![Page 1: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/1.jpg)
Real Time AirborneViable Particle Detection
TechnologyOverviewandApplicabilityToPharmaceuticalQualitySystems
![Page 2: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/2.jpg)
© 2012, TSI Incorporated
OverviewIntroductionTechnologyOverviewLaserInducedFluorescenceCriticalOperationalParameters
ApplicationScenariosApplicationroadmapReal‐timeevaluationofaseptictechniquesRootcauseinvestigationsQualityriskmanagement
![Page 3: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/3.jpg)
© 2012, TSI Incorporated
Introduction: The Desired State“Amaximallyefficient,agile,flexiblepharmaceuticalmanufacturingsectorthatreliablyproduceshighqualitydrugproductswithoutextensiveregulatoryoversight”
Anopportunityforalessrestrictiveregulatoryparadigmanddrivesinnovationandcontinualimprovementinmanufacturingprocesses
QbD ICH Q8(R2)
QRM ICH Q9
PQS ICH Q10
PQSICH Q10
QbDICHQ8(R2) QRM
ICH Q9
PAT
PAT
![Page 4: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/4.jpg)
© 2012, TSI Incorporated
Pursuit of Faster Results:Traditional Microbial Monitoring
Process Environment
Review Levels
Above LimitBelow Limit
Sampling Incubation Results Reaction
2 - 14 Days
![Page 5: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/5.jpg)
© 2012, TSI Incorporated
Pursuit of Faster Results:Lab-Based RMM
2-14 Days
Majority of RMM’s Reduce Incubation Time
8-48 hours
![Page 6: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/6.jpg)
© 2012, TSI Incorporated
Pursuit of Faster Results:Real-Time Viable Particle Detection
+ BioTrak™Detector• NewcategoryofRMM• Opticalspectroscopy• LaserInducedFluorescence(LIF)
+ Real‐timeresults
Seconds
![Page 7: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/7.jpg)
© 2012, TSI Incorporated
Risk Analysis
Root Cause Investigations
Segregate Product
TrainingQualification
QRM ICH Q9
CAPA
QbD ICH Q8
PAT
Environmental Monitoring in Real-Time
Real-time
viable particle
data
KnowledgeInformation
FMS
Respond in real-
time
![Page 8: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/8.jpg)
© 2012, TSI Incorporated
ISO 21501-4 compliant particle counter 28.3 lpm
Three Stages of TSI’s BioTrak™ Real-Time Particle Counter
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
37mm Collection Filter for speciation of analyzed particles
Viability detector-Laser Induced Fluorescence
![Page 9: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/9.jpg)
© 2012, TSI Incorporated
• Proven TSI AeroTrak 9510 Airborne Particle Counter Engine
• ISO 21501-4 Compliant Specification and Calibration
• Full Airborne Particle Counter Functionality
Airborne Particle Counter
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
![Page 10: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/10.jpg)
© 2012, TSI Incorporated
Particle Concentrator
Clean air and small particles
Most particles of interest
The flow is slowed down for the viability detector using a concentrator
28.3 lpm flow is too fast viability detection
![Page 11: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/11.jpg)
© 2012, TSI Incorporated
Particle Concentrator
What impact does the concentrator have on the particle transport
efficiency?
Air Samplers also have particle transport
efficiency concerns.
