USABILITY TESTING OF MEDICAL DEVICES AND SOFTWARE
Boston UXPA 2017
![Page 2: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/2.jpg)
2
Agenda
Medical device usability testing
Why is human factors critical for medical devices?
Q & A
Software and apps as medical devices
The medical device HF process
1 4
2
3
5
![Page 3: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/3.jpg)
3
Why is HF critical for medical devices?
![Page 4: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/4.jpg)
4
Because it’s required!
![Page 5: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/5.jpg)
5
Laws, Standards, and Guidance
• FDA Quality System Regulations21 CFR 820.30, Subpart C
Design Controls, paragraphs c, f, and g
• IEC 62366-1:2015Application of Usability Engineering to Medical Devices
• ANSI/AAMI HE75:2009 Human Factors Engineering – Design of Medical Devices
• FDA Guidance Document (2016)Applying Human Factors and Usability Engineering to
Medical Device Design
![Page 6: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/6.jpg)
6
The Medical Device HF Process
![Page 7: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/7.jpg)
7
The Medical Device HF Processfrom IEC 62366-1 & FDA Guidance
HF/UE planning User researchUse specification
Preliminary risk analysis
User interface design
Formative usability testing
Risk analysis/ mitigation
HF/UE validation
![Page 8: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/8.jpg)
8
User Interface Definition
• Physical device and accessories
• Software UI
• Labeling
• Instructions for use
• Training
Includes ALL communication with the user!
![Page 9: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/9.jpg)
9
The HF Process Applies to…
• Medical devices: all types of hardware from scalpels to surgical robots.
• Combination products: device plus drug or biologic.
• In-vitro diagnostic (IVD) devices.
• Software as a medical device.
![Page 10: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/10.jpg)
10
Software and Apps as Medical Devices
![Page 11: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/11.jpg)
11
Definition of a Medical Device
• A medical device is "an instrument, apparatus, implement, machine, contrivance, implant, in vitro reagent, or other similar or related article, including a component part, or accessory which is:
– …intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, in man or other animals, or
– intended to affect the structure or any function of the body of man or other animals…”
![Page 12: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/12.jpg)
12
• Software in a medical device (“embedded”)
• Software as a medical device (SaMD) (“stand alone”)
➢ Definition of SaMD is irrespective of software technology and/or platform (e.g., mobile app, cloud).1
Software and Apps as Medical Devices
What FDA does regulate
2 FDA (2015): Guidance on Mobile Medical Applications
1 International Medical Device Regulatory Forum (IMDRF) (2015): Software as a Medical Device: Key Definitions
![Page 13: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/13.jpg)
13
• Software in a medical device (“embedded”)
• Software as a medical device (SaMD) (“stand alone”)
➢ Definition of SaMD is irrespective of software technology and/or platform (e.g., mobile app, cloud).1
Software and Apps as Medical Devices
What FDA does regulate
2 FDA (2015): Guidance on Mobile Medical Applications
1 International Medical Device Regulatory Forum (IMDRF) (2015): Software as a Medical Device: Key Definitions
FDA:
“When the intended use of a mobile app is for the diagnosis of disease or other conditions, or the cure, mitigation, treatment, or prevention of disease, or is intended to affect the structure or any function of the body of man, the mobile app is a device.”2
![Page 14: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/14.jpg)
14
• Software used to make or maintain a device (testing, source code, servicing, etc.).
• Software that simply retrieves and/or organizes data.
• Mobile apps that are electronic copies of medical textbooks, teaching aids or reference materials, etc.
• Mobile apps that are solely used to log, record, track, evaluate, or make decisions related to developing or maintaining general health and wellness.
Software and Apps as Medical Devices
What FDA does NOT regulate
![Page 15: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/15.jpg)
15
• Software used to make or maintain a device (testing, source code, servicing, etc.).
• Software that simply retrieves and/or organizes data.
• Mobile apps that are electronic copies of medical textbooks, teaching aids or reference materials, etc.
• Mobile apps that are solely used to log, record, track, evaluate, or make decisions related to developing or maintaining general health and wellness.
Software and Apps as Medical Devices
What FDA does NOT regulate
FDA:
“…we intend to apply this oversight authority only to those mobile apps whose functionality could pose a risk to a patient’s safety if the mobile app were to not function as intended.”
