1. introduction - shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/6423/7/07... · 2015. 12....
TRANSCRIPT
1 INTRODUCTION
1. INTRODUCTION
1.1 ANALYTICAL CHEMISTRY
Analytical Chemistry(1)
seeks ever improved means of measuring the chemical
composition of natural and artificial materials. The techniques of this science are used to
identify the substances which may be present in a material and to determine the exact
amounts of the identified substances. Analytical chemists work to improve the reliability
of existing techniques to meet the demands for better chemical measurements.
Analytical chemistry serves the needs of many fields in industry; analytical chemistry
provides the means of testing raw materials and for assuring the quality of finished
products. Chemical analysis may be qualitative or quantitative.(2)
a. Qualitative:
Qualitative analysis involves attempting to identify what materials are present in a
sample. Answering the question: “What is it?”
b. Quantitative:
Quantitative analysis involves determining how much of a material is present in a sample.
Answering the question: “How much is present?”
Modern analytical methods make it possible to identify hundreds of components in a
single sample and to detect specific substances present in less than one part per million.
Analytical chemistry is the science of making quantitative measurements. In practice,
quantifying an analyte in a complex sample becomes an exercise in problem solving. To
be efficient and effective, an analytical chemist must know the tools that are available to
tackle a wide variety of problems. To be effective and efficient, analyzing samples
requires expertise in:
a. The chemistry that can occur in a sample
b. Analysis and sample handling methods for a wide variety of problems (the tools-
of-the-trade)
c. Proper data analysis and record keeping to meet these needs, Analytical
Chemistry courses usually emphasize equilibrium, spectroscopic and
electrochemical analysis, separations and statistics.
2 INTRODUCTION
Specific Technologies And Instrumentation3
:
A) Spectrometric techniques
o Ultraviolet and visible Spectrophotometry
o Fluorescence and phosphorescence Spectrophotometry
o Atomic Spectrometry (emission and absorption)
o Infrared Spectrophotometry
o Raman Spectroscopy
o X-Ray Spectroscopy
o Radiochemical Techniques including activation analysis
o Nuclear Magnetic Resonance Spectroscopy
o Electron Spin Resonance Spectroscopy
B) Electrochemical Techniques
o Potentiometry
o Voltametry
o Voltametric Techniques
o Amperometric Techniques
o Colorimetry
o Electrogravimetry
o Conductance Techniques
C) Chromatographic Techniques
o Gas Chromatography
o High performance Liquid Chromatography
o Thin Layer Chromatography
D) Miscellaneous Techniques
o Thermal Analysis
o Mass Spectrometry
o Kinetic Techniques
E) Hyphenated Techniques
o GC-MS (Gas Chromatography – Mass Spectrometry)
o ICP-MS (Inductivity Coupled Plasma Mass Spectrometry)
3 INTRODUCTION
o GC-IR (Gas Chromatography – Infrared Spectroscopy)
o MS-MS (Mass Spectrometry – Mass Spectrometry)
1.2. ANALYTICAL METHOD DEVELOPMENT
Methods are developed for new products when no official methods are available.
Alternate methods for existing (non-pharmacopoeial) products are developed to reduce
the cost and time for better precision and ruggedness. Trial runs are conducted, method is
optimized and validated. When alternate method proposed is intended to replace the
existing procedure comparative laboratory data including merit / demerits are made
available.
1.2.1. Steps of method development4
Documentation starts at the very beginning of the development process, a system for full
documentation of the development studies must be established. All data relating to these
studies must be recorded in laboratory notebook or an electronic database.
1. Analyte standard characterization
a) All known information about the analyte and its structure is collected i.e., physical
and chemical properties, toxicity, purity, hygroscopic nature, solubility and
stability.
b) The standard analyte (100% purity) is obtained. Necessary arrangement is made
for the proper storage (refrigerator, desicators, freezer).
c) When multiple components are to be analyzed in the sample matrix, the number of
components is noted, data is assembled and the availability of standards for each
one is determined.
d) Only those methods (MS, GC, HPLC etc.,) that are compatible with sample
stability are considered.
2. Method requirements
The goals or requirements of the analytical method that need to be developed are
considered and the analytical figures of merit are defined. The required detection limits,
selectivity, linearity, range, accuracy and precision are defined.
4 INTRODUCTION
2. Literature search and prior methodology
The literature for all types of information related to the analyte is surveyed. For
synthesis, physical and chemical properties, solubility and relevant analytical
methods. Books, periodicals, chemical manufacturers and regulatory agency
compendia such as USP / NF, AOAC and ASTM publications are reviewed.
Chemical Abstracts Service (CAS) automated computerized literature searches are
convenient.
4. Choosing a method
a) Using the information in the literatures and prints, methodology is adapted. The
methods are modified wherever necessary. Sometimes it is necessary to acquire
additional instrumentation to reproduce, modify, improve or validate existing
methods for in-house analytes and samples.
b) If there are no prior methods for the analyte in the literature, from analogy, the
compounds that are similar in structure and chemical properties are investigated
and are worked out. There is usually one compound for which analytical method
already exist that is similar to the analyte of interest.
5. Instrumental setup and initial studies
The required instrumentation is setup. Installation, operational and performance
qualification of instrumentation using laboratory standard operating procedures (SOP’s)
are verified. Always new consumables (e.g. solvents, filters and gases) are used, for
example, method development is never started, on a HPLC column that has been used
earlier. The analyte standard in a suitable injection / introduction solution and in known
concentrations and solvents are prepared. It is important to start with an authentic,
known standard rather than with a complex sample matrix. If the sample is extremely
close to the standard (e.g., bulk drug), then it is possible to start work with the actual
sample. Analysis is done using analytical conditions described in the existing literature.
6. Optimization
During optimization one parameter is changed at a time, and set of conditions are
isolated, rather than using a trial and error approach. Work has been done from an
organized methodical plan, and every step is documented (in a lab notebook) in case of
dead ends.
