To NeSSi or Not to NeSSi:A Fast and Micro GC Experience Review
IFPAC® Annual Meeting January 26, 2016
Grand H
NeSSi
Arlington, VA (Washington DC), U.S.A
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 2
Abstract
• By now there is substantial experience using NeSSi systems with fast and micro gas chromatography. All are successful but some are more successful than others. On the surface, the modular fluid delivery sample systems are a perfect fit for fast and micro gas chromatography. However, just because one can use NeSSi doesn’t necessarily mean one should use NeSSi in every case. Technical issues such as temperature, pressure and especially flow rate and even aesthetic issues must be considered carefully before plowing ahead. This presentation will review the current status of 3 installations and discuss the lessons learned.
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 3
What Is A Fast & Micro Gas Chromatograph?
• … a GC having internal dimensions measured in microns not millimeters, retention times in seconds, not minutes and none of the limitations of silicon chip based GCs• One tenth the size of conventional GCs – 30 lbs vs 300
lbs
• One tenth the power of conventional GCs – 300 watts vs 3000 watts
• Analysis times 10 – 50 times faster than conventional GCs – Fixed gases and Hydrocarbons up to C50
• Using resistively heated steel capillary column modules eliminating air bath ovens
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 4
… MUST have…all the major attributes of conventional GCs
• Split/splitless injector with a variety of sampling options• Manual syringes as well as
syringe based autosamplers
• Fully automated gas & liquid sample valves
• Range of sample pretreatment options: traps, headspace, pyrolysis and…
• Process sample systems for automated process analysis &
control… even NeSSI
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 5
… MUST have…all the major attributes of conventional GCs
• Programmed temperature and/or isothermal columns• Up to 2 modules, each having 1
and up to 8 meters of stainless steel capillary GC columns
• Independently controlled from 5oC above ambient to the limit of the column type or 400oC whichever is lower
• Can include one column switching valve
NO air bath oven!
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 6
… MUST have…all the major attributes of conventional GCs
• Detectors• Thermal Conductivity
• Flame Ionization
• Flame Photometric
• Dielectric Barrier Discharge• Electron Capture for halogens
• Photo Ionization for aromatics & olefins
• Helium Ionization for everything
• Even… High Pressure Mass Spec Detector
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 7
… AND has…modern software and communications technology
• …to match the analytical and automation requirements• Sensor rich GC system &
firmware
• Chromperfect Ethernet/PC based data system
• Application modules to fit• Process analyzer HMI
• TC/PIP communications
• MODBUS via Ethernet
• BTU calculations
• Links to other Windows applications
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 8
… AND has…modern software and communications technology
• …to match the analytical and automation requirements• Simulated Distillation
applications
• Infometrix chemometricapplications• LineUp for retention time
alignment
• InStep for runtime chemometricmodels
• PCA
• PLS & other Pirouette apps
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 9
Falcon Calidus Intellectual Property
Trans-Configurable Modular Chromatographic
Assembly US Patent 8336366
Fast Micro Gas Chromatograph System
US Patent 8414832
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 10
*
* Dielectric Barrier Discharge Detector
Modular Calidus GC Single Injections
• Models are made of modules
• Currently there are 7 Calidus models are available• 101, single column & detector
• 201, dual serial columns, single detector
• 301, dual parallel columns & detectors
• CS, dual columns & detectors
• Heartcut
• Trap/bypass
• Backflush
• GS & GS-T for environmental
• MT for transportable fraud detection
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 11
Modules Make InstrumentsInstruments Make Analyzers
Analyzers plus Sample Systems Make Solutions… One Solution
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 12
Network
Micro-PROCESS
Photometer
Micro LABMass
Spectometer
Micro LABOptical
Spectrometer
Micro LABPhotometer
Micro-PROCESS
GC
Micro-PROCESS
MassSpectrometer
Micro LABGC
At-LinePackaging
Pro
cess
Co
ntr
ol N
etw
ork
Lab
Inst
rum
en
t N
etw
ork Transportable
Packaging
Network
Micro-PROCESSOpticalSpectrometer
ERP
Copyright Falcon Analytical 2015
Old School
New Thinking…CONVERGENCE!
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 13
FalconFAST Analytical Solutions and the Promise of NeSSi
• Easy to use: Sophisticated instrument/analyzer pairs employed with intelligence, agility and transparency to the user - thereby unifying the measurement results from the laboratory, at-line, on-line or in the field
• Ultrafast gas chromatography instrumentation with applications technology designed to perform the job combined with
• NeSSI sampling,
• Advanced data processing and
• Enlightened interpretation
• “Customers need and want the answers…Not the hassles associated with analytical chemistry!”
Laboratories
Process
Field
LoadingDock
InfometrixChemo-metrics
ParkerNeSSI
Sampling
FalconFast GC
ChromperfectData System
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 14
Modularity… Modularity… Modularity
• Modularity from the analytical elements to the entirety of the system• Calidus
• Process Enclosure
• NeSSI Modular Fluid Delivery Systems
• Even software
• Calidus firmware
• Chromperfect Engine
• Chromperfect Apps
• Pirouette Engine
• LineUp
• InStep
• Pirouette models
• From sample probe to the DCS connection… modularity makes jobs easier
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 15
Just because one can use NeSSi doesn’t necessarily mean one should use NeSSi.
• Sample systems STILL need the right fundamental engineering design must be done• Temperature, pressure and flow rate
• Preconditioning for particulate or condensate or, or, or must be properly done
• Transport distance and mechanism
• Sample stability and materials in contact
• And all the others
•What follows are some examples in FalconFast experience.
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 16
The Right Stuff – full scale plant system
• 3 Calidus Process Analyzers
• 3 Single Stream Parker Intraflow systems with autocal
• Chromperfect operational controls and data system & communications
• Infometrix retention time alignment for rock solid stability
• 3 process streams: dimer, trimer & tetramers of propylene
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 17
What about Stability? Single System Exampleat the End of the 1st Quarter
Q 1 Visit: part of last run showing 60 to 140 seconds
101 seconds
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 18
Single System Exampleat the End of the 1st and 2nd Quarter Overlaid
Q 1 & 2 Visits Overlaid: part of last run showing 60 to 140 seconds
101 seconds
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 19
Single System Exampleat the End of the 1st,2nd and 3rd Quarter Overlaid
Q 1, 2 & 3 Visits Overlaid: part of last run showing 60 to 140 seconds
101 seconds
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 20
Single System Example after a Whole Year!End of the 1st,2nd,3rd and 4th Quarter Visits Overlaid
Q 1, 2, 3 & 4 Visits Overlaid: part of last run showing 60 to 140 seconds
Of course operating processes show compositional variance. However, theretention times don’t!
4th chromatogram is after > 175,000 runs with ZERO downtime, no sample valve rotor nor column change nor detector maintenance…
ZIP
One year RT drift using
Infometrix LineUP… ZERO
Peak to peak RAW (without LineUP) drift 0.6 seconds for a WHOLE YEAR!!!
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 21
Works OK… after rework
• Pilot plant operations• The measurement goal: help find
optimum processing conditions• Somewhat reactive samples
• Analysis is for ppm level impurities in product
• Process development campaigns mean different compositions at different times
• The issues: • Very small process volume
• Sample system rework to return analyzer loop to the process WITHOUT contamination (disturbing the process)
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 22
Works Just Fine… but should it have been done?
• 8 Small Bioreactors• Off gas analysis for the
product concentration
• Indicates performance of the microbes and nutrient and air feed rate requirements
• Heat tracing absolutely required to get the C10
product through the tubing to the sample system & GC sample valve
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 23
Why the Second Guessing? LOST ORDERS!
• Small reactors• Probably should have used a
rotary selector valve
• Small diameter tubing would have fit in a single heated bundle
• Biological culture not really in tune with our industrial culture
• It all works great BUT… it isn’t really a good fit in this lab…• Well… they said they would
ultimately put it on very large fermentors
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 24
The Next Big Thing… not NeSSI
• Highly reactive monomer production
• Sample must be kept above 200 C… or polymer fills the tubing and sets like concrete!
• Sample system components become “a disposable item”
• We will see how this works out soon
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 25
IFPAC 2016, NeSSi
Have you been listening?
No one asks Falcon to do C1/C2 ratios.
26
Old SchoolPropylene Purity Application Single Oven
• Ethylene purity would be very similar to this application.
SBFV SBFV SEL
SBFV SBFV
FID
TCD
methanizer
Ref. Low ppm H2
Mea. %
C3H8
ppm C0,
C02
ppm
MeOH
27
New School, Ethylene Purity Application CALIDUD CS, Dual Column, TCD & FID
Column 1
MXT-U-BondColumn 2
MXT-Molesieve 5A
FID TCD
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 28
All Leads to…
Data
Equivalency
[email protected] Tel: (304) 647-5860Copyright Falcon Analytical 2/24/2016
433 AEI Drive – Lewisburg, WV 24901 29
Industrial Implications (leaving you with food for thought)
Analytical Results Enable
• Lab to Process equivalency
• Using same instrument, same method eliminates the variables with the exception of sampling
• Possibilities for online product certificates of analysis
• Plant to plant equivalency
• Improves corporate product consistency, quality and tighter process control
• Inter plant process trouble shooting
• Work can now be focused on process variances by elimination of analytical variances
System Modularity Enables• Significant maintenance reductions
• Substitute whole instruments as a module, eliminating field repairs and costs
• Repairs/refurbishment in the lab or return to factory… like a premium car
• Packaging appropriate for the installation site
• Purpose built for the application
• Deploy only what is needed for the job
• Potential for shelterless installations
• Lower costs
• Capital costs
• Operational costs
• Long term maintenance costs