Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________ _____________Basic Operation of a TRACE MS Plus for Engineers ___________________ 1 1. Basic Operation of the TRACE MS PLUS for Engineers Introduction This document assumes the reader already has a working familiarity with TRACE MS/Voyager/MD800 control via the Xcalibur data system. TRACE MS Plus requires Xcalibur 1.3. Configuring TRACE MS Plus Run the Xcalibur Instrument Configurator. Select Instrument Configurator from the Windows Start menu.

Upload: gc-power

Post on 11-Apr-2015

42 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________1

1. Basic Operation of the TRACE MS PLUS for Engineers

Introduction This document assumes the reader already has a working familiarity with TRACE MS/Voyager/MD800 control via the Xcalibur data system. TRACE MS Plus requires Xcalibur 1.3.

Configuring TRACE MS Plus Run the Xcalibur Instrument Configurator.

Select Instrument Configurator from the Windows Start menu.

Page 2: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

2 __________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Double-click the TRACE MS Plus button that appears in the instruments list on the right-hand-side of the dialog.

At time of writing, there were no configurational settings for TRACE MS Plus. It can be anticipated that there will be some adjustable parameters in a later release version, to enable such options as 800amu (for ToxLab) and CI capability.

Launching the Server The Server can be launched in 2 ways:

1. Running the Tuning application for TRACE MS Plus.

2. Launching Xcalibur.

If Xcalibur is launched, it is assumed that the instrument configuration has declared the TRACE MS Plus as selected.

Page 3: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________3

When the Server is running, there will be an icon in the Windows Start menu, as shown below.

The Icon changes colour depending on the state of the MS.

Yellow indicates Pumped Down and in Standby.

Green Indicates the MS is in Operate.

Red indicates the system is Vented, and therefore in Standby.

Perform a right-click on the icon to open the Server menu.

It is possible to access the Manual Tune dialog from here if it is not already open.

Other operations that are possible from this menu are:

Automated Tuning and/or Automated Mass Axis Calibration

Manual Mass Calibration, via the Instrument Tune and Calibration… menu item

Reconfiguration of the MS Inlet (GC, DIP or DCI)

Venting/Pump-down operations

Page 4: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

4 __________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Tune Dialog

Opening the Tune dialog

Tune can be launched from its Shortcut on the Windows Desktop, or if not visible, through the Start Menu, as shown below.

Page 5: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________5

Default Tune dialog

The Tune dialog represents a radical departure from earlier generation user interfaces. When it first launches the display is seen to be organised into two regions.

Graphical Display

The top of the dialog is occupied by one or more mass range graphical displays - dependent on how many have been specified/activated – and each will have annotation indicating Absolute Intensity, Centroided Mass Measurement and Display Gain (Zoom) of the individual mass selection.

Tabular Display

The bottom region of the dialog is composed of a table of Scan Functions. Each entry (row) describes a mass with defined mass span, scan time,

Page 6: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

6 __________________ Basic Operation of a TRACE MS Plus for Engineers_____________

detector voltage setting and peak format and each may be selected in the graphical region above.

In the example caption, it will be evident that a level of customisation is possible. Each Scan Function can have an alias name applied to it, for familiarity. The total number of rows visible on-screen is defined in a configurational dialog that opens from the View | Options… menu item. It will be discussed later.

The example shows 7 Scan Functions that have been compiled and together these define the most routine interactive procedures associated with maintenance of a MS.

FS1 through FS3 show individual ions from the spectrum of PFTBA (Heptacosa); these would be used to review sensitivity and mass resolution for routine analysis.

FS4 has been designed to show a wide mass range and therefore allowing inspection of the instrument ion background; this would be useful to monitor contaminants such as column and septum bleeds, hydrocarbon background from dirty samples and plasticizer ions that could potentially be introduced during column installations and/or source cleaning.

FS5 through FS7 have been set up to monitor for potential air leaks, referenced against Helium carrier gas.

Collectively, the table of Scan Functions replaces the earlier technique of swapping saved parameter display files to accomplish these fundamental tasks.

The table also permits time windows to be applied to any or all Scan Functions. Although the default setting is zero, a time range, say 0.00min to 1.00min would enable an acquisition to be performed from the tune dialog, independent of Xcalibur. This is useful for performing diagnostic manual mass calibrations very rapidly.

Page 7: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________7

Displaying Tune Parameters The Windows slider control marked Show Per Method Parameters, at top right of the dialog will reveal the tune parameters, as shown below.

These parameters are global to all Scan Functions.

Method Editor dialog

The Method Editor dialog for TRACE MS Plus is launched from Xcalibur’s Instrument Setup dialog

Page 8: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

8 __________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Pumping Down and Venting This is achieved by right-clicking the Server icon at the right hand side of the Windows taskbar.

Pumping Down

When the instrument is vented, the Server icon will be red (Vented).

To pump down, perform a right-click on the icon to display its menu.