5 INTRODUCTION
7. Documentation of analytical figures of merit
The originally determined analytical figures of merit limit of quantitation (LOQ), Limit
of detection (LOD), linearity, time per analysis, cost, sample preparation etc., are
documented.
8. Evaluation of method development with actual samples
The sample solution should lead to unequivocal, absolute identification of the analyte
peak of interest apart from all other matrix components.
9. Determination of percent recovery of actual sample and demonstration of
quantitative sample analysis:
Percent recovery of spiked, authentic standard analyte into a sample matrix that is shown
to contain no analyte is determined. Reproducibility of recovery (average +/- standard
deviation) from sample to sample and whether recovery has been optimized has been
shown. It is not necessary to obtain 100% recovery as long as the results are reproducible
and known with a high degree of certainty. The validity of analytical method can be
verified only by laboratory studies. Therefore documentation of the successful
completion of such studies is a basic requirement for determining whether a method is
suitable for its intended applications.
1.2.2. Introduction to validation
Validation of an analytical method is the process by which it is established, by laboratory
studies, that the performance characteristics of the method meet the requirements for the
intended analytical applications5.
Validation is defined as follows by different agencies:
Food and Drug administration (FDA)
Establishing documentation evidence, which provides a high degree of assurance that
specific process, will consistently produce a product meeting its predetermined
specification and quality attributes.
World Health Organization (WHO)
Action of providing that any procedure, process, equipment, material, activity, or system
actually leads to the expected results.
6 INTRODUCTION
European Committee (EC)
Action of providing in accordance with the principles of good manufacturing practice that
any procedure, process, equipment, material, activity or system actually leads to the
expected results. In brief validation is a key process for effective Quality Assurance6
1.2.3.Reasons For Validation7:
There are two important reasons for validating assays in the pharmaceutical industry. The
first, and by for the most important, is that assay validation is an integral part of the
quality-control system. The second is that current good manufacturing practice regulation
requires assay validation.
Steps followed for validation procedures
1. Proposed protocols or parameters for validations are established.
2. Experimental studies are conducted.
3. Analytical results are evaluated
4. Statistical evaluation is carried out.
5. Report is prepared documenting all the results.
1.3. Objective and Parameters of Analytical Method Validation
The objective of validation of an analytical procedure is to demonstrate that it is suitable
for its intended purpose. According to ICH guidelines, typical analytical performance
characteristics that should be considered in the validation of the types of methods are8.
1. Accuracy
2. Precision
3. Specificity
4. Detection Limit
5. Quantitation Limit
6. Linearity
7. Range.
8. Ruggedness
9. Robustness
7 INTRODUCTION
1) Accuracy
The accuracy is the closeness of the measured value to the true value for the sample.
The ICH documents recommended that accuracy should be assessed using a minimum of
nine determinations over a minimum of three concentrations levels the specified range (
i.e, three concentrations and three replicates of each concentration)
Accuracy was tested (% Recovery and % RSD of individual measurements) by analyzing
samples at least in triplicate, at each level (80,100 and 120 % of label claim) is
recommended. For each determination fresh samples were prepared and assay value is
calculated. Recovery was calculated from slope and y intercept of the calibration curve
obtained in linearity study. Accuracy was determined from the mean relative error for a
set of replicate analysis (i.e. the difference between measured and nominal concentration)
for spiked samples.
2) Precision
The precision of an analytical procedure expresses the closeness of agreement between
series of measurements obtained from multiple sampling of the same homogenous
sample under the prescribed conditions. Precision of an analytical method is usually
expressed as the standard deviation, realative standard deviation or coefficient of
variation of a series of measurement. The ICH documents recommend the repeatability
should be assessed using a minimum of nine determinations covering specified range of
procedure. Precision may be measure of either the degree of reproducibility or of
repeatability of the analytical method under normal operating conditions.
Repeatability: Repeatability expresses the precision under the same operating conditions
over a short interval of time. Repeatability is also termed intra –assay precision.
Intermediate Prescision: Intermediate precison expresses with in laboratories variations:
different days, different analyst, different equipment.
Reproduciblity: When the procedure is carried out by different analyst in different
laboratories using different equipment, regents and laboratories setting. Reproducibility
was determined by measuring repeatability and intermediate precision. Reproducibility is
assessed by means of an interlaboratory trial.
8 INTRODUCTION
Fig No.1.1: Difference between Accuracy and Precision illustrated by using target
diagram
3) Specificity
An investigation of specificity should be conducted during the validation of identification
tests, the determination of impurities.ICH documents defines specificity as the ability to
assess unequivocally the analyte in the presence compounds that may be expected to
products and matrix components.
Examples of the extraneous peaks are as followings:
For the drug substance or raw material, the related substances to consider are process
impurities (which include isomeric impurities) from the synthetic process, residual
pesticide, solvents and other extraneous components from extracts of natural orgin.
For the drug product, the related substance may be impurities present in an active drug,
degradation product, interaction of drug with the expients, extraneous components e.g
residual solvents from the expients or manufacturing process, leachable or extractables
from the container and closure system or from the manufacturing process.
The definition has the following implications:
Identification test: Suitable identification tests should be able to discriminate
compounds of closely related structure which are likely to be present .Ensure identity of
an analyte. The analyte should have no interference from other extraneous components
and be well resolved from them.
9 INTRODUCTION
Purity Test: To ensure that all the analytical procedures performed allow an accurate
statement of the content of impurity of the content of impurity of an analyte i.e related
substances test, heavy metals, residual solvents etc.
Assay: To provide an exact result this allows an accurate statement on the content or
potency of the analyte in a sample.
4) Detection Limit (LOD)
It is the lowest amount of analyte in a sample that can be deceted, but not necessarily
quantities as an exact value, under the stated experimental conditions. The detection limit
is usually expressed as the concentration of analyte (Percentage parts per million) in the
sample. The quantitative limit is generally determined by the analysis of samples with
known concentration of analyte and by establishing the minimum level can be accuracy
and precision.
