Vivid S70 product datasheet – June 2015 – DOC1662324 Page 1 of 18 excellent image quality with little manipulation along with automated tools like 2D Auto EF, AFI Productivity Package and Scan Assist Pro. Ease of use in 4D imaging is accom- plished with a number of GEHC innovations, including Single Beat 4D, 4D Views, Advanced 4D User Toolbox, Advanced 4D User Quantification Package and 4D Auto LVQ. Ergonomic features include the “FlexFit” mechanism enabling continu- ous pivoting height adjustment of the control panel, allowing the user to adjust distance to the control panel while providing the adequate legroom for standing or sitting positions. In addition, the articulating monitor arm (horizontal and vertical), and lightweight transducers combine to make the Vivid S70 an extremely ergonomic-friendly cardio- vascular ultrasound system. Portability – The Vivid S70’s compact size and light weight, combined with a fold-down monitor, enables easy transportation and promotes scanning at the patient site. The battery option provides a transportation mode that keeps the system ready to scan within a few seconds of being connected to a power outlet. The cSound platform takes GE’s Raw Data to a new level. For image processing and reconstruction, the Vivid S70 utilizes more than 100 times the data compared to the Vivid S6. Additionally, the Vivid S70 uses the proven Raw data format technology that allows for advanced processing on archived images by applying many of the same scan controls and advanced quantitative tools as are available during the original exam. cSound delivers volume sizes suited for full volume single-beat and multi-beat 4D acquisition (option). Using both coherent and harmonic image processing, the system provides computational power, ease of imaging, workflow flexibility and product upgradeability. The Vivid S70 excels in the following areas: Exceptional image quality on the Vivid S70 is created through the use of Ultra Definition Clarity filtering, elevation compound imaging (considering a wider slice for 2D imaging) with the 6VT probe, HD Imaging (balanced resolution, penetration and image uniformity) and virtual apex (larger field-of-view) for the FPA probes. Probe Technology – The XDclear series of probes are designed to help deliver powerful and efficient sound waves, with high bandwidth and efficiency. XDclear probe technology provides impressive deep penetration and high sensitivity while maintaining high spatial resolution. The combination of Single Crystal, Acoustic Amplifier and Cool Stack technologies is the core technology of the XDclear series of probes. Ease of use features make Vivid S70 an extremely productive 2D and 4D cardiovascular ultrasound system. The combination of the touch screen control with conventional (tactile) buttons provide intuitive controls, helping the operator maintain focus on the patient and the ultrasound images during the exam. The touch screen can also provide alpha-numeric keyboard entry or a dedicated A/N keyboard option can be purchased. Ease of use for the operator in 2D imaging is provided by the cSound technology delivering auto optimized Product Description The Vivid S70 combines the proven breadth and performance of the Vivid product line with a new and innovative software-based image processing platform: cSound. The Vivid S70 is GE cardiovascular ultrasound’s high-end scanner. The cSound architecture benefits all Vivid S70 probes and applications. The Vivid S70 supports the following clinical applications: Adult cardiac (2D and 4D transesophageal), coronary, pediatric, fetal heart, obstetrics, gynecology, abdominal, small parts, thyroid, adult and neonatal cephalic, peripheral vascular, musculoskeletal, nerves, and urology. System Architecture GE’s innovative, programmable and flexible beamforming technology, cSound, provides exceptional image quality and power compared to conventional GE hardware-based beam forming technology. In 4D, Vivid S70

Upload: lamtu

Post on 06-Jun-2018

216 views

Category:

Documents


1 download

TRANSCRIPT

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 1 of 18

excellent image quality with little manipulation along with automated tools like 2D Auto EF, AFI Productivity Package and Scan Assist Pro.

Ease of use in 4D imaging is accom-plished with a number of GEHC innovations, including Single Beat 4D, 4D Views, Advanced 4D User Toolbox, Advanced 4D User Quantification Package and 4D Auto LVQ.

Ergonomic features include the “FlexFit” mechanism enabling continu-ous pivoting height adjustment of the control panel, allowing the user to adjust distance to the control panel while providing the adequate legroom for standing or sitting positions. In addition, the articulating monitor arm (horizontal and vertical), and lightweight transducers combine to make the Vivid S70 an extremely ergonomic-friendly cardio-vascular ultrasound system.

Portability – The Vivid S70’s compact size and light weight, combined with a fold-down monitor, enables easy transportation and promotes scanning at the patient site. The battery option provides a transportation mode that keeps the system ready to scan within a few seconds of being connected to a power outlet.

The cSound platform takes GE’s Raw Data to a new level. For image processing and reconstruction, the Vivid S70 utilizes more than 100 times the data compared to the Vivid S6.

Additionally, the Vivid S70 uses the proven Raw data format technology that allows for advanced processing on archived images by applying many of the same scan controls and advanced quantitative tools as are available during the original exam.

cSound delivers volume sizes suited for full volume single-beat and multi-beat 4D acquisition (option). Using both coherent and harmonic image processing, the system provides computational power, ease of imaging, workflow flexibility and product upgradeability.

The Vivid S70 excels in the following areas:

Exceptional image quality on the Vivid S70 is created through the use of Ultra Definition Clarity filtering, elevation compound imaging (considering a wider slice for 2D imaging) with the 6VT probe, HD Imaging (balanced resolution, penetration and image uniformity) and virtual apex (larger field-of-view) for the FPA probes.

Probe Technology – The XDclear™ series of probes are designed to help deliver powerful and efficient sound waves, with high bandwidth and efficiency. XDclear probe technology provides impressive deep penetration and high sensitivity while maintaining high spatial resolution. The combination of Single Crystal, Acoustic Amplifier and Cool Stack technologies is the core technology of the XDclear series of probes.

Ease of use features make Vivid S70 an extremely productive 2D and 4D cardiovascular ultrasound system.

The combination of the touch screen control with conventional (tactile) buttons provide intuitive controls, helping the operator maintain focus on the patient and the ultrasound images during the exam. The touch screen can also provide alpha-numeric keyboard entry or a dedicated A/N keyboard option can be purchased.

Ease of use for the operator in 2D imaging is provided by the cSound technology delivering auto optimized

Product DescriptionThe Vivid™ S70 combines the proven breadth and performance of the Vivid product line with a new and innovative software-based image processing platform: cSound.™ The Vivid S70 is GE cardiovascular ultrasound’s high-end scanner.

The cSound architecture benefits all Vivid S70 probes and applications. The Vivid S70 supports the following clinical applications: Adult cardiac (2D and 4D transesophageal), coronary, pediatric, fetal heart, obstetrics, gynecology, abdominal, small parts, thyroid, adult and neonatal cephalic, peripheral vascular, musculoskeletal, nerves, and urology.

System Architecture GE’s innovative, programmable and flexible beamforming technology, cSound, provides exceptional image quality and power compared to conventional GE hardware-based beam forming technology. In 4D,

Vivid S70

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 2 of 18

General Specifications

Dimensions and Weight• Width:54cm,21.4"

• Depth:76cm,30.2"

• Height:138cm–168cm, 54.4"–66.7"

• Minimumheightwith foldedscreen:112cm,44.4"

• Weight:<73 kg, 161 lbs

Electrical Power• Nominalinputvoltage:

100-240 VAC, 50/60 Hz

• Ratedpowerconsumption:500VA

Operating System• Windows® 7

Uninterruptible Power Supply• Batterybackupforstandby

• Incaseofpowerfailureoraccidentalshutdown, when power is restored within less than 20 minutes, the system automatically turns on instantly, maintaining exact system state prior to shutdown

• Forlongerperiodsofpowerinterrup-tions, the system automatically saves data and changes into “Standby” state

Console Design• Fiveactiveprobeports

• ECGport

• IntegratedHDD

• MultipleUSBports(front/back)

• IntegratedDVD-Rmultidrive(optional )

• On-boardstorageforB/W thermal printer

• Integratedspeakersfor premium sound

• Fourswivelwheels–frontwheelbreaks, rear wheels direction lock

• Integratedcablemanagement

• Easilyaccessibleremovable air filters for cleaning

• Frontandrearhandles

• Rearstoragetrays/baskets

• Handrest

• LCDtranslation (independent of console)

- 160 mm horizontal bidirectional

- 100 mm vertical height adjustment

- Swivel to any viewing direction

• Folddownandrotationlock mechanism for transportation

• Horizontalviewingangleofmorethan 170°

• Resolution:1680x1050pixels

• Automaticormanualdigitalbrightness and contrast adjustment for excellent viewing in different ambient light conditions (light sensor)

• Tintandbacklightadjustments

• Separateadjustmentforexternalmonitor brightness/contrast

System Overview

Probe Presets• Cardiac

• Stress(optional )

• Abdominal

• Peripheralvascular

• Fetalheart

• Pediatrics

• Neonatalcephalic

• Adultcephalic

• Smallparts

• Thyroid

• Musculoskeletal

• Urology

• Rodent(optional )

• Transesophageal

• OB/GYN

• Coronary

• LVOcontrast(optional )

• Nerves

Operating Modes• 2Dtissue

• 4Dtissue(optional )

• 2Dcolorflow

• 4Dcolorflow(optional )

