Varian SpectrAA 50/55/110/220

AA

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 1

Page 2: Varian SpectrAA 50/55/110/220 brochure

2

Varian, Inc. is a world leader ininstrumentation for atomic spectroscopy.From our development of the firstcomponents for atomic absorptionspectrometers in the 1950s to the recentground breaking Vista series ofsimultaneous ICP emission spectrometers,Varian, Inc. has over 40 years experience of innovation and development in all areas of atomic spectroscopy.

SpectrAA Series—because experience counts

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 2

Page 3: Varian SpectrAA 50/55/110/220 brochure

Since their introduction in 1985, Varian’s SpectrAA series atomic absorptioninstruments have been chosen by thousandsof analysts around the world for their ease-of-use, reliability and performancewith challenging samples. The newgeneration of SpectrAA series instrumentscontinues this proud tradition, offering thesuperb performance, simplicity of operation,and industry leading software expectedfrom a world leader in atomic spectroscopy.

Varian’s SpectrAA series provides you withthe ideal solution to your atomic absorptionneeds. The SpectrAA family ranges fromstand-alone systems (SpectrAA 50/55) andPC-controlled instruments with varyinglevels of automation (SpectrAA 110/220),through to the innovative Fast SequentialAA (SpectrAA 220 FS) and our high-performance SpectrAA 880. When combinedwith the modular design and completeaccessory options, the SpectrAA series canbe configured to meet all your analyticaland budgetary needs.

The SpectrAA Family comprises:

SpectrAA 50Stand-alone single beam AA spectrometer

SpectrAA 55Stand-alone double beam AA spectrometer

SpectrAA 110PC-controlled single beam AA spectrometer

SpectrAA 220PC-controlleddouble beam AA spectrometer

SpectrAA 220 FS*A fully automateddouble beam system with FastSequential operationfor improvedproductivity

SpectrAA 880*Our highestperformance system featuring fullyautomated hardwareand research grade optics

* Refer to separatebrochure for details

3

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 3

Page 4: Varian SpectrAA 50/55/110/220 brochure

High performance entry level AA

4

Varian’s SpectrAA 50/55 series instrumentsare ideal for any laboratory requiring anentry level AA system with advancedcapabilities. The single beam SpectrAA 50and the double beam SpectrAA 55 feature a LCD screen and dedicated keyboardenabling simple, stand-alone operation.

On-screen displayThe SpectrAA 50/55 interface enables you to view and edit instrument parameters on-screen, ensuring fast, easy operation.Complete results, including concentration,precision and replicate readings aredisplayed. This allows you to confirm results as each reading is completed. After calibration, you can view thecalibration graph to check goodness of fit,or read absorbance or concentration values from the graph.

Multi language supportThe interface is available in a choice oflanguages. Language selection can bechanged at any time, making this instrumentideal for use in a multilingual environment.

Easy setupAdvanced features such as automatedwavelength and slit selection ensure easysetup, even for novice users. Two lamppositions are provided so you can completeone analysis while pre-warming the lampfor the next element in sequence.

Cookbook and custom methodsCookbook methods with recommendedinstrument parameters are provided for allelements. These methods can be recalledand modified at any time. You can also saveup to 30 customized methods, providingeasy access to your most common analyses.When you recall a method, the fullyoptimized instrument conditions are loadedinto the system for immediate use.

The SpectrAA 50/55series can be used withthe revolutionary SIPS(Sample IntroductionPump System). SIPS enables you toperform onlinecalibration from asingle standard and fast online dilution of over range samples.

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 4

Page 5: Varian SpectrAA 50/55/110/220 brochure

5

Background correctionThe optional, fast response deuteriumbackground correction features a 2 msresponse time for accurate correction of transient background signals. The deuterium lamp is located in the lamp compartment for easy optimizationand fast replacement by the user (when required).

Upgrade to PC ControlThe SpectrAA 50/55 series can be upgraded from stand-alone operation to aPC controlled system for semi-automated,sequential, multi-element analysis. Upgrades can be performed at any time,ensuring that the instrument you purchase today will continue to meet yourrequirements in the future.

Full automation with SpectrAA 110/220 The SpectrAA 110/220 series offers youenhanced capabilities, including full PCcontrol and automated operation asstandard. Based on the reliable SpectrAA50/55 hardware, the SpectrAA 110/220 seriesis available with a range of options,including automatic lamp selection andprogrammable gas control, providing fullyautomated operation.

