welch allyn cp 50 electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... ·...
TRANSCRIPT
![Page 1: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/1.jpg)
Welch Allyn CP 50™ Electrocardiograph
For more information, contact your local Welch Allyn representative or visit www.welchallyn.com/CP50
Advanced ECG in a compact design that fit your facility and your budget. > Offered at an affordable price to help maximize your budget> Easy-to-use, with one-button operation, makes performing ECG tests efficient> Full-color, touch-screen display to help improve staff satisfaction> High-resolution thermal printer provides quick outputs> Battery operation lets you easily transport your device to the patient > Optional Mobile Stand and Carrying Case are also available
CP 50 ELECTROCARDIOGRAPH
SPECIFICATIONS
CP 50™ ECG Plus with ConnectivityIncludes all of the features of the CP 50, plus • Flexible connectivity options• External printing capability• Programmable patient entry fields
CP 50™ ECG• Reliable touch-screen display• Optional MEANS interpretation• High-resolution thermal printer• ‘Instant-on’ button• Internet storage
![Page 2: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/2.jpg)
CP 50 Technical SpecificationsDimensions L x H x W 12.8 in (32.5 cm) x 3.1 in (8 cm) x 7.4 in (18.8 cm)
Weight 4.4 lb (2.0 kg)
Keyboard type (power, auto ECG, and rhythm ECG buttons) Polyester overlay
DisplayType TFT, 4.3 in (11 cm) color touch screen
Resolution WQVGA, 480 x 272
Thermal paperZ-fold 4.5 in (11.4 cm) x 250 sheets
Roll 4.5 in (11.4 cm) x 65.6 ft (20 m)
Thermal printer (internal) Computer-controlled dot array, 8 dots/mm
Thermal chart paper speeds 10, 25, 50 mm/s
Gain settingsAuto ECGs 2.5, 5, 10, 20 mm/mV, AUTO
Rhythm ECGs 2.5, 5, 10, 20 mm/mV
Lead configurations Standard, Cabrera
Report print formats—Internal thermal printer (PCL 3 or 5 only)
Auto ECGs 3x4, 3x4 + 1R, 2x6
Rhythm ECGs 3 leads
Average cycles 3x4 + 3R @ 25 mm/s, 3x4 + 3R @ 50 mm/s, 2x6 + 1R @ 25 mm/s, 2x6 + 1R @ 50 mm/s
Report print formats—Compatible External Inkjet or LaserAuto ECGs 3x4, 3x4 + 1R, 2x6 , 3x4 + 3R, 6x2, 2x6, 12x1
Average cycles 3x4 +3R @ 25 mm/s, 3x4 + 3R @ 50 mm/s, 2x6 + 1R @ 25 mm/s, 2x6 + 1R @ 50 mm/s , 3x4 + 2R @ 25 mm/s, 3x4 + 2R @ 50 mm/s, 6x2 + 2R @ 50 mm/s
Frequency range 0.3 to 150 Hz
Digital sampling rate > 1,000 samples/second/channel
Pacemaker detection ANSI/AAMI EC11
Power requirement Universal AC power supply ~110–240 V, ~50/60 Hz, 1.5 A maximum
AC fuses Time-log type, 2.0-amp 250-V rating, Littlefuse 0218002P or equivalent
Rechargeable battery 10.8 V, 2000 mAh, 3-cells Li-Ion “smart” battery. Recharge time < 3 hours
Filters
High-performance baseline 0.5 Hz
Muscle tremor 35 Hz
AC interference 50 Hz or 60 Hz
Internal ECG Storage Up to 50 ECGs
External ECG Storage Unlimited via USB connection
Safety, EMC, and regulatory compliance
ANSI/AAMI EC11* UL60601-1
CAN/CSA C22.2 No. 601.1 IEC/EN 60601-1
CAN/CSA C22.2 No. 601.1.1 IEC/EN 60601-1-1
CAN/CSA C22.2 No. 601.1.2 IEC/EN 60601-1-2
CAN/CSA C22.2 No. 601.1.4 IEC/EN 60601-1-4
CAN/CSA C22.2 No. 601.2.25 IEC/EN 60601-1-6 IEC/EN 60601-2-25IEC/EN 60601-2-51 (2x6 lead arrangement)
Standard connectivity Mini USB Client and USB Host (CP 50 Plus version only)
Ethernet (CP 50 Plus version only)
Connectivity with electronic medical records (CP 50 Plus only) Through the Welch Allyn CardioPerfect™ Workstation software
Electrodes Rigorously tested for conductivity, adhesion, and hypoallergenic qualities; exceeds all AAMI standards