User Interface

Operator Keyboard• ErgonomicFlexFitdesignwith

left/right swivel and up/down arm-mobility of keyboard and monitor permitting both physiological sitting or standing operation

• Touchkeyboardwithsupportforcharacters in 12 languages

• DrawertypeA/Nkeyboardwith adjustable backlight (option)

• Physicalkeyboardsupportfor international characters in 12 languages (option)

• Ergonomichardkeylayout

• Interactiveback-lightingof application-specific push buttons – adjustable back-light intensity

• Integratedgelholders

• Easy-to-learnuserinterfacewithintelligent keyboard

• Dedicatedrotaryforoverallgain for 2D-mode

• DedicatedgainrotaryforM-mode,CFMorDopplercontrolledby active mode

• Imagemanageronthetouch screen for quick review of image clipboard contents

Touch Screen• 12"ultra-high-resolution,widescreen

format, color, multi-touch LCD screen

• Interactiveuser-configurable dynamic software menu

• Backlightadjustment–automaticbylight sensor or manual

• Touch-panelcontrolofTGCsliders

• Touch-panelcontrolscontentcan be set to routine or extended usage

LCD Monitor• 19"widescreenHigh-Definition(HD)

flicker-free LCD display

• 256shadesofgrayand16.7millionsimultaneous colors available

• Articulatedmonitorarm

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 3 of 18

• 2Dangioflow

• ColorM-mode

• TissuevelocityM-mode

• ContinuouswaveDoppler

• TissueM-mode

• PulsedwaveDoppler

• AnatomicalM-mode

• CurvedanatomicalM-mode

• Tissuevelocityimaging

• Tissuetracking

• Tissuesynchronizationimaging (optional )

• Strainimaging(optional )

• Strainrateimaging(optional )

• TissuevelocityDoppler

• Bloodflowimaging(option)

• B-flow(option)

• 2Dstress(optional )

• AFIAutomatedFunctionImaging (optional )

• AutoEF(optional )

• 2Dvirtualapeximaging

• Bi-plane(optional )

• Tri-plane(optional )

• Bi-andtri-planewithcolor(optional )

• Codedphaseinversion

• Compoundimaging

• Extendedfield-of-view(LOGIQView)

• 4Dfull-volumeacanning–singlebeat and multi beat (optional)

Scanning Methods• Electronicsector

• Electronicvolume(option)

• Electronicconvex

• Electroniclinear

• CWpencil

Transducer Types• Sectorphasedarray

• Convexarray

• Lineararray

• Singlecrystalmatrixarray

• 2Dmatrixarray(option)

Accessories (options)

• InterfacecableforexternalECGandexternal respiratory

• ECGadapterforDIN-typepediatricselectrode leads

• Cablestoragebox

Display Modes• Liveandstoreddisplayformat:

Full size and split screen, both with thumbnails, for still and cine

• Instant-reviewscreendisplays12 simultaneous loops/images for a quick study review

• Selectabledisplayconfigurationofduplex and triplex modes: Side-by-side or top-bottom during live, digital replay and clipboard image recall

• Single,dualandquad-screenview

• Simultaneouscapability

- 2D+PW/CW

- 2D+CFM/TVI+PW

- 2D+CFM+CW

- 2D+CFM/Angio/TVI/SRI/TT/SI/TSI

- 2D+M/AMM/CAMM

- 2D+CFM/Angio/TVI/SRI/TT/SI/TSI+M/AMM/CAMM

- Real-time duplex or triplex mode

- Compound+M/CFM/PW

- 4D+CFM(optionalwith6VTprobeand 4D option)

- 2D + Bi-plane (optional with 6VT probe and 4D option)

- 2D+Bi-plane+CFM/TVI/SRI/TT/SI/TSI/AMM/CAMM(optionalwith 6VT probe and 4D option)

- 2D + Tri-plane (optional with 6VT probe and 4D option)

- 2D+Tri-plane+FM/TVI/SRI/TT/SI/TSI/AMM/CAMM(optionalwith 6VT probe and 4D option)

- 2D + color split screen (simultaneous mode)

• Selectablealternatingmodes

- 2Dorcompound+PW

- 2D+CW

- 2Dorcompound+CFM/PW

- 2D+CFM+CW

Standard 4D Features (available with the 4D option)• Single,dualormultiplecyclevolume

acquisition

• Bi-planeacquisitionincludestiltandrotate, and bi-plane prepare

• Tri-planeacquisition

• Multi-dimensional (bi-plane/tri-plane) color acquisition

• FlexiSlicewithdepthmode

• QuickRotate/Rotate

• 2Clickcrop

• Flipcrop

• Viewcrop

• Dynamicviewcrop

• Measurementonrender

• FlexiZoom

• Laserlines

• Depthcolorrender

• Automated4Dleftventricular quantification (LV volume and EF)

Advanced 4D Options• Dynamicmulti-sliceviews

• Livemulti-sliceviews

• Dynamiccrop

• Measurementonrender

Optional 4D Feature• MVassessment(TomTec®)

Peripheral OptionsInternal peripherals

• USBB/Wvideoprinterwithcontrolfrom system (optional )

External peripherals

• Networkprinters

- Color laser printer

- Color video printer with control from system

• 16GBencryptedmemorystick

• Three-pedalconfigurablefootswitch

External outputs

• DVI-I

• Ethernet–10Mbps,100Mbps, 1 Gbps electrically isolated

• MultipleUSB2.0ports

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 4 of 18

• Multi-image(split/quadscreen)

- Live and/or frozen

- Independent cine playback

• Timelinedisplay

- Independent 2D (or compound) + PW/CW/Mdisplay

- A choice of display formats with varioussizesof2D+PW/CW/M

• Top/bottomselectableformat

• Side/sideselectableformat

• 4Ddisplay(optionalwith6VTprobeand 4D option)

- Two + one slice and render view

- Quad view (three slice and render)

- Single render view

- Slice-only view

- Live multi-slice

- FlexiSlice (live and replay)

- Bi-plane side/side view

- Tri-plane view (quad including geometry viewer)

- Crop view (three orthogonal slice + render)

- Apical slice view (three 60 degrees view + render)

- Cine rotate render view

- Bi-plane prepare (two slice + render)

Display Annotation• Patientname:First,lastandmiddle

• PatientID

• AdditionalpatientID

• Age,sexandbirthdate

• Hospitalname

• Dateformat:Twotypesselectable– MM/DD/YY,DD/MM/YY

• Timeformat:Twotypesselectable– 24 hours, 12 hours

• GestationalagefromLMP/EDD/GA

• Probename

• Mapnames

• Probeorientation

• Depthscalemarker

• Focalzonemarkers

• Imagedepth

• Zoomdepth

• Activemodedisplay

• Stressprotocolparameters

• Parameterannotationfollow ASE standard

• Freetextwithwordlibrary

• 4Dsliceintersectionmarkers (with 6VT probe and 4D option)

• 4Dgauge(with6VTprobeoption)

• 4Dviewinganglearrows (with 6VT probe and 4D option)

• 4Dgeometryviewer (with 6VT probe and 4D option)

• 4Dnumberofcycles (with 6VT probe and 4D option)

• Scanplanepositionindicatorandprobe temperature are displayed with all TEE probes

• Imageorientationmarker

General System Parameters

System Setup• Pre-programmableM&A and

annotation categories

• Differentuserpresetsper probe/application may be stored for quick access

• Userprogrammablepresetcapabilitywith administrator preset protection

• Factorydefaultpresetdata,protected against modification

• UserInterfacelanguages:English,LA Spanish, French, German, Italian, Portuguese (European and Brazilian), Russian, Swedish, Norwegian, Danish, Dutch, Finnish

• User-definedannotations

• Bodypatterns

• Customizedcommenthomeposition

Comprehensive User Manual Available on Board Available through touch-panel utility page. User manual and service manual are included on a USB memory device with each system. A printed user manual is provided.