The SpectrAA 110/220 optics feature four lamp positions with automatedwavelength and slit selection, and highintensity deuterium background correctionas standard. Lamps are mounted in fixedpositions, and a rotating mirror is used to select the operating lamp, eliminating the need for a conventional turret. For maximum stability in an automatedsequence, a software controlled next-lampwarm up facility is provided

The SpectrAA 50/55 and the SpectrAA-110/220 are compatible with Varian’scomplete range of AA accessories.

Upgrade the SpectrAA50/55 to external PCcontrol for fullaccessory support.Experience the benefitsof semi-automatedanalysis, includingautomatic samplepresentation using the SPS-5 autosampleror trace analysis using the GTA-110graphite furnace.

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 5

Page 6: Varian SpectrAA 50/55/110/220 brochure

Innovative worksheet software

6

Status display provides an instantsummary of instrumentconditions for theselected element and keeps you informed of progressduring analysis.

View concentrationresults, precision andabsorbance results asthe analysis progresses.Color coding and flags(displayed after thesample result) highlightout-of-range results,uncalibrated results orother errors.

Using OLE-2, resultscan be exported on-line during analysisto other commercialsoftware programs(e.g., Microsoft Excel),for custom datamanipulation.

Autocopy and uniquepattern recognitioncapabilities simplifymanual label entry.Alternatively, you canimport directly fromLIMS or use bar codereaders to scan sample labels directlyinto the software.

The message line informs you of the system status and provides operational hints.

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 6

Page 7: Varian SpectrAA 50/55/110/220 brochure

7

The PC controlled single beam SpectrAA 110and true double beam SpectrAA 220combine the rugged and reliable hardwareof the SpectrAA 50/55 series with Varian’sacclaimed worksheet software. Using themulti-tasking capabilities of Windows, you can start today’s analysis whilesimultaneously preparing a report onyesterday’s results. Imagine the productivity improvements you will achieve. The spreadsheet concept at theheart of the SpectrAA software and thesimplified interface also means minimaltraining is required.

The Worksheet AAFinal concentration results are displayed(after weight/volume correction if required)in a ‘spreadsheet’ concept common to manybusiness applications. Results for flame,vapor and furnace determinations aredisplayed simultaneously so that you canview results for the entire batch of samples.

Quality ControlExtensive rate and label driven QualityControl Protocols are provided to validatedata as it is collected. The optional QualityControl Protocols software provides evengreater capability for full compliance withall international regulatory requirements.This includes the capability to create custom QC tests.

Which operating system?Choose from Windows 95/98, Windows NTor Windows 2000 operating systems.

Click on an individualresult to display thesignal graphics, wherestored and detailedresults can be viewed.You can customize thedisplay to view only theinformation you need.For example, select fromthe worksheet, signalgraphics, calibrationgraph with fullcalibration data or the data log withcomplete analyticalresults. The data log isdate and time stampedand includes allmeasurement parameters,raw data, final resultsand error messages.

Comprehensive multi-media Help includesstep-by-step audio andvideo instructions onhardware setup androutine maintenance.

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 7

Page 8: Varian SpectrAA 50/55/110/220 brochure

Still the world’s leading flame atomization system

8

Varian, Inc. flame atomic absorption systems have always set the performancebenchmarks, and the Mark 7 design fitted to the SpectrAA series offers considerableimprovements over competitive systems.

ConstructionHigh performance and unrivalled flexibilityare combined with new ease of use features, such as ‘twist and lock’ assembly to simplify maintenance. Corrosion resistantcomponents are standard, providing evenlonger durability.

Reduced blockageThe contoured design of the Mark 7 burnerprovides a stable flame with outstandingresistance to blockage from dissolved solids or carbon build-up. This ensureslonger running times, even with the mostdifficult samples.

High sensitivityHigh sensitivity and good precision areassured. Typical performance exceeds 0.75 absorbance units for 5 mg/L copperwith precision of less than 0.5 % RSDroutinely achieved.

Key advantages of the Mark 7 systeminclude:

• High sensitivity: typically > 0.75 Abs from 5 mg/L Cu

• Excellent precision: < 0.5 % RSD from ten 5 sec integrations

• An externally adjustable impact bead thatenables you to tune the system formaximum sensitivity or to achieve extendedoperation with difficult samples

• Precise, accurate determinations withreduced interferences

• Minimal burner blockage, even with themost difficult samples

• Excellent drainage without spiking,memory effects, or long wash-out times

• Rapid change-over to organics operation,by simply installing organic resistant O rings

The externallyadjustable impact beadenables you to easilyoptimize sensitivity and precision.

