tips & tricks gpc/sec: calibration using broad · pdf filetips & tricks gpc/sec:...

6
Tips & Tricks GPC/SEC: Calibration Using Broad Standards For polymeric materials where no narrowly distributed molar mass standards are available, reliable and true molar masses can be obtained using well characterized samples of broad molar mass distribution. This approach does not require additional instrumentation and offers an alternative to using molar mass sensitive detectors such as on-line light scattering or on-line viscometry detectors. Wolfgang Radke and Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany. Gel permeation chromatography/ size-exclusion chromatography (GPC/SEC) is an easy and robust method to determine molar masses and molar mass distributions (MMD) of macromolecules. The separation is based on the size of macromolecules in solution — the larger the macromolecule, the earlier it elutes from the column. The primary information generated by GPC/SEC is the chromatogram showing the detector signal as a function of elution volume. To derive the molar mass distribution, the chromatogram has to be converted using a calibration curve that relates elution volume to the molar mass of the eluting polymer. Calibration curves are usually established using narrowly distributed polymer standards; however, chemically different polymers of the same molar mass often exhibit very different sizes in solution. Correct molar masses are therefore only obtained if the chemical structure of the calibrant matches the structure of the analyte. Unfortunately, for a large number of important polymers (for example, polyamides, polyesters, or polyolefins), narrow standards are not commercially available and alternative methods are required. The use of a molar mass sensitive detector is one option but it requires additional expensive instrumentation, more precise and accurate sample preparation, more time, and a greater level of knowledge about data evaluation. These limitations are especially crucial for quality control laboratories because fast, robust, and easy-to use calibration methods are required. A suitable, cost-effective solution is the use of broadly distributed standards. Photo Credit: CGinspiration/Getty Images 10 Pittcon 2015 Preview 2 News 7 Tips & Tricks: GPC/SEC 10 Q&A: Haselberg 16 7 10 0 The LCGC Blog 20 CHROMacademy 23 Training & Events 24 Staff 25 23 23 24 2

Upload: ngodang

Post on 31-Mar-2018

243 views

Category:

Documents


3 download

TRANSCRIPT

Tips & Tricks GPC/SEC: Calibration Using Broad Standards

For polymeric materials where no narrowly distributed molar mass standards are available, reliable and true molar masses

can be obtained using well characterized samples of broad molar mass distribution. This approach does not require additional

instrumentation and offers an alternative to using molar mass sensitive detectors such as on-line light scattering or on-line

viscometry detectors.

Wolfgang Radke and Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany.

Gel permeation chromatography/

size-exclusion chromatography (GPC/SEC)

is an easy and robust method to determine

molar masses and molar mass distributions

(MMD) of macromolecules. The separation

is based on the size of macromolecules in

solution — the larger the macromolecule,

the earlier it elutes from the column.

The primary information generated by

GPC/SEC is the chromatogram showing

the detector signal as a function of

elution volume. To derive the molar mass

distribution, the chromatogram has to be

converted using a calibration curve that

relates elution volume to the molar mass

of the eluting polymer. Calibration curves

are usually established using narrowly

distributed polymer standards; however,

chemically different polymers of the same

molar mass often exhibit very different

sizes in solution. Correct molar masses are

therefore only obtained if the chemical

structure of the calibrant matches the

structure of the analyte.

Unfortunately, for a large number

of important polymers (for example,

polyamides, polyesters, or polyolefins),

narrow standards are not commercially

available and alternative methods are

required. The use of a molar mass

sensitive detector is one option

but it requires additional expensive

instrumentation, more precise and

accurate sample preparation, more time,

and a greater level of knowledge about

data evaluation.

These limitations are especially crucial

for quality control laboratories because

fast, robust, and easy-to use calibration

methods are required. A suitable,

cost-effective solution is the use of broadly

distributed standards. Ph

oto

Cre

dit

: C

Gin

spir

ati

on

/Ge

tty I

ma

ge

s

10

Pittcon 2015 Preview2 News7 Tips & Tricks: GPC/SEC10 Q&A: Haselberg167 100The LCGC Blog20 CHROMacademy23 Training & Events24 Staff252323 242

The Column www.chromatographyonline.com

Calibration Using Broad Samples —

How Does It Work?

For a narrowly distributed polymer, the

maximum peak of the chromatogram

can be easily identifi ed and assigned. In

addition, we can assume Mw ≈ Mn ≈ Mp.