![Page 16: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/16.jpg)
16
Medical Device Usability Testing
![Page 17: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/17.jpg)
17
Important Test Inputs
• Intended user profiles.
• Intended use environments.
• Detailed task analysis.
• A user interface design.
• Use-related hazards and risks analysis (at least preliminary).
![Page 18: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/18.jpg)
18
Formative Usability Testing
• Done early and often during development.
• To refine the design.
• Uncovers additional user requirements.
• Identifies additional use hazards.
![Page 19: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/19.jpg)
19
HF Validation Testing
• Also called “summative testing.”
• Required for FDA and CE approval.
• Based on use-related risk analysis.
• Must be conducted in a specific way.
![Page 20: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/20.jpg)
20
Determining User Profiles
• How you define user groups is key!
• “When their characteristics could affect their interactions with the device, or when their tasks are different.”
– Different age groups
– Different roles
– Trained vs. untrained users
![Page 21: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/21.jpg)
21
Sample Sizes for Validation
3 Loring & Wan (2017): Recruiting Patients with Rare Diseases and Their Caregivers
• Minimum of 15 participants per user profile.
• Sometimes means huge sample sizes – 60, 90, 120.
• Especially challenging for hard-to-find users.3
![Page 22: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/22.jpg)
22
Select Tasks Based on Risks
• Ensure the tasks tested include critical tasks (at a minimum).
• Define success and failure at the task level.
• Tasks that require the user to respond to alerts/alarms are considered critical tasks and should be tested.
• Warnings and cautions in the product labeling imply critical tasks and should be assessed through ‘knowledge questions.’
![Page 23: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/23.jpg)
23
Actual or Simulated Environment
• Mimic real world conditions:
– Lighting, acoustics, vibration
– Room layout and other equipment
– Interruptions and distractions
– Mobility and portability
• You may need to rent a medical simulation facility.
• Multiple use environments?
![Page 24: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/24.jpg)
24
Simulating Emergency Use Conditions
• Testing conditions should mimic emergency situation.
• Simulation of stress-induced environments can include:
– Continuous telephone ringing
– A beeping timer that increases in frequency and loudness
– Multiple people in a confined space
![Page 25: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/25.jpg)
25
Representative Training
• Training in the study should be the same as training in actual use.
• If training will not be provided consistently for every user, it may be OK to only test untrainedusers.
• ‘Decay of training’ period is often required.
![Page 26: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/26.jpg)
26
Moderating a Validation Test
• It’s different than what you may be used to!
• Simulate realistic interactions:
– Don’t tell participants to read the Instructions for Use.
– No thinking aloud (unless it’s spontaneous).
– No interruptions while participants are performing the tasks (unless they could get hurt).
– Don’t debrief until all tasks are done.
![Page 27: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/27.jpg)
27
Analyzing the Data
• Observational data (pass/fail, close calls, operational difficulties, and unanticipated/unknown problems).
• Interview data or subjective assessment from study participants.
• Answers to knowledge questions.
• Perform a root cause analysis.4
4 Wiklund, et al (2015): Medical Device Use Error: Root Cause Analysis
![Page 28: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/28.jpg)
28
The HF/UE Validation Report
• Present the data in a summary table for FDA review:
• Show data by distinct user profiles.
• Provide a detailed discussion of subjective data and root cause analysis.
TasksC=CriticalE=Essential
# of Task Failures/Use Errors
# of Close Calls/Operational Difficulties
Descriptions of Use Errors, Close Calls, Operational Difficulties
Root Cause Analysis
![Page 29: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/29.jpg)
29
Do You Need to Re-test?
• Evaluate what aspects of the UI may have caused or contributed to the issues seen in the study.
• Determine options to further optimize the user interface.
• Implement additional UI improvements.
• Update the use-related risk analysis….then
• Decide whether to perform additional HF studies to demonstrate that the mitigations addressed the errors and no new risks were introduced.
![Page 30: Usability Validation Testing of Medical Devices and Software](https://reader031.vdocuments.mx/reader031/viewer/2022021815/5a65f8e47f8b9aaf638b6a17/html5/thumbnails/30.jpg)
30
Final Conclusion
“The <Product> has been found to be reasonably safe and effective for the intended users, uses and use environments.
And…
Any residual risk that remains after the validation testing would not be further reduced by modifications of design of the user interface (including any accessories and the IFU), is not needed, and is outweighed by the benefits that may be derived from the device’s use.”