High salt content insamples can rapidlyinduce burner blockage.The Mark 7 atomizationsystem achievesexcellent long termstability and reducedblockage even withcontinuous aspirationof a 7.5 % sodiumchloride solution.

120

100

80

60

40

20

01 5 9 13 17 21 25 29 33 37 41 45 49 53 57 61

Rela

tive

Abs

orba

nce

Time (minutes)

Mark 7

5 µg/L Cu in 7.5% NaCl

Other

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 8

Page 9: Varian SpectrAA 50/55/110/220 brochure

9

The adjustable nebulizeris compatible with all common solventsand enables you toconveniently set therequired flow rate. The platinum-iridiumcapillary and PEEKventuri offeroutstanding resistanceto corrosive solutions.

The Mark 7 burnerprovides a stable flamewith outstandingresistance to salt andcarbon build-up. This translates intolonger running timesand higher samplethroughput.

The safety systemconstantly monitors all interlocks, includingthe liquid trap floatand pressure reliefbung. They eitherprevent flame ignitionor instantaneouslyextinguish the flame if a fault is detected,ensuring safe andreliable operation.

Disassembly of thespraychamber for cleaning is easy.Simply rotate the liquid trap to remove it and then twist thespray chamber toseparate it from thenebulizer block. No tools are neededand there are no safetychains or exposed gas hoses to remove.

The twin headed mixing paddles remove unfavorablylarge droplets, reducing interferencesand minimizing burner blockage, even when aspiratinghigh dissolved solids solutions.

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 9

Page 10: Varian SpectrAA 50/55/110/220 brochure

10

Flexible sample preparation

More than autosampling!Automate your analysis with the SPS-5(Sample Preparation System) and enhancethe productivity of your laboratory.

The SPS-5 offers so much more than anyother autosampler, with unmatched samplecapacity for extended operation, off-linesample preparation capabilities, and aflexible choice of racks accommodatingdifferent test tube types. You even have theflexibility to use your own racks.

Key capabilities include:

• Capacity for up to 500 solutions, loaded in convenient sample racks

• The ability to determine more solutions,simply by changing racks

• A choice of five inert rack types fordifferent vessel sizes

• A flexible ‘Type 25’ rack with adjustablesettings to accommodate different tube heights and low cost rack overlays for alternative tube diameters

• Automatic rack identification that reduces rack-changing errors. All rackscontain a unique magnetic code thatenables the software to identify racks loaded during analysis.

• A custom rack definition feature thatenables you to program the SPS-5 to acceptany rack type in any position.

• A programmable probe height that enableslayers within a solution to be sampled

• Three rack positions are supported as standard

• An optional rack extension supports twoadditional racks

The SpectrAA 110/220can be combined with the revolutionarySIPS and the SPS-5flame autosampler fora fully automatedsystem. SIPS can alsobe used with manualsample introduction.

Off-line sample preparationBy combining the SPS-5 with the optionalDiluter and Roboprep software, you canprevent the problems associated with samplepreparation. Tedious, time-consumingsolution preparation is eliminated, as arevolumetric based dilutions that are prone toerror. SPS-Roboprep achieves realproductivity gains by providing a powerfulsample preparation system that is costeffective and practical. The SPS-Roboprepcan be used off-line, while the laboratory isunattended, for preparation of:

• Single or multi-element calibrationstandards

• Single or multi-element addition standards

• Spiked samples for spike recovery studies

• Single or serial dilution of samples withaddition of up to 2 reagents to maintainsample chemistry

Solutions are automatically stirred withSmart Mixing to ensure homogeneity.

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 10

Page 11: Varian SpectrAA 50/55/110/220 brochure

11

SIPS—A revolution in sampleintroductionVarian’s SIPS (Sample Introduction PumpSystem) is an innovative sampleintroduction and dilution system for flameAA. Using the revolutionary SIPS, you caneliminate multiple standard preparationsand dramatically simplify over-range sample dilution.

SIPS is available in two versions— the singlepump SIPS-10 and the dual pump SIPS-20.When used with the SpectrAA flame series instruments, both systems provide:

• Multi-point calibration from a single standard

• Intelligent on-line dilution of over range samples

• Simplified operation and improvedproductivity

The dual pump SIPS-20 provides evengreater capability. When used with theSpectrAA 110/220, the SIPS-20 can:

• Automate the tedious task of flamestandard additions calibrations

• Perform on-line spiking of samples forspike recovery studies

• Add modifiers such as ionizationsuppressants on-line

The benefits of SIPS• Provides on-line multi-point calibrationfrom a single standard

• Eliminates tedious multiple standardpreparation

• Eliminates dilution errors during standardpreparation

• Provides fast on-line dilution of overrange samples with ‘Smart Rinse’ to reducememory effects

• Eliminates manual dilution and re-analysisof over-range samples

• Provides excellent dilution accuracy withless than 2% error

• Extends the linear dynamic range of flameAA, approaching that of ICP-OES

• Improves the productivity of flame AA determinations

On-line calibrationUsing SIPS, only one calibration standard isrequired. Simply define the standardconcentration and the number of calibrationpoints required. SIPS does the rest for you—automatically preparing the calibrationby diluting your standard on-line.

Fast on-line dilutionSIPS will also automatically dilute overrange samples without any operatorintervention. If any solution is over range,measurement is terminated automaticallyafter the first replicate. A ‘Smart Rinse’ isthen performed, where diluent is aspiratedto wash out the atomization system. After checking for possible contamination,the over range sample is re-measured withdilution applied. ‘Smart Rinse’ optimizesrinse times, minimizes memory effects andincreases sample throughput.

SIPS provides superbaccuracy withperformance exceedingthat achieved usingmanual dilutions.Typical dilutionperformance across thedilution range is shown.

On-line calibration & dilution

105

103

101

99

97

950 20 40 60 80 100 120 140 160

Rela

tive

Acc

urac

y (%

)Dilution Factor

SIPS Dilution Accuracy

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 11

Page 12: Varian SpectrAA 50/55/110/220 brochure

12

Constant Temperature Zone designVarian’s furnace research has demonstratedthat the best graphite furnace performanceis achieved using a Constant TemperatureZone (CTZ) graphite furnace. The CTZ iscreated by carefully matching the design of the GTA-110 power supply to low mass,end-heated graphite tubes. The GTA-110 also features the most advanced furnacesampling systems.

When coupled with the SpectrAA series,Varian’s longitudinally heated GTA-110 CTZgraphite furnace provides a fully automatedsystem capable of tackling difficult samplesand achieving exceptional performance.

Benefits of the CTZ designThe proof of Varian’s CTZ is performance.Varian’s CTZ design GTA-110 graphitefurnace provides:

• Rapid, even heating of the low mass, end heated graphite tubes

• High sensitivity, from the long atom celland CTZ design, ensures the lowest possibledetection limits

• Reduced memory effects, even for thedifficult refractory elements

• Accurate and reproducible temperaturecontrol, achieved using our patentedDynamic Feedback Temperature Control system

• Optimized inert gas flows ensure extended graphite tube lifetimes areachieved, providing economical operation.

Using Varian’s GTA-110, typical tubelifetimes easily exceed 1000 firings,even for the difficultrefractory elements.

Abso

rban

ce

Number of FiringsV at 2600 deg. C.

2000

0.2

0.4

0.6

0.8

1.0

1.2

1.4

400 600 800 1000 1200 1400 1680

Mn at 2200 deg. C.

Superior furnace performancefor difficult samples

Typical performanceElement Wavelength Slit Width Characteristic Detection Limit (3 sigma)

(nm) (nm) Mass (pg) Peak Height Peak Area(µg/L) (µg/L)

AI 396.2 0.5 5.0 0.2 0.2

As* 193.7 0.5 6.0 0.25 0.4

Cd 228.8 0.5 0.2 0.01 0.01

Cr 357.9 0.2 R 1.5 0.075 0.075

Ni 232.0 0.5 4.8 0.5 0.5

Pb* 283.3 0.5 3.0 0.2 0.25

Se* 196.0 1.0 10 0.3 0.3

Ti 364.3 0.5 R 50 0.75 0.85

V 318.5 0.5 22 0.5 0.75

* Determined using an UltrAA lampR = reduced slit height

Typical characteristic mass figures (the mass of analyte required to produce 0.0044 absorbance) and 3 sigma detection limits are shown.

SAA_50.55.110.220_Bro 25/9/00 1:05 PM Page 12

Page 13: Varian SpectrAA 50/55/110/220 brochure

13

Whether you are using flame, furnace orvapor AA, Varian’s range of UltrAA lampsincrease sensitivity and reduce detectionlimits, allowing determinations of tracemetals at even lower levels. The UltrAA lampseries of high intensity, boosted discharge,hollow cathode lamps can be used with allSpectrAA series instruments, for the mostdemanding applications.

Lamp operationUltrAA lamps use the standard lamp current, but apply a second discharge within the lamp. This boost current issupplied from an external control module.The resulting sharper emission profilereduces self-absorption and line broadening,increasing sensitivity of AA determinationsby up to 40%. Emission intensity is alsoincreased by three to five times, reducingbaseline noise levels. The lower baselinenoise levels and the higher sensitivity ensurelower detection limits can be achieved.

Benefits of the UltrAA lamp• Increased sensitivity—comparative signals for a 75 µg/L Se standard are shown(in the graph opposite)

• Lower detection limits

• Enhanced calibration linearity

• Longer lamp lifetimes—typical lamplifetimes exceed 8000 mA hours of operation

• Simple installation—lamps are mounteddirectly into the socket (without any flyingleads), just like conventional lamps

• A fixed boost current eliminates difficultmanual adjustments

• Automatic lamp recognition enables thesoftware to identify the lamp and select therecommended operating conditions

• Enables simultaneous operation of twoUltrAA lamps—one lamp can be in use whilethe other is being pre-warmed

UltrAA lamps—lower detection limits

Typical detection limits achieved using UltrAA lampsElement Wavelength Slit Width Detection Limit (3 sigma)

(nm) (nm) Conventional UltrAA lamp lamp (µg/L) (µg/L)

As 193.7 0.5 1.4 0.25

Pb 283.3 0.5 0.8 0.20

Se 196.0 1.0 1.1 0.30

Abso

rban

ce

Tem

pera

ture

Time (sec)

0.00

0.50 0.50

0.00

Standard 3

Conventional HCL

UltrAA lamp

72.0 74.5

0

2500

Back

grou

nd

SAA_50.55.110.220_Bro 25/9/00 1:06 PM Page 13

Page 14: Varian SpectrAA 50/55/110/220 brochure

14

High productivity vapor generation

High sample throughputVarian’s fully automated, modular VGA-77(Vapor Generation Accessory) is ideal forrapid trace level analyses of mercury (usingthe proven cold vapor technique) and otherhydride forming elements at part per billion(µg/L) concentrations. The VGA-77 offers:

• Maximum sensitivity

• High sample throughput of up to 70 samples per hour

• Excellent precision of 1-2% RSD

• Low sample consumption of less than 8 mL per element during analysis

• Modular construction for rapidchangeover and reduced contaminationwhen switching between elements withconflicting chemistries

Advantages of Continuous Flow Vapor GenerationThe continuous flow VGA-77 offers thefollowing advantages over competitivetechniques, such as Flow Injection:

• Maximum sensitivity

• Enhanced precision—the VGA-77 producesa steady state signal, which can bemeasured using integration. Flow Injectionproduces transient signals, requiringmultiple injections for precise results.

The ETC-60 enhancesproductivity of hydridegeneration AA byproviding unattended,flameless atomization ofthe hydride elements.

• Greater productivity—results are obtainedusing the VGA-77 in less than a minute

• Simple and automatic operation—eachsample is automatically combined with acontinuous flow of acid and reagents

Unattended hydride analysisFor even greater productivity, or increasedsensitivity from the vapor technique,combine the VGA-77 with Varian’s SPS-5autosampler and ETC-60 ElectrothermalTemperature Controller. The ETC-60 is anelectrically heated quartz cell, whicheliminates flame heating of the atomizationcell, enabling safe, unattended, flamelesshydride analyses.

The lower atomization temperatures usedwith the ETC-60 (typically <1000 ˚C)increase sensitivity by up to 30%, ensuringsub µg/L detection limits are achieved.

SAA_50.55.110.220_Bro 25/9/00 1:06 PM Page 14

Page 15: Varian SpectrAA 50/55/110/220 brochure

15

Reducing the cost of ownershipThe SpectrAA series were designed with cost of ownership in mind. The easy-to-usesoftware will minimize training and get your lab up and running faster.

For protection in dusty or corrosiveenvironments, the optics are completelysealed and all mirrors are quartz overcoatedfor long term, stable performance. When connected to a supply of clean, dry air, the internal air purge ensures totalprotection from attack in corrosive, humid or dust-filled environments.

Comprehensive diagnostics assist rapidtrouble shooting and fault diagnosis,maximizing instrument uptime.