This is not the case for a broadly distributed

sample. Here, the maximum can be

diffi cult to precisely identify because it can

depend on the resolution of the separation

columns. It is therefore not recommended

to establish the calibration curve by simply

plotting the molar masses versus the elution

volume at peak maximum.

The generally accepted concept of

universal calibration1 in GPC/SEC results

in the following relationship for the

calibrations curves of the calibrant and

the analyte:

[1]log M analyte log M calibrantA �+= B

106

Elution volume (mL)

Mo

lar

ma

ss (

g/m

ol) 105

104

103

10 15 20 25

Figure 1: The effect of parameters A and B on the calibration curve and subsequently MMD. Solid turquoise: Original (base) calibration curve and MMD (left) derived therefrom. Blue: Calibration curve and MMD derived from turquoise curve by changing parameter B (slope). Orange: Calibration curve and MWD derived from blue curve by changing parameter A (vertical shift).

Tips & Tricks: GPC/SEC

11

Pittcon 2015 Preview2 News7 Tips & Tricks: GPC/SEC10 Q&A: Haselberg167 100The LCGC Blog20 CHROMacademy23 Training & Events24 Staff252323 242

LCGC Web VideosTUNE IN TODAY!

ChromTube is LCGC’s platform

for informative videos from

premier suppliers in the

separation science field. Watch

videos in your areas of interest:

Mass Spec, Sample Prep, GC,

and HPLC/UHPLC .

LCGC TV’s editorial videos bring

you insights from leading

analytical scientists, updates on

new research and methods as

well as best practices and

practical advice you can apply

in your lab.

WATCH NOW!LCGC TV: www.chromatographyonline.com/lcgctv

ChromTube: www.chromatographyonline.com/chromtube

The Column www.chromatographyonline.com

This is based on the assumption that

both calibrant and analyte elute without

interacting with the stationary phase.

The parameter A shifts the calibration

curve vertically relative to the calibration

curve, while a change in B varies the

slope of the calibration curve (Figure 1).

The parameter A and B are related to the

Mark–Houwink parameters of the analyte

and calibrant in the respective solvent.2–5

Because of the effect of parameters A and

B on the calibration curve, the MMD and

consequently the molar mass averages

calculated for a given chromatogram

change as a function of the parameters

as well.

This is illustrated in Table 1 where two

chromatograms of broadly distributed

poly(lactic acid) (PLA) samples were

evaluated using a polystyrene calibration

curve (base calibration) and different sets

of A and B. Application of the polystyrene

calibration curve overestimates the true

molar masses by approximately a factor

of 3. Changing the parameters brings

the molar mass closer to the true value;

however, both parameters need to be

changed to get reliable results for different

samples. For the optimized parameter set

A and B, given in the last line, a good

agreement with the reference values from

light scattering is observed.

Table 1 also explains how the parameters

A and B can be determined from the

chromatograms and known average molar

masses of broadly distributed standards. To

obtain the A and B, they are systematically

varied, until the best agreement

Table 1: Effect of variation of parameters A and B on the molar masses derived from a polystyrene calibration for two samples of polylactic acid (PLA).

PLA-Sample 1Mw [g/mol]

PLA-Sample 2Mw [g/mol]

Reference value from light scattering 37,000 500,000

Polystyrene calibration - (A = B = 1) 82,300 1,247,000

A = 0.5, B = 1 35,400 580,000

A = 0.5; B = 8 3790 355,000

Optimized values from software

A = 0.717; B = 0.95736,950 500,700

Tips & Tricks: GPC/SEC

12

Pittcon 2015 Preview2 News7 Tips & Tricks: GPC/SEC10 Q&A: Haselberg167 100The LCGC Blog20 CHROMacademy23 Training & Events24 Staff252323 242

5

3rdAnnual

CHROMATOGRAPHY COMMUNITY MIXERJoin chromatographers from around the

world as we gather in the spirit of

fellowship and collaboration!

EVENT DETAILS<��&.02*"-�40�4)&�-"4&��&02(&3��5*0$)0/

<��*6&�.53*$�#9��2"..9�/0.*/"4&%�=�04����2"33��"/%>�

<�=�&$0/%��*/&>�+"::�1"2"%&

<��"33&%�)023�%>0&562&3�"/%�$2&0-&�$5*3*/&�#5''&4

<��1&/�#"2�

<��0..&.02"4*6&�(*'4�#"(

WHEN�5&3%"9���"2$)����������'20.� ���� ��1.�

WHERE�&/&2"4*0/3��"--�������/%2&7��*((*/3��2���&7��2-&"/3����������

TICKETS REQUIRED FOR ENTRY *3*4�"/9�0'�052�12&.*&2�310/3023�"4��*44$0/�40�$0--&$4�9052�4*$,&4 �

FOR MORE INFORMATION, CONTACT:

�0/"4)"/��%&-."/�;��&- ���������������."*- ��*44$0/.*8&2�(."*-�$0.