• B-mode

- Gain

- Imaging frequency

- Frame averaging

• M-mode

- Gain

- Frequency

- Time scale

• Dopplermode

- Gain

- Angle

- Sample volume size and position

- Wallfilter

- Velocity and/or frequency scale

- Spectrum inversion

• Timescale

- PRF

- Doppler frequency

• ColorflowDopplermode

- Frame rate

- Sample volume size

- Color scale

- Power

- Color baseline

- Color threshold marker

- Color gain

• Spectruminversion

• Acousticframerate

• CINEgauge,imagenumber/framenumber

• Bodymarks:Multiplehuman anatomical structures

• Application/presetname

• Measurementresults

• Operatormessage

• Displayedacousticoutput

- TIS: Thermal Index Soft Tissue

- TIC: Thermal Index Cranial (Bone)

- TIB: Thermal Index Bone

• MI:Mechanicalindex

• PoweroutputindB

• Biopsyguidelineandzone

• Heartrate

• Trackball-drivenannotationarrows

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 5 of 18

• UserManualLanguages:English,French, German, Spanish, Italian, Portuguese (European and Brazilian), Swedish, Danish, Dutch, Norwegian, Japanese, Chinese, Polish, Finnish, Greek, Russian, Hungarian, Slovak, Romanian, Czech, Latvian, Lithuanian, Turkish, Estonian, Korean, Serbian, Bulgarian, Croatian, Indonesian, Kazakh, Ukraine

CINE Memory/Image Memory• 2GBofcinememory

• Selectablecinesequencefor cine review

• Measurements/calculationsand annotations on cine playback

• Scrollingtimelinememory

• Dual-imagecinedisplay

• Quad-imagecinedisplay

• CINEgaugeandcineimage number display

• CINEreviewloop

• CINEreviewspeed

Image Storage• 4Dvirtualstore(with6VTprobe

and 4D option) for efficient 4D image management

• On-boarddatabaseofpatient information from past exams

• User-selectableECGandtime gated acquisition available on touch panel during live

• User-selectableprospectiveor retrospective capture in config

• Storageformats:

- DICOM®-compressed or uncom-pressed, single/multi-frame, with/without raw data, storage via clipboard and/or seamlessly directly to destination device

- Transfer/“SaveAs”JPEG,MPEG, AVI and VolDicom (with 6VT probe and 4D option) formats

• Dataformatfullycompatiblewithoffline EchoPAC review/reporting stations of same or newer vintage

• Instantaccesstoultrasoundrawdata provided by the system

• Advancedpost-processinganalysis

• Threeuserlevelshelporganizingdata security requirements

• E-signoffcompatibility,withclearindications in patient management screens and report screen that a report was signed off, and by whom and at what time. The signed off report and exam cannot be changed. The “Diagnosing Physician” field is automatically assigned to the user that did the sign-off

Image and Data Management• Exceptionalworkflowwithinstant

access data management

• DICOM3.0support–seeDICOM conformance statement for details

• Supportfortransferoftheproprietary rawdatafileswithintheDICOMstandard

• 2D,CFMorTVIdataatmaximumframe rate may be reviewed by scrolling or by running cine loops (can contain more than 1000 images for imaging modes)

• Imageclipboardforstamp-size storage and review of stored images and loops

• Built-inpatientarchivewith images/loops, patient information, measurements and reports

• DICOM-SRStandardstructured reporting mechanism (option)

• Structuredfindingsreporttoolssupport efficient text entries with direct editing of findings text, usability improvements, new configuration options and conclusion section

• Usercanenternormalvalues which are then compared to actual measurements

• Storagedevices:

- USB memory stick: 16 GB

- CD-RWstorage:700MB (DVD option required)

- DVD storage: -R (4.7 GB) (DVD option required)

- Hard drive image storage: 0.5 TB

• Compareoldimageswith current exam

• Reloadofarchiveddatasets

• ActivationcontrolofUSBdevices (for security)

Connectivity and DICOM • Ethernetnetworkconnection

• DICOM3.0

• Verify(optional)

• Print(optional)

• Store(optional)

• Modalityworklist(optional)

• Storagecommitment(optional)

• ModalityPerformedProcedureStep(MPPS)(optional)

• Mediaexchange

• DICOMspooler(optional)

• DICOMquery/retrieve(optional)

• Structuredreporting–compatiblewith adult cardiac and vascular (optional)

• Mediastoreofstructuredreporting

• InSite™ ExC capability for remote service/access

• Supportoftwopatients’IDsinDICOM(optional)

• SeparateDICOMSRandimage storage destinations (optional)

• SimultaneoustransferofDICOM to multiple destinations (optional)

Patient Archive

EchoPAC™/Patient Archive• IntegratedEchoPACfunctionality

adds connectivity and image analysis capability to scanner

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 6 of 18

• ConfigurableHTML-based report function

• Reporttemplatescanbecustomizedon board

• ASE-baseddefaulttextmodules (English), user-customizable

• Internalarchivedatacanbe exported to removable image storagethroughDICOMmedia

• Internalharddisk–forstoring programs, application defaults, ultrasound images and patient archive

• Alldatastorageisbasedon ultrasound raw data, allowing to change gain, baseline, color maps, sweep speeds, etc., for recalled images and loops

• DICOMmedia–read/writeimages onDICOMformat

• DICOMviewerembeddedonmedia(optional and selectable in Config)

• AlphanumericdatacanbeexportedinXMLformat

• JPEGexport(“SaveAs”)forstillframes

• AVIandMPEGexport(“SaveAs”) for cineloops

• Specializedfileformat“SaveAs” VolDICOMfeaturetoallowdata import into TomTec Research Arena free-standing workstation (with 6VT probe and 4D option)

eVue/MPEGvue (optional)

• Allowsinteractiveviewingof images, loops or full exams from remote location

• UsingMPEGvue,examsmaybestored onto removable media or on remote networked system together withintegratedMPEGvueplayerforviewing on standard PC

• Smartemailfeatureallowstransparent transmission of images via email using resident Outlook™ email client

• Widthrange:10–120degrees

• Continuousdynamicreceivefocus/continuous dynamic receive aperture

• Adjustabledynamicrange,infiniteupper level

• Imagereverse:Right/left

• Imagerotationof0,°180°

Tissue ImagingGeneral

• Variabletransmitfrequenciesforresolution/penetration optimization

• Displayzoomwithzoomareacontrol

• High-Resolution(HR)zoom– concentrates all image acquisition power into selected Region of Interest (ROI)

• Variablecontourfiltering–foredgeenhancement

• Depthrangeupto36cm– probe specific

• Selectablegrayscaleparameters:Gain, reject, DDP, clarity, dynamic range and compress – can be adjusted in live, digital replay and image clipboard recall (probe dependent)

• AutomaticallycalculatedTGCcurvesreduces operator interaction

• Automaticallycalculatedlateralgain

• Smartdepth:automaticallyoptimizing transmit pattern parameters according to scan-depth setting

2D Mode

• Sectortiltandwidthcontrol

• Framerateinexcessof1500fps,depending on probe, settings and applications

• Codedoctaveimagingwithcodedphase inversion – 3rd generation harmonic tissue imaging providing improved lateral and contrast resolu-tion over conventional fundamental imaging. Features help reduce noise, improve wall definition, and axial resolution, making it well suited for a wide variety of patient groups

Self-contained DICOM Viewer (optional)

• ExamscanbetransferredtoCD/DVDor USB media with an integrated “EZDICOMCDviewer™ ”

• Self-contained“EZDICOMCDviewer™ ” allows to review exams from media on a standard PC, without installing anything on the host PC

Insite™ Express Connection (ExC) Enables Remote Service and Training

• Easy,flexibleandsecureconnectivity configuration. The “Contact GE” on-screen button directly generates a real-time service request to the GE online engineering or application specialist. It takes a snapshot (e.g., error logs, setup files) of the system at the time of the service request to enable analysis of problem before customer contact

• VirtualConsoleObservation(VCO) enables the customer to allow desktop screens to be viewed and controlled remotely over the encrypted tunnel to enable real-time training, device configuration and clinical application support

• OperationofInsiteExpressConnection is dependent on the infrastructure being available – check with your local GE service representative

• Filetransferenablesthecustomer(biomed or clinician) to directly transfer system information (e.g., system logs, images, parametric data) to GE product engineering teams (no patient data transferred)

• Softwarereloadprovidesremote application reconstruction and recovery capabilities in the event of system corruption

Scanning Parameters• Infinitenumberofeffectivechannels

• Minimumfield-of-viewrange(depth):0 – 2 cm (zoom) (probe dependent)

• Maximumfield-of-viewrange(depth):0 – 36 cm (probe dependent)

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 7 of 18

• Confocalimaging–allowsfor multiple transmit focal zones over range of view and a high vector density, probes dependent

• Automatictissueoptimization– single keystroke optimizes immedi-ately automatically and dynamically different grayscale settings with the goal of signal independent uniform gain and contrast distribution

• UDclarityandUDspecklereduceimaging – an advanced image processing technique to remove speckle in real-time examining the relative difference between neigh-boring pixel values and determining whether the grayscale variations have a sharp difference, follow a trend, or are random in nature

• HDimaging–real-timesimultaneous acquisition at dual frequencies compounded to help reduce speckle and noise while enhancing resolution and contrast

• Multiple-anglecompoundimaging– multiple co-planar images from dif-ferent angles combined into a single image in real-time to help enhance border definition and contrast resolution, as well as reduce angular dependence of border or edge as compared to no-compound imaging

• Elevationcompounding (with 6VT probe)

• LOGIQView:Providestheabilitytoconstruct and view a static 2D image with wider field-of-view of a given transducer. This allows viewing and measurements of anatomy that is larger than what would fit in a single image

• Virtualapexprovidesawider field-of-view with phased array probes, effective at certain imaging views where a wide near field may be preferred

• L/Randup/downinvert,inlive,digitalreplay or image clipboard recall

• Dynamicmulti-sliceenables positioning of the multi-slice, short-axis cut-planes at same anatomical position throughout the heart cycle (advanced 4D option)

• Livemulti-slicelayoutsavailable during live 4D acquisition (advanced 4D option)

• FlexiSliceforinteractiveslicing, cropping and navigation designed to provide the user with a flexible, yet intuitive way of extracting 2D slices from 4D data sets