![Page 12: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/12.jpg)
© 2012, TSI Incorporated
Laser Induced FluorescenceEx
cita
tion
Wav
elen
gth
Emission Wavelength
Laser Induced Fluorescence (LIF)
Viable cell metabolites tryptophan, NADH, and flavin’s (riboflavin) fluoresce when
excited by ultra-violet light
“Advanced Trigger Development” Jeyes et al, Lincoln Laboratory Journal Volume 17, Number 1, 2007
![Page 13: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/13.jpg)
© 2012, TSI Incorporated
Laser Induced Fluorescence• Wanttodifferentiatethemicroorganismsfromthenonviableparticleswithfluorescenceproperties
• Difficulttaskevenwithfullexcitationandemissionspectruminformation
• Notpossibletoimplementfullspectroscopyinaffordableportableinstruments
• TSICollaborativedevelopmentsince1990’s
• TSIhasfundamentalsingleparticleLIFpatents*
*TSI Patents 5,701,012; 5,895,922; 6,831,279
![Page 14: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/14.jpg)
© 2012, TSI Incorporated
Laser Induced Fluorescence
![Page 15: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/15.jpg)
© 2012, TSI Incorporated
Viability Detector
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
• Patented* Laser Induced Fluorescence (LIF) Viability Detection
• TSI collaborative technology development over the past 20 years
*Patents 5,701,012; 5,895,922 and 6,831,279
![Page 16: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/16.jpg)
© 2012, TSI Incorporated
Discrimination-Impact of Additional Optical Parameters
1610/18/2012© TSI Incorporated
Single Fluorescence Channel
Two Fluorescence Channels
![Page 17: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/17.jpg)
© 2012, TSI Incorporated
Discrimination-Another way to visualize
1710/18/2012© TSI Incorporated
+ Algorithms are used to calculate a discrimination parameter+ Particles with a discrimination value higher than a threshold
are classified as viable+ Particles with optical signatures near the intersection have
potential to be misclassified (False Positive or False Negative)
A
B
AB
Discrimination Parameter
![Page 18: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/18.jpg)
© 2012, TSI Incorporated
Discrimination
1810/18/2012© TSI IncorporatedDiscrimination Parameter
BetterDiscrimination+ Numberofanalysis
parameters+ Algorithms+ Orthogonal
measurements
![Page 19: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/19.jpg)
© 2012, TSI Incorporated
Sensitivity
Discrimination Parameter
Discrimination Parameter
![Page 20: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/20.jpg)
© 2012, TSI Incorporated
LIF Dual Sensitivity
+ Abilitytoadjustinstrumentresponsetoapplicationspace
“Clean” Environments• Low number of viable particles• Low number of total particles• Lower number of particles with fluorescent characteristics• Unacceptable to miss viable particles• HIGH SENSITIVITY
“Dirty” Environments• Higher number of viable particles• Higher number of total particles• Don’t want excessive false positives• Higher limit of viable particles acceptable to miss some• LOW SENSITIVITY
![Page 21: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/21.jpg)
© 2012, TSI Incorporated
Collection Filter
37mm Collection Filter: Provides sample for speciation of optically analyzed particles
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
![Page 22: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/22.jpg)
© 2012, TSI Incorporated
Real-Time Viable Particle Counting -Important Considerations+ Needtounderstandtheproduct’scapabilitiesandlimitations
• EffectivenessofLIFMeasurement‐ Discriminationcapability‐ FalsePositivesversusFalseNegatives
• Functionality‐ Easeofuse‐ Totalparticulatesizingcapability‐ Viableparticledetectioncapability‐ Dataavailability
• Particlecapturecapability• RegulatoryChallenges
![Page 23: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/23.jpg)
© 2012, TSI Incorporated
Characterization of Real-Time Viable Particle Counter
+ Needstobecharacterizedwithexistingmethods
+ Comparabilitystudies:efficiencyoftheinstrumentundertestandthereferencesamplermustbeconsidered
+ Desirecorrelationtoexistingmethods
![Page 24: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/24.jpg)
© 2012, TSI Incorporated
Regulatory Guidance-Challenges
Compendial Method
Rapid Method
ISO 14698-1 USP 1223, EP 5.1.6, PDA TR33
Real-Time Aerosol Viable Detection
Active Air Sampling Comparison: Equivalent or Better
![Page 25: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/25.jpg)
© 2012, TSI Incorporated
Applicability+ LIFresultswillbedifferentthancurrentactivesamplerbasedresults
+Whatisthevalue?• Real–timedata• Continuousdata• Characterizationofprocesses• Trainingfeedback
+ Evolutiontoreal‐timerelease
![Page 26: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/26.jpg)
© 2012, TSI Incorporated
Application Road Map“The Innovation Highway”
Vendor Validation
Comparability Studies
Operator Training
Real-time
release
Root cause Investigations
Monitoring Locations+ Zone Qualification
Validated in Process
Measurement
Immediate benefit Long term benefit
Confiden
ce in
the measuremen
t
![Page 27: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/27.jpg)
© 2012, TSI Incorporated
Risk Analysis
Root Cause Investigations
Segregate Product
TrainingQualification
QRM ICH Q9
CAPA
QbD ICH Q8
PAT
Environmental Monitoring in Real-Time
Real-time
viable particle
data
KnowledgeInformation
FMS
Respond in real-
time
![Page 28: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/28.jpg)
© 2012, TSI Incorporated
New Information: Grade C Example
Grade C Monitoring LocationIn Operation
![Page 29: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/29.jpg)
© 2012, TSI Incorporated
0.00
500.00
1000.00
1500.00
2000.00
2500.00
3000.00
3500.00
4000.00
4500.00
5000.00
08:24 09:36 10:48 12:00 13:12
TCNT
#/M^3
11 AM
ParticleCountscomplianttoISO21501‐4
0.5 µm Counts
5 µm Counts
Root Cause Investigations
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
![Page 30: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/30.jpg)
© 2012, TSI Incorporated
0.00
10.00
20.00
8:24 9:36 10:48 12:00 13:12
VCNT
Viable Counts
12:30PM
Particleviabilitydetector
Root Cause Investigations
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
#/M^3
![Page 31: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/31.jpg)
© 2012, TSI Incorporated
Root Cause investigations
0.00
500.00
1000.00
1500.00
2000.00
2500.00
3000.00
3500.00
4000.00
4500.00
5000.00
08:24 09:36 10:48 12:00 13:12
TCNT
0.00
10.00
20.00
8:24 9:36 10:48 12:00 13:12
VCNT
12:30 PM11 AM
2 different conclusions…..