ORGANIZING SPONSORS:

��� ��!����������������������������������

MEDIA SPONSOR:

The Column www.chromatographyonline.com

between the known molar masses and

the ones calculated from the respective

chromatograms, the base calibration, and

the parameters is obtained.

Fortunately this variation can be

automated with GPC/SEC software

packages. With the obtained parameters,

a suitable calibration curve for unknown

samples of the same structure as the

analyte can be easily derived from the

base calibration.

In contrast to other approaches for broad

standards calibration, the above approach

benefi ts from the fact that the base

calibration reveals the pore size distribution

of the column packing and therefore

the true shape of the calibration curve.

In particular the method preserves the

non-linear shape, the exclusion volume, and

the separation limit of the base calibration

curve. An additional advantage is that more

than one sample can be used to create the

Figure 2: Example for a base calibration for two samples of polylactic acid (PLA) with known Mw from light scattering. To increase the accuracy and the molar mass range, up to eight different samples can be used.

Tips & Tricks: GPC/SEC

13

Pittcon 2015 Preview2 News7 Tips & Tricks: GPC/SEC10 Q&A: Haselberg167 100The LCGC Blog20 CHROMacademy23 Training & Events24 Staff252323 242

The Column www.chromatographyonline.com

calibration curve. This increases accuracy

and allows a wider molar mass range

to be covered.

How to Establish a Calibration Using

Broad Standards

Establishing a calibration curve by the

above procedure requires:

t� A base curve established using narrowly

distributed standards of arbitrary

chemical structure, such as polystyrene

for many organic solvents or pullulan of

poly(ethylene oxide)/poly(ethylene glycol)

(PEO/PEG) for aqueous applications.

t� At least one broadly distributed sample

with known Mw and Mn values of

the same chemical structure as the

analytes to be characterized. If two

or more broadly distributed samples

are used one molar mass average

Figure 3: The blue calibration curve is the pullulan base calibration curve. If this curve is used for evaluation molar masses are overestimated. The red curve is the resulting calibration curve calculated using the optimized A and B values from the samples with the broad molar mass distribution.

Tips & Tricks: GPC/SEC

14

Pittcon 2015 Preview2 News7 Tips & Tricks: GPC/SEC10 Q&A: Haselberg167 100The LCGC Blog20 CHROMacademy23 Training & Events24 Staff252323 242

Lifetime Achievement and Emerging Leader in Chromatography Awards

8th Annual LCGC Awards at Pittcon 2015

Visit LCGC, LCGC’s CHROMacademy and Spectroscopy at Booth #2401

For more information, please visit www.chromatographyonline.com

The winner of the 2015 Lifetime Achievement award

is Jack Kirkland.

The Lifetime Achievement award

honors a seasoned professional for

a lifetime of contributions to the advancement of

chromatographic techniques and applications.

The winner of the 2015 Emerging Leader Award is Caroline West.

This award recognizes the achievements and aspirations of a talented young separation scientist who has made strides early in his or her career toward the advancement of chromatographic techniques and applications.

JOIN LCGC AT A NEW PITTCON AWARDS SYMPOSIUMMonday, MARCH 9, 1:30PM – 4:45PM

The Column www.chromatographyonline.com

(such as Mw or Mn) per sample is

suffi cient.

The procedure then is straightforward:

t� Run the narrowly and the broadly

distributed standards on your GPC/SEC

system.

t� Establish the base calibration using the

molar masses (Mp) and peak maxima of

the narrowly distributed calibrants.6

t� Use a software tool to shift the

calibration curve and adjust the slope

so that the calibration provides correct

results for the broadly distributed

samples. This means that the software

internally determines the optimized

parameters A and B using the

chromatograms of the broad standards

and applying the known molar mass

averages.

t� Save this calibration curve so that it can

be used to determine the complete molar

mass distribution and the true molar

mass averages of samples with unknown

molar masses.

Figure 2 shows the chromatograms of

the broadly distributed standards as well as

what is required to use the broad standards

calibration concept. To increase the accuracy

of the calibration, more than one broadly

distributed standard should be used. If a

wide molar mass range needs to be covered

the use of several samples is recommended.