• Viewcropsettingfortogglecontrol of view plane vs. crop plane

• 2Clickcropforquickandeasy extraction of standard and non- standard views for visualization of 4D structures seen during or after the examination

• Stereovisionin4D(option)

• Laserlinestohelpimprovethevisuallinkage between the 4D-rendered view and the 2D slices

• Widerangeofdepthcolor rendering maps

• QuickRotateandRotateforaflexibleand easily accessible way of obtaining the desired single- or multi-plane, two-dimensional views

Multi-dimensional Mode (with 6VT-D probe and 4D option)

• Bi-planescanning:Twoindependentsimultaneous scan planes where one of them can be rotated and tilted freely

• Bi-planepreparemodeforeaseofobtaining bi-plane views from 4D render data sets

• Tri-plane:Threeindependent simultaneous scan planes that can be rotated freely

• Bothbi-planeandtri-planescanningis possible in all color Doppler modes

• Digitalreplayforretrospective review or automatic looping of images, allowing for adjustment of parameters such as gain, reject, anatomicalM-mode,persistence and replay speed

• Datadependentprocessingperformstemporal processing which helps reduce random noise but leaves motion of significant tissue structures largely unaffected – can be adjusted even in digital replay

• 256shadesofgray

• Colorized2D-mode,user-selectablein real-time, digital replay

4D Mode (option)

• Flexi-volumeswithcustomizable acquisition for volume size, volume rate or resolution

• Single-beat4Dscanningwith real-time volume rendering display

• Multi-beat4Dscanningfor high-resolution scanning

• Adjustablevolumesizesforbothsingle- and multi-beat scanning

• Adjustablevolumeshapecontrol

• Pre-definedvolumesizesforquickvolume setup

• Adjustablenumberofcyclesformulti-beat scanning

• FlexiZoomforeasy4Dvisualization of structures of interest

• 4Dscanningsupportingvariableoctave and fundamental frequencies

• Coherentvolumeprocessingwithmotion compensation for seamless and artifact-free 4D and 2D slices

• Variableframeratesettingsavailable

• Volumeoptimizecontrolfor volume rendering transparency and quality setting

• Flipcropavailableforchanging 4D view direction 180 degrees with mirrored crop volume

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 8 of 18

M-mode

• TrackballsteersM-modeline available with all imaging probes – max steering angle is probe dependent

• Simultaneousreal-time 2D-andM-mode

• M-modePRF1kHz–imagedata acquired is combined to give high-quality recording regardless of display scroll speed

• Digitalreplayforretrospectivereviewof spectral data

• Severaltop-bottomformats,side-by-side format and time-motion-only format – can be adjusted in live or digital replay

• Selectablehorizontalscrollspeed: 1, 2, 3, 4, 6, 8, 12, 16 seconds across display

• Horizontalscrollcanbeadjusted in live or digital replay

Anatomical M-mode

• M-modecursorcanbeadjusted at any plane

• CurvedanatomicalM-mode– free(curved)drawingofM-modegenerated from the cursor independent from the axial plane

• Canbeactivatedfromlive,digitalreplay or image clipboard recall

• Anatomicalcolorandtissue velocityM-mode

• M&A capability

Color Doppler ImagingGeneral

• SteerablecolorDoppleravailablewith all imaging probes – max steering angle is probe dependent

• Trackball-controlledROI

• Removalofcolormapfromthe tissue during digital replay

• Digitalreplayforretrospective reviewofcolororcolorM-mode data allowing for adjustment of parameters such as encoding principle, color priority and color gain even on stored data

• Digital replay for retrospective review or automatic looping of color images, allowing for adjustment of parameters such as DDP, encoding principle, baseline shift, color maps, color priority and color gain even on frozen/recalled data

• Application-dependent,multi-variatemotion discriminator helps reduce flash artifacts

• Dedicatedcoronaryflowapplication

• Multiple-anglecompoundimagingin2D mode is maintained while in color Doppler mode

4D Color Doppler Imaging (with 6VT-D probe and 4D option)• Single-beat4Dcolorflowscanning

• Volumesizecontroltochangethesize of the color ROI

• Multi-beat4Dcolorflowscanningusing ECG stitching for increased volume rate

• Pre-definedvolumesizesforquickvolume setup

• Adjustablenumberofcyclesfor multi beat scanning

• Variablevolumeratesettings available

• Flipcropavailableforchanging 4D view direction 180 degrees with mirrored crop volume

• View-cropsettingfortogglecontrolof view plane vs. crop plane

• Stereovisionin4Dcolor

• Tissuetransparencycontrol

• Flowtransparencycontrol

Multi-Dimensional Color Doppler Imaging (with 6VT-D probe and 4D option)• Bi-planeandtri-planescanning

with all color Doppler and tissue velocity modes

Color Angio• Angle-independentmodefor

visualization of small vessels with increased sensitivity compared to standard color flow of previous GE products

• PRFsettings–user-selectable

• Advancedregressionwallfiltergivesefficient suppression of wall clutter

• Foreachencodingprinciple,multiplecolor maps can be selected in live and digital replay – variance maps available

• Morethan65,000simultaneouscolors processed, providing a smooth display two-dimensional color maps containing a multitude of color hues

• Simultaneousdisplayofgrayscale 2D and 2D with color flow

• Colorinvert–user-selectableinliveand digital replay

• Variablecolorbaseline– user-selectable in live and digital replay

• Multi-variatecolorpriorityfunctiongives delineation of disturbed flows even across bright areas of the 2D-mode image

• ColorDopplerfrequencycanbechanged independently from 2D

Color Flow Imaging• ThecSoundplatformwithitsparallel

beamformer architecture allows a combination of ultra-high frame rate and increased lateral resolution compared to previous generation GE scanners

• Veryhighdigitalsignalprocessingpower, maintaining high frame rates with large ROI's even for very low PRF settings

• Framerateinexcessof700fps, depending on probe and settings

• VariableROIsizeinwidthanddepth

• User-selectableradialandlateralaveraging to help reduce statistical uncertainty in the color velocity and variance estimates

• DataDependentProcessing(DDP)performs temporal processing and display smoothing to help reduce loss of transient events of hemodynamic significance

Vivid S70 product datasheet – June2015–DOC1662324 Page9of18

Color M-mode• VariableROIlengthandposition–

user-selectable

• User-selectableradialaveragingtohelp reduce statistical uncertainty in the color velocity and variance estimates

• Selectablehorizontalscrollspeed: 1, 2, 3, 4, 6, 8, 12, 16 seconds across display – can be adjusted during live, digital replay or image clipboard recall

• Real-time2Dimagewhilein colorM-mode

• Samecontrolsandfunctionsavailable as in standard 2D color Doppler

Anatomical Color M-mode• GE-patented,anyplanecolor M-modedisplayderivedfrom color Doppler cine loop

• Alsoapplicabletotissue velocity Imaging

• M&A capability

B-flow• B-flowisadigitalimagingtechnique

that provides real-time visualization of vascular hemodynamics by directly visualizing blood reflectors and presenting this information in a grayscale display

• UseofGE-patentedtechniquestoboost blood echoes, and to help preferentially suppress non-moving tissue signals

• B-flowisavailableformostvascularand shared service applications

Blood Flow Imaging• CombinescolorDopplerwith

grayscale speckle imaging

• Helpsimprovedelineationofbloodflow without bleeding into tissue or vessel wall

Blood Flow Angio Imaging• Combinesangiowithgrayscale

speckle imaging

• EfficientsegmentspecificTSItimemeasurements

• Immediatebulls-eyereport

• AutomaticcalculatedTSI synchrony indexes

• TSIsurfacemapping

• LVsynchronizationreporttemplate

• CRTprogrammingprotocol

Strain/Strain Rate Mode (option, enabled by Advanced QScan)

• Tissuedeformation(strain)andrateof deformation (strain rate) are cal-culated and displayed as real-time, color-coded overlay on the 2D image

• Cinecompoundcalculatesanddisplays cineloops generated from a temporal averaging of multiple consecutive heart cycles

• AnatomicalM-modeandcurved anatomicalM-modedisplays (SI and SRI)

Spectral DopplerGeneral

• OperatesinPW,HPRFandCWmodes

• TrackballsteerableDoppleravailablewith all imaging probes – max steering angle is probe dependent

• SelectableDopplerfrequencyforenhanced optimization

• High-quality,real-timeduplexor triplex operation in all Doppler modes,CWandPW,andfor all velocity settings

• Frameratecontrolforoptimized use of acquisition power between spectrum, 2D and color Doppler modes in duplex or triplex modes

• Veryfastandflexiblespectrumanalysis with an equivalent DFT rate of 0.2 ms

• AutomaticSpectrumOptimization(ASO) provides a single press, automatic, real-time optimization ofPWorCWspectrumscale,andbaseline display

Tissue Velocity ImagingTissue Velocity Imaging Mode

• MyocardialDopplerimagingwithcolor overlay on tissue image

• TissueDopplerdatacanbeacquired in background during regular 2D imaging

• Thevelocityofmyocardialsegmentsafter entire heart cycle can be displayed in one single image

• Tissuecoloroverlaycanberemovedtoshow just the 2D image, still retaining the tissue velocity information