Power cable Meets or exceeds Type SJT
Patient cable and leads Meets or exceeds ANSI/AAMI EC53, EN/IEC 60601-2-25 and EN/IEC 60601-2-51
Environmental operating conditions
Temperature +50° F to +104° F (+10° C to +40° C )
Relative humidity 15 - 95% noncondensing (30 - 70% for printing)
Atmospheric air-pressure limits 700 – 1060 hPa
Environmental storage conditions
Temperature -4° F to +122° F (-20° C to +50° C )
Relative humidity 15 - 95% noncondensing
Atmospheric air-pressure limits 700 – 1060 hPa
Protection against electric shock Class I, internally powered Type BF
Mode of operation Continuous
* Per AAMI EC11:1991/(R)2001 Diagnostic Electrocardiographic Devices, Section 3.1.2.1 Disclosure of cautionary information/performance characteristics paragraph c) Accuracy of input signal reproduction, the manufacturer shall disclose the methods used to establish overall system error and frequency response. Welch Allyn has used methods A & D, as prescribed in section 3.2.7.2 and 4.2.7.2 of this same standard, to verify overall system error and frequency response. Because of the sampling characteristics and the asynchronism between sample rate and signal rate, digital ECG systems such as the CP 50 may produce a noticeable modulating effect from one cycle to the next, particularly in pediatric recordings. This phenomenon is not physiologic. Specifications are subject to change without notice.
©2013 Welch Allyn MC10387Welch Allyn Corporate Headquarters, 4341 State Street Road, Skaneateles Falls, New York 13153 USA (p) +1.800.535.6663
![Page 3: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/3.jpg)
Welch Allyn CP 150™ Electrocardiograph with optional SpirometrySimple. Fast. Connected.
![Page 4: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/4.jpg)
Simple, streamlined ECG testing with a range of advanced features to help improve your workflow.The CP 150 combines the best traditional-box ECG features like full-size
printouts and clinical decision support with MEANS interpretation
with a simple touch-screen interface, fast one-button operation,
and flexible connectivity options to help you keep your staff
productive and your tests accurate. In fact, 81% of clinicians
prefer the CP 150 over the competition.*
*Source: Walker Information, Inc.
Simple
The 7-inch, touch-screen display with on-screen prompts guides staff through a test, to help enhance accuracy, minimize training and save time.
Avoid repeat ECGs with the lead quality display that is designed to help technicians quickly and easily identify lead connection problems before completing a test.
Keep your practice running with the Welch Allyn CP 150 Resting ECG
of CP 150 users say the touchscreen display and simple user interface has helped them improve their practice efficiency. Source: Survey by Walker Information, Inc. of clinicians who have purchased the CP 150.
94%
![Page 5: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/5.jpg)
Fast
Instant-On FeaturePowers the device up quickly, and
eliminates the typical startup delay
to take an ECG.
12-Lead ECG PreviewLarge display provides three, six,
or 12-lead ECG preview options.
ECG Print PreviewSave time and reduce paper costs
by verifying the ECG test quality
prior to printout.
The CP 150 spirometry option features four incentive screens to help educate and coach your patients:
Spirometry OptionThe CP 150 lets you perform two tests with one device, to help make detection at the point of care easier and more efficient.