Based on Signal-to-Noise Approach: This approach can only be applied to analytical
procedures that exhibits baseline noise. Determination of the signal-to-noise ratio is
performed by compairing measured signals from samples with known low concentration
of analyte with those of blank samples and by establishing the minimum concentration at
which the analyte can be really quantified. A typical signal-to-noise ratio is 10:1.
Based on the Standard deviation of the response and the slope:
LOD = 3 * SD / slope of calibration curve
SD = Standard deviation of intercepts
5) Limit of quantification (LOQ)
It is the lowest amount of analyte in a samples that can be determined with acceptable
precision and accuracy under the stated experimental conditions. Quantification limit is
expressed as the concentration of analyte (e.g- % ppm) in the sample.
LOQ = 10 * SD / slope of calibration curve
SD = Standard deviation of intercepts
Based on S.D of the blank: Measurement of the magnitude of analytical background
response is performed by analyzing an appropriate number of blank samples and
calculating the standard deviation of these responses.
10 INTRODUCTION
Based on the calibration curve: A specific calibration curve should be studied using
samples, containing an analyte in the range of QL. The residuals S.D of regression line or
the S.D of intercepts of regression lines, may be used as the S.D. The quantitative limit is
a parameter of quantitative assay for low levels of compounds in sample matrics, and is
use particularly for the determination of impurities or degradation products.
6) Linearity and Range
The linearity of an analytical procedure is its ability to obtain test results which are
directly proportional to concentration of the analyte in the sample. The range of an
analytical is the intervals between the upper and lower concentration (amounts) of analyte
in the sample for which it has been demonstrated which it has been demonstrated that the
analytical procedure has a suitable level of precision accuracy and linearity. For
determination of linearity standard were tested six times. A calibration curve was
constructed and the proposed method was evaluated by its correlation coefficient and
intercept value, calculated in the corresponding statistical study (ANOVA). Calibration
plots were constructed for drug and related substances
in either standard solution or synthetic mixtures of drug product components by plotting
the concentration of compounds versus peak area response using least square method.
The ICH recommends that for the establishment of linearity a minimum of five
concentrations normally used.
7) Ruggedness
Degree of reproducibility of test results obtained by the analysis of the same samples
under a variety of condition such as different laboratories, different analysts, different
instruments etc, normally expressed as the lack of influence on test results of operational
and environmental variable of the analytical method. Ruggedness is a measurement of
reproducibility of test results under the variation in condition normally expected from
laboratory to laboratory and from analyst to analyst. Degree of represenatative of test
results is then determined as a function of the assay variable. By analysis of aliquots from
homogenous lots in different laboratories, by different analyst, using operational and
environmental conditions that may differs but are still with in the specified parameter of
the assay variable.
11 INTRODUCTION
8) Robustness
Robustness of an analytical method is measure of its capacity to remain unaffected
small but deleribate variations in method parameters and provides an indication of its
reliability during normal usage.
Testing varing some or all condition:
Age of column
Column temperature
PH of buffer in mobile phase
Reagents and flow rate
9) System Suitability
System suitability tests are based on the concept that the equipment, electronics,
analytical operations and samples constitute an integral system that can be evaluated as a
whole. According to USP system suitability are an integral part of chromatographic
methods. These tests verify that the resolution and reproducibility of the system are
adequate for the analysis to be performed. One consequence of the evaluation of
robustness and ruggedness should be that a series of system suitability parameters is
established to ensure that the validity of the analytical method is maintained whenever
used. The parameters which are recommended by International Committee of
harmonization to be validated for different types of assays are shown in Table No. 1.2.
12 INTRODUCTION
Table No.1.2:
Validation Parameters Recommended by
International Committee on Harmonization (ICH)8
ASSAY TYPE VALIDATIONS
Identification tests are intended to ensure the identity of an
analyte in a sample. This is normally achieved by
comparison of a property of the sample to that of a
reference standard.
Specificity
Impurities quantitations are intended to accurately reflect
the purity characteristic of the sample. Different validation
characteristics are required for a quantitative test than for a
limit test.
Accuracy
Precision
Specificity
Detection limit
Quantitation
limit
Linearity
Range
Impurities Limit are intended to reflect the purity
characteristics of the sample.
Specificity
Detection limit
Content / Potency, Dissolution are intended to measure the
analyte present in a given sample. A quantitative
measurement of the major component (s) in the drug
substance.
Accuracy
Precision
Specificity
Linearity
Range
13 INTRODUCTION
Table No– 1.3
Validation Characteristics Versus Type of Analytical Procedures Test of
Impurities9:
Type
of Procedure Identification Quantitation Limit
Dissolution
Measurement
(Content /
Potency)
Accuracy No Yes No Yes
Precision
/ Repeatability
No Yes No Yes
Intermediate
Precision
No Yesa No Yes
a
Specificity Yes Yes Yes Yes
Detection Limit No Nob Yes No
Quantitation
Limit
No Yes No No
Linearity No Yes No Yes
Range No Yes No Yes
a. When reproducibility is performed, intermediate precision is not
needed.
b. May be needed in some cases.
The comparison of different official guidelines in case of parameters required to be
validated for different assays is shown in Table – 1.3
14 INTRODUCTION
Table – 1.4
Comparative Table Representing FDA, USP and
ICH Requirements
Criteria GMP FDA USP ICH
Accuracy x X x x
Reproducibility x x
Sensitivity x
Specificity x X x x
Linearity X x x
Precision X x x
Detection Limit x x
Quantitation Limit x x
Range x x
Recovery X
Ruggedness X x
1.3.1. Benefits Of Method Validation7:
A fully validated process may require less in-process control and end –product testing.
It deepens the understanding of processes, decrease the risks of processing problems, and
thus assure the smooth running of the process.
Quality: Customer – patient satisfaction. It has been built into the product.
1.3.2. Cost Reduction
Increased efficiency, shortening lead time resulting in lower inventories.
Fewer rejects and reworks.
15 INTRODUCTION
Longer equipment life by operating the equipment as per manufacturer’s
specifications and the establishing of cost effective preventive maintenance
schedules.
Possible reduced testing of raw materials bulk formulations and finished
products.