• QuantitativeprofilesforTVI,tissuetracking, strain and strain rate can be derived

• Timemarkersforvalveeventsderived fromanyTMmodehelpsimplifyunderstanding of signals in velocity tracesorcurvedanatomicalM-mode

Tissue Tracking Mode

• Real-timedisplayofthetimeintegral of TVI for quantitative display of myocardial systolic displacement

• Myocardialdisplacementiscalculated and displayed as a color-coded overlayonthegrayscaleandM-mode image – different colors represent different displacement ranges

Tissue Synchronization Imaging Mode (option, enabled by Advanced QScan)

• Parametricimagingwhichgivesinformation about synchronicity of myocardial motion

• Myocardialsegmentscolored according to time to peak velocity, green for early and red for late peak

• Waveformtraceavailableto obtain quantitative time to peak measurement from TSI Image

• Availableinlivescanning,aswellasan offline calculation derived from tissue Doppler data

• Additionalfeaturesincombinationwith multi-dimensional imaging option

• Simultaneousacquisitionoftri-planeTSI images covering all standard segments in apical views

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 10 of 18

• Dynamicgaincompensationfordisplay of flows with varying signal strengths over the cardiac cycle to help improve ease of use

• Dynamicrejectgivesconsistent suppression of background – user-selectable in real-time, digital replay or image clipboard recall

• Digitalreplayforretrospective review of spectral Doppler data

• Severaltop-bottomformats, side-by-side format and time- motion-only format – can be adjusted in live or digital replay

• Selectablehorizontalscrollspeed:1, 2, 3, 4, 6, 8, 12, 16 seconds across display – can be adjusted in live or digital replay

• AdjustablespectralDopplerdisplayparameters: Gain, reject, compress, color maps – can be adjusted in live or digital replay

• User-adjustablebaselineshift– in live, digital replay and image clipboard recall

• Adjustablevelocityscale

• Wallfilterswithrange10-2000Hz(velocity scale dependent)

• Anglecorrectionwithautomaticadjustment of velocity scale – in live, digital replay and image clipboard recall

• AutoDopplerangle

• Stereospeakersmountedinthe front panel

• Displayannotationsoffrequency,mode, scales, Nyquist limit, wall filter setting, angle correction, acoustic power indices

• Compoundinduplex

PW/HPRF Doppler

• AutomaticHPRFDopplermaintainsits sensitivity even for shallow depths and with the highest PRF's

• DigitalvelocitytrackingDoppler employs processing in range and time for high-quality spectral displays

• Adjustablesamplevolumesizeof1-16 mm (probe dependent)

• Maximumsamplevolumedepth30cm

• Configurationpackagetosetup a customized set and sequence of measurements to use, defining user-defined measurements and changing settings for the factory-defined measurements

• Stressechosupportallowingwallmotion scoring and automatic stress level labeling of measurements

• Supportformeasuringon DICOMimages

• AutomaticDopplertracefunctionality for use in non-cardiac applications in both live and replay

• Worksheetforreview,editand deletion of performed measurements

• Reportingsupportallowinga configurable set of measurements to be shown in the exam report

• DICOMSRexportofmeasurementdata

Intima Media Thickness (IMT) Measurements (optional )

• Automaticmeasurements (patent pending) of carotid artery Intima-MediaThickness(IMT)on any acquired frame

• On-boardIMTpackagefacilitatesnon-interrupted workflow – fully integratedwithM&A, worksheet, archiving and reporting functions

• Algorithmprovidesrobust,quick, reliable measurements which can be stored to the on-board archive for review and reporting

• IMTmeasurementcanbemadefromfrozen images or images retrieved from archive

• IMTpackagesupportsmeasure-ments of different regions of the intima in the carotid vessel (e.g., Lt./Rt./CCA/ICA etc.)

• FrameforIMTmeasurementcan be selected in relation to the ECG waveform

Z-Scores

• Limitedimplementationofz-scores for a set of predefined pediatric dimension measurements

CW Doppler

• HighlysensitivesteerableCWavailable with all phased array probes

• TissuevelocityDoppler

Physiological Traces• Integratedthree-leadECGmodule

• AutomaticQRScomplexdetection

• ExternalECGleadinput

• Internallygeneratedrespiratory trace using ECG leads

• ECGleadselection

• AdjustableECGQRSmarkers

Automatic Optimization• DynamicoptimizationofB-mode

image to improve contrast resolution, TGC and grayscale (soft or sharp, user-selectable)

• Auto-spectraloptimize–dynamicadjustments of baseline, and PRF (on live image) and angle correction

Measurement and Analysis (M&A)• Personalizedmeasurement

protocols allow individual set andorderofM&A items

• Measurementscanbelabeled seamlessly by using protocols or post assignments

• Measurementsassignableto protocol capability

• Parameterannotationfollow ASE standard

• Seamlessdatastorageand report creation

• User-assignableparameters

• Comprehensivesetofcardiac measurements and calculations to help assess dimensions, flow properties and other functional parameters of the heart

• Comprehensivesetofsharedservicemeasurements and calculations covering vascular, abdominal, obstetrics and other application areas

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 11 of 18

4D Auto LVQ (with 6VT probe and 4D option)

• AutomatedmeasurementofLV volume and EF from volumetric data

• Automatedidentificationof standard views

• Validationofdetectedboundaries

• LVvolumewaveformforentire cardiac cycle

• EDandESautomaticallyselectedfrom volume waveform (max/min)

• Editingbypointandclick

• Userapprovaloffinalresults

• FullyintegratedintoM&A system with results in worksheet

Mitral Valve Assessment (with 6VT probe and 4D option) (optional )

• Thesemi-automatedMVassessmenttool from TomTec provides the ability to include quantitative results from the mitral valve apparatus into the patient exam

Quantitative Analysis Package (Q-Analysis) (optional )

• Tracesforvelocityorderived parameters (strain rate, strain, displacement) inside defined regions of interest as function of time

• Contrastanalysiswithtracesforgrayscale intensity or angio power inside defined regions of interest as function of time

• CurvedanatomicalM-modedisplayallowinganM-modealongan arbitrary curve in a 2D image

• Sample-areapointsmaybedynami-cally anchored to move with the tissue when running the cineloop

• Cinecompounddisplayscineloopsgenerated from a temporal averaging of multiple consecutive heart cycles

Automated Function Imaging (AFI) (optional )

• Parametricimagingtoolwhich gives quantitative data for global and segmental wall motion

• Allowscomprehensiveassessment at a glance by combining three longitudinal views into one comprehensive bulls-eye view

• User-selectablefetalgrowth parameters based on European, American or Asian methods charts

• GYNpackageforovaryanduterusmeasurements and reporting

OB Measurements/Calculations

• Gestationalageby:

- GS (Gestational Sac)

- CRL (Crown Rump Length)

- FL (Femur Length)

- BPD (Biparietal Diameter)

- AC (Abdominal Circumference)

- HC (Head Circumference)

- APTD x TTD (Anterior/Posterior Trunk Diameter by Transverse Trunk Diameter)

- LV (Length of Vertebra)

- FTA (Fetal Trunk Cross-sectional Area)

- HL (Humerus Length)

- BD (Binocular Distance)

- FT (Foot Length)

- OFD (Occipital Frontal Diameter)

- TAD (Transverse Abdominal Diameter)

- TCD (Transverse Cerebellum Diameter)

- THD (Thorax Transverse Diameter)

- TIB (Tibia Length)

- ULNA (Ulna Length)

• EstimatedFetalWeight(EFW)by:

- AC, BPD

- AC, BPD, FL

- AC, BPD, FL, HC

- AC, FL

- AC, FL, HC

- AC, HC

- EFBW

• CalculationsandRatios

- FL/BPD

- FL/AC

- FL/HC

- HC/AC

- CI (Cephalic Index)

- AFI (Amniotic Fluid Index)

- CTAR (Cardio-Thoracic Area Ratio)

• IntegratedintoM&A package with specialized report templates

• 2Dstrainbaseddatamovesintoclinical practice

• Simplifiedworkflowwithfully automated ROI tracing (if configured), quick tips and combined display of traces from all segments

Automated Ejection-Fraction Calculation (AutoEF) (optional )

• AutomatedEFmeasurementtoolbased on 2D-speckle tracking algorithm and on Simpson

• IntegratedintoM&A package with worksheet summary

Generic Measurements

• BSA(BodySurfaceArea)

• MaxPG(MaximumPressureGradient)

• MeanPG(MeanPressureGradient)

• %Stenosis(StenosisRatio)

• PI(PulsatilityIndex)

• RI(ResistivityIndex)

• HR(HeartRate)–beats/minute

• A/BRatio(VelocitiesRatio)

• TAMAX(TimeAveragedMaximumVelocity) – Trace method is Peak orManual

• TAMIN(TimeAveragedMinimumVelocity) – Trace method is Floor

• TAMEAN(TimeAveragedMean Velocity)–TracemethodisMean

• Volume

OB/GYN Application Module (option)

• OBpackageforfetalgrowth analysis containing more than 100 biometry tables

• DedicatedOB/GYNreports

• Fetalgraphicalgrowthcharts

• Growthpercentiles

• Multi-gestationalcalculations (up to four)