> Perform FVC and pre- and post-BD testing with
automatic interpretation and reversibility analysis
> Customize report formats and screen displays to
fit your needs and workflow
> Performs spirometry tests typically required for health
and safety purposes including ATS and NIOSH
> Automatic lung age calculation to help with
smoking cessation programs
> Integrated ambient sensors that eliminate
the need for the user to find and enter values
for temperature, pressure and humidity
![Page 6: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/6.jpg)
Device Configurations:CP150A-1ENB CP 150 Interpretive ElectrocardiographCP150—1ENB CP 150 Non-Interpretive Electrograph CP150AS-1ENB CP 150 Interpretive Electrocardiograph
with SpirometryCP150S-1ENB CP 150 Non-Interpretive Electrocardiograph
with Spirometry
Accessories:715006 ECG Multifunction Electrode Adaptor 45008-0000 Resting Tab Electrodes (box of 1000) 105353 CP 150 Recording Paper
(case, 1000 sheets)719653 10-lead Patient Cable, AHA, Banana,
CP 150 105341 CP 150 Office Cart105342 CP 150 Hospital Cart105343 CP 150 Cable Arm & Shelf Cart
Spirometry Accessories:105660 Spirometry Upgrade for CP 150
(works with any standard CP150A-1ENB or CP150-1ENB configuration)
720705 Disposable Flow Transducers, 25 pk, CP 150720706 Disposable Flow Transducers, 100 pk,
CP 150 / Spiro
Partners in CareSM Service Options:S1-CP150 CP 150 Comprehensive Partner ProgramS-1-CP150-2 CP 150 Comprehensive Partner Program, 2 YRS1-CP150S CP 150 Spiro Comprehensive Partner ProgramS1-CP150S-2 CP 150 Spiro Comprehensive Partner
Program, 2 YR
Warranty Information:CP 150 Electrocardiograph—3 yearsPatient Cable/Battery—90 daysSpirometer—1 year
Simple, Streamlined EMR Integration • Save ECG tests in PDF format• Export tests to USB and drag-and-drop file into folder in EMR• Use the integrated ethernet connection and export PDF directly to a folder on your network
Bring your cardiopulmonary data together with the Welch Allyn CardioPerfect® Workstation• Store, access, review, compare and edit diagnostic test data from any computer
on your network, in any exam room, or outside the office• Ability to interface with more than 40 Electronic Health Record systems
Take your CP 150 on the go• Attractive, ergonomic mobile cart features optional hospital
grade wheels and patient cable arm and shelf• Long-lasting lithium ion battery gives you more on-time,
letting you transport your device directly to the patient
Ordering Information
Connectivity Options that mean you’ll never have to scan and shred an ECG test again.The CP 150 features a range of connectivity options to match your workflow
© 2014 WELCH ALLYN MC11412 SM4091 REV D
Welch Allyn Corporate Headquarters4341 State Street Road, P.O. Box 220 Skaneateles Falls, NY 13153-0220 USA(p) 800.535.6663 (f) 315.685.3361
WWW.WELCHALLYN.COM
Contact your Welch Allyn representative or visit www.welchallyn.com/cp150 today for more information.
![Page 7: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/7.jpg)
The Welch Allyn PC-Based Resting ECG
Spend less time with paperwork and more time with your patients.
![Page 8: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/8.jpg)
Convert to digital PC-based ECG technology and boost your efficiency with the Welch Allyn PC-Based Resting ECG.
Tired of printing, scanning and shredding ECG records?
Improve Workflow and Efficiency> Preview a complete 12-lead report on large screen prior to printing
and eliminate repeat ECGs
> ECG measurements speed physician review and interpretation
> Flexible reports that you can print on plain (standard) paper using an inkjet or laser printer. No need for expensive thermal paper. No more fading problems
Make Clinical Decisions with Confidence > Serial comparison of previous ECGs to highlight changes
> Easily monitor patients with intermittent arrhythmias
> Connect with your EHR and eliminate the need to sort through lengthy paper reports or interpret hard-to-read faxed copies of test results
“The process of patient data gathering using the Welch Allyn product is excellent—now I don’t have to read 30 and 40 pages of ECGs. Now with the Welch Allyn devices, you hit one button
and everything is there for you. The process is so streamlined now, and that allows me to focus more on my patients.”
— Sasha Van Dyke, Exercise Physiologist, Cardiology of Tulsa
All the benefits of a full-featured, PC-based ECG in an affordable solution that speeds the turnaround of test results, makes patient diagnostic information more
readily available and enhances communication for you, your staff, and your patients.
![Page 9: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/9.jpg)
Convert to digital PC-based ECG technology and boost your efficiency with the Welch Allyn PC-Based Resting ECG.
Tired of printing, scanning and shredding ECG records?
Efficient Diagnosis Optional MEANS ECG interpretation software available for both adults and pediatric patients. Use on-screen report editor to change or add a diagnosis in seconds. Or select from preset menus customized to your needs, for simpli-fied review and editing of diagnostic tests.
Improved Condition Management Easily monitor trends and changes in patient history. Retrieve previous records with the touch of a key—no more digging through paper files looking for the right information.
Connect to Your EHRNo printing or scanning, as data is already in the chart. Fax urgent referrals in no time directly from the system. And the database stores thousands of test records.
Bidirectional Connectivity Right at Your Fingertips Most devices only push information out to the EHR. When you use one of our certified EHR partners, you benefit from our bidirectional connectivity. This enables immediate sharing of patient demographics and critical test data between the EHR and the ECG, so information is readily available for review anytime, anywhere. The benefit? You reduce transcription errors because you don’t have to continually re-enter data.
Maximize the power of your PC-Based ECG with the paperless efficiency and connectivity of the Welch Allyn CardioPerfect™ Workstation software.
