AN ISO 9001-2008 COMPANY Where innovation never stops..... 301, Punit Indl. Premises, Turbhe Naka, Navi Mumbai - 400 705. Tel. : +91 22 6139 3000 (100 lines) Fax : +91 22 2761 8421 E-mail : [email protected] / [email protected] Website: www.contechindia.com HANDHELD DETECTORS (Si-PIN / SDD) CGX-104 X-RAY FLUORESCENCE SPECTROMETERS XRF is used to detect and analyze noble metals like Gold ( Au), Silver (Ag), Platinum (Pt), Iridium (Ir), Osmium (Os), Palladium (Pd), Rhodium (Rh), Ruthenium (Ru), Copper (Cu), Zinc (Zn), Nickel (Ni), Cadmium (Cd), Iron (Fe), Cobalt (Co), Indium (In), Tin (Sn), in the Jewellery Industry. Energy Dispersive X-ray Fluorescence (EDXRF) technique relies on the detector and the detection circuit capability to resolve spectral peaks due to different energy X-Rays emitted when a primary X-Ray beam strikes a metal alloy corresponding to different elements in the sample, intensity of the spectrum radiation proportional to the concentration of element. The spectrum is then processed through the powerful computer software which calculates and reports the various elements & their composition in the sample. XRF Gold Purity Tester consists of X-Ray tube, high voltage power supply for X-Ray source, X-Ray tube filters, X-Ray fluorescence Detector, processing electronics and system interface & an external computer. A camera is incorporated in the system to view the sample on the computer screen. XRF is a reference technique, hence standards are required for quantitative results. Standards are analyzed, spectral intensities generated are used to obtain a intensity v/s concentration plot for calibration. This calibration is used to evaluate the spectral intensities of unknown samples. How to get best results from XRF Spectrometer : XRF is a reference method, proper standardization of the equipment is required for accurate results. In order to get most appropriate results, testing at different points is necessary especially when the jewellery design is intricate. Adequate warm up period, sturdy working surface, stable power supply, proper grounding and conditioned atmosphere is required to achieve the best product performance. For powder sample, proper grinding of sample minimizes scattering effect. For solid sample, surface polishing minimizes the scattering effect. Flat surface of the sample yields the best results. Due to different melting points, atomic structure and crafting techniques, not all the elements can be melted with noble metals. OPTIONAL LAB STAND SETUP CGX-104 Handheld X-ray analyzer works on a rechargeable 7.2V Li-ion baery with a backup period of 8 hours under normal use. XRF analyzer has a detectable range from Ti to U. CGX-104 operates under control of the HP iPAQ PDA with 4 inch high resoluon touch-screen display. The oponal lab stand allows using of handheld analyzer as a tabletop instrument under PC or laptop control. Baery Operaon Time : Up to 8 hours Dimensions : 242 mm (L) x 78 mm (W) x 230 mm (H) Weight (with baery) : 1230 g (1440 g) X-ray tube Detector type Acve area Energy resoluon Display Supported operang systems Tungsten anode, 40 kV , 100 µA with 5 filter posions solid state Si-pin-diode with thermo-electrical cooling (oponal SDD) 2 6 mm <165 eV at 5.9 keV (<140 eV for SDD) High resoluon 4 inch color touch-screen Windows CE/Mobile for PDA Windows XP/Vista/7 for PC Specificaons

Upload: others

Post on 18-Nov-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: X-RAY FLUORESCENCE SPECTROMETERS4.imimg.com/data4/LP/IB/MY-1598947/handheld-detector-si...Energy Dispersive X-ray Fluorescence (EDXRF) technique relies on the detector and the detection

AN ISO 9001-2008 COMPANY

Where innovation never stops.....

301, Punit Indl. Premises, Turbhe Naka, Navi Mumbai - 400 705. Tel. :+91 22 6139 3000 (100 lines) Fax :+91 22 2761 8421 E-mail : [email protected] / [email protected]: www.contechindia.com

HANDHELD DETECTORS (Si-PIN / SDD) CGX-104

X-RAY FLUORESCENCE SPECTROMETERS

XRF is used to detect and analyze noble metals like Gold ( Au), Silver (Ag), Platinum (Pt),Iridium (Ir), Osmium (Os), Palladium (Pd), Rhodium (Rh), Ruthenium (Ru), Copper (Cu), Zinc (Zn), Nickel (Ni), Cadmium (Cd), Iron (Fe), Cobalt (Co), Indium (In), Tin (Sn), in theJewellery Industry.

Energy Dispersive X-ray Fluorescence (EDXRF) technique relies on the detector and thedetection circuit capability to resolve spectral peaks due to different energy X-Rays emittedwhen a primary X-Ray beam strikes a metal alloy corresponding to different elements in thesample, intensity of the spectrum radiation proportional to the concentration of element.The spectrum is then processed through the powerful computer software which calculatesand reports the various elements & their composition in the sample.

XRF Gold Purity Tester consists of X-Ray tube, high voltage power supply for X-Ray source,X-Ray tube filters, X-Ray fluorescence Detector, processing electronics and system interface& an external computer. A camera is incorporated in the system to view the sample on thecomputer screen.

XRF is a reference technique, hence standards are required for quantitative results. Standards are analyzed, spectral intensities generated are used to obtain a intensity v/s concentration plot for calibration. This calibration is used to evaluate the spectral intensities of unknown samples.

How to get best results from XRF Spectrometer :XRF is a reference method, proper standardization of the equipment is required for accurate results.

In order to get most appropriate results, testing at different points is necessary especially when the jewellery design is intricate.

Adequate warm up period, sturdy working surface, stable power supply, proper grounding and conditioned atmosphere is required to achieve the best product performance.

For powder sample, proper grinding of sample minimizes scattering effect.

For solid sample, surface polishing minimizes the scattering effect.

Flat surface of the sample yields the best results.

Due to different melting points, atomic structure and crafting techniques, not all the elements can be melted with noble metals.

OPTIONAL LAB STAND SETUP

CGX-1

04

Handheld X-ray analyzer works on a rechargeable 7.2V Li-ion ba�ery with a backupperiod of 8 hours under normal use. XRF analyzer has a detectable range from Ti to U.CGX-104 operates under control of the HP iPAQ PDA with 4 inch high resolu�ontouch-screen display. The op�onal lab stand allows using of handheld analyzer as atabletop instrument under PC or laptop control.

Ba�ery Opera�on Time : Up to 8 hours

Dimensions : 242 mm (L) x 78 mm (W) x 230 mm (H)

Weight (with ba�ery) : 1230 g (1440 g)

X-ray tube

Detector type

Ac�ve area

Energy resolu�on

Display

Supported opera�ng systems

Tungsten anode, 40 kV , 100 µA with 5 filter posi�ons

solid state Si-pin-diode with thermo-electrical cooling(op�onal SDD)

26 mm

<165 eV at 5.9 keV (<140 eV for SDD)

High resolu�on 4 inch color touch-screen

Windows CE/Mobile for PDAWindows XP/Vista/7 for PC

Specifica�ons