• ProgrammableOBtables

• Expandedworksheets

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 12 of 18

• Measurements/calculationsby:ASUM,ASUM2001,Berkowitz,Bertagnoli, Brenner, Campbell, CFEF, Chitty, Eik-Nes, Ericksen, Goldstein, Hadlock, Hansmann, Hellman, Hill, Hohler,Jeanty,JSUM,Kurtz,Mayden,Mercer,Merz,Moore,Nelson,OsakaUniversity, Paris, Rempen, Robinson, Shepard,Shepard/Warsoff,TokyoUniversity,Tokyo/Shinozuka,Yarkoni

• Fetalgraphicaltrending

• Growthpercentiles

• Multi-gestationalcalculations(four)

• Fetalqualitativedescription (anatomical survey)

• Fetalenvironmentaldescription (biophysical profile)

• ProgrammableOBtables

• Over20selectableOBcalculations

• Expandedworksheets

GYN Measurements/Calculations

• Rightovarylength,width,height

• Leftovarylength,width,height

• Uteruslength,width,height

• Cervixlength,trace

• Ovarianvolume

• ENDO(endometrialthickness)

• OvarianRI

• UterineRI

• Follicularmeasurements

• Summaryreports

Vascular Calculations• RTECA(RightExternalCarotid

Artery Velocity)

• RTCCA(RightCommonCarotid Artery Velocity)

• RTBIFURC(RightCarotid Bifurcation Velocity)

• RTICA(RightInternalCarotid Artery Velocity)

• RTICA/CCA(RightInternalCarotid Artery Velocity/Common Carotid Artery Velocity Ratio)

• LTECA,LTCCA,LTBIFURC,LTICA, LT ICA/CCA (same as above, for Left Carotid Artery)

• AVEOAIVmax(AorticValve Effective Orifice Area Index by Continuity Equation Peak V)

• AVCO(CardiacOutputbyAorticFlow)

• AVCusp (Aortic Valve Cusp Separation, 2D)

• AVDecTime (Aortic Valve Deceleration Time)

• AVDiam(AorticDiameter,2D)

• AVmaxPG (Aortic Valve Peak Pressure Gradient)

• AVmeanPG (AorticValveMeanPressureGradient)

• AVSV(StrokeVolumebyAorticFlow)

• AVVmax(AorticValvePeakVelocity)

• AVVmean(AVMeanVelocity)

• AVVTI (Aortic Valve Velocity Time Integral)

• AVA(Vmax)(AVAreabyContinuityEquation by Peak V)

• AVA(VTI)(AVAreabyContinuity Equation VTI)

• AVAPlanimetry(AorticValveArea)

• AVET(AorticValveEjectionTime)

• CO(Teich) (CardiacOutput,M-mode,Teicholtz)

• D-EExcursion (MVAnteriorLeafletExcursion)

• E’Avg(Averagedearlydiastolic mitral valve annular velocity)

• E’Lat(Earlydiastolicmitralvalvelateral annular velocity)

• E’Sept(Earlydiastolicmitralvalveseptal annular velocity)

• E/E’Avg (MitralinflowEvelocitytoE’Avgratio)

• E/E’Lat (MitralinflowEvelocitytoE’Latratio)

• E/E’Sept (MitralinflowEvelocitytoE’Septratio)

• EDV(Cube)(LeftVentricleVolume,Diastolic, 2D, Cubic)

• EF(A-LA2C)(EjectionFraction2CH,Single Plane, Area-Length)

• E-FSlope(MitralValveE-FSlope)

• EPSS(E-Point-to-SeptumSeparation,M-mode)

• A/BRatio(VelocitiesRatio)

• %Stenosis(StenosisRatio)

• S/DRatio(SystolicVelocity/DiastolicVelocities Ratio)

• PI(PulsatilityIndex)

• RI(ResistivityIndex)

• HR(HeartRate)–beats/minute

Cardiac Measurements• %FS(LVFractionalShortening)

• %IVSThck(IVSFractionalShortening)

• %LVPWThck(LVPosterior WallFractionalShortening)

• AoArchDiam(AorticArchDiameter)

• Aoasc(AscendingAorticDiameter)

• AoDescDiam (Descending Aortic Diameter)

• AoIsthmus(AorticIsthmus)

• AoRootDiam(AorticRootDiameter)

• ARERO (PISA: Regurgitant Orifice Area)

• ARFlow(PISA:RegurgitantFlow)

• ARPHT(AVInsuf.PressureHalfTime)

• ARRad(PISA:RadiusofAliasedPoint)

• ARRF(RegurgitantFraction over the Aortic Valve)

• ARRV (PISA: Regurgitant Volume Flow)

• ARVel(PISA:AliasedVelocity)

• ARVmax(AorticInsuf.PeakVelocity)

• ARVTI (Aortic Insuf. Velocity Time Integral)

• ARedmaxPG(AorticInsuf. End-Diastole Pressure Gradient)

• ARedVmax(AorticInsuf. End-Diastolic Velocity)

• AVAccSlope (Aortic Valve Flow Acceleration)

• AVAccTime (Aortic Valve Acceleration Time)

• AVAccT/ET(AVAcceleration to Ejection Time Ratio)

• AVEOAI(VTI)(AorticValve Effective Orifice Area Index by Continuity Equation VTI)

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 13 of 18

• ERO(EffectiveRegurgitantOrifice)

• ESV(Cube)(LeftVentricleVolume,Systolic, 2D, Cubic)

• HR(HeartRate,2D,Teicholtz)

• IVC(InferiorVenaCava)

• IVCT(IsovolumicContractionTime)

• IVRT(IsovolumicRelaxationTime)

• IVSd(InterventricularSeptum Thickness, Diastolic, 2D)

• VSs(InterventricularSeptum Thickness, Systolic, 2D)

• LADiam(LeftAtriumDiameter,2D)

• LAMajor(LeftAtriumMajor)

• LAMinor(LeftAtriumMinor)

• LA/Ao(LADiameterto AoRoot Diameter Ratio, 2D)

• LAAd(A2C)(LeftAtriumArea,Apical2C)

• LAEDV(A-L) (LA End Diastolic Volume, Area-Length)

• LAEDVIndex(A-L)(LAEndDiastolicVolume Index, Area-Length)

• LAESV(A-L)(LAEndSystolicVolume,Area-Length)

• LAESVIndex(A-L)(LAEndSystolicVolume Index, Area-Length)

• LAEDVMOD (LAEndDiastolicVolumeMOD)

• LAESVMOD (LAEndSystolicVolumeMOD)

• LIMP(LeftIndexof MyocardialPerformance)

• LVA(s)(LeftVentricularArea, Systolic, 2CH)

• LVAd(A2C)(LeftVentricularArea,Diastolic, 2CH)

• LVAd(sax)(LVarea,SAX,Diastolic)

• LVAend(d)(LVEndocardialArea,SAX)

• LVAepi(d)(LVEpicardialArea,SAX)

• LVAs(A4C) (Left Ventricular Area, Systolic, 4CH)

• LVAs(sax)(LVarea,SAX,Systolic)

• LVdMass(LVMass,Diastolic,2D)

• LVdMass(LVMass,Diastolic,M-mode)

• LVdMassIndex (LVMassIndex,Diastolic,2D)

• MRmaxPG (MitralRegurg.PeakPressureGradient)

• MRRad(PISA:RadiusofAliasedPoint)

• MRRF(Regurgitantfraction overtheMitralValve)

• MRRV(PISA:RegurgitantVolumeFlow)

• MRVel(PISA:AliasedVelocity)

• MRVmax(MitralRegurg.PeakVelocity)

• MRVmean (MitralRegurg.MeanVelocity)

• MRVTI (MitralRegurg.VelocityTimeIntegral)

• MVADur (MitralValveA-WaveDuration)

• MVAVelocity(MVVelocityPeakA)

• MVAccSlope (MitralValveFlowAcceleration)

• MVAccTime (MitralValveAccelerationTime)

• MVAcc/DecTime (MV:Acc.Time/Decel.TimeRatio)

• MVandiam (MitralValveAnnulusDiameter,2D)

• MVCO(CardiacOutputbyMitralFlow)

• MVDecSlope (MitralValveFlowDeceleration)

• MVDecTime (MitralValveDecelerationTime)

• MVEVelocity(MVVelocityPeakE)

• MVE/ARatio (MitralValveE-PeaktoA-PeakRatio)

• MVmaxPG (MitralValvePeakPressureGradient)

• MVmeanPG (MitralValveMeanPressureGradient)

• MVPHT (MitralValvePressureHalfTime)

• MVRegFrac (MitralValveRegurgitantFraction)

• MVSI (StrokeVolumeIndexbyMitralFlow)

• MVSV(StrokeVolumebyMitralFlow)

• MVTimetoPeak (MitralValveTimetoPeak)

• MVVmax(MitralValvePeakVelocity)

• LVEDV(A-LA2C)(LVVolume,Diastolic,2CH, Area-Length)

• LVESV(A-LA2C)(LVVolume,Systolic,2CH, Area-Length)