Figure 3 shows a comparison of the base

calibration (blue) and the resulting calibration

curve after determination of A and B (red).

This calibration curve is then used for the

evaluation of unknown samples and it provides

true molar masses for all samples of this type.

How to Get Broad Standards

A limited number of broad standards for

a variety of chemical structures is available

from suppliers. If no broad standards are

commercially available, sample testing

laboratories or the R&D department can

determine true molar mass using conventional

batch measurements for example, by light

scattering, nuclear magnetic resonance

(NMR), or osmometry. On-line determination

using multidetection GPC/SEC can also solve

this problem.

Summary

t�Calibration using broad standards is a fast

and easy-to-use method for quality control

laboratories.

t�Standards with a broad molar mass

distribution can be easily obtained from

contract analysis laboratories or the R&D

department using batch methods. This

allows the measurement of precise and

accurate molar masses worldwide.

E-mail: [email protected]: www.pss-polymer.com

References

1. Z. Grubisic, P. Rempp, and H. Benoît, Journal

of Polymer Science, Part B: Polymer Letters

5(9), 753–759 (1967).

2. S. Mori, Anal. Chem. 53, 1813–1818 (1981).

3. H.K. Mahabadi and K.F. O´Driscoll, J. Appl.

Polym. Sci. 21, 1283–1287 (1977).

4. A.R. Weiss and E. Cohn-Ginsberg, Journal of

Polymer Science Part B: Polymer Letters 7(5),

379–381 (1969).

5. W. Radke, in Macromolecular Engineering,

K.G. Matyjaszewski and L. Leibler, Eds.

(Wiley-VCH, Weinheim, Germany vol. 3), pp.

1881–1936.

6. D. Held, The Column 4(6), 18–21 (2008).

Daniela Held studied polymer chemistry

in Mainz, Germany, and works in the

PSS solutions department. She is also

responsible for PSS webinars, education

programmes, and customer training.

Wolfgang Radke studied polymer

chemistry in Mainz, Germany, and in

Amherst, Massachusetts, USA. He is now

head of the PSS application development

department. He is also responsible for the

PSS customer training programme and for

customized trainings.

Tips & Tricks: GPC/SEC

15

Pittcon 2015 Preview2 News7 Tips & Tricks: GPC/SEC10 Q&A: Haselberg167 100The LCGC Blog20 CHROMacademy23 Training & Events24 Staff252323 242

Make Your Lab Productivity Soar With the Right

Chromatography Data SystemON-DEMAND WEBCAST(originally aired January 27, 2015)

Register for free atwww.chromatographyonline.com/soar

EVENT OVERVIEW:

Today’s chromatography laboratory is faced with a growing

number of business challenges: reducing time-to-results, meeting

increasing requirements for regulatory compliance, maximizing

productivity through optimum instrument asset utilization, and

reducing operational

costs. Having the right

chromatography data

system (CDS) software

is critical to addressing

these challenges. Of

course, replacing an

existing CDS is a big

decision and requires lab

decision-makers to take

a number of key factors

into account.What are the factors one needs to consider when

selecting a CDS? In this webinar, we will share with you what is

involved in making the change and how you can prepare for it.

Additionally, you’ll learn how you can increase productivity in

your lab by using the significantly improved data analysis and

reporting features that the right CDS has to offer. Key questions

that indicate that you should tune into this web seminar include:

■ Is your software running on

an out-of-support operating

system?

■ Do you use programs like

Microsoft Excel for calculations

and reporting?

■ Are you unable to take

advantage of the latest

instrumentation due to lack of

software support?

Who Should Attend:

■ Decision Makers:

Analytical Services, Method

Development, Production,

QA/QC, R&D, Lab Manager,

Manager/Dept. Head,

Scientist/ Chemist, Technical

Decision Maker

■ This information is valuable for all companies planning to

upgrade the software in their labs

Presenters

Thomas SchmidtOpenLAB CDS Product Manager

Michael SpeziaSenior Professional Services Engineer

Lisa Sprague,Software and Informatics Product ManagerTechnology Transitions

Moderator

Laura BushEditorial Director, LCGC

Key Learning Objectives:

■ Learn of the many advantages a

move to the right CDS can offer

■ How you can reduce complexity

and improve efficiency in your lab

■ How to boost lab productivity by

simplifying operations

Sponsored by Presented by

For questions, contact Kristen Moore at [email protected]