Select the Pump menu item to start the pump down procedure. The Server icon will flash yellow until high vacuum has been achieved AND turbomolecular pumps achieve adequate frequency of rotation, where after it will stop flashing and adopt a continuous yellow appearance.

Page 9: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________9

Venting

When the instrument is pumped down, the Server icon will be yellow (Operate OFF) or green (Operate ON). The right-click operation will reveal the following menu:

Selecting the Vent item will display a confirmation prompt, as follows:

Select Yes to Vent the MS, if this is required. Venting will be immediate, with the deactivation of the Vent Valve solenoid attached to the front (or only) turbomolecular pump.

Page 10: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

10_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Tuning and Calibrating

Autotuning

Select the Instrument Tune and Calibration menu item from the Server icon.

The following dialog will be displayed:

Select the option for Full System Autotune.

A status dialog will be generated showing a log of events that will eventually lead to a fully tuned and calibrated instrument.

Page 11: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 11

The Tune dialog can be seen to update with the Autotune algorithm.

Page 12: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

12_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Page 13: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 13

Ultimately, the log of events can be copied to the clipboard for pasting into a document. This may be useful for diagnostics, if a problem is suspected, or as a teaching aid for learning manual tuning.

Eventually, the system will be tuned and calibration will follow:

Page 14: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

14_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

The Tune dialog will show a full scan across the reference ions of the internal tuning compound, PFTBA (FC43).

Page 15: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 15

And here is the final outcome:

Print the report, if required, otherwise hit the Finish button.

The Reference gas will be turned OFF automatically and Operate will be set back to its previous state.

Page 16: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

16_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Tuning in Normal Mode Select the Manual Tune menu item to launch the Tune dialog.

This dialog shows the Standard Tune view. There are no Tune parameters to adjust. It is assumed that Autotune is used to set up the instrument.

Page 17: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 17

Selecting Ionisation Mode Different ionisation modes are selectable from the combo bow as shown below:

Page 18: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

18_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Tuning in Engineers Mode To turn ON Engineers Mode, open the Options dialog from the View menu:

The Options dialog will appear as shown:

Page 19: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 19

Select the radio button marked Engineer to reveal all parameters.

When the Tune page is re-opened you will see additional Per Method Parameters.

It is possible to view many readbacks, power supply rails and perform various diagnostic tests.

Page 20: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

20_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Performing Mass Calibration This can be achieved from the menu attached to the Server icon.

Select the Standard Mass Scale Calibration option and follow the instructions:

Page 21: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 21

The dialog will show status information as the instrument is calibrated.

Page 22: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

22_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

The Tune dialog will show a scanning acquisition occurring.

Ultimately the process will terminate:

There will be an option to print off a report prior to closing the dialog.

Page 23: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 23

Manual Mass Calibration Only necessary for diagnostic purposes. Uses an acquired file, and a selected calibration reference file:

Follow the instructions in the wizard, as shown below:

The Raw File must contain spectra acquired over an adequate mass range with satisfactory mass resolution.

Page 24: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

24_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Saving Tune Parameters to a named file Very straightforward, merely open the File | Save As… menu item and name your tune file.

Page 25: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 25

Setting up MS-only acquisition

Acquiring from Tune Dialog

Need a scan function defining an adequate mass range, as in example below:

Page 26: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

26_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Specify a file name and location in the options dialog:

Use the File | Start Acquiring menu item to commence:

Page 27: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 27

You will be able to monitor the elapsed time.

Acquiring from Xcalibur This is same as with previous systems. Define parameters in Instrument Method Editor and run from a Sample Sequence in Home Page.

Performing GC/MS acquisitions

Solvent Delays

There is no longer a limit on the positioning or number of solvent delays within an acquisition.

Page 28: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

28_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Scanning acquisition This is the default mode of acquisition. To add a scan function, literally hit the Add Full command.

Page 29: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 29

SIM acquisition There are no constraints on how many ions get acquired in a data file. In reality, there is an upper limit of 511 individual ions which may be grouped into any number of SIM groups.

The start and end times can be defined in the table or by moving the graphical bars with the mouse.

At the time of writing, the grouping control was unfinished for TRACE MS Plus.

Mixed Scan and SIM acquisition TRACE MS Plus can be used analytically to perform scanning acquisitions as well as SIM. There is complete freedom of overlap, simultaneous or discreet groupings.

Page 30: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

30_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Use the Add Full and Add SIM HTML links to add Scan Functions:

Now use the Preview | All Scans to view all Scan Functions.

Page 31: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 31

They may be edited as needed. There are no constraint on start and end times, and Scan Functions can overlap.

If the green scan function is clicked-and-dragged, then its position and times can be varied with the mouse.

Toggle between Scan and SIM functions with the slide controls:

Page 32: Trace MS Plus Basic Operation for Engineers

Basic Operation of the TRACE MS PLUS for Engineers _____________________________________________________________________________

32_________________ Basic Operation of a TRACE MS Plus for Engineers_____________

When you click on any graphical scan function, its properties are shown in a pop-up window: