESTINGSAMPLES

METHODVALIDATION

DOCUMENTATION

DEMONSTRATEDQUALITY ON

SIMILAR PROJECTS

METHOD REVIEW& COMPARABILITY

EXTERNAL PE PROGRAMRESULTS

ONSITEAUDIT

DESK AUDIT

ADMINISTRATIVE / PROJECTMANAGEMENT

29

MQO - Method Uncertainty RequirementFlexibility in Data Review

• When Project Planners establish the MQO for method uncertainty for method selection and development, the maximum allowable standard deviation, uMR at the UBGR is specified.

• During data evaluation, the measurement uncertainty at any sample analyte concentration UBGR should not exceed uMR

• During data evaluation, the measurement uncertainty at any sample analyte concentration > UBGR should not exceed uMR /

UBGR or MR

30

Monitoring A Lab’s Quantitative Performance

• Premise

– Ensuring the Lab meets the method uncertainty requirement

• uMR = / 10

MR = uMR / UBGR

• No method bias is assumed

31

Use of Internal and External QC Samples

Laboratory Control Samples

% D = 100 * (SSR - SA) / SA

where SSR is measured result and SA is the spike value. Assumes the uncertainty in SA is negligible compared to the uncertainty in SSR.

Warning Limits: ( 2 MR ) * 100

Control Limits: ( 3 MR ) * 100

Plot on control chart for trending - no action based on single measurement

32

Laboratory Control Samples - Example

UBGR= 5 Bq/kg, uMR = 0.35 Bq/kg, MR = 0.07

LCS is prepared with SA = 10.0 Bq/kg and analytical result of SSR = 11.61 0.75 Bq/kg. % D = 100 * (SSR - SA) / SA = 100 * (11.61 - 10.0) / 10 = 16.1

Warning Limits: ( 2 MR ) * 100 or 14 %Control Limits: ( 3 MR ) * 100 or 21 %

% D is above the warning limit but below the control limit

33

Use of Internal and External QC Samples

Duplicate Samples - assuming significant activity

Xav = ( X1 + X2 ) / 2

When Xav < UBGR; test the statistic | X1 - X2 |

Warning Limit: 2.83 * uMR

Control Limit: 4.24 * uMR

Plot on control chart for trending - no action based on single measurement

34

Use of Internal and External QC Samples

Duplicate Samples - assuming significant activity

Xav = ( X1 + X2 ) / 2

When Xav > UBGR;

Test the statistic RPD = | X1 - X2 | * 100 / Xav

Warning Limit: 2.83 * MR Control Limit: 4.24 * MR

Plot on control chart for trending - no action based on single measurement

35

Use of Internal and External QC Samples

Method Blanks - testing criteria are provided.

However, the target value for an analytical blank is zero. For the sample specific MDC, the critical level should be used to determine if a blank is statistically positive.

Testing Criteria Related to the uMR:

Test the statistic: Measured Concentration ValueWarning Limits: 2 uMR

Control Limits: 3 uMR

Plot on control chart for trending - no action based on single measurement

36

Method Blanks - Example

UBGR= 5 Bq/kg, uMR = 0.35 Bq/kg, MR = 0.07

Analytical Result: X = 0.20 0.10 Bq/kg

Testing Criteria Related to the uMR:

Warning Limits: 2 uMR or 0.70 Bq/kgControl Limits: 3 uMR or 1.05 Bq/kg

Analytical result is below the warning limit. Although the test allows for a certain degree of contamination in comparison to the action level, the target value is zero and any blank value > the critical vale would be considered to be positive

37

Use of Internal and External QC Samples

Matrix Spikes - testing criteria are provided.

Test the statistic: Z

Z = ( SSR - SR - SA ) / {MR ( SSR2 + max(SR, UBGR)2)1/2}

where SSR sample result, SA is the spike value, SR is the unspiked value. (Assumes the uncertainty of the spike value is insignificant compared to the uncertainty of the measurements)

Warning Limits: 2 Control Limits: 3

Plot on control chart for trending - no action based on single measurement

38

Matrix Spike ExampleUBGR= 5 Bq/kg, uMR = 0.35 Bq/kg, MR = 0.07SR = 3.5 0.29 Bq/kg, SA = 10.1 0.31 Bq/kg, SSR = 11.2 0.55 Bq/kg. Since SR is less than UBGR, max(SR, UBGR) = UBGR = 5 Bq/kg.

Z = ( SSR - SR - SA ) / {MR ( SSR2 + max(SR, UBGR)2)1/2}

Z = ( 11.2 - 3.5 - 10.1 ) / ( 0.07 * ( 11.22 + 52 )1/2 = - 2.80

Warning Limits: 2 Control Limits: 3

Z is less than the lower warning limit (-2) but slightly greater than the lower control limit (-3)

39

MQO - Method Uncertainty RequirementFlexibility in Data Review

• When Project Planners establish the MQO for method uncertainty for method selection and development, the maximum allowable standard deviation, uMR at the UBGR is specified.

• During data evaluation, the measurement uncertainty at any sample analyte concentration UBGR should not exceed uMR

• During data evaluation, the measurement uncertainty at any sample analyte concentration > UBGR should not exceed uMR /

UBGR or MR

40

Example - Evaluation of Data Below and Above the UBGR

UBGR= 1 Bq/L, LBGR = 0.5 Bq/L, uMR = 0.17 Bq/L, MR = 0.17 or 17%

Any result 1 Bq/L should have a measurement ( Combined Standard ) uncertainty no more than 0.17 Bq/L.

Any result > 1 Bq/L should have a relative combined standard uncertainty no greater than 17%.