1.3.3. STATISTICAL ANALYSIS10
:
Statistical procedures and representative calculations:
The consistency and suitability of the developed method are substantiated through the
statistical analysis like standard deviation, relative standard deviation and theoretical
plates per meter.
For Accuracy: Standard deviation = =1
)xx(2
i
n
Where, x = sample, xi = mean value of samples, n = number of samples
Relative Standard Deviation = /xi × 100
Molar extinction coefficient (mol-1
cm-1
) = A/C × L
Where, A= Absorbance of drug, C= concentration of drug, L= Path length
Sandell, s sensitivity (µg/cm2/0.001 absorbance units) = C/A×0.001
Where, C= concentration of drug, A= Absorbance of drug
16 INTRODUCTION
1.5. HPLC system is composed of several components. These components are11
:
o Sample delivery system, which includes pump, associated pressure, and flow
controls and filters on the inlet side.
o Sample injection function.
o The column.
o The detector
o Data processor.
o Recorder.6
Fig. No.1.5: Diagram of a Typical HPLC
17 INTRODUCTION
1.4.1. HPLC method validation12
Everyday many chromatographers face the need to develop a HPLC separation whereas
individual approaches may exhibit considerable diversity; method development often
follows the series of steps summarized in the following fig.
Fig 1.6: - Steps involved in HPLC method validation
1. Introduction on sample
Define separation goals
2. Need for special HPLC
Procedure, sample, pretreatment, etc
3. Choose detector and
Detector settings
4. Choose LC method;
Preliminary run; estimate
best separation conditions
5. Optimize separation
condition
6. Check for problems or requirements
for special procedure
7a. Recover
Purified material
7c. Qualitative
Method
7b.Quantitative
Calibration
8. Validation method for release to
routine laboratory
18 INTRODUCTION
1.4.2. Selection of Column:
The HPLC column is the heart of the method, critical performing the separation. The
column must possess the selectivity, efficiency and reproducibility to provide good
separation. Commonly used reversed phase columns are C18 (octadecyl silane,), C8 (octyl
silane,) phenyl and cyano. They are chemically different bonded phases and demonstrate
significant changes in the selectivity using the same mobile phase. During method
development selection of column can be streamlined by starting with shorter columns
(150,100 or even 50mm long.). By selecting a shorter column with an appropriate phase
run time can be minimized so that an elution order and an optimum mobile phase can be
quickly determined. It is also advantageous to consider the column internal diameter,
many laboratories use 4.6mm ID as standard, but it is worth considering use of 4mm ID
column as an alternative. This requires only 75% of the solvent flow than that of 4.6mm
column.Selecting an appropriate stationary phase can also help to improve the efficiency
of method development. For example, a octyl phase (C8) can provide time saving over a
octadecyl (C18) as it doesn’t retain analytes as strongly as the C18 phase. For normal phase
applications Cyano phases are the most versatile. C18 (250*4.6mm) column are more
often used in the laboratory. These columns are able to resolve a wide variety of
compounds due to their selectivity and high plate counts.
1.4.3. Role of Flow rate
Flow rate, more for isocratic than gradient separation, can sometimes be useful and
readily utilized to increase the resolution, although its effect is very modest. The slower
flow rate will also decrease the column back pressure. The disadvantage is that when
flow rate is decreased, to increase the resolution slightly, there is a corresponding
increase in the run time. Method development involves considerable trial and error
procedures. Optimization can be started only after a reasonable chromatogram has been
obtained. A reasonable chromatogram means that all the compounds are detected by
more or less symmetrical peaks on the chromatogram.
19 INTRODUCTION
The parameters that are affected by the changes in chromatographic conditions are,
Capacity factor (k’),
Selectivity (),
Column efficiency (N) and
Peak asymmetry factor (As).
1. Capacity factor (k')
k' = (t R- t0) / t 0
The capacity factor is a measure of the degree of retention of an analyte relative to an
unretained peak, where tR is the retention time for the sample peak and to is the retention
time for an unretained peak.
Recommendations:
The peak should be well-resolved from other peaks and the void volume. Generally the
value of k' is > 2.
2. Backpressure
The pressure required to pump the mobile phase through the column. It is related to
mobile phase viscosity (η), flow rate (F), column length (L), and diameter (dc), and
particle size (dp) by the following equation:
ΔP α FLη / dp2
dc2
3. Resolution (Rs)
Ability of a column to separate chromatographic peaks, It can also be expressed in terms
of the separation of the apex of two peaks divided by the tangential width average of the
peaks.
Rs = ΔtR / W1/2 + W2/2;
Where ΔtR = t2 – t1
For reliable quantitation, well-separated peaks are essential for quantitation.
20 INTRODUCTION
Resolution can be improved by increasing column length, decreasing particle size,
increasing temperature, changing the eluent or stationary phase
Recommendations:
Rs > 2 between the peak of interest and the closest potential interfering peak (impurity,
excipient, degradation product, internal standard, etc.) are desirable.
4. Tailing factor (T)
A measure of the symmetry of a peak, given by the following equation where W0.05 is the
peak width at 5% height and f is the distance from peak front to apex point at 5% height.
Ideally, peaks should be Gaussian in shape or totally symmetrical.
T = W0.05 / 2f
The accuracy of quantitation decreases with increase in peak tailing because of the
difficulties encountered by the integrator in determining where/when the peak ends and
hence the calculation of the area under the peak. Integrator variables are preset by the
analyst for optimum calculation of the area for the peak of interest.
Recommendations:
T of ≤ 2
5. Theoretical plate number / Efficiency (N)
Theoretical plate number is a measure of column efficiency, that is, how many peaks can
be located per unit run-time of the chromatogram. A measure of peak band spreading
determined by various methods, some of which are sensitive to peak asymmetry. The
most common are shown here, with the ones most sensitive to peak shape shown first:
4-sigma / tangential: N = 16 (tR / W) 2
= L / H
Half height: N = 5.54 (tR / W) 2 = L / H
Where tR is the retention time for the sample peak and W is the peak width.
21 INTRODUCTION
N is fairly constant for each peak on a chromatogram with a fixed set of operating
conditions. H, or HETP, the height equivalent of a theoretical plate, measures the column
efficiency per unit length (L) of the column. Parameters which can affect N or H include
Peak position, particle size in column, flow-rate of mobile phase, column temperature,
viscosity of mobile phase, and molecular weight of the analyte
Recommendations:
The theoretical plate number depends on elution time but in general should be > 2000.
General Recommendation:
The amount of testing required will depend on the purpose of the test method. For
dissolution or release profile test methods using an external standard method, k', T and
RSD are minimum recommended system suitability tests. For acceptance, release,
stability, or impurities/degradation methods using external or internal standards, k', T, Rs
and RSD are recommended as minimum system suitability testing parameters. System
suitability testing is essential for the assurance of the quality performance of the
chromatographic system13
.
22 INTRODUCTION
1.5 HIGH PERFORMANCE THIN LAYER CHROMATOGRAPHY (HPTLC) (14-
19) :
1.5.1. INTRODUCTION:
The term “thin layer chromatography” was coined by Egon Stahl in Germany in the late
1950’s. Stahl’s greatest contribution to the field was standardization of materials,
procedures and nomenclature and description of selective solvent systems for the
resolution of many important compound classes. In present day HPTLC is without doubt
one of the most versatile and widely used separation methods in chromatography.
Commercially, many sorbents on a variety of backings and mobile phases are now
available and allows the handling of a large number of samples in one chromatographic
run. It has found use in a wide range of application areas as the concept of HPTLC is so
simple and samples usually require only minimal pretreatment, detecting components at
low nanogram sensitivities. HPTLC technique is much more reliable and reproducible
method for the standardization of both single and compound herbal and allopathic
formulations. HPTLC is a separation technique based on a stationary phase and a liquid
mobile phase. Separations are achieved by adsorption. The stationary phase is a flat thin
layer coated onto an inert support (glass, aluminum or polyester). In its simplest form,
HPTLC costs little, but even including the more sophisticated instrumentation, it still
remains less expensive per sample analysis than HPLC. The modern HPTLC has seen a
strong move in the direction of plate scanning and video imaging as a means of providing
sensitive and reliably accurate results and a more permanent record of chromatogram.
The HPTLC is the improved method of TLC which utilizes the conventional technique of
TLC in more optimized way. It is also known as planar chromatography or Flat-bed
chromatography.
Features of HPTLC
1. Simultaneous processing of sample and standard - better analytical precision and
accuracy less need for Internal Standard
2. Several analysts work simultaneously
3. Lower analysis time and less cost per analysis
4. Low maintenance cost
5. Simple sample preparation - handle samples of divergent nature
23 INTRODUCTION
6. No prior treatment for solvents like filtration and degassing
7. Low mobile phase consumption per sample
8. No interference from previous analysis - fresh stationary and mobile phases for
each analysis - no contamination
9. Visual detection possible - open system
10. Non UV absorbing compounds detected by post-chromatographic derivatization
Various steps involved in HPTLC:
1. Selection of HPTLC plates and adsorbent
2. Sample preparation
3. Application of sample
4. Development
5. Detection and post-chromatographic derivatisation
6. Quantification
7. Documentation
24 INTRODUCTION
Fig. No.1.7: Schematic procedure for HPTLC
Validation method for release to
routine laboratory
25 INTRODUCTION
1.5.2. Selection of HPTLC plates:
Pre coated plates:
The plates are used which have different support materials and sorbent layers with
different format and thickness.
HPTLC Plates:
HPTLC grew out of improvements in the quality of sorbents and consistency of plate
manufacture. HPTLC plates are characterized by smaller particles (<10 μm), thinner
layers (<150 μm) and smaller plates (<10 cm developing distance). Also, the particle size
distribution of the sorbent is narrower than for conventional TLC layers. HPTLC plates
provide more resolving power per unit distance, develop faster and consume less solvent.
However, sample diffusion in the direction of migration must be kept to a minimum to
prevent band broadening, and the thin sorbent layer dictates sample volume in HPTLC be
kept to a maximum of 1 µL per stroke.
Supports:
Table 1.8: List of supports materials for HPTLC plate
Materials Advantage Disadvantage
Glass 1. resistant to heat and chemicals.
2. easy to handle and offers superior
flat surface for work.
1. fragile.
2. costs more for
additional packaging.
Polyester sheets
(0.2 mm thick)
1. more economical as produced
even in roll forms.
2. unbreakable.
3. less packing material.
4. spots can be cut and eluted thus
eliminates dust from scrapping
1. charring reactions if
temperature exceeds
120oC as the plates are
dimensionally unstable
beyond this temperature.
Aluminum Sheets
(0.1 mm)
1. increasesed temperature
resistance.
1. eluents containing high
concentration of mineral
acids or ammonia can
attack chemically on
aluminum.
26 INTRODUCTION
Plate size:
10X20 cm, 10X10 cm, 5X10 cm and 5X7.5 cm
The plates should have good cut edges of sheets to obtain constant Rf values if not flaking
off the layer on one or both sides of cut edges which leads for capillary cracks between
chromatographic layer and the foil in which mobile phase will travel much more rapidly
forward than it does in the center of chromatogram. The cracks causes mobile phase to
migrate from edges of the layer to the center thus causing deformation of zones and
distortion of tracks.
Pre-washing of pre-coated plates:
Pre-washing is required because sorbents with large surface area absorb not only water
vapors and other impurities from atmosphere but also other volatile substances which
often condense particularly after the packing has been opened and exposed to laboratory
atmosphere for a long time. Such impurities include elutable components of the binder
usually give dirty zones and fail to give reproducible results. So all the times pre-coated
plates are always packed with glass or aluminum foil side upward.
