torque control device - straumannnordic.straumann.com/content/dam/internet/straumann_nordic... ·...
TRANSCRIPT
THIS IS GOING TO BE TORQUED:
TORQUE NEWTON CENTIMETER (NCM)
CREATECH
Bridge, implant level, straight channel
35 NCM
Bridge, abutment level, straight channel
15 NCM
Bridge, implant or abutment level, angulated channel 15 NCM
ABUTMENT GROUP TORQUECROWN/BRIDGE ON THE ABUTMENT IS TORQUED TO:
EXCEPTION
SynOcta e.g., 048;601,605,612-618 35 NCM 15 NCM Cares ZrO2 RN comes with a synocta 1,5 (048.601). In this case both
the abutment and the zirconia crown shall be torqued to 35 NCM.
Variobase family
35 NCM n/a Does not include the Variobase coping for SRA. It shall be torqued to 15 NCM. It utilizes an occlusal screw and not a basal screw.
SRA
35 NCM 15 NCM
Caresall materials 35 NCM n/a Cares ZrO2 RN comes with a synocta 1,5 (048.601). In this case both
the abutment and the zirconia crown shall be torqued to 35 NCM.
AnatomicalZrO2 and Ti 35 NCM n/a
Torque control device
Using the torque control device
Torque control device.Being used together with the SCS screwdriver in order to torque abutments and restorations from Straumann®.
Storing the torque control deviceThe torque control must be disassembled for cleaning and sterilization. It shall be stored disassembled and only assembled before use.
Using the torque control devicePlace a finger on the SCS screwdriver for stability. Push/pull the drop shaped handle only, until it has reached the correct torque marking. See page 1 for details. Observe! The arrow shall always point clockwise.
Abutment The abutment is inserted into the implant and tightened by hand first. With the use of the torque control device you torque the abutment to the correct torque with one uninterrupted movement. The screw channel can be sealed with e.g., guttapercha/silicontube and composite.
Restoration.The restoration is inserted on the abutment/into the implant and tightened by hand first. With the use of the torque control device you torque the it to the correct torque with one uninterrupted movement.The screw channel can be sealed with e.g., guttapercha/silicontube and composite.
The torque in this guide is only valid for Straumann® Original or Createch prosthetics on Straumann® implants. Regarding prosthetic components or restorations which is not manufactured by Straumann®. Please contact the supplier/manufacturer for the correct torque it is the abutment/screw that decides the torque not the implant.