Data Access Where and When You Need It
![Page 10: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/10.jpg)
© 2010 Welch Allyn SM4006 Rev B
4341 State Street Road, PO Box 220, Skaneateles Falls, NY 13153-0220(p) 1.800.535.6663 (f) 1.315.685.2174 www.welchallyn.com
Technical Specifications Digital ECG module• Dimensions (L x W x D): 5.8" x 4.3" x 1.3" (147 mm x 108 mm x 32.5 mm)• Weight: 222 g, 0.48 lbs• Integrated defibrillation protection• Computer interface: USB• 10-lead patient cable with banana electrode adapters• Frequency response: 0.05 – 150 Hz• Sampling rate: 600 sps
Filters• 25 to 35 Hz muscle tremor filter• Selectable 50/60 Hz
Software• Monitoring formats: 1 x 12, 4 x 3, 2 x 6• Print formats: 4 x 3, 2 x 6, average complexes, expanded
measurements matrix and rhythm strips• Print speeds: 5, 10, 12.5, 25, 50 and 100 mm/sec• Lead sensitivity: 5, 10 and 20 mm/mV
• Selectable QTc formula: Bazett, Hodges or Fridericia
Minimum Computer Requirements (Recommended)•� Configuration: Tower or Desktop PC preferred; Laptop or Tablet.
Some tablet computers may not support data recording• CPU: Pentium 4 (or equivalent) or higher• Hard disk: 20 GB with 5 GB free space• RAM: 512 MB, 1 GB or higher recommended• Monitor Resolution: 1024 x 768 True Color, 1280 x 1024 recommended,
15” or higher• Ports: 2 USB• Operating system: Windows XP SP2 (32 bit) or later, Vista Business SP1
or later (32 and 64 bit), Windows 7 Home Premium, Professional and Ultimate (32 and 64 bit)
• Thin Client/Virtual Channel Client: -Citrix V. 6 to 11 -Microsoft Terminal Services (RDP) with above mentioned OS
• Built-in PerfectLink™ technology allows you to use your regular ProLink USB cable for optimal thin-client performance
• Peripherals: Compatible with inkjet and laser prints
NOTES: Internet connection required for telemedicine option or to allow Welch Allyn to provide on-line technical support.PC running only CardioPerfect Workstation and EHR application during tests.
Ordering Information
CPR-UN-UB-D PC-Based Resting ECG, noninterpretive software—12-lead resting ECG system including: CardioPerfect™ Workstation software; digital ECG patient cable with USB connection; software for resting ECG—ECG measurements, rhythm/manual mode, average complexes, serial compari-sons of ECGs and measurements; 500 resting tab electrodes and electronic operator’s manual
CPR-UI-UB-D PC-Based Resting ECG, interpretive software—12-lead resting ECG system including: CardioPerfect™ Workstation software; digital ECG patient cable with USB connection; software for resting ECG—ECG measurements/interpretation, rhythm/manual mode, average complexes, serial comparisons of ECGs and measurements; 500 resting tab electrodes and electronic operator’s manual
101442 CPWS Resting ECG, interpretive software, and Spirometry Module Bundle— Includes all hardware components included in CPR-UI-UB-D and SPIRO-S
101444 CPWS Resting ECG, interpretive software, and ABPM-6100S Module Bundle— Includes all hardware components included in CPR-UI-UB-D and ABPM-6100S
101455 CPWS Resting ECG, interpretive software, and SpiroPerfect and ABPM-6100S Module Bundle—Includes all hardware components included in CPR-UI-UB-D, SPIRO-S and ABPM-6100S
Welch Allyn PC-Based Resting ECG
CPT® Codes93000 Electrocardiogram, routine ECG with at least 12 leads; with interpretation and report
93005 Tracing only, without interpretation and report
93017 Interpretation and report only
93040 Rhythm ECG—1 to 3 leads, with interpretation and report
CPT is a registered trademark of the American Medical Association.
Shown with #401393 ECG Cart, #102794 Cable Arm and Shelf
![Page 11: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/11.jpg)
VALUE WITHOUT COMPROMISE
• Complete functionality in a compact, portable, device. Easy-to-use and feature-rich, the Burdick ELI 230 offers acost-effective ECG solution with capabilities such as digital sampling, accurate and reliable adult and pediatric ECGinterpretation, and full-size printouts.
• High-resolution color display provides real-time preview of 3, 8, or 12-leads of ECG data, as well as post-acquisitionreview of acquired ECG.
• VERITAS™ resting ECG interpretation algorithm uses gender specific and adult and pediatric criteria to provide a silentsecond opinion for diagnostic interpretation.