Some common methods involved in pre-washing:
a) Ascending method: In this technique the chromatographic plates are run blank
(i.e. before application of the sample) with suitable solvent / mobile phase. The
solvent/mobile phase carries the impurities to the top of the plate. It takes longer
time but cleaning effect is superior. The disadvantage of this technique is active
dirt gets accumulated at the solvent front.
b) Dipping method: In this technique, the chromatographic plate is dipped in a
suitable solvent for specified period of time, removed from the chamber and finally
dried. Dipping method is quicker and yields uniform layer but cleaning effect is
often not as good as Ascending method.
c) Continuous method: In this technique, the plate to be washed is placed in
chamber having an entrance and exit slits. The solvent is made to flow
continuously through the chamber that carries the impurities from the plate. The
wanted plates should always be stored in a dust free atmosphere, under ambient
conditions. Usually desiccators of suitable size are used for storage of plates.
27 INTRODUCTION
Solvents used for pre-washing:
a. Methanol
b. Chloroform: methanol
c. Chloroform: Methanol: Ammonia
d. Methylene chloride: Methanol
e. Ammonia solution
The plates are then dried for 20 minutes in a clean drying oven at 120°C. The plates are
then removed and equilibrated with lab atmosphere (temperature, relative humidity) in a
suitable container providing protection from dust and fumes. Further then the plates are
handled either on both side edges or on the top edge. The pre-washed plates are kept in
desiccators without applying grease for sealing of edges.
1.5.4. Sample Preparation:
1. Sampling, mechanical crushing, extraction, filtration and enrichment of minor
compounds are the several steps for sample preparation.
2. Sample and reference substances should be dissolved in the same solvent to ensure
comparable distribution at starting zones.
Application of sample:
The selection of a sample application technique and the device to be used depends
primarily on:
Sample volume
Number of samples to be applied
Required precision and degree of automation.
In order to maximize the separation, it is important to restrict the size of the sample origin
in the direction of chromatography to a minimum. The maximum sample volume that can
be applied spot wise (in one stroke) are 5 µL L on conventional layers and 1 µL L on
HPTLC layers. Spot wise application of larger volumes requires a device with
controllable delivery speed.
Sample application in the form of bands or rectangles:
Spraying-on samples as narrow bands allows the application of larger volumes. Narrow
bands as starting zones always ensure the highest resolution. Very large sample volumes
28 INTRODUCTION
or samples with high matrix content can be sprayed-on in the form of rectangles. All
types of Thin-Layer Chromatography, whether qualitative, quantitative or preparative,
benefit from optimized resolution as a result of the appropriate sample application
technique. Sample application is the first step of instrumental Thin-Layer
Chromatography and thus determines the quality of the analysis. For application, the
sample should be dissolved in the solvent of lowest suitable solvent strength in order to
minimize the effects of spot broadening during application. For application by spraying,
the speed has to be adjusted to them properties of chosen solvent. In order to avoid
volume errors highly volatile solvents should not be used to prepare and apply standards
and samples for quantitative analyses. Samples are applied as bands by spray-on
technique or as spots using the following parameters:
Table 1.9: Parameters for HPTLC
Parameters HPTLC
Distance from lower edge of plate in for TLC 8 mm
Minimum distance from left and right edge of plate 10 mm
Minimum space in mm between bands / spots 4 mm
Band length 8 mm
Maximum diameter of application spot 5 mm
1.5.5. Chromatogram Development: Thin-layer chromatography differs from all other
chromatographic techniques in the fact that in addition to stationary and mobile phase a
gas phase is present. This gas phase can significantly influence the result of the
separation.
Processes in the Developing Chamber:
The classical way of developing a chromatogram is to place the plate in a chamber, which
contains a sufficient amount of developing solvent. The lower end of the plate should be
immersed several millimeters. Driven by capillary action the developing solvent moves
up the layer until the desired running distance is reached and chromatography is stopped.
The following considerations primarily concern silica gel as stationary phase, which can
29 INTRODUCTION
be described as adsorption chromatography. The process occurring within the
development chamber is as follows:
a. Between the components of the developing solvent and their vapor, equilibrium is
established eventually. This equilibrium is called chamber saturation. Depending
on the vapor pressure of the individual components the composition of the gas
phase can differ significantly from that of the developing solvent.
b. While still dry, the stationary phase adsorbs molecules from the gas phase. This
process of adsorptive saturation is also approaching an equilibrium in which the
polar components will be withdrawn from the gas phase and loaded onto the
surface of the stationary phase.
c. Simultaneously a part of the layer which is already wetted with mobile phase
interacts with the gas phase. Thereby especially the less polar components of the
liquid are released into in the gas phase. This process is not as much governed by
vapor pressure as by adsorption forces.
d. During migration, the components of the mobile phase can be separated by the
stationary phase under certain conditions, causing the formation of secondary
fronts.
The problems occurring due to this can be avoided by keeping increasing the chamber
saturation time and this can be achieved by as follows: keeping the chamber more or less
completely with filter paper soaked with developing solvent, waiting a certain time
between the introduction of developing solvent into the chamber and beginning of
chromatography. The advantages of chamber saturation is that the, components of the
developing solvent, which have been loaded onto the dry layer via the gas phase, the gas
phase are pushed ahead of the mobile phase with invisible solvent front. The exceptions
are very polar components such as water, methanol, acids, or bases. This results in Rf
values being lower in saturated chambers and particularly on pre-conditioned layers, than
in unsaturated chamber. The consequences of not allowing chamber saturation proceeds
in non-equilibrium between stationary, mobile, and gas phase. For this reason it is very
difficult to describe the conditions in a developing chamber. Reproducible
chromatographic results can only be expected when all parameters are kept as constant as
30 INTRODUCTION
possible. Chamber shape and saturation are playing a predominant role in this regard.
Unfortunately this means that the chromatographic result is different in each chamber.
Choosing a developing chamber:
The selection of the proper developing chamber is done during method development and
generally follows practical considerations such as which chamber is available, which one
must be used due to sop, or which one has been used in the past if a results comparison is
to be made. However, a focus should also be on economical aspects such as time
requirement and solvent consumption. A selection of glass chambers can be done by as
follows:
The Horizontal Developing Chambers: It is economical, flexible and reproducible
in operation. Although designed for applications where the plate is developed from two
sides, they are also suitable for single-sided developments in unsaturated, saturated and
sandwich configuration as well as for preconditioning of HPTLC plates.
Twin trough chamber: Twin trough chambers offer several ways to improve the
results of TLC/HPTLC developing techniques. The trough is divided into two divisions,
where the mobile phase is added to one division and allowed for the plate saturation. This
reduces the time for chamber saturation when compared to flat bottom chambers. The
development is started only when developing solvent is introduced into the trough with
the plate. The main advantage in this is the amount of mobile phase required is not more
than 20 mL for 20X20 cm plate. This not only saves solvent, but also reduces waste
disposal problem.
Automatic Developing Chamber (ADC): In this system, the chromatographic
development can be fully automated and can control the chamber saturation, pre-
conditioning of the layer and final drying. This instrument not only eliminate any effects
of the operator when introducing the plate into a saturated chamber, but also the activity
of the layer prior to start of chromatography can be set and drying of the
chromatographed plate is rapid and complete. The main advantage in this system is that
one can control the humidity required for the development of chromatographic plate.
Automated Multiple Development (AMD): This is used In case the sample contains
polar and non-polar components, which must be separated in the same analysis.
31 INTRODUCTION
Development is performed on the basis of a solvent gradient from polar to non-polar over
several steps with intermediate drying.
1.5.6. Mobile phase (M P):
The Mobile phase should be of high grade,
1. Chemical properties and analytes and sorbent layer factors should be considered
while selection of mobile phase.
2. Use of mobile phase containing more than three or four components should
normally be avoided as it is often difficult to get reproducible ratios of different to
get reproducible ratios of different components.
Properties of Mobile phase:
1. M P optimization is necessary while performing HPTLC.
2. Various components of M P should be measured separately and then placed in mixing
vessel. This prevents contamination of solvents and also error arising from volumes
expansion or contraction on mixing.
3. Trough chambers are used in which smaller volumes of M P usually 10-15 mL is
required. Usually volumetric pipettes are used.
4. Different components of M P are mixed first in mixing vessel and then transferred to
developing chambers.
5. Chambers containing multi component M P are not generally used for re-use for any
future development, due to differential evaporation and adsorption by layer and also
once the chamber is opened, solvents evaporate disproportionally depending on their
volatilities.
1.5.7. Post-chromatographic Derivatization:
After the development, it’s an inherent advantage of TLC that fractions remain stored on
the plate and can be derivatized after chromatography. By derivatization substances that
do not respond to visible or UV light can be rendered detectable. In many cases,
substances or classes of substances can be identified by specific reagents. Derivatization
can be achieved with gas, by liquid spraying or dipping (immersion). In any case the
reagent needs to be homogenously transferred to the chromatogram. By immersing a
TLC plate into the derivatizing reagent a very homogenous reagent transfer can be
achieved. Dipping and withdrawing has to be performed smoothly in order to avoid
32 INTRODUCTION
tidemarks. Using the suitable Chromatogram Immersion Device the reproducibility of the
derivatization step can be significantly improved when compared to spraying.
Furthermore, no fumes are generated during this derivatization technique and the
exposure to hazardous chemicals is limited. If the reagent is suitable, dipping should be
preferred over spraying. However, the fact is that spraying is most widely used for
reagent transfer onto the TLC plate because it is simple and quick. No expensive
equipment is necessary and only small volumes of reagent are needed. In addition
spraying is very flexible and indispensable when reagents have to be applied in sequence.
Also during method development, when searching for the most suitable reagent, spraying
is more frequently mentioned. The two principal heating devices are ovens and plate
heaters.
Chromatogram evaluation with classical densitometry or electronic image
acquisition: Quantitative evaluation of a TLC/HPTLC plate is always performed
densitometrically, either in absorption or fluorescence mode. The signal of each
substance zone is compared to the substance free plate background. For calibration and
result calculation the obtained peak data of the unknowns are compared against data
obtained for standards on the same plate. Quantitative evaluation can be performed with
data from classical densitometry or with those from electronic image acquisition.
Classical densitometry uses monochromatic light and a slit of selectable length and width
to scan the tracks of a chromatogram, measuring the diffusely reflected light. The latest
available instrument uses the entire spectral range from 190 to 800 nm with high spectral
selectivity for data acquisition. Absorption spectra for substance identification and for
selection of the most suitable measurement wavelength can be recorded within this range.
The strengths of classical densitometry are the spectral resolution of the light source and
the higher reproducibility of quantitative determinations. Currently electronic image
acquisition works only in the visible range. The UV region – exceptionally useful for
Thin-Layer Chromatography – is only indirectly accessible to image acquisition. In this
respect this technology exactly parallels the human eye.
The advantages of spectral selectivity are as follows: Spectral selectivity is only
accessible through the classical densitometer. UV-absorbing substances can be detected
by image acquisition technology. The quenching of a fluorescence indicator embedded
33 INTRODUCTION
in the layer, i.e. detection is shifted to the visible region. The more a substance to be
quantified absorbs at or near the excitation maximum of the fluorescence indicator (254
nm), the higher is sensitivity and accuracy of image processing but the lower the
absorbance at 254 nm, the less sensitive and less accurate it becomes. Recording of UV
spectra with their information about absorption maxima, identity and purity, is only
possible with classical densitometry.
Electronic image acquisition uses polychromatic light (white light, UV 254 or UV 366)
to illuminate the entire object and to capture an electronic image with a digital camera.
For the documentation of Thin-Layer Chromatograms, electronic image acquisition has
essentially replaced photography. Electronic images can easily be archived and can be
retrieved at any time, for review or to perform a quantitative evaluation. The strength of
electronic image acquisition is the view of the complete image of the chromatogram. This
possibility to get a visual impression of the chromatogram is one of the principal
advantages of Thin-Layer Chromatography over all other chromatographic techniques.
1.5.8. Applications of HPTLC:
1. Pharmaceutical industry: Quality control, content uniformity, uniformity test,
identity/purity check.
