Troubleshooting the
NextSeq System
2
After completing this lesson, you will be able to:
Troubleshoot pre-run check errors
Identify helpful troubleshooting files
Locate troubleshooting resources
Learning Objectives
3
Agenda
ReagentsBaseSpace
ErrorsPre-Run and
System ChecksRe-hyb
WorkflowTroubleshooting
Files
4
BaseSpace Errors
5
• Ensure that the reagent barcode entered in Prep tab matches that of the reagent cartridge
• Or leave Reagent Barcode field blank
Reagent Barcode was entered incorrectly
Planned Run Not Available
6
• Option to “Automatically adjust for Daylight Savings Time” under the computer date and time settings is unchecked. Ensure that the following are CHECKED:
• “Automatically adjust for Daylight Savings Time”
• Network Time Server
• Correct local time zone of instrument
Daylight Savings Option
• Ensure network upload speed is at least 10 mbps
Network upload speed is too slow
BaseSpace Connection Errors
7
NextSeq Pre-Run Checks
8
Status Icons
Status icons signal any change
in the conditions during run
setup or during the run
When a change in condition
occurs, the icon image changes
and blinks
9
NextSeq Sequencing Pre-Run checks
10
Pre-Check Errors: Systems Checks
• Ensure that buffer compartment and reagent compartment doors are closed
Doors Closed
• Consumable sensors do not register
• Make sure that each consumable is properly loaded and repeat run set-up
Consumables Loaded
• Critical components of the software are missing
• Perform a manual software update to restore all software components
Required Software
• There is not enough disk space
• Free up disk space on the Network or remove any previously saved runs on the instrument
Instrument or Network Disk Space
• The network connection has been interrupted
• Confirm network status and physical network connection
Network Connection
11
NextSeq Sequencing Pre-Run checks
12
Pre-Check Errors: Temperature
• Checks the temperature ramp time of the flow cell heater plate
• If this check fails, contact Illumina Technical Support
Temperature Ramp
• Checks the speed of system fans in pulse per minute (PPM) to confirm fans are functioning
• If this check fails, contact Illumina Technical Support
Fans
• Checks the average temperature of each thermal sensor
• If this check fails, contact Illumina Technical Support
Temperature Sensors
13
NextSeq Sequencing Pre-Run checks
14
Pre-Check Errors: Imaging System
• Sends stages to their sensor limits
• If this check fails, contact Illumina Technical Support
Imaging Limits
• Measures ability for the 6 z-stages to move and stop moving in a set time
• If this check fails, contact Illumina Technical Support
Z Step-and-Settle
• Each camera sends a bit to the computer to make sure the circuits to each of the camera are working properly
• If this check fails, contact Illumina Technical Support
Bit Error Rate
• Verifies registration of the flow cell in the correct imaging position
• If fails, it is possible that the flow cell is not properly seated
• Unload and reload the flow cell to make sure that it is seated properly
Flow Cell Registration
15
NextSeq Sequencing Pre-Run checks
16
Pre-Check Errors: Reagent Delivery
• Tests the accuracy of valve and pump syringe rate movements
• If this check fails, contact Illumina Technical Support
Valve Response
• Moves both pump syringes to their min and max positions
• If this check fails, contact Illumina Technical Support
Pump
• Moves the sipper assay in the Buffer Straw Mechanism to it’s min and max sensor limits
• If this check fails, contact Illumina Technical Support
Buffer Mechanism
• Verifies spent reagents tray is empty
• If fails, empty spent reagents tray and return to the instrument
Spent Reagents Empty
21
NextSeq System Check
22
System Check
• Used during Live Help session with assistance from an Illumina representative
• Not required during normal operation or for instrument maintenance
When to use
• Select System Check from Manage Instrument screen
• Load consumables when prompted:
o Used flow cell
o Empty spent reagents tray
o Buffer wash tray with lab-grade H2O
o Empty reagent wash tray unless otherwise prompted
• Select tests to run
How to use
23
Live Help
Online assistance tool enabling
an Illumina Tech Support
representative to view the
NextSeq screen with permission
Obtain unique access code from
representative
Support.illumina.com/share_desktop.html
24
The following tests can be performed without assistance of an Illumina field representative:
• Motion tests
• Fluidics tests
• Thermal test
System Checks Tests
Note: Selecting Systems Check from the management screen will close the NextSeq Control
Software (NCS) and will open the NextSeq Service Software (NSS)
25
Rehybridization Workflow
26
Rehybridization Run
• Rehybridization of the Read 1 sequencing primer onto clusters
What is it
• If metrics generated during the first few cycles show intensities below 2500• Note: some low-diversity libraries can show intensities
<1000 which is expected
When to use it
• NextSeq is using NCS v1.3 or later
• Run must be stopped before paired-end synthesis
• Requires new NextSeq 500 kit
• Not all poor quality runs can be corrected with a rehybridization run
Considerations
27
End Current
Run
Save Flow Cell
Perform Manual Wash
Set up New Run on Prep
Tab
Start Run on
Instrument
Rehybridization Workflow
28
End Current Run
Select End Run
Select Yes to
save the flow cell
Brief wait while flow
cell is placed in safe state
RFID of saved flow
cell is unlocked
Automatic post-run wash is
bypassed
NOTE: The End Run command is final. The run cannot be resumed, run
consumables cannot be reused, and sequencing data from the run is not saved.