• LVET(LeftVentricleEjectionTime)

• LVIDd (LV Internal Dimension, Diastolic, 2D)

• LVIDs (LV Internal Dimension, Systolic, 2D)

• LVLd(apical) (Left Ventricular Length, Diastolic, 2D)

• LVLs(apical) (Left Ventricular Length, Systolic, 2D)

• LVOTArea (Left Ventricle Outflow Tract Area)

• LVOTCO (Cardiac Output by Aortic Flow)

• LVOTDiam(LeftVentricularOutflowTract Diameter)

• LVOTmaxPG (LVOT Peak Pressure Gradient)

• LVOTmeanPG (LVOTMeanPressureGradient)

• LVOTSI (Stroke Volume Index by Aortic Flow)

• LVOTSV (Stroke Volume by Aortic Flow)

• LVOTVmax(LVOTPeakVelocity)

• LVOTVmean(LVOTMeanVelocity)

• LVOTVTI(LVOTVelocityTimeIntegral)

• LVPWd(LeftVentricularPosteriorWallThickness, Diastolic, 2D)

• LVPWs(LeftVentricularPosteriorWallThickness, Systolic, 2D)

• LVsMass(LVMass,Systolic,2D)

• LVsMassIndex (LVMassIndex,Systolic,2D)

• LAAd(A2C)(LeftAtriumArea,Apical2C)

• MCO(MitralValveclosuretoOpening)

• MPArea(MitralValveProsthesis)

• MRAccTime (MVRegurg.FlowAcceleration)

• MRERO (PISA: Regurgitant Orifice Area)

• MRFlow(PISA:RegurgitantFlow)

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 14 of 18

• MVVmean(MVMeanVelocity)

• MVVTI (MitralValveVelocityTimeIntegral)

• MVA(MitralValveArea)

• MVAByPHT (MitralValveAreaaccordingtoPHT)

• MVAbyplan(MitralValveArea,2D)

• MVET(MitralValveEjectionTime)

• PVeinA(PulmonaryVeinVelocityPeak A) – reverse

• PVeinADur (PulmonaryVeinA-WaveDuration)

• PVeinD(PulmonaryVein End-Diastolic Peak Velocity)

• PVeinS(PulmonaryVein Systolic Peak Velocity)

• PAEDP (Pulmonary Artery Diastolic Pressure)

• PE(d)(PericardEffusion,M-mode)

• PEs(PericardEffusion,2D)

• PRmaxPG(PulmonicInsuf. Peak Pressure Gradient)

• PRmeanPG(PulmonicInsuf. MeanPressureGradient)

• PRPHT (Pulmonic Insuf. Pressure Half Time)

• PRVmax (Pulmonic Insuf. Peak Velocity)

• PRVTI(PulmonicInsuf. Velocity Time Integral)

• PRendmaxPG(PulmonicInsuf. End-Diastole Pressure Gradient)

• PRendVmax(PulmonicInsuf. End-Diastolic Velocity)

• PulmonicDiam (Pulmonary Artery Diameter, 2D)

• PVAccSlope (Pulmonic Valve Flow Acceleration)

• PVAccTime (Pulmonic Valve Acceleration Time)

• PVAccTime/ETRatio(PVAcceleration to Ejection Time Ratio)

• PVandiam (Pulmonic Valve Annulus Diameter, 2D)

• PVAnnArea(PulmonicValveArea)

• PVCO (Cardiac Output by Pulmonic Flow)

• RVET(RightVentricleEjectionTime)

• RVIDd(RightVentricleDiameter, Diastolic, 2D)

• RVIDs (Right Ventricle Diameter, Systolic, 2D)

• RVOTArea (Right Ventricle Outflow Tract Area)

• RVOTDiam (RV Output Tract Diameter, 2D)

• RVOTDiam(RVOutput TractDiameter,M-Mode)

• RVOTmaxPG (RVOT Peak Pressure Gradient)

• RVOTmeanPG (RVOTMeanPressureGradient)

• RVOTSI(LVStrokeVolumeIndex by Pulmonic Flow)

• RVOTSV(StrokeVolume by Pulmonic Flow)

• RVOTVmax(RVOTPeakVelocity)

• RVOTVmean(RVOTMeanVelocity)

• RVOTVTI(RVOTVelocityTimeIntegral)

• RVSP (Right Ventricle Systolic Pressure)

• RVWd(RightVentricleWallThickness,Diastolic,M-mode)

• RVWs(RightVentricleWallThickness,Systolic,M-mode)

• RAA(d) (Right Atrium Area, 2D, Diastole)

• RAA(s) (Right Atrium Area, 2D, Systole)

• SI(A-LA2C)(LVStrokeIndex, Single Plane, 2CH, Area-Length)

• SI(A-LA4C)(LVStrokeIndex, Single Plane, 4CH, Area-Length)

• SI(Bi-plane) (LVStrokeIndex,Bi-plane,MOD)

• SI(bullet) (LV Stroke Index, Bi-plane, Bullet)

• SI(MODA2C)(LVStrokeIndex, SinglePlane,2CH,MOD)

• SI(MODA4C)(LVStrokeIndex, SinglePlane,4CH,MOD)

• SI(Teich)(LVStrokeIndex, Teicholtz, 2D)

• PVmaxPG(PulmonicValve Peak Pressure Gradient)

• PVmeanPG(PulmonicValve MeanPressureGradient)

• PVSV (Stroke Volume by Pulmonic Flow)

• PVVmax (Pulmonary Artery Peak Velocity)

• PVVmean(PVMeanVelocity)

• PVVTI (Pulmonic Valve Velocity Time Integral)

• PVA(VTI)(PulmonaryArtery Velocity Time Integral)

• PVeinS/DRatio (Pulmonary Vein SD Ratio)

• PVET(PulmonicValveEjectionTime)

• PVPEP (Pulmonic Valve Pre-Ejection Period)

• PVPEP/ETRatio(PVPre-Ejection to Ejection Time Ratio)

• Qp/Qs (Pulmonic-to-Systemic Flow Ratio)

• RAMajor(RightAtriumMajor,2D)

• RAMinor(RightAtriumMinor,2D)

• RAA(d) (Right Atrium Area, 2D, Diastole)

• RAA(s) (Right Atrium Area, 2D, Systole)

• RAEDVA2C(RightAtriumEnd Diastolic Volume, Apical 2 Chamber)

• RAESVA-L (RA End Systole Volume [A-L])

• RALd(RightAtriumLength,Diastole)

• RALs(RALength,Systole)

• RIMP(RightIndexof MyocardialPerformance)

• RJA(A4C)(RegurgitantJetArea)

• RJA/LAA (Regurgitant Jet Area ratio RJA/LAA)

• RVMajor(RightVentricleMajor)

• RVMinor(RightVentricleMinor)

• RVS’(Tricuspidannulussystolic excursion velocity)

• RVAWd(RightVentricle WallThickness,Diastolic,2D)

• RVAWs(RightVentricle WallThickness,Systolic,2D)

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 15 of 18

• SI(Teich)(LVStrokeIndex, Teicholtz,M-mode)

• SV(A-LA2C)(LVStrokeVolume, Single Plane, 2CH, Area-Length)

• SV(A-LA4C)(LVStrokeVolume, Single Plane, 4CH, Area-Length)

• SV(Bi-plane)(LVStrokeVolume, Bi-plane,MOD)

• SV(bullet)(LVStrokeVolume, Bi-plane, Bullet)

• SV(MODA2C)(LVStrokeVolume,Single-plane,2CH,MOD)–Simpson

• SV(MODA4C)(LVStrokeVolume,Single-plane,4CH,MOD)–Simpson

• SV(Cube)(LVStrokeVolume,2D,Cubic)

• SV(Cube) (LVStrokeVolume,M-mode,Cubic)

• SV(Teich) (LV Stroke Volume, 2D, Teicholtz)

• SV(Teich) (LVStrokeVolume,M-mode,Teicholtz)

• SystemicDiam (Systemic Vein Diameter, 2D)

• SystemicVmax (Systemic Vein Peak Velocity)

• SystemicVTI (Systemic Vein Velocity Time Integral)

• TAPSE(TricuspidAnnularPlane Systolic Excursion)

• TCO (Tricuspid Valve Closure to Opening)

• TRmaxPG(TricuspidRegurg. Peak Pressure Gradient)

• TRmeanPG(TricuspidRegurg. MeanPressureGradient)

• TRVmax (Tricuspid Regurg. Peak Velocity)

• TRVmean (TricuspidRegurg.MeanVelocity)

• TRVTI(TricuspidRegurgitation Velocity Time Integral)

• TVAdur (TricuspidValveA-WaveDuration)

• TVAVelocity(TricuspidValveAVelocity)

• TVAccTime (Tricuspid Valve Time to Peak)

• TVAnnArea(TricuspidValveArea)

• Supportsselectionofallmodes,allmeasurements and dual annotations

• Imagingattributes:Octave,Steer,Dual/Quad screen, Compound, LogiqView,Zoom,Depth,Scale and Baseline