SMALL IN STATURE, BIG IN CAPABILITY
• Choose between the innovative WAM™ wireless acquisition module or the AM12™ acquisition module. Both includereplaceable lead wires, lead fail indicator, and remote control with buttons for ECG acquisition and rhythm printing.
• Our unique Best 10 feature automatically selects and displays the best 10 seconds of ECG data from the memory,reducing clinical review time and reducing the need for repeat ECGs caused by patient movement or other sourcesof ECG artifact.
• Digital sampling technology provides in-depth ECG signal assessment including 40,000 sps per channel for unsurpassedpacemaker spike detection.
ELI 23012-lead RestingElectrocardiograph
PRIMARY BENEFITS:
• Portable solution. Compact and lightweight, theELI™ 230 provides comprehensive functionalityin a portable device.
• Best 10 technology. Capture the best 10 secondsof ECG data from system memory, reducing theneed for repeat ECGs.
• Wireless data acquisition. Increase workspaceflexibility by acquiring 12-lead data without theneed for a tethered cable.
COMPREHENSIVE AND PORTABLE ELECTROCARDIOGRAPH
![Page 12: Welch Allyn CP 50 Electrocardiographi2admin02.webstorepackage.com/allmed/virtualweb/images... · 2017-02-21 · CP 50 Technical Specifications Dimensions L x H x W 12.8 in (32.5 cm)](https://reader033.vdocuments.mx/reader033/viewer/2022042914/5f4d2eaa12a0d14da47b7254/html5/thumbnails/12.jpg)
MORTARA INSTRUMENT, INC.
MORTARA RANGONI EUROPE, SRL
MORTARA INSTRUMENT AUSTRALIA
MORTARA DOLBY UK LTD.
www.mortara.com ISO 13485 CERTIFIED
WARRANTY + SERVICE
Ref: 9513-175-02-ENG Rev 1
*Specifications subject to change without notice.
Mortara®, AM12™, Burdick®, ELI™, VERITAS™, andWAM™ are trademarks or registered trademarks ofMortara Instrument, Inc.
ELI™ 230 ELECTROCARDIOGRAPH
FEATURE SPECIFICATION*
INSTRUMENT TYPE 12-lead electrocardiograph
INPUT CHANNELS Simultaneous acquisition of all 12 leads
STANDARD LEADS ACQUIRED I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
WAVEFORM DISPLAY Backlit, 1/4 VGA 320 x 240 LCD color display4+4 or 6+6 lead presentation
INPUT IMPEDANCE Meets or exceeds the requirements of ANSI/AAMI EC11INPUT DYNAMIC RANGEELECTRODE OFFSET TOLERANCECOMMON MODE REJECTION
PATIENT LEAKAGE CURRENT Meets or exceeds the requirements of ANSI/AAMI ES1CHASSIS LEAKAGE CURRENT
DIGITAL SAMPLING RATE 40,000 s/sec/channel used for pacemaker spike detection;1000 s/sec/channel used for recording and analysis
RESOLUTION 1.875 microvolt LSB
A/D CONVERSION 20 bits
FREQUENCY RESPONSE 0.05 to 300 Hz
FILTERS High-performance baseline filter; AC interference filter 50/60Hz; low-pass filters 40 Hz, 150 Hz, or 300 Hz
LEAD REVERSAL DETECTION Notification of possible lead reversal conditions, for both limband chest leads, indicated on the display, prior to printing
PACEMAKER SPIKE DETECTION Pacemaker spike markers, for both atrial and ventricularrhythms, indicated on printouts
OPTIONAL FUNCTIONS Optional Mortara VERITAS resting ECG interpretation with ageand gender specific algorithm
PAPER Thermal roll paper; 210 mm (8.25”) wide
THERMAL PRINTER Computer-controlled dot array; 8 dots/mm
THERMAL PRINTER SPEEDS 5, 10, 25, or 50 mm/s
GAIN SETTINGS 5, 10, or 20 mm/mV
REPORT PRINT FORMATS Standard or Cabrera; 12, 6, or 3+1 channel
RHYTHM PRINT FORMATS 12, 6, or 3 channel with configurable lead groups
DEVICE CLASSIFICATION Class I, Type CF defibrillation-proof applied parts
ECG STORAGE Internal storage up to 20 ECG records; external storage toUSB memory stick
CONNECTIVITY Review or print from ELI Link V4.xx
WEIGHT 5.8 lbs. (2.63 kg) including battery (without paper)
DIMENSIONS 11.25 x 7.5 x 2.75" (28.58 x 19 x 7 cm)
POWER REQUIREMENTS Universal AC power supply (100-240 VAC at 50/60 Hz) 110 VA;internally rechargeable battery