total pressure measurementrtvide.cnrs.fr/img/pdf/capteurs_vide_secondaire_et_uhv.pdf · total...
TRANSCRIPT
A P A S S I O N F O R P E R F E C T I O N
Total Pressure Measurement
Prepared by: Andreas Schopphoff
Date: 2012-10
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 2
Content
� Basics of Total Pressure Measurement
� Measurement Ranges
� Gas Type Dependent Measurement
� Medium Vacuum Measurement
� Pirani
� Ultra High Vacuum Measurement
� Hot Cathode Gauges
� Cold Cathode Gauges
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 3
Overview
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 4
Measurement Ranges
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 5
Measurement independent of gas type
� Pressure defined
� p=F/A
� F=force
� A=Area
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 6
Commonly Used Units in Vacuum Technology
� SI Pascal Pa=N/m2
� Units used in vacuum technology
� 1 mbar = 100 Pa
� 1 Torr = 133,322 Pa
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 7
Measurement dependent on gas type
� Pressure defined
� p = n KT
� n = particle number density
� K = Boltzmann’s constant
� T = thermodynamic temperature
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 8
Measurement Ranges
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 9
Content
� Basics of Total Pressure Measurement
� Measurement Ranges
� Gas Type Dependent Measurement
� Medium Vacuum Measurement
� Pirani
� Ultra High Vacuum Measurement
� Hot Cathode Gauges
� Cold Cathode Gauges
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 10
Pirani Gauge, Principle
� Heated wire (tungsten) 100 to 130°C
� Heat loss due to thermal transfer of gas
� Constant heating
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 11
Characteristic Curve Pirani Gauge
0,0001
0,001
0,01
0,1
1
10
100
1000
0,0001 0,001 0,01 0,1 1 10 100 1000
True pressure (mbar)
Indic
ate
d p
ressure
(m
bar)
Stickstoff
Luft
Wasserstoff
Helium
Argon
Kohlendioxyd
Nitrogen
Air
Hydrogen
Helium
Argon
Carbondioxide
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 12
Summary: Pirani
� Measurement range: E-4 to 1000 mbar
� Accuracy + - 15 %
� Easy to use
� Can be mounted in any orientation
� Gas type dependent
� Easy to clean
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 13
Content
� Basics of Total Pressure Measurement
� Measurement Ranges
� Gas Type Dependent Measurement
� Medium Vacuum Measurement
� Pirani
� Ultra High Vacuum Measurement
� Hot Cathode Gauges
� Cold Cathode Gauges
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 14
Hot Cathode, Principle
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 15
Differential Ionization
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 16
Bayard Alpert: X-Ray Effect
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 17
Ion Source: Trajectories
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 18
Bayard Alpert: ESD Ions
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 19
Extractor Ion Gauge
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 20
Summary: Hot Cathode Gauge
� Measurement Range: E-12 to E-1mbar
� High Accuracy + - 10 to 15 %
� Sensitive to conatmination
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 21
Cold Cathode Sensor: Principle
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 22
Cold Cathode Sensor
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 23
Magnetron
� According to Redhead
� Additional electrodes
� p = E-13 mbar
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 24
Summary: Cold Cathode Gauge
� Measurement Range E-11 to E-2 mbar
� Accuracy + - 30%
� Rugged
� Easy to clean
� Not useful with sputter applications
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 25
The Right Gauge for the Application
� Measurement Range?
� Accuracy?
� Gases?
� Environment?
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 26
Gauges with external Controller
GaugeElektronic TPG 300
Up to 500m
A P A S S I O N F O R P E R F E C T I O N
© Pfeiffer Vacuum 2012 • Market Management R&D • Andreas Schopphoff • 2012-06-21 ■ 27
End
� Thank you for your attention!