• On-lineoroff-lineprotocoleditor

• Imageacquisitionaccordingto predefined protocol templates

• Variousfactoryprotocoltemplates

• User-configurableprotocoltemplates

Stress Echo (optional)Supported Protocol Examinations

• 2Dpharmacologicalstressecho

• 2Dbicyclestressecho

• 2Dcontinuouscapturestressecho(treadmill stress echo)

• Cardiacresynchronizationtherapyprogramming protocols (available with the Advanced QScan option)

Protocol Examinations Features(enabled with stress option)

• Wallmotionscoring:Analysis by wall motion in individual myocardial segments

• Showreference:Showareferenceimage from baseline or previous level during acquisition

• Smartstress:Automaticallysetupvarious scanning parameters (for instance geometry, frequency, gain, etc.) according to same projection on previous level

• Scanmodesettings:Scanmode may be specified for individual views in the protocol

• Previewofstore:Showrunning loops as preview before storing to the examination

Continuous Capture

• Continuouslyacquirelargeamountsof 2D image data, and selection of pro-jection views for analysis afterwards

• Theentirecontinuouscapturerecording may be kept in memory while it is possible to store new images outside the protocol template, or the entire recording can be stored to file

• TVanndiam(TricuspidValveAnnulusDiameter, 2D)

• TVArea(TricuspidValveArea,2D)

• TVCO (Cardiac Output by Tricuspid Flow)

• TVDecSlope (Tricuspid Valve Flow Deceleration)

• TVEVelocity(TricuspidValveEVelocity)

• TVE/ARatio(TricuspidValve E-Peak to A-Peak Ratio)

• TVmaxPG(TricuspidValve Peak Pressure Gradient)

• TVmeanPG(TricuspidValve MeanPressureGradient)

• TVmeanPG(TricuspidValve MeanPressureGradient)

• TVPHT (Tricuspid Valve Pressure Half Time)

• TVSV (Stroke Volume by Tricuspid Flow)

• TVVmean(TVMeanVelocity)

• TVVTI (Tricuspid Valve Velocity Time Integral)

• VSDmaxPG (VSD Peak Pressure Gradient)

• VSDVmax(VSDPeakVelocity)

PleaserefertotheReferenceManualfor the full list of measurements and calculations for all applications.

AnnotationsBody Marks

• Bodymarkiconsforlocationandposition of probe

• Easyselectionofbodymarksfromtouch panel

Text Annotations

• Easyselectionoftextannotationsfrom touch panel

Scan Assist Pro (optional)• Customizableautomationsthat

assist the user through each step of the scan

• Facilitates consistency and reduce keystrokes

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 16 of 18

• Selectionofprojectionviewsonscanner or EchoPAC when the entire recording is stored to file

Wall Motion Scoring

• Aspartofthemeasurementandanalysis package one can access a wall motion assessment module, providing analysis/scoring of individual myocardial segments

• Forusewithallstressmodalities

Cardiac Resynchronization Therapy (CRT) Programming Protocols• CRTprotocolsrequireStressand

Advanced QScan

• Tailoredacquisitionprotocolfordataneeded for programming of AV and VV delays in biventricular pacemakers

• Imageacquisitionofasetofprojection views with various scan mode settings

• Templateeditor

• User-configurableprotocoltemplates

• Configureprotocolname,number of levels and views, name of level and views and several other protocol settings (smart stress, show reference, scan mode, preview of store, timer handling, etc.)

4D Analysis Tools (with 6VT probe and 4D option)

4D Views

• Autoalignmenttodefinestandardorientation of acquired 4D data

• Standardviews,suchas4CH,2CH,LAX, mitral valve and aortic valve, are defined from the standard orientation

• Automaticdisplayofvolume renderings and 2D cut planes from standard views

4D Data Cropping

• Flexibletoolforstandardordynamiccropping (with Advanced 4D option) 4D data using up to six different crop planes

• Eachcropplanecanbemoved without any restrictions

• TheVividS70ultrasoundunit meetstheEMCrequirementsin IEC/EN60601-1-2:2007 Class B

Virus ProtectionTo reduce virus vulnerability, Vivid S70 is configured with a minimal set of open ports and with all network services not actively used by the system closed down. This helps to significantly reduce the risk of a virus attack on Vivid S70.

GE is continuously judging the need for additional actions to reduce vulnerability of equipment; this includes vulnerability scanning of our products and evaluation of new security patches for the 3rd-party technologyused.Microsoft® (and other) security patches that address serious issues with Vivid S70 will be made available to customers after GE verification of those patches.

TransducersM5Sc-D XDclearActiveMatrixSingleCrystal Phased Array Probe

• Probepresets:Cardiacadult,pediatric, abdomen, fetal heart, transcranial, coronary, stress, LVO contrast, LVO stress,OB/GYN,vascular

• Biopsyguide:Multi-angledisposablewith a reusable bracket

6S-D Phased Array Probe

• Probepresets:Pediatric,cardiac, coronary, neonatal head, fetal heart, abdominal

12S-D Phased Array Probe

• Probepresets:Pediatric,cardiac,coronary, neonatal head, rodent, vascular (incl. carotid, LEA, LEV, UEA, UEV), abdomen

9L-D Linear Array Probe

• Vascular(incl.carotid,LEA,LEV,UEA,UEV), musculoskeletal, nerves, small parts, thyroid

• Biopsyguide:Multi-angledisposablewith a reusable bracket

• Thecropplanepositionsarevisible in both the volume rendering and in the 2D cut plane displays

Depth Render

• Volumevisualizationwherethe color hue changes according to the distance into the image

• Wideselectionofdifferent render maps

Stereo Render

• Volumevisualizationbystereoscopicdisplay necessitates the use of red/cyan stereoscopic glasses for StereoVision

MultiSlice

• Simultaneousdisplayof5,6,7,9 or 12 slices extracted from the 4D volume data (tissue and/or color)

• Combinationofshortaxisandlongaxis standard views.

• Availableinlive(withAdvanced4Doption) and replay

FlexiSlice

• Simultaneousdisplayofthree independent random slices through the 4D volume (tissue and color)

Safety Conformance• TheVividS70isbuilttomeet

the requirements of:

• IEC60601-2-37

• IEC60601-1

• IEC60601-1-2

• IEC60601-1-6

• UL60601-1

• NEMAUD3

• TheEuropeanMedicalDevices Directive(MDD)93/42/EEC(CEMark)

• Directive2011/65/EUonthe restriction of use of certain hazardous substances

• TheVividS70ultrasoundunitisaClass I device, with BF (probes) and CF (ECG leads) applied parts according to IEC60601-1

Vivid S70 product datasheet – June 2015 – DOC1662324 Page 17 of 18

C1-6-D XDclear Single Crystal Curved Array Probe (Convex)

• Probepresets:Abdomen,OB/GYN,urology, vascular, fetal heart

• Biopsyguide:Multi-angle,disposablewith a reusable bracket

P2D Pencil Probe

• Probepresets:Cardiac

6Tc-RS TEE Probe

• Probepresets:Cardiac,coronary, LVO contrast

6VT-DActiveMatrix4DVolumeTEEProbe (4D and multi-plane only with 4D option)

• Probepresets:Cardiac, LVO contrast, coronary

9T-RS TEE Probe

• Probepresets:Pediatric

Wideband Probes• Electronicselectionbetweenfour

solid-state and one stand-alone Doppler probe connectors

• ThreeprobesocketsareDLPtype

• OneRSprobesocketforsupport of TEE

PROBE FREQUENCY RANGE CATALOG #

M5Sc-D(Sector) 1.5–4.6MHz H44901AE

6S-D (Sector) 2.4–8.0MHz H45021RR

12S-D (Sector) 4.0–12.0MHz H45021RT

9L-D(Linear) 2.4–10.0MHz H40442LM

C1-6-D (Convex) 1.6–6.0MHz H40472LT

P2D (Pencil) 2.0MHz H4830JE

6Tc-RS (TEE) 3.0–8.0MHz H45551ZE

6VT-D (Volume TEE)1 3.0–8.0MHz H45581BJ

9T-RS(TEE) 3.3–10.0MHz H45531YM

1 Also 6VT-D with catalog # H45561TA is supported.

DOC1662324

About GE HealthcareGE Healthcare provides transformational medical technologies and services to meet the demand for increased access, enhanced quality and more affordable healthcarearoundtheworld.GE(NYSE:GE)worksonthings that matter – great people and technologies taking on tough challenges. From medical imaging, software & IT, patient monitoring and diagnostics to drug discovery, biopharmaceutical manufacturing technologies and performance improvement solutions, GE Healthcare helps medical professionals deliver great healthcare to their patients.

GE Healthcare9900InnovationDriveWauwatosa,WI53226U.S.A.

www.gehealthcare.com

©2015 General Electric Company – All rights reserved.

June 2015

General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE representative for the most current information.

GE, GE monogram, Vivid, XDclear, EchoPAC, InSite and cSound are trademarks of General Electric Company or one of its subsidiaries.

DICOMistheregisteredtrademarkoftheNationalElectricalManufacturersAssociationforitsstandardspublications relating to digital communications of medical information.

WindowsandMicrosoftareregisteredtrademarksof MicrosoftCorporation.