V-730 Lenses Spectrophotometer Solution for analysis of sunglasses and ophthalmic lenses

Upload: others

Post on 22-Feb-2020

24 views

Category:

Documents


1 download

TRANSCRIPT

V-730 Lenses Spectrophotometer

Solution for analysis of sunglasses and ophthalmic lenses

V-730 Lenses Spectrophotometer

JASCO V-730/Lenses UV/VIS double beam Spectrophotometer is an optimised system to estimate compliance and driving suitability of sunglasses and ophthalmic lenses according with current regulations. • European ISO EN 12312-1:2015 • European EN 1836:2005+A1:2007 • American ANSI Z80.3-2015 • Australian/New Zealand AS//NZS 1067.1:2016 • American ANSI Z87.1 (only for V-770

spectrophotometer)

V-730/Lenses is also a UV-VIS double beam spectrophotometer which may be used for research and routine analysis like spectrum measurement, kinetics, quantitative analysis, multi-wavelength and many other kinds of spectroscopic applications. Fully PC controlled by new JASCO Spectra Manager II 64-bit Windows program that allows prompt and simple use.

Why choose JASCO V-730/Lenses

• Self-coloured, degr.coloured and photochromatic

lenses compliance according to current regulations (EN, ANSI; AS/NSZ)

• User Friendly interface and prompt analysis results

• Save analysis results • Customized printing report (single regulation or

all regulations in a single report) • Standard spectrophotometric measurement,

colour analysis, system Validation GLP/GMP compliance

• Optimization of your production, ideal to your quality control analysis

JASCO Lenses Program

Spectrum Measurement

SPECIFICATIONS

Optical system Double beam, single monochromator with a 1200 grooves/mm concave grating and modified Rowland mount

Spectral bandwidth 1 nm

Wavelength range 190 – 1100 nm

Light sources Deuterium lamp for UV range (190 – 350nm) Halogen lamp for VIS range (350 – 1100nm) (automatic switching)

Wavelength repeatability ± 0.1nm

Wavelength accuracy ± 0.5nm

Photometric mode Abs, %T, %R

Photometric range (guaranteed on the whole spectral range)

-3 ÷ 3Abs

Photometric reproducibility ± 0.001Abs

Photometric accuracy ± 0.0015Abs (obtained using NIST SRM 930D)

Stray light 0.02%T

Wavelength scan speed 10 to 8,000nm/min

Wavelenght slew speed 24,000nm/min

Baseline stability ± 0.0004Abs/hour

Baseline flatness ± 0.0005Abs (within 200 – 1000nm range)

Response time VQuick, Quick, Fast, Medium, Slow

Detector Silicon photo-diode (S1337)

RMS noise 0.00006Abs (500nm)

Power requirement 105W

Dimensions 486(W) x 441(D) x 216(H) mm

Net weight approx. 15Kg

STANDARD ACCESSORIES

V-730 basic unit, Holmium glass filter, USB 2.0 cable, AC power cable, spare fuses, hexagonal wrench, Spectra Manager II spectroscopy software, Jasco-Lenses software (Microsoft Excel 2003, 2007, 2010, 2013 required), instruction manual.

ORDERING INFORMATION

7066-J021A - V-730 UV/VIS Spectrophotometer (basic unit, including Spectra Manager II Software) J-LENTI – Jasco-Lenses software package

JASCO Europe S.r.l. - Via L. Cadorna, 1 23894 Cremella (Lecco) - Italy Tel. +39-039-9215811 Fax +39-039-9215835

e-mail: [email protected]

Carpi Branch - P.le della Repubblica, 2 41012 Carpi (Modena) - Italy Tel/Fax +39-059-699.530

e-mail: [email protected]

www.jasco-europe.com

V-730 Lenses Spectrophotometer

JASC

O V

-73

0-L

ense

s B

roch

ure

– R

ev.1

JASCO Europe S.r.l.

Via Cadorna, 1 - 23894 Cremella (LC)

[email protected]

www.jasco-europe.com

DISCLAIMER The contents of this publication are for reference and illustrative purposes only. Information, descriptions, and specifications in this document are subject to change without notice and cannot be used from third parts for data comparison and/or performance comparison. JASCO assumes no responsibility and will not be liable for any errors or omissions contained herein or for incidental, consequential damages or losses in connection with the furnishing, performance or use of this material.