2. Food Analysis: Quality control, additives, pesticides, stability testing, analysis of
sub-micron levels of toxins etc
3. Clinical Applications: Metabolism studies, drug screening, stability testing etc
4. Industrial Applications: Process development and optimization, In-process check,
validation etc.
5. Forensic: Poisoning investigations.
34 INTRODUCTION
1.6 DRUG PROFILE (20-25)
1.6.1. PIOGLITAZONE ( PIO )
Structural formula
IUPAC name : 5-[[4-[2-(5-ethylpyridin-2-yl)ethoxy]phenyl]methyl]-1,3-thiazolidine-2,4-
dione
Formula: C19H20N2O3S
Mol.Weight: 356.4390 g mol -1
Appearance: A crystalline solid
Approved Name: Pioglitazone
Category: Antidiabetic agent, Hypoglycemic Agents
Mechanism of action: Pioglitazone acts as an agonist at peroxisome proliferator activated
receptors (PPAR) in target tissues for insulin action such as adipose tissue, skeletal
muscle, and liver. Activation of PPAR-gamma receptors regulates the transcription of
insulin-responsive genes involved in the control of glucose production, transport, and
utilization. In this way, pioglitazone enhances tissue sensitivity to insulin.
Half life: 3-7 hour
Uses: Treatment of Type II diabetes mellitus
35 INTRODUCTION
1.6.2. ROSIGLITAZONE ( ROS)
Structural formula
IUPAC name: 5-[[4-[2-(methyl-pyridin-2-ylamino)ethoxy]phenyl]methyl]-1,3-
thiazolidine-2,4-dione
Formula: C18H19N3O3S
Mol. Weight: 357.42 g mol -1
Appearance: A crystalline solid
Approved Name: Rosigliazone maleate, Rosiglitazone maleate, Rosiglitazone
Category: Antidiabetic agent ,Hypoglycemic Agent
Mechanism of action: Rosiglitazone acts as an agonist at peroxisome proliferator
activated receptors (PPAR) in target tissues for insulin action such as adipose tissue,
skeletal muscle, and liver. Activation of PPAR-gamma receptors regulates the
transcription of insulin-responsive genes involved in the control of glucose production,
transport, and utilization. In this way, rosiglitazone enhances tissue sensitivity to insulin
Half life: 3-4 hours
Uses: For the treatment of Type II diabetes mellitus
36 INTRODUCTION
1.6.3. METFORMIN ( MET)
Structural formula
IUPAC name: 3-(diaminomethylidene)-1,1-dimethylguanidine
Formula: C4H11N5
Mol. weight: 129.16 g mol -1
Approved Name: Metformin hydrochloride
Appearance: A crystalline solid
Category: Antidiabetic agent , Hypoglycemic Agent
Mechanism of action: Metformin's pharmacologic mechanisms of action are different
from other classes of oral antihyperglycemic agents. Metformin decreases hepatic
glucose production, decreases intestinal absorption of glucose, and improves insulin
sensitivity by increasing peripheral glucose uptake and utilization
Half life: 6.2 hours
Uses: Metformin is an antihyperglycemic agent, which improves glucose tolerance in
patients with type 2 diabetes, lowering both basal and postprandial plasma glucose.
Metformin is not chemically or pharmacologically related to any other classes of oral
antihyperglycemic agents.
37 INTRODUCTION
1.6.4. GLIMEPIRIDE ( GLP )
Structural formula
Systematic(IUPAC)name:
3-ethyl-4-methyl-N-[2-[4[(4methylcyclohexyl)carbamoylsulfamoyl]phenyl]ethyl]-
2-oxo-5H-pyrrole-1-carboxamide
Formula: C24H34N4O5S
Mol. weight: 490.61 g mol -1
Approved Name: Glimepride
Appearance: A crystalline solid
Category: Antidiabetic agent , Hypoglycemic Agent
Mechanism of action: The mechanism of action of glimepiride in lowering blood glucose
appears to be dependent on stimulating the release of insulin from functioning pancreatic
beta cells, and increasing sensitivity of peripheral tissues to insulin. Glimepiride likely
binds to ATP-sensitive potassium channel receptors on the pancreatic cell surface,
reducing potassium conductance and causing depolarization of the membrane. Membrane
depolarization stimulates calcium ion influx through voltage-sensitive calcium channels.
This increase in intracellular calcium ion concentration induces the secretion of insulin.
Half life: 5 hours
Uses: Glimepiride is used with diet to lower blood glucose by increasing the secretion of
insulin from pancreas and increasing the sensitivity of peripheral tissue for concomitant
use with insulin for the treatment of noninsulin-dependent (type 2) diabetes mellitus..
Glimepiride, like glyburide and glipizide, is a "second-generation" sulfonylurea agents.
38 INTRODUCTION
1.6.5. REPGLINIDE ( REP )
Structural formula
Systematic(IUPAC)name:
2-ethoxy-4-[2-[[(1S)-3-methyl-1-(2-piperidin-1-ylphenyl)butyl]amino]-2-
oxoethyl]benzoic acid
Formula: C27H36N2O4
Mol. Weight: 452.58 g mol -1
Appearance: A crystalline solid
Approved Name: Repaglinide
Category: Antidiabetic agent, Hypoglycemic Agent
Mechanism of action: Repaglinide closes ATP-dependent potassium channels in the b-
cell membrane by binding at characterizable sites. This potassium channel blockade
depolarizes the b-cell, which leads to an opening of calcium channels. The resulting
increased calcium influx induces insulin secretion. The ion channel mechanism is highly
tissue selective with low affinity for heart and skeletal muscle.
Half life: 3-7 hours
Uses: For the treatment of Type II diabetes mellitus. Repaglinide is an oral blood
glucose-lowering drug of the meglitinide class used in the management of type 2 diabetes
mellitus (also known as non-insulin dependent diabetes mellitus or NIDDM).
39 INTRODUCTION
Repaglinide lowers blood glucose levels by stimulating the release of insulin from the
pancreas. This action is dependent upon functioning beta cells in the pancreatic islets.
Insulin release is glucose-dependent and diminishes at low glucose concentrations.