OR
#/M^3#/M^3
![Page 32: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/32.jpg)
© 2012, TSI Incorporated
0.00
500.00
1000.00
1500.00
2000.00
2500.00
3000.00
3500.00
4000.00
4500.00
5000.00
08:24 09:36 10:48 12:00 13:12
TCNT
RootcauseInvestigationsRoot Cause Investigations
#/M^3
![Page 33: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/33.jpg)
© 2012, TSI Incorporated
Root Cause Investigations Collection Filter
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
![Page 34: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/34.jpg)
© 2012, TSI Incorporated
RootCauseInvestigations
ICH Q8(R2) – Improved Process Knowledge
0.00
10.00
20.00
8:24 10:48 13:12
The BioTrak detector high temporal resolution means you know when a microbial
excursion occurred
Leads toImproved root cause
investigations
![Page 35: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/35.jpg)
© 2012, TSI Incorporated
IdentifyingMicrobiologicalContaminationSources
ICH Q8(R2) – Improved Process Knowledge
InvestigationToolQuickly determinethesourceofairbornemicrobiologicalcontamination“GeigerCounter“mode– UserconfigurableBeepforViablecounts
![Page 36: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/36.jpg)
© 2012, TSI Incorporated
RootCauseInvestigations
Leads to
ICH Q8(R2) – Improved Process Knowledge
Understanding the source of airborne microbiological contamination will help
determine the cause
Improved root cause
investigations
![Page 37: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/37.jpg)
© 2012, TSI Incorporated
RootCauseInvestigations
Leads to
ICH Q8(R2) – Improved Process Knowledge
Improved root cause
investigations
More effective CAPA
![Page 38: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/38.jpg)
© 2012, TSI Incorporated
QualificationTool:Earlyriskbasedreleaseof(manufacturing)areas‐Supportedbyexistingmethods:resultsdayslater
ICH Q9 – Quality Risk Management
Loc 2Loc 1
Loc 3 Loc 4
Loc 5
6m
Grade B
Loc 6
6mBioTrak
Real-time OPC and
Airborne Viables
OPC Monitoring
location
Environmental Monitoring
location
![Page 39: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/39.jpg)
© 2012, TSI Incorporated
RiskAnalysis:DatatosupportmonitoringpositionrationaleforSettlePlates,ActiveAirSamplingandParticleMonitoring:Monitoringlocationfinetuning
ICH Q9 – Quality Risk Management
Loc 1Loc 1
Grade BBioTrak
Real-time OPC andAirborne Viables
?OPC Monitoring
location
Environmental Monitoring
location
![Page 40: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/40.jpg)
© 2012, TSI Incorporated
Training‐ ImmediatenotificationofpoorpracticeGowning
ICH Q9: Quality Risk Management
![Page 41: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/41.jpg)
© 2012, TSI Incorporated
Training‐ ImmediatenotificationofpoorpracticeAseptictechnique
ICH Q9: Quality Risk Management
![Page 42: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/42.jpg)
© 2012, TSI Incorporated
Training‐ ImmediatenotificationofpoorpracticePoorcleanroompractice
ICH Q9: Quality Risk Management
![Page 43: Webinar - Real Time Airborne Viable Particle Detection 2012...Data to support monitoring position rationale for Settle Plates, Active Air Sampling and Particle Monitoring : Monitoring](https://reader034.vdocuments.mx/reader034/viewer/2022042711/5f6d7ebae34f494fb6370a01/html5/thumbnails/43.jpg)
© 2012, TSI Incorporated
Innovation HighwayReal-Time Release
Vendor Validation
Comparability Studies
Operator Training
Goal
Root cause Investigations
Monitoring Locations+ Zone Qualification
Validated in Process
Measurement
Short to medium term benefit Long term benefit
Confiden
ce in
the measuremen
t