29
Perform a Manual Post-Run Wash
Reagent and buffer cartridges from the stopped run can be used for the
manual wash if they have not yet been removed
Perform Manual Wash
30
Set up a new run using the same parameters as the original run
Click the Pools tab, select the appropriate Pool ID to retain the previous
run settings, and then assign a unique name for the new run
Set Up New Run in Prep Tab
31
Prepare a new reagent cartridge
Allow saved flow cell to reach room temperature (15-
30’)
Load saved flow cell
Empty spent reagents
Load the new buffer and reagent
cartridges
Select the run from the run set
up screen*
Select Next to proceed to the
pre-run check and start the run
Start New Run on Instrument
*If using Standalone Mode: enter the
name of the run and specify the same
parameters as the original run
32 COMPANY CONFIDENTIAL – INTERNAL USE ONLY
Troubleshooting Files
33 COMPANY CONFIDENTIAL – INTERNAL USE ONLY
An Illumina Technical Support representative
might request copies of run-specific files for
troubleshooting purposes
– These files can be found from the run
folder
Troubleshooting files
Run folder in Windows
Run folder in BaseSpace
Links to BaseSpace runs
are also very useful for
Technical Support
Troubleshooting
34 COMPANY CONFIDENTIAL – INTERNAL USE ONLY
Run Information and Run Parameters
Location: root folder
Run Info.xml description: contains name of the run, number of cycles in the run, number of cycles in each read, whether the read is an indexed read, and the number of swatch and tiles on the flow cell
Runparameters.xml description: contains a summary of run parameters and information about run components, such as the RFID of the flow cell and reagents associated with the run
Run information (RunInfo.xml)
35 COMPANY CONFIDENTIAL – INTERNAL USE ONLY
RTA Configuration File
Location: root folder
Description: contains the RTA configuration settings for the run. The RTAConfiguration.xml file is created at the beginning of the run
RTA configuration file
36 COMPANY CONFIDENTIAL – INTERNAL USE ONLY
InterOp files
Location: root folder
Description: Binary reporting files used for Sequencing Analysis Viewer.
InterOp files are updated throughout the run.
InterOp files
37 COMPANY CONFIDENTIAL – INTERNAL USE ONLY
RTA Logs
Location: RTA Logs
Description: Log of RTA errors. Error log files are updated whenever an error occurs.
Location: RTA Logs
Description: Log of all RTA events. Global log files are updated throughout the run.
Error log files
Global log files
Location: RTA Logs
Description: Logs RTA processing events. Lane log files are updated throughout the run.
Lane log files
RTA Logs
38 COMPANY CONFIDENTIAL – INTERNAL USE ONLY
Some pre-run check errors are easily resolvable by the user
– See the Troubleshooting section of the NextSeq System User Guide for more detail
System Checks are typically performed during a Live Help session with
assistance from an Illumina Technical Support representative
Rehybridization of read 1 sequencing primer can be performed as potential run
rescue
Troubleshooting files found in the run folder help Illumina representatives
diagnose sequencing run issues more effectively
Session Summary
39
Questions?