Varian’s Telediagnostics option connects you to our worldwide network of supportpersonnel for on-line assistance. This remote diagnosis service minimizes on-site calls and reduces down-time.Telediagnostics also gives you the freedomto access your data at any time. You canview the progress of your analysis from yourhome, exchange methods or results withother users or connect to other laboratoriesin your company.

GLP Compliance and ValidationThe SpectrAA software complies with manyinternational standards for Good LaboratoryPractice (GLP). If you need to validate yourSpectrAA system, Validation documentationis available for SpectrAA instruments,software and accessories. Varian, Inc. service organizations around the worldsupport validation of our instruments in anumber of ways, including trainingprograms, support agreements, hotlines,Telediagnostics, service contracts andcertification. An overview of the Validationdocumentation and services Varian, Inc.provides is available from your local Varian, Inc. office.

SafetyIt is Varian’s policy to manufacture safeproducts and to meet all legal requirementsgoverning the design, manufacture and saleof safe products. As with all similarproducts, the following hazards may bepresent: high temperatures, high pressuregases, explosive gases, UV and visible lightand electricity. Each product is designed toprotect operators from potential hazards.

Varian, Inc. supplies instructions thatdescribe the correct procedures for theoperation and maintenance of each product.

For clarity, some images in this brochuremay depict the spectrometer being usedwithout safety covers. Ensure all safetycovers, chimney etc. are in place for normal operation.

SpectrAA atomic absorption spectrometersare designed to determine theconcentrations of trace and major elementsin solution.

From ownership to partnership

World Wide WebVarian’s home pagehttp://www.varianinc.com is another part of our on-going commitment to customersupport. You can access application notes,newsletters and updates about latestproduct developments or you can email us if you have a question. Use our on-line Parts and Supplies catalog to identify anyconsumables you need. Immediate access to information, applications support andtechnical service is never more than amouse click, or a phone call away.

The SpectrAA series instruments are certified to comply with the

requirements of the EMC and LV directivesof the European Union.

SAA_50.55.110.220_Bro 25/9/00 1:06 PM Page 15

Page 16: Varian SpectrAA 50/55/110/220 brochure

Prin

ted

on e

nviro

nmen

tally

frie

ndly

pap

er. D

esig

n: C

ruis

e De

sign

Par

tner

ship

. Pho

togr

aphy

: Ear

l Car

ter

Phot

ogra

phy.

Visit Varian atwww.varianinc.com

Sales offices ArgentinaTel (54) 11 4783 5306

Australia Tel 1 300 368 274

Austria Tel (43) 1 69 99 669

Belgium Tel (32) 15 55 6460

Brazil Tel (55) 11 3845 0444

Canada Tel 1800 387 2216 or (1) 950 819 8181

ChinaTel (86) 21 6375 6969

Europe (other) Tel (31) 11 867 1000

France Tel (33) 1 6986 3838

Germany Tel (49) 6151 7030

IndiaTel (91) 22 857 0181

Italy Tel (39) 11 997 9111

Japan Tel (8) 3 5232 1239

Korea Tel (82) 345 22452

Mexico Tel (52) 5 533 5985

Netherlands Tel (31) 164 282800

Russian Federation Tel (7) 502 9374280

Spain Tel (34) 1 472 7612

Sweden Tel (46) 8 445 1600

Switzerland Tel (41) 41 749 8844

TaiwanTel (886) 22 698 9555

United Kingdom & Eire Tel (44) 1932 898000

United States Tel 1800 926 3000 or (1) 925 939 2400

VenezuelaTel (58) 41 257 608

Other countries Australia Tel (61) 3 9560 7133 Fax (61) 3 9560 7950

Specifications subject tochange without notice.

© Varian Australia Pty LtdACN 004 559 540

Partnumber 8510155100Australia 9/2000

®Windows is a registered trademark of Microsoft Corp.

Varian Analytical Instruments, serving worldwide markets in:

AgricultureBasic ChemicalBiotechnologyClinicalElectronicsEnvironmentalPhotonicsToxicologyPharmaceuticalFood and BeverageMetals and MiningPetroleum and Petrochemical

GC, GC/MS, HPLC, AAS, ICP, ICP-MS, UV-Vis-NIR,Fluorescence, NMR, andAnalytical Supplies

Varian is committed to aprocess of continuousimprovement whichdemands that weunderstand and then meetor exceed the needs andexpectations of ourcustomers—both inside andoutside the company—ineverything we do.

SAA_50.55.110.220_Bro 25/9/00 